Search Filters

Search Results

Found 11 results

510(k) Data Aggregation

    K Number
    K251845

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-07-16

    (30 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Kerecis® Marigen Wound Extra Autologous Hydration, Kerecis Silicone Autologous Hydration and Kerecis Parvus Autologous Hydration

    Management of wounds including:

    • Partial thickness wounds
    • Full thickness wounds
    • Pressure ulcers
    • Venous ulcers
    • Chronic vascular ulcers
    • Diabetic ulcers
    • Trauma wounds (abrasions, lacerations, partial thickness burns, skin tears)
    • Surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence)
    • Draining wounds
    Device Description

    All three subject devices of this bundled submission are part of a family of devices manufactured by Kerecis Limited. The subject devices can be seen in Table 1. They are lyophilized, terminally sterilized, fish skin medical devices comprised of biocompatible, resorbable fish skin (Wild North Atlantic Cod) for wound management. The devices are intended for single use only. The devices are applied to the wound bed to maintain a moist wound environment. The primary predicate device is Marigen Wound Extra (K190528) and the additional predicate devices are Kerecis Silicone (K213231), and Kerecis Parvus (K241080). Marigen Wound Extra is commercially available under the names Kerecis MariGen, Kerecis GraftGuide, and Kerecis SurgiClose. Kerecis Silicone is commercially available under the names Kerecis Shield and Kerecis SurgiClose Silicone. For clarity, this submission will refer to the devices under their commercially available names, except when specifically referring to the primary predicate device. This information is also shown in Table 1.
    Although the subject devices differ from each other in terms of device indications and dimensional specifications, each one remains physically identical to its primary predicate device, both in design and packaging, as well as for indications for use. The only difference between each subject device and its respective primary predicate device is in the device labeling, with the subject devices having additional rehydration fluid options included in their instructions for use (IFUs).

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary describe the acceptance criteria and a study to prove the device meets these criteria. However, it's important to note that this submission is for a modification to an existing device (Kerecis Marigen Wound Extra, Kerecis Silicone, Kerecis Parvus) and not for an entirely novel device. The modification specifically addresses the inclusion of additional rehydration fluids. Therefore, the "study" is focused on verifying the device's performance with these new rehydration fluids, rather than establishing initial clinical effectiveness.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the device's performance with the original rehydration fluid (saline) and the need for the device to perform comparably with the new rehydration fluids. The performance is assessed through specific bench tests.

    Acceptance Criteria CategorySpecific Criteria/TestReported Device Performance
    RehydrationDevice rehydration with lactated Ringer's solutionPerformance remains consistent and comparable to primary predicate devices.
    Suture RetentionDevice suture retention with autologous body fluidPerformance remains consistent and comparable to primary predicate devices.
    General Device PerformanceMaintenance of intended use, materials, manufacturing processes, and fundamental scientific technology despite new rehydration fluid options.Device performance remains consistent and comparable to primary predicate devices, supporting substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size for the rehydration and suture retention tests. It mentions "bench testing" was performed.

    Data provenance: Not explicitly stated, but bench testing typically involves laboratory-controlled conditions. It is not patient data from a specific country, nor is it referred to as retrospective or prospective in a clinical trial sense.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    This type of information (expert consensus for ground truth) is typically associated with clinical studies or diagnostic device clearances where a "truth" is established through expert review of patient data (e.g., radiologist opinions on images). Since this submission focuses on bench testing for a modification to rehydration fluids, this information is not relevant or provided. The "ground truth" here is the prior established performance of the predicate device under saline rehydration.

    4. Adjudication Method for the Test Set

    Adjudication methods (e.g., 2+1, 3+1) are relevant for studies involving human interpretation or clinical endpoints. As this submission describes bench testing for material and process compatibility, an adjudication method is not applicable and therefore not provided.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically for diagnostic devices where multiple readers evaluate cases to assess performance with and without AI assistance. This submission is for wound dressings and focuses on physical and biological compatibility with different rehydration fluids.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The device is a wound dressing, not a software algorithm or an AI-based system. Therefore, there is no "algorithm only" performance to evaluate.

    7. The Type of Ground Truth Used

    For the specific tests conducted for this modification (rehydration and suture retention with new fluids), the ground truth is implicitly the established performance characteristics of the predicate device when rehydrated with saline. The goal of the new tests was to show that these characteristics are maintained or are comparable when using lactated Ringer's solution and autologous body fluids.

    8. The Sample Size for the Training Set

    This concept is not applicable as this is not a machine learning or AI device. The "training set" for a traditional medical device would refer to the data used to design and develop the device prior to its initial submission. The summary doesn't provide this detail for the original device development, only that performance testing was "leveraged from the Kerecis primary predicate devices."

    9. How the Ground Truth for the Training Set was Established

    Again, "training set" and its "ground truth" are terms typically used in AI/ML contexts. For a medical device like a wound dressing, the "ground truth" during initial development (analogous to a training phase) would involve extensive material testing, biocompatibility studies, mechanical property evaluations, and potentially pre-clinical and clinical studies to establish its safety and effectiveness for wound management. The summary indicates that for this modification, the ground truth is based on the previously established performance of the predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K251844

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-07-15

    (29 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair, or reinforcement in plastic or reconstructive surgery.

    Device Description

    Kerecis SurgiBind (50241) is a part of a family of devices manufactured by Kerecis® Limited. The device is lyophilized, terminally sterilized, fish skin medical device comprised of biocompatible, resorbable fish skin (Wild North Atlantic Cod). The device is intended for single use only.

    AI/ML Overview

    Analysis of Kerecis SurgiBind (K251844) Acceptance Criteria and Study

    This FDA 510(k) clearance letter details the Kerecis SurgiBind device, which is essentially a re-submission of the Kerecis Reconstruct (K202430) with an updated label to include additional rehydration fluid options. Due to this, the performance testing for this submission primarily focuses on demonstrating that the change (new rehydration fluids) does not negatively impact the device's performance.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal "acceptance criteria" in a quantitative sense for this 510(k) submission, as it largely leverages prior testing of the predicate device. Instead, the focus is on demonstrating that the new rehydration fluids do not compromise the device's existing performance characteristics. The implicit acceptance criterion is that performance using the expanded rehydration fluids (including autologous body fluids) is comparable to performance using the previously cleared rehydration fluids (saline and Ringer's solution).

    Acceptance Criteria (Implicit)Reported Device Performance (with autologous body fluids)
    Device performance remains consistent and comparable to the predicate device."The results confirmed that device performance remains consistent and comparable to the predicate device, supporting a determination of substantial equivalence."
    Suture retention remains acceptable.Suture retention tests using autologous body fluid were conducted.
    Rehydration properties remain acceptable.Rehydration tests using autologous body fluid were conducted.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "Performance testing... was largely leveraged from the Kerecis predicate device... As a result, performance testing is not required, except for rehydration and suture retention tests using autologous body fluid."

    • Sample Size for this submission's additional tests: Not explicitly stated. The nature of these tests (bench testing) suggests a relatively small, controlled sample size, but the exact number is not provided.
    • Data Provenance: The new tests (rehydration and suture retention with autologous body fluids) are presumed to be prospective bench testing conducted specifically for this submission. The origin of the "autologous body fluids (such as blood)" used for testing is not detailed (e.g., human, animal, synthetic proxy).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Experts: Not applicable for this type of bench testing. The "ground truth" for these tests would be the established scientific and engineering principles for material properties and device function.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. Performance testing for this submission involves objective physical measurements (e.g., rehydration rate, suture pull-out force) rather than subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • MRMC Study: No. This device is a biological surgical mesh (fish skin), not an AI/software device. Therefore, MRMC studies are not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Standalone Performance: Not applicable. This is not an algorithm or AI device.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: For the specific tests conducted for this submission (rehydration and suture retention with autologous body fluids), the ground truth is based on bench testing measurements against established parameters or comparison to the predicate device's performance. For the broader substantial equivalence claim, the ground truth is the performance and safety profile of the legally marketed predicate device (K202430).

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable. This is a medical device, not an AI/machine learning model that requires a "training set." The manufacturing process for the fish skin material itself has an inherent "training" through its development and validation, but this is not typically referred to as a "training set" in this context.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K243843

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-06-04

    (173 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Kerecis® Tendon Protect is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.

    Device Description

    Kerecis® Tendon Protect is a medical device made from solid fish skin (North Atlantic Cod) sheets of thickness no less than 0.4 mm. The product is intended to provide a covering around the newly repaired tendon for the management and protection of tendon injuries in which there has been no substantial loss of tissue. The device with natural structure and proteins from fish skin is applied around the newly repaired tendon. The structure of the fish skin provides a supportive environment for new host tissue deposition as part of wound healing while still providing an initial barrier facilitating the sliding of the repaired tendon within the surrounding tissue during tendon use. The biomechanical strength for the tendon itself is provided separately by sutures used to repair the tendon tear and/or by sutures or bone anchors used to attach the tendon tissue to the bone.

    The fish skin is fully integrated into the surrounding tissue over time, with corresponding new host tissue deposition. The device is biocompatible, non-crosslinked, bioresorbable, strong, and pliable. Its tensile strength supports fixation by sutures customary for tendon protection surgical procedures. The device is intended for one-time use and supplied sterile in labelled Tyvek® pouch packaging.

    The subject device is available in following sizes:

    • 3x5 cm size
    • 6x9 cm size

    Kerecis® Tendon Protect is a standalone, single use medical device made from fish skin that is technologically similar to the predicate device Tendon Wrap Tendon Protector (K053655) (Substantial Equivalence) and identical to the Kerecis® Reconstruct™ (K202430).

    AI/ML Overview

    The provided text is a 510(k) clearance letter and summary for a medical device called "Tendon Protect." It details the device's characteristics, indications for use, and a summary of the non-clinical performance testing conducted to demonstrate its safety and effectiveness and substantial equivalence to predicate devices. However, it does not contain the specific detailed information typically found in acceptance criteria tables or a comprehensive study report, particularly regarding a multi-reader multi-case (MRMC) study, standalone algorithm performance, or the detailed methodology for ground truth establishment that would be associated with AI/ML-based medical devices.

    The document describes a physical medical device (surgical mesh made from fish skin) rather than an AI/ML-based diagnostic or assistive software. Therefore, many of the questions related to AI/ML specific criteria (such as sample size for training set, number of experts for ground truth, adjudication methods, MRMC studies, and standalone algorithm performance) are not applicable to this particular device submission.

    Here's a breakdown of the applicable information found in the document:

    1. A table of acceptance criteria and the reported device performance:

    The document mentions that the device "met all the pre-defined acceptance criteria" for various tests, but it does not provide a table detailing the specific numerical or quantitative acceptance criteria and the corresponding reported performance values. It only states that the tests were "conducted and passed successfully according to the test specifications."

    Test CategoryTest NameAcceptance CriteriaReported Device Performance
    BiocompatibilityCytotoxicity (ISO 10993-5)(Not specified, but implied "pass")Passed
    Sensitization (10993-10)(Not specified, but implied "pass")Passed
    Irritation or intracutaneous reactivity (ISO 10993-10)(Not specified, but implied "pass")Passed
    Material mediated pyrogenicity (ISO 10993-11)(Not specified, but implied "pass")Passed
    Implantation Effects (ISO 10993-6)(Not specified, but implied "pass")Passed
    Genotoxicity (ISO 10993-3)(Not specified, but implied "pass")Passed
    Acute systemic toxicity (ISO 10993-11)(Not specified, but implied "pass")Passed
    Subacute/subchronic toxicity (ISO 10993-11)(Not specified, but implied "pass")Passed
    Bench TestingTendon Summative Usability Evaluation (ISO 62366-2)Pre-defined acceptance criteria for safety and effectivenessMet all pre-defined acceptance criteria
    Ball burst testing(Not specified, but implied "met acceptance criteria")Met acceptance criteria
    Suture retention testing(Not specified, but implied "met acceptance criteria")Met acceptance criteria
    Tensile strength testing(Not specified, but implied "met acceptance criteria")Met acceptance criteria
    Animal Testing (GLP Chicken Study)Histopathology evaluation(Not specified, but implied "met acceptance criteria")Acceptance criteria met; exhibited similar overall performance to predicate
    Macroscopic evaluations(Not specified, but implied "met acceptance criteria")Acceptance criteria met; exhibited similar overall performance to predicate
    Tendon biomechanical function assessment(Not specified, but implied "met acceptance criteria")Acceptance criteria met; exhibited similar overall performance to predicate
    Overall animal health(Not specified, but implied "met acceptance criteria")Acceptance criteria met; exhibited similar overall performance to predicate

    2. Sample sizes used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated for most tests (e.g., biocompatibility bench tests). For the Tendon GLP Chicken Study, it is a chronic model evaluating early, mid-term, and late time points, comparing the subject device to a control and a negative control. The specific number of animals or tendons tested is not provided.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be pre-clinical (bench and animal testing), not human clinical data, and are likely prospective given they are part of a regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable as the device is a physical surgical mesh, not a diagnostic or AI-based device requiring expert interpretation of images or signals for ground truth. Ground truth for the animal study would be based on objective scientific measurements and observations (e.g., histology, biomechanics).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This is not applicable for the reasons stated above.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable as the device is a physical surgical mesh, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable as the device is a physical surgical mesh, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the animal study:

    • Histopathology evaluation
    • Macroscopic evaluations
    • Tendon biomechanical function assessment
    • Overall animal health

    For bench testing, ground truth would be established by the physical and mechanical properties of the material itself, measured against predefined specifications.

    8. The sample size for the training set:

    This is not applicable as the device is a physical surgical mesh, not an AI/ML system requiring a training set.

    9. How the ground truth for the training set was established:

    This is not applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241080

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-21

    (124 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Management of wounds including:

    • Partial and full-thickness wounds
    • Pressure ulcers
    • Venous ulcers
    • Chronic vascular ulcers
    • Diabetic ulcers
    • Trauma wounds (abrasions, lacerations, partial thickness burns, skin tears)
    • Surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence)
    • Draining wounds
    Device Description

    Kerecis "Parvus™ is a lyophilized, terminally sterilized, acellular, particulate fish skin medical device comprised of biocompatible, non-crosslinked, resorbable, acellular fish skin (North Atlantic Cod) for wound management. The device is intended for single use only.

    The subject device is packaged in the following weights:
    100mg (4 cm2) 200mg (8 cm2) 500mg (19 cm²) 1000mg (38 cm²) 2,500mg (95 cm2 ) 3,000mg (114 cm2)

    AI/ML Overview

    This document is a 510(k) Premarket Notification summary for Kerecis® Parvus™, a wound dressing. It is not an AI/ML device, and therefore does not contain acceptance criteria for device performance related to AI/ML or a study proving those criteria are met. The document states that the Kerecis® Parvus™ is substantially equivalent to a predicate device, Kerecis MariGen Wound Extra (K190528), in terms of indications for use, intended use, raw material origin and composition, device performance, packaging material, and sterilization methods. The primary difference is an additional cutting and sieving step in the manufacturing process to convert the fish skin from intact sheets to fragmented pieces ≤2.0mm.

    Here's a breakdown of the requested information based on the provided document, noting that many items are not applicable (N/A) due to the nature of the device:

    1. A table of acceptance criteria and the reported device performance

      This document does not provide specific performance acceptance criteria or reported performance data in the traditional sense of a clinical or analytical study with defined metrics (e.g., accuracy, sensitivity, specificity). Instead, it asserts substantial equivalence to a predicate device based on material properties and manufacturing processes. The "Performance Characteristics" section for the subject device largely mirrors the predicate device, with differences primarily in the physical form (fragmented vs. sheets) and nominal sizes.

      CharacteristicAcceptance Criteria (Implied by Substantial Equivalence Goal)Reported Device Performance (as described)
      Source OriginWild Caught Atlantic Cod FishWild Caught Atlantic Cod Fish (Same as predicate)
      Tissue sourceFish SkinFish Skin (Same as predicate)
      Nominal SizesFragmented, size controlled to ≤2.00mmIrregular shaped 3D fragmented fish skin, size controlled to ≤2.00mm and packaged in various weights (100mg to 3,000mg). (Differs from predicate's sheet sizes, but subject device is cut from predicate device sizes).
      PresentationLyophilized, sterilized, fragmented fish skinLyophilized, sterilized, Fragmented, fish skin in a Tyvek peel pouch. (Differs from predicate's sheet presentation due to cutting and sieving).
      PackagingTyvek Single and Double Peel PouchTyvek Single and Double Peel Pouch (Same as predicate)
      SterilizationEthylene OxideEthylene Oxide (Same as predicate)
      Sterility Assurance Level (SAL)SAL 10^-6^SAL 10^-6^ (Same as predicate)
      Endotoxin limit≤20 EU/device≤20 EU/device (Same as predicate)
      Shelf LifeAt least 1 year (matching current reported)1 year (Real-time shelf life testing in progress for the subject device; predicate has 3 years).
      BiocompatibilityDemonstratedAdditional testing for biocompatibility completed (for the subject device).
      Fragmented size characterizationDemonstratedAdditional testing for fragmented size characterization completed (for the subject device).
      Ethylene oxide residualsDemonstratedAdditional testing for ethylene oxide residuals completed (for the subject device).
      Metal contaminationDemonstratedAdditional testing for metal contamination completed (for the subject device).
      Protein analysisDemonstratedAdditional testing for protein analysis completed (for the subject device).
      BioburdenDemonstratedAdditional testing for bioburden completed (for the subject device).
      Residual limitsDemonstratedAdditional testing for residual limits completed (for the subject device).
    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      N/A. This is a 510(k) submission for a non-AI/ML medical device establishing substantial equivalence through material and manufacturing process comparisons, not a clinical study involving a test set for performance evaluation. The "additional testing" mentioned is likely laboratory bench testing on the device materials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      N/A. Not an AI/ML device requiring expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      N/A. Not an AI/ML device requiring adjudication of a test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      N/A. This device is not an AI/ML product and does not involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

      N/A. This device is not an AI/ML product.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

      N/A. Not an AI/ML device requiring ground truth for performance evaluation. The substantial equivalence argument relies on material characteristics and manufacturing processes.

    8. The sample size for the training set

      N/A. This is not an AI/ML device with a training set.

    9. How the ground truth for the training set was established

      N/A. This is not an AI/ML device with a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213904

    Validate with FDA (Live)

    Device Name
    Kerecis Oral
    Manufacturer
    Date Cleared
    2022-09-30

    (290 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • Localized gingival augmentation to increase keratinized tissue (KT) around teeth or implants;
    • Covering of implants placed in immediate extraction sockets;
    • Covering of implants placed in delayed extraction sockets;
    • Covering of bone defects after root resection and removal of retained teeth; and
    • Guided tissue regeneration procedures in periodontal and recession defects.
    Device Description

    The subject device is an acellular resorbable fish dermal matrix, intended for use in periodontal surgical procedures to aid in soft tissue and bone regeneration. It is obtained from cod fish skin by a standardized controlled manufacturing process, and supplied in terminally sterile peel-pouch packaging in the following solid sizes:

    • 15mm x 20mm
    • 20mm x 30mm
    • . 30mm x 40mm
      It is biocompatible, non-cross linked, and therefore resorbable, strong, flexible, and supports fixation by sutures and pins.
    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the Kerecis® Oral device. It details the device's characteristics, intended use, and a comparison to predicate and reference devices to demonstrate substantial equivalence.

    Based on the provided text, the device itself (Kerecis® Oral) is a biological material (acellular resorbable fish dermal matrix) used for various oral surgical procedures to aid in soft tissue and bone regeneration. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the bench testing and animal studies performed to demonstrate that the Kerecis® Oral device is substantially equivalent to existing predicate devices, thereby clearing it for market.

    Here's an analysis of the provided information to answer your questions:

    1. Table of acceptance criteria and the reported device performance

    The document doesn't present a single, explicit "acceptance criteria" table with pass/fail thresholds for the entire device in the way a software or AI model might have. Instead, it outlines a series of tests performed on the material properties and functional characteristics of the Kerecis® Oral device and compares them to predicate devices. The "acceptance" is implied by demonstrating substantial equivalence.

    Here's a table summarizing the comparison and performance points, where "acceptance" means demonstrating substantial equivalence or meeting/exceeding predicate device performance:

    Feature/TestAcceptance Criteria (Implied: Substantial Equivalence to Predicate/Reference)Reported Device Performance (Kerecis® Oral)
    Technological Characteristics
    Device NameN/A (Identification)Kerecis Oral
    Regulation21 CFR 872.3930 (Bone Grafting Material)21 CFR 872.3930 (Same Regulation as Predicate)
    Device ClassClass IIClass II (Same Class as Predicate)
    Product CodeNPLNPL (Same Product Code as Predicate)
    Device ClassificationBarrier, Animal Source, IntraoralBarrier, Animal Source, Intraoral (Same Classification as Predicate)
    Intended UseIncludes elements of predicate and reference devices, supported by performance testing.Covers localized gingival augmentation, implant covering (immediate and delayed), bone defect covering, GTR.
    Animal Origin MaterialSimilar origin to reference device (fish) or acceptable alternative to predicate (porcine).Fish: skin tissue, single layer sheet (Same animal species as reference)
    BiocompatibilityYes (demonstrated per ISO 10993 series)Yes (Biocompatibility testing performed per ISO 10993 series standards)
    Non-PyrogenicYes (demonstrated safe per ISO 10993 series)Yes (Materials-mediated pyrogenicity safe per ISO 10993 series standard)
    Tensile StrengthComparable to predicate device (4.6 MPa for predicate)14.3 MPa (Meets/Exceeds tensile strength value of the predicate device per ASTM D638)
    ResorbableYesYes (All devices are resorbable per comparative performance data)
    SizesEquivalent to predicate/reference sizes (15x20mm, 20x30mm, 30x40mm)15x20mm, 20x30mm, 30x40mm (Equivalent sizes by dimensional analysis)
    SterilizationTraditional Sterilization Methods (EO or Gamma)Ethylene Oxide (Traditional Sterilization Methods per ISO 11135 and ISO 11737-1. EO residual testing per ISO 10993-7)
    Sterility Assurance Level (SAL)10^-610^-6 (Equivalent SAL per ISO 11137 and ISO 11737)
    Shelf Life3 years3 years (Equivalent shelf life per ASTM F1980 and Q5C (R2)[ICH])
    Mode of ActionFixationFixation (Equivalent mode of action per ANSI/AAMI/ISO 7198 and ASTM F-1839-08)
    Performance Testing - Bench
    Morphology Observation (H&E, SEM)Rich in collagen, porous, favoring cellular infiltration, preserved collagen structure.Rich in collagen and porous, favoring cellular infiltration. SEM shows equivalent preserved collagen structure.
    Cellular Ingrowth Comparison (Fibroblast)Favorable cellular infiltration after 14 days.Favorable cellular infiltration of fibroblasts after 14 days.
    Heavy Metal AnalysisAcceptable limits per ICH guidelines (Q3D Elemental Impurities).Limits of Cd, Pb, As, Hg acceptable under ICH guidelines.
    Dimensional ValidationQuality limits for capability index > 1.33.All tested parameters met or exceeded the set goal (>1.33) on three lots.
    Stability in Simulated Physiological Env.Structurally stable compared to predicate, dissolves slower at neutral pH.Structurally more stable than predicate, dissolved slower than predicate at neutral pH 7.
    Suture Pull-Out StrengthMeets or exceeds predicate with >95% confidence.Meets or exceeds predicate with >95% confidence (Consistent with ANSI/AAMI/ISO 7198).
    Pin Pull-Out StrengthExceeds predicate with >95% confidence.Exceeds predicate with >95% confidence (Consistent with ASTM F-1839-08).
    Compression (Peak-Load, Load-at-Break, etc.)Meet or exceed predicate with >95% confidence.Meet or exceed predicate with >95% confidence (Consistent with ASTM-F-1306).
    User Evaluation (Cutting and Shaping)Favorable usability, substantially equivalent to predicate.Favorable usability, substantially equivalent to predicate for cutting and shaping (via questionnaire).
    PackagingCompliant with ISO 11607-series, ASTM F88 and ASTM F1886.Compliant.
    Animal Origin and Viral InactivationCompliant with ISO 22422 series.Compliant.
    Performance Testing - Animal
    New Bone Formation, Xenograft Resorption, Soft Tissue InfiltrationNo qualitative or significant differences compared to predicate.No noticeable or significant differences between the two membranes at any time point (Histologic and micro-CT volumetric analysis).

    2. Sample sizes used for the test set and the data provenance

    • Bench Testing: The specific sample sizes for each bench test (e.g., number of samples for tensile strength, compression, dimensional validation) are not explicitly stated in the provided text. It mentions "representative product samples" and "three lots" for dimensional validation. The provenance is implied to be laboratory testing of the manufactured Kerecis® Oral device.
    • Animal Testing: The sample size for the animal study is a "canine (beagle bilateral mandibular bony defect model)." This implies multiple beagle dogs, but the exact number is not specified. The study was conducted in AAALC accredited facilities in compliance with GLP (Good Laboratory Practices) 21 CFR §58, suggesting a controlled, prospective study. The location (country) is not specified, but GLP compliance implies a rigorous study design.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • For the animal study, the evaluation involved "gross pathology and histological, and quantitative micro-CT volumetric analysis." This type of analysis would typically be performed by veterinary pathologists and possibly radiologists/imaging specialists. However, the number and specific qualifications of these experts are not mentioned in the text.
    • For the user evaluation of cutting and shaping, it was done "using a questionnaire." This implies that "ground truth" was established by user feedback, not by experts in the context of diagnostic accuracy. The number and qualifications of these users are not stated.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • None explicitly mentioned. For the animal studies, "no qualitative or significant differences was observed" suggests direct comparison by the evaluators rather than an explicit adjudication process among multiple independent reviewers, though this is not definitively stated. For the bench tests, the data itself (e.g., tensile strength, dimensional measurements) would be objective, not typically requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This product is a biological medical device (a tissue graft), not a diagnostic AI device that assists human readers. Therefore, the concept of "human readers improving with AI assistance" is not applicable here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical medical device, not an algorithm. The "performance" assessment is based on its material properties and biological interaction in an in vitro and in vivo (animal) setting, not an algorithm's output.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • For Bench Testing: The "ground truth" is defined by objective material property measurements (e.g., tensile strength in MPa, dimensional measurements in mm, pH and conductivity values) and established international standards (e.g., ISO, ASTM, ICH).
    • For Animal Testing: The "ground truth" was established through histologic and micro-CT volumetric analysis of animal tissues collected at different time points post-surgery. This involves pathological assessment (histology) and quantitative imaging analysis (micro-CT) to assess bone formation, xenograft resorption, and soft tissue infiltration. While the specific experts aren't named, this is a form of expert assessment of biological outcomes.

    8. The sample size for the training set

    • Not applicable. This is a physical medical device, not a machine learning model, so there is no "training set" in the context of an AI/ML algorithm.

    9. How the ground truth for the training set was established

    • Not applicable. As a physical device, there is no "training set" or "ground truth for the training set."
    Ask a Question

    Ask a specific question about this device

    K Number
    K213231

    Validate with FDA (Live)

    Device Name
    Kerecis Silicone
    Manufacturer
    Date Cleared
    2022-06-29

    (272 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Kerecis Silicone is indicated for the management of wounds including:

    • Partial and full-thickness wounds
    • Pressure ulcers
    • Venous ulcers
    • Chronic vascular ulcers
    • Diabetic ulcers
    • Trauma wounds (abrasions, lacerations, partial-thickness burns, skin tears)
    • Surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence)
    • Draining Wounds
    Device Description

    The subject device is a bilayer of processed resorbable acellular fish dermal matrix adhered to a thin, transparent, porous, soft silicone layer.
    The subject device is obtained from fish skin via standardized controlled GMP manufacturing process. The fish dermal matrix layer is approximately 1 mm in thickness and is porous.
    The silicone layer is a transparent polyurethane film single-coated with soft, medical grade silicone that is attached to the scaly side of the fish dermal matrix. The silicone layer is porous, soft and conformable to the wound surface.
    The subject device is supplied as a sterile intact sheet offered in two configurations: a) with the silicone layer extending beyond the borders of the fish dermal matrix and b) the silicone layer having the same dimension as the fish dermal matrix with no silicone layer extending beyond the fish dermal matrix.
    The silicone acts as: protection for the fish dermal matrix layer, as additional wound coverage, and in configuration (a), as an adhesive contact layer to the skin surrounding the wound.
    The silicone layer can be peeled off as the fish dermal matrix is resorbed.
    The device is intended for single use only.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device (Kerecis Silicone) and does not describe a study involving an AI/Machine Learning device or a human-in-the-loop study.

    This document describes a medical device, Kerecis Silicone, which is a bilayer wound dressing. The substantial equivalence determination is based on the technological characteristics and performance testing of the device itself (biocompatibility, physical properties, etc.), not on the performance of a software algorithm or AI.

    Therefore, most of the requested information regarding AI/ML acceptance criteria, ground truth establishment, sample sizes for AI training/testing, expert adjudication, or MRMC studies for AI assistance is not applicable to this document.

    However, I can extract the relevant information regarding the device's acceptance criteria (in terms of performance testing) and the study proving it meets these criteria based on the provided text.


    Device: Kerecis Silicone (a bilayer wound dressing)
    Regulatory Status: 510(k) clearance (K213231)

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this medical device are based on demonstrating biocompatibility and physical/chemical characteristics that are comparable to or safe for its intended use, typically by meeting the requirements of established ISO and ASTM standards. The reported device performance is indicated by a "Pass" for each test.

    TestStandard / MethodAcceptance Criteria (Implied: Meeting Standard Requirements)Reported Device Performance
    CytotoxicityISO 10993-5, Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicityNon-cytotoxicPass
    SensitizationISO 10993-10, Biological evaluation of medical devices — Part 10: Tests for irritation and skin sensitizationNon-sensitizingPass
    Intracutaneous IrritationISO 10993-10, Biological evaluation of medical devices — Part 10: Tests for irritation and skin sensitizationNon-irritatingPass
    Acute Systemic ToxicityISO 10993-11:2017, Biological evaluation of medical devices — Part 11: Tests for systemic toxicityNon-systemically toxicPass
    Material Mediated PyrogenicityISO 10993-11:2017, Biological evaluation of medical devices — Part 11: Tests for systemic toxicityNon-pyrogenicPass
    Subacute Systemic Toxicity and Implantation EffectsISO 10993-6, Biological evaluation of medical devices — Part 6: Tests for local effects after implantation and ISO 10993-11:2017, Biological evaluation of medical devices — Part 11: Tests for systemic toxicityAcceptable local and systemic effectsPass
    Endotoxin Validation and AnalysisANSI/AAMI ST72 Bacterial Endotoxins - Test Methods, Routine Monitoring, And Alternatives To Batch TestingAcceptable endotoxin levelsPass
    HydrationN/A (Internal method implied)Adequate hydration properties (implied)Pass
    Shelf Life and StabilityN/A (Internal method implied)Stable for 36 monthsPass
    Tensile strengthASTM D638-14 Tensile Properties of PlasticsAdequate tensile strength (implied)Pass
    Implantation (for fish dermal matrix alone)ISO 10993-6, Biological evaluation of medical devices — Part 6: Tests for local effects after implantationAcceptable local effects post-implantationPass
    Genotoxicity (for fish dermal matrix alone)ISO 10993-3, Biological evaluation of medical devices — Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicityNon-genotoxicPass
    Genotoxicity – Chromosomal Aberration (for fish dermal matrix alone)ISO 10993-3, Biological evaluation of medical devices — Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicityNo chromosomal aberrationsPass
    Subchronic Toxicity (for fish dermal matrix alone)ISO 10993-11:2017, Biological evaluation of medical devices — Part 11: Tests for systemic toxicityNon-systemically toxic (subchronic)Pass
    Cellular Ingrowth (for fish dermal matrix alone)N/A (Internal method implied)Evidence of cellular ingrowth (implied)Pass
    Cell Remains and Collagen Structure (for fish dermal matrix alone)N/A (Internal method implied)Intact collagen structure, minimal cell remains (implied)Pass
    Hemostatic Properties (for fish dermal matrix alone)N/A (Internal method implied)Acceptable hemostatic properties (implied)Pass
    Resorption in Sprague-Dawley Rats (for fish dermal matrix alone)N/A (Internal method implied)Acceptable resorption profile (implied)Pass
    Collagen Induced Arthritis Mouse Modal (for fish dermal matrix alone)N/A (Internal method implied)No adverse inflammatory response in model (implied)Pass

    Study Proving Device Meets Acceptance Criteria

    The study proving the device meets the acceptance criteria is detailed under "7. PERFORMANCE TESTING" in the 510(k) summary. These are pre-clinical (biocompatibility, physical, and chemical) tests.

    1. Sample size used for the test set and the data provenance:

      • The document does not explicitly state the sample sizes for each specific biological or physical test. These tests generally use a small number of samples (e.g., typically n=3 or n=5 for physical tests, or a defined number of animals for in-vivo biocompatibility tests) as per the specific ISO/ASTM standards referenced.
      • The data provenance is not specified in terms of country of origin, but it is implied to be from laboratory testing supporting the device's manufacturing and regulatory submission. The tests are retrospective to the submission date.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable as this document does not describe a study involving image interpretation by experts or a clinical diagnostic AI. The "ground truth" here is established by validated laboratory testing methods governed by the cited ISO/ASTM standards, and the results are interpreted by qualified laboratory personnel.
    3. Adjudication method for the test set:

      • Not applicable. There is no "adjudication" in the sense of expert consensus for laboratory tests. The results are typically quantitative or qualitative (e.g., cytotoxic/non-cytotoxic) based on direct measurement and predefined criteria within the respective standards.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

      • No. This is not an AI/ML diagnostic device, so an MRMC study comparing human readers with and without AI assistance is irrelevant and was not performed.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical medical device, not a software algorithm.
    6. The type of ground truth used:

      • The ground truth for this device's performance is established through adherence to recognized international standards (ISO, ASTM, ANSI/AAMI) for biological safety and physical properties. This involves direct laboratory measurements and observations under controlled conditions to determine if the device elicits specific biological responses or possesses particular physical characteristics as defined by these standards.
    7. The sample size for the training set:

      • Not applicable. This concept applies to machine learning models, not to the pre-clinical performance testing of a physical medical device. The "training" for such a device involves product development and manufacturing process controls, not data-driven algorithm training.
    8. How the ground truth for the training set was established:

      • Not applicable. As above, this pertains to AI/ML context.
    Ask a Question

    Ask a specific question about this device

    K Number
    K202430

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-05-02

    (250 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The subject device is indicated for: For implantation to reinforce soft tissue where weakness exists, in patients requiring soft tissue repair, or reinforcement in plastic or reconstructive surgery.

    Device Description

    The subject device is a fish skin medical device indicated for physical reinforcement of a soft tissue defect or weakness. The subject device is obtained from cod fish skin by a standardized controlled manufacturing process and supplied in a peel-pouch terminally sterile packaging in the following rectangular solid sizes: 4x7 cm, 7x10 cm, 7x20 cm. The subject device is biocompatible, non-crosslinked, and therefore resorbable, strong, flexible, and supports fixation.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called "Kerecis Reconstruct," which is a surgical mesh. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than detailed clinical effectiveness studies with explicit acceptance criteria for performance metrics like sensitivity, specificity, or reader improvement. Therefore, the information requested in the prompt, which is typical for AI/ML-based diagnostic devices, is not directly available in this document for the Kerecis Reconstruct device.

    However, I can extract the information that is present, particularly regarding the studies and comparisons performed to demonstrate substantial equivalence, which serves a similar purpose to meeting "acceptance criteria" in the context of a 510(k) submission for this type of device.

    Here's an analysis based on the provided text, structured to address your points where possible, and noting where the information is not applicable or not provided.

    Device: Kerecis Reconstruct (Surgical Mesh)

    1. Table of Acceptance Criteria and Reported Device Performance

    For a surgical mesh, the "acceptance criteria" for 510(k) clearance are primarily focused on demonstrating that the new device is as safe and effective as a predicate device, based on similar technological characteristics and performance. Formal quantitative performance metrics (like sensitivity/specificity for AI, or specific effect sizes for human reader improvement) are not typically applicable or reported for this type of device in a 510(k) summary in the same way they would be for an AI diagnostic algorithm.

    Instead, the "acceptance criteria" implicitly relate to meeting similar material properties, functional performance (e.g., tensile strength, suturability), and biocompatibility as the predicate device. The "reported device performance" is demonstrated through bench testing and animal studies, showing equivalence.

    Acceptance Criteria (Implicit for Surgical Mesh Equivalence)Reported Device Performance (as summarized)
    Biocompatibility & Safety: Safe tissue response, no adverse effects.Animal testing in a GLP laboratory showed: - Veterinarian's assessment of animal health (safety) was equivalent. - Pathologist's assessment of tissue response (safety and efficacy) was equivalent.
    Material Properties: Similar physical and mechanical properties to predicate (e.g., non-crosslinked, resorbable, strong, flexible).Described as biocompatible, non-crosslinked, resorbable, strong, and flexible. Derived from cod fish skin (similar to predicate being porcine).
    Mechanical Performance: Equivalent functional characteristics (e.g., tensile strength, suturability, stiffness, hydration).Bench testing (tensile strength, suturability, hydration time, thickness, weight, stiffness/bend test, microscopic structure) performed on hydrated, sterile, ready-for-market devices confirmed performance as expected. - Specific quantitative values for these tests are not provided in the summary. - In-vitro performance data tensile strength testing of the test and control explants side-by-side (efficacy) showed equivalence.
    Intended Use Compatibility: Works for reinforcing soft tissue in plastic/reconstructive surgery.Predicate device's intended use is identical: "to reinforce soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in plastic or reconstructive surgery." Subject device claims identical intended use.
    Sterility & Shelf Life: Meets established standards.Sterilization Method: Ethylene Oxide (identical to predicate/references). Shelf-Life: 3 years. Predicate was 1.5 years, but reference devices were 3 years. This seems to be accepted.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated numerically for the bench testing. The animal study involved at least one "test device" (Kerecis Reconstruct) and a "control device," but the number of animals/samples is not specified in this summary.
    • Data Provenance:
      • Bench testing: Performed internally or by a testing laboratory as part of the submission per FDA guidance.
      • Animal testing: Performed in a "GLP laboratory" (Good Laboratory Practice). The origin (country/retrospective/prospective) is not specified, but GLP implies a prospective, controlled study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable in the context of this 510(k) submission for a non-diagnostic medical device. "Ground truth" in this case is established through objective physical measurements (bench testing) and histological/veterinary assessments in the animal model. The animal study mentions a "veterinarian's assessment" and a "pathologist's assessment," but the number and specific qualifications (beyond "veterinarian" and "pathologist") are not provided.

    4. Adjudication Method for the Test Set

    Not applicable. This is not a study involving human readers/interpreters needing adjudication. The evaluation relies on objective physical/mechanical tests and expert animal health and pathology assessments.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance?

    Not applicable. This device is a surgical mesh, not an AI diagnostic tool or an imaging product that would involve human readers or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a surgical mesh, not an algorithm.

    7. The Type of Ground Truth Used

    • For bench testing: Established by validated testing methods against industry standards or direct comparison to the predicate device. This is essentially objective physical and mechanical property measurement.
    • For animal study: Established by GLP-compliant veterinary examination and histopathological analysis. This can be considered a form of "pathology" ground truth for tissue response and "outcomes data" for animal health.

    8. The Sample Size for the Training Set

    Not applicable. This is a manufactured medical device, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.


    Summary of what the document does provide regarding the "study that proves the device meets the acceptance criteria" (i.e., demonstrates substantial equivalence):

    The submission for Kerecis Reconstruct justifies substantial equivalence to the predicate device (Biodesign Plastic Surgery Matrix, K191696) and reference devices (Kerecis SecureMesh K153364, Biodesign Hernia Graft K133306) through:

    • Analysis of 510(k) Substantial Equivalence Decision-Making Process: Following FDA's guidance document.
    • Bench Testing: Performed according to FDA guidance "Preparation of a Premarket Notification Application for a Surgical Mesh" (1999). Tests included: tensile strength, suturability, hydration time, thickness, weight, stiffness (bend test), and microscopic structure analysis. These tests were performed on hydrated, sterile, ready-for-market devices. The summary states these tests "confirmed that the devices perform as expected under clinical conditions, and there were no negative effects on the mechanical properties."
    • Animal Testing: Performed in a GLP laboratory using the subject device and a control device. The results demonstrated "equivalence in safety and efficacy" to the control device, based on:
      • Veterinarian's assessment of animal health (safety).
      • Pathologist's assessment of the tissue response to the device (safety and efficacy).
      • Histology.
      • In-vitro performance data (tensile strength testing of the test and control explants side-by-side) (efficacy).

    The document concludes that the "data provided within this submission support substantial equivalence of the subject device to the predicate device with regards to intended use, technological characteristics including principles of operation, performance characteristics, and device safety."

    Ask a Question

    Ask a specific question about this device

    K Number
    K192612

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-11-13

    (420 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Kerecis Gingiva Graft is indicated for:

    • Localized gingival augmentation to increase keratinized tissue (KT) around teeth or implants.
    Device Description

    The subject device is fish skin medical device indicated gingiva augmentation.
    The subject device is obtained from cod fish skin by a standardized controlled manufacturing process and supplied in a peel-pouch terminally sterile packaging in the following rectangular sizes:

    • 15mm x 20mm
    • 20mm x 30mm
    • 30mm x 40mm
      The subject device becomes completely integrated into the surrounding tissue over time, with corresponding new host tissue deposition. The physical properties of the subject device allow cellular ingrowth for augmentation of keratinized tissue.
      The subject device is biocompatible, non-crosslinked, bioresorbable, strong, pliable and supports fixation by sutures.
    AI/ML Overview

    The provided text describes the Kerecis Gingiva Graft device, a collagen membrane intended for localized gingival augmentation. This is a 510(k) submission, meaning the device seeks clearance based on its substantial equivalence to a legally marketed predicate device, not necessarily on meeting quantitative acceptance criteria established by the FDA for novel devices.

    Therefore, the information typically requested in questions 1 through 9 (acceptance criteria, sample sizes, expert ground truth, MRMC studies, standalone performance, training sets) is not applicable in the context of this 510(k) summary, as it describes a different type of regulatory submission process. The study presented here aims to demonstrate equivalence, not to quantify the device's performance against pre-defined thresholds.

    However, I can extract information related to the "study that proves the device meets the acceptance criteria" in the context of proving substantial equivalence to the predicate device.

    Here's the relevant information based on the provided text, reinterpreting "acceptance criteria" as "criteria for demonstrating substantial equivalence" for this 510(k) submission:

    1. A table of (Substantial Equivalence) Criteria and the Reported Device Performance:

    The document establishes substantial equivalence by comparing the Kerecis Gingiva Graft (subject device) to the MUCOGRAFT® Collagen Matrix (predicate device) and other reference devices across various characteristics and performance tests. Rather than explicit "acceptance criteria" with numerical thresholds, the "performance" is the demonstration of comparability or superiority to the predicate device.

    Characteristic / Performance ElementSubstantial Equivalence Criterion (Implicit)Reported Device Performance
    Intended UseSame as, or subset of, predicate device."Subset of the intended use of the predicate device." (Kerecis: "a biocompatible, sterile collagen membrane intended for augmentation and regeneration of soft tissue in oral surgical settings." Predicate: includes broader indications like "guided tissue regeneration and multiple oral tissue defect regeneration in oral surgical settings.")
    Indications for UseSame as, or subset of, predicate device."Subset of the indications of the predicate device." (Kerecis: "Localized gingival augmentation to increase Keratinized tissue (KT) around teeth and implants." Predicate: includes this, plus "Covering of implants placed in immediate extraction sockets," "Alveolar ridge reconstruction for prosthetic treatment," and other GTR procedures.)
    Regulation, Product Code, ClassSame as predicate."Same as predicate." (21 CFR 872.3930, NPL, Class II)
    BiocompatibilityMeets ISO 10993 series standards. Comparable to predicate."Yes" (Same as predicate). Leveraged testing from applicant's own predicate devices (K190528 and K153364) for Cytotoxicity, Sensitization, Irritation, Acute/Subacute/Sub-chronic/Chronic Toxicity, Genotoxicity, Implantation, Materials-Mediated Pyrogenicity, Carcinogenicity.
    NON-PyrogenicYes."Yes" (Same as predicate). Endotoxin validation (<20 EU/device) per LAL turbidimetric kinetic method following ISO 10993-11.
    ResorbableYes."Yes" (Same as predicate).
    Sterilization & SALValidated sterilization method, SAL 10-6.Ethylene Oxide sterilization with SAL 10-6. "Traditional Sterilization Method." Ethylene Oxide residual test following ISO 10993-7.
    Shelf LifeValidated, comparable to predicate."3 years" (Same as predicate). Validated per ASTM F1980 and Q5C (R2)[ICH] using accelerated and real-time aged samples. Packaging per ISO 11607-series, ASTM F88 and ASTM F1886.
    Animal Origin MaterialSafe and suitable (via viral inactivation, etc.).North Atlantic Cod fish: skin tissue, single layer sheet. Predicate is Porcine skin and connective tissue, double layer sheet. "Different animal source, same anatomical tissue." Viral inactivation per ISO 22422 series.
    Morphology (H&E, SEM)Rich in collagen, porous, favoring cellular infiltration, preserved collagen structure.Both subject and predicate "rich in collagen and porous, therefore favoring cellular infiltration." SEM "shows equivalent preserved collagen structure." Cross section shows porous surfaces allowing tissue adherence and promoting cellular ingrowth.
    Cellular IngrowthFavorable cellular infiltration in vitro.Both materials "showed favorable cellular infiltration of fibroblasts after 14 days which is a key component for tissue augmentation and re-epithelization of defected keratinized tissue."
    Tensile StrengthComparable to predicate."Determined to be comparable to the predicate device measured by ultimate tensile strength."
    Heavy Metal AnalysisWithin ICH guidelines: Q3D Elemental Impurities.Limits of cadmium (Cd), lead (Pb), arsenic (As), and mercury (Hg) were "acceptable under the ICH guidelines."
    Stability (Simulated Physiological)Stable in artificial saliva buffer, comparable to predicate's effect on pH/conductivity.Buffer incubated with both products was "stable over time." Subject device "raises the pH slightly," predicate "decreases it slightly." Subject device "is structurally more stable than the predicate device since it dissolved slower than the predicate device at neutral pH 7."
    Suture Pull-Out StrengthMeets or exceeds predicate with 95% confidence."Meets or exceeds that of the predicate with a confidence of greater than 95%." Deemed equivalent for oral surgery and gingival augmentation.
    Pin Pull-Out StrengthExceeds predicate with 95% confidence."Exceeds that of the predicate at a confidence level of greater than 95%." Deemed equivalent for oral surgery and gingival augmentation.
    CompressionCompressive Peak-Load, Load-at-Break, Probe Penetration-at-Break, Energy-to-Break meet or exceed predicate with 95% confidence."Meet or exceed those of the predicate device, with a confidence of greater than 95%." Deemed equivalent for oral surgery and gingival augmentation.
    User Evaluation (Cutting/Shaping)Favorable usability, substantially equivalent to predicate.Evaluated by four dental clinicians using questionnaire. "Showed a favorable usability that was substantially equivalent to the predicate for cutting and shaping the device for use as a dental barrier membrane." Assessed ease of placement, stability, robustness, handling satisfaction.
    Animal (Canine) Study - SafetyAcceptable safety profile for the subject device."All animals remained in good general health throughout the duration of the study and gross pathological findings suggest an acceptable safety profile for the subject device."
    Animal (Canine) Study - EfficacyNo statistically significant difference from predicate in key outcomes.For all parameters (KT thickness, KT length, root coverage, inflammation, membrane degradation), "no statistically significant difference was detected between the subject and predicate devices after 90 days of healing." Both devices "were able to increase keratinized tissue around teeth."
    Clinical Study - SafetyNo adverse effects or complications."There were no adverse effects or complications during the duration of the clinical study."
    Clinical Study - EfficacyIncreased KT, similar to predicate's published results, long-term stability."An increased in the in the width of the KT was noticed for all treated sites. This average gain has been reported as adequate to maintain long-term periodontal health." "The device showed stability as the KT slowly replaced the piscine xenograft." Results "similar to published results reported for the predicate device."

    2. Sample sizes used for the test set and the data provenance:

    • Bench Testing: "representative products" used for various tests. Specific numbers for each test are not provided, only statements like "with a confidence of greater than 95%" (implying statistical analysis on a certain sample size).
    • Animal Study: 10 canines (3 terminated at 30 days, 3 at 60 days, 4 at 90 days). Data provenance: Preclinical testing in canines (implied prospective). No country of origin specified.
    • Clinical Study: 6 human individuals. Data provenance: Open label, noncomparative study (prospective). No country of origin specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Bench Testing (User Evaluation): Four "dental clinicians" evaluated the device for cutting and shaping. Their specific qualifications (e.g., years of experience, specialization) are not detailed.
    • Animal Study: Outcomes included "Histomorphometric measurements" and "Histological analysis." This implies expert pathologists/histologists reviewed the samples, but the number and qualifications are not specified.
    • Clinical Study: Ground truth was the observation of clinical outcomes (KT increase) by clinicians. The number and qualifications of the clinicians performing assessments in the clinical study are not specified beyond the study being "clinical."

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • No specific adjudication method is mentioned for any of the studies (bench, animal, or clinical). For a 510(k) submission focused on equivalence, such detailed adjudication methods are less commonly emphasized compared to novel device approvals.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a physical gingival graft, not an AI or imaging diagnostic device. Therefore, MRMC studies and the concept of human readers improving with AI assistance are not relevant to this product.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a physical gingival graft. There is no "algorithm" or standalone software component. Its performance is inherent to its physical properties and biological interaction.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Bench Testing: Laboratory measurements (tensile strength, compression, etc.), and subjective "user evaluation" feedback from dental clinicians.
    • Animal Study: Histomorphometric measurements (objective), histological analysis (pathology, expert interpretation), in vivo and postmortem measurements by periodontal probe.
    • Clinical Study: Clinical observations of "increased in the width of the KT" and overall "stability" reported by the study investigators/clinicians as outcomes data.

    8. The sample size for the training set:

    • Not applicable. This is a physical medical device, not a machine learning or AI algorithm, so there is no concept of a "training set" in the traditional sense for algorithm development. The "training" here would be the manufacturing process and quality control methods to ensure consistent device properties.

    9. How the ground truth for the training set was established:

    • Not applicable for the same reason as point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190528

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-07-10

    (128 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MariGen Wound Extra is indicated for the management of wounds, including:

    • Partial and full-thickness wounds
    • Pressure ulcers
    • Venous ulcers
    • Chronic vascular ulcers
    • Diabetic ulcers
    • Trauma wounds (second degree burn, abrasions, lacerations, skin tears),
    • Surgical wounds (donor sites/grafts, post-Mohs surgery, post laser surgery, podiatric, wound dehiscence),
    • Draining wounds.
    Device Description

    The subject device is processed fish dermal matrix composed of fish collagen and is supplied as a sterile intact, or meshed sheet ranging in sizes up to 20 x 30 cm. The subject device is obtained from fish skin via standardized controlled GMP manufacturing process and supplied in terminally sterile packaging. The subject device is biocompatible, pliable, and non-cross linked.
    The device is intended for single use only.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA, evaluating Kerecis Limited's MariGen Wound Extra device. The core of this document is to establish "substantial equivalence" to a predicate device, rather than to prove new performance criteria through a study involving AI. Therefore, the information requested about acceptance criteria and a study proving device performance, especially related to AI, is not present in the provided text.

    The document states that "The subject device is identical to the predicate device apart from being offered in sizes up to 600cm²." This is a key statement explaining why extensive new performance studies (like those typically associated with AI devices) were not required. The approval is based on the device being a larger version of an already approved product made of the same material.

    Here's how the requested information relates to the provided text:

    1. A table of acceptance criteria and the reported device performance:

      • Not Applicable. This document does not describe acceptance criteria for a new performance study like an AI model. Instead, it aims to show substantial equivalence to an existing cleared device. The "performance data" mentioned (elemental impurities and chemical residual analysis) are for basic safety and material characterization, not for clinical performance demonstration as would be expected for an AI device. The table provided is a "Summary Table of Substantial Equivalence," comparing features like product codes, intended use, indications, resource origin, tissue resource/scaffold base, nominal sizes, presentation, sterilization, and shelf life between the subject and predicate devices. The acceptance criteria for each of these features is "Equivalent," meaning the subject device needs to be essentially the same or functionally similar to the predicate.
    2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

      • Not Applicable. No test set or data provenance for a performance study (as would be done for an AI device) is described. The rationale for approval is similarity to an existing device.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

      • Not Applicable. Ground truth establishment by experts for a test set is not described as part of this submission, as it is not an AI device or a device requiring a de novo clinical performance study against expert reads.
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not Applicable. No adjudication method is mentioned as there is no clinical performance test set described.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not Applicable. This device is a wound dressing, not an AI-assisted diagnostic or treatment device. Therefore, no MRMC study or AI assistance evaluation was performed or is relevant to this submission.
    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Not Applicable. This is a physical wound dressing, not a software algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not Applicable. The "ground truth" here is the established safety and efficacy of the predicate device, which the subject device is deemed substantially equivalent to due to material and functional similarity. There's no new "ground truth" established for clinical performance via a new study.
    8. The sample size for the training set:

      • Not Applicable. There is no AI model or training set described.
    9. How the ground truth for the training set was established:

      • Not Applicable. There is no AI model or training set described.

    In summary, the provided FDA 510(k) document for K190528, MariGen Wound Extra, is for a physical wound dressing and does not involve AI or new clinical performance studies to prove its efficacy. Its clearance is based on its substantial equivalence to a previously cleared predicate device (MariGen Wound) and a reference device (SecureMesh), primarily due to sharing the same material and functional design, with the only significant difference being larger available sizes.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153364

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-08-19

    (273 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    0 - 65
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Kerecis SecureMesh is intended for use as a prosthesis when staple line reinforcement is needed in surgical repair of soft tissue deficiencies using surgical staplers.

    Kerecis SecureMesh can be used for reinforcement of staple lines during:

    • Bariatric surgical procedures
    • Colorectal and Colon surgical procedures
    • Gastric, small bowels and mesentery procedures
    • Lung and bronchus resections
    Device Description

    Subject Device is formed of intact acellular fish skin and is intended to be used as a staple line buttress with surgical staplers. The buttress material is supplied in a double pouch sterile packaging (EtO), where a pair of buttresses is provided on a pair of carriers to be used by a healthcare professional, to transfer the buttresses to the opposed jaws of a surgical stapler in the operating theatre prior to a stapling procedure. One pair of carrier and a buttress is used for each stapler firing.

    AI/ML Overview

    The provided text is a 510(k) summary for the Kerecis SecureMesh, a surgical mesh. This document describes the device, its intended use, comparison with predicate devices, and performance testing to demonstrate substantial equivalence for regulatory clearance.

    However, the questions you've asked are typically relevant to the development and validation of an AI/Machine Learning medical device, particularly concerning diagnostic or interpretative capabilities. The Kerecis SecureMesh is a physical surgical implant, not a software or AI-driven diagnostic tool. Therefore, the concepts of "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of AI performance metrics (like sensitivity, specificity, MRMC studies, ground truth establishment by experts, training/test sets) do not apply to this medical device.

    The document describes performance testing for a surgical mesh, which includes:

    1. Bench Testing:

      • Biocompatibility (by leveraging prior testing on a similar product).
      • Viral-inactivation evaluation (also leveraged from similar product).
      • Tensile strength comparison with predicate.
      • Usability (loading time, transfer to stapler, ease of use, force to fire, slippage).
      • Performance (leak at staple line, leak pressure, seal burst).
    2. Animal Studies:

      • An acute porcine ex-vivo study comparing the Subject Device to a predicate device and un-buttressed stapling (negative control). The conclusion was "no significant difference."
    3. Clinical Studies:

      • "No clinical testing was included in this submission." This means no human clinical trials were performed for the 510(k) clearance based on substantial equivalence to predicate devices and the non-clinical testing.

    Given the nature of the device and the provided text, I cannot answer the questions about AI/ML device validation. The "acceptance criteria" for this device would be established engineering and performance specifications, and the "proof" is the bench and animal testing demonstrating equivalence to a legally marketed predicate device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2