Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K222948

    Validate with FDA (Live)

    Date Cleared
    2023-09-19

    (357 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Q Distal Access Catheter is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the peripheral, coronary and neuro vascular systems.

    Device Description

    The Q Distal Access Catheter is a single-lumen, variable stiffness catheter with radiopaque markers on the distal and proximal end of the catheter portion for angiographic visualization. The proximal portion of the device is a stainless-steel control wire. The distal portion of the device is a coiled/braided catheter shaft with hydrophilic coating along the entire length to reduce friction during use. The Q Distal Access Catheter may be introduced via an 8F guide catheter/6F guide sheath and over a guidewire/microcatheter into the arterial vasculature until the desired vessel is reached. The pin vise may be used to advance the catheter. The syringe and flush tool components may aid in the flushing of the Q Distal Access Catheter.

    AI/ML Overview

    The medical device described, the Q Distal Access Catheter, is not an AI/ML powered device, therefore, the requested information regarding AI/ML related aspects (such as effect size, standalone performance, training set details, and expert qualifications for ground truth) is not applicable. The provided text details the non-clinical performance testing for a physical medical catheter.

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Q Distal Access Catheter are derived from various non-clinical performance tests, including bench testing and biocompatibility testing. The "Results" column from the provided tables serves as the reported device performance, indicating that the device "Pass"ed each criterion.

    Test CategorySpecific TestAcceptance Criterion (Test Method Summary)Reported Device Performance (Result)
    Bench TestingLabel IntegrityDevice labels were evaluated to ensure labels are intact and legible after environmental conditioning, distribution simulation and aging.Pass
    Bubble LeakDevice packaging was evaluated to detect gross leaks in packaging.Pass
    Pouch Seal StrengthPouch was evaluated for the mechanical strength of the seal.Pass
    Coating ParticulateThe number and size of particles generated during simulated use were measured.Pass
    Dimensional VerificationDimensional specifications were verified to ensure device meets all the requirements.Pass
    Tip InspectionCatheter tip was evaluated to ensure the complete retrieval of tip during its removal.Pass
    Surface IntegrityDevice visually inspected to ensure surface is free of defects that can cause tissue trauma.Pass
    Heat Shrink InspectionTo ensure heat shrink identifier remains intact and legible after environmental conditioning, distribution simulation and aging.Pass
    Coating UniformityTo evaluate the devices for uniform coating.Pass
    Simulate Use/CompatibilityThe catheter performance was evaluated when used in conjunction with other devices used in standard procedures without sustaining damage or kinks and without causing damage to the other devices.Pass
    Coating AdhesionTo visually categorize the integrity of coating on the device.Pass
    Push/TrackTo ensure device is able to be tracked without kink under normal conditions in tortuous anatomy.Pass
    Kink ResistanceThe distal shaft of the device was evaluated for kink resistance when subjected to a bend radius of 1.0 cm.Pass
    Liquid Leak Under PressureTo ensure device does not exhibit leakage of liquid under a minimum pressure of 44 psi for a minimum of 30 seconds.Pass
    Static BurstTo ensure device withstands a minimum pressure of 100 psi prior to burst.Pass
    Tensile Strength - Push WireTo ensure device meets the required tensile strength between the push wire and catheter body: 15 N pull force minimum.Pass
    Tensile Strength - Distal SectionTo ensure device meets the required tensile strength of the distal section of the shaft: 5N pull force minimum.Pass
    BiocompatibilityCytotoxicityBiological reactivity Grade 0 on the test article after 48 hrs. incubation with mouse fibroblast. The response on the positive and negative control article extracts confirmed the suitability of the tests. The test article (TA) is considered non-cytotoxic.Pass
    Hemolysis (Direct and Indirect)The test article in direct contact with blood had a hemolytic index of 0.3%, and the test article extract had a hemolytic index of 0.2%. Both the test article in direct contact with blood and the test article extract were non-hemolytic.Pass
    SterilizationEthylene Oxide (EO) ResidualsIn accordance with ISO 10993-7.Pass (Implied by Conclusion)
    Limulus Amoebocyte Lysate (LAL)In accordance with USP 24 chapter 85 and FDA guidance.Pass (Implied by Conclusion)
    Bioburden TestingIn accordance with relevant standards (not explicitly listed but implied alongside EO residuals and LAL).Pass (Implied by Conclusion)
    Shelf-lifeShelf-life Verification (3 years)The device will be labeled with an expiration date of 3 years from the date of sterilization.Pass (Implied by Conclusion)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for each specific bench or biocompatibility test. The evaluation involved "design verification testing" and "additional biocompatibility testing." The provenance of the data is from internal testing conducted by MIVI Neuroscience, Inc., as part of their 510(k) submission. This is considered retrospective for the purpose of the submission, as the tests were performed to demonstrate compliance of an already designed and manufactured device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This question is not applicable as the Q Distal Access Catheter is a physical medical device, not an AI/ML powered device that relies on expert interpretation for "ground truth" establishment in its performance evaluation. The "ground truth" in this context refers to established engineering specifications, material science standards, and biocompatibility guidelines.

    4. Adjudication Method for the Test Set

    This question is not applicable for a physical medical device. The "adjudication method" concept is typically used for subjective assessments in clinical or AI/ML studies where human interpretation of ambiguous data is involved. For engineering and scientific testing of a physical device, results are typically objective and determined by established protocols and measurement techniques.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This question is not applicable. The Q Distal Access Catheter is a physical medical device and does not involve AI assistance for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. The Q Distal Access Catheter is a physical medical device and does not have a standalone algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the Q Distal Access Catheter's performance evaluation is based on a combination of:

    • Design Specifications: Predefined engineering requirements and dimensional tolerances for the catheter.
    • Established Industry Standards: Such as ISO 11135:2014 for sterilization, ISO 10993-7 for biocompatibility residuals, ASTM F756 for hemolysis, and USP 24 chapter 85 for bacterial endotoxins.
    • Visual and Mechanical Inspection Criteria: Standardized methods to assess physical integrity, coating uniformity, and functional performance.
    • Physiological Response Criteria: For biocompatibility, established thresholds for cytotoxicity and hemolysis.

    8. The Sample Size for the Training Set

    This question is not applicable. The Q Distal Access Catheter is a physical medical device and does not involve an AI/ML training set.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable. The Q Distal Access Catheter is a physical medical device and does not involve an AI/ML training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212402

    Validate with FDA (Live)

    Date Cleared
    2021-12-16

    (136 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MIVI Q Distal Access Catheter is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the peripheral, coronary and neuro vascular systems.

    Device Description

    The MIVI Q Distal Access Catheter (Q Catheter) is a single-lumen, variable stiffness catheter with radiopaque markers on the distal and proximal end of the catheter portion for angiographic visualization. The catheter shaft has a hydrophilic coating to reduce friction during use. The proximal portion of the catheter is a stainless-steel control (push) wire. The Q Catheter may be introduced via an 8F guide catheter/6F guide sheath and over a guidewire/microcatheter into the arterial vasculature until the desired vessel is reached. The Q Docking Station may be used to facilitate insertion and extraction of the Q Catheter through a hemostasis valve attached to the 8F guide catheter/6F guide sheath. The pin vise may be used to advance the catheter.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the MIVI Q Distal Access Catheter:

    The document provided is a 510(k) Summary for the MIVI Q Distal Access Catheter, which is a premarket notification to the FDA for a medical device seeking substantial equivalence to a predicate device. This type of submission focuses on demonstrating that the new device is as safe and effective as a legally marketed predicate, rather than proving novel effectiveness.

    Given the nature of a 510(k) submission for a catheter, the "acceptance criteria" and "device performance" in this context refer to the ability of the device to meet specified engineering and biocompatibility standards, and to perform its intended function without significant differences from the predicate. There is no clinical study in the traditional sense (e.g., patient trials with efficacy endpoints) described for this device, as it's a substantially equivalent determination based on non-clinical testing.

    Here's the information extracted and organized according to your request:


    Acceptance Criteria and Device Performance for MIVI Q Distal Access Catheter

    1. Table of Acceptance Criteria and Reported Device Performance

    For Bench Testing (Non-Clinical Performance):

    TestAcceptance CriteriaReported Device Performance
    Dimensional VerificationThe device must meet dimensional specifications.Pass
    System Introduction (Simulated Use)The device must function as intended.Pass
    Liquid Leakage under PressureThe device must hold a hydrostatic pressure.Pass
    Tensile StrengthTensile strength pull force minimum must be met.Pass

    For Biocompatibility Testing:

    TestAcceptance Criteria (Implied)Reported Device Performance
    CytotoxicityNon-toxic to cells (determined by specific extract and cell line).PASS
    Irritation or Intracutaneous ReactivityNo significant irritation or reactivity in vivo (animal model).PASS
    SensitizationNo significant sensitization reaction (animal model).PASS
    Acute Systemic ToxicityNo acute systemic toxicity (animal model).PASS
    Pyrogenicity (Material-mediated)Non-pyrogenic (no fever-inducing substances in vivo).PASS
    HemolysisNo significant hemolysis (destruction of red blood cells) in vitro.PASS
    Unactivated Partial Thromboplastin TimeNo significant impact on blood clotting time (in vitro).PASS
    Complement ActivationNo significant complement activation (immune response) in vitro.PASS

    For Sterilization and Shelf Life:

    TestAcceptance Criteria (Implied)Reported Device Performance
    SterilizationConfirmed sterility of the final packaging configuration and method.Met specification
    Shelf Life and Expiration DatingMaintain performance and integrity for the specified period.3 years (same as predicate)

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify the exact sample sizes used for each individual non-clinical (bench and biocompatibility) test. Such details are typically found in the full test reports, not in the summary document.

    Data Provenance: The tests are non-clinical, meaning they were conducted in a laboratory setting (in vitro or in animal models), not on human subjects. Therefore, there is no country of origin for human patient data. The provenance of the data is the laboratory where the tests were performed. The studies are prospective in the sense that they were designed and executed to test the characteristics of the MIVI Q Distal Access Catheter.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This question is not applicable to the provided document. The "ground truth" in this context is established by objective measurements and standardized test methods (e.g., ISO standards for biocompatibility, engineering specifications for bench tests), not by expert consensus on clinical data. No human "experts" were used to establish ground truth for this type of non-clinical testing.


    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods (like 2+1, 3+1) are used in clinical studies, particularly for interpreting imaging or clinical outcomes where there might be inter-reader variability. For non-clinical bench and biocompatibility tests, results are typically objective (e.g., a tensile strength value, a cytotoxicity rating) and do not require expert adjudication in this manner.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The MIVI Q Distal Access Catheter is a physical medical device (a catheter), not an AI algorithm, and therefore does not involve human readers interpreting output from an AI system. The submission is for substantial equivalence of a physical device.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. As explained above, this device is a physical catheter, not an AI algorithm.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical tests described:

    • Bench Testing: Ground truth is defined by engineering specifications and objective physical measurements (e.g., specified dimensions, force thresholds, pressure integrity).
    • Biocompatibility Testing: Ground truth is established by standardized biological assays and observation in animal models and in vitro systems, as per recognized international standards (e.g., ISO 10993 series). The results are objective measurements against defined pass/fail criteria.

    8. The sample size for the training set

    This is not applicable. There is no AI algorithm involved that requires a training set. The "study" here consists of non-clinical performance and biocompatibility testing of the physical catheter.


    9. How the ground truth for the training set was established

    This is not applicable, as there is no AI algorithm or training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163233

    Validate with FDA (Live)

    Date Cleared
    2017-04-06

    (140 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MIVI Mi-EXT Catheter is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the peripheral, coronary and neuro vascular systems.

    Device Description

    Not Found

    AI/ML Overview

    There is no information in the provided text about acceptance criteria or a study that proves a device meets those criteria. The document is a 510(k) clearance letter for the MIVI Mi-EXT Catheter, indicating it has been found substantially equivalent to a predicate device. It specifies the intended use of the catheter but does not detail performance metrics, study design, or any of the other requested information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151396

    Validate with FDA (Live)

    Date Cleared
    2015-12-09

    (197 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MIVI Mi-Axus™ Guide Catheter is indicated for use in facilitating the insection and guidance of microcatheters into a selected blood vessel in the peripheral, coronary and neuro vascular systems.

    Device Description

    The MIVI Mi-Axus 8F Guide Catheter consists of a single lumen, braided, variable stiffness shaft designed for use in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the peripheral, coronary, or neurovascular system. A radiopaque marker is included on the distal end for angiographic visualization. The catheter shaft has a hydrophilic coating to reduce friction during use. Additionally, the distal segment of the PEBAX shaft of the 8F device is impregnated with Barium Sulfate (BaS04) for increased radiopacity. A luer hub on the proximal end allows attachments for flushing, insertion of catheters, and aspiration. It is used in junction with a rotating hemostatic valve with side-arm adapter for flushing, catheter insertion and aspiration. The MIVI Mi-Axus Guide Catheter has a straight distal tip and is available in lengths of 80 or 90 cm and an 8F diameter.

    AI/ML Overview

    The provided text describes a 510(k) summary for the MIVI Mi-Axus™ Guide Catheter. It does not contain information about a study with acceptance criteria and device performance in the context of an AI/ML medical device. Instead, it details the functional, biocompatibility, and sterilization testing performed for a guide catheter.

    Therefore, I cannot provide the requested information. The document focuses on establishing substantial equivalence for a medical device (a guide catheter) based on its physical properties, materials, intended use, and comparison to predicate devices, rather than on the performance metrics of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151825

    Validate with FDA (Live)

    Date Cleared
    2015-11-16

    (133 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Viradius™ Neurowire is intended for selective placement of catheters within the neuro and peripheral vasculature.

    Device Description

    The Viradius™ Neurowire consists of a stainless steel wire core and a radiopaque platinum/tungsten coil on the distal tip. The distal section of the guidewire is coated with a hydrophilic coating and the proximal portion of the device is coated with a hydrophobic coating. The guidewire is inserted into a polyethylene hoop and then placed along with a torque device into a labeled poly/tyvek pouch and ethylene oxide sterilized. The pouch, along with an Instruction for Use paper insert is placed into a labeled shipping box.

    AI/ML Overview

    The Viradius™ Neurowire is a guidewire intended for selective placement of catheters within the neuro and peripheral vasculature. The information provided outlines the acceptance criteria and the studies conducted to demonstrate its substantial equivalence to a legally marketed predicate device (NeuroScout Steerable Guidewire, K100351).

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Viradius™ Neurowire are generally that the device "meets acceptance criteria" or performs comparably to the predicate device or selected competitor guidewires, or that the results are adequate/acceptable. The reported device performance indicates that all devices met these criteria for all tested parameters.

    Test / Performance CharacteristicAcceptance CriteriaReported Device Performance
    Bond and Joint Tensile StrengthThe guidewire, including all joints, must not suffer damage when subjected to tensile force.All devices met acceptance criteria.
    Ultimate Torque ResponseThe guidewire must withstand a minimum complete rotations without physical damage.All devices met acceptance criteria.
    Torque ResponseDistal tip rotation shall be equivalent to predicate guidewire.All devices met acceptance criteria.
    Tip DeflectionTip deflection force shall be comparable to selected competitor guidewires.All devices met acceptance criteria.
    Push Track – Anatomical ModelForce required to advance guidewire through simulated tortuous neurovascular arteries shall be comparable to predicate device.All devices met acceptance criteria.
    Surface IntegrityGuidewire surface shall not have any particulate visible to the naked eye under close inspection without magnification.All devices met acceptance criteria.
    Guidewire DiameterThe diameter was measured and met specifications (implicitly, as it passed; also noted 0.012" distal, 0.014" proximal).All devices met acceptance criteria.
    Guidewire LengthThe length was measured and met specifications (implicitly, as it passed; also noted 205 cm / 300 cm).All devices met acceptance criteria.
    Catheter CompatibilityGuidewire shall be compatible with microcatheters, distal access catheters, guide catheters, and fluids.All devices met acceptance criteria.
    Kink ResistanceProximal shaft shall not kink when subjected to a bend radius.All devices met acceptance criteria.
    Flex TestIn accordance with ISO 11070:2014 8.5 the guidewire shall not fracture, loosen, or fail in such a manner that any section of the coil is left free to stretch, a sharp, or potentially traumatic fracture surface is exposed, any part of the device becomes separated such that it would not be removable by withdrawing the device from use, or coated guidewires show flaking of the coating.All devices met acceptance criteria.
    Distal Fracture TestIn accordance with ISO 11070:2014 8.4 the guidewire shall not fracture, loosen, or fail in such a manner that any section of the coil is left free to stretch, a sharp, or potentially traumatic fracture surface is exposed, or any part of the device becomes separated such that it would not be removable by withdrawing the device from use.All devices met acceptance criteria.
    Tip Bending and RetentionAfter insertion through a competitive microcatheter, the bend radius shall not degrade.All devices met acceptance criteria.
    Device Coating AdhesionAfter simulated use, the guidewire coating shall not exhibit any scratches, flaking, or generate any loose coating material.All devices met acceptance criteria.
    Particulate TestingThe guidewire was evaluated for particulate generation under simulated use in a representative tortuous model, with an implicit acceptance criterion of an "adequate" number and size of particulates.Number and size of particulates generated was adequate.
    Corrosion TestThe guidewire shall not exhibit visual corrosion when immersed in sodium chloride solution.All devices met acceptance criteria.
    Biocompatibility (Cytotoxicity)PassPass
    Biocompatibility (Sensitization)PassPass
    Biocompatibility (Intracutaneous)PassPass
    Biocompatibility (Systemic Tox)PassPass
    Biocompatibility (Pyrogen)PassPass
    Biocompatibility (Hemocompat.)PassPass
    Sterilization (SAL)SAL of 10^-6 in accordance with ISO 11135-1:2007.Validated.
    Shelf Life (1 year)Product and packaging remain functional and maintain sterility for up to one year; packaging integrity, seal strength, and device functionality met all acceptance criteria.Met all acceptance criteria through aging studies.

    2. Sample Size and Data Provenance (Test Set)

    The document does not explicitly state the sample sizes (number of units or iterations) used for each specific bench test mentioned. However, for all performance tests, it states that "All devices met acceptance criteria." This implies that a sufficient number of devices were tested to draw this conclusion based on the test methods used.

    The data provenance for the test set is retrospective/bench (laboratory) data, generated from direct testing of the Viradius™ Neurowire and comparison to the predicate device or competitor guidewires. There is no indication of country of origin as it applies to clinical data, as no clinical studies were performed.

    3. Number of Experts and Qualifications (Ground Truth for Test Set)

    This section is not applicable. The studies described are primarily bench (laboratory) performance tests and biocompatibility testing, not studies that involve human interpretation of medical images or diagnoses requiring expert ground truth in the traditional sense. The "ground truth" for these tests is the objective physical and chemical properties and performance of the device relative to established standards or a predicate device.

    4. Adjudication Method (Test Set)

    This section is not applicable. As the studies were bench tests demonstrating physical performance and material safety, there was no adjudication method involving multiple human reviewers to establish a consensus or resolve discrepancies, which is typical for clinical or imaging studies.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was performed. The submission explicitly states: "No clinical studies were performed as there are no change to the indications for use or the fundamental scientific technology of the device." Therefore, there is no information on the effect size of human readers improving with or without AI assistance. This device is a medical instrument, not an AI or imaging diagnostic tool.

    6. Standalone (Algorithm Only) Performance

    No standalone performance study was done. This device is a medical guidewire, a physical instrument, not an algorithm or software. Therefore, the concept of "standalone performance" for an algorithm without human-in-the-loop is not relevant to this submission.

    7. Type of Ground Truth Used

    The "ground truth" for the performance testing was based on:

    • Engineering/Performance Standards: Adherence to established ISO standards (e.g., ISO 11070:2014, ISO 11135-1:2007) and internal company specifications.
    • Comparison to Predicate Device: Performance (e.g., torque response, push track force) comparable to the NeuroScout Steerable Guidewire (K100351).
    • Objective Laboratory Measurements: Direct measurements of physical dimensions (diameter, length), visual inspection (surface integrity, corrosion), and chemical tests (biocompatibility).
    • Simulated Use Models: Performance in simulated anatomical models (e.g., push track test, particulate testing in tortuous model).

    8. Sample Size for the Training Set

    No training set was used. This device is a physical medical instrument, not a machine learning model or AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1