K Number
K161831

Validate with FDA (Live)

Date Cleared
2016-11-15

(133 days)

Product Code
Regulation Number
862.1825
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IDS iSYS 25 VilD assay is intended for the quantitative determination of total 25-hydroxyvitamin D (125(OH)D) in hyman serum or plasma on the IDS iSYS Multi-Discipline Automated System. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in an adult population.

The IDS-iSYS 25 VitD Control Set is used for quality control of the IDS-iSYS 25 VitD assay on the IDS-iSYS Multi-Discipline Automated System.

Device Description

The IDS-iSYS 25 VitDs reagent kit consists of one (1) Immunoassay Cartridge, one (1) vial each of 2 concentration Calibrator levels and a mini-CD containing the Instructions For Use (IFU), CRY files and Certificate of Analysis.

IDS-iSYS 25 VitDS Cartridge, sufficient for 100 tests, consists of reagents provided in individual compartment within a plastic container called the Cartridge. IDS-iSYS 25 VitDS Cartridge contains the following ready to use reagents:

  • Magnetic particles coated with streptavidin in a phosphate . buffer containing sodium azide.
  • . Sodium hydroxide solution.
  • . 25(OH)D labelled with an acridinium ester derivative, in buffer containing bovine serum albumin with sodium azide.
  • . Anti-25(OH)D sheep polyclonal antibody labelled with an biotin, in buffer containing bovine, sheep and mouse proteins and sodium azide.
  • . Assay buffer containing proprietary displacing compounds, methanol and sodium azide.

The IDS-iSYS 25 VitD® Calibrators are included in the reagent kit. The ready to use calibrators contains human serum buffer matrix with two defined concentrations of 25(OH)D and sodium azide as preservative (<0.1 %), 1 bottle per concentration level.

The IDS-iSYS 25VitDe Control Set contains human serum in a buffer matrix with two defined concentrations of 25(OH)D and sodium azide as a preservative; 3 bottles per concentration level.

AI/ML Overview

The provided document describes the IDS-iSYS 25VitD8 assay, a device for the quantitative determination of total 25-hydroxyvitamin D [25(OH)D] in human serum or plasma. Below is an analysis of its acceptance criteria and the studies performed.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state formal "acceptance criteria" for all performance characteristics in a single, consolidated table. However, it presents performance data for various analytical characteristics, and in some cases, implicitly defines acceptable ranges or desired outcomes for these studies. For instance, in the specificity study, an acceptance criterion of "<10% bias between the test and control samples" is stated.

Based on the provided information, a table can be constructed to show the reported device performance and implicitly derived or explicitly stated acceptance criteria.

Performance CharacteristicAcceptance Criteria (Implicitly or Explicitly Stated)Reported Device Performance
PrecisionNot explicitly stated for all samples, but generally seeks low CV%See table in section 9 for detailed per-sample SD and %CV. Example: Sample 1 (13.9 ng/mL) had a Total %CV of 7.4%. Sample 9 (103 ng/mL) had a Total %CV of 6.1%.
Linearity/Reportable RangeLinear over the tested range; reported range to be well-defined.Linear over 3.58 to 136 ng/mL. Reported range: 4 to 110 ng/mL. Values below 4 ng/mL reported as "<4 ng/mL".
Detection Limits (LoB, LoD, LoQ)Specific low values for LoB, LoD, LoQ.LoB: 1.31 ng/mL; LoD: 1.98 ng/mL; LoQ: 3.53 ng/mL.
Analytical Specificity (Interference)Non-significant interference of <10% bias between test and control samples.All listed interferents (e.g., Triglycerides, Hemoglobin, Bilirubin, HAMA) showed <10% bias.
Analytical Specificity (Cross-reactivity)Not explicitly stated, but implies that related compounds should not significantly cross-react negatively, while 25(OH)D3 and D2 should show high cross-reactivity.25(OH)D3: 101%; 25(OH)D2: 105%; 24,25 dihydroxyvitamin D3: 197%; 1,25 dihydroxyvitamin D3: 3%.
Method Comparison (vs. Reference Method)Correlation coefficient (r) indicative of good agreement; slope close to 1 and intercept close to 0 in Passing-Bablok regression.r = 0.97; Slope = 0.99 (95% CI: 0.94 to 1.05); Intercept = -0.51 ng/mL (95% CI: -1.93 to 0.75).
Matrix ComparisonSlope close to 1 and intercept close to 0 in Passing-Bablok regression, with strong correlation (r) across different sample types vs. control serum.SST: Slope 0.98, r 0.99; EDTA: Slope 0.96, r 1.00; Lithium Heparin: Slope 0.98, r 1.00; Sodium Heparin: Slope 0.99, r 0.99.
Calibration Interval7 days for the candidate device (change from 14 days for predicate).7 days.
In-use Reagent Stability42 days (change from 21 days for predicate).42 days.
Shelf Life (accelerated studies)Minimum 12 months for kit and kit controls.12 months.

2. Sample Size Used for the Test Set and Data Provenance

  • Precision/Reproducibility: Nine serum-based samples at different 25(OH)D concentration levels were used. Each sample was likely tested multiple times over several days and systems (as indicated by the methodology for LoB/LoD/LoQ, which involved 3 batches, multiple runs, and operators, though precision specifics aren't identical).
  • Linearity: One high and one low human serum sample were used, along with 9 evenly spaced dilutions created by mixing them. (Total of 11 unique concentrations tested).
  • Detection Limits (LoB, LoD, LoQ):
    • LoB: 3 manufacturing batches (MB1, MB2B, MB3). Each batch ran 6 assays over 4-5 days, with 2 duplicates per assay. Total 60 replicates per batch.
    • LoD: 6 very low 25(OH)D samples. Each batch ran 6 assays over 3-5 days, with 2 duplicates per assay. Total 72 replicates for each manufacture batch.
    • LoQ: 10 low 25(OH)D samples. Each batch ran 6 assays over 3-5 days, with 2 duplicates per assay. Total 120 replicates for each manufacture batch (118 for MB1 due to mislabeling).
  • Analytical Specificity (Interference): Two serum samples at two different 25(OH)D concentrations for each interferent tested. Each condition tested with 26 replicates (spiked vs. control).
  • Analytical Specificity (Cross-reactivity): Samples spiked with relevant cross-reactants. Specific numbers of replicates or distinct samples are not detailed for each cross-reactant. Endogenous cross-reactants were tested with samples having established ID-LC-MS/MS values.
  • Method Comparison: A total of 136 human serum samples with a wide range of 25(OH)D concentrations (5.6 to 110 ng/mL).
  • Expected Values/Reference Range: 392 apparently healthy donors (200 males, 192 females) from three diverse regions of the United States (North, Central, South), sampled in the winter.
  • Matrix Comparison: 57 native samples plus 10 treated (spiked or diluted) samples, covering a range of 4.8 to 108 ng/mL from serum without additives, compared across various tube types.

Data Provenance:

  • For the Expected Values/Reference Range study, data was collected from the United States (North, Central, South regions).
  • For other studies (Precision, Linearity, Detection Limits, Specificity, Method Comparison, Matrix Comparison), the country of origin is not explicitly stated, but given Immunodiagnostic Systems Limited is based in the UK, it is plausible some studies might have been conducted there or in collaboration with international labs. The method comparison refers to NIST/Ghent ID-LC-MS/MS, suggesting international standards and potentially international lab involvement. Based on the 510(k) submission, the studies were conducted by the manufacturer for the purpose of demonstrating substantial equivalence to a US-marketed device. All data is retrospective as it was collected before the submission for marketing clearance.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

There were no experts used to establish ground truth in the traditional sense for these studies. This device is an in vitro diagnostic (IVD) for quantitative measurement. The "ground truth" for the test set was established by:

  • Reference Methods: The primary ground truth for the Method Comparison study was the NIST/Ghent ID-LC-MS/MS 25(OH) Vitamin D Reference Method Procedure. This is a highly accurate and standardized analytical method, not a human expert.
  • Known Concentrations/Spiked Samples: For studies like linearity, detection limits, analytical specificity, and cross-reactivity, ground truth was based on:
    • Known concentrations of analytes (e.g., purified 25(OH)D, interferents, cross-reactants) used to spike samples.
    • Samples with established values from other validated methods, like ID-LC-MS/MS for endogenous cross-reactants.
  • Population Studies: For Expected Values/Reference Range, the "ground truth" are the measured values from a characterized "apparently healthy" donor population, with outliers and specific exclusions defined.

4. Adjudication Method for the Test Set

There was no adjudication method for these studies in the context of expert review, as the ground truth was based on analytical reference methods or predefined sample concentrations, not subjective interpretations requiring consensus from multiple human readers.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

No MRMC comparative effectiveness study was done. This device is an automated in vitro diagnostic (IVD) assay designed for quantitative measurement of a biomarker (25(OH)D). It does not involve human readers interpreting images or data that would be assisted by AI. The device directly yields a numerical result. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This device operates as a standalone algorithm (assay) in the sense that it performs the measurement without human intervention influencing the result of that specific measurement. The IDS-iSYS Multi-Discipline Automated System performs the assay automatically. The results are then used by clinicians in conjunction with other data, but the measurement itself is automated. The performance characteristics described (precision, linearity, detection limits, specificity, method comparison, matrix comparison) are all "standalone" performance metrics of the assay system itself.

7. The Type of Ground Truth Used

The ground truth used was primarily:

  • Reference Method Data: Specifically, the isotope dilution-liquid chromatography/tandem mass spectrometry (ID-LC-MS/MS) 25(OH)D Reference Method Procedure (RMP), particularly for method comparison and traceability. This RMP is itself traceable to the National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 2972.
  • Known/Assigned Concentrations: For stability, calibrator, and control value assignments, internal stock solutions, secondary standards (Internal Reference Calibrators, IRs), and manufactured kit calibrators/controls had values assigned back to the CDC VDSP aligned secondary standards or by using validated kit combinations over multiple instruments and runs.
  • Spiked Samples: For linearity, analytical specificity (interference and cross-reactivity) studies, samples were spiked with known concentrations of analytes, interferents, or cross-reactants.

8. The Sample Size for the Training Set

The document does not explicitly mention a separate "training set" in the context of machine learning or AI development. This device is an immunoassay, not a machine learning model. The various studies described (precision, linearity, specificity, method comparison, etc.) are validation studies performed on the final assay formulation and instrument system.

However, the closest analogous concepts would be:

  • Calibrator and Control Value Assignment: This involves multiple runs (minimum 20 assay runs for calibrators, minimum 21 assay runs for controls) and systems to establish and verify values. These essentially "train" the system by defining its response curve and quality control ranges.
  • Assay Development: During the initial development of the assay, numerous samples and experiments would have been conducted to optimize reagents, conditions, and performance characteristics. This "development data" would serve a similar purpose to a training set but is not explicitly detailed as such in this 510(k) summary, which focuses on validation of the final device.

9. How the Ground Truth for the Training Set Was Established

As noted above, there's no explicit "training set" in the AI sense. For the closest analogous processes:

  • Calibrator and Control Value Assignment:
    • Traceability: The assay is traceable to the ID-LC-MS/MS 25(OH)D Reference Method Procedure (RMP), which was used to assign target values for the Vitamin D Standardization Program (VDSP) samples. This RMP is further traceable to the NIST SRM 2972.
    • Internal Standards: An internal stock solution ("Top Dose") of 25-hydroxyvitamin D3 had its concentration assigned by performing dilutions and testing in approved, previously QC-released kits.
    • Secondary Standards (IRs): These were manufactured and value-assigned using the CDC VDSP aligned secondary standards.
    • Kit Calibrators: Tested as unknowns in a minimum of 20 assay runs on one IDS-iSYS instrument, using these secondary standards (IRs) to establish their values through Excel data reduction and Prism for curve parameters. Values were then verified over 3 assays.
    • Kit Controls: Tested as unknowns in a minimum of 21 assay runs using multiple systems and approved kit combinations, with values calculated using the IDS-iSYS instrument's on-board software. Values were then verified in a single assay.

Essentially, the "ground truth" for the calibrators and controls used to define the assay's operating curve and quality parameters is established through a hierarchical traceability chain leading back to NIST-traceable reference methods.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 15, 2016

IMMUNODIAGNOSTIC SYSTEMS LTD MICK HENDERSON REGULATORY AFFAIRS OFFICER 10 DIDCOT WAY, BOLDON BUSINESS PARK BOLDON NE35 9PD GREAT BRITAIN

Re: K161831

Trade/Device Name: IDS-iSYS 25VitD8, IDS-iSYS 25VitD8 Control Set Regulation Number: 21 CFR 862.1825 Regulation Name: Vitamin D test system Regulatory Class: II Product Code: MRG, JJX Dated: September 30, 2016 Received: October 4, 2016

Dear Mick Henderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For : Courtney H. Lias, Ph.D.

Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161831

Device Name IDS-iSYS 25VitDy IDS-iSYS 25VitDS Control Set

Indications for Use (Describe)

The IDS iSYS 25 VilD assay is intended for the quantitative determination of total 25-hydroxyvitamin D (125(OH)D) in hyman serum or plasma on the IDS iSYS Multi-Discipline Automated System. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in an adult population.

The IDS-iSYS 25 VitD Control Set is used for quality control of the IDS-iSYS 25 VitD assay on the IDS-iSYS Multi-Discipline Automated System.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

D Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Immunodiagnostic Systems (IDS). The logo features the letters "ids" in a stylized, connected script, with a red circle above the "i". Below the letters, the words "immunodiagnosticsystems" are written in a smaller, red font. The overall design is modern and clean, suggesting a company focused on diagnostics.

510(k) SUMMARY

510k NumberK161831
IntroductionAccording to the requirements of 21CFR807.92, the followinginformation provides sufficient detail to understand the basis for adetermination of substantial equivalence.
SubmitterImmunodiagnostic Systems Limited10 Didcot WayBoldon Business ParkBoldonTyne and WearNE35 9PDUnited Kingdom
Contact Person: Mick HendersonPhone: +44 191 5190660Fax: +44 191 5190760Email: mick.henderson@idsplc.com
Secondary Contact: Alexandra BennettPhone: +44 191 5190660Fax: +44 191 5190760Email: alexandra.bennett@idsplc.com
Date prepared: 11th November 2016
Device NameProprietary names: IDS-iSYS 25VitDSIDS-iSYS 25VitDS Control Set
Common names: As above
Classification: 21CFR862.1825 Vitamin D Test System,Class II21CFR862.1660 Quality Control Material(Assayed and Unassayed),Class I, reserved
Product Code: MRGIIX

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Immunodiagnostic Systems (IDS). The logo features the letters 'ids' in a stylized, interconnected design, with the 'i' having a red circular dot above it. Below the gray 'ids' is the text 'immunodiagnosticsystems' in red, presented in a smaller, sans-serif font.

Predicate Device The IDS-iSYS 25 VitDS is substantially equivalent to other products in commercial distribution intended for similar use. We claim equivalency to the currently marketed IDS-iSYS 25-Hydroxy Vitamin DS (K140554).

IDS-iSYS 25 VitDS Device Description

The IDS-iSYS 25 VitDs reagent kit consists of one (1) Immunoassay Cartridge, one (1) vial each of 2 concentration Calibrator levels and a mini-CD containing the Instructions For Use (IFU), CRY files and Certificate of Analysis.

IDS-iSYS 25 VitDS Cartridge, sufficient for 100 tests, consists of reagents provided in individual compartment within a plastic container called the Cartridge. IDS-iSYS 25 VitDS Cartridge contains the following ready to use reagents:

  • Magnetic particles coated with streptavidin in a phosphate . buffer containing sodium azide.
  • . Sodium hydroxide solution.
  • . 25(OH)D labelled with an acridinium ester derivative, in buffer containing bovine serum albumin with sodium azide.
  • . Anti-25(OH)D sheep polyclonal antibody labelled with an biotin, in buffer containing bovine, sheep and mouse proteins and sodium azide.
  • . Assay buffer containing proprietary displacing compounds, methanol and sodium azide.

The IDS-iSYS 25 VitD® Calibrators are included in the reagent kit. The ready to use calibrators contains human serum buffer matrix with two defined concentrations of 25(OH)D and sodium azide as preservative (<0.1 %), 1 bottle per concentration level.

IDS-iSYS 25VitD® Control Set

The IDS-iSYS 25VitDe Control Set contains human serum in a buffer matrix with two defined concentrations of 25(OH)D and sodium azide as a preservative; 3 bottles per concentration level.

The intended use and composition of the calibration verifiers to be used with this assay are identical to the ones previously cleared in K111650 and will not form any part of this 510(k). However they will be now known as IDS-iSYS 25 VitDS Calibration Verifiers and not IDS-iSYS 25 Hydroxy Vitamin D Calibration Verifiers.

  • The IDS-iSYS 25 VitDe Assay is intended for the quantitative Indications for Use determination of total 25-hydroxyvitamin D [25(OH)D] in human

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Immunodiagnostic Systems (IDS). The logo features the letters "ids" in a stylized, sans-serif font, with a red circle above the "i". Below the letters, the words "immunodiagnosticsystems" are written in a smaller, red font. The letters "ids" are in gray.

serum or plasma on the IDS-iSYS Multi-Discipline Automated System. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in an adult population.

The IDS-iSYS 25 VitDo Control Set is used for quality control of the IDS-iSYS 25 VitDS assay on the IDS-iSYS Multi-Discipline Automated System.

For in vitro diagnostic use only. Conditions for use: Rx Only

Special instrument Requirements:

IDS-iSYS Multi-Discipline Automated System (K091849)

Comparison Tables

Similarities compared to the chosen (FDA cleared; marketed) predicate device (K140554)

ItemPredicate DeviceIDS-iSYS 25-Hydroxy Vitamin DS(K140554)Candidate DeviceIDS-iSYS 25 VitDS
Indications for useResults are to be used in conjunctionwith other clinical and laboratory datato assist the clinician in theassessment of vitamin D sufficiencyin an adult population.Same
Intended UseFor the quantitative measurement oftotal 25-hydroxyvitamin DSame
Analyte25-hydroxyvitamin D [25(OH)D]Same
Reagent storage2-8 °CSame
Samplepreparation(pre-treatment)Performed on-board the systemSame
Sample volume10μLSame
Method ofdetection (Testmethodology)Chemiluminescent immunoassayusing magnetic-particle solid phaseand acridinium labelSame
AutomationFully automated assaySame
CalibrationprocedureUser-initiated 2 point calibration toadjust the batch related master curve.The system stores the calibration forthe interval specified in the kit IFU.Same
Traceability/StandardizationTraceable to the isotope dilution-liquid chromatography/tandem massspectrometry (ID-LC-/MS/MS)Same

Assay

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Immunodiagnostic Systems (IDS). The logo consists of the letters "ids" in a stylized, connected font, with a red dot above the "i". Below the letters, the words "immunodiagnosticsystems" are written in a smaller, red font. The letters "ids" are in gray.

25(OH)D Reference MethodProcedure (RMP) which was used inassigning the target value for theVDSP samples.
The ID-LC-MS/MS RMP is traceableto the National Institute of Standardsand Technology Standard ReferenceMaterial (SRM) 2972.
On board thesystem reagentstability21 daysSame

Kit Controls similarities

ItemPredicate DeviceIDS-iSYS 25-Hydroxy Vitamin DSControl Set(K140554)Candidate DeviceIDS-iSYS 25 VitDSControl Set
Intended UseThe quality control of the assay onthe IDS-iSYS.Same
StabilityAfter opening at 2 - 8 °C: To theexpiry dateSame
Reagent storage2-8 °CSame

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Immunodiagnostic Systems (IDS). The logo features the letters "ids" in a stylized, rounded font, with the "i" having a red circular dot above it. Below the letters, the words "immunodiagnosticsystems" are written in a smaller, red font, forming a single word. The overall design is clean and modern, suggesting a company in the healthcare or diagnostics field.

Differences compared to the chosen predicate device (K140554)

Assay

ItemPredicate DeviceIDS-iSYS 25-Hydroxy Vitamin DS(K140554)Candidate DeviceIDS-iSYS 25 VitDS
Calibrator matrixEquine serum buffer matrix with twodefined concentrations of 25(OH)Dand sodium azide as a preservative.Human serum buffer matrixwith two definedconcentrations of 25(OH)Dand sodium azide as apreservative.
Kit reagentcomponentsReagent cartridge (1 vial each ofMPV1, CONJ, NaOH & BUF), twoconcentration levels of calibrators(A&B) (1 vial of each) & a mini CDReagent cartridge (1 vialeach of MPV1, NaOH, 25D-ACR, Ab-BIOT & BUF),two concentration levels ofcalibrators (A&B) (1 vial ofeach) & a mini CD
Quality ControlRequires three controls to validate thecalibrationRequires two controls tovalidate the calibration
Kit reagentcomponentvolumesReagent cartridge (1 vial each):MPV1 (2.0mL), CONJ (10.1mL),NaOH (13mL) & BUF (26.0mL)Reagent cartridge (1 vialeach): MPV1 (2.5mL),NaOH (13mL), 25D-ACR(9.0mL), Ab-BIOT (9.0mL)& BUF (23.0mL)
AntibodiesAnti-25 OH D Sheep Polyclonal IgGSame, but with a differentsource of antibody pool
Calibration interval14 days7 days
Range of assay7 - 125 ng/mL4- 110 ng/mL
SensitivityLoB: 0.6 ng/mLLoD: 2.6 ng/mLLoQ: 7.0 ng/mLLoB: 1.31 ng/mLLoD: 1.98 ng/mLLoQ: 3.53 ng/mL
Expected values12.7 to 64.2 ng/mL10.4 to 59.5 ng/mL
Reference rangeOverall: 8.2 to 91.9 ng/mLOverall: 10.4 to 59.5 ng/mL
In use (afteropening at 2-8°C)reagent stability21 days42 days

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Immunodiagnostic Systems (IDS). The logo features the letters "ids" in a stylized, interconnected design, with the "i" having a red circular dot above it. Below the letters, the words "immunodiagnosticsystems" are written in a smaller, sans-serif font, with the color red.

MethodcomparisonAgainst ID-LC-MS/MSReference Method Procedure(n = 99)Against ID-LC-MS/MSReference Method Procedure(n = 136)
Passing-Bablok regression:IDS-iSYS = $0.95(x) + 0.8$ ng/mLPassing-Bablok regression:IDS-iSYS = $0.99(x) - 0.51$ ng/mL
Correlation coefficient (r) = 0.925Correlation coefficient (r) = 0.97
Sample matrixSerumSerum (standard sampling tubesor tubes containing serumseparating gel) or plasma (K2EDTA, lithium heparin, sodiumheparin)
Sample tube typeSecondary tubePrimary or Secondary tube

Kit Controls differences:

ItemPredicate deviceIDS-iSYS 25-Hydroxy VitaminDS Control Set(K140554)Candidate deviceIDS-iSYS 25 VitDSControl Set
Control matrixEquine serum buffer matrix withthree defined concentrations of25(OH)D and sodium azide as apreservative.Human serum buffer matrix withtwo defined concentrations of25(OH)D and sodium azide as apreservative.
Control KitcomponentsThree concentration levels ofcontrols (3 vials of each) & amini CDTwo concentration levels ofcontrols (3 vials of each) & a miniCD
Quality ControlRequires three buffer basedcontrols for daily Quality ControlRequires two serum basedcontrols for daily Quality Control
StabilityOn board the system: 2.5 hoursOn board the system: 4 hours

Performance Characteristics (if/when applicable):

    1. Analytical performance:
  • a. Precision/Reproducibility:

Precision was determined in accordance with CLSI EP5-A2, "Evaluation of Precision Performance of Quantitative Measurement Methods". Assessment was made for the following variables: within run precision, total precision.

Nine serum based samples at different 25(OH)D concentration levels, ranging from approximately 13.9 ng/mL to approximately 103 ng/mL, to ensure that the assay range of the IDS-iSYS 25 VitDe was covered were used to assess precision of the IDS-iSYS 25 VitD® kit.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Immuno Diagnostic Systems (IDS). The logo features the letters 'ids' in a stylized, interconnected design, with the 'i' having a red circular dot above it. Below the main letters, in smaller red font, is the text 'immunodiagnosticsystems'.

SampleMean Conc.Within RunTotal
ng/mLnmol/LSD%CVSD%CV
ng/mLnmol/Lng/mLnmol/L
113.934.80.71.85.01.02.57.4
216.942.30.71.83.91.23.07.2
328.671.51.12.83.91.94.86.5
437.092.51.33.33.51.84.54.9
549.81252.05.04.02.97.35.8
661.11532.46.03.93.79.36.0
762.11552.87.04.44.010.06.4
893.02333.58.83.85.814.56.2
91032584.812.04.76.315.86.1

Final claims data reported is representative of one manufacture batch (MB3) where 20 days were available for analysis. Results from one representative lot:

b. Linearity/assay reportable range:

A linearity study was conducted based on a modified version of CLSI EP6-A. A high human serum sample (pool of two endogenous spiked serum samples) and a low human serum sample (low serum sample diluted in zero matrix) were analyzed in addition to 9 evenly spaced dilutions which were created by mixing the high and low sample to cover the range of the assay (4 - 110 ng/mL) as indicated below:

SampleDilutionDilution Factor (%)
1:Low (L)0
2:0.90L + 0.10H10
3:0.80L + 0.20H20
4:0.70L + 0.30H30
5:0.60L + 0.40H40
6:0.50L + 0.50H50
7:0.40L + 0.60H60
8:0.30L + 0.70H70
9:0.20L + 0.80H80
10:0.10L + 0.90H90
11:High (H)100

Observed = 1.00 x (Expected) + 3.75 ng/mL Observed = 1.00 x (Expected) + 9.38 nmol/L Regression coefficient R2: 1.00

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Immunodiagnostic Systems (IDS). The logo features the letters 'ids' in a stylized, connected font, with a red circle above the 'i'. Below the letters, in a smaller, red font, is the full name 'immunodiagnosticsystems'. The logo is simple and modern, using a limited color palette.

The IDS-iSYS 25 VitDe assay is linear over the range tested, 3.58 to 136 ng/mL.

The measuring (reportable) range of the assay is from 4 to 110 ng/mL. Any value that read below 4 ng/mL (10 nmol/L) should be reported as "<4 ng/mL" or "<10 nmol/L".

Traceability, Stability, Expected values (controls, calibrators, or methods): C.

Calibrator traceability and value assignment

An internal stock solution (Top Dose) is prepared by reconstituting a vial of 25hydroxyvitamin D3 with ethanol. The concentration of the Top Dose when used in the assay is then assigned by performing dilutions of the Top Dose low charcoal stripped lipid stripped human serum (calibrator serum) or normal non lipid stripped human serum (kit control serum) and testing in an approved kit combination, for example a previously QC released kit. The concentration of the Top Dose is calculated back to the stock concentration by correction of the dilution with serum. This figure is an accurate determination and is assigned to the CDC VDSP aligned calibrators.

A set of secondary standards (Internal reference calibrators, IRs) were manufactured using the same materials as those used in the kit calibrators, however these cover a wider assay range (0 – 168 ng/mL). Kit Calibrators are manufactured and value assigned using the CDC VDSP aligned secondary standards.

Kit Controls are manufactured and value assigned using at least one approved kit combination over three IDS-iSYS instruments using the kit calibrators that were assigned against CDC VDSP aligned internal references.

For kit calibrator value assignment, the kit calibrators are tested as unknowns in a minimum of 20 assay runs on one IDS-iSYS instrument. Each run uses secondary standards (IRs), and the concentrations of the Kit Calibrators are calculated from these secondary standards (IRs) by using Excel data reduction for outliers followed by Prism to establish the curve parameters following an issued QC procedure. Following assignment the kit calibrator values are then verified over 3 assays using the full curve parameters in addition to the established calibrator values. The values must fall within specified acceptable ranges. The kit calibrator concentrations are reagent batch specific and linked together.

For kit control value assignment, the kit controls are tested as unknowns in a minimum of 21 assay runs using multiple systems. Each run uses an approved kit combination e.g. QC previously released kit, and the concentrations of the Kit Controls are calculated as unknown samples using the IDS-iSYS instrument on-board software following an issued QC procedure. The established values are then verified in a single assay in an approved kit combination. The values must fall within specified acceptable ranges.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Immunodiagnostic Systems (IDS). The logo features the letters "ids" in a stylized, lowercase font, with a red circle above the "i". Below the letters, the words "immunodiagnosticsystems" are written in a smaller, red font. The letters "ids" are in gray.

The IDS-iSYS 25 VitDo assay is traceable to the isotope dilution-liguid chromatography/tandem mass spectrometry (ID-LCMS/MS) 25(OH)D Reference Method Procedure (RMP) which was used in assigning the target value for the Vitamin D Standarization Program (VDSP) samples. The ID-LCMS/MS RMP is traceable to the National Institute of Standards and Technology Standard Reference Material (NIST SRM) 2972.

Stability

The kit stability was performed in which the kit calibrators were tested in combination with all kit combination reagents. The stability based on accelerated studies determined a minimum shelf life of 12 months. Real time stability studies are on-going and shelf life will be amended accordingly.

Kit control stability is based on accelerated studies where a minimum shelf life of 12 months was determined.

d. Detection limit:

The limit of blank (LoB), limit of detection (LoD) and limit of quantitation (LoQ) were determined with guidance from CLSI EP17-A, "Protocols for Determination of Limits of Detection and Limits of Quantitation". For the measurement of LoB, LoD and LoQ three manufacture batches (MB1, MB2B and MB3) were tested.

Each LoB sample was run in duplicate. Lot MB1 and MB2B each ran a total of 6 assays over 4 days and MB3 ran 6 assays over 5 days, for a total of 60 replicates per batch. Two operators were used and where two assays were performed in one day the set up time between assays start time was at least 2 hours. Each manufacture batch was tested on a different system.

The LoD was 6 very low 25(OH)D samples. Each LoD sample was run in duplicate. For MB1 a total of 6 assays were run over 3 days, MB2B ran 6 assays over 4 days and MB3 ran 6 assays over 5 days, for a total of 72 replicates for each manufacture batch. Two operators were used and where two assays were performed in one day the set up time between assays start time was at least 2 hours. Each manufacture batch was tested on a different system.

For calculation of LoQ 10 samples were used, these all had low levels of 25(OH)D in them. Each LoO sample was run in duplicate. For MB1 a total of 6 assays were run over 3 days, MB2B ran 6 assays over 4 days and MB3 ran 6 assays over 5 days, for a total of 120 replicates for each manufacture batch. For MB1, 1 sample was miss-labelled and excluded from data analysis resulting in a total of 118 replicates being available. Two operators were used for the LoQ assays and where two assays were performed in one day the set up time between assays start time was at least 2 hours. Each manufacture batch was tested on a different system.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Immunodiagnostic Systems (IDS). The logo features the letters "ids" in a stylized, connected font, with a red circle above the "i". Below the letters, the words "immunodiagnosticsystems" are written in a smaller, red font. The logo is simple and modern, with a focus on the company's name and its area of expertise.

SensitivityConcentration
LoB1.31 ng/mL (3.28 nmol/L)
LoD1.98 ng/mL (4.95 nmol/L)
LoQ3.53ng/mL (8.83 nmol/L)

e. Analytical specificity:

Interference and cross-reactivity studies were performed in accordance with the CLSI EP7-A2"Interference testing in clinical chemistry".

To determine potential interference in the specific detection of 25(OH)D, two serum samples at two different concentrations of 25(OH)D were spiked with the potential interferent. Control samples (blank) were spiked with a volume of Phosphate Buffer saline (PBS) (0 ng/mL) or relevant diluent equal to that of the spiked interferent. The mean of 26 replicate assays, for both spiked and control samples, were then compared. The differences observed between the mean spiked and control sample values were examined and assessed according to acceptance criteria of non-significant interference of <10% between the test and control samples.

% Interference was calculated using the formula below:

$$% \text{ Therefore} = \underbrace{(\text{mean spilled concentration} - \text{mean un-spilled concentration})}_{\text{mean un-spilled concentration}} \ge 100$$

Cholesterol interference was tested by recovery of 25(OH)D from a high serum pool with 25(OH)D at 107.3 ng/mL spiked into a serum sample with a known cholesterol levels. 10% and 20% of high 25(OH)D was spiked into the cholesterol sample. The percent recovery was calculated using the formula below:

Recovery value = Observed mean spiked value – Observed mean un-spiked value % Recovery = (Recovery value / Spike value) x 100

Rf Interference was tested by spiking a sample with known levels of Rf into two serum base samples. The 25(OH)D in the base samples were calculated at 32.6 ng/mL and 54.9 ng/mL. The 25(OH)D in the Rf sample was calculated to be 30.7 ng/mL. The level of interference was calculated using the formula below:

Observed concentration Interference % = (Expected conc. from base sample + Expected conc. from Rf sample)

The expected values were calculated by taking into account the amount each sample was diluted by following the equation below:

Expected concentration of 25(OH)D = 25(OH)D concentration in neat sample (Total sample volume / volume Rf or base sample)

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Immunodiagnostic Systems (IDS). The logo features the letters 'ids' in a stylized, lowercase font, with a red dot above the 'i'. Below the letters, in a smaller, red font, is the text 'immunodiagnosticsystems'. The overall design is clean and modern.

The following table provides the levels of interferents tested and the design acceptance criteria for specificity of ≤10% bias between the test and control samples:

Potential interferentHighest concentrationtested that demonstratedno significant interference25(OH)D concentration(ng/mL)
low controlhigh control
Triglycerides (Intralipid)500 mg/dL31.461.4
Hemoglobin500 mg/dL33.254.8
Bilirubin, conjugated20 mg/dL30.359.4
Bilirubin, unconjugated20 mg/dL23.267.4
Total Protein10 g/dL32.030.5
Human Anti MouseAntibody (HAMA)1000 ng/mL33.865.6
Red Blood Cells0.2%29.857.7
Vitamin D BindingProtein (GC Globulin)2000 ng/mL30.061.7
Biotin200nM34.964.6
Acetaminophen200 µg/mL34.760.2
Ibuprofen140 µg/mL32.666.7
Carbamazepine30 µg/mL28.367.4
Phenytoin50 µg/mL32.262.2
Rheumatoid Factor600 IU/mL32.654.9
Cholesterol Total300 mg/dL8.6N/A

Exogenous and Endogenous cross-reactants where tested:

Exogenous cross-reactants were prepared by manufacturing a 'top dose' of the relevant analyte (in ethanol typically) which was then diluted down to a range of stock concentrations in undetectable 25(OH)D matrix. The stock cross reactant or undetectable 25(OH)D matrix was spiked directly into zero serum (no measurable 25(OH)D), a low sample and a high sample. The cross reactivity was then determined using the following equation:

% cross reactivity =

(Mean concentration of spiked sample - mean concentration of un-spiked sample) x100% Spike concentration

Where:

Mean concentration of spiked sample is determined by the candidate device Mean concentration of un-spiked sample is determined by the candidate device Spiked concentration is calculated by dilution of known concentration Top Dose

Endogenous cross-reactants (25(OH)D3 and 25(OH)D2) were prepared by spiking samples with known levels of 25(OH)D3 and 25(OH)D2 analytes with sample also with known levels of 25(OH)D3 and 25(OH)D2 analytes. The samples which must be used for endogenous cross reactivity tests are supplied where the levels of 25(OH)D3 and

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for Immunodiagnostic Systems (IDS). The logo features the letters 'ids' in a stylized, lowercase font, with a red circle above the 'i'. Below the letters, the words 'immunodiagnosticsystems' are written in a smaller, red font. The overall design is modern and clean.

25(OH)D2 have been established ID-LCMS/MS. The cross reactivity was then determined using the following equation:

% cross reactivity =

(Mean concentration of spiked sample - mean concentration of un-spiked sample) x100% Spike concentration

Where:

Mean concentration of spiked sample is determined by the candidate device Mean concentration of un-spiked sample is determined by LC-MS/MS Spiked concentration is determined by LC-MS/MS

Potential cross-reactantSpike concentration tested (ng/mL)% Cross Reactivity
25 (OH)D320101%
25 (OH)D220105%
24,25 dihydroxyvitamin D325197%
24,25 dihydroxyvitamin D22537%
25,26 dihydroxyvitamin D31054%
1,25 dihydroxyvitamin D3203%
1,25 dihydroxyvitamin D2208%
3-epi-25-OH vitamin D31000%
3-epi-25-OH vitamin D21002%
Alfacalcidol5000%
Cholecalciferol10001%
Ergocalciferol1005%
Paricalcitol100-2%

2. Comparison studies:

Method comparison:

The IDS-iSYS 25 VitDo assay was compared against the NIST/Ghent ID-LC-MS/MS 25(OH) Vitamin D Reference Method Procedure, following CLSI EP-9A2 "Method Comparison and Bias Estimation Using Patient Samples". A total of 136 human serum samples, selected to represent a wide range of 25(OH)D concentrations (5.6 to 110 ng/mL, 14.0 to 275 nmol/L), were tested in on the IDS-iSYS 25 VitDe assay.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for Immunodiagnostic Systems (IDS). The logo features the letters 'ids' in a stylized, gray font, with a red circle above the 'i'. Below the letters, the words 'immunodiagnosticsystems' are written in a smaller, red font. The logo is simple and modern, with a focus on the company's name and its field of expertise.

UnitnSlope95% CIIntercept95% CICorrelationCoefficient (r)
ng/mL1360.990.94 to 1.05-0.51-1.93 to 0.750.97
nmol/L1360.990.94 to 1.05-1.28-4.83 to 1.880.97

Passing-Bablok regression analysis was performed on the comparative data:

3. Expected values/Reference range:

There is no universal agreement on the optimal concentration of 25(OH)D. Ranges should be based on clinical decision values that apply to both sexes of all ages rather than population based reference ranges for 25(OH)D.

In the case of 25(OH)D, there are also many other factors that may influence values: diet, time of day, sun exposure, season of year, geographic location, age, use of sunscreen and/or protective clothing and skin pigmentation.

The 25(OH)D concentration were measured in serum samples collected from three hundred and ninety-two (392) apparently healthy donors using the IDS-iSYS 25 VitD assay. To represent a broad spectrum of UV light exposure, the study population included 200 males and 192 females from three diverse regions of the United States (North, Central and South) that were sampled in the winter and were representative of the overall United Stated population in term of gender, race and ethnicity. Individuals who were pregnant, had bariatric surgery, had personal history of kidney disease and GI disease, had family history of parathyroid or calcium regulatory disease, had abnormal serum calcium, phosphorus, PTH or TSH level were excluded from the study. The study cohort included subjects from 21 to 89 years of age and 40% reported taking vitamin supplements.

The observed ranges were established, according to CLSI C28-A3c "Defining. Establishing and Verifying Reference Intervals in the Clinical Laboratory" is summarised in the table below:

nMeanConcentrationMedianConcentrationObserved Range(2.5th to 97.5th percentile)
39230.7 ng/mL29.5 ng/mL10.4 ng/mL to 59.5 ng/mL
39276.8 nmol/L73.8 nmol/L26.0 nmol/L to 149 nmol/L

The above ranges should be considered as guidelines only. It is recommended that each laboratory to establish its own expected range, representative of its typical population.

4. Matrix Comparison Study:

The IDS-iSYS 25 VitDS matrix comparison study was performed to evaluate the difference across tube types (serum separator tubes (SST), lithium heparin plasma, sodium heparin plasma and K2 EDTA plasma) versus the control samples (red top serum, without additive) following the CLSI EP9-A3 guideline. Fifty seven (57) native samples

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the logo for Immunodiagnostic Systems (IDS). The logo features the letters 'ids' in a stylized, lowercase font, with the 'i' having a red circular dot above it. Below the 'ids' is the full name 'immunodiagnosticsystems' in a smaller, red font. The overall design is clean and modern.

were tested with ten (10) treated (spiked or diluted) samples to cover the range of 4.8 to 108 ng/mL (serum without additives).

Sample TypenSlope95% Cl.Intercept(ng/mL)95% Cl.Corr.Coeff. (r)
SST670.980.94 to 1.020.18-0.31 to 1.120.99
EDTA670.960.94 to 0.980.05-0.31 to 0.591.00
Lithium Heparin670.980.93 to 1.010.22-0.52 to 0.861.00
Sodium Heparin670.990.95 to 1.020.05-0.66 to 0.710.99

Passing-Bablok regression analysis was performed on the comparative data:

The IDS-iSYS 25 VitD assay is suitable for use with human serum (red top serum without additive, serum separator tubes SST) and plasma (lithium heparin, sodium heparin and K2 EDTA).

Conclusion:

The IDS-iSYS 25 VitDe assay and IDS-iSYS 25 VitD8 Control Set data presented and provided is complete and supports the basis for substantial equivalence to the predicate device.

§ 862.1825 Vitamin D test system.

(a)
Identification. A vitamin D test system is a device intended for use in clinical laboratories for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used in the assessment of vitamin D sufficiency.(b)
Classification. Class II (special controls). Vitamin D test systems must comply with the following special controls:(1) Labeling in conformance with 21 CFR 809.10 and
(2) Compliance with existing standards of the National Committee on Clinical Laboratory Standards.