Search Filters

Search Results

Found 35 results

510(k) Data Aggregation

    K Number
    K212894

    Validate with FDA (Live)

    Device Name
    PhasTIPP
    Date Cleared
    2022-03-31

    (202 days)

    Product Code
    Regulation Number
    870.4885
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PhasTIPP System is indicated for use in ambulatory phlebectomy procedures for the resection and ablation of varicose veins. The Illuminator is also indicated for use without the Resector for visualization of varicose veins and infusion of tumescent solution during an ambulatory phlebectomy procedure.

    Device Description

    The PhasTIPP System consists of an Illuminator and a Resector. During the surgical procedure, the illuminator shaft is placed through an incision under the skin with its light-emitting end to provide intense light for a better visualization of the varicose veins. A port that runs along the illuminator shaft is used to infuse tumescence solution to cause the contraction of varicosities, which aids vein visualization by creating a subcutaneous fluid pocket through which the illuminator's light can disperse. A powered resector, positioned through a different incision can then be used to morcellate and aspirate the varicosities.

    The PhasTIPP Illuminator consists of two devices: a reusable Illuminator Handpiece and a disposable Illuminator.

    The Handpiece provides illumination controls and contains an LED to provide intense light. The Handpiece is powered with new batteries for each new surgical procedure (Duracell CR 123a cell). The handpiece is provided non-sterile and must be covered with the microbial barrier sheath attached to the Disposable Illuminator in the sterile field before the procedure begins.

    The Disposable Illuminator includes, in addition to the microbial barrier sheath, a distal stainless steel fiber optic light shaft. When connected to a peristaltic pump, the Disposable Illuminator can also infuse tumescence.

    The PhasTIPP Resector consists of two devices: a reusable Resector Handpiece and a Resector Disposable (available in two diameters, 4.5mm and 5.5mm). The Disposable Resector also connects to a peristaltic suction pump to remove the resected varicosities.

    The Resector Handpiece provides controls for the operation of the rotation blades on the Disposable Resector and is powered with a set of new batteries (TLM-1550 HPM cell), for each surgical procedure. The Resector Handpiece is provided non-sterile and must be covered with the microbial barrier sheath attached to the Disposable Resector in the sterile field.

    The Control Unit that was a large equipment in the predicate device is displaced in the subject device. Instead, the Resector Handpiece and the Illuminator Handpiece control the disposable resector and illuminator in the subject device.

    AI/ML Overview

    This document describes the PhasTIPP System, an external vein stripper. The FDA's 510(k) summary provides details about the device, its intended use, and the testing conducted to establish its substantial equivalence to a predicate device.

    1. A table of acceptance criteria and the reported device performance

    The provided document doesn't explicitly state "acceptance criteria" in a tabulated format alongside "reported device performance" in the way one might expect for a clinical study with primary endpoints. Instead, it lists various design validation, system validation, and performance tests that were completed to demonstrate the device meets product performance specifications and for substantial equivalence.

    For the purpose of this request, I will infer the "acceptance criteria" from the types of tests performed and the "reported device performance" as the overall statement that the device meets these specifications.

    Acceptance Criteria (Inferred from Tests)Reported Device Performance (Summary)
    Illuminator Design Validation:The Illuminator's design, including shaft dimensions, flow rate, light output, leak testing, sheath integrity and usability, and tensile/torque strengths of components, met specifications.
    Shaft diameter, length, IV spike to pump, pump tube, pump to hub lengthsMet specifications
    Flow Rate (ml/min)Met specifications
    Light Output (lumens)Met specifications
    Leak TestingPassed
    Sheath Length (in)Met specifications
    Usability Through SheathConfirmed
    Sheath Maintains Microbial BarrierConfirmed
    Tensile Strength: Hub to Shaft (lbf), Disposable (lbf), Pump tube to Connectors, IV Spike to Tubing (lbf)Met specifications
    Torque Strength: Disposable (in-lb)Met specifications
    Kink Test: Infusion TubingPassed
    Illuminator System Validation:The Illuminator's system functions, including basic operation and low battery performance, were verified.
    Basic FunctionVerified
    Low Battery FunctionalityVerified
    Fault Insertion and Power Supply Fuse testingVerified
    Resector System Validation:The Resector's system functions, including voltage regulation, motor RPM, home position, communications, diagnostics, embedded processor, and battery critical function, were verified.
    Voltage RegulationVerified
    Motor RPMVerified
    Home Position and Home Position DriftVerified
    Communications and Diagnostic LoggingVerified
    Embedded Processor Hardware Basic FunctionVerified
    Jam ClearVerified
    Battery Critical FunctionVerified
    Embedded processor/Software loadVerified
    Resector Software Integration Validation:The Resector's software integration for timer, low battery, hardware configuration, motor state machine, oscillation, anti-stall, high current fault, motor time save/retrieve, and RTC were validated.
    Timer functionalityValidated
    Low battery functionalityValidated
    Low level hardware configuration of the micro-controller hardwareValidated
    Motor state machine runningValidated
    Functionality of the oscillation feature and softwareValidated
    Fault states related to the "anti-stall routineValidated
    High motor current fault by simulating excessive motor currentValidated
    Motor time save and retrieve functionValidated
    Real Time Clock (RTC) against the system clockValidated
    Battery critical lockout featureValidated
    Resector Design Validation:The Resector's design, including shaft diameters and lengths, aspiration tube characteristics, sterile sheath integrity, and tensile/torque strengths of components, met specifications.
    Shaft Diameter (4.5mm & 5.5mm)Met specifications
    Shaft Length (4.5mm & 5.5mm)Met specifications
    Aspiration Tube LengthMet specifications
    Aspiration Tube tensile strength to ConnectorMet specifications
    Kink Testing & Leak TestingPassed
    Bond Tubing to Back LidMet specifications
    Aspiration Flow Rate Comparison (4.5 & 5.5)Met specifications
    Motor Functionality DataMet specifications
    Sterile Sheath: Adequate length, Microbial Barrier Testing, Maintains Contamination BarrierConfirmed
    Tensile Strength: Disposable Hub to Main Shaft (4.5mm & 5.5mm), Driven Gear to the resector inner shaft (4.5mm & 5.5mm), Disposable Assembly to HandpieceMet specifications
    Torque Strength: Disposable Hub to Shaft assembly, Driven Gear to the resector inner shaftMet specifications
    Bonded joint between proximal hub and the backlidMet specifications
    Distal Hub Pinned to proximal resectorMet specifications
    Microbial Sheath Barrier Design Validation:The sheath's bubble leak test confirmed its barrier integrity.
    Sheath Bubble Leak TestPassed
    PhasTIPP Illuminator Injection Line Design Validation:The injection line's pressurized leak testing met specifications.
    Pressurized Leak TestingPassed
    PhasTIPP Illuminator Injection Line Shelf Life Validation (2 years Accelerated aging):The injection line maintained performance after accelerated aging.
    Pressurized Leak TestingPassed
    Flow Rate MeasurementMet specifications
    Luer Fitting Tensile Strength testMet specifications
    Overall LengthMet specifications
    PhasTIPP Performance Testing:System performance was confirmed in a 3D in-vitro bench test.
    System performance in 3D in-vitro bench testConfirmed
    PhasTIPP Illuminator Heat Evaluation:The heat produced by the Illuminator was evaluated.
    Evaluation of heat produced by PhasTIPP IlluminatorEvaluated (implies within acceptable limits for a medical device)
    PhasTIPP Testing to IEC 60601 Medical Electrical Equipment:The device met relevant IEC 60601 standards for electromagnetic compatibility.
    Radiated EmissionsCompliant
    Electro-Static DischargeCompliant
    Radiated Field ImmunityCompliant
    Power Frequency Magnetic FieldCompliant
    Biocompatibility (ISO 10993-1:2018):The device was found to be biocompatible for its intended use.
    External Communicating Device, Contact Circulating Blood, limited use (<24h)Demonstrated biocompatible
    Sterilization (EtO gas):The sterilization process was validated according to industry standards.
    Per EN ISO 11135-1:2014 and AAMI TIR28:2016Validated

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not provide specific sample sizes for the individual engineering and performance tests listed. The data provenance is not specified beyond being "product testing" and "verification activities." These are typically conducted in-house by the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This type of information is not applicable to the tests described in this 510(k) summary. The listed tests are primarily engineering and performance validations, not clinical studies requiring expert interpretation of medical images or patient outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Adjudication methods are typically relevant for clinical studies involving human interpretation or subjective assessments. This 510(k) submission focuses on device engineering and performance validation, so "adjudication method" is not applicable here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such MRMC study was mentioned. The device described is an external vein stripper, not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the PhasTIPP System is a physical medical device (an external vein stripper) used in surgery, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the engineering and performance tests described, the "ground truth" or reference standard would be the established engineering specifications, industry standards (e.g., IEC 60601, ISO 10993-1, EN ISO 11135-1), and the performance of the predicate device. These are objective measures or recognized benchmarks. For example:

    • Mechanical properties (tensile strength, torque strength): Established material specifications and engineering design tolerances.
    • Flow rate, light output: Measured against design specifications.
    • Sterilization: Validated against sterility assurance levels defined by standards.
    • Biocompatibility: Assessed against cytotoxicity, sensitization, irritation tests as per ISO 10993.

    8. The sample size for the training set

    The concept of a "training set" is generally applicable to machine learning or AI models. This 510(k) summary does not concern an AI/ML device, so a training set as such is not relevant. The device undergoes design verification and validation testing.

    9. How the ground truth for the training set was established

    As there is no "training set" in the context of this device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190882

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-02-13

    (315 days)

    Product Code
    Regulation Number
    870.3470
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XenoSure Biologic Patch is intended for use as a surgical patch material for cardiac reconstruction and repair, soft tissue deficiency repair and reinforcing the suture line during general surgical procedures.

    Device Description

    The XenoSure consists of one piece of bovine pericardial tissue that has been selected for minimal tissue blemishes. The tissue is treated with a glutaraldehyde process which crosslinks the collagen fibers and minimizes antigenicity. XenoSure patch is liquid chemical sterilized and packaged in a plastic jar containing sterile glutaraldehyde storage solution.

    AI/ML Overview

    The provided text is a 510(k) summary for the XenoSure Biologic Patch, which is a medical device. This document details the pre-clinical testing performed to demonstrate substantial equivalence to a predicate device, specifically to support a labeling change allowing the device to be exposed to temperatures below 0°C (i.e., frozen).

    The request asks for information typically associated with the acceptance criteria and proof of performance for an AI/ML-driven medical device, including aspects like human reader studies, training/test set sample sizes, and ground truth establishment by experts. However, the provided document does not describe an AI/ML device; it is for a physical biologic patch used as surgical material. Therefore, information about AI-specific aspects (like MRMC studies, human-in-the-loop performance, training/test set ground truth for AI, number of experts for ground truth, etc.) is not applicable to this document.

    Instead, the document details laboratory and animal studies to prove the physical and biological characteristics of the XenoSure Biologic Patch are not adversely affected by freezing.

    Here's an analysis of the provided text in the context of the device's performance and acceptance criteria:


    Acceptance Criteria and Study for XenoSure Biologic Patch

    The study described here aims to demonstrate that a frozen XenoSure Biologic Patch is substantially equivalent to the unfrozen predicate XenoSure Biologic Patch. The acceptance criteria are based on mechanical properties and biological performance, ensuring the frozen device retains its intended function and safety.

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance (Frozen Device)
    Longitudinal Tensile≥ 2 MPaMean tensile strength: 12.1 MPa. All patches passed. No statistical difference from unfrozen predicate (12.9 MPa).
    Elongation5% - 50% elongationMean elongation: 22.3%. All patches passed. No statistical difference from unfrozen predicate (21.8%).
    Burst Strength≥ 12 PSIMean burst strength: 113 PSI. All patches passed. Statistically different from unfrozen predicate (134 PSI), but the difference is not clinically significant as both are much higher than the 12 PSI clinical specification.
    Suture Retention≥ 300 gfMean suture retention: 1235 gf. All patches passed. No statistical difference from unfrozen predicate (1233 gf).
    Cross LinkingNo acceptance criteria stated for this test, but comparison to predicate is implicit for substantial equivalence.28 ppm free amine site per gram. No statistical difference from unfrozen predicate (29 ppm free amine site per gram).
    Collagenase DigestionNo acceptance criteria stated for this test, but comparison to predicate is implicit for substantial equivalence.0.18%. No statistical difference from unfrozen predicate (0.16%).
    Water Permeability< 0.1 ml/cm² minAll samples for both predicate and proposed (frozen) devices recorded zero water permeability. No statistical difference.
    Delamination (Visual Inspection)No delaminationNo delamination identified for both predicate and proposed (frozen) devices.
    In-vivo Animal Study (Safety & Comparability)Favorable local tissue responses (e.g., endothelialization, tissue integration, absence of adverse effects) comparable to control device; no effect on vessel patency; bioequivalence.Six of seven animals survived the full 90-day duration. One animal's early euthanasia was unrelated to the patches. All carotid patches achieved appropriate sealing (no leaks). Both test (frozen) and control (unfrozen) patches did not affect vessel patency based on angiography and ultrasound. Histologically, the frozen patches showed favorable local tissue responses comparable to the control, with no significant differences in histological parameters. Displayed bioequivalence and comparable characteristics in vascular safety, local tissue response, endothelialization, and tissue integration.

    2. Sample Sizes and Data Provenance:

    • Test Set (Physical Properties):

      • Specific numbers of samples tested for each physical property (e.g., tensile strength, burst strength) are not explicitly stated in the summary, but the results indicate that "All XenoSure patches" or "All samples for both predicate and proposed devices" were tested and passed, implying a sufficient sample size was used for each test to demonstrate statistical equivalence or compliance with criteria.
      • Data Provenance: The tests are laboratory-based, performed on the manufactured device. The document does not specify the country of origin of the data, but it is for a US FDA submission, implying adherence to relevant standards for medical device testing. The studies are pre-clinical performance testing.
    • Test Set (In-vivo Animal Study):

      • Sample Size: Seven Polypay sheep (6 survived to the scheduled time point). Each animal received both a frozen (test) and unfrozen (control) patch, allowing for within-subject comparison.
      • Data Provenance: The study was conducted in animals, a pre-clinical, prospective study design. The country of origin of the animal study isn't specified, but it was performed to support a US FDA submission.

    3. Number of Experts and Qualifications for Ground Truth:

    • Not applicable for an AI/ML device. For this physical device, "ground truth" is established through standardized laboratory testing methods (e.g., Instron for mechanical properties, spectrophotometry for chemical properties) and histological analysis by qualified veterinary pathologists in the animal study. The document does not specify the number or qualifications of individuals interpreting the histology, but it is standard for such studies to involve board-certified pathologists.

    4. Adjudication Method:

    • Not applicable for an AI/ML device. For the physical testing, the results are quantitative measurements or visual inspections against defined criteria. For the animal study, histological evaluation would typically follow a standardized protocol, often with consensus or independent review by pathologists, but no formal "adjudication method" in the sense of reader disagreements for an AI study is described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • Not applicable. This device is a physical patch, not an AI system. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was performed.

    6. Standalone Performance (Algorithm Only):

    • Not applicable. This pertains to an AI algorithm. The device is a physical biologic patch. The "standalone performance" of the device is assessed through the physical and biological tests described in the table above.

    7. Type of Ground Truth Used:

    • Laboratory Measurements: For mechanical and chemical properties, the "ground truth" is derived from direct quantitative measurements using validated test methods (e.g., Instron for force, UV-Vis Spectrophotometry for chemical levels).
    • Visual Inspection: For delamination, ground truth is established by direct visual inspection.
    • In-vivo Histopathology and Clinical Outcomes: For the animal study, the "ground truth" for safety and biological integration is based on gross observations, angiographic and ultrasound analysis of vessel patency, and detailed histopathologic evaluation of explanted tissues by qualified personnel.

    8. Sample Size for Training Set:

    • Not applicable. There is no "training set" as this is not an AI/ML device.

    9. How Ground Truth for Training Set was Established:

    • Not applicable. There is no "training set" for this physical medical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190267

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-10-30

    (264 days)

    Product Code
    Regulation Number
    870.4885
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EZE SIT Valvulotome is intended to render venous valves incompetent during in situ bypass procedures. This includes distal infrainguinal bypass when a non-anatomic position is required (e.g., profunda to anterior tibial artery), composite vein infrainguinal bypass, or aorto-renal bypass procedures.

    Device Description

    The EZE SIT Valvulotome is a medical device utilized for disruption of venous valves, and is used in patients that suffer from peripheral vascular disease, for which an attending physician determines the relative benefit of in situ bypass to restore blood flow to the extremities. The EZE-SIT Valvulotome is comprised of a catheter with a luer-lock hub at the proximal end and a threaded connector at the distal end. The catheter is designed with a lumen that extends throughout its entire length. The threaded connector allows attachment of 2 mm, 3 mm and 4 mm diameter cutting heads. These cutting heads are also configured with a central lumen. This design allows the physician to select the cutter head diameter that is best suited for the particular vessel being prepared. The central lumen allows irrigation during the procedure. The cutting head design minimizes vessel wall contact while effectively disrupting valve leaflet tissue. The EZE SIT Valvulotome is surgically invasive and intended for less than 24-hour use. This medical device has no software components. This device is not an in vitro diagnostic device.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the EZE SIT Valvulotome. It asserts substantial equivalence to a predicate device (Tru-Incise Valvulotome K930011).

    Crucially, this document states that no performance data, including clinical studies or data demonstrating adherence to acceptance criteria for an AI/algorithm-driven device, were required or provided. The submission is intended "to simply indicate a new manufacturer for a currently cleared device design that has not changed since its clearance (K930011)."

    Therefore, the acceptance criteria and study proving the device meets those criteria, as requested in the prompt, are not present in this document. This device does not have software components and is not an in vitro diagnostic device, and thus the type of AI/algorithm performance study envisioned by the prompt is not applicable to this submission.

    The document only includes:

    • Nonclinical data: Sterilization, Shelf Life/Aging, and Biocompatibility, all of which confirmed substantial equivalence to the predicate device.
    • Sterilization: Validated with half-cycle using ISO 11135-1: 2014, achieving a Sterility Assurance Level (SAL) of 10-6. Residuals measured: 0.6 mg EtO, <0.1 mg Ethylene Chlorohydrin (ECH). Endotoxin measured by LAL method.
    • Shelf Life/Aging: Extended shelf life to 6 years (from 2.5 years for the predicate) based on accelerated aging.
    • Biocompatibility: Passed all tests in the biocompatibility battery, recapitulating tests performed for the predicate device.

    In summary, none of the requested information regarding AI/algorithm performance (acceptance criteria table, sample sizes, expert ground truth, MRMC study, standalone performance, training data, etc.) can be extracted from this document, as it pertains to a non-AI/non-software medical device that underwent a 510(k) submission based on substantial equivalence without requiring new performance data for design changes.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183513

    Validate with FDA (Live)

    Date Cleared
    2019-06-13

    (177 days)

    Product Code
    Regulation Number
    882.5910
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DuraSure Biologic Patch is intended for use as a surgical patch material to close dura mater during neurosurgery.

    Device Description

    The DuraSure consists of one piece of bovine pericardial tissue that has been selected for minimal tissue blemishes. The tissue is treated with a glutaraldehyde process which crosslinks the collagen fibers and minimizes antigenicity. The DuraSure is liquid chemical sterilized and packaged in a plastic jar containing sterile glutaraldehyde storage solution. The DuraSure is designed to repair the body's natural organs.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the DuraSure Biologic Patch, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaDuraSure Biologic Patch PerformancePredicate Device Performance
    Tensile Strength≥ 2 MPaMean: 11.9 MPaNot explicitly stated (implied to meet acceptance criteria, but specific value for predicate device not given, only "mean of tensile strength of the predicate device was [value not provided]" in text)
    Elongation5% - 50% elongationMean: 25% elongationMean: 32% elongation
    Burst Strength≥ 12 PSIMean: 127 PSIMean: 59 PSI
    Suture Retention≥ 300 gfMean: 970 gfMean: 978 gf
    Thickness0.35 mm - 0.75 mmPassed acceptance criteria0.32 mm - 0.71 mm
    BiocompatibilitySatisfactory biocompatibility (implied, by comparison to predicate)Satisfactory biocompatibility resultsEstablished biocompatibility
    SterilizationChemically sterilized according to ISO14160: 2011 with 10⁻⁶ SALChemically sterilized according to ISO14160: 2011 with 10⁻⁶ SALChemically sterilized with 10⁻⁶ SAL
    Animal Study ConclusionTest article is locally non-toxic and performs equivalently to control in dural repair model.Test article is locally non-toxic; performed equivalently to control.Control article deemed locally non-toxic and performed equivalently to test article.

    2. Sample Size Used for the Test Set and Data Provenance

    The document focuses on "pre-clinical" testing, which includes both in-vitro (bench) testing and an in-vivo animal study.

    • Bench Testing (Tensile, Elongation, Burst Strength, Suture Retention, Thickness):

      • Sample Size: Not explicitly stated for each test (e.g., "All DuraSure patches passed the acceptance criteria"). However, results are given as "mean" values, implying multiple samples were tested for each characteristic.
      • Data Provenance: Not specified, but generally refers to laboratory testing conducted by the manufacturer or contracted labs. The country of origin is not mentioned. This is retrospective data collected for the 510(k) submission.
    • In-vivo Animal Study:

      • Sample Size: Forty-nine (49) rabbits.
      • Data Provenance: Not explicitly stated, but implies a controlled laboratory setting for animal research. This is prospective data collected during the study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Bench Testing: No experts are typically used to establish ground truth for this type of quantitative functional testing. The "ground truth" is defined by the physical measurement methods and the device specifications.
    • In-vivo Animal Study:
      • Number of Experts: At least one. The document states, "All slides were evaluated by a veterinary pathologist for neuropathological changes in brain tissue, dural integrity, neoduralization and local tissue reaction according to ISO 10993-6 and FDA Guidance."
      • Qualifications: "veterinary pathologist." No further details on years of experience or specific sub-specialties are provided.

    4. Adjudication Method for the Test Set

    • Bench Testing: Not applicable. These are objective measurements following established test methods.
    • In-vivo Animal Study: Not explicitly described. It states "All slides were evaluated by a veterinary pathologist." This implies a single expert assessment. If multiple pathologists were involved, no method for resolving disagreements (e.g., 2+1, 3+1) is mentioned.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The studies presented are pre-clinical (bench and animal studies) and do not involve human readers evaluating cases.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    No, this question is not applicable. The DuraSure Biologic Patch is a physical medical device (surgical patch material), not an AI algorithm or software device. Therefore, a standalone performance evaluation of an algorithm is not relevant.

    7. The Type of Ground Truth Used

    • Bench Testing: The ground truth is based on objective quantitative measurements obtained using standardized test methods (e.g., Instron for tensile/elongation/suture retention, pressure sensors for burst strength, thickness gauges).
    • In-vivo Animal Study: The ground truth for the animal study was established through histopathological evaluation by a veterinary pathologist, based on internationally recognized standards (ISO 10993-6) and FDA guidance for dural substitute devices. This includes macroscopic and microscopic evaluations, with scoring.

    8. The Sample Size for the Training Set

    This question is not applicable. The DuraSure Biologic Patch is a physical medical device, not a machine learning model. Therefore, there is no "training set" in the context of AI.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182916

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-11-16

    (29 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pruitt F3 Carotid Shunt is indicated for use in carotid endarterectomy as a temporary conduit to allow for blood flow between the common and internal carotid arteries.

    Device Description

    The Pruitt F3 Carotid Shunt is used in patients with carotid artery disease in instances where removal of vascular occlusions is the therapeutic objective. The Pruitt F3 Carotid Shunt is employed in Carotid Endarterectomy (CEA), where the shunt is used as a short-term conduit for blood flow between the common carotid and the internal carotid arteries. The Pruitt F3 Shunt is provided sterile for single use, and is provided with two 3 ml syringes used to inflate the balloons (accessory syringes have been previously cleared in K051067).

    The Pruitt F3 Carotid Shunt is designed to serve as an artificial passage connecting two blood vessels, allowing blood flow from one vessel to another. This is accomplished by using a clear, plastic, sterile conduit that is held in place by a stabilization technique (balloons) on both ends of the conduit.

    The Pruitt F3 Shunts are multi-lumen devices with balloons on both the distal (internal carotid) and proximal (common carotid) ends of the shunt. The balloons, when inflated independently, act as a stabilization mechanism to maintain the position of the shunt when it is placed within the common and internal carotid arteries.

    The Pruitt F3 Carotid Shunt has features to aid the user during shunt insertion and balloon inflation. The inflation path of the proximal (common carotid) balloon is color-coded. Sterile saline is injected from the blue stopcock, through the blue lumen and into the blue common carotid balloon.

    AI/ML Overview

    The provided K182916 document for the Pruitt F3 Carotid Shunt is a 510(k) premarket notification. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials as would be required for a PMA.

    Therefore, the document does not contain the information requested regarding acceptance criteria, device performance, sample sizes for test/training sets, data provenance, expert ground truth establishment, adjudication methods, or MRMC studies.

    The document primarily describes bench test data for engineering parameters, not clinical performance metrics tied to specific acceptance criteria.

    Here's what can be extracted, and what is explicitly not available:

    1. A table of acceptance criteria and the reported device performance

    • Not Available directly in the provided text. The document lists "Bench Test Data" items like "Main Lumen Flow Test," "Inflated Common Balloon Diameter," "Deflation Time," etc. However, it does not provide the specific acceptance criteria (e.g., "Deflation Time shall be < X seconds") nor the reported device performance values against these criteria. It only states that these tests were performed to support substantial equivalence.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Available. The document refers to "engineering studies" and "bench test data" but does not specify sample sizes for these tests or the provenance of any data (e.g., if any clinical data was indirectly used, which it doesn't appear to be for performance metrics).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable / Not Available. This device is a mechanical shunt, and its performance is evaluated through engineering bench tests, not clinical image interpretation or diagnosis that would require expert ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable / Not Available. Similar to point 3, adjudication methods are relevant for subjective assessments, typically in clinical studies involving human interpretation. Bench tests for mechanical devices do not use such methods.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not an AI-assisted diagnostic device. It is a physical medical device (carotid shunt). MRMC studies are not relevant for this product.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable. This is not an algorithm-based device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable. For the bench tests performed, "ground truth" would be established by direct measurement against engineering specifications or known physical properties, not clinical outcomes or expert consensus.

    8. The sample size for the training set

    • Not Applicable. There is no "training set" as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set, this question is not relevant.

    In summary, the provided K182916 document focuses on demonstrating substantial equivalence through a comparison of technological characteristics and bench test data for a physical medical device. It does not provide the type of detailed performance criteria and study results that would be expected for an AI/CADe device or a device requiring extensive clinical outcome data for a PMA.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173323

    Validate with FDA (Live)

    Date Cleared
    2018-03-08

    (139 days)

    Product Code
    Regulation Number
    878.4300
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AnastoClip GC is intended for use in the creation of everting anastomoses in blood vessels and other small tubular structures when tissue penetration is desired. The Applier is also intended for approximation of durotomies following open craniotomy and open spinal laminectomy procedures.

    Device Description

    AnastoClip GC Closure System is designed to create everting anastomoses of tissue. It includes the applier, clip remover, and atraumatic forceps. The AnastoClip GC applier consists of a rotating shaft and an integral cartridge containing titanium clips. As the levers of the applier are squeezed together, the clip is closed around the everted tissue edges. As the levers are released, a new clip is automatically loaded into the clip applier jaws. It is recommended, with each procedure, to use the atraumatic everting forceps to aid in the everting of the tissue edges. It is also recommended, with each procedure, to use the AnastoClip Remover for the removal of any AnastoClip GC clips when needed.

    AI/ML Overview

    The provided text describes the AnastoClip GC Closure System, a medical device. Based on the information presented, here's a breakdown of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative "acceptance criteria" in a table format with corresponding "reported device performance" values for a specific study. Instead, it describes a preclinical study that aimed to demonstrate safety and substantial equivalence to a predicate device for an expanded indication. The "acceptance criteria" can be inferred from the successful outcomes of this study.

    Here's an inferred table based on the preclinical study's reported findings:

    Acceptance Criteria (Inferred from successful study outcomes)Reported Device Performance (AnastoClip GC)
    Absence of CSF leaks at cranial and spinal dural sitesNo CSF leaks observed at both cranial and spinal dural sites after application.
    No clinically significant changes in body weight, body condition score, or clinical pathology parametersNo clinically significant changes in body weight, body condition score, or clinical pathology parameters in any animal over the course of the study.
    Absence of adverse responses (inflammation, infection) in CSF samplesNo CSF sample exhibited any indication of adverse responses such as inflammation or infection.
    Absence of hydrocephalus or other abnormalities via CT scanNo signs of hydrocephalus or other abnormalities noted from the CT scan evaluations.
    Comparable histopathologic findings (minimal fibrosis, vascularization, hemorrhage, no necrosis, infection, or foreign body reaction) compared to predicateHistologic findings were comparable to the predicate device (AnastoClip AC), with minimal fibrosis and vascularization, minimal hemorrhage, and no implant-associated necrosis, infection or foreign body reaction.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: 6 porcine models.
    • Data Provenance: The study was a "Pre-clinical GLP safety evaluation" performed in porcine models. The country of origin is not specified, but it's a prospective animal study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not specify the number of experts or their qualifications for establishing ground truth. It states that a "comprehensive necropsy was performed" and "histomorphologic evaluation" of cranial and spinal implant sites was conducted. These evaluations would typically be done by veterinary pathologists, but their number and specific qualifications are not detailed.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not provide details on any adjudication method for the test set. It simply states that a "comprehensive necropsy" and "histomorphologic evaluation" were performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document describes a preclinical animal study for a medical device (surgical clip system), not an AI-powered diagnostic tool. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance was not performed and is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the device is a physical surgical closure system, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the preclinical study was established through a combination of:

    • Direct observation: Absence of CSF leaks.
    • Clinical observation: Body weight, body condition score, clinical pathology parameters in animals.
    • Laboratory analysis: CSF sample analysis (for inflammation/infection).
    • Imaging: CT scan evaluations (for hydrocephalus/abnormalities).
    • Pathology/Histomorphology: Microscopic evaluation of cranial and spinal implant sites.

    8. The sample size for the training set

    This question is not applicable as the device is a physical medical device and does not involve AI or machine learning that requires a training set.

    9. How the ground truth for the training set was established

    This question is not applicable as the device is a physical medical device and does not involve AI or machine learning that requires a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152833

    Validate with FDA (Live)

    Date Cleared
    2016-01-21

    (114 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pruitt F3-S Polyurethane Carotid Shunt is in carotid endarterectomy as a temporary conduit to allow for blood flow between the common and internal carotid arteries.

    Device Description

    The Pruitt F3-S Polyurethane Carotid Shunt is designed to serve as an artificial passage connecting two blood vessels allowing blood flow from one vessel to another. This is accomplished by using a clear, plastic, sterile conduit that is held in place by a stabilization technique (balloons) on both ends of the conduit. devices with balloons on both the distal (internal carotid) and proximal (common carotid) ends of the shunt. The balloons, when inflated independently, act as a stabilization mechanism to maintain the position of the shunt when it is placed within the common and internal carotid arteries. The Pruitt F3-S Polyurethane Carotid Shunt has features to aid the user during shunt insertion and balloon inflation. The inflation path of the proximal (common carotid) balloon is color-coded. Sterile saline is injected from the blue stopcock, through the blue lumen and into the blue common carotid balloon. The Pruitt F3-S Polyurethane Carotid Shunt is not made of natural rubber latex.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device (Pruitt F3-S Polyurethane Carotid Shunt) and does not contain information about the performance of an AI/ML device. Therefore, the requested information regarding acceptance criteria and study details for an AI-powered device cannot be extracted.

    The document focuses on demonstrating substantial equivalence to a predicate device, as required for 510(k) submissions. This involves comparing the new device's technological characteristics, indications for use, and performance data (e.g., sterilization, biocompatibility, functional testing) to those of the predicate device. The functional testing described is for the physical device itself (shunts and balloons), not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143454

    Validate with FDA (Live)

    Date Cleared
    2015-08-27

    (267 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pruitt F3-S Carotid Shunt is indicated for use in carotid endarterectomy as a temporary conduit to allow for blood flow between the common and internal carotid arteries.

    Device Description

    The Pruitt F3-S Carotid Shunt is designed to serve as an artificial passage connecting two blood vessels, allowing blood flow from one vessel to another. This is accomplished by using a clear, plastic, sterile conduit that is held in place by a stabilization technique (balloons) on both ends of the conduit. The Pruitt F3-S Shunts are multi-lumen devices with balloons on both the distal (internal carotid) and proximal (common carotid) ends of the shunt. The balloons, when inflated independently, act as a stabilization mechanism to maintain the position of the shunt when it is placed within the common and internal carotid arteries. The Pruitt F3-S Carotid Shunt has features to aid the user during shunt insertion and balloon inflation. The inflation path of the proximal (common carotid) balloon is color-coded. Sterile saline is injected from the blue stopcock, through the blue lumen and into the blue common carotid balloon.

    AI/ML Overview

    The provided document describes the LeMaitre Vascular Pruitt F3-S Carotid Shunt (K143454) and its substantial equivalence to a predicate device. This is a medical device submission, not a study of an AI algorithm, so many of the requested categories (e.g., sample size for test set, number of experts for ground truth, MRMC study, training set sample size) are not applicable or do not directly translate. However, I can extract information related to the device's performance testing and acceptance criteria as described in the summary.

    Acceptance Criteria and Device Performance for Pruitt F3-S Carotid Shunt

    The document focuses on demonstrating substantial equivalence to a predicate device (Pruitt F3 Carotid Shunt K051067) by comparing technological characteristics and performing functional/safety testing. It does not provide explicit numerical acceptance criteria values or specific performance results for each test in a table format, but rather states that the device "meets the product performance requirements of the device specifications."

    However, based on the "Summary of Product Testing," we can infer the aspects of performance that were evaluated.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (Inferred from "Meets product performance requirements of the device specifications")Reported Device Performance
    Shunt lumen diameterDevice demonstrates appropriate lumen diameter for its intended function of allowing blood flow between carotid arteries, as per design specifications."meets the product performance requirements of the device specifications"
    Balloon diameterBalloons inflate to the specified diameters to effectively stabilize the shunt within the common and internal carotid arteries, as per design specifications and IFU data."meets the product performance requirements of the device specifications"
    Balloon deflation timeBalloons deflate within an acceptable timeframe to allow for proper removal or adjustment, as per design specifications."meets the product performance requirements of the device specifications"
    Balloon radial forceBalloons exert sufficient radial force to anchor the shunt securely without causing undue trauma, as per design specifications."meets the product performance requirements of the device specifications"
    Kink testThe shunt resists kinking under normal physiological and surgical usage conditions to maintain patency, as per design specifications."meets the product performance requirements of the device specifications"
    Balloon inflation curveThe relationship between inflation volume and balloon diameter is consistent with specifications, guiding users for proper inflation."meets the product performance requirements of the device specifications"
    SterilizationDevice is sterile as defined by ANSI/AAMI/ISO 11135-1:2007."is ethylene oxide (EO) sterilized according to ANSI/AAMI/ISO 11135-1:2007"
    BiocompatibilityMaterials are biocompatible and do not elicit adverse biological responses."The materials used... are identical to those in the predicate device which has established biocompatibility."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a distinct "test set" for a statistical study in the way an AI algorithm might. The testing described is pre-clinical functional and safety testing on the device itself.

    • Sample Size: Not explicitly stated for each test, but typically these tests involve a representative sample of devices.
    • Data Provenance: The tests are performed by the manufacturer, LeMaitre Vascular, Inc., likely at their facilities in Burlington, MA, USA, as part of their internal verification process. The data is prospective, generated specifically for this submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This concept is not directly applicable. The "ground truth" for a medical device's functional integrity is established by engineering specifications, recognized industry standards (like ISO 11135-1 for sterilization), and comparison to predicate devices, rather than expert consensus on a diagnostic outcome. The "experts" involved would be the engineers and quality assurance personnel at LeMaitre Vascular.

    4. Adjudication Method for the Test Set

    Not applicable. This type of testing relies on objective measurements against predefined engineering specifications, not subjective interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, an MRMC study was not done. This type of study is relevant for diagnostic devices where human readers interpret patient data (e.g., images) with and without AI assistance. The Pruitt F3-S Carotid Shunt is a physical medical device, not a diagnostic AI algorithm.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable, as this is not an AI algorithm.

    7. The Type of Ground Truth Used

    The ground truth for this device's performance is based on engineering specifications, design requirements, and established performance characteristics of the predicate device. For biocompatibility, it's the known biocompatibility profile of the materials used in the predicate device. For sterilization, it's compliance with an international standard (ANSI/AAMI/ISO 11135-1:2007).

    8. The Sample Size for the Training Set

    Not applicable, as this is not an AI algorithm requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as this is not an AI algorithm requiring a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142660

    Validate with FDA (Live)

    Date Cleared
    2014-10-17

    (29 days)

    Product Code
    Regulation Number
    870.4885
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LeMills Valvulotomes are intended to cut venous valves.

    Device Description

    The Antegrade LeMills Valvulotome consists of small metal antegrade cutting blade with atraumatic distal edge. The blade is a part of a long stainless steel wire that allows it to be inserted into the venous anatomy. It is held by a plastic handle. It is designed for cutting the venous valves. Once the valves have been rendered ineffectual, the vein can then be utilized as an arterial conduit.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Antegrade LeMills Valvulotome, a medical device. This type of document focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than providing extensive details on novel performance studies and acceptance criteria as would typically be found for AI/ML-based devices or those requiring clinical trials for efficacy.

    Based on the content provided, it is not possible to extract the information requested regarding acceptance criteria and a study proving the device meets them in the context of AI/ML performance, multi-reader multi-case studies, or complex ground truth establishment. This is because the device is a physical surgical tool (a valvulotome) designed to cut venous valves, and its approval relies on bench testing and comparison to existing predicate devices, not on diagnostic performance or AI algorithm validation.

    However, I can extract the available information related to performance testing that was conducted:

    1. A table of acceptance criteria and the reported device performance:

    The document mentions "Functional/Safety Testing" and "Summary of Product Testing" but does not provide a specific table with acceptance criteria and corresponding reported device performance values. It generally states that the device meets performance requirements.

    Acceptance CriteriaReported Device Performance
    Not explicitly defined in the document for specific numerical thresholds.The verification activities conducted indicate that Antegrade LeMills Valvulotome device meets the product performance requirements of the device specifications and does not raise any additional safety issues.
    Tests listed:
    Dimensional Comparison(Implied to meet specifications)
    Sharpness Test (Effectiveness Test)(Implied to be effective for cutting venous valves)
    Visual Inspection(Implied to meet quality standards)
    Joint Tensile Strength (Pull Test)(Implied to meet required strength)

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not specified in the document. The tests appear to be bench-top physical tests performed on the device itself, rather than studies involving patient data.
    • Data Provenance: Not applicable as it's not a data-driven diagnostic or AI device. The tests are for the physical properties and function of the surgical tool.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts/Qualifications: Not applicable. Ground truth for physical device testing is typically established through engineering specifications, material standards, and functional performance benchmarks, not expert consensus on medical images or patient outcomes.

    4. Adjudication method for the test set:

    • Adjudication Method: Not applicable. This concept is relevant for interpreting ambiguous clinical data, not for physical device testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No. This device is a surgical tool, not a diagnostic imaging device or an AI-assisted diagnostic tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Standalone Performance: Not applicable. There is no algorithm or AI component in this device.

    7. The type of ground truth used:

    • Ground Truth Type: For the physical tests mentioned (Dimensional Comparison, Sharpness Test, Visual Inspection, Joint Tensile Strength), the 'ground truth' would be the engineering specifications and design requirements for the device. For biocompatibility and sterilization, the ground truth is adherence to recognized standards (ISO 10993, ANSI/AAMI/ISO 11135-1:2007).

    8. The sample size for the training set:

    • Training Set Sample Size: Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    • Ground Truth Establishment for Training Set: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140042

    Validate with FDA (Live)

    Date Cleared
    2014-04-10

    (92 days)

    Product Code
    Regulation Number
    870.4885
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is used for the treatment of vascular disorders, and more particularly for excising or disrupting venous valves.

    Device Description

    The 1.5mm HYDRO Expandable LeMaitre Valvulotome is a self-centering and self-sizing valvulotome device used for cutting vein valves. The centering hoops keep the device centered in the vein. The size of the centering hoops and cutting blades adjust to the internal diameter of the vein as the valvulotome is being drawn through the vessel cutting the valves and rendering them ineffectual. The modifications to the predicate device - Expandable LeMaitre Valvulotome (ELV)- include the addition of hydrophilic coatings and co-extruded sheath to create 1.5mm HYDRO Expandable LeMaitre Valvulotome. The hydrophilic coatings provide increased lubricity and allow the device to navigating veins easier.

    AI/ML Overview

    The provided document describes a 510(k) submission for the 1.5mm HYDRO Expandable LeMaitre Valvulotome. This is a medical device, and the submission focuses on demonstrating "substantial equivalence" to a predicate device, rather than proving performance against specific acceptance criteria in the way a new diagnostic or AI-driven device often does.

    Therefore, the structure of the response will be tailored to the information provided in a 510(k) summary for a physical medical device. Many of the requested categories (like sample size for test sets, number of experts for ground truth, MRMC studies, standalone performance of an algorithm, training set details) are not applicable to this type of submission because it's not a diagnostic or AI device.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this device, the "acceptance criteria" are implied by meeting product performance specifications and not raising additional safety issues, as compared to its predicate device. Specific quantitative acceptance criteria with corresponding performance values are generally not provided in this type of 510(k) summary for functional device modifications. The tests performed are primarily engineering and bench tests, with a cadaver study, to show that the modified device functions similarly and safely.

    Acceptance Criterion (Implied)Reported Device Performance
    Meet product performance specifications (general)"meets the product performance specifications"
    Modifications do not raise any additional safety issues"modifications do not raise any additional safety issues"
    Achieve intended use (excising/disrupting venous valves)Demonstrated through Cadaver Study, functional testing.
    Biocompatibility for blood contact"is biocompatible"
    Sterilization validated"is validated for ethylene oxide (EO) sterilization"
    Maintain fundamental scientific technology as predicate"maintains the same intended use and fundamental scientific technology as the predicate device(s)"
    Dimensional accuracy"Dimensional Verification" completed
    Durability/Reliability (e.g., fatigue)"Fatigue Testing" completed
    Fluid dynamics/Lubricity (hydrophilic coating benefit)"Flushability" completed
    Ease of use (insertion/removal)"Force to insert and remove device" completed
    Mechanism function (opening/closing)"Force to open and close device" completed
    Navigability in tortuous anatomy"Tortuous Sheathing" completed
    Functionality in a biological environment (cutting valves)"Cadaver Study" completed

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: Not explicitly stated for specific tests (e.g., how many devices for dimensional, fatigue, force tests). For the Cadaver Study, the sample size is not specified.
    • Data Provenance: Not specified. Standard practice for bench testing is typically done in a lab environment. Cadaver studies are usually conducted by the manufacturer or a contract research organization (CRO) in a controlled setting. No country of origin for the data is mentioned. The study is prospective for the device's performance given the modifications.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This concept is not applicable here as this is a physical device, not an AI or diagnostic device that requires expert-established ground truth on a dataset. The "ground truth" for a medical device's performance is typically established through engineering standards, functional tests, and in vivo/cadaver studies where success is defined by the device performing its intended mechanical function.
    • For the cadaver study, it's implied that medical professionals (surgeons or similar) would have conducted or observed the study to assess functionality, but their number and specific qualifications are not detailed.

    4. Adjudication method for the test set

    • Not applicable in the context of an AI/diagnostic device's performance evaluation. For engineering tests, results are typically assessed against pre-defined engineering specifications. For the Cadaver Study, the "adjudication" would be whether the device successfully performed its function (cutting valves) and navigated the anatomy, likely assessed by the operating physician(s).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a medical instrument, not an AI system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a medical instrument, not an AI algorithm.

    7. The type of ground truth used

    • The "ground truth" for this device is based on functional performance metrics derived from engineering tests (e.g., force measurements, dimension checks, fatigue cycles) and observations from the cadaver study confirming the device's ability to navigate and effectively cut venous valves in a simulated biological environment. There is also biocompatibility testing against ISO 10993 guidelines.

    8. The sample size for the training set

    • Not applicable. This device does not use a "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 4