Search Filters

Search Results

Found 21 results

510(k) Data Aggregation

    K Number
    K202136

    Validate with FDA (Live)

    Device Name
    IDS Cortisol
    Date Cleared
    2021-04-13

    (256 days)

    Product Code
    Regulation Number
    862.1205
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDS Cortisol assay is an in vitro diagnostic device intended for the quantitative determination of cortisol in human serum and plasma on the IDS system. Results are to be used in conjunction with other clinical and laboratory data to assist clinicians in the diagnosis and treatment of disorders of the adrenal gland.

    Device Description

    The IDS Cortisol assay consists of a reagent cartridge. The reagent cartridge contains multiple reagents:

    • MPE1:Magnetic particles coated rat anti-mouse monoclonal antibody in a phosphate buffer with Proclin as preservative.
    • CONJ: Cortisol coupled with an acridinium ester derivative in = phosphate buffer with Proclin as a preservative.
    • mAb: Mouse anti-cortisol monoclonal antibody in phosphate buffer with Proclin as a preservative .;
    • BUF: HEPES buffer containing Proclin as preservative .
    AI/ML Overview

    This document describes the analytical performance of the IDS Cortisol assay, an in vitro diagnostic device, and demonstrates its substantial equivalence to a predicate device (Roche Elecsys Cortisol II). The acceptance criteria and the study proving the device meets these criteria are detailed below.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this in vitro diagnostic device are primarily based on demonstrating analytical performance that is comparable to, or meets specified standards relative to, established laboratory methods and a predicate device.

    Performance CharacteristicAcceptance Criteria (from context/implied standard)Reported Device Performance (IDS Cortisol)
    PrecisionRepeatability (Within-run): Lower CV%Within-Run / Repeatability (n=80 per sample, 1 lot, 1 system): - 0.94 µg/dL: 7.8% CV - 1.84 µg/dL: 4.6% CV - 5.75 µg/dL: 2.4% CV - 13.06 µg/dL: 2.4% CV - 19.94 µg/dL: 1.8% CV - 44.63 µg/dL: 1.9% CV
    Intermediate Precision (Within-System/Total): Lower CV%Within-System (n=80 per sample, 1 lot, 1 system): - 0.94 µg/dL: 16.2% CV - 1.84 µg/dL: 10.9% CV - 5.75 µg/dL: 5.2% CV - 13.06 µg/dL: 3.9% CV - 19.94 µg/dL: 5.1% CV - 44.63 µg/dL: 4.2% CV Total (Combind 3 lots, 3 systems, n=240 per sample): - 0.88 µg/dL: 15.3% CV - 1.78 µg/dL: 10.1% CV - 5.75 µg/dL: 4.5% CV - 13.09 µg/dL: 3.3% CV - 20.22 µg/dL: 4.8% CV - 44.48 µg/dL: 5.0% CV
    Linearity/Reportable RangeData should demonstrate linearity across the claimed measuring range.Measuring Range: 0.59 to 45.00 µg/dL. Regression: Observed = 1.01 * Expected + 0.01 µg/dL; R²: 1.00 (Accepted based on R² close to 1 and slope ~1, intercept ~0)
    Detection Limits (LoB, LoD, LoQ)Specific quantifiable low limits.LoB: 0.10 µg/dL LoD: 0.24 µg/dL LoQ: 0.59 µg/dL
    Analytical Specificity (Interference)Non-significant bias (<10%) for tested interfering substances at specified concentrations.Interference (<10% bias): - Acetaminophen: 200 µg/mL - Bilirubin (Conj/Unconj): 40 mg/dL - Biotin: 6 µg/mL - Carbamazepine: 30 µg/mL - Haemoglobin: 500 mg/dL - HAMA: 1000 ng/mL - Ibuprofen: 500 µg/mL - Phenytoin: 50 µg/mL - RhF: 2000 IU/mL - Total protein: 12 g/dL - Triglycerides: 3000 mg/dL (All tested compounds found not to interfere significantly.)
    Method Comparison (vs. Predicate Device)Correlation coefficient (r) close to 1, slope close to 1, intercept close to 0, indicating strong agreement.vs. Roche Elecsys Cortisol II (K152227): - N: 194 samples - Slope: 1.06 (95% CI: 1.04 to 1.07) - Intercept: -0.10 µg/dL (95% CI: -0.39 to 0.04) - Correlation Coefficient (r): 0.99 (Demonstrates strong agreement with predicate)
    Matrix ComparisonSlopes close to 1 and intercepts close to 0, with high correlation coefficients, for various sample types vs. control.vs. Red Top Serum: - SST (n=45): Slope 1.02, Int. -0.08, r 1.00 - K2 EDTA (n=45): Slope 1.03, Int. -0.11, r 1.00 - K3 EDTA (n=45): Slope 1.02, Int. -0.10, r 1.00 - Lithium Heparin (n=45): Slope 1.00, Int. 0.15, r 1.00 - Sodium Heparin (n=45): Slope 1.01, Int. 0.00, r 1.00 (All demonstrate strong agreement across sample types)
    StabilityDemonstrates shelf-life stability.Shelf life: 12 months (based on accelerated stability studies).

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision Testing:
      • Per sample/concentration level (for one representative lot on one system): N=80 replicates.
      • Combined (3 lots on 3 systems): N=240 replicates per sample/concentration level.
      • Provenance: Not explicitly stated, but clinical laboratory studies typically use samples from diverse patient populations to cover the measurement range. The study mentions "human serum samples" and "low level samples."
    • Linearity Testing:
      • A high human serum sample and a low human serum sample were used, along with 12 evenly spaced dilutions.
      • Provenance: "high patient sample with a low patient sample."
    • Detection Limit Testing (LoB, LoD, LoQ):
      • 60 blank replicates and 13 low-level samples.
      • Provenance: Not explicitly stated, but based on typical laboratory practices.
    • Analytical Specificity (Interference):
      • Human serum samples with low and high cortisol concentrations.
      • For Rheumatoid Factor, a serum sample with low Cortisol and high Rheumatoid Factor was diluted 1:2 and 1:4. Each was assayed in duplicate.
      • For cross-reactivity, serum samples spiked with cross-reactants and unspiked controls were assayed in 26 replicates each.
    • Method Comparison (vs. Predicate Device):
      • N: 194 samples.
      • Provenance: Samples were "selected to represent a wide range of Cortisol concentrations" (0.64 to 44.66 ug/dL), implying patient samples. Country of origin not specified, but usually derived from laboratory settings.
    • Matrix Comparison:
      • N: 45 samples (36 native, 9 spiked or diluted) per matrix type.
      • Provenance: "human samples" to cover the range of 0.89 to 42.38 ug/dL.
    • Expected Values/Reference Range:
      • N: 307 apparently healthy donors.
      • Provenance: "serum samples collected from 307 apparently healthy donors" from 21 to 65 years of age. Details imply a prospective collection for this specific study, but geographical location not explicitly stated beyond "local population" for general reference ranges.

    All studies mentioned state they followed CLSI (Clinical and Laboratory Standards Institute) guidelines, which implies these were prospective analytical performance studies conducted under controlled laboratory conditions.


    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    For an in vitro diagnostic (IVD) device like the IDS Cortisol assay, "ground truth" for the test set is established through reference methods or highly accurate comparative assays, not typically by human expert consensus or adjudicated readings. The nature of this device is a quantitative immunoassay meant to measure a biomarker.

    • Ground Truth Establishment for Analytical Performance:

      • Traceability: The IDS Cortisol assay is traceable to the LC-MS/MS Candidate Reference Measurement Procedure (cRMP) Total Serum Cortisol. This is a highly accurate chemical analytical method, considered a "higher-order" reference method. The ground truth in this context is the quantitative value determined by this cRMP, often involving specialized laboratories and expert analytical chemists.
      • Method Comparison: The test device's performance is compared against the Roche Elecsys Cortisol II (K152227), a previously cleared and marketed predicate device. The values obtained from the predicate device serve as the comparative "truth" for demonstrating substantial equivalence.
      • Precision, Linearity, Detection Limits, Analytical Specificity, Matrix Comparison: These performance characteristics are evaluated against pre-defined statistical criteria (e.g., CV thresholds, R² values, bias limits) using controlled samples, calibrators, and reference materials. The "truth" is the known concentration or behavior of these materials, verified through standard laboratory practices and reference methods.
    • Number of Experts & Qualifications: Not applicable in the context of human expert adjudication for image interpretation. The "experts" would be the skilled laboratory personnel and analytical chemists performing the LC-MS/MS and predicate device measurements, as well as those defining the CLSI protocols for analytical testing.


    4. Adjudication Method for the Test Set

    Not applicable. This is an in vitro quantitative diagnostic device, not an imaging device requiring human-in-the-loop reading and adjudication.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    Not applicable. This is an in vitro quantitative diagnostic device, not an imaging device requiring human readers or MRMC studies.


    6. If a Standalone Performance (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the entire analytical performance evaluation (precision, linearity, detection limits, interference, method comparison, and matrix comparison) represents the standalone performance of the IDS Cortisol assay as an algorithm (chemical reaction + instrument measurement and calculation) without human-in-the-loop intervention for result generation. The human role is in operating the instrument and interpreting the results, but the measurement itself is automated.


    7. The Type of Ground Truth Used

    The primary ground truth used for this quantitative in vitro diagnostic device is:

    • Reference Measurement Procedure: The IDS Cortisol assay is traceable to the LC-MS/MS Candidate Reference Measurement Procedure (cRMP) Total Serum Cortisol, which represents a highly accurate and standardized method for determining cortisol concentrations. This is a form of outcomes data in the sense of a definitive analytical outcome.
    • Predicate Device Comparison: The values generated by the Roche Elecsys Cortisol II assay serve as a comparative ground truth for demonstrating substantial equivalence.
    • Known Concentrations of Reference Materials/Spiked Samples: For analytical performance studies like precision, linearity, and detection limit, the ground truth is often established by using precisely prepared calibrators or samples with known, verified concentrations of the analyte.

    8. The Sample Size for the Training Set

    This document describes the analytical validation of a commercial diagnostic kit, not a machine learning algorithm that typically undergoes distinct "training" datasets. Therefore, there isn't a "training set" in the sense of data used to iteratively optimize an AI model.

    The "development" or "design" of the assay (reagents, methodology) would have involved extensive R&D and internal testing, which could be considered analogous to a training phase in a broader sense, but no specific sample size for such a phase is provided in this regulatory submission. The data presented are for the validation of the final device.


    9. How the Ground Truth for the Training Set Was Established

    As noted in point 8, there isn't a distinct "training set" in the context of an AI/ML algorithm as described in this document. The "ground truth" for the development of the assay itself would have been established through a combination of:

    • Biochemical principles: Understanding the cortisol molecule and its interactions with antibodies.
    • Standard laboratory methods: Using established analytical techniques (like spectrophotometry, chromatography, or existing cortisol assays) to characterize reagents and ensure proper reaction kinetics.
    • Reference materials: Utilizing internationally recognized reference materials and calibrators for initial assay calibration and optimization.
    • Iterative development and testing: Repeated internal testing with various concentrations of cortisol in different matrices to optimize assay components (antibodies, conjugates, buffers) and instrument parameters to achieve desired performance characteristics (sensitivity, specificity, dynamic range). This iterative process involves comparing results against known concentrations or a "gold standard" method (like LC-MS/MS) during development.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200475

    Validate with FDA (Live)

    Date Cleared
    2020-09-30

    (217 days)

    Product Code
    Regulation Number
    862.1050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDS-iSYS Ostase® BAP assay is an in vitro diagnostic device intended for the quantitative determination of bone-specific alkaline phosphatase (BAP), an indicator of osteoblastic activity, in human serum on the IDS system. Results are to be used in conjunction with other clinical and laboratory data to aid the clinician in the management of postmenopausal osteoporosis and Paget's disease.

    Device Description

    The IDS-iSYS Ostase® BAP assay consists of one reagent cartridge and one set of calibrators (CAL A & CAL B).

    The reagent cartridge contains multiple reagents:

    • MPM1 (Magnetic particles coated with streptavidin in a phosphate buffer with sodium azide as preservative);
    • Ab-BIOT Monoclonal anti-BAP labelled with biotin, in buffer containing horse serum with bovine and mouse proteins and sodium azide as a preservative (<0.1 %)
    • -SUBS (p-nitrophenyl phosphate in a stabilising buffer containing preservatives).

    Calibrators A and B are buffered bovine protein matrix containing human BAP with sodium azide as preservative (<0.1 %).

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a medical device called "IDS-iSYS Ostase® BAP." This document is primarily concerned with demonstrating the substantial equivalence of the new device to a predicate device, rather than focusing on the acceptance criteria and study proving performance for an AI/ML-based medical device.

    Therefore, the requested information regarding AI/ML-specific acceptance criteria and validation studies (like sample size for test set, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details) is not present in this document because it describes an in vitro diagnostic (IVD) test for quantitative determination of bone-specific alkaline phosphatase, not an AI/ML-driven imaging or diagnostic algorithm.

    However, I can extract the relevant performance characteristics that are analogous to "acceptance criteria" for this type of IVD device and the study data proving it meets those.


    Device Description

    The IDS-iSYS Ostase® BAP assay is an in vitro diagnostic device intended for the quantitative determination of bone-specific alkaline phosphatase (BAP), an indicator of osteoblastic activity, in human serum on the IDS system. Results are to be used in conjunction with other clinical and laboratory data to aid the clinician in the management of postmenopausal osteoporosis and Paget's disease.


    1. Table of Acceptance Criteria (Performance Characteristics) and Reported Device Performance

    For an in vitro diagnostic device, "acceptance criteria" are typically defined by various analytical performance characteristics that demonstrate the device's accuracy, precision, and reliability for its intended use. While explicit acceptance ranges are not always presented as a separate "criteria" table in a 510(k), the studies conducted implicitly aim to demonstrate performance within acceptable ranges for IVDs. The comparison to the predicate device and adherence to CLSI guidelines are key for demonstrating substantial equivalence.

    Here's a table summarizing the analytical performance characteristics and the reported device performance, which serve as the "proof" that the device meets the implied acceptance criteria for an IVD:

    Performance CharacteristicAcceptance Criteria (Implied / Comparator)Reported Device Performance (IDS-iSYS Ostase® BAP)
    PrecisionTypically aims for low %CV (Coefficient of Variation) within and between runs, indicating reproducibility. Compared to predicate: - Predicate Within Run: 2.6% to 6.5% (7.4 to 79.5 μg/L) - Predicate Between Run: 2% to 6.4% (8.4 to 81.1 µg/L)Repeatability (Within Run): From 1.7% to 2.8% in the concentration range 6.2 to 59.8 µg/L (N=80 data points for each sample, for one representative lot). From 1.7% to 2.8% for combined 3 lots across samples 1-10 (6.2-59.8 μg/L). Within Laboratory (Between Run/Total Precision): From 3.0% to 7.6% in the concentration range 6.2 to 59.8 µg/L (N=80 data points for each sample, for one representative lot). From 3.0% to 7.2% for combined 3 lots across samples 1-10 (6.2-59.8 μg/L). Overall, shows robust precision comparable or improved relative to the predicate.
    Linearity / Reportable RangeThe assay should demonstrate linearity across its claimed measuring range. Expected a high correlation coefficient (R²) close to 1.0. - Predicate Range: 0.7 – 90 µg/L.Linear Range: 0.9 to 78.5 µg/L. Measuring (Reportable) Range: 3 to 70 µg/L. Regression: Observed = 0.98 x (Expected) - 0.9 ng/mL. Regression coefficient R²: 1.00. The high R² demonstrates excellent linearity across the range.
    Detection Limits (LoB, LoD, LoQ)Low values for LoB (Limit of Blank), LoD (Limit of Detection), and LoQ (Limit of Quantitation) are desirable to demonstrate the ability to detect very low concentrations of the analyte. - Predicate: LoD 0.7 µg/L. Other values N/A.LoB (Limit of Blank): 0.3 µg/L LoD (Limit of Detection): 0.4 µg/L LoQ (Limit of Quantitation): 0.5 µg/L Demonstrates improved or comparable sensitivity to the predicate.
    Analytical Specificity (Interference)Bias due to common interfering substances should be non-significant, typically defined as <10% bias between test and control samples.Non-significant interference (<10% bias) observed for: Acetaminophen, Alendronate, Bilirubin (Conjugated & Unconjugated), Biotin, Calcium Chloride, Cholesterol, Estradiol, Etidronate, Haemoglobin, HAMA, Ibuprofen, Pamidronate, Progesterone, PTH 1-34, PTH 1-84, Raloxifene, Red Blood Cells, Rheumatoid Factor (RF), Risedronate, Salicylic Acid (Asprin), Salmon Calcitonin, Total Protein, Triglycerides, 25-hydroxyvitamin D. Tested at specified high concentrations (e.g., Bilirubin 40 mg/dL, Biotin 400 ng/mL, Haemoglobin 300 mg/dL etc.).
    Analytical Specificity (Cross-Reactivity)Low percent cross-reactivity with structurally similar substances is desired to ensure specificity to BAP.Low cross-reactivity observed for: - Liver ALP (745 µg/L spiked): 0.1% - Placental ALP (90 U/L spiked): 0.5% - Intestinal ALP (500 µg/L spiked): Undetectable Demonstrates high specificity to BAP.
    Method ComparisonStrong correlation and agreement with the legally marketed predicate device (Ostase® BAP EIA) is required for substantial equivalence. A slope close to 1.0 and an intercept close to 0, with a high correlation coefficient (r) close to 1.0, are expected. - Predicate comparison to Tandem-R Ostase: slope 1.02, intercept 0.28, r = 0.97.Comparison against Ostase® BAP EIA: - N: 150 samples - Slope: 0.99 (95% CI: 0.97 to 1.02) - Intercept: 0.17 µg/L (95% CI: -0.1 to 0.5) - Correlation Coefficient (r): 0.99 These results demonstrate excellent agreement and strong correlation, supporting substantial equivalence.
    StabilityThe device should maintain its performance over a defined shelf life.Shelf Life: 12 months, determined based on real-time studies.
    Expected Values / Reference RangeEstablished ranges for different populations. Consistency with established clinical understanding.Population-specific BAP Concentrations (µg/L) for IDS-iSYS Ostase® BAP: - Males (N=140): Mean 13.7, Median 13.0, Observed Range 7.9 to 23.5 - Pre-menopausal (N=140): Mean 11.5, Median 11.1, Observed Range 5.9 to 20.5 - Post-menopausal (N=139): Mean 15.7, Median 14.3, Observed Range 7.9 to 34.2 These values are consistent with the general understanding of BAP levels in these populations, though laboratories are advised to establish their own ranges.

    Study Details (Based on the Provided Document)

    As this is an IVD device, not an AI/ML system, many of the AI/ML-specific questions are not applicable. However, I will address what is implied or directly stated in the context of an IVD submission.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Precision/Reproducibility:
      • 10 native serum samples.
      • Each sample assayed 80 replicates (in duplicate, twice per day for 20 days) on one system (representative lot) and 240 replicates (combined from 3 lots, 3 systems).
      • Data provenance not explicitly stated (e.g., country of origin, retrospective/prospective), but implied to be laboratory-based analytical studies.
    • Linearity:
      • Not specified how many serum samples (a high and a low human serum sample were used to create 11 dilutions, but the N for testing these dilutions is not given).
    • Detection Limits (LoB, LoD, LoQ):
      • LoB: 60 replicates of blank per kit lot (total 180 over 3 lots).
      • LoD & LoQ: 10 low BAP concentration samples tested per kit lot (details on number of replicates not fully clear, mentions tested in duplicate for 5 days per lot).
    • Analytical Specificity (Interference/Cross-Reactivity):
      • For interference: Two serum samples at two different BAP concentrations were spiked with potential interferents. For each condition, 26 replicate assays were performed for spiked and control samples.
      • For cross-reactivity: Not explicitly stated, but substances were spiked into serum samples and measured.
    • Method Comparison:
      • N = 150 samples.
      • Samples were "selected to represent a wide range of BAP concentrations [3.0 to 67.6 ug/L]".
      • Data provenance not explicitly stated (e.g., country of origin, retrospective/prospective).
    • Expected Values/Reference Range:
      • N = 419 apparently healthy donors from the United States.
      • 140 males (35 to 75 years of age)
      • 140 pre-menopausal women (35 to 45 years of age)
      • 139 post-menopausal women (55 to 75 years of age).
      • This is a prospective collection of reference interval data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • Not applicable in the AI/ML sense. For IVD tests measuring a biomarker like BAP, the "ground truth" is typically the quantitative value obtained from a reference method or the assigned values of traceable calibrator materials. No human expert "adjudication" or "ground truth labeling" of the BAP level in a sample is described as a part of establishing test set ground truth; rather, it’s about the measured BAP concentration.
    • The calibrators for the new device are traceable to the predicate device via internal standards that were value-assigned using the predicate assay procedure. This method of traceability establishes the "truth" for calibration.

    4. Adjudication Method for the Test Set:

    • Not applicable. As above, for an IVD, the measurement itself is the "truth" within the context of the assay's analytical performance. No human adjudication is involved in determining the concentration of BAP in a blood sample for these types of studies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No. MRMC studies are specific to image-based diagnostic aids where multiple human readers' performance is evaluated and compared with and without AI assistance. This document describes an in vitro diagnostic assay that quantifies a substance in human serum, not an imaging device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • Yes, in essence. For an IVD, the "standalone" performance is the analytical performance of the assay itself – its precision, linearity, detection limits, specificity, and comparison to a predicate device. The performance characteristics described in Section 1 (Precision, Linearity, Detection Limits, Analytical Specificity, Method Comparison) represent the standalone performance of the IDS-iSYS Ostase® BAP assay. It is designed to provide a quantitative result without direct human "interpretation" of the assay's output in the way an AI would assist a radiologist.

    7. The Type of Ground Truth Used:

    • Analytical Ground Truth (for analytical performance): For precision, linearity, detection limits, and specificity, the "ground truth" refers to:
      • The known properties of control materials or spiked samples.
      • The measured values from samples used in reproducibility studies.
      • The established values/performance of the predicate device (for comparison studies).
    • Reference Intervals (for clinical context): For expected values, the "ground truth" is derived from a defined cohort of apparently healthy individuals, and ranges are statistically determined (e.g., 2.5th to 97.5th percentile).

    8. The Sample Size for the Training Set:

    • Not applicable (in the AI/ML sense). This device is an IVD assay, not an AI/ML model that requires training on a vast dataset. The "training" for an IVD involves assay development, reagent optimization, and establishing manufacturing controls, not statistical model training based on patient data.
    • The calibrators for the device are value-assigned using a method traceable to the predicate device. This process involves testing calibrators in multiple runs (minimum 20 assay runs on one system) to determine their assigned values. This could be considered analogous to "training" the assay system to accurately report concentrations based on known standards.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable (in the AI/ML sense). However, for an IVD, the "ground truth" for calibrator and control materials is established through:
      • Traceability: The IDS-iSYS Ostase® BAP kit calibrators are value assigned against in-house secondary standards (IRs). These IRs are themselves value assigned against the predicate device (Ostase® BAP EIA assay) using the predicate assay procedure. This establishes traceability to a previously cleared, legally marketed device.
      • Verification: The assigned kit calibrator values are then verified by running the assay on three different IDS systems and analyzing internal quality control (IQC) samples of known values across the range of the assay.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190121

    Validate with FDA (Live)

    Device Name
    IDS SHBG
    Date Cleared
    2019-06-17

    (143 days)

    Product Code
    Regulation Number
    862.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDS SHBG assay is an in vitro diagnostic device intended for the quantitative determination of SHBG in human serum or plasma on the IDS System. Results are to be used as an aid in the diagnosis of androgen disorders

    Device Description

    The IDS SHBG assay is an in vitro diagnostic device intended for the quantitative determination of sex hormone binding globulin (SHBG) in human serum and plasma on the IDS-iSYS Multi-Discipline Automated System. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the diagnosis of androgen disorders.

    The assay is based on chemiluminescence technology. 5 uL of patient sample or calibrators are incubated with biotinylated monoclonal anti-SHBG antibody, an acridinium labelled monoclonal anti-SHBG conjugate and streptavidin labelled magnetic particles. The magnetic particles are captured using a magnet and a wash step performed to remove any unbound analyte. Trigger reagents are added; the resulting light emitted by the acridinium label is directly proportional to the concentration of analyte in the original sample.

    The IDS SHBG assay is an in vitro diagnostic device consisting of ready to use reagents provided in individual compartments within the reagent cartridge.

    The reagent cartridge contains:

    • Magnetic particles magnetic particles coated with streptavidin in a phosphate buffer containing preservatives
    • -Biotin antibody - monoclonal anti-SHBG labelled with biotin in a buffer containing proteins and preservatives
    • Conjugate monoclonal anti-SHBG labelled with an acridinium ester derivative in a buffer containing proteins and preservatives The calibrators consist of:
    • Calibrators A and B are included in the assay kit. The calibrators consist of a human serum matrix with defined concentrations of SHBG and preservatives. Together with a lot specific master calibration curve, the calibrators will be used to perform adjustment of the master calibration curve.
    AI/ML Overview

    The provided document is a 510(k) summary for the IDS SHBG assay, an in vitro diagnostic device for the quantitative determination of Sex Hormone Binding Globulin (SHBG). The document details the device's performance characteristics and compares it to a predicate device to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Key Takeaway: This document describes the validation of an in vitro diagnostic (IVD) assay, not an AI/ML-based diagnostic device that typically involves human readers or image analysis. Therefore, many of the requested categories (e.g., number of experts, adjudication method, MRMC studies, human reader improvement with AI, standalone AI performance) are not applicable to this type of device and study. The ground truth in this context is established by reference methods or validated reference materials, typical for IVD assays.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes several analytical performance characteristics that serve as "acceptance criteria" for the IDS SHBG assay. These are primarily related to the accuracy, precision, limits of detection, and specificity of the assay.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    Precision (Reproducibility)Within-run and total precision (CV%) to be within acceptable limits for a quantitative assay.Within Run CV: 1.7% to 3.7% across various SHBG concentration levels (e.g., IQC 1: 1.7% at 5.57 nmol/L; CV 3: 3.7% at 201.67 nmol/L). Total CV: 3.2% to 5.0% across various SHBG concentration levels (e.g., IQC 3: 3.2% at 96.82 nmol/L; CTL3: 5.0% at 93.43 nmol/L). (Compared favorably to predicate which reported 2.5-3.8% within run and 3.1-6.5% total precision.)
    Linearity/Reportable RangeAssay to be linear over its claimed measuring range; accuracy demonstrated for automated dilution.Measuring Range: Linear from 1.60 to 180.00 nmol/L for serum and K2 EDTA plasma (R2 values of 0.999 and 0.998 respectively). Reportable Range: 0.30 to 720.00 nmol/L with automated 1:4 dilution for samples >180 nmol/L. Recovery (Automated Dilution): 87% to 100% when compared to predicate for samples in the 180-720 nmol/L range (Mean recovery: 93%).
    Detection Limits (LoB, LoD, LoQ)Limits should be sufficiently low to meet diagnostic needs.LoB: 0.01 nmol/L LoD: 0.15 nmol/L LoQ: 0.30 nmol/L
    Traceability of CalibratorCalibrator values must fall within specified acceptable ranges and internal quality controls within their respective ranges with defined precision.Calibrator A: 0.10 to 0.30 nmol/L, precision CV ≤ 11%. Verified through internal QC procedures. Calibrator B: 110.00 to 130.00 nmol/L, precision CV ≤ 8%. Verified through internal QC procedures. Internal QC controls: CV ≤ 11% for IQC1, ≤ 8% for IQC2 and IQC3. Results from Table 11 show these were met (e.g., IQC1 Total CV 4.1%, IQC2 Total CV 3.4%, IQC3 Total CV 3.2%).
    Analytical Specificity (Interference)No significant interference from common biological substances and exogenous compounds up to tested concentrations.No significant interference found for: - Triglycerides: up to 3000 mg/dL - Haemoglobin: up to 500 mg/dL - Bilirubin (conjugated/unconjugated): up to 40 mg/dL - Total Protein: up to 12 g/dL - Biotin: up to 6000 ng/mL (and 1500 ng/mL) - Rheumatoid Factor: up to 7000 IU/mL - HAMA: up to 3000 ng/mL - Cholesterol: up to 456 mg/dL - Various common drugs (e.g., Acetaminophen, Ibuprofen, Ascorbic acid, Creatinine, Dopamine, Tetracycline, Tolbutamide, Tolazamide, Uric Acid).
    Analytical Specificity (Cross-Reactivity)No significant cross-reactivity with structurally similar compounds or other biological substances up to tested concentrations.Low/No significant cross-reactivity observed (<1% typically) for: - AFP (-0.3% to 0.5%) - Thyroxin binding globulin (0.0% to -0.1%) - Transferrin (0.0%) - Cortisol (0.0%) - 11-deoxycortisol (0.0% to -0.1%) - 5a-dihydroxytestosterone (0.0%) - Testosterone (0.0%) - Fibrinogen (0.0%) - Corticosteroid binding globulin (0.0%) - Thyrotropin (TSH) (0.0%) - Plasminogen (0.0%) - Human IgA (0.0%) - Human IgG (0.0%) Notable exceptions: Estradiol (-7.3% to -3.2%) and Thyroglobulin (5.1% to 90.2%), where the higher value indicates specific interference for Thyroglobulin.
    Method ComparisonGood correlation and agreement with a predicate device.Correlation Coefficient (r): 0.989 (n=136 samples, range 2.54 - 172.12 nmol/L). Passing-Bablok Regression: Slope 0.9112, Intercept 0.1556 nmol/L.
    Matrix ComparisonPerformance across different sample matrices (serum, serum gel, K2 EDTA plasma) should be comparable.Good correlation: - Gel tube vs. Serum: r=0.999, Mean bias 0.7% - K2 EDTA vs. Serum: r=0.998, Mean bias -1.3% (n=69 samples, range 0.51 to 238.45 nmol/L)
    Expected Values/Reference RangeEstablished typical ranges for different populations.Males 21-49y: 11.47 – 58.07 nmol/L (n=165) Males >50y: 14.85 – 65.21 nmol/L (n=180) Premenopausal Females: 20.30 – 140.18 nmol/L (n=206) Postmenopausal Females: 11.30 – 127.31 nmol/L (n=120) (Based on 671 apparently healthy adults from the United States).

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision/Reproducibility: 14 serum-based samples at different SHBG concentration levels covering the assay range. Tested across 21 days for one kit lot. (Table 11 shows individual sample IDs, CTLs, CALBs, IQCs)
    • Linearity/Assay Reportable Range:
      • Linearity: A high human serum sample and a low human serum sample, plus 14 evenly spaced dilutions created by mixing high and low samples.
      • Automated Post-dilution Accuracy: 9 native samples with known SHBG concentrations obtained from the predicate device.
    • Traceability of Calibrator:
      • Value assignment of Internal Reference Standards (IRs): At least 24 runs using three iSYS instruments, 2 kit lots, 3 replicates for each run. Serial dilutions of WHO 2nd international standard 08/266 used.
      • Value assignment of kit calibrators: At least 20 runs (14 runs using previous IR lot, 6 runs using international standard) using one iSYS instrument, 2 replicates for each run.
      • Value assignment verification: Internal quality controls at 3 SHBG levels tested in five replicates in one run and on each of three different iSYS instruments.
      • Correlation study (2-point calibration vs. IS 08/266 curve): 189 samples (139 serum, 50 K2 EDTA plasmas). Tested in two replicates using three lots (MB1, MB2, MB3) on one iSYS instrument.
    • Detection Limit (LoB, LoD, LoQ):
      • LoB: LoB sample run in 12 replicates for each of 5 runs over 3 days, by one operator, on one different instrument for each of 3 manufacture batches (MB1, MB2B, MB3) = total 60 replicates per lot.
      • LoD: 7 LoD samples measured in duplicate. For each of 3 kit lots, 5 assays over 3 days by one operator on a different instrument = total 70 replicates per lot.
      • LoQ: Panel of 9 samples measured in singlicate two times per day. For each of 3 kit lots, 10 assays over 5 days by one operator on a different instrument = total 90 replicates per lot.
    • Analytical Specificity (Interference): Two serum samples (low and high SHBG conc.) spiked with potential interferents. Control samples were also run. Number of replicates for each specific test not detailed but implied to be sufficient for statistical comparison.
    • Analytical Specificity (Cross-Reactivity): Low and high SHBG samples spiked with various cross-reactants. Number of replicates for each specific test not detailed.
    • Method Comparison: 136 samples, selected to represent a wide range of SHBG concentrations (2.54 - 172.12 nmol/L).
    • Matrix Comparison: 69 samples (68 native, 1 diluted) covering a range of 0.51 to 238.45 nmol/L.
    • Expected Values/Reference Range: 671 serum samples from apparently healthy adults (21-77 years old). Data Provenance: From the United States. Retrospective/Prospective: Not explicitly stated, but typically such studies for establishing reference ranges are retrospective collections of banked samples or a prospective study designed to collect samples from a healthy population.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not Applicable (N/A).

    This is an in vitro diagnostic (IVD) assay quantifying a biomarker (SHBG). "Ground truth" for an IVD assay is established through:

    • Reference Methods: Such as the WHO 2nd international standard 08/266 for SHBG, against which the calibrators are standardized.
    • Known Concentrations: Use of accurately prepared standards, spiked samples, or samples extensively characterized by reference methods.
    • Clinical Data: For expected values, SHBG concentrations are measured in samples from defined healthy populations.

    Therefore, the concept of "experts establishing ground truth" in the way it applies to image interpretation or AI-assisted diagnostics (e.g., radiologists labeling images) is not relevant here. The ground truth is analytical and based on metrological traceability to international standards.


    4. Adjudication Method for the Test Set

    N/A.

    Adjudication methods (like 2+1, 3+1) are relevant for subjective interpretations (e.g., radiology reads) where discrepancies between readers need to be resolved to establish a definitive ground truth. For a quantitative IVD assay like IDS SHBG, the "reading" is a numerical output from the instrument based on chemical reactions. Accuracy is determined by comparison to reference materials or established methods, not by human adjudication of interpretations.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A.

    This is an IVD assay, not an AI/ML diagnostic for human interpretation. No MRMC study or human reader improvement with AI assistance is applicable.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    N/A.

    This is not an AI/ML algorithm. The "device" is a fully automated immunoassay system (IDS-iSYS Multi-Discipline Automated System) that performs the biochemical analysis and reports a quantitative result without human "interpretation" of a signal beyond reading the numerical output. Its performance is evaluated as a standalone analytical system.


    7. The Type of Ground Truth Used

    The ground truth for the analytical performance studies (precision, linearity, detection limits, specificity, method comparison) is based on:

    • International Reference Standards: Specifically, the WHO 2nd international standard for SHBG (IS 08/266) for traceability and calibration. This is the primary reference.
    • Known Concentrations/Spiked Samples: Samples with defined, known concentrations of SHBG or interferents, prepared in a laboratory setting.
    • Comparison to a Legally Marketed Predicate Device: The Siemens ADVIA Centaur SHBG assay (K151986) was used as a comparative method to assess agreement and accuracy (e.g., for method comparison and automated dilution accuracy).
    • Healthy Population Data: For expected values/reference ranges, a large cohort of apparently healthy individuals were tested to establish population-specific normal ranges.

    8. The Sample Size for the Training Set

    N/A (for AI/ML 'training set' in the traditional sense).

    For an IVD assay, the equivalent of a "training set" would be the samples and calibrators used during the assay's development and optimization phases to set parameters, establish reagent formulations, and fine-tune the system. This information is typically proprietary development data and is not explicitly detailed as a 'training set' in 510(k) summaries, which focus on the final validation/test data.

    The closest analogous "training" or "calibration" process mentioned is the traceability and value assignment of calibrators:

    • Value assignment of secondary standards (IRs) involves at least 24 runs (using 3 instruments, 2 kit lots, 3 replicates) comparing to the WHO international standard.
    • Value assignment of IDS SHBG kit calibrators A and B involves at least 20 runs (1 instrument, 2 replicates) using the secondary standards and IS-08/266.
      These processes are used to establish the calibration curve for the assay.

    9. How the Ground Truth for the Training Set Was Established

    N/A (for AI/ML 'training set').

    As explained in point 8, the concept of a "training set" for an IVD assay is different from that for AI/ML. The "ground truth" for establishing the calibration and parameters of the assay is based on:

    • Metrological traceability to the WHO 2nd international standard for SHBG (IS 08/266). This involved serial dilutions of the international standard in SHBG-depleted human serum to create reference points.
    • Use of internal reference calibrators (IRs) that were themselves value-assigned against the WHO standard.

    The goal is that the assay's measurements accurately reflect the true concentration of SHBG as defined by an international reference.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161158

    Validate with FDA (Live)

    Date Cleared
    2017-01-31

    (281 days)

    Product Code
    Regulation Number
    862.1545
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDS-iSYS Intact PTH+ assay is an in vitro diagnostic device intended for the quantitative determination of intact PTH in human serum or plasma on the IDS-iSYS Multi-Discipline Automated System. Results are to be used in the differential diagnosis of hypercalcemia and hypocalcemia resulting from disorders of calcium metabolism.

    Device Description

    The IDS-iSYS Intact PTH" assay is based on chemiluminescence technology. 100 uL of patient sample is incubated with a biotinylated polyclonal anti-PTH (39-84) antibody and an acridinium labelled PTH (13-34) antibody. Streptavidin labelled magnetic particles are added prior to a second incubation step. The magnetic particles are captured using a magnet and a wash step performed to remove any unbound analyte. Trigger reagents are added; the resulting light emitted by the acridinium label is directly proportional to the concentration of Intact PTH in the original sample.

    IDS-iSYS Intact PTH reagent kit consists of one (1) Immunoassay Cartridge, two (2) vials each of 2 concentration Calibrator levels and a mini-CD containing the Instructions For Use (IFU), CRY files and Certificate of Analysis.

    IDS-iSYS Intact PTHN Cartridge, sufficient for 100 tests, consists of reagents provided in individual compartment within a plastic container called the Cartridge. The IDS-iSYS Intact PTHN Cartridge contains the following ready to use reagents:

    • · Magnetic particles coated with streptavidin in a phosphate buffer containing sodium azide as preservative (<0.1%), 1 bottle.
    • Goat polyclonal antibody against 13-34 PTH labelled with an acridinium ester derivative, in buffer containing goat serum with sodium azide as preservative (<0.1%), 1 bottle.
    • Goat polyclonal antibody against 39-84 PTH 1abelled with biotin, in buffer containing bovine and goat proteins with sodium azide as preservative (<0.1%), 1 bottle.

    IDS-iSYS Intact PTH Calibrators are included in the reagent kit.

    • · Calibrator A: Two vials of lyophilized porcine serum matrix buffer containing low level PTH and sodium azide as preservative >1% (w/w%).
    • · Calibrator B: Two vials of lyophilized porcine serum matrix buffer containing high level PTH and sodium azide as preservative >1% (w/w%).

    The submission is due to a new source of antibody for the assay.

    AI/ML Overview

    This document describes the performance characteristics of the IDS-iSYS Intact PTHN assay, an in vitro diagnostic device for quantitative determination of intact PTH.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as distinct numerical targets but are implicitly defined by the reported performance and the comparison to the predicate device. The information below summarizes the performance characteristics provided, which in a regulatory context, demonstrate the device meets its intended use and is substantially equivalent to the predicate.

    Performance CharacteristicPredicate Device (K103325) PerformanceCandidate Device (IDS-iSYS Intact PTHN) PerformanceImplied Acceptance Criteria (Demonstrated)
    Method Comparison (against a commercially available method)N/An=312 samples; Slope: 1.02 (95% CI: 0.99 to 1.04); Intercept: -0.7 pg/mL (95% CI: -5.4 to 4.1); Correlation coefficient (r): 1.00Demonstrate strong correlation and agreement with a commercially available quantitative chemiluminescence Intact PTH assay (e.g., r close to 1, slope close to 1, intercept close to 0) as per CLSI EP-9A2 guidelines.
    Sample Matrix Comparison (vs. K2 EDTA plasma)N/A
    - Serum (Red Top)N/An=52; Slope: 0.94 (95% CI: 0.92 to 0.97); Intercept: 2.55 (95% CI: 0.86 to 3.16); r: 1.00Demonstrate acceptable agreement across different sample matrices (serum, lithium heparin plasma) compared to the control (K2 EDTA plasma) as per CLSI EP9-A3 guidelines. Slopes close to 1 and intercepts close to 0 with good correlation coefficients.
    - Serum (SST)N/An=52; Slope: 0.93 (95% CI: 0.91 to 0.96); Intercept: 2.38 (95% CI: 1.25 to 3.15); r: 1.00
    - Lithium HeparinN/An=52; Slope: 0.98 (95% CI: 0.95 to 0.99); Intercept: 0.42 (95% CI: -0.43 to 1.63); r: 1.00
    Reference Interval11.5 to 78.4 pg/mL10.3 to 80.5 pg/mLEstablish a 95% reference interval for a healthy population, consistent with clinical expectations and CLSI C28-A3 guidelines. The new range should be comparable to the predicate.
    Sensitivity
    - Limit of Blank (LoB)1.2 pg/mL0.9 pg/mLDemonstrate low limits of blank, detection, and quantitation, indicating the ability to detect very low concentrations of PTH, as per CLSI EP17-A guidelines. Should be comparable or better than the predicate.
    - Limit of Detection (LoD)2.5 pg/mL2.3 pg/mL
    - Limit of Quantitation (LoQ)4.5 pg/mL4.5 pg/mL
    Precision
    - Within-run CV%1.1% to 6.3%1.5% to 9.9%Demonstrate consistent and reproducible results across different concentration levels, within acceptable ranges for clinical diagnostic assays, as per CLSI EP-5A2 guidelines. Total CV should be within acceptable limits (e.g., <10-15% depending on concentration).
    - Total CV%4.1% to 8.2%1.8% to 9.9%
    Linearity
    - K2 EDTA Plasma$y = 1.002 x – 4.748; R^2 = 1.00$$y = 0.96 x − 0.1; R^2 = 1.00$Demonstrate linearity across the claimed measurement range (5 to 3500 pg/mL) with a regression coefficient ($R^2$) close to 1 and minimal deviation from expected values, as per CLSI EP-6A guidelines.
    - SerumN/A (Only K2 EDTA provided for predicate)$y = 1.02x – 0.2; R^2 = 1.00$
    High Dose Hook EffectNot observed up to 95,000 pg/mLNot observed up to 100,000 pg/mLNo significant hook effect should be observed within a clinically relevant and extended high concentration range to prevent falsely low results in patients with very high PTH levels. The demonstrated range should be comparable or wider than the predicate.
    Interferences (≤±10% bias)
    - Bilirubin, conjugatedN/A22 mg/dLNo significant interference (≤±10% bias) from common endogenous and exogenous interfering substances at clinically relevant concentrations, as per CLSI EP7-A2 guidelines. The tested concentrations should be comparable or improved compared to the predicate device.
    - Bilirubin, unconjugated20 mg/dL40 mg/dL
    - Biotin300 nmol/L300 nmol/L
    - CholesterolN/A395 mg/dL
    - HAMA1000 ng/mL1000 ng/mL
    - Hemoglobin250 mg/dL500 mg/dL
    - Rheumatoid Factor1500 IU/mL1836 IU/mL
    - Total ProteinN/A10 g/dL
    - Triglycerides3000 mg/dL3000 mg/dL
    - AcetaminophenN/A200 µg/mL
    - CarbamazepineN/A200 µg/mL
    - IbuprofenN/A500 µg/mL
    Cross-Reactivity (vs. PTH (1-84))
    - PTH (1-84)100%100%Demonstrate high specificity to intact PTH (1-84) and minimal cross-reactivity with PTH fragments or structurally similar proteins, improving upon the predicate where possible, as per CLSI EP7-A2 guidelines.
    - PTH (7-84)60%83.6%
    - PTH (1-34)0.5%<0.01%
    - PTH (39-84)Not detectable<0.01%
    - PTH (53-84)2%<0.01%
    - PTH (39-68)N/A<0.01%
    - PTH (44-68)2%<0.01%
    - Human Calcitonin4%<0.01%
    - CTX-1 (β CrossLaps)2%<0.01%
    - Osteocalcin2%<0.01%

    2. Sample Sizes and Data Provenance for Test Sets

    • Method Comparison:

      • Sample Size: 312 serum samples (291 native, 21 spiked).
      • Data Provenance: Not explicitly stated but clinical samples are generally considered prospective or retrospective from a clinical laboratory setting. The use of "commercially available quantitative chemiluminescence Intact PTH assay" as the comparator suggests real-world clinical samples.
    • Sample Matrix Comparison:

      • Sample Size: 52 matched samples (45 native, 7 spiked).
      • Data Provenance: Not explicitly stated, but clinical samples are typically used for this type of evaluation.
    • Reference Interval:

      • Sample Size: 129 individuals (67 males, 62 females; 21 to 89 years of age).
      • Data Provenance: K2 EDTA plasma samples collected from individuals with normal calcium, phosphate, and TSH values from the north, central and south regions of the United States. This is prospective or a well-characterized retrospective collection.
    • Sensitivity (LoB, LoD, LoQ):

      • Sample Size: 60 blanks and 10 low-level serum samples.
      • Data Provenance: Lab-prepared samples and low-level serum samples, likely from internal or commercially acquired biobanks.
    • Precision:

      • Sample Size: 7 samples (tested multiple times, N=80 for each sample).
      • Data Provenance: Lab-prepared controls and potentially pooled clinical samples ("Serum Sample," "EDTA Plasma Sample").
    • Linearity:

      • Sample Size: Not explicitly stated as a single number, but involved diluting high patient samples with low patient samples. Each sample was measured in 4 replicates.
      • Data Provenance: High and low intact PTH human K2 EDTA plasma and serum samples.
    • High Dose Hook Effect:

      • Sample Size: 3 samples (1 EDTA Plasma, 2 Serum) per spiked concentration.
      • Data Provenance: Spiked samples, likely from a laboratory setting using clinical matrices.
    • Interferences:

      • Sample Size: Two K2-EDTA samples (low and high Intact PTH levels) and two serum samples (different Intact PTH concentrations).
      • Data Provenance: Spiked samples, using clinical matrices (K2-EDTA and serum).
    • Cross-Reactivity:

      • Sample Size: A serum sample and a zero calibrator matrix.
      • Data Provenance: Spiked samples, using clinical serum matrix.

    3. Number of Experts and Qualifications for Ground Truth

    This device is an in vitro diagnostic assay, where the "ground truth" for performance characteristics such as sensitivity, precision, linearity, and interference is established by:

    • Reference Methods: Comparison against established, commercially available, and often FDA-cleared or gold-standard assays (e.g., the predicate device or a "commercially available quantitative chemiluminescence Intact PTH assay" for method comparison).
    • Analytical Standards/Spikes: Precisely known concentrations of analyte or interfering substances are used to create "spiked" samples, where the "true" concentration is known by design.
    • Clinical Laboratory Standard Institute (CLSI) Guidelines: Adherence to these guidelines (EP-9A2, EP9-A3, C28-A3, EP17-A, EP-5A2, EP-6A, EP7-A2) dictates the experimental design and statistical analysis which are considered "best practice" for establishing performance characteristics in IVDs.

    There are no "experts" in the sense of physicians or radiologists directly assessing images or patient data to establish ground truth for this type of analytical device performance. The "ground truth" is analytical and derived from established laboratory practices and reference methods.

    4. Adjudication Method for the Test Set

    Not applicable for this type of in vitro diagnostic device study. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies, particularly for diagnostic imaging or subjective clinical assessments, where consensus among multiple expert readers is needed to establish a ground truth for a disease state. For IVD analytical performance, the ground truth is either the value from a reference method or a gravimetrically/volumetrically determined "true" concentration.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This type of study is primarily relevant for diagnostic imaging AI algorithms where multiple human readers interpret cases with and without AI assistance. This document describes an in vitro diagnostic assay for a biochemical marker, not an imaging device or an AI application that assists human interpretation.

    6. Standalone Performance

    Yes, the studies described are standalone performance assessments of the IDS-iSYS Intact PTHN assay. The device is an automated system for measuring PTH levels, and the performance characteristics (method comparison, precision, linearity, etc.) directly evaluate the algorithm/assay's ability to accurately quantify PTH in patient samples. There is no human-in-the-loop component being evaluated in these tests; the device's output is directly compared to established analytical standards or reference methods.

    7. Type of Ground Truth Used

    • Reference Method Results: For method comparison, the results from a "commercially available quantitative chemiluminescence Intact PTH assay" served as the ground truth comparator.
    • Assigned Values of Controls/Standards: For precision, linearity, sensitivity, interference, and cross-reactivity studies, the ground truth was based on:
      • Known concentrations: Spiked samples with precisely known concentrations of analyte or interfering substances.
      • Expected values: Derived from dilution series or preparation of controls.
      • Blank matrices: For limit of blank determination.
    • Clinical Criteria: For the reference interval study, the ground truth for "normal" individuals was based on clinical criteria (normal calcium, phosphate, and TSH values).

    8. Sample Size for the Training Set

    The document does not specify a "training set" in the context of machine learning. This is an analytical medical device, not an AI/ML algorithm that is "trained" on data. The characteristics and parameters of the assay (e.g., antibodies, reagent concentrations, calibration curve algorithm) are developed and optimized during the product development phase, which is analogous to "training" in AI, but it's not described in terms of a distinct dataset.

    9. How Ground Truth for the Training Set Was Established

    As explained above, there isn't a "training set" in the common AI/ML sense. The "ground truth" for the development and optimization of the assay's chemical and instrumental parameters typically comes from:

    • Biochemical principles: Understanding of antigen-antibody interactions, chemiluminescence, and PTH biology.
    • Calibration materials: Use of international standards or highly characterized primary calibrators with assigned values.
    • Internal validation studies: Extensive testing during R&D to optimize reagent formulations, incubation times, and instrumental settings to achieve desired performance characteristics against known samples and reference methods.
    • Iterative development and testing: Performance data from pilot studies or earlier assay versions would inform adjustments to achieve acceptable analytical performance compared to the established medical consensus for PTH measurement.
    Ask a Question

    Ask a specific question about this device

    K Number
    K161831

    Validate with FDA (Live)

    Date Cleared
    2016-11-15

    (133 days)

    Product Code
    Regulation Number
    862.1825
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDS iSYS 25 VilD assay is intended for the quantitative determination of total 25-hydroxyvitamin D (125(OH)D) in hyman serum or plasma on the IDS iSYS Multi-Discipline Automated System. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in an adult population.

    The IDS-iSYS 25 VitD Control Set is used for quality control of the IDS-iSYS 25 VitD assay on the IDS-iSYS Multi-Discipline Automated System.

    Device Description

    The IDS-iSYS 25 VitDs reagent kit consists of one (1) Immunoassay Cartridge, one (1) vial each of 2 concentration Calibrator levels and a mini-CD containing the Instructions For Use (IFU), CRY files and Certificate of Analysis.

    IDS-iSYS 25 VitDS Cartridge, sufficient for 100 tests, consists of reagents provided in individual compartment within a plastic container called the Cartridge. IDS-iSYS 25 VitDS Cartridge contains the following ready to use reagents:

    • Magnetic particles coated with streptavidin in a phosphate . buffer containing sodium azide.
    • . Sodium hydroxide solution.
    • . 25(OH)D labelled with an acridinium ester derivative, in buffer containing bovine serum albumin with sodium azide.
    • . Anti-25(OH)D sheep polyclonal antibody labelled with an biotin, in buffer containing bovine, sheep and mouse proteins and sodium azide.
    • . Assay buffer containing proprietary displacing compounds, methanol and sodium azide.

    The IDS-iSYS 25 VitD® Calibrators are included in the reagent kit. The ready to use calibrators contains human serum buffer matrix with two defined concentrations of 25(OH)D and sodium azide as preservative (<0.1 %), 1 bottle per concentration level.

    The IDS-iSYS 25VitDe Control Set contains human serum in a buffer matrix with two defined concentrations of 25(OH)D and sodium azide as a preservative; 3 bottles per concentration level.

    AI/ML Overview

    The provided document describes the IDS-iSYS 25VitD8 assay, a device for the quantitative determination of total 25-hydroxyvitamin D [25(OH)D] in human serum or plasma. Below is an analysis of its acceptance criteria and the studies performed.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal "acceptance criteria" for all performance characteristics in a single, consolidated table. However, it presents performance data for various analytical characteristics, and in some cases, implicitly defines acceptable ranges or desired outcomes for these studies. For instance, in the specificity study, an acceptance criterion of "<10% bias between the test and control samples" is stated.

    Based on the provided information, a table can be constructed to show the reported device performance and implicitly derived or explicitly stated acceptance criteria.

    Performance CharacteristicAcceptance Criteria (Implicitly or Explicitly Stated)Reported Device Performance
    PrecisionNot explicitly stated for all samples, but generally seeks low CV%See table in section 9 for detailed per-sample SD and %CV. Example: Sample 1 (13.9 ng/mL) had a Total %CV of 7.4%. Sample 9 (103 ng/mL) had a Total %CV of 6.1%.
    Linearity/Reportable RangeLinear over the tested range; reported range to be well-defined.Linear over 3.58 to 136 ng/mL. Reported range: 4 to 110 ng/mL. Values below 4 ng/mL reported as "<4 ng/mL".
    Detection Limits (LoB, LoD, LoQ)Specific low values for LoB, LoD, LoQ.LoB: 1.31 ng/mL; LoD: 1.98 ng/mL; LoQ: 3.53 ng/mL.
    Analytical Specificity (Interference)Non-significant interference of <10% bias between test and control samples.All listed interferents (e.g., Triglycerides, Hemoglobin, Bilirubin, HAMA) showed <10% bias.
    Analytical Specificity (Cross-reactivity)Not explicitly stated, but implies that related compounds should not significantly cross-react negatively, while 25(OH)D3 and D2 should show high cross-reactivity.25(OH)D3: 101%; 25(OH)D2: 105%; 24,25 dihydroxyvitamin D3: 197%; 1,25 dihydroxyvitamin D3: 3%.
    Method Comparison (vs. Reference Method)Correlation coefficient (r) indicative of good agreement; slope close to 1 and intercept close to 0 in Passing-Bablok regression.r = 0.97; Slope = 0.99 (95% CI: 0.94 to 1.05); Intercept = -0.51 ng/mL (95% CI: -1.93 to 0.75).
    Matrix ComparisonSlope close to 1 and intercept close to 0 in Passing-Bablok regression, with strong correlation (r) across different sample types vs. control serum.SST: Slope 0.98, r 0.99; EDTA: Slope 0.96, r 1.00; Lithium Heparin: Slope 0.98, r 1.00; Sodium Heparin: Slope 0.99, r 0.99.
    Calibration Interval7 days for the candidate device (change from 14 days for predicate).7 days.
    In-use Reagent Stability42 days (change from 21 days for predicate).42 days.
    Shelf Life (accelerated studies)Minimum 12 months for kit and kit controls.12 months.

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision/Reproducibility: Nine serum-based samples at different 25(OH)D concentration levels were used. Each sample was likely tested multiple times over several days and systems (as indicated by the methodology for LoB/LoD/LoQ, which involved 3 batches, multiple runs, and operators, though precision specifics aren't identical).
    • Linearity: One high and one low human serum sample were used, along with 9 evenly spaced dilutions created by mixing them. (Total of 11 unique concentrations tested).
    • Detection Limits (LoB, LoD, LoQ):
      • LoB: 3 manufacturing batches (MB1, MB2B, MB3). Each batch ran 6 assays over 4-5 days, with 2 duplicates per assay. Total 60 replicates per batch.
      • LoD: 6 very low 25(OH)D samples. Each batch ran 6 assays over 3-5 days, with 2 duplicates per assay. Total 72 replicates for each manufacture batch.
      • LoQ: 10 low 25(OH)D samples. Each batch ran 6 assays over 3-5 days, with 2 duplicates per assay. Total 120 replicates for each manufacture batch (118 for MB1 due to mislabeling).
    • Analytical Specificity (Interference): Two serum samples at two different 25(OH)D concentrations for each interferent tested. Each condition tested with 26 replicates (spiked vs. control).
    • Analytical Specificity (Cross-reactivity): Samples spiked with relevant cross-reactants. Specific numbers of replicates or distinct samples are not detailed for each cross-reactant. Endogenous cross-reactants were tested with samples having established ID-LC-MS/MS values.
    • Method Comparison: A total of 136 human serum samples with a wide range of 25(OH)D concentrations (5.6 to 110 ng/mL).
    • Expected Values/Reference Range: 392 apparently healthy donors (200 males, 192 females) from three diverse regions of the United States (North, Central, South), sampled in the winter.
    • Matrix Comparison: 57 native samples plus 10 treated (spiked or diluted) samples, covering a range of 4.8 to 108 ng/mL from serum without additives, compared across various tube types.

    Data Provenance:

    • For the Expected Values/Reference Range study, data was collected from the United States (North, Central, South regions).
    • For other studies (Precision, Linearity, Detection Limits, Specificity, Method Comparison, Matrix Comparison), the country of origin is not explicitly stated, but given Immunodiagnostic Systems Limited is based in the UK, it is plausible some studies might have been conducted there or in collaboration with international labs. The method comparison refers to NIST/Ghent ID-LC-MS/MS, suggesting international standards and potentially international lab involvement. Based on the 510(k) submission, the studies were conducted by the manufacturer for the purpose of demonstrating substantial equivalence to a US-marketed device. All data is retrospective as it was collected before the submission for marketing clearance.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    There were no experts used to establish ground truth in the traditional sense for these studies. This device is an in vitro diagnostic (IVD) for quantitative measurement. The "ground truth" for the test set was established by:

    • Reference Methods: The primary ground truth for the Method Comparison study was the NIST/Ghent ID-LC-MS/MS 25(OH) Vitamin D Reference Method Procedure. This is a highly accurate and standardized analytical method, not a human expert.
    • Known Concentrations/Spiked Samples: For studies like linearity, detection limits, analytical specificity, and cross-reactivity, ground truth was based on:
      • Known concentrations of analytes (e.g., purified 25(OH)D, interferents, cross-reactants) used to spike samples.
      • Samples with established values from other validated methods, like ID-LC-MS/MS for endogenous cross-reactants.
    • Population Studies: For Expected Values/Reference Range, the "ground truth" are the measured values from a characterized "apparently healthy" donor population, with outliers and specific exclusions defined.

    4. Adjudication Method for the Test Set

    There was no adjudication method for these studies in the context of expert review, as the ground truth was based on analytical reference methods or predefined sample concentrations, not subjective interpretations requiring consensus from multiple human readers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

    No MRMC comparative effectiveness study was done. This device is an automated in vitro diagnostic (IVD) assay designed for quantitative measurement of a biomarker (25(OH)D). It does not involve human readers interpreting images or data that would be assisted by AI. The device directly yields a numerical result. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This device operates as a standalone algorithm (assay) in the sense that it performs the measurement without human intervention influencing the result of that specific measurement. The IDS-iSYS Multi-Discipline Automated System performs the assay automatically. The results are then used by clinicians in conjunction with other data, but the measurement itself is automated. The performance characteristics described (precision, linearity, detection limits, specificity, method comparison, matrix comparison) are all "standalone" performance metrics of the assay system itself.

    7. The Type of Ground Truth Used

    The ground truth used was primarily:

    • Reference Method Data: Specifically, the isotope dilution-liquid chromatography/tandem mass spectrometry (ID-LC-MS/MS) 25(OH)D Reference Method Procedure (RMP), particularly for method comparison and traceability. This RMP is itself traceable to the National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 2972.
    • Known/Assigned Concentrations: For stability, calibrator, and control value assignments, internal stock solutions, secondary standards (Internal Reference Calibrators, IRs), and manufactured kit calibrators/controls had values assigned back to the CDC VDSP aligned secondary standards or by using validated kit combinations over multiple instruments and runs.
    • Spiked Samples: For linearity, analytical specificity (interference and cross-reactivity) studies, samples were spiked with known concentrations of analytes, interferents, or cross-reactants.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a separate "training set" in the context of machine learning or AI development. This device is an immunoassay, not a machine learning model. The various studies described (precision, linearity, specificity, method comparison, etc.) are validation studies performed on the final assay formulation and instrument system.

    However, the closest analogous concepts would be:

    • Calibrator and Control Value Assignment: This involves multiple runs (minimum 20 assay runs for calibrators, minimum 21 assay runs for controls) and systems to establish and verify values. These essentially "train" the system by defining its response curve and quality control ranges.
    • Assay Development: During the initial development of the assay, numerous samples and experiments would have been conducted to optimize reagents, conditions, and performance characteristics. This "development data" would serve a similar purpose to a training set but is not explicitly detailed as such in this 510(k) summary, which focuses on validation of the final device.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, there's no explicit "training set" in the AI sense. For the closest analogous processes:

    • Calibrator and Control Value Assignment:
      • Traceability: The assay is traceable to the ID-LC-MS/MS 25(OH)D Reference Method Procedure (RMP), which was used to assign target values for the Vitamin D Standardization Program (VDSP) samples. This RMP is further traceable to the NIST SRM 2972.
      • Internal Standards: An internal stock solution ("Top Dose") of 25-hydroxyvitamin D3 had its concentration assigned by performing dilutions and testing in approved, previously QC-released kits.
      • Secondary Standards (IRs): These were manufactured and value-assigned using the CDC VDSP aligned secondary standards.
      • Kit Calibrators: Tested as unknowns in a minimum of 20 assay runs on one IDS-iSYS instrument, using these secondary standards (IRs) to establish their values through Excel data reduction and Prism for curve parameters. Values were then verified over 3 assays.
      • Kit Controls: Tested as unknowns in a minimum of 21 assay runs using multiple systems and approved kit combinations, with values calculated using the IDS-iSYS instrument's on-board software. Values were then verified in a single assay.

    Essentially, the "ground truth" for the calibrators and controls used to define the assay's operating curve and quality parameters is established through a hierarchical traceability chain leading back to NIST-traceable reference methods.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161082

    Validate with FDA (Live)

    Date Cleared
    2016-05-17

    (29 days)

    Product Code
    Regulation Number
    862.1660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IDS-iSYS 17-OH Progesterone Control Set
    The IDS-iSYS 17-OH Progesterone Control Set is for in vitro diagnostic use, for the quality control of the IDS-iSYS 17-OH Progesterone on the IDS-iSYS Multi-Discipline Automated System.
    Rx Only.

    IDS-iSYS 17-OH Progesterone Calibration Verifiers
    The IDS-iSYS 17-OH Progesterone Calibration Verifiers are an in vitro diagnostic device intended for medical purposes in the quantitative verification of assay calibration and measuring range of the IDS-iSYS 17-OH Progesterone assay, when performed on the IDS-iSYS Multi Discipline Automated System.
    Rx Only.

    Device Description

    The IDS-iSYS 17-OH Progesterone Control Set consists of Two sets of three vials, 1.0 mL each in liquid form. Human serum containing 17-OH Progesterone and sodium azide as a preservative (<0.1%), with three concentration levels.
    The IDS-iSYS 17-OH Progesterone Calibration Verifiers consists of Four calibration verifier levels in liquid form (two 1.0mL vials for level 0 and one 1.0mL vial for levels 1, 2 & 3). Human serum containing 17-OH Progesterone and sodium azide as a preservative (<0.1%), with three concentration levels.

    AI/ML Overview

    The provided text describes the acceptance criteria and supporting studies for two in vitro diagnostic devices: the IDS-iSYS 17-OH Progesterone Control Set and the IDS-iSYS 17-OH Progesterone Calibration Verifiers.

    1. Table of acceptance criteria and the reported device performance:

    Study TypeAcceptance CriteriaReported Device Performance
    Control Set
    Value AssignmentAssigned target value defined as the mean of all runs for the IDS-iSYS 17-OH Progesterone assay and analyzer.Expected values: Low control: 2.0 ng/mL, Medium control: 5.0 ng/mL, High control: 10.0 ng/mL. (The study provides these as the expected values, which implies they were achieved, but does not explicitly re-state them as performance against the criteria beyond the definition of the target value.)
    Closed Vial Stability- Mean concentration must be within QC ranges (as stated in Certificate of Analysis)- Precision: CV ≤ 10% for low concentration, ≤ 8% for middle and high concentration- Accelerated stability studies (CLSI guideline EP25-A) support a stability claim of 9 months when stored at 2-8°C.- Real-time studies are ongoing. (No explicit statement if current readings meet the mean concentration and precision criteria for the 9-month claim, only that it "supports" it).
    Open Vial StabilityPercent recoveries within 10% of the reference material concentration.Data supports the open vial stability claim of 49 days when stored at 2-8°C. (Implies the 10% recovery criterion was met).
    On-Board StabilityCompared to a reference material run at time 0. (Specific quantitative criteria not explicitly stated, assumes comparison ensures acceptable performance.)The on-board stability data supports the claimed on-board stability of 4 hours. (Implies acceptable comparison results).
    Calibration Verifiers
    Value AssignmentAssigned target value defined as the mean of all runs for the IDS-iSYS 17-OH Progesterone assay and analyzer.Expected values: Cal Ver 0: Undetectable, Cal Ver 1: 2.0 ng/mL, Cal Ver 2: 8.0 ng/mL, Cal Ver 3: 17.0 ng/mL. (Similar to control set, implies these were achieved.)
    Closed Vial Stability- Mean concentration must be within QC ranges (as stated in Certificate of Analysis)- Precision: CV ≤ 10% for low concentration, ≤ 8% for middle and high concentration- Accelerated stability studies (CLSI guideline EP25-A) support a stability claim of 9 months when stored at 2-8°C.- Real-time studies are ongoing. (No explicit statement if current readings meet the mean concentration and precision criteria for the 9-month claim, only that it "supports" it).
    On-Board StabilityCompared to a reference material run at time 0. (Specific quantitative criteria not explicitly stated, assumes comparison ensures acceptable performance.)The on-board stability data supports the claimed on-board stability of 4 hours. (Implies acceptable comparison results).

    2. Sample size used for the test set and the data provenance:

    • IDS-iSYS 17-OH Progesterone Control Set - Value Assignment:
      • Sample Size: A minimum of 21 runs using cartridge batches (number of batches not specified) tested in triplicate on each of three IDS-iSYS Multi Discipline Automated Systems. Control solutions were prepared gravimetrically.
      • Data Provenance: Not explicitly stated as retrospective or prospective, nor country of origin. Given it's a device manufacturer's study to establish product characteristics, it's inherently prospective in nature.
    • IDS-iSYS 17-OH Progesterone Control Set - Stability (All types):
      • Closed Vial (Real-time): Three lots of controls, tested in pentaplicate (5 replicates) at 2-month intervals for up to a minimum of 15 months. Each tested vial compared to reference material stored at -20°C.
      • Closed Vial (Accelerated): Performed according to CLSI guideline EP25-A.
      • Open Vial: Tested in duplicate at time points stated in the stability protocol (time points not detailed), against unopened vials.
      • On-Board: Three batches of controls, using three IDS-iSYS instruments. Tested at 0, 2, 4, 6, and 8 hours compared to a reference material at time 0.
      • Data Provenance: Same as above, likely prospective internal testing.
    • IDS-iSYS 17-OH Progesterone Calibration Verifiers - Value Assignment:
      • Sample Size: Minimum of five runs for each cartridge batch (number of batches not specified) tested in triplicate on each of three IDS-iSYS Multi Discipline Automated Systems. Calibration Verifier solutions were prepared gravimetrically.
      • Data Provenance: Same as above, likely prospective internal testing.
    • IDS-iSYS 17-OH Progesterone Calibration Verifiers - Stability (Closed Vial & On-Board):
      • Closed Vial (Real-time): Three lots of calibration verifiers, tested in pentaplicate (5 replicates) at 2-month intervals for up to a minimum of 15 months.
      • Closed Vial (Accelerated): Performed according to CLSI guideline EP25-A.
      • On-Board: Three batches of calibration verifiers, using three IDS-iSYS instruments. Tested at 0, 2, 4, 6, and 8 hours compared to a reference material at time 0.
      • Data Provenance: Same as above, likely prospective internal testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    This device is an in vitro diagnostic (IVD) quality control material and calibration verifier for automated systems, not an imaging AI device. The "ground truth" here refers to the reference values established for the control and calibration materials themselves, rather than diagnosis or interpretation by human experts.

    • Value Assignment: The "ground truth" (assigned target values) for both the control set and calibration verifiers is established by the mean of all runs from multiple tests on three IDS-iSYS Multi Discipline Automated Systems, using cartridge batches, and confirmed by immunologic analysis using the IDS-iSYS 17-OH Progesterone assay. The solutions themselves are prepared gravimetrically from intermediate stock solutions.
    • There are no "experts" in the traditional sense (e.g., radiologists) involved in establishing the ground truth for this type of chemical/immunological assay. The ground truth relies on the precision and accuracy of the analytical methods and the instrument itself.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. This is not a study involving human interpretation or subjective assessment of data that would require an adjudication method. The device's performance is determined by quantitative measurements and statistical analysis against predefined criteria.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is an IVD quality control and calibration material, not an AI-powered diagnostic tool for human readers. No MRMC study was conducted or is relevant for this product.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Yes, the studies are inherently standalone as they evaluate the performance of the control and calibration materials on the IDS-iSYS Multi-Discipline Automated System. The system/assay performs the measurements without human intervention in the interpretation of the control/verifier results themselves beyond setting up the experiment and analyzing statistical compliance. The performance reported (e.g., mean concentration, CV, percent recovery) is that of the material as measured by the automated system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The ground truth used is based on:

    • Gravimetric preparation: For the initial stock solutions and preparation of control/calibration verifier samples.
    • Immunologic analysis: Confirmation of concentrations using the IDS-iSYS 17-OH Progesterone assay itself.
    • Statistical averaging: The mean of multiple runs on multiple instruments to establish the assigned target values, treated as the reference ground truth for quality control and calibration verification.
    • Reference material comparison: For stability studies, comparison to a reference control material stored under optimal conditions (-20°C).

    This is a form of analytical ground truth derived from established quantitative laboratory methods and statistical analysis rather than clinical consensus or pathological diagnosis.

    8. The sample size for the training set:

    Not applicable. This is not an AI/machine learning device that requires a training set. The values and stability characteristics are determined through direct analytical testing, not model training.

    9. How the ground truth for the training set was established:

    Not applicable. As there is no training set for an AI/machine learning model, no ground truth needed to be established for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142351

    Validate with FDA (Live)

    Date Cleared
    2015-08-25

    (368 days)

    Product Code
    Regulation Number
    862.1825
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 25-Hydroxy Vitamin DS EIA assay is intended for the quantitative determination of 25-hydroxyvitamin D [25(OH)D] and other hydroxylated metabolites in human serum or plasma. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in an adult population.

    Device Description

    The 25-Hydroxy Vitamin De EIA is a manual assay test and does not require the use of an automated system. The whole assay is performed in a microtitre plate and requires a user to perform each step. The 25-Hydroxy Vitamin De assay is an enzymeimmunoassay for the quantitation of 25(OH)D and other hydroxylated metabolites in serum or plasma. 25uL of each calibrators, controls and samples are diluted with biotin labelled 25(OH)D. The diluted samples are incubated in microtitre wells which are coated with a highly specific sheep 25(OH)D at room temperature before aspiration and washing. Enzyme (horseradish peroxidase) labelled avidin, is added and binds selectively to complexed biotin and, following a further wash step, colour is developed using a chromogenic substrate (TMB). The absorbance of the stopped reaction mixtures are read in a microtitre plate reader, colour intensity developed being inversely proportional to the concentration of 25(OH)D.

    AI/ML Overview

    The provided document describes the performance characteristics of the 25-Hydroxy Vitamin D EIA assay, but it focuses on analytical and comparative performance rather than a study proving the device meets general acceptance criteria in a clinical setting with human readers. The document is for an in-vitro diagnostic (IVD) device, so the criteria and studies described are relevant to laboratory performance.

    Here's an analysis of the acceptance criteria and supporting studies based on the provided text, adapted for an IVD context where appropriate:

    1. A table of acceptance criteria and the reported device performance

    For IVD devices, "acceptance criteria" are usually internal performance goals set by the manufacturer to demonstrate analytical and clinical performance required for regulatory clearance. The document lists several performance characteristics and provides the corresponding measurements for the candidate device and, in some cases, the predicate device.

    Performance CharacteristicAcceptance Criteria (from text or implied)Reported Device Performance (Candidate Device)
    TraceabilityTraceable to ID-LCMS/MS 25(OH) Vitamin D RMP (VDSP standard)Traceable to ID-LCMS/MS RMP (VDSP human serum panel) and certified by CDC VDSP
    Stability (Kit)8 months when stored at 2-8°C8 months when stored at 2-8°C
    Stability (Open Vial)(Implied: adequate for practical use)Reconstituted Biotin 25(OH)D Solution: 8 weeks (2-8°C, dark) Unused Ab. coated plate strip: 8 weeks (2-8°C, foil pouch) Calibrators, Controls: 8 weeks (2-8°C) Wash Solution: 8 weeks (room temp, after prep)
    Limit of Blank (LoB)(Implied: sufficiently low)1.3 ng/mL (4.5 nmol/L)
    Limit of Detection (LoD)(Implied: sufficiently low)2.7 ng/mL (6.9 nmol/L)
    Limit of Quantitation (LoQ)(Implied: sufficiently low)4.8 ng/mL (12.0 nmol/L)
    Cross-Reactivity (25(OH)D3)(Implied: high expected, near 100%)95%
    Cross-Reactivity (25(OH)D2)(Implied: high expected, near 100%)109%
    Interference (Triglycerides)No interference up to 475 mg/dLNo interference up to 475 mg/dL
    Interference (Bilirubin conjugated)No interference up to 20 mg/dLNo interference up to 20 mg/dL
    Interference (Haemoglobin)No interference up to 400 mg/dLNo interference up to 400 mg/dL
    Interference (HAMA)No interference up to 1000 ng/mLNo interference up to 1000 ng/mL
    Interference (Rheumatoid Factor)No interference up to 800 IU/mLNo interference up to 800 IU/mL
    Interference (Red Blood Cells)No interference up to 0.4%No interference up to 0.4%
    Interference (Vitamin D Binding Proteins)No interference up to 2000 ng/dLNo interference up to 2000 ng/dL
    Interference (Total Protein)No interference up to 9.2 g/dLNo interference up to 9.2 g/dL
    Interference (Cholesterol, Total)No interference up to 500 mg/dLNo interference up to 500 mg/dL
    Interference (Biotin)No interference up to 200 nmol/LNo interference up to 200 nmol/L
    Method Comparison (vs. Predicate) Slope0.85 to 1.150.88 (Passing Bablok), 0.82 (Linear Reg)
    Method Comparison (vs. Predicate) Intercept+/- 7 ng/mL3.23 ng/mL (Passing Bablok), 4.86 ng/mL (Linear Reg)
    Method Comparison (vs. Predicate) Correlation (r)≥ 0.900.97
    Method Comparison (vs. Reference) Slope0.88 to 1.02 (95% CI from study)0.96 (Passing Bablok), 0.97 (Deming Reg)
    Method Comparison (vs. Reference) Intercept-1.62 to 1.88 ng/mL (95% CI from study)-0.11 ng/mL (Passing Bablok), -0.84 ng/mL (Deming Reg)
    Method Comparison (vs. Reference) Correlation (r)0.947 (from study)0.947
    Precision (Total %CV)(Implied: acceptable for diagnostic assay)3.7% to 11.5% across 11.7 to 65.1 ng/mL
    Linearity (Regression Coefficient R²)(Implied: close to 1.00)1.00
    Linearity (Max Deviation)(Implied: acceptable limits)-8.8% (>20ng/mL), 2.37 ng/ml (<20ng/mL)
    Reportable Range6.5 to 100 ng/mL6.5 to 100 ng/mL (16.3 to 250 nmol/L)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Precision/Reproducibility: 10 serum samples, assayed using 3 lots of reagents in duplicate, twice per day for a minimum of 20 days (n ≥ 80 replicates per sample). The origin of these samples is not specified.
    • Linearity/Assay Reportable Range: Samples containing varying concentrations of 25-hydroxyvitamin D were assayed in replicate of four. The number of unique samples for this study is not explicitly stated, but they were patient samples (high patient sample diluted with a low patient sample). Origin not specified.
    • Analytical Specificity (Cross-reactivity): Vitamin D serum samples were spiked with various cross-reactants. Number of samples not specified. Not specified if retrospective or prospective. Origin not specified.
    • Analytical Specificity (Interference): Substances spiked into the assay. Number of samples/experiments not specified. Not specified if retrospective or prospective. Origin not specified.
    • Method Comparison (vs. Predicate Device):
      • Sample Size: n = 195 (patient samples).
      • Provenance: Not specified (country/retrospective/prospective).
    • Method Comparison (vs. Reference Measurement Procedure - RMP):
      • Sample Size: n = 109 independent native serum samples.
      • Provenance: Samples were value assigned by the Ghent reference method procedure. Not specified if retrospective or prospective, or country of origin for the samples themselves.
    • Matrix Comparison:
      • Sample Sizes: SST (n=28), EDTA plasma (n=38), Sodium Heparin plasma (n=28), Lithium Heparin plasma (n=28), Citrate plasma (n=28).
      • Provenance: Not specified (country/retrospective/prospective).
    • Expected Values/Reference Range:
      • Sample Size: 280 apparently healthy adult individuals.
      • Provenance: Retrospective. Individuals were "living in geographical diverse regions of the United States to represent a broad spectrum of UV light exposure".

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    For this IVD device (25-Hydroxy Vitamin D EIA assay), the "ground truth" is established by reference methods or the true concentration of the analyte, not typically by expert interpretation from medical images or clinical judgment directly.

    • Method Comparison (vs. Reference Measurement Procedure): The "ground truth" for this comparison was established by the Ghent reference method procedure for 25-Hydroxy Vitamin D, and the assay is traceable to the isotope dilution-liquid chromatography/tandem mass spectrometry (ID-LCMS/MS) 25(OH) Vitamin D reference method procedure (RMP) which was used in assigning the target value for the Vitamin D Standardization Program (VDSP single donor human serum panel, itself traceable to NIST SRM 2972). This indicates a highly standardized and validated analytical method, which serves as the "expert" or definitive reference for concentration. The "experts" are the developers and maintainers of these reference methods, typically highly qualified analytical chemists or metrologists, not clinical experts for interpreting results in a diagnostic imaging sense.
    • Other studies: For precision, linearity, and specificity, the "ground truth" is inherent to the prepared samples or known concentrations, not established by human experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Adjudication methods like "2+1" or "3+1" are typically used in studies involving human interpretation (e.g., radiology reads) where disagreements need to be resolved. This is not applicable to the analytical performance studies of an IVD assay where quantitative results are compared to a reference method or known values. Therefore, none was used in the sense of clinical expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done, as this device is an in-vitro diagnostic assay for quantitative laboratory measurement, not an AI-assisted diagnostic imaging tool that involves human readers interpreting images.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This device is a standalone assay. It provides a numerical result for 25-Hydroxy Vitamin D concentration. The performance studies detailed (precision, linearity, LoD, LoQ, method comparisons, interference, cross-reactivity) are all evaluations of this standalone analytical performance, quantifying its accuracy and reliability in measuring the analyte in patient samples. There is no "human-in-the-loop" once the sample is processed by the assay; a user performs the assay steps, but the result generated is direct from the assay's chemical reactions and spectrophotometric reading.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used primarily consists of:

    • Reference Measurement Procedures (RMP): Specifically, the ID-LCMS/MS method (traceable to NIST SRM 2972) and the Ghent reference method procedure. These are highly accurate and standardized analytical methods for determining the true concentration of 25-Hydroxy Vitamin D.
    • Known Concentrations: For studies like linearity, LoB, LoD, LoQ, interference, and cross-reactivity, ground truth is established by preparing samples with known, precise concentrations of the analyte or interfering substances.

    8. The sample size for the training set

    This document describes a premarket notification (510(k)) for a medical device (an in-vitro diagnostic assay). The concept of a "training set" is typically associated with machine learning or AI models. For an IVD assay, raw materials and assay design are developed through R&D, and then validated with studies like those described. There isn't a "training set" in the sense of data used to train an algorithm. The development and optimization of the assay's components (e.g., antibodies) would occur during the manufacturing process, but the document doesn't provide details on sample sizes used during that phase.

    9. How the ground truth for the training set was established

    As there isn't a "training set" in the AI/ML context for this IVD assay, this question is not directly applicable. The "ground truth" for the analytical performance studies (as described in point 7) was established using reference measurement procedures and known concentrations.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142994

    Validate with FDA (Live)

    Date Cleared
    2015-04-21

    (187 days)

    Product Code
    Regulation Number
    862.1045
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDS-iSYS Aldosterone assay (IS-3300) is a device intended for use in clinical laboratories for the quantitative determination of Aldosterone in human EDTA plasma on the IDS-iSYS Multi-Discipline Automated System. Aldosterone measurements are used in the diagnosis and treatment of primary aldosteronism (a disorder caused by excessive secretion of Aldosterone by the adrenal gland), hypertension caused by primary aldosteronism, selective hyperaldosteronism, edematous states and other conditions of electrolyte balance.

    The IDS-iSYS Aldosterone Control Set (IS-3330) is intended for use as assaved quality control samples to monitor the accuracy of the IDS-iSYS Aldosterone assay on the IDS-iSYS Multi-Discipline Automated System.

    The IDS-iSYS Aldosterone Calibration Verifiers (IS-3335) are intended for medical purposes for use in the quantitative verification of calibration of the IDS-iSYS Aldosterone assay on the IDS-iSYS Multi-Discipline Automated System.

    Device Description

    The IDS-iSYS Aldosterone assay is based on chemiluminescence technology. A biotinylated monoclonal anti-Aldosterone antibody is incubated with the sample, after an incubation step an Aldosterone acridinium conjugate is added and after a further incubation step streptavidin coated magnetic particles are added. Following a third incubation step the particles are "captured" using a magnet. After a washing step and addition of trigger reagents, the light emitted by the acridinium label is inversely proportional to the concentration of Aldosterone in the original sample.

    AI/ML Overview

    Here's an analysis of the provided text to extract the requested information about acceptance criteria and the supporting study:

    The provided document is a 510(k) Summary for the IDS-iSYS Aldosterone assay, control set, and calibration verifiers. It focuses on demonstrating substantial equivalence to a predicate device, as required for FDA clearance. The document details performance characteristics but does not explicitly state acceptance criteria in a pass/fail format typical of formal acceptance criteria documents. Instead, it reports performance values found during validation studies.

    Here's the breakdown of the information you requested, based on what's available in the document:


    1. Table of Acceptance Criteria and Reported Device Performance

    As mentioned, explicit "acceptance criteria" are not listed in a single table. However, the performance characteristics are reported, and implicitly, these values are considered acceptable for demonstrating substantial equivalence. The table below compiles the reported performance data from the document.

    Note: The document does not provide a column for "Acceptance Criteria" as a separate, pre-defined target. The "Reported Device Performance" is the outcome of the studies aiming to demonstrate acceptable performance.

    Performance CharacteristicImplicit or Explicit Acceptance Threshold (Not explicitly stated as "acceptance criteria" but implied targets from regulatory guidance or industry standards)Reported Device Performance (IDS-iSYS Aldosterone Assay)Source of Data
    Precision (CV%)Implied: Generally low CV% indicating good reproducibility (e.g., <15-20% for clinical assays, lower for critical ranges)Within-Run: 2.2% - 8.4% (at 98.7 ng/dL to 7.5 ng/dL)Table (p.8)
    Total: 5.2% - 12.8% (at 98.7 ng/dL to 7.5 ng/dL)Table (p.8)
    Linearity (R²)Implied: High R² value (e.g., >0.98 or >0.99) indicating a strong linear relationship.R² = 1.00Text (p.9)
    Regression Equation (y=mx+b)Implied: Slope (m) close to 1, intercept (b) close to 0.y = 1.00x - 1.24 (slope = 1.00, intercept = -1.24)Text (p.9)
    Limit of Blank (LoB)Implied: Low value to correctly identify absence of analyte.2.0 ng/dLTable (p.11)
    Limit of Detection (LoD)Implied: Low value to correctly detect presence of analyte.3.2 ng/dLTable (p.11)
    Limit of Quantitation (LoQ)Implied: Low value with acceptable precision (e.g., typically ≤20% CV).3.9 ng/dL (at 20% CV)Table (p.11)
    Interference (Concentration Bias)Explicitly stated criterion in study: ≤10% concentration bias to the unspiked sample.All tested interferents met this criterion.Text (p.11)
    Cross-ReactivityImplied: Low percentage for non-target analytes, 100% for target analyte.Aldosterone: 100%. Other analytes: 0.0003% to 3.1% (3α, 5β-Tetrahydroaldosterone: 3.1%)Table (p.13)
    Method Comparison (Slope vs. Predicate)Implied for substantial equivalence: Slope close to 1, intercept close to 0, high R².Linear Regression: Slope = 1.053, Intercept = 0.09 ng/dL, R² = 0.980Text (p.14)
    Passing-Bablok: Slope = 1.070, Intercept = -0.29Text (p.14)

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision/Reproducibility:

      • Sample Size: Six EDTA plasma samples (9 "samples" were run over 20 days, generating 80 replicates per sample concentration).
      • Data Provenance: Not explicitly stated, but clinical laboratory setting is implied. The phrase "two sites using three analyzers" could suggest multi-center testing, potentially from different locations/countries, but this is not specified. The studies were performed for the manufacturer for regulatory submission.
    • Linearity/Assay Reportable Range:

      • Sample Size: One high plasma sample diluted with a low sample to create 11 dilution samples. Run in quadruplicate.
      • Data Provenance: Not specified, likely internal laboratory data for the manufacturer.
    • Detection Limit (LoB, LoD, LoQ):

      • Sample Size:
        • LoB (Lot 1): One zero-aldosterone plasma sample, 10 replicates for 5 days (total 50 replicates).
        • LoD/LoQ (Lot 1): 7 samples, 2 replicates per sample, once per day for 8 days.
        • LoB (Lot 2): One zero-aldosterone plasma sample, 6 replicates for 5 days (total 30 replicates).
        • LoD (Lot 2): 8 samples, 2 replicates per sample, once per day for 5 days.
        • LoQ (Lot 2): 7 samples, 2 replicates per sample, once per day for 5 days.
      • Data Provenance: Not specified, likely internal laboratory data for the manufacturer. Two different sites and two different analyzers were used.
    • Analytical Specificity (Interference and Cross-Reactivity):

      • Sample Size:
        • Interference: Two base plasma samples ("Low" and "High" Aldosterone concentrations), spiked with various potential interferents. For each interferent, 26 replicates for blank and spiked samples were compared.
        • Cross-reactivity: Stock solutions of various compounds diluted serially to create 7-point standard curves for each substance.
      • Data Provenance: Not specified, likely internal laboratory data for the manufacturer.
    • Method Comparison:

      • Sample Size: 161 samples (including 12 altered samples).
      • Data Provenance: Not specified, implied to be clinical or patient samples. Whether these were retrospective or prospective, or their country of origin, is not mentioned.
    • Reference Range Study (Expected Values):

      • Sample Size: 228 Caucasian adult samples.
      • Data Provenance: Collected in the US, prospective (samples collected under specific conditions: 7-10 am after overnight fasting, upright/supine positions).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This type of information is generally NOT applicable to in vitro diagnostic (IVD) devices like the Aldosterone assay described. For IVD assays, "ground truth" is typically established by:

    • Reference methods/predicate devices (e.g., Diasorin Liaison Aldosterone assay for method comparison).
    • Gravimetric preparation of standards and calibrators for traceability.
    • Defined sample characteristics (e.g., "zero aldosterone plasma," "high plasma sample").

    There were no human experts assessing images or making diagnoses that would require adjudication.


    4. Adjudication Method for the Test Set

    Not applicable, as this is an IVD assay, not a device requiring human interpretation of results in a diagnostic context that would call for adjudication of different interpretations.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an IVD assay for quantitative determination of a biomarker; there are no "human readers" interpreting results in the way an MRMC study would be designed for an imaging AI device.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the studies conducted (Precision, Linearity, LoD/LoQ, Analytical Specificity, Method Comparison) represent the standalone performance of the IDS-iSYS Aldosterone assay system. The device quantifies aldosterone in plasma samples directly, without human interpretation of results being a variable in its core analytical performance.


    7. The Type of Ground Truth Used

    The ground truth used for various studies includes:

    • Reference materials: Aldosterone (≥95% HPLC; A9477Sigma-Aldrich) dissolved in Dioxane, with concentration calculated by UV quantitation using molar extinction coefficient (for calibrators and traceability).
    • Predicate device results: The Diasorin Liaison Aldosterone assay (K130321) for method comparison.
    • Spiked samples: For linearity, interference, and cross-reactivity studies, known concentrations of analyte or interferents were added to base samples to create "true" concentrations.
    • Patient samples: For method comparison and reference range studies, patient samples were used. Their "truth" for method comparison was established by the predicate device. For reference ranges, the "truth" was the measured value distributed among a healthy population.

    8. The Sample Size for the Training Set

    The document describes performance studies (validation) but does not explicitly mention a "training set" in the context of machine learning. For an IVD assay, the development process involves reagent optimization, calibration curve fitting, etc., which conceptually use data, but this is not typically termed a "training set" like in AI/ML.

    However, the "Master curve" and "two-point calibration" system is mentioned. The "logistic parameters of the Master calibration curve" are generated using "data of 20 runs Internal Reference Calibrators" (p.9). This could be considered analogous to a training process for establishing the core measurement algorithm, but it's not a training set for an AI inference model.


    9. How the Ground Truth for the Training Set Was Established

    If we consider the generation of the "Master calibration curve" as a "training" process:

    • Ground Truth: The "Internal Reference Calibrators" referenced (p.9) would serve as the ground truth. These calibrators are traceable to gravimetrically prepared Aldosterone standards using high-purity Aldosterone and UV quantitation (p.9).
    • Establishment Method: The "new kit calibrator sets are run as 'unknowns' in duplicate in at least 20 assays on one analyser" (p.9). "The data of 20 runs Internal Reference Calibrators are used to generate the logistic parameters of the Master calibration curve by using Prism software package" (p.9).
    Ask a Question

    Ask a specific question about this device

    K Number
    K143324

    Validate with FDA (Live)

    Date Cleared
    2015-03-30

    (131 days)

    Product Code
    Regulation Number
    862.1660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDS-iSYS CTX-1 (CrossLaps ®) Calibration Verifiers is a device intended for the verification of the IDS-iSYS CTX-I (CrossLaps ®) Assay when performed on the IDS-iSYS Multi-Discipline Automated Analyzer

    Device Description

    The IDS-iSYS CTX-I (CrossLaps®) Calibration Verifiers consists of one set of four vials, 2.5 mL each in liquid form, containing horse serum with <0.1% (w/w) sodium azide as a preservative, with four concentration levels of human CTX-I:
    Cal. Ver. 0: Undetectable
    Cal. Ver. 1: 0.12 - 0.16 ng/mL
    Cal. Ver. 2: 2.4 - 3.2 ng/mL
    Cal. Ver. 3: 5.6 - 6.6 ng/mL

    AI/ML Overview

    The document is a 510(k) Summary for the IDS-iSYS CTX-I (CrossLaps®) Calibration Verifiers, a device intended for the verification of calibration of the IDS-iSYS CTX-I (CrossLaps®) Assay when performed on the IDS-iSYS Multi-Discipline Automated Analyzer.

    Here's an analysis of the provided information to answer your request:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" for the device's performance in a table format. Instead, it defines the "Target Range" for each Calibration Verifier level, derived from the mean ± 2 standard deviations during value assignment. This target range implicitly serves as the acceptance criteria for individual measurements when verifying calibration.

    CVMTarget Mean (ng/mL)Standard Deviation (SD)Target Range (ng/mL) (Acceptance Criteria)Reported Device Performance (Target Mean)
    Cal. Ver. 00 (Undetectable)NANA0 (Undetectable)
    Cal. Ver. 10.140.0090.12 - 0.160.14
    Cal. Ver. 22.80.1682.4 - 3.22.8
    Cal. Ver. 36.10.3665.6 - 6.66.1

    Note: The "Reported Device Performance" here is the target mean derived during the value assignment process, which aligns with the center of the acceptance range. The study ensures that the device can consistently achieve these target means and fall within the defined ranges.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document describes the "value assignment" process, which functions as the primary method for establishing the performance characteristics and implicitly, the "test set" for the verifiers.

    • Sample Size: "Each lot-specific value assignment was tested in five runs on at least three different IDS-iSYS analyzers in triplicate, for a total of 45 replicates." This refers to 45 individual measurements for each calibrator verifier (Cal. Ver. 0, 1, 2, 3).
    • Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective. Given that Immunodiagnostic Systems Ltd is based in the United Kingdom, it's likely the studies were conducted there. The value assignment process described suggests a prospective study design for establishing lot-specific values.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable to this type of device. The device is a "Quality control material (assayed and unassayed)" for an in vitro diagnostic assay. Its "ground truth" (target values) is established through rigorous analytical testing and standardization against in-house reference standards, not by human expert interpretation of signals or images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As explained in point 3, the ground truth is established through analytical methods, not through expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not a study involving human readers or AI assistance. The device is a calibration verifier for an automated immunoassay.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of the IDS-iSYS CTX-I assay on the IDS-iSYS Multi-Discipline Automated Analyzer when calibrated and verified using these verifiers. The "value assignment" study described is a standalone performance assessment of the verifiers themselves, demonstrating their ability to consistently produce expected concentration values within a statistically defined range. The objective of the verifiers is to ensure the assay's standalone performance is accurate.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the calibrator verifiers is their assigned concentration values, which are established through:

    • Standardization against in-house reference standards (CTX-I in horse serum).
    • Derivation from the mean of multiple replicates across multiple runs and instruments. The target ranges (e.g., 0.12 - 0.16 ng/mL for Cal. Ver. 1) are determined statistically (mean ± 2SD).

    8. The sample size for the training set

    This concept is not directly applicable to a calibration verifier in the same way it would be for an AI algorithm. The "training" here refers to the process of standardizing the manufacturing of the verifiers and establishing their lot-specific values. The "value assignment" study (45 replicates per verifier level) serves to establish these "ground truth" values and ranges.

    9. How the ground truth for the training set was established

    For each lot, the "ground truth" (target mean values and ranges) was established through:

    • Extensive testing: "Each lot-specific value assignment was tested in five runs on at least three different IDS-iSYS analyzers in triplicate," generating a substantial dataset for statistical analysis.
    • Statistical analysis: The "assigned target value of each calibrator verifier was defined as the mean of all the runs for each calibrator verifier. The guideline target range is defined as the mean of all runs ± 2SD."
    • Standardization: The IDS-iSYS CTX-I assay (for which these verifiers are used) is itself "standardized against in-house reference standards (CTX-I in horse serum)." This implies a chain of traceability to primary or secondary reference materials.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140554

    Validate with FDA (Live)

    Date Cleared
    2014-12-19

    (290 days)

    Product Code
    Regulation Number
    862.1825
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDS iSYS 25-Hydroxy Vitamin DS Assay (IDS-iSYS 250HD*) is intended for the quantitative determination of 25-hydroxyvitamin D (25OHD) and other hydroxylated metabolites in human serum on the IDS iSYS Multi-Discipline Automated System. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in an adult population.

    The IDS-iSYS 25-Hydroxy Vitamin DS (250HDS) Control Set is used for quality control of the IDSiSYS 25-Hydroxy Vitamin DS assay on the IDS-iSYS Multi-Discipline Automated System.

    Device Description

    The IDS-iSYS 25-Hydroxy Vitamin DS assay consists of a reagent cartridge and one set of calibrators (Calibrators A & B or CAL A & CAL B). The reagent cartridge contains multiple reagents: MPV1 (Magnetic particles coated with 25-OH D in a phosphate buffer containing methanol with sodium azide as preservative), CONJ (Anti- 25-OH D sheep polyclonal antibody labelled with an acridinium ester derivative, in buffer containing bovine, sheep, rabbit and mouse proteins with sodium azide as preservative), NaOH (Sodium hydroxide solution <0.5 M), and BUF (Assay buffer containing proprietary displacing compounds, methanol, and sodium azide as preservative). Calibrators A and B contain horse serum in a buffer matrix with two defined concentrations of 25-OH D and sodium azide as a preservative. The IDS-iSYS 25-Hydroxy Vitamin DS Control Set contains horse serum in a buffer matrix with three defined concentrations of 25-OH D and sodium azide as a preservative.

    AI/ML Overview

    This is an analysis of a 510(k) summary for the IDS-iSYS 25-Hydroxy Vitamin Dˢ Assay and IDS-iSYS 25-Hydroxy Vitamin Dˢ Control Set.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally established by internal specifications of the manufacturer, often guided by CLSI (Clinical and Laboratory Standards Institute) guidelines or regulatory precedents. The document presents the reported performance without explicitly stating pre-defined acceptance criteria for each measurement, but rather implies acceptability by presenting results typically expected for such assays. For some sections, such as precision/reproducibility and interference, the acceptable range is mentioned within the text (e.g., ±10% bias for interference).

    Performance CharacteristicAcceptance Criteria (Implied/Stated)Reported Device Performance (IDS-iSYS 25-Hydroxy Vitamin Dˢ)
    Precision/Reproducibility (Total CV%)No explicit overall criterion, typically <10-15% for clinical assays.Serum 1 (13.2 ng/mL): 10.6%
    (Based on CLSI EP5 A2)Serum 2 (27.2 ng/mL): 9.0%
    Serum 3 (38.9 ng/mL): 9.1%
    Serum 4 (54.5 ng/mL): 9.1%
    Serum 5 (77.2 ng/mL): 8.5%
    Serum 6 (119.9 ng/mL): 7.2%
    Linearity (R²)Typically ≥ 0.99 for good linearity.R² = 1.00
    Reportable RangeDefined by linearity study.7-125 ng/mL
    Limit of Blank (LoB)As determined by CLSI EP-17A.0.6 ng/mL
    Limit of Detection (LoD)As determined by CLSI EP-17A.2.6 ng/mL
    Limit of Quantitation (LoQ)As determined by CLSI EP-17A (at 20% precision CV).7.0 ng/mL
    Analytical Specificity (Cross-reactivity - Endogenous)Percentage cross-reactivity with specified metabolites.25(OH)D3: 97%
    25(OH)D2: 120%
    24,25(OH)2D3: 124%
    Analytical Specificity (Cross-reactivity - Exogenous)Percentage cross-reactivity with specified metabolites.3-epi-25(OH)D3: 1%
    3-epi-25(OH)D2: 1%
    1,25-(OH)2 D3: -23%
    1,25-(OH)2 D2: 9%
    Vitamin D3: 0%
    Vitamin D2: 0%
    Paricalcitol: 0%
    Interference (Bias %)< ±10% bias the control sample to the test sample.Triglycerides: -6% & -5% (at 500mg/dL)
    Bilirubin, conjugated: 9.2% & 6.4% (at 30mg/dL)
    Haemoglobin: -8% & -2% (at 40mg/dL)
    Biotin: 1% & -3% (at 300nmol/L)
    HAMA: -5% & -2% (at 500ng/mL)
    Red Blood Cells: -10% & -2% (at 0.2%)
    Vitamin DBP: 0% & 0% (at 2000ng/mL)
    Method Comparison (vs. ID-LC-MS/MS RMP)Slope typically close to 1, intercept close to 0, high R or r.Passing-Bablok: Slope = 0.95 (CI: 0.86 to 1.04), Intercept = 0.80 ng/mL (CI: -1.32 to 3.08 ng/mL)
    Deming: Slope = 0.94 (CI: 0.86 to 1.01), Intercept = 1.34 ng/mL (CI: -0.78 to 3.45 ng/mL)
    Pearson r: 0.925
    Method Comparison (vs. Predicate device)Slope typically close to 1, intercept close to 0, high R or r.Passing-Bablok: Slope = 0.96 (CI: 0.91 to 1.01), Intercept = 1.1 ng/mL (CI: -0.3 to 2.3 ng/mL)
    Correlation Coefficient, r: 0.94

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision/Reproducibility: 6 serum samples, 80 measurements each (total = 480 measurements across 3 lots, 2 internal replicates, 2 external replicates, 20 days, 3 analyzers). Data provenance: Not explicitly stated, but typically internal studies.
    • Linearity/Assay Reportable Range: 11 concentration levels from a high and low serum sample dilution, each assayed in duplicate from one manufacturing batch (total 22 concentration levels). Data provenance: Not explicitly stated, typically internal studies.
    • Detection Limit:
      • LoB: Zero calibrator assayed as 10 replicates over 10 assays on 10 separate days (total of 100 measurements).
      • LoD: 10 samples (native and/or diluted) with very low vitamin D concentrations, in duplicate over 12 assays spanning multiple days (total of 240 data points).
      • LoQ: 13 low samples (native and/or diluted) in duplicate over 7 individual assays spanning multiple days (total of 182 data points).
        Data provenance for all detection limits: Not explicitly stated, typically internal studies.
    • Analytical Specificity (Cross-reactivity): Not explicitly stated numbers of samples per cross-reactant, but uses vitamin D serum samples, some spiked. Data provenance: Not explicitly stated, typically internal studies.
    • Interference: Not explicitly stated numbers of samples per interfering agent, but uses test vs. control samples, and in one case, 26 measurements for each condition. Data provenance: Not explicitly stated, typically internal studies.
    • Method Comparison with Predicate Device (vs. ID-LC-MS/MS RMP): 99 samples. Data provenance: Not explicitly stated, but implies connection to CDC VDSP program for traceability.
    • Method Comparison with Internal Predicate Device: 283 European-sourced serum samples. Data provenance: European, retrospective.
    • Expected Values/Reference Range: 275 apparently healthy adults (light skin and dark skin, male and female, aged 21-77 years) from geographically diverse regions of the United States. Data provenance: Prospective, United States.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    For this in vitro diagnostic device (IVD), the "ground truth" is typically established by reference methods or accepted techniques for measuring the analyte. No human expert "adjudication" in the sense of image review is performed.

    • For Traceability/Standardization: The device is traceable to the isotope dilution-liquid chromatography/tandem mass spectrometry (ID-LC-MS/MS) 25(OH)D Reference Method Procedure (RMP) which was used in assigning the target value for the VDSP samples. The ID-LC-MS/MS RMP is traceable to the National Institute of Standards and Technology Standard Reference Material (SRM) 2972. This is considered the reference standard, not "expert concensus" in this context.
    • For Method Comparison with Predicate Device (vs. ID-LC-MS/MS RMP): The ID-LC-MS/MS RMP serves as the reference method and thus the "ground truth" in this comparison.

    4. Adjudication Method for the Test Set

    Not applicable. This is an IVD device, and the ground truth is established by chemical reference methods and measurements, not by human expert adjudication of interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an in vitro diagnostic assay that directly quantifies an analyte in serum, not an imaging device requiring human interpretation, which is where MRMC studies are typically employed. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply here.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies presented (precision, linearity, detection limits, specificity, method comparisons) evaluate the standalone performance of the IDS-iSYS 25-Hydroxy Vitamin Dˢ Assay. This device is an automated system designed to provide quantitative results directly from serum samples without direct human interpretation of the assay's output for diagnosis, beyond basic quality control and clinical interpretation of the quantitative value.

    7. The Type of Ground Truth Used

    The ground truth for the analytical performance studies (e.g., linearity, detection limit, cross-reactivity, interference) is established by the design of the experiments, using known concentrations of analytes, spiked samples, or reference materials.

    For the method comparison studies, the ground truth is:

    • ID-LC-MS/MS RMP (Reference Method Procedure): This is considered the gold standard or highly accurate reference method for 25(OH)D measurement, traceable to NIST SRM 2972.
    • The currently marketed IDS-iSYS 25-Hydroxy Vitamin D (K091849) predicate device: This device's results are used as a comparative "truth" to demonstrate substantial equivalence, although the new device is also aligned with the ID-LC-MS/MS RMP.

    8. The Sample Size for the Training Set

    The document does not explicitly delineate a separate "training set" in the context of machine learning. For IVD devices like this, method development and optimization phases involve extensive testing with various samples, which could be considered analogous to a training process, but it's not typically formally described as such.

    • Calibrator Traceability and Value Assignment: Master calibrators and kit calibrators are developed using stock solutions, and their values are adjusted based on results from running calibrators in multiple assays (minimum of 20 assay runs for kit calibrators) and multiple analyzers. Final values are adjusted for batch-to-batch consistency and alignment to the CDC VDSP program based on multiple correlation assays of an "established patient library panel". The size of this patient library panel is not specified.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, a formal "training set" with ground truth in the machine learning sense is not explicitly described. However, the development and calibration of the assay involve:

    • Traceability to U.V. quantification and then ID-LC-MS/MS RMP:
      • Internal stock solutions are prepared, and their potency is initially based on absorbance at 264nm of an ethanolic 25D solution.
      • Master calibrators are prepared, and their final value assignment is based on extensive testing against the assay itself (running them in multiple assays on multiple analyzers).
      • The kit calibrators are tested as unknowns in a minimum of 20 assay runs calibrated with master calibrators, and values are adjusted to ensure consistency and alignment to the CDC VDSP program based on multiple correlation assays of an established patient library panel. This patient library panel's "ground truth" would have been established by the ID-LC-MS/MS RMP.

    In summary, for this IVD, the ground truth primarily relies on established analytical standards and reference methods (ID-LC-MS/MS, NIST SRMs), rather than expert consensus or pathology reports. The studies demonstrate the assay's analytical performance and its comparability to both the ID-LC-MS/MS RMP and a predicate device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 3