Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K242170

    Validate with FDA (Live)

    Date Cleared
    2025-04-18

    (268 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    K-ASSAY® CRP (Ver.2) is intended to be used for the quantitative determination of C-reactive protein (CRP) in human serum and plasma (potassium-EDTA or lithium-heparin) by immunoturbidimetric assay. Measurement of CRP aids in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. FOR IN VITRO DIAGNOSTIC USE.

    Device Description

    The K-ASSAY® CRP (Ver.2) assay quantifies C-reactive protein based on immunoturbidimetric assay. The reagent uses latex combined with goat polyclonal antibody specific to human CRP. By adsorbing CRP in the sample to the surface of the latex particles and reacting it with the anti-CRP antibody, specific aggregation corresponding to the CRP concentration occurs. Since the absorbance of the reaction changes in proportion to the amount of aggregation, the concentration of CRP in the sample is determined based on the calibration curve prepared using a standard of known CRP concentrations. The K-ASSAY® CRP (Ver.2) assay can be run using a chemistry analyzer. 6 levels of calibrators from the K-ASSAY® CRP Calibrator (Ver.2) calibrators are used for quantifying the levels of CRP present in the patient's sample.

    AI/ML Overview

    This document describes the FDA 510(k) clearance for the K-ASSAY CRP (Ver.2) IVD device. This device is an in-vitro diagnostic test system, which means it analyzes biological samples in a lab setting rather than directly interacting with a patient.

    Therefore, the concepts of "human readers," "AI assistance," "effect size," "multi reader multi case (MRMC) comparative effectiveness study," and "standalone (i.e. algorithm only without human-in-the loop performance)" are not applicable in this context. These terms are typically used for medical imaging AI/ML devices where human interpretation is involved.

    For this in-vitro diagnostic device, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are related to its analytical performance characteristics when compared to a predicate device, and the reported device performance refers to the results of these analytical studies.

    Here's the breakdown of the information provided within the scope of this IVD device:

    1. Table of Acceptance Criteria and Reported Device Performance

    For an IVD device like the K-ASSAY® CRP (Ver.2), the acceptance criteria are generally established by demonstrating equivalent or superior analytical performance compared to a legally marketed predicate device, and meeting established CLSI guidelines for accuracy, precision, linearity, and interference. The "reported device performance" refers to the actual study results for these characteristics.

    Acceptance Criteria (Implicit from predicate comparison and CLSI guidelines)Reported Device Performance (K-ASSAY® CRP (Ver.2))
    Intended Use Equivalence: Quantitative determination of CRP in human serum and plasma for detection/evaluation of infection, tissue injury, inflammatory disorders.K-ASSAY® CRP (Ver.2) is intended for the quantitative determination of CRP in human serum and plasma (potassium-EDTA or lithium-heparin) by immunoturbidimetric assay. Measurement of CRP aids in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases.
    Methodology Equivalence: Latex-enhanced immunoturbidimetric assay.Latex-enhanced (immuno)turbidimetric assay.
    Calibration Equivalence/Validation: Appropriate calibrator levels for the intended range.6 levels of calibrators (0.0, 10.0, 50.0, 150.0, 300.0 and 400.0 mg/L).
    Assay Range (Comparable to predicate): Demonstrates clinically relevant measuring range.Claimed Assay Range: 5.0 - 400.0 mg/L. (Predicate: 0.2 – 480 mg/L)
    Precision: Demonstrate acceptable repeatability, within-run, between-run, within-day, between-day, within-laboratory, between-site, and reproducibility CVs as per CLSI EP05-A3 guidelines.Single Site Precision (Combined Data From 3 Lots):- Within-Run CV: 0.8% to 2.2%- Between-Run CV: 0.0% to 0.6%- Within-Lot CV: 0.9% to 1.9%- Between-Lot CV: 0.0% to 1.2%- Total CV: 0.9% to 2.3%Multisite Precision (Combined Data From 3 Analyzers):- Repeatability CV: 0.5% to 1.5%- Between-Run CV: 0.0% to 1.1%- Between-Day CV: 0.6% to 1.8%- Between-Site CV: 0.9% to 1.8%- Reproducibility CV: 1.1% to 2.5%
    Interference: No significant interference from common endogenous and exogenous substances (recovery within 10% of initial value).Endogenous Substances (up to listed concentrations, no significant interference): Bilirubin C (40 mg/dL), Bilirubin F (40 mg/dL), Cholesterol (300 mg/dL), Hemoglobin (1,000 mg/dL), Intralipid (500 mg/dL), Rheumatoid Factor (1,000 IU/mL), Triglycerides (1,000 mg/dL).Exogenous Substances (up to listed concentrations, no significant interference): Acetaminophen (1.5 mM), Amoxicillin (400 µmol/L), Aspirin (3.6 mM), Cephalexin (360 µmol/L), Fluconazole (480 µmol/L), Ibuprofen (2.5 mg/dL), Methotrexate (1,400 µmol/L), Prednisolone (2 µmol/L), Vitamin C (500 mg/L).
    Method Comparison: Strong correlation and minimal bias against the predicate device.Regression Equation: y = 1.005x - 0.002, r = 0.999 (n = 175 clinical native serum samples), where y = K-ASSAY® CRP (Ver.2), x = predicate device.
    Linearity: Demonstrates linearity across the claimed assay range.Regression Equation: y = 0.9709x - 1.095, r = 0.999 (tested range: 4.6 - 441.2 mg/L).
    Limit of Quantitation (LoQ): Clinically acceptable LoQ (within-laboratory precision ≤ 20% CV).Reported LoQ: 1.0 mg/L (highest observed across 3 reagent lots). Claimed LoQ: 5.0 mg/L.
    Matrix Comparison: No significant difference in results across different sample matrices (serum vs. plasma).K2-EDTA Plasma vs Serum: y = 1.007x - 0.141, r = 0.999Li-Heparin Plasma vs Serum: y = 0.972x + 0.074, r = 0.999
    Expected Values (Verification): Distribution in normal population consistent with clinical literature.168 normal U.S. serum samples tested; 4 out of 168 samples (>5.0 mg/L, 2.4%) consistent with literature (≤ 5 mg/L indicates apparently healthy).

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Method Comparison: n = 175 clinical native serum samples.
    • Linearity: The number of samples/dilutions used is not explicitly stated, but the tested range was 4.6 - 441.2 mg/L.
    • Limit of Quantitation (LoQ): Not specified in terms of sample size for the LoQ determination itself (typically involves low-concentration samples measured multiple times).
    • Matrix Comparison: 42 donor samples (each collected into 3 different tubes: serum, K2-EDTA plasma, Li-Heparin plasma).
    • Precision (Single-site): For each of the 7 samples/controls, N=240 individual measurements (2 runs/day x 2 replicates/run x 20 days x 3 lots).
    • Precision (Multisite): For each of the 7 samples/controls, N=75 individual measurements (1 run/day x 5 replicates/run x 5 days x 3 analyzers).
    • Interference: The number of unique samples tested for interference is not explicitly stated. Typically, a few samples (e.g., low, medium, high concentration) are spiked with interferents and compared to unspiked controls.
    • Expected Values: 168 normal serum samples.
    • Data Provenance: The document states for "Expected Values" that 168 normal serum samples were "taken from healthy individuals in the U.S." This indicates a U.S. origin for at least this specific study. For other studies (Method Comparison, Precision, Linearity, Interference, Matrix Comparison), the specific country of origin is not mentioned. All studies are retrospectively analyzed in the sense that they were completed performance validation studies submitted to FDA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    For an in-vitro diagnostic device like this CRP assay, the "ground truth" is established through highly accurate and precise reference methods or established predicate devices, adhering to rigorous analytical chemistry best practices and guidelines (e.g., CLSI standards). There isn't a panel of human "experts" like radiologists interpreting images. The "ground truth" is quantitative and objective, derived from reference measurements.

    In this case:

    • Method Comparison: The predicate device (K-ASSAY® CRP (3), K023828) serves as the comparator for method comparison, which represents the established "ground truth" for clinical samples.
    • Linearity, LoQ, Precision, Interference, Matrix Comparison: The "ground truth" for these analytical performance studies is established by rigorous laboratory protocols, highly calibrated equipment, reference materials, and adherence to CLSI guidelines. The performance is assessed against predefined statistical criteria rather than expert consensus on individual cases.
    • Expected Values: The ground truth comes from established clinical literature and verified normal ranges based on studies of healthy populations.

    4. Adjudication Method for the Test Set

    Not applicable for an IVD device. Adjudication methods (e.g., 2+1, 3+1) are used in studies where subjective human interpretation (e.g., image reading) requires consensus building for ground truth establishment. For quantitative IVD assays, the results are numerical and objectively measured; therefore, there's no need for adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. As explained above, this is an in-vitro diagnostic device, not an imaging AI/ML device involving human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in essence, all the analytical performance studies (Method Comparison, Linearity, LoQ, Precision, Interference, Matrix Comparison) are "standalone" in nature for an IVD. The device's performance is evaluated based solely on its ability to accurately and precisely measure CRP concentrations in samples, without any human interpretation of the measurement itself or the involvement of an "algorithm" in the AI/ML sense. The device directly processes the sample and outputs a quantitative result.

    7. The Type of Ground Truth Used

    The ground truth for this IVD device's performance studies is primarily based on:

    • Reference Method/Predicate Device Comparison: For the method comparison study, the predicate device serves as the reference against which the new device's measurements are compared.
    • Reference Materials and Calibrators: For linearity, LoQ, and precision studies, the ground truth for concentration values is established using certified reference materials and meticulously prepared calibrators with known concentrations.
    • Spiked Samples: For interference studies, known amounts of interfering substances are added to samples, and the true CRP concentration (before interference) serves as the baseline ground truth.
    • Clinical Literature/Established Norms: For "Expected Values," the ground truth is derived from widely accepted clinical ranges and population studies cited in the literature.

    8. The Sample Size for the Training Set

    This document describes a 510(k) submission for a traditional IVD device, not an AI/ML-based device. Therefore, there is no "training set" in the context of machine learning. The device's performance is based on its chemical and physical principles (latex-enhanced immunoturbidimetric assay) and validated through the analytical studies detailed.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this type of IVD device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192727

    Validate with FDA (Live)

    Date Cleared
    2020-05-20

    (236 days)

    Product Code
    Regulation Number
    866.5775
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The K-ASSAY® RF (Ver.2) assay is for the quantitative determination of human IgG rheumatoid factor antibodies in patient serum or plasma (citric acid, EDTA, or lithium heparin) based on immunoturbidimetric assay. The presence of IgG RF antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of rheumatord arthritis (RA). FOR IN VITRO DIAGNOSTIC USE.

    The K-ASSAY® RF Calibrator (Ver.2) is intended to be used to calibrate the K-ASSAY® RF (Ver.2) immunoturbidimetric assay. FOR IN VITRO DIAGNOSTIC USE.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) premarket notification letter from the FDA for a medical device called "K-ASSAY® RF (Ver.2), K-ASSAY® RF Calibrator (Ver.2)". This device is an in-vitro diagnostic test for quantitative determination of human IgG rheumatoid factor antibodies.

    The letter explicitly states that the device is a "Rheumatoid Factor Immunological Test System" (Regulation Name: 21 CFR 866.5775) and is a "Class II" device.

    Crucially, this document is an FDA clearance letter for a laboratory test (an in-vitro diagnostic device), not an AI/ML-based medical device that would have acceptance criteria based on performance metrics like sensitivity, specificity, or AUC, or involve human readers and their improvement with AI assistance.

    Therefore, most of the requested information regarding acceptance criteria, study design for AI/ML, human readers, ground truth establishment, etc., is not applicable to this type of device and will not be found in this document.

    The "acceptance criteria" for this type of device typically revolve around demonstrating substantial equivalence to a predicate device, and performance measures like precision, accuracy, linearity, measuring range, interference, and agreement studies for clinical performance. These are standard validation tests for IVD products, not AI performance metrics.

    In summary, based on the provided text, I cannot provide the requested information because it pertains to an AI/ML medical device, which the K-ASSAY® RF (Ver.2) is not.

    I can, however, extract the following relevant information from the document as it pertains to a medical device:

    1. Device Name: K-ASSAY® RF (Ver.2), K-ASSAY® RF Calibrator (Ver.2)
    2. Regulation Number: 21 CFR 866.5775
    3. Regulation Name: Rheumatoid Factor Immunological Test System
    4. Regulatory Class: Class II
    5. Indications for Use: The K-ASSAY® RF (Ver.2) assay is for the quantitative determination of human IgG rheumatoid factor antibodies in patient serum or plasma (citric acid, EDTA, or lithium heparin) based on immunoturbidimetric assay. The presence of IgG RF antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of rheumatoid arthritis (RA). FOR IN VITRO DIAGNOSTIC USE. The K-ASSAY® RF Calibrator (Ver.2) is intended to be used to calibrate the K-ASSAY® RF (Ver.2) immunoturbidimetric assay. FOR IN VITRO DIAGNOSTIC USE.
    6. Type of Use: Prescription Use
    7. Predicate Device: The letter states the device is "substantially equivalent" to legally marketed predicate devices, but the specific predicate device is not named in this letter.

    To reiterate, the questions about acceptance criteria, study design, sample sizes, experts, ground truth, MRMC studies, standalone performance, and training sets are not applicable to this medical device submission as described in the provided FDA 510(k) clearance letter.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121232

    Validate with FDA (Live)

    Date Cleared
    2012-05-24

    (30 days)

    Product Code
    Regulation Number
    866.5340
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The K-ASSAY® Ferritin (2nd Gen.) assay is an in vitro diagnostic reagent for the quantitative determination of ferritin (an iron-storing protein) in human serum and plasma by immunoturbidimetric assay on the Beckman AU680 analyzer. Measurements of ferritin aid in the diagnosis of diseases affecting iron overload and iron deficiency anemia. For in vitro diagnostic use.

    The K-ASSAY® Ferritin Calibrator (2nd Gen.) is an in vitro diagnostic reagent for calibration of the K-ASSAY® Ferritin (2nd Gen.) assay. For in vitro diagnostic use.

    Device Description

    The K-ASSAY® Ferritin (2nd Gen.) assay is a latex enhanced immuno-turbidimetric assay for the quantitative in vitro determination of ferritin levels in serum and plasma (EDTA and heparin) samples.

    The K-ASSAY® Ferritin (2nd Gen.) consists of two reagents. Reagent 1 contains HEPES buffer solution (50 mmol/L) and Reagent 2 contains HEPES buffer solution (50 mmol/L) and a solution of latex suspension with mouse monoclonal anti-human ferritin antibodies. Both reagents also contain less than 0:01 w/v% of sodium azide as a preservative.

    The K-ASSAY® Ferritin Calibrators (2nd Gen.) are liquid stable products consisting of a human serum matrix and known quantities of human ferritin at 6 levels ranging from 0 -1,000 ng/mL (0, 25, 250, 500, 750, 1,000 ng/mL). The calibrators also contain less than 0.1 w/v% of sodium azide as a preservative.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the studies that prove the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    PrecisionBetween-lot CV% < 1%Group 1: 0.99%, Group 2: 0.37%, Group 3: 0.39%, Group 4: 0.42% (All < 1%)
    LinearityLinear throughout the claimed measuring range of 5 - 1,000 ng/mLSlope: 0.9999 (95% CI: 0.9976 to 1.0022), Intercept: 0.1249 (95% CI: -0.9049 to 1.1546), Demonstrated to be linear with <10% deviation
    TraceabilityCalibrator values traceable to WHO 2nd International Ferritin Standard, lot 80/578Ferritin values assigned to calibrators are traceable to WHO 2nd International Ferritin Standard, lot 80/578
    Stability (Reagent)Unopened shelf life: 1 year (2-8°C)1 year (2-8°C)
    Open-vial stability: 1 month (2-8°C)1 month (2-8°C)
    On-board stability: 1 monthAt least 1 month
    Stability (Calibrator)Unopened shelf life: 1 year (2-8°C)1 year (2-8°C)
    Open-vial stability: 1 month (2-8°C)1 month (2-8°C)
    Detection Limits (LoB)Claimed Value 1.039 ng/mLReported result: 1.039 ng/mL (matches claimed value)
    Detection Limits (LoD)Claimed Value 1.825 ng/mLReported result: 1.825 ng/mL (matches claimed value)
    Detection Limits (LoQ)Claimed Value 5.0 ng/mLReported result: 5.0 ng/mL (matches claimed value)
    Analytical Specificity (Interference)No significant interference with specified substances within ±10% deviationBilirubin C (20 mg/dL), Bilirubin F (20 mg/dL), Hemoglobin (500 mg/dL), Rheumatoid Factor (1,000 IU/mL), Sodium Citrate (1,000 mg/dL), Sodium EDTA (500 mg/dL), Sodium Fluoride (1,000 mg/dL), Sodium Heparin (20 mg/dL), Turbidity (2,000 FTU) all showed no significant interference
    Analytical Specificity (Antigen Excess/Prozone)Measured value does not decrease to less than 1,000 ng/mL up to 80,000 ng/mLUp to 80,000 ng/mL, the measured value does not decrease to less than 1,000 ng/mL
    Method Comparison (vs. Predicate)Favorable regression analysis indicating substantial equivalencey = 1.0057x + 0.0968; Slope 95% CI: 1.0022 to 1.0092; Y-Intercept 95% CI: -1.1690 to 1.3625
    Matrix Comparison (Serum vs. EDTA Plasma)Favorable regression analysis indicating acceptabilityy = 1.0199x - 0.0813; Slope 95% CI: 1.0105 to 1.0294; Y-Intercept 95% CI: -4.1542 to 3.9916
    Matrix Comparison (Serum vs. Heparin Plasma)Favorable regression analysis indicating acceptabilityy = 0.9825x + 1.3632; Slope 95% CI: 0.9711 to 0.9939; Y-Intercept 95% CI: -2.8804 to 5.6067

    Study Details for Demonstrating Acceptance Criteria:

    1. Sample Size Used for the Test Set and Data Provenance:

    • Precision:
      • 6 samples (3 controls, 3 human serum samples) for within-device precision (assayed in duplicate twice a day for 20 days).
      • 50 patient samples (ranging from 5.7 - 884.7 ng/mL) for between-lot precision (tested in duplicate with 3 different lots of reagent).
      • Data provenance is not explicitly stated (e.g., country of origin, retrospective/prospective).
    • Linearity / Analytical Range:
      • 11 test concentrations prepared from analyte-spiked normal human serum samples diluted with analyte-depleted normal human serum.
      • Each sample run 4 times.
      • Data provenance not explicitly stated.
    • Detection Limits (LoB, LoD, LoQ):
      • LoB: 60 replicates of blank (analyte depleted human serum).
      • LoD: 6 samples diluted to concentrations from LoB to 4xLoB (1.05 - 6.16 ng/mL).
      • LoQ: 10 samples diluted to concentrations ranging from 1.0 to 10.0 ng/mL.
      • Data provenance not explicitly stated.
    • Analytical Specificity (Endogenous Interference):
      • Two serum samples containing ferritin at ~82 ng/mL and ~445 ng/mL.
      • Data provenance not explicitly stated.
    • Analytical Specificity (Antigen Excess / Prozone):
      • A set of 9 concentrations made by diluting analyte-spiked normal human serum (~80,000 ng/mL).
      • Data provenance not explicitly stated.
    • Method Comparison:
      • 54 unaltered human serum samples (ferritin concentrations ranging from 5.9 to 994.4 ng/mL).
      • Data provenance not explicitly stated.
    • Matrix Comparison (Serum vs. EDTA Plasma):
      • 65 matched serum and EDTA plasma patient samples (one sample spiked to extend range).
      • Data provenance not explicitly stated.
    • Matrix Comparison (Serum vs. Heparin Plasma):
      • 59 matched serum and heparin plasma patient samples (14 samples spiked to extend range).
      • Data provenance not explicitly stated.

    2. Number of Experts and Qualifications for Ground Truth:

    • Not applicable. This device is an in vitro diagnostic (IVD) assay for quantitative measurement of a biomarker. The "ground truth" for the performance studies is based on the inherent analytical properties of the samples themselves and comparison to a referenced standard or predicate device, rather than expert interpretation of images or clinical cases.

    3. Adjudication Method for the Test Set:

    • Not applicable. As an IVD assay, the performance is assessed through quantitative measurements and statistical analysis against predefined analytical criteria and/or predicate device results. There is no human adjudication process involved in establishing the "ground truth" for these analytical measurements.

    4. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

    • No. This is an in vitro diagnostic device for quantitative biomarker measurement. MRMC studies are typically used for imaging or diagnostic devices that involve human interpretation.

    5. Standalone (Algorithm Only) Performance:

    • Yes, implicitly. All described performance characteristics (precision, linearity, detection limits, analytical specificity, method comparison, matrix comparison) reflect the standalone performance of the K-ASSAY® Ferritin (2nd Gen.) assay on the Beckman/Olympus AU680 analyzer, without direct "human-in-the-loop" interaction for interpreting the results in these analytical performance studies. The output is a quantitative value, not an interpretation requiring a human reader.

    6. Type of Ground Truth Used:

    • Referenced Standards and Analytical Measurements:
      • For Traceability, the ground truth for calibrator values is established against the WHO 2nd International Ferritin Standard, lot 80/578.
      • For Precision, Linearity, and Detection Limits, the ground truth is based on the actual measured concentrations of analytes in well-characterized samples (e.g., controls, diluted samples, blank samples) and the statistical models used to derive these parameters according to CLSI guidelines.
      • For Analytical Specificity, the ground truth for interference studies is the known concentration of the interferent added to samples and the observed effect on ferritin measurement.
      • For Method Comparison, the ground truth is the result obtained from the predicate device (K-ASSAY® Ferritin (2) from K050944).
      • For Matrix Comparison, the ground truth for serum, EDTA plasma, and heparin plasma is the measured ferritin concentration in matched samples.

    7. Sample Size for the Training Set:

    • Not explicitly stated/Not applicable in the conventional sense. This is a chemical/immunological assay, not a machine learning or AI-based device that would typically have a "training set" for an algorithm. The development of reagents and optimization of the assay protocol would involve numerous experiments and adjustments, but these are not referred to as statistical "training sets" in this context.

    8. How the Ground Truth for the Training Set Was Established:

    • Not applicable. As explained above, this device does not utilize a training set in the machine learning sense. The assay principle is based on immunoturbidimetry, where the signal generated directly correlates with the ferritin concentration, using a calibration curve established with known calibrator concentrations traceable to an international standard.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1