Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-sterile, Low Dermatitis Potential. Tested for Use with Chemotherapy Drugs and Fentanyl Citrate.
K223375 · Kossan International Sdn Bhd3 · LZA · Mar 15, 2023 · General Hospital
Device Facts
| Record ID | K223375 |
| Device Name | Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-sterile, Low Dermatitis Potential. Tested for Use with Chemotherapy Drugs and Fentanyl Citrate. |
| Applicant | Kossan International Sdn Bhd3 |
| Product Code | LZA · General Hospital |
| Decision Date | Mar 15, 2023 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.6250 |
| Device Class | Class 1 |
Intended Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Device Story
Disposable nitrile patient examination glove; blue colored; non-sterile. Worn by healthcare professionals during patient examinations to provide a protective barrier against contamination. Tested for permeation resistance against various chemotherapy agents and Fentanyl Citrate per ASTM D6978-05. Provides protection for clinicians handling hazardous drugs; however, specific permeation limits apply for certain agents (e.g., Carmustine, Thiotepa). Used in clinical settings; point-of-care. Benefits include reduced risk of exposure to infectious materials and specific chemical agents during routine medical procedures.
Clinical Evidence
No clinical data. Bench testing only. Permeation testing performed per ASTM D6978-05 (Reapproved 2019) to determine breakthrough detection times for a panel of chemotherapy drugs and Fentanyl Citrate.
Technological Characteristics
Material: Nitrile. Form factor: Blue, powder-free, non-sterile patient examination glove. Testing standards: ASTM D6978-05 (Reapproved 2019) for chemotherapy drug permeation. No electronic components, software, or energy sources.
Indications for Use
Indicated for use as a disposable patient examination glove to prevent cross-contamination between patient and examiner. Suitable for use with specific chemotherapy drugs and Fentanyl Citrate. Contraindicated for use with Carmustine (BCNU).
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
- K181857 — Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Low Dermatitis Potential. Tested for Use with Chemotherapy Drugs and Fentanyl Citrate · Kossan International Sdn Bhd · Nov 20, 2018
- K161944 — Synthetic Nitrile Patient Examination Gloves, Powder Free, Blue Color, and Tested for Use with Chemotherapy Drugs · Shandong Baisheng Medical Products Co., Ltd. · Nov 21, 2016
- K220943 — Dual Color Blue/White Powder Free Nitrile Examination Glove with Chemotherapy Drug and Fentanyl Test Claim · Eco Medi Glove Sdn. Bhd. · Jul 19, 2022
- K211810 — Shamrock Powder Free Blue Nitrile Examination Gloves (Tested for use with Chemotherapy Drugs and Fentanyl) · Pt. Shamrock Manufacturing Corpora · May 23, 2022
- K213408 — BPG ChemoPlus Nitrile Patient Examination Gloves, Powder Free · Meditech Gloves Sdn Bhd · Apr 17, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
March 15, 2023
Kossan International Sdn Bhd3 Cho Sow Fong Senior Manager Regulatory Affairs Wisma Kossan, Lot 782, Jalan Sungai Putus Off Batu 3 3/4, Jalan Kapar Klang, Selangor 42100 Malaysia
Re: K223375
Trade/Device Name: Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-sterile, Low Dermatitis Potential. Tested for Use with Chemotherapy Drugs and Fentanyl Citrate Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, OPJ, QDO Dated: January 31, 2023 Received: February 2, 2023
Dear Cho Sow Fong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bifeng Qian -S
Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223375
#### Device Name
Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Low Dermatitis Potential. Tested for Use with Chemotherapy Drugs and Fentanyl Citrate.
#### Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
These gloves were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time in Minutes |
|---------------------------------------|------------------------------------------------|
| Bendamustine HCI (Treanda) (5 mg/ml) | >240 |
| Bleomycin Sulfate (15 mg/ml) | >240 |
| Busulfan (6 mg/ml) | >240 |
| Carboplatin (10 mg/ml) | >240 |
| Carfilzomib (2 mg/ml) | >240 |
| Carmustine (BCNU) (3.3 mg/ml) | 26.2 |
| Cetuximab (Erbitux) (2 mg/ml) | >240 |
| Cisplatin (1 mg/ml) | >240 |
| Cladribine (1 mg/ml) | >240 |
| Cyclosporin A (100 mg/ml) | >240 |
| Cyclophosphamide (Cytoxan) (20 mg/ml) | >240 |
| Cytarabine (100 mg/ml) | >240 |
| Cytovene (Ganciclovir) (10 mg/ml) | >240 |
| Dacarbazine (DTIC) (10 mg/ml) | >240 |
| Daunorubicin HCI (5 mg/ml) | >240 |
| Decitabine (5 mg/ml) | >240 |
| Docetaxel (Taxotere) (20 mg/ml) | >240 |
| Doxorubicin Hydrochloride (2 mg/ml) | >240 |
| Epirubicin HCI (Ellence) (2 mg/ml) | >240 |
| Etoposide (20 mg/ml) | >240 |
| Fludarabine (25 mg/ml) | >240 |
| Fluorouracil (50 mg/ml) | >240 |
| Gemcitabine (38 mg/ml) | >240 |
| Idarubicin HCI (1 mg/ml) | >240 |
| Ifosfamide (50 mg/ml) | >240 |
| Irinotecan (20 mg/ml) | >240 |
| Mechlorethamine HCI (1 mg/ml) | >240 |
| Melphalan (5 mg/ml) | >240 |
| Methotrexate (25 mg/ml) | >240 |
| Mitomycin C (0.5 mg/ml) | >240 |
| Mitoxantrone (2 mg/ml) | >240 |
| Oxaliplatin (5 mg/ml) | >240 |
| Paclitaxel (Taxol) (6 mg/ml) | >240 |
| Pemetrexed (25 mg/ml) | >240 |
| Raltitrexed (0.5 mg/ml) | >240 |
| Retrovir (Zidovudine) (10 mg/ml) | >240 |
| Rituximab (10 mg/ml) | >240 |
| Thiotepa (10 mg/ml) | 59.1 |
| Topotecan (1 mg/ml) | >240 |
| Trisenox (Arsenic Trioxide) (1 mg/ml) | >240 |
| Velcade (Bortezomib) (1 mg/ml) | >240 |
| Vidaza (Azacytidine) (25 mg/ml) | >240 |
| Vinblastine (1 mg/ml) | >240 |
| Vincristine Sulfate (1 mg/ml) | >240 |
| Vinorelbine (10 mg/ml) | >240 |
| Zoledronic Acid (1 mg/25ml) | >240 |
{3}------------------------------------------------
Please note that Carmustine (BCNU) and Thiotepa has low permeation times of 26.2 and 59.1 minutes respectively. Warning: Do Not Use with Carmustine (BCNU)
Caution: Thiotepa has low permeation times of 59.1 minutes
| Opioid and Concentration | Minimum Breakthrough Detection Time in Minutes |
|------------------------------------------|------------------------------------------------|
| Fentanyl Citrate Injection (100 mcg/2ml) | >240 |
| Simulated Gastric Acid Fluid | Minimum Breakthrough Detection Time in Minutes |
Simulated Gastric Acid Fluid
>240
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."