K Number
K253754

Validate with FDA (Live)

Date Cleared
2025-12-23

(28 days)

Product Code
Regulation Number
878.4810
Age Range
18 - 65
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Philips Lumea IPL is indicated for the removal of unwanted hair for a single user. It is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

Device Description

Philips Lumea IPL is an Intense Pulsed Light (IPL) device intended for removal of unwanted hair. The device is to be used by persons in the age range of 18-65 years with fair up to medium brown skin tones, and naturally dark blond, brown, dark brown and black hair.

The Philips Lumea IPL applies spectrally adjusted light to the user skin, which is absorbed by the melanin and after some applications hair sheds naturally and growth is delayed or inhibited.

The Philips Lumea IPL is an Over-the-counter hand-held device powered by a wall mount adapter. It is provided in a non-sterile state and intended to be used by single person for multiple uses in homecare environment.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - Philips Lumea IPL

Page 1

December 23, 2025

Philips Consumer Lifestyle B.V.
℅ Shaylee Masilunas
Sr. Regulatory Affairs Project Manager
Philips Personal Health, Philips North America, LLC
1600 Summer St.
5th Floor
Stamford, Connecticut 06905

Re: K253754
Trade/Device Name: Philips Lumea IPL
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHT, GEX
Dated: November 25, 2025
Received: November 25, 2025

Dear Shaylee Masilunas:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

Page 2

K253754 - Shaylee Masilunas Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

Page 3

K253754 - Shaylee Masilunas Page 3

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MARK MACIOS -S
Digitally signed by MARK MACIOS -S
Date: 2025.12.23 12:36:05 -0500

for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

FieldValue
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.K253754
Please provide the device trade name(s).Philips Lumea IPL

Please provide your Indications for Use below.

Philips Lumea IPL is indicated for the removal of unwanted hair for a single user. It is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

Please select the types of uses (select one or both, as applicable).

  • ☐ Prescription Use (21 CFR 801 Subpart D)
  • ☑ Over-The-Counter Use (21 CFR 801 Subpart C)

Page 5

510(k) Summary - K253754

(As required by 21CFR807.92)

I. SUBMITTER

FieldValue
Submission typeSpecial 510(k)
Company Name and AddressPhilips Consumer Lifestyle B.V.Tussendiepen 49206 AD DrachtenThe Netherlands
Primary Correspondent/Consultant InformationMrs. Shaylee MasilunasSr. Regulatory Affairs Project ManagerEmail: shaylee.masilunas@philips.com
Contact PersonMr. Martijn HalbesmaGroup Lead G&B Regulatory & ComplianceEmail: pclcerts@philips.com
Date Prepare24 November 2025

II. SUBJECT DEVICE

FieldValue
Trade NamePhilips Lumea IPL
Classification Regulation21CFR§878.4810
Regulation DescriptionLaser surgical instrument for use in general and plastic surgery and in dermatology
Classification Product CodeOHT: Light Based Over-The-Counter Hair Removal,GEX: Laser Surgical Instrument for Use in General and Plastic Surgery and in Dermatology (subsequent product code)
Device Classification2
Regulation medical specialtyGeneral & Plastic Surgery
Premarket reviewOffice of Surgical and Infection Control Devices

III. PREDICATE DEVICE

FieldValue
Trade NamePhilips Lumea IPL (K243453)
Classification Regulation21CFR§878.4810
Regulation DescriptionLaser surgical instrument for use in general and plastic surgery and in dermatology
Classification Product CodeOHT: Light Based Over-The-Counter Hair Removal,GEX: Laser Surgical Instrument for Use in General and Plastic Surgery and in Dermatology (subsequent product code)
Device Classification2
Regulation medical specialtyGeneral & Plastic Surgery
Premarket reviewOffice of Surgical and Infection Control Devices

IV. DEVICE DESCRIPTION

Philips Lumea IPL is an Intense Pulsed Light (IPL) device intended for removal of unwanted hair. The device is to be used by persons in the age range of 18-65 years with fair up to medium brown skin tones, and naturally dark blond, brown, dark brown and black hair.

The Philips Lumea IPL applies spectrally adjusted light to the user skin, which is absorbed by the melanin and after some applications hair sheds naturally and growth is delayed or inhibited.

The Philips Lumea IPL is an Over-the-counter hand-held device powered by a wall mount adapter. It is provided in a non-sterile state and intended to be used by single person for multiple uses in homecare environment.

Page 6

V. INDICATIONS FOR USE

Philips Lumea IPL is indicated for the removal of unwanted hair for a single user. It is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES

The table below provides a side-by-side comparison of the subject device and predicate device. The difference between the subject device and the predicate device are outlined in Table 1.

Table 1: Feature / Characteristics comparison of Subject device and Predicate device

Feature / Characteristic for comparisonSubject Device - Philips Lumea IPL (9900 Pro)Predicate Device – Philips Lumea IPL (K243453)Similarity between Subject and Predicate Device
Classification
Legal ManufacturerPhilips Consumer Lifestyle B.V.Philips Consumer Lifestyle B.V.Identical
510(k) NumberIn this applicationK243453Not applicable
Device Regulation21CFR878.481021CFR878.4810Identical
Classification Product CodeOHT, GEX (Subsequent Product Code)OHT, GEX (Subsequent Product Code)Identical
Device Classification NameOHT: Light Based Over-The-Counter Hair RemovalOHT: Light Based Over-The-Counter Hair RemovalIdentical
GEX: Laser Surgical Instrument for Use in General and Plastic Surgery and in DermatologyGEX: Laser Surgical Instrument for Use in General and Plastic Surgery and in DermatologyIdentical
Device Class22Identical
Classification PanelGeneral & Plastic SurgeryGeneral & Plastic SurgeryIdentical
Indications for Use
Indications for UsePhilips Lumea IPL is indicated for the removal of unwanted hair for a single user. It is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.Philips Lumea IPL is indicated for the removal of unwanted hair for a single user. It is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.Identical
Operating Environment Characteristics
AvailabilityOver-The-Counter Use (21 CFR 801 Subpart C)Over-The-Counter Use (21 CFR 801 Subpart C)Identical
Intended User Group18 to 65 Years18 to 65 YearsIdentical
Treatment AreasSkin tone: The device is intended for use on fair up to medium brown skin tones.Skin tone: The device is intended for use on fair up to medium brown skin tones.Identical
Hair color: Naturally dark blond, brown, dark brown and black hairHair color: Naturally dark blond, brown, dark brown and black hairIdentical
ReusabilitySingle Patient Multiple UseSingle Patient Multiple UseIdentical
Environment of UseHomeHomeIdentical

Page 7

Feature / Characteristic for comparisonSubject Device - Philips Lumea IPL (9900 Pro)Predicate Device – Philips Lumea IPL (K243453)Similarity between Subject and Predicate Device
Operating EnvironmentTemperature range: 5°C to 40°C (41°F to 104°F)Temperature range: 5°C to 40°C (41°F to 104°F)Identical
Relative humidity range: 15% to 90% RH (noncondensing)Relative humidity range: 15% to 90% RH (noncondensing)Identical
Atmospheric pressure range: 700hPa to 1060hPa.Atmospheric pressure range: 700hPa to 1060hPa.Identical
Storage and Transport EnvironmentTemperature range: -25°C to 70°C (-13°F to 158°F)Temperature range: -25°C to 70°C (-13°F to 158°F)Identical
Relative humidity: upto 95% RH (noncondensing)Relative humidity: upto 95% RH (noncondensing)Identical
Atmospheric pressure range: 700hPa to 1060hPaAtmospheric pressure range: 700hPa to 1060hPaIdentical
Technology & Safety Characteristics
Operating PrincipleIntense pulsed light hair removalIntense pulsed light hair removalIdentical
Light sourceXenon Arc FlashlampXenon Arc FlashlampIdentical
Power SourceExternal power supplyExternal power supplyIdentical
User InterfaceFlash ButtonLED indicatorsFlash ButtonLED indicatorsIdentical
Safety FeaturesSkin detection SensorSkin detection SensorIdentical
Skin tone detection SensorSkin tone detection SensorIdentical
Wireless ConnectivityBluetooth Low Energy (BLE) (Wireless connection with Mobile App – Lumea)No connectivity available (No connection with Mobile App – Lumea)Equivalent
Real-time motion guidanceReal-time motion guidance using Inertial Measurement Unit (IMU) SensorNo motion sensing availableEquivalent
Over-the-air software updatesYesNoEquivalent
Software / Firmware / Microprocessor controlYesYesIdentical
Performance Specifications
# of AttachmentsBody AttachmentFace AttachmentPrecision AttachmentBody AttachmentFace AttachmentPrecision AttachmentIdentical
Spot Size (cm²)Body Attachment: 4.1 cm²Precision Attachment: 3 cm²Face Attachment: 2 cm²Body Attachment: 4.1 cm²Precision Attachment: 3 cm²Face Attachment: 2 cm²Identical
Emitted Light Spectrum (nm)520nm - 1200nm (max.)520nm - 1200nm (max.)Identical
Energy Density (J/cm²)6 J/cm²(Max)6 J/cm²(Max)Identical
Pulse duration1.25±0.4 ms (FWHM)1.25±0.4 ms (FWHM)Identical
Optical Pulse Interval0.6 - 2.4 s1.0 - 2.4 sEquivalent
Electrical System Characteristics
Supply Frequency50/60 Hz.50/60 Hz.Identical
Supply Voltage100-240 VAC100-240 VACIdentical
International Standards
IEC 60601-1YesYesIdentical
IEC 60601-1-2YesYesIdentical
IEC 60601-1-6YesYesIdentical
IEC 60601-1-11YesYesIdentical
IEC 62366-1YesYesIdentical
IEC 62304YesYesIdentical
IEC 60601-2-57YesYesIdentical

Page 8

Feature / Characteristic for comparisonSubject Device - Philips Lumea IPL (9900 Pro)Predicate Device – Philips Lumea IPL (K243453)Similarity between Subject and Predicate Device
IEC 60601-2-83YesYesIdentical
ISO 10993-1YesYesIdentical
ISO 10993-5YesYesIdentical
ISO 10993-10YesYesIdentical

Page 9

VII. PERFORMANCE DATA

Summary of non-clinical tests

The Philips Lumea IPL has been tested to the applicable requirements of the following standards:

Standards and Designation NumberStandards Title
ANSI/AAMI ES60601-1:2005/A2:2021Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance - Amendment 2
ANSI/AAMI IEC 60601-1-2:2014 + A1:2021Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Disturbances – Requirements and Tests
IEC 60601-1-6:2010 + A1:2013 + A2:2020Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
ANSI AAMI IEC 60601-1-11:2015 + A1:2021Medical electrical equipment and medical electrical systems used in the home healthcare environment
IEC 60601-2-83:2019 + A1:2022Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
IEC 60601-2-57:2011Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
ISTA 3A: 2018Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb) or Less
ANSI AAMI IEC 62304 2006 + A1:2016Medical device software — Software life cycle processes
IEC 82304-1 ed.1 2016-10Health Sofware – Part 1: General requirements for product safety
IEC 81001-5-1 ed.1.0 2021-12Health software and health IT systems safety, effectiveness and security - Part 5-1: Security - Activities in the product life cycle
ANSI AAMI IEC 62366-1:2015 + A1:2020Medical devices — Part 1: Application of usability engineering to medical devices
ANSI AAMI ISO 10993-1:2018Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process
ANSI AAMI ISO 10993-5:2009/(R)2014Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2021-11Biological evaluation of medical devices — Part 10: Tests for irritation and skin sensitization

VIII. CONCLUSIONS

Philips concludes that the supporting information and justifications provided herein conclusively demonstrate that the subject device, Philips Lumea IPL (9900 Pro) is substantially equivalent to the predicate device, Philips Lumea IPL (K243453).

Software verification and validation testing were conducted and basic level of documentation was provided as recommended by FDA'S Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions (2023)."

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.