    K Number
    K100313

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-06-14

    (495 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MiMedx HydroFix™ Surgical Sheet is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. The surgical sheet minimizes tissue attachment to the device in case of direct contact with the tissues.

    Device Description

    The HydroFix™ Surgical Sheet is a flexible sheet of 30 Wt.% polyvinyl alcohol (PVA) material with dimensions 60 ±6 mm X 50 ±5 mm and 60 ±6 mm X 100 ±10 mm with a thickness of 1.0 ±0.2 mm. The comers of the sheet are rounded. There are no holes or perforations. There are no markings on either side of the sheet, raised (embossed) or printed. The sheet is provided sterile and hydrated in saline solution.

    The HydroFix™ Surgical Sheet will be provided in other shapes and sizes as needed for particular surgical procedures.

    AI/ML Overview

    This document describes the 510(k) submission for the HydroFix™ Surgical Sheet, which aims to demonstrate its substantial equivalence to previously cleared devices. It primarily focuses on comparing the new device's performance and characteristics to those of predicate devices rather than establishing specific acceptance criteria for standalone performance.

    Here's an analysis of the provided information based on your requested points:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state numerical acceptance criteria with pass/fail thresholds for the HydroFix™ Surgical Sheet. Instead, it states that the device's performance was judged as "comparable to the predicate devices."

    Performance CriterionReported Device Performance (vs. Predicate)
    In Vitro Testing
    Suture Pull outComparable
    Tensile StrengthComparable
    Burst StrengthComparable
    Tear resistanceComparable
    In Vivo Testing
    Tissue attachmentComparable
    Ability to suture to tissueComparable
    Ability to cut the sheetComparable
    Ability to secure to tissueComparable
    Ability to manage and protect tendon injuriesComparable

    2. Sample size used for the test set and the data provenance

    • Sample size for test set: Not specified for in vitro tests. For in vivo tests, it mentions studies conducted in "both sheep and rabbit models," implying a sample size of at least two different animal types, but the number of individual animals within each model is not provided.
    • Data provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, in vivo studies typically imply prospective animal studies conducted in a controlled environment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided. The document refers to "in vitro" and "in vivo" testing, implying laboratory and animal model evaluations, not expert-based ground truth for interpretation as might be found in diagnostic imaging studies.

    4. Adjudication method for the test set

    This information is not provided, as the evaluation type doesn't involve human interpretation that would require an adjudication method.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This device is a surgical sheet, not an AI-assisted diagnostic tool, so such a study would not be applicable.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This concept is not applicable. The device is a surgical sheet, not an algorithm. The testing performed was to evaluate the physical and biological characteristics of the sheet.

    7. The type of ground truth used

    • For In Vitro Testing: The ground truth is inherent in the physical properties and mechanical measurements themselves (e.g., actual tensile strength, burst strength, etc., measured according to established material testing standards).
    • For In Vivo Testing: The ground truth is the observed biological response and mechanical performance within the animal models as assessed by researchers (e.g., visual assessment of tissue attachment, successful suturing, etc.). This would likely involve histological examination and gross observations post-sacrifice.

    8. The sample size for the training set

    This concept is not applicable. The device is a physical product (surgical sheet), not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This concept is not applicable, as there is no training set for a physical surgical sheet.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101805

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-04-08

    (284 days)

    Product Code
    Regulation Number
    870.3470
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HydroFix Vaso Shield™ is indicated for use as a cover for vessels following anterior vertebral surgery.

    Device Description

    The MiMedx HydroFix™ Vaso Shield is a flexible sheet of polyvinyl alcohol (PVA) material provided in various dimensions. It is a slippery, clear pliable, conformable, nanoporous provided in valled over that minimizes tissue contact between the vessels and the spine. There are no holes or perforations. There are no markings on either side of the sheet, raised (embossed) or printed. The sheet is provided sterile and hydrated.

    AI/ML Overview

    Acceptance Criteria and Study for HydroFix™ Vaso Shield (K101805)

    The provided document describes the 510(k) premarket notification for the HydroFix™ Vaso Shield. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than establishing novel safety and effectiveness criteria through a new study with predefined acceptance criteria. Instead, the "acceptance criteria" are implied by the demonstration of comparable performance to existing, legally marketed devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Technological Characteristics and Functionality comparable to predicate devices.The device is a flexible sheet of polyvinyl alcohol (PVA). It is described as slippery, clear, pliable, conformable, and nanoporous, designed to minimize tissue contact between vessels and the spine. It has no holes or perforations, no markings, and is provided sterile and hydrated. This aligns with the basic function of a "vessel guard."
    Similar Indications for Use as predicate devices.Indicated as a cover for vessels during anterior vertebral surgery. This is directly comparable to the stated intended use of the Paradis Vaso Shield™ (K090022 and K093551) and Gore PRECLUDE® Vessel Guard (K061727).
    Demonstrated reduction in risk of potential vessel damage in vivo.A sheep model (animal study) was conducted to demonstrate the reduction in the risk of potential vessel damage during anterior vertebral revision surgery. The implanted devices (HydroFix™ Vaso Shield) allowed a separation of the soft tissue adjacent to the study site from the study site. This provides evidence for the device's protective function.
    Safety profile comparable to predicate devices.While not explicitly detailed, the 510(k) process inherently requires a demonstration that the new device does not introduce new or different questions of safety or effectiveness compared to predicate devices. The absence of adverse findings in the animal study and the comparison to established predicate devices serve as the basis for this.
    Materials comparable to predicate devices or acceptable for medical use.The device is made of polyvinyl alcohol (PVA). The overall assessment of substantial equivalence suggests this material is acceptable for the intended purpose, aligning with the materials of predicate devices (polypropylene, polyethylene terephthalate, or polytetrafluoroethylene for the general class, though specific predicate materials are not detailed for comparison).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document mentions a "Large Animal Implantation A sheep model (animal study)." It does not specify the exact number of sheep used in the study.
    • Data Provenance: The study was an in vivo animal study conducted on sheep. The country of origin for the data is not specified. It is a prospective study as an experiment was conducted to generate the data for the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not provide details on the number or qualifications of experts used to establish ground truth for the animal study. Veterinary surgeons or researchers would likely have been involved in the implantation and assessment of the devices in the sheep model.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the animal study data. The results would likely have been analyzed and interpreted by the research team involved.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. An MRMC comparative effectiveness study was not performed. This type of study is typically used for diagnostic devices involving human interpretation of images or other data, which is not applicable to this physical implantable device.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No. This is not
    applicable. The HydroFix™ Vaso Shield is a physical implantable device, not an algorithm or software. Its performance is assessed through its physical interaction with tissues in vivo.

    7. Type of Ground Truth Used

    The ground truth for the in vivo testing was established through direct observation and assessment of the implanted devices in the animal model. This would likely involve:

    • Direct observation/histology: Assessing the physical separation of soft tissue, the presence or absence of vessel damage, and tissue responses at the implant site.
    • Pathology: Likely the most definitive ground truth, involving histological examination of tissues around the implant to confirm device presence, tissue separation, and absence of adverse reactions or damage.

    8. Sample Size for the Training Set

    No training set is mentioned or applicable. This device's performance was evaluated through an in vivo study, not through a machine learning or algorithm-based approach that requires training data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101826

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-07-28

    (27 days)

    Product Code
    Regulation Number
    870.3470
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K100905

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-06-02

    (62 days)

    Product Code
    Regulation Number
    870.3470
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MiMedx Group, Inc. HydroFix™ Vaso Shield is indicated as a cover for vessels during anterior vertebral surgery.

    Device Description

    The MiMedx HydroFix™ Vaso Shield is a flexible sheet of polyvinyl alcohol (PVA) material provided in various dimensions. There are no holes or perforations. There are no markings on either side of the sheet, raised (embossed) or printed. The sheet is provided sterile and hydrated.

    AI/ML Overview

    The provided text describes a 510(k) summary for the MiMedx HydroFix™ Vaso Shield, focusing on demonstrating its substantial equivalence to predicate devices rather than establishing novel acceptance criteria for device performance as an AI device. Therefore, a direct table of "acceptance criteria" and "reported device performance" in the context of an AI device's metrics (like sensitivity, specificity, etc.) is not applicable here.

    However, I can extract the performance criteria that were evaluated to show substantial equivalence and explain the study conducted.

    Here's a breakdown of the requested information based on the provided text, interpreted for a medical device (not an AI device):

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative "acceptance criteria" with numerical thresholds for performance metrics. Instead, it describes comparative performance against predicate devices to demonstrate "comparable" or "equivalent" performance.

    Performance Criteria (Evaluated Attributes)Device Performance (Reported Outcome)
    In Vitro Testing:
    Suture pull outResults demonstrate comparable technological characteristics and performance to predicate devices.
    In Vivo Testing:
    Reflectivity of light from the surgical sheetResults show the MiMedx HydroFix™ Vaso Shield is comparable to predicate devices and can perform equivalently for the intended use.
    Edge sharpnessResults show the MiMedx HydroFix™ Vaso Shield is comparable to predicate devices and can perform equivalently for the intended use.
    Ability to suture to tissueResults show the MiMedx HydroFix™ Vaso Shield is comparable to predicate devices and can perform equivalently for the intended use.
    Ability to secure to tissueResults show the MiMedx HydroFix™ Vaso Shield is comparable to predicate devices and can perform equivalently for the intended use.
    Ability to cut the sheetResults show the MiMedx HydroFix™ Vaso Shield is comparable to predicate devices and can perform equivalently for the intended use.

    2. Sample size used for the test set and the data provenance

    • Sample size for in vitro testing: Not specified.
    • Sample size for in vivo testing: Not specified, only mentions "a porcine model," implying one or more animals were used.
    • Data provenance: The in vivo study was conducted in a "porcine model," indicating animal data. The location or specific details of where the in vitro testing occurred are not provided. Both types of studies appear to be prospective, as they were conducted to demonstrate substantial equivalence for this specific 510(k) application.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable and not provided in the document. The studies conducted (in vitro and in vivo animal studies) do not involve human expert interpretation for "ground truth" in the way an AI device study would. The performance was likely assessed by researchers and possibly veterinary professionals in the case of the porcine model.

    4. Adjudication method for the test set

    Not applicable and not provided. This concept is typically relevant for human-interpreted diagnostic findings, not for the physical performance testing of a medical device described here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device, not an AI device. No MRMC study involving human readers or AI assistance was conducted or mentioned.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm or AI.

    7. The type of ground truth used

    For the in vitro and in vivo studies, the "ground truth" would be objective measurements and observations of the device's physical and functional characteristics (e.g., how well it holds a suture, its light reflectivity) under controlled experimental conditions. It is not based on expert consensus, pathology, or outcomes data in the typical sense of diagnostic ground truth.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As above, this is a physical medical device, not an AI model.