Search Results
Found 227 results
510(k) Data Aggregation
(176 days)
The Hudson RCI® Infant Dri-Tech™ Breathing Circuits are intended for neonatal, infant, and pediatric patients weighing 32kg or less in a professional healthcare environment as a conduit for warmed and humidified respiratory gases between a patient and a ventilator and includes heated wires for use with the Hudson RCI Neptune Heated Humidifier.
The Hudson RCI® Infant Dri-Tech Breathing Circuits are intended for neonatal, infant, and pediatric patients weighing 32kg or less in professional healthcare environments as a conduit for respiratory gas between a patient and a ventilator and includes heated wire(s) for use with the Hudson RCI Neptune Heated Humidifier (K131912). The heated wires are intended to aid in maintaining the set patient temperature and minimize condensation in the ventilator tubing.
The Dri-Tech Infant Breathing Circuits are designed to be used with various patient interfaces such as oral tracheal tubes, nasal tracheal tubes, endo-bronchial tubes, and masks (for noninvasive ventilation therapies such as CPAP, BiPAP, Cannula and Mask usage). As such, Dri-Tech Breathing Circuits are designed and will be manufactured to conform to the test requirements of ISO 5367:2014 "Anaesthetic and respiratory equipment -- Breathing sets and connectors".
The dual limb configuration consists of 12mm ID inspiratory and expiratory limbs with 22mm machine side connectors that are compliant to ISO 5356-1:2015 Anaesthetic and respiratory equipment – Conical connectors – Part 1: Cones and sockets, and 12mm patient side connectors that utilize the same taper and engagement length as ISO 5356-1:2015 connectors. The infant circuits have two length configurations, a short inspiratory that is 1.12m or a long inspiratory that is 1.6m, either of which can be used with an expiratory limb that is 1.7m, and may be kitted with the accessories described below:
- Accessory Tee with Closure Cap: Optional tubing adaptor for various connections as required.
- Remote Port Tubing, 6" and 12": Unheated sections of tubing for use with incubators, isolettes, or other ambient heating to ensure the temperature probe is not heated by the environment.
- Unheated Ventilator Tubing: Required tubing and adaptors for connecting the ventilator inhalation port to the heated humidifier.
The single limb configuration is identical to the long version of the inspiratory limb of the dual limb configuration, however, is packaged with a different set of adaptors to facilitate connections and use with non-invasive therapies as described below:
- Unheated Ventilator Tubing: Optional tubing and adaptors for connecting the ventilator inhalation port to the heated humidifier.
- Pressure Monitoring Line: Optional accessory tubing for use with units without an Airway Pressure Monitor.
All Hudson RCI Dri-Tech breathing circuit configurations are packaged in polyethylene (PE) bags to protect the devices from environmental influences like dust and moisture. Included in the PE bags are additional PE bags containing ConchaSmart columns (-KIT configurations only) and included accessories. The breathing circuit PE bags are packaged ten (10) to a cardboard carton for all configurations. They are sold non-sterile.
N/A
Ask a specific question about this device
(268 days)
The Konig Bell Circumcision clamps are reusable instruments used in a medical procedure to compress the foreskin of the penis during the circumcision of male infant or child. Use of the Konig Bell Circumcision clamps in adults is to assist in Human Immunodeficiency Virus (HIV) prevention.
The Konig Bell Circumcision Clamp is intended to be used in a medical procedure to compress the foreskin of the penis during circumcision of a male infant, child or adult. The device is manufactured from Germangrade 304 stainless steel and is composed of four parts: base, bell, hook arm and thumb screw. The device is offered in the following bell sizes 0.8cm, 1.1cm, 1.3cm, 1.45cm, 1.6cm, 2.1cm, 2.6cm, 2.9cm, 3.2cm and 3.5cm.
The Konig Bell Circumcision Clamp is packaged in a low-density polyethylene film pouch. The device is provided non-sterile and is intended to be cleaned and sterilized at the healthcare facility prior to use. Validated cleaning and steam sterilization instructions are included in the instructions for use.
N/A
Ask a specific question about this device
(58 days)
The Namic Radial Arm Band is indicated to apply compression in order to achieve hemostasis while allowing the user to maintain patency of the radial artery after a transradial procedure (patent hemostasis).
The Namic Radial Arm Band is a tourniquet-style device consisting of a soft wristband with a secure hook and loop adjustable fastener on each end, two compression balloons, a support plate, flexible tubing with the input valve, and an inflation device. The device is offered in two lengths, 23 cm and 26 cm.
The use of the Namic Radial Arm Band begins after the transradial catheterization procedure is concluded and as the introducer sheath is gradually withdrawn from the patient. The wristband is placed around the patient's wrist with the adjustable hook and loop fastener. Once the Radial Arm Band is secured in place, the inflation device is connected via the input valve and tubing. The inflation device is used to apply pressure to the patient's access site. Both radial compression balloons (RCB) are filled with air at the same time to allow for efficient compression for hemostasis. The volume of air can be increased or decreased using the inflation device, thus allowing the physician to adjust the pressure in the Radial Arm Band. The Namic Radial Arm Band is a disposable, single use only device.
N/A
Ask a specific question about this device
(260 days)
Intended for the aseptic dispensing of solutions from IV containers. For use in transferring IV fluids/medication from a bag to an IV fluid administration device.
The Medline Bag Decanter is a non-pyrogenic, single use, disposable device, which is supplied sterile. The device comprises of a one-piece transfer device with protective flexible caps at either end. The device is an injection molded hollow tube with a spiked end used to access the source container and withdraw fluid. The spike component is designed with a built-in splash guard. The device is designed for use in transferring IV fluids/medication from a bag to an IV fluid administration device.
N/A
Ask a specific question about this device
(149 days)
Hudson RCI® Comfort Flo Nasal Cannulas Premature, Infant:
Hudson RCI® Comfort Flo Nasal Cannulas Premature, Infant are intended to be used in conjunction with the Comfort Flo Humidification System to provide a continuous flow of heated and humidified air/oxygen mixtures to spontaneously breathing patients.
It is indicated for single use by premature and infant (birth to 2 years) patients in professional healthcare environments.
Hudson RCI® Comfort Flo® Plus Cannulas and Comfort Flo® Soft Plus Cannulas Extra Small:
Hudson RCI® Comfort Flo® Plus Cannulas and Comfort Flo® Soft Plus Cannulas Extra Small are intended to be used in conjunction with the Humidification System to provide a continuous flow of heated and humidified air/oxygen mixtures to spontaneously breathing patients.
It is indicated for single use by adult and pediatric (12 years and above) patients in professional healthcare environments.
Hudson RCI® Comfort Flo Nasal Cannulas Premature, Infant
Hudson RCI® Comfort Flo Plus Cannulas Extra Small
Hudson RCI® Comfort Flo Soft Plus Cannulas Extra Small
This is a 510(k) clearance letter for a medical device (nasal cannula), not an AI/ML-driven device. Therefore, the information typically requested for AI/ML device validation (such as expert consensus, MRMC studies, or training/test set details) is not applicable.
The document describes the device, its intended use, and a summary of non-clinical testing performed to demonstrate substantial equivalence to predicate devices. It clearly states that no clinical testing was performed.
Here's a breakdown of the requested information based on the provided document, highlighting why many points are not applicable for this type of device clearance:
-
A table of acceptance criteria and the reported device performance
- The document lists various non-clinical performance tests but does not provide a table of specific acceptance criteria with corresponding reported device performance values. Instead, it generally states that testing was conducted "to demonstrate substantial equivalence." For example, it lists "Relevant Humidity Output testing" but doesn't state the specific humidity output range considered acceptable or the measured output. This is typical for a 510(k) submission where broad equivalence is the goal rather than meeting precise performance thresholds for novel functionality.
-
Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not applicable. This is a physical medical device. Testing involved bench tests on manufactured units, not data sets in the AI/ML sense. Data provenance, retrospective/prospective, and sample sizes for test sets (in the context of patient data) are not relevant as no human or clinical data was used for validation.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. No "ground truth" was established by experts in the context of diagnostic performance, image interpretation, or similar AI/ML applications, as this is a physical medical device.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. There was no test set requiring expert adjudication for performance evaluation.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is not an AI-assisted device, and no MRMC study was conducted. The document explicitly states "No clinical testing was performed."
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not applicable. This is not an algorithm-based device.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not applicable. The concept of "ground truth" as it applies to AI/ML model validation is not relevant here. The device's performance was evaluated against engineering specifications and industry standards for physical device functionality (e.g., flow rates, material compatibility, strength).
-
The sample size for the training set
- Not applicable. There is no AI/ML model, and thus no training set.
-
How the ground truth for the training set was established
- Not applicable. There is no AI/ML model, and thus no training set or ground truth establishment for it.
In summary, the provided document is a 510(k) clearance for a physical medical device, not an AI/ML-driven product. Therefore, most of the requested information related to AI/ML validation (ground truth, expert studies, training/test sets) is irrelevant to this submission.
The acceptance criteria for this device focus on:
- Biocompatibility: Meeting ISO standards (e.g., ISO 10993 series, ISO 18562 series) to ensure the materials are safe for patient contact and breathing gas pathways. The document lists the specific ISO standards used.
- Performance Bench Testing: Covering aspects like visual inspection, humidity output, thermal overshoot, connection strength, leak testing, headgear testing, shelf life, and useful life. While the specific numerical acceptance criteria for each test (e.g., "humidity output must be within X amount") are not explicitly listed in this summary, the document indicates these tests were performed and implicitly met the company's internal acceptance criteria for demonstrating substantial equivalence.
- Substantial Equivalence: The overarching acceptance criterion for a 510(k) is demonstrating that the device is "as safe and as effective" as a legally marketed predicate device, based on the non-clinical testing performed.
The "Study that proves the device meets the acceptance criteria" for this product consists of the various Non-Clinical Testing described on pages 16-17, primarily Biocompatibility Testing and Performance Testing (Bench), which were conducted to support the claim of substantial equivalence to the predicate devices.
Ask a specific question about this device
(114 days)
The Medline UNITE® REFLEX® Hybrid Nitinol Implants are intended to provide fixation for fractures, fusions or osteotomies of the bones of short (tarsals) and long (metatarsals, phalanges, distal tibia and fibula) bones comprising the foot and ankle such as: First metatarsal-cuneiform arthrodesis, First metatarsophalangeal arthrodesis, Talo-Navicular Fusion, Lis-Franc arthrodesis, Scarf and Chevron osteotomies. Implants are intended for single use only.
The Medline UNITE® Locking and Non-Locking Cortical Screws are indicated for use with the Medline UNITE® REFLEX® Hybrid Nitinol Implant. The Non-Locking Cortical Screws are also indicated for bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
The Medline UNITE® REFLEX® Hybrid Nitinol Implants are manufactured from nickel titanium alloy (nitinol). The implants utilize chemical etching and passivation to form a protective oxidation layer on the outer surface. Chemical etching and passivation are common processes to create a uniform oxidation layer on the surface of the implant. The system includes implants offered in various styles, sizes, and options; each designed for specific anatomy and procedures. The implants can accommodate Ø2.7mm, Ø3.5mm, and Ø4.0mm locking and non-locking cortical screws to be used with the polyaxial locking holes and compression slots. The system also includes reusable and disposable instrumentation necessary for implantation of the REFLEX® Hybrid Nitinol Implant.
This 510(k) clearance letter is for a medical implant system, not an AI/software device, so the typical AI-related acceptance criteria and study details (like sample size for test set, experts for ground truth, MRMC study, standalone performance, training set details) are not applicable to this document. The provided text describes the regulatory clearance process for a physical medical device.
The document focuses on demonstrating substantial equivalence to a previously cleared predicate device, the Medline UNITE® REFLEX® Nitinol Staple System (K231885). This is achieved primarily through non-clinical (bench) testing and engineering analysis, not clinical trials or AI performance evaluations.
Here's a breakdown of the requested information based on the provided document, noting where specific AI-related questions are not applicable:
Acceptance Criteria and Device Performance (for a physical medical implant)
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Device Performance (Medline UNITE® REFLEX® Hybrid Nitinol Implant System) |
|---|---|
| Substantial Equivalence to Predicate Device (K231885) | Demonstrated through engineering analysis and performance testing. The subject device is concluded to be as safe and effective for its intended use as the predicate device. |
| Primary Intended Use (Fixation for fractures, fusions, osteotomies of short and long bones of foot and ankle) | Same as predicate. |
| Indications for Use | Similar to predicate, with additional indications for locking and non-locking cortical screws (cleared under K234031) used with polyaxial locking holes and compression slots. Overall indications remain within the scope of the predicate. |
| Material Composition (Nickel Titanium Alloy, Nitinol, ASTM F2063 compliance) | Same as predicate. |
| Sterility (Provided non-sterile, steam sterilized prior to use) | Same as predicate. |
| Elastic Static Bending Strength | Engineering analysis determined the subject device does not represent a new worst-case compared to the predicate, demonstrating substantial equivalence. |
| Axial Pullout Strength | Engineering analysis determined the subject device does not represent a new worst-case compared to the predicate, demonstrating substantial equivalence. |
| Corrosion Susceptibility (ASTM F2129) | Testing was conducted to meet predefined acceptance criteria. Initial testing showed "not all samples met the minimum acceptance criteria for electrostatic breakdown potential," leading to additional nickel ion release testing. |
| Nickel-ion Release (ASTM F3306) | The total Ni-ion release and maximum daily Ni-ion release for any of the three (3) samples tested over 60 days was significantly lower than the FDA guidance for the Tolerable Intake (TI) value for parenteral exposure to nickel (0.5 μg/kg/day, or 35 μg/day for a 70 kg adult). |
| Galvanic Corrosion (ASTM F3044) | Microscopic examination of all three (3) test samples revealed no pitting or indications of corrosion. The average calculated material release from mass loss was significantly less than the FDA guidance for the Tolerable Intake (TI) value for parenteral exposure to nickel (0.5 μg/kg/day, or 35 μg/day for a 70 kg adult). |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set:
- For Corrosion Susceptibility: Mentioned "not all samples," implying multiple samples were tested, but an exact number for the "test set" is not specified for the initial pass/fail.
- For Nickel-ion Release: "three (3) samples tested over 60-days."
- For Galvanic Corrosion: "all three (3) test samples."
- Data Provenance: The testing was non-clinical (bench testing) and performed according to ASTM standards and FDA guidance documents. There is no mention of country of origin of data as it's not human or patient data. It is inherently "prospective" in the sense that the tests were conducted specifically for this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. This is a physical medical device. Ground truth as typically defined in AI/software evaluations (e.g., expert consensus on medical images) is not relevant here. The "ground truth" for the device's performance is established by standardized material and mechanical testing specifications (ASTM standards and FDA guidance).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. This is a physical medical device. Adjudication methods are typically used for expert consensus on challenging cases in AI/software evaluations.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is a physical medical device, not an AI-assisted diagnostic tool. No MRMC study was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for this physical device is defined by established engineering and material science standards (ASTM F2129, ASTM F3306, ASTM F3044) and FDA guidance on biocompatibility and material safety (specifically for Nitinol implants and nickel release). Device performance is measured against these quantitative and qualitative benchmarks.
8. The sample size for the training set
- Not Applicable. This is a physical medical device. There is no "training set" in the context of machine learning.
9. How the ground truth for the training set was established
- Not Applicable. This is a physical medical device. There is no "training set" or corresponding ground truth establishment process in the context of machine learning.
Ask a specific question about this device
(159 days)
The Medline Level 4 Surgical Gown with Breathable Sleeves is a sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter. The Medline Level 4 Surgical Gown with Breathable Sleeves meets the Level 4 requirements of ANSI/AAMIPB70:2022 Liquid barrier performance and classification of protective apparel and drapes intended for use in healthcare facilities.
The Medline Level 4 Surgical Gown with Breathable Sleeves is also sold in a non-sterile version to repackagers/kitting entities, to be sterilized using the validated ethylene oxide sterilization method according to ISO 11135-1 in its final finished form, as a kit component.
The Medline Level 4 Surgical Gown with Breathable Sleeves is a single-use, sterile, disposable medical device that is provided in a variety of sizes (Large, X-Large, 3X-Large, 3X-Large and X-Long (only Sirus)) and styles (Sirus, Aurora and Eclipse). The Medline Level 4 Surgical Gown with Breathable Sleeves is constructed from nonwoven SMS (spunbond, meltblown, spunbond), a poly-fabric reinforcement on the chest front panel, and a breathable film (Polyester ether and polyester) for the sleeves. Each gown is designed with a hook and loop closure at the neck, ties at the waist, knitted cuffs at the end of the sleeves as well as a transfer tab for ease in donning. The Medline Level 4 Surgical Gown with Breathable Sleeves has been tested according to ANSI/AAMI PB70:2022 and meets AAMI Level 4 barrier level protection for a surgical gown.
This document describes the performance testing for the Medline Level 4 Surgical Gown with Breathable Sleeves (K242844), a surgical apparel device. The information provided is based on non-clinical testing only, as no clinical testing was performed for this 510(k) submission.
Here's an analysis of the acceptance criteria and the study proving the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance:
| Standard | Standard Title | Acceptance Criteria | Reported Device Performance (Results) |
|---|---|---|---|
| ISO 10993-5 Cytotoxicity | ISO MEM Elution Using L-929 Mouse Fibroblast Cells | Non-cytotoxic | Pass - Non-cytotoxic |
| ISO 10993-23 Irritation | ISO Intracutaneous Irritation Test | Non-irritating | Pass - Non-irritating |
| ISO 10993-10 Sensitization | ISO Guinea Pig Maximization Sensitization Test | Non-sensitizing | Pass - Non-sensitizing |
| ASTM F1671 | Resistance of Materials Used in Protective Clothing to Penetration by Blood-Borne Pathogens | 4.0 AQL | Pass |
| AATCC 42 | Water Resistance: Impact Penetration Test | ≤4.5 g | Pass |
| AATCC 127 | Water Resistance: Hydrostatic Pressure Test | ≥50 cmH20 | Pass |
| ASTM D5034-09 (2013) | Breaking Strength and Elongation of Textile Fabrics (Grab Test) | ≥20 N | Pass |
| ASTM D5587-15 | Tearing Strength of Fabrics by Trapezoid Procedure | ≥20 N | Pass |
| ASTM D1683 | Standard Method for Failure in Sewn Seams of Woven Apparel Fabrics | ≥20 N | Pass |
| 16 CFR 1610 | Flammability of Clothing Textiles | Meets Class 1 Requirements | Pass |
| ASTM E96 | Water Vapor Transmission of Materials | Reinforced Outside Material >800 g/m²/24 hrs | Pass |
| Sleeve Material >1200 g/m²/24 hrs | Pass | ||
| ANSI/AAMI PB70:2022 | Liquid Barrier performance and classification of protective apparel and drapes intended for use in healthcare facilities. | Meets ANSI/AAMI PB 70:2012 Level 3 and Level 4 Liquid Barrier requirements | Pass |
2. Sample Size Used for the Test Set and Data Provenance:
The document describes non-clinical tests conducted to verify design specifications. For each test listed in the table, the "Results" column indicates a "Pass," meaning the acceptance criteria were met.
- Sample Size: The specific sample sizes for each non-clinical test (e.g., number of fabric specimens tested for ASTM F1671) are not detailed in this summary. The 510(k) summary typically provides a high-level overview, and the detailed test reports (which FDA reviews) would contain this information.
- Data Provenance: The tests are generally performed by accredited testing laboratories or the manufacturer's own qualified labs. The document does not specify the country of origin of the data. Given it's an FDA submission, the tests would typically adhere to recognized international standards like ISO and ASTM. These are prospective tests performed specifically to support the regulatory submission of this new device.
3. Number of Experts Used to Establish Ground Truth and Qualifications:
- Not Applicable. For this type of medical device (surgical gowns), "ground truth" is established through adherence to standardized, objective performance tests (e.g., measuring water resistance, breaking strength, biocompatibility). These are not studies that rely on expert interpretation of images or clinical outcomes to establish ground truth in the way an AI diagnostic device would.
4. Adjudication Method for the Test Set:
- None. Adjudication methods (like 2+1, 3+1) are relevant for studies involving human interpretation or clinical endpoints, often in the context of diagnostic accuracy or clinical trials. As this is a non-clinical performance evaluation against objective standards, no adjudication method is used. The results are quantitative measurements interpreted against pre-defined thresholds.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done:
- No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices (especially those involving image interpretation by human readers) to assess how an AI algorithm impacts reader performance. The Medline Level 4 Surgical Gown is a protective apparel, not a diagnostic device, and its performance is evaluated through physical and material property tests, not human reader studies.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not Applicable. This question typically pertains to AI/ML software as a medical device (SaMD). The Medline Level 4 Surgical Gown is a physical medical device. Its performance is inherent in its material properties and manufacturing, not an algorithm.
7. The Type of Ground Truth Used:
The "ground truth" for this device's performance is established by objective, quantitative measurements against published national and international standards. This includes:
- Biocompatibility Standards: ISO 10993 series (Cytotoxicity, Irritation, Sensitization).
- Barrier Performance Standards: ASTM F1671 (blood-borne pathogens), AATCC 42 (impact penetration), AATCC 127 (hydrostatic pressure), and ANSI/AAMI PB70:2022 (overall liquid barrier performance classification).
- Material Strength Standards: ASTM D5034 (breaking strength), ASTM D5587 (tearing strength), ASTM D1683 (sewn seam failure).
- Other Performance Standards: 16 CFR 1610 (flammability), ASTM E96 (water vapor transmission/breathability).
The "ground truth" is therefore the measured physical and chemical properties of the gown, compared against the predefined acceptance criteria of these standard test methods.
8. The Sample Size for the Training Set:
- Not Applicable. The concept of a "training set" is relevant for machine learning algorithms. This device is a manufactured product, not an AI algorithm. Its design and manufacturing process would involve material selection and engineering, not machine learning training.
9. How the Ground Truth for the Training Set Was Established:
- Not Applicable. As there is no training set for an AI algorithm, this question is not relevant.
Ask a specific question about this device
(182 days)
The Hudson RCI Comfort Flo® CubCannula™ is a single use nasal cannula intended to administer respiratory gas to a spontaneously breathing patient through both nostrils delivering flow rates ranging from 1 to 25 LPM.
CubCannula™ is a patient interface used to deliver heated and humidified gas to the following populations:
- · Neonates, birth to 1 month of age
- · Infants, 1 month to 2 years of age
- · Children, 2 to 12 years of age
CubCannula™ is designed for use in professional healthcare environments and must be prescribed by a physician.
CubCannula is used in conjunction with the Humidification System intended to provide a continuous flow of heated and humidified air/oxygen mixtures to spontaneously breathing patients.
The Comfort Flo® CubCannula™ is a single patient use nasal cannula containing two-prongs intended to administer gas to a spontaneously breathing patient through both nostrils. CubCannula™ is a patient interface used as a conduit to deliver heated and humidified gas to the following pediations: neonates, infants, children. The nasal cannula incorporates soft prongs, a soft, silicone, kink and crush resistant lariat, in addition to a color-coded bolo for easy size identification. The CubCannulas will be available in five sizes: Extra Small (XS), Small (XS), Small (XS), Small (XS), Medium (M), Large (L), and Extra Large (XL). The nasal cannula will be secured to the patient's face via use of Hydrocolloid pads (CubPads™) to assist with easy positioning and change out of the cannula during use.
The CubCannula™ will be individually packaged with a pair (2) of CubPad™ replacements will be available in a separate package.
Based on the provided text, the device in question is the Hudson RCI Comfort Flo® CubCannula™, a nasal cannula. The document is a 510(k) Premarket Notification, which demonstrates substantial equivalence to a legally marketed predicate device, rather than proving a device meets specific acceptance criteria through a clinical study as might be done for a novel device or pre-market approval (PMA) pathway.
Therefore, the information regarding "acceptance criteria" and a "study that proves the device meets the acceptance criteria" in the traditional sense of a clinical trial with pre-defined primary and secondary endpoints is not explicitly present for this 510(k) submission. Instead, the focus is on demonstrating equivalence through non-clinical testing (biocompatibility and bench testing).
Here's an analysis of the provided information relative to your questions, with specific callouts where information is not available due to the nature of a 510(k) substantial equivalence submission:
1. A table of acceptance criteria and the reported device performance
The document does not present a formal table of acceptance criteria with corresponding performance results in the way a clinical study report would. Instead, it lists the non-clinical tests performed and states that they revealed "passing results" or that "emissions are well below the health-based thresholds." The "acceptance criteria" are implied by the standards cited and the statement of compliance.
Implied Acceptance (Compliance) and Reported Performance (Statement of Passing):
| Test Category | Implied Acceptance (Standard/Requirement) | Reported Device Performance |
|---|---|---|
| Biocompatibility | ISO 10993-1:2018 (General Evaluation), ISO 10993-5: 2009 (Cytotoxicity), ISO 10993:10: 2010 (Irritation & Sensitization), ISO 10993-11:2017 (Systemic Toxicity), ISO 10993-6:2016 (Local Effects after Implantation), ISO 10993-3:2014 (Genotoxicity, Carcinogenicity, Reproductive Toxicity), ISO 10993-12:2012 (Sample Prep), ISO 10993-17:2002 (Allowable Limits for Leachables), ISO 10993-18:2005 (Chemical Char.); ISO 18562-1:2017 (Breathing Gas), ISO 18562-2:2017 (Particulate Matter), ISO 18562-3:2017 (VOCs). | "All biological endpoint testing including cytotoxicity, sensitization, irritation, acute systemic toxicity, and genotoxicity revealed passing results.""The extractables/leachables testing on the humidified gas pathway components revealed MOS values greater than 1.0 after clinically relevant considerations and condensate attenuation.""VOC and Particulates that the CubCannula emissions are well below the health-based thresholds for both volatile organic compounds and particulates.""the CubCannula referenced in this document is biocompatible and does not present a foreseen biological and/or toxicological risk." |
| Performance (Bench) | ISO 5356-1:2015 (Conical connectors), ISO 80601-2-74:2017 (Respiratory devices), ASTM F1980-21 (Accelerated Aging), ASTM D4169-22 (Shipping Containers). | Performed tests include: Post aging Visual Inspection, Relevant Humidity Output testing, Thermal Overshoot testing, Connection strength testing, Flow Leak Test/Gas path leak testing, Shelf Life Testing, Useful Life Testing, Peel Testing, Transportation simulation, ISO gauging. All are implicitly stated as having demonstrated substantial equivalence. Specific quantitative results are not provided in this summary. For "Useful Life," it states: "Testing has been performed to validate the 10-day useful life (safety factor included)." |
| Product Characteristics | Maintain functional equivalence (e.g., flow rates, patient population, counter-indications) to predicate device. | Flow rates (1-25 LPM) are similar to the predicate (0.5-25 LPM, max 36LPM for XXL size not included). Useful life (10 days) has been validated vs predicate (7 days). Shelf life (2 years) vs predicate (3 years). These differences were assessed not to raise new safety/effectiveness concerns. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the document as it pertains to a 510(k) submission based on non-clinical (bench) testing, not a clinical study. The "test set" would refer to the physical units of the device tested in the lab, but the number of units is not specified. There is no patient data or country of origin for the data as no clinical testing was performed ("No clinical testing was performed").
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable as there was no clinical study, no patient data, and therefore no "ground truth" established by experts in the context of clinical interpretation or outcomes. The ground truth for the non-clinical tests would be the established performance standards set by the cited ISO and ASTM documents.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable as there was no clinical study or expert review for adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device is a physical medical device (nasal cannula), not an AI-powered diagnostic or assistive tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable, as it is not an algorithm-based device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Since no clinical study comparing the new device's performance against a "ground truth" in patients was conducted per the document ("No clinical testing was performed"), the concept of "ground truth" in a clinical sense is not applicable. For the non-clinical testing, the "ground truth" is adherence to the specified performance characteristics and safety profiles as measured against the requirements of the cited ISO and ASTM standards. These standards define the acceptable range for each test parameter (e.g., cytotoxicity, leachable substances thresholds, connection strength, leak rates).
8. The sample size for the training set
This is not applicable. This is a physical device, not an AI/machine learning model that requires a training set.
9. How the ground truth for the training set was established
This is not applicable for the reasons stated in point 8.
Ask a specific question about this device
(132 days)
The Medline Microdissection Needle is a monopolar electrosurgical instrument used for precision soft tissue dissection. It is a single-use device intended for cutting, dissecting, and cauterizing soft tissue. The Medline Microdissection Needle is not intended for use in the central nervous system or in the central circulatory system.
The Medline Microdissection Needle is a sterile, active electrode that is also referred to as a needle. It is designed for precision cutting, dissecting, and cauterization. The device is comprised of stainless steel, tungsten, and polyolefin heat shrink and will be offered in two different lengths, 4 cm (140936/ESE104A) and 3 cm (140937/ESE103A). The Medline Microdissection Needle is intended for connection with multiple brands of standard cautery pencils, including 510(k)-cleared Medline Cautery Pencil (K190643).
The Medline Microdissection Needle is a medical device and is not an AI/ML powered device, therefore no information about AI/ML acceptance criteria or studies is available in the provided text.
The provided document describes the 510(k) premarket notification for the Medline Microdissection Needle, comparing it to a predicate device. This submission process focuses on demonstrating substantial equivalence to a legally marketed device rather than establishing new acceptance criteria or conducting studies in the way one might for a novel AI/ML product.
Here's a breakdown of the non-clinical testing performed to demonstrate substantial equivalence, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't present a table of specific acceptance criteria with numerical performance targets in the way an AI/ML product might. Instead, it demonstrates equivalence through a comparison of technological characteristics and various non-clinical tests. The "Comparison Analysis" column in Table 1 serves as the "reported device performance" against the predicate device's characteristics.
| Device Characteristic | Proposed Device Performance | Predicate Device Characteristic | Comparison Analysis |
|---|---|---|---|
| Product Name | Medline Microdissection Needle | Stryker Colorado Microdissection Needle® | N/A |
| 510(k) Reference | K242370 | K033232 | N/A |
| Manufacturer | Applied Medical Coatings (for proposed device) | Stryker | N/A |
| Product Code | GEI | GEI | Same |
| Classification Panel | General & Plastic Surgery | General & Plastic Surgery | Same |
| Classification Name | Electrosurgical Cutting & Coagulation & Accessories | Electrosurgical Cutting & Coagulation & Accessories | Same |
| Regulation Number | 878.4400 | 878.4400 | Same |
| Indication for Use | Monopolar electrosurgical instrument for precision soft tissue dissection; single-use for cutting, dissecting, and cauterizing soft tissue; not for CNS or central circulatory system. | Monopolar electrosurgical instrument for precision soft tissue dissection (e.g., tonsillectomy, blepharoplasty); single-use for cutting, dissecting, and cauterizing soft tissue; not for CNS or central circulatory system. | Same – Both devices for precision soft tissue dissection and for cutting, dissecting, and cauterizing soft tissues. |
| Materials | Tube: Stainless Steel; Tip/Needle: Tungsten; Insulation: Polyolefin heat shrink; Tip Protector: PVC Compound; Tip Protector Colorant: Ethylene Vinyl Acetate | Tube: Stainless Steel; Tip/Needle: Tungsten; Insulation: Polyolefin (latex free); Non-stick coating: Isolation PTFE | Similar: Minor differences in materials of construction are addressed via biological safety testing per ISO 10993-1 and functional testing. The differences do not alter the safety and effectiveness of the product. |
| Design Features | Ultra-sharp (5 µm) tungsten needle that transfers current from electrosurgical generator to very small tissue surface area. | Ultra-sharp (5 µm) needle that transfers current from electrosurgical generator to very small tissue surface area. PTFE coating on tip to reduce eschar build up. | Similar: Functional testing, including thermal zone damage testing and tip/pencil compatibility testing on the subject device demonstrate the differences do not alter the safety and effectiveness of the product. |
| Active Accessory | Compatible with a variety of standard monopolar electrosurgical hand pieces. | Compatible with a variety of standard monopolar electrosurgical hand pieces. | Same |
| Single Use vs. Reusable | Single-Use | Single-Use | Same |
| Sterile vs. Non-Sterile | Sterile (EO) | Sterile (EO) | Same |
| OTC vs Rx | Rx only | Rx only | Same |
| Shelf Life | 3 years | 5 years | Different: Shelf-life testing supports the proposed expiration date through evaluation of the packaging integrity for maintaining device sterility and/or evaluation of any changes to device performance or functionality. |
| Dimensions | ESE103A: 51.75 ± 1.78 mm; ESE104A: 61.75 ± 1.78 mm; Outer Shaft Diameter: 2.36 ± 0.1 mm | Not explicitly stated for predicate in table, but discussed in comparison. | Functional testing, including thermal zone damage testing and tip/pencil compatibility, was performed on the subject device to ensure the differences do not alter the safety and effectiveness of the product. |
| Connector Type | Stainless Steel Tube | Stainless Steel Tube | Same |
| Performance Specifications | Rated Supply Voltage: 4000Vpeak; Max Power Output: 30 watts | Rated Supply Voltage: 5400 V; Max Power Output: 50 watts | Functional testing, including thermal zone damage testing and IEC 60601-2-2 was performed on the subject device to ensure the differences do not alter the safety and effectiveness of the product. |
| Compatibility | IFU advises verifying compatibility of all instruments and accessories, always with monopolar electrosurgical generators and accessories tested to IEC 60601 and EN 60601 standards. Compatible with any active handle with same or lower rated accessory voltage and complies with IEC 60601-2-2 given proper and secure fit. | IFU advises using with regulatory cleared and approved generators compatible with monopolar electrosurgical devices. Fits most cautery handpieces with a standard 2.4 mm or 1.6 mm shaft. | Similar |
2. Sample Size Used for the Test Set and Data Provenance:
The document primarily describes non-clinical bench testing, biocompatibility testing, electrical safety, and stability testing. It does not involve a "test set" of patient data or clinical images in the context of an AI/ML device. Therefore, a sample size for a test set and data provenance (country of origin, retrospective/prospective) are not applicable as typically understood for AI/ML studies.
For bench tests, sample sizes would relate to the number of devices tested to meet statistical significance for particular tests (e.g., thermal damage, compression, dimensional measurements). These specific numbers are not detailed in the provided text.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
Not applicable. This is not an AI/ML study involving human expert ground truth. The "ground truth" for this device's performance is established by adherence to recognized standards and validated physical/electrical performance tests.
4. Adjudication Method for the Test Set:
Not applicable. There is no human adjudication process described as this is not an AI/ML study.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
No, an MRMC comparative effectiveness study was not done. This type of study is typically conducted for diagnostic devices, especially those incorporating AI, to assess the impact of the device on human reader performance. The Medline Microdissection Needle is a surgical instrument.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:
Not applicable. The Medline Microdissection Needle is a physical electrosurgical instrument used by a surgeon, not an algorithm.
7. The Type of Ground Truth Used:
The "ground truth" for the Medline Microdissection Needle is established through:
- Adherence to recognized international standards: e.g., ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-11, ISO 11135, ISO 11607, ASTM F1980, ANSI/AAMI ST 72, USP <85>, IEC 60601 series.
- Physical and Electrical Performance Specifications: Demonstrated through bench testing (Thermal Damage Zone Testing, Tip/Pencil Compatibility, Appearance, Dimensional Measurements, Compression Test, Electrical Safety, EMC testing).
- Biocompatibility Testing: Evaluating potential biological responses through various ISO 10993 tests.
- Stability (Shelf Life) Testing: Demonstrating integrity over the proposed shelf life.
8. The Sample Size for the Training Set:
Not applicable. This is not an AI/ML product.
9. How the Ground Truth for the Training Set was Established:
Not applicable. This is not an AI/ML product.
In summary, the provided document focuses on demonstrating that the Medline Microdissection Needle is substantially equivalent to a predicate device through comprehensive non-clinical testing against established engineering and biological standards, rather than through AI/ML specific performance metrics like accuracy, sensitivity, or human reader improvement.
Ask a specific question about this device
(249 days)
Hudson RCI® Comfort Flo nasal cannulas are intended to be used in conjunction with the Comfort Flo Humidification System to provide a continuous flow of heated and humidified air/oxygen mixtures to spontaneously breathing patients.
It is indicated for single use by adult and pediatic (2 years and above) patients in professional healthcare environments.
Hudson RCI® Comfort Flo® Plus cannulas are intended to be used in conjunction with the Humidification System to provide a continuous flow of heated and humidified air/oxygen mixtures to spontaneously breathing patients.
It is indicated for single use by adult and pediative (12 years and above) patients in professional healthcare environments.
Hudson RCI® Comfort Flo Nasal Cannulas:
Comfort Flo nasal cannulas are intended to be used in conjunction with the Comfort Flo Humidification System to provide a continuous flow of heated humidified air to spontaneously breathing patients. Sizes available include adult and pediatric.
Hudson RCI® Comfort Flo Plus Cannulas:
Cannula tube is assembled with other components to create a product named Comfort Flo® Plus Cannula with/ without Chin Strap. The Comfort Flo Plus cannulas are intended to be used in conjunction with the Humidification system intended to provide a continuous flow of heated and humidified air/oxygen mixtures to spontaneously breathing patients. It allows for the safe and effective delivery of heated, humidified oxygen therapy to the patients.
The Comfort Flo Plus Cannulas are available in small, medium and large configurations. The chin strap is also offered individually.
Hudson RCI® Comfort Flo Plus Cannulas:
Comfort Flo Soft Plus cannulas are the same as Comfort Flo Plus cannulas except for head strap. These have a soft/cushioned head strap for patient comfort.
It is used as an alternative to oxygen face mask and allows the patient to continue to talk, eat and drink while receiving the therapy.
Hudson RCI® Tracheostomy Adaptor:
The Hudson RCI® Tracheostomy Adapters are intended to be used in conjunction with tracheostomies in order to provide a secure connection to spontaneously breathing tracheostomy patients for oxygen therapy. It allows for the safe and effective delivery of oxygen therapy to the patient's trachea.
The Tracheostomy Adaptor consists of 15mm (female) connector with cap deflector which enables it to be attached to 15M conical connector of tracheostomy tube/ patient connector port and swivel connector (15F/22M) at another end in order to connect with breathing circuit/ humidification system.
It is indicated for single use by adult and pediatric (12 years and above) patients in professional healthcare environments.
The provided text is related to a 510(k) premarket notification for a medical device (nasal cannulas) and does NOT contain information about acceptance criteria or a study proving that the device meets specific performance criteria, especially not in the context of an AI/ML-driven medical device.
The document describes the device, its intended use, comparison to predicate devices, and a summary of non-clinical (bench) performance testing and biocompatibility testing. However, it explicitly states "This section does not apply. No animal testing was performed" and "This section does not apply. No clinical testing was performed" under "Performance Testing (Animal)" and "Performance Testing (Clinical)" respectively.
Therefore, I cannot extract the requested information regarding acceptance criteria, device performance from a test set, sample sizes for test/training sets, data provenance, expert numbers/qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth establishment relevant to an AI/ML device.
The document discusses tests like:
- Post aging Visual Inspection
- Relevant Humidity Output testing
- Thermal Overshoot testing
- Gas path and headgear joint and connection strength testing
- Flow Leak Test/ Gas path leak testing
- Head Gear Testing
- Shelf Life Testing
- Useful Life Testing
- Transportation simulation
- ISO gauging
These are all bench tests for a mechanical/physical device (nasal cannula), not studies that would involve AI/ML performance metrics like sensitivity, specificity, AUC, or the involvement of human readers/experts for ground truth.
In summary, the provided document does not contain the information required to answer your request about acceptance criteria and study details for an AI/ML medical device.
Ask a specific question about this device
Page 1 of 23