Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K223375

    Validate with FDA (Live)

    Date Cleared
    2023-03-15

    (131 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Low Dermatitis Potential. Tested for Use with Chemotherapy Drugs and Fentanyl Citrate.

    AI/ML Overview

    This document describes the acceptance criteria and performance study for the Powder Free Nitrile Patient Examination Glove.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device, particularly for its resistance to chemotherapy drugs and Fentanyl Citrate, are defined by the "Minimum Breakthrough Detection Time in Minutes" as specified in the ASTM D6978-05 (Reapproved 2019) standard. The reported device performance is indicated by the measured breakthrough times against various substances.

    Chemotherapy Drug and ConcentrationAcceptance Criteria (Minimum Breakthrough Detection Time in Minutes)Reported Device Performance (Minimum Breakthrough Detection Time in Minutes)
    Bendamustine HCI (Treanda) (5 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05. A "pass" would typically be >240 minutes for standard chemotherapy glove claims.>240
    Bleomycin Sulfate (15 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Busulfan (6 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Carboplatin (10 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Carfilzomib (2 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Carmustine (BCNU) (3.3 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.26.2
    Cetuximab (Erbitux) (2 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Cisplatin (1 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Cladribine (1 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Cyclosporin A (100 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Cyclophosphamide (Cytoxan) (20 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Cytarabine (100 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Cytovene (Ganciclovir) (10 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Dacarbazine (DTIC) (10 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Daunorubicin HCI (5 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Decitabine (5 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Docetaxel (Taxotere) (20 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Doxorubicin Hydrochloride (2 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Epirubicin HCI (Ellence) (2 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Etoposide (20 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Fludarabine (25 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Fluorouracil (50 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Gemcitabine (38 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Idarubicin HCI (1 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Ifosfamide (50 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Irinotecan (20 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Mechlorethamine HCI (1 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Melphalan (5 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Methotrexate (25 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Mitomycin C (0.5 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Mitoxantrone (2 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Oxaliplatin (5 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Paclitaxel (Taxol) (6 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Pemetrexed (25 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Raltitrexed (0.5 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Retrovir (Zidovudine) (10 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Rituximab (10 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Thiotepa (10 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.59.1
    Topotecan (1 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Trisenox (Arsenic Trioxide) (1 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Velcade (Bortezomib) (1 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Vidaza (Azacytidine) (25 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Vinblastine (1 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Vincristine Sulfate (1 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Vinorelbine (10 mg/ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Zoledronic Acid (1 mg/25ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Fentanyl Citrate Injection (100 mcg/2ml)Not explicitly stated as a minimum, but tested to ASTM D6978-05.>240
    Simulated Gastric Acid FluidNot explicitly stated as a minimum, but tested to ASTM D6978-05.>240

    Note: For Carmustine (BCNU) and Thiotepa, the reported times (26.2 and 59.1 minutes respectively) are below the ">240" observed for most other drugs. This results in explicit warnings not to use with Carmustine (BCNU) and caution when using with Thiotepa due to low permeation times.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size (number of gloves or individual tests) used for the permeation test for each chemical. It only provides the resulting "Minimum Breakthrough Detection Time in Minutes."

    The data provenance is not specified in terms of country of origin. The study appears to be a prospective test specifically designed to evaluate the performance of these gloves against the listed chemicals, rather than a retrospective analysis of existing data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of study (chemical permeation test as per ASTM D6978-05) does not involve human experts establishing "ground truth" in the way that image interpretation or diagnostic studies do. The "ground truth" is the objective measurement of chemical breakthrough time, determined in a laboratory setting according to a standardized protocol (ASTM D6978-05). Therefore, there are no medical experts (e.g., radiologists) involved in establishing this particular ground truth.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, this is a laboratory-based chemical permeation test, not a subjective assessment requiring adjudication among experts. The results are objective measurements based on the ASTM D6978-05 standard.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is not a study involving human readers or AI. It is a chemical resistance test for a medical device (gloves).

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is not an algorithm or AI study. It is a physical performance test of a medical device.

    7. The Type of Ground Truth Used

    The ground truth used is objective laboratory measurement based on a standardized protocol, specifically "Assessment of Medical Gloves to Permeation by Chemotherapy Drugs" as per ASTM D6978-05 (Reapproved 2019). The outcome is a quantified "Minimum Breakthrough Detection Time in Minutes."

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning study, so there is no concept of a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for this type of physical performance study, no ground truth was established for it.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1