AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ADVIA Centaur® Testosterone II (TSTII)

The ADVIA Centaur Testosterone II (TSTII) assay is for in vitro diagnostic use in the quantitative determination of total testosterone (bound and unbound) in human serum and plasma using the ADVIA Centaur XP system.

Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females, hirsutism (excessive hair) and virilization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

ADVIA Centaur® Testosterone II (TSTII) Master Curve Material

The ADVIA Centaur Testosterone II (TSTI) Master Curve Material is for in vitro diagnostic use in the verification of calibration and reportable range of the ADVIA Centaur Testosterone II (TSTII) assay.

ADVIA Centaur® SHBG

The ADVIA Centaur SHBG assay is an in vitro diagnostic immunoassay for the quantitative determination of sex hormone-binding globulin (SHBG) in human serum and plasma using the ADVIA Centaur XP system.

The ADVIA Centaur SHBG assay is intended for use as an aid in the diagnosis of androgen disorders.

ADVIA Centaur® SHBG Calibrator

For in vitro diagnostic use in calibrating the ADVIA Centaur SHBG assay on the ADVIA Centaur systems.

ADVIA Centaur® SHBG Master Curve Material

The ADVIA Centaur SHBG Master Curve Material is for in vitro diagnostic use in the verification of calibration and reportable range of the ADVIA Centaur SHBG assay.

Device Description

The submission describes several components:

ADVIA Centaur TSTII Assay Components: Includes ADVIA Centaur TSTII Primary Reagent ReadyPack (Lite Reagent and Solid Phase Reagent), ADVIA Centaur TSTII Ancillary Reagent ReadyPack (Releasing Agent), and ADVIA Centaur TSTII Calibrator (Low and High Calibrators). The ingredients are listed in Table 6.

ADVIA Centaur TSTII Master Curve Material: Includes ADVIA Centaur TSTII MCM1 and MCM2-7. The ingredients are listed in Table 6.

ADVIA Centaur SHBG Assay Components: Includes ADVIA Centaur SHBG Primary Reagent ReadyPack (Lite Reagent and Solid Phase Reagent), and ADVIA Centaur SHBG Ancillary Well Reagent. The ingredients are listed in Table 7.

ADVIA Centaur SHBG Calibrator: Includes ADVIA Centaur SHBG Low and High Calibrators. The ingredients are listed in Table 7.

ADVIA Centaur SHBG Master Curve Material: Includes ADVIA Centaur SHBG MCM1-5. The ingredients are listed in Table 8.

The ADVIA Centaur TSTII assay uses a chemiluminescence competitive immunoassay methodology. The ADVIA Centaur SHBG assay uses a chemiluminescence sandwich immunoassay methodology. Both assays are intended for use on the ADVIA Centaur XP system.

AI/ML Overview

The provided document describes the acceptance criteria and study results for the ADVIA Centaur Testosterone II (TSTII) assay and the ADVIA Centaur SHBG assay.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

Since the document does not explicitly present a table of predetermined acceptance criteria with a pass/fail column, I will infer the acceptance criteria from the context provided (e.g., studies performed and expected deviations/recoveries). The "Reported Device Performance" column will contain the results from the studies presented.

ADVIA Centaur Testosterone II (TSTII) Assay:

Performance CharacteristicInferred Acceptance Criteria (Based on typical IVD expectations and study narrative)Reported Device Performance
PrecisionCVs generally <5-10% at relevant concentrationsRepeatability: 3.1-8.5% CV (Lowest at 768.74 ng/dL, highest at 10.58 ng/mL) Within-Lab: 4.5-12.6% CV (Lowest at 226.34 ng/mL, highest at 10.58 ng/mL)
LinearityBias from linear fit estimate < 10% (or specified acceptable range)Bias from linear fit estimate: < 10% for most samples, up to 12.71% at the lowest end (6.72 ng/dL) Regression: Observed = 0.92(Expected) - 0.22 ng/dL
Dilution RecoveryRecoveries within an acceptable range (e.g., 90-110%)Range: 93.9% to 114.2% Mean: 106.7%
Method ComparisonGood correlation (r > 0.95) and minimal bias compared to reference method/predicateCDC HoSt Testosterone ID-LC-MS/MS: ADVIA Centaur TSTII = 0.97(ID-LC-MS/MS) + 1.94 ng/dL (r = 0.98) Dimension Vista TST (Adult & Pediatric): ADVIA Centaur TSTII = 1.01(Dimension Vista TST) + 0.99 ng/dL (r = 0.99) Dimension Vista TST (Pediatric): ADVIA Centaur TSTII = 1.04(Dimension Vista TST) + 0.22 ng/dL (r = 0.99)
Matrix ComparisonNo significant difference or acceptable correlation between tube typesSlope: 0.94-0.98, Intercept: -1.61 to 4.50 ng/dL, r: 0.991-0.996 (No significant difference observed)
Detection Limits (LoQ)LoQ should be clinically relevant and achievableLoQ: 7.00 ng/dL
InterferenceAcceptable interference percentages for common interferentsHemoglobin (-0.4 to 5.8%), Triglycerides (-8.5 to 1.2%), Conjugated Bilirubin (-4.1 to 4.2%), Unconjugated Bilirubin (3.2 to 9.6%), Cholesterol (-9.4 to 6.4%), Rheumatoid Factor (0.7 to 4.6%). Biotin: 10% interference at 30 ng/mL. Conjugated Bilirubin: >15 mg/dL causes erroneous results (~-14.4% change at 20 mg/dL). Unconjugated Bilirubin: >20 mg/dL causes erroneous results (~20.1% change at 30 mg/dL).
Heterophile InterferenceMean recovery close to 100% (e.g., 90-110%)Mean recovery for ~30 ng/dL: 98.69% Mean recovery for ~300 ng/dL: 94.55%
Cross-ReactivitySpecificity for testosterone with minimal cross-reactivity from other steroidsVaried, with some substances showing higher cross-reactivity (e.g., 11β-Hydroxytestosterone, Ethisterone, Testosterone propionate). Specific percentages are provided in the table.

ADVIA Centaur SHBG Assay:

Performance CharacteristicInferred Acceptance Criteria (Based on typical IVD expectations and study narrative)Reported Device Performance
PrecisionCVs generally <5-10% at relevant concentrationsRepeatability: 1.8-3.0% CV Within-Lab: 1.9-4.4% CV
LinearityBias from linear fit estimate < 15% (or specified acceptable range)Bias from linear fit estimate: ≤15% across the range Regression: Observed = 0.984(Expected) + 0.245 ng/dL
Dilution RecoveryRecoveries within an acceptable range (e.g., 90-110%)Range: 87% to 101% Mean: 92%
Method ComparisonGood correlation (r > 0.95) and minimal bias compared to predicateADVIA Centaur SHBG = 0.99(Elecsys) - 0.11 nmol/L (r = 0.99)
Detection Limits (LoQ)LoQ should be clinically relevant and achievableLoQ: 1.8 nmol/L

2. Sample sizes used for the test set and the data provenance:

  • ADVIA Centaur Testosterone II (TSTII) Assay:

    • Precision: 5 serum patient pools, 3 control levels, 1 female patient serum specimen. Assayed twice a day in replicates of 2 for 20 days (n = 80 replicates per sample).
    • Linearity: 11 serially diluted samples. Each sample tested in duplicate.
    • Dilution Recovery: 5 serum samples.
    • Method Comparison:
      • Against CDC HoSt Testosterone ID-LC-MS/MS: 128 serum samples.
      • Against Dimension Vista Total Testosterone: 117 serum samples (adults) and 61 samples (pediatrics). The pediatric samples consisted of 27 females (age 7-18) and 34 males (age 9-18).
    • Matrix Comparison: Minimum of 50 matched specimens for each tube type (serum red top, serum separator tube, plasma separator tube, dipotassium EDTA, lithium heparin, and sodium heparin).
    • Reference Intervals:
      • Adults: 119-167 samples per group (Males Age <50, Males Age ≥50, Females Pre-Menopause, Females Post-Menopause).
      • Pediatrics: 20-130 samples per group (based on age and Tanner stage). Samples were collected prospectively from apparently healthy pediatric subjects.
    • Interference: Two sample pools (approx. 30 ng/dL and 300 ng/dL testosterone), spiked with potential interferents.
    • Heterophile Interference: 19 HAMA-positive serum samples.
    • Cross-Reactivity: Two sample pools (approx. 0 ng/dL and 300 ng/dL testosterone), spiked with potential cross-reactants.
    • Data Provenance: Not explicitly stated for all studies, but reference interval samples for pediatrics were "collected prospectively". It is generally inferred for in vitro diagnostic device submissions that data is from a clinical laboratory setting, likely in the US or regions where regulatory standards are aligned. No specific country of origin is mentioned for all samples.
  • ADVIA Centaur SHBG Assay:

    • Precision: 3 patient serum specimens, 2 patient pools, 3 control levels. Assayed once a day in replicates of 5 for 5 days (n = 25 replicates per sample).
    • Linearity: 11 serially diluted samples. Each sample tested in duplicate.
    • Dilution Recovery: 5 serum samples.
    • Method Comparison: 174 serum samples against Elecsys SHBG assay.
    • Data Provenance: Not explicitly stated.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document. For in vitro diagnostic assays like these, the "ground truth" is typically established through a reference method (e.g., ID-LC-MS/MS for testosterone) or a well-established predicate device, rather than through expert consensus in the same way it might be for image analysis. The document details traceability to international standards (e.g., WHO 2nd International Standard for SHBG) and reference methods.

4. Adjudication method for the test set:

This information is not applicable as the context is an in vitro diagnostic assay evaluating quantitative measurements against reference methods or predicate devices, not subjective assessments requiring expert adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. The device is an in vitro diagnostic assay for quantitative determination of hormones, not an AI-powered diagnostic tool requiring human reader studies or MRMC studies.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This is not applicable as the device is a laboratory assay system. Its performance characteristics (precision, linearity, accuracy, etc.) are inherently "standalone" in the sense that the system itself provides the quantitative result. There is no human-in-the-loop component in terms of result interpretation that would typically apply to AI/imaging diagnostics.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • ADVIA Centaur Testosterone II (TSTII) Assay:

    • Traceability/Standardization: ID-LC-MS/MS (CDC HoSt Testosterone Reference Measurement Procedure), which is traceable to the primary testosterone standard National Measurement Institute (NMI) M914. This is a highly accurate chemical reference method.
    • Method Comparison: Comparison against the CDC HoSt ID-LC-MS/MS method (reference method) and another legally marketed predicate device (Dimension Vista Total Testosterone assay).
    • Reference Intervals: Established using samples from "apparently healthy" individuals, with methodologies aligned with CLSI guideline EP28-A3c.
  • ADVIA Centaur SHBG Assay:

    • Traceability/Standardization: Standardized to World Health Organization (WHO) 2nd International Standard for SHBG, NIBSC code 08/266.
    • Method Comparison: Comparison against a legally marketed predicate device (Elecsys SHBG assay).

8. The sample size for the training set:

This information is not provided in the document. For in vitro diagnostic assays like these, there isn't typically a "training set" in the machine learning sense. Instead, assay development involves extensive internal testing and optimization (sometimes referred to as development or verification lots) before the formal validation studies presented in this document. The sample sizes listed above are for the validation (test) phase.

9. How the ground truth for the training set was established:

As mentioned in point 8, a "training set" in the machine learning context is not applicable here. The development and optimization of the assay would rely on established chemical and biological principles and internal reference materials, which would themselves be traced to primary standards or reference methods as described in point 7 for the "ground truth" of the final product.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other to create a sense of depth and connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

SIEMENS HEALTHCARE DIAGNOSTICS INC. MATTHEW GEE SENIOR MANAGER, REGULATORY AFFAIRS 511 BENEDICT AVENUE TARRYTOWN NY 10591-5097

June 29, 2016

Re: K151986 Trade/Device Name: ADVIA Centaur Testosterone II (TSTII), ADVIA Centaur Testosterone II (TSTII) Master Curve Material, ADVIA Centaur SHBG. ADVIA Centaur SHBG Calibrator, ADVIA Centaur SHBG Master Curve Material Regulation Number: 21 CFR 862.1680 Regulation Name: Testosterone test system Regulatory Class: I, reserved Product Code: CDZ, JIT, JJX Dated: June 3, 2016 Received: June 7, 2016

Dear Mr. Matthew Gee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements

{1}------------------------------------------------

as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For:

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151986

Device Name

ADVIA Centaur® Testosterone II (TSTII) ADVIA Centaur® Testosterone II (TSTII) Master Curve Material

Indications for Use (Describe) ADVIA Centaur® Testosterone II (TSTII)

The ADVIA Centaur Testosterone II (TSTII) assay is for in vitro diagnostic use in the quantitative determination of total testosterone (bound and unbound) in human serum and plasma using the ADVIA Centaur XP system.

Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females, hirsutism (excessive hair) and virilization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

ADVIA Centaur® Testosterone II (TSTII) Master Curve Material

The ADVIA Centaur Testosterone II (TSTI) Master Curve Material is for in vitro diagnostic use in the verification of calibration and reportable range of the ADVIA Centaur Testosterone II (TSTII) assay.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K151986

Device Name ADVIA Centaur® SHBG ADVIA Centaur® SHBG Calibrator ADVIA Centaur® SHBG Master Curve Material

Indications for Use (Describe) ADVIA Centaur® SHBG

The ADVIA Centaur SHBG assay is an in vitro diagnostic immunoassay for the quantitative determination of sex hormone-binding globulin (SHBG) in human serum and plasma using the ADVIA Centaur XP system.

The ADVIA Centaur SHBG assay is intended for use as an aid in the diagnosis of androgen disorders.

ADVIA Centaur® SHBG Calibrator

For in vitro diagnostic use in calibrating the ADVIA Centaur SHBG assay on the ADVIA Centaur systems.

ADVIA Centaur® SHBG Master Curve Material

The ADVIA Centaur SHBG Master Curve Material is for in vitro diagnostic use in the verification of calibration and reportable range of the ADVIA Centaur SHBG assay.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92 and the Safe Medical Device Act of 1990.

The assigned 510(k) Number is: K151986

Date Prepared 1.

June 27, 2016

Applicant Information 2.

Contact:Matthew Gee, M.Sc.Senior Manager, Regulatory Affairs
Address:Siemens Healthcare Diagnostics Inc.511 Benedict AvenueTarrytown, NY 10591-5097
Phone:914-524-2099
Fax:914-524-3579

Email: matthew.gee@siemens.com

3. Regulatory Information

Table 1. Regulatory Information for ADVIA Centaur TSTII Assay

Trade NameADVIA Centaur® Testosterone II (TSTII)
Model Numbers10696862 (1-pack); 10696863 (5-pack)
Common NameRadioimmunoassay, testosterones and dihydrotestosterone
Classification NameTestosterone test system
FDA ClassificationClass I (Reserved)
Review PanelClinical Chemistry (75)
Product CodeCDZ
Regulation Number862.1680

Table 2. Regulatory Information for ADVIA Centaur TSTII Master Curve Material

Trade NameADVIA Centaur® Testosterone II (TSTII) Master Curve Material
Model Numbers10492623
Common NameSingle (specified) analyte controls (assayed and unassayed)
Classification NameQuality control material (assayed and unassayed)
FDA ClassificationClass I
Review PanelClinical Chemistry (75)
Product CodeJJX
Regulation Number862.1660

{5}------------------------------------------------

Table 3. Regulatory Information for ADVIA Centaur SHBG Assay

Trade NameADVIA Centaur® SHBG
Model Numbers10997191 (1-pack)
Common NameRadioimmunoassay, testosterones and dihydrotestosterone
Classification NameTestosterone test system
FDA ClassificationClass I (Reserved)
Review PanelClinical Chemistry (75)
Product CodeCDZ
Regulation Number862.1680

Table 4. Regulatory Information for ADVIA Centaur SHBG Calibrator

Trade NameADVIA Centaur® SHBG Calibrator
Model Numbers10997192
Common NameCalibrator, Secondary
Classification NameCalibrator
FDA ClassificationClass II
Review PanelClinical Chemistry (75)
Product CodeJIT
Regulation Number862.1150

Table 5. Regulatory Information for ADVIA Centaur SHBG Master Curve Material

Trade NameADVIA Centaur® SHBG Master Curve Material
Model Numbers10997193
Common NameSingle (specified) analyte controls (assayed and unassayed)
Classification NameQuality control material (assayed and unassayed)
FDA ClassificationClass I
Review PanelClinical Chemistry (75)
Product CodeJJX
Regulation Number862.1660

Predicate Device Information 4.

ADVIA Centaur Testosterone II (TSTII)

Predicate Device Name: Elecys Testosterone II

510(k) Number: K093421

ADVIA Centaur Testosterone II (TSTII) Master Curve Material

Predicate Device Name: ADVIA Centaur Testosterone (TSTO) Master Curve Material

510(k) Number: K140505

{6}------------------------------------------------

ADVIA Centaur SHBG

Predicate Device Name: ADVIA Centaur SHBG

510(k) Number: K091867

ADVIA Centaur SHBG Calibrator

Predicate Device Name: ADVIA Centaur SHBG Calibrator

510(k) Number: K091867

ADVIA Centaur SHBG Master Curve Material

Predicate Device Name: ADVIA Centaur SHBG Master Curve Material

510(k) Number: K091867

Intended Use / Indications for Use 5.

ADVIA Centaur Testosterone II (TSTII)

The ADVIA Centaur Testosterone II (TSTII) assay is for in vitro diagnostic use in the quantitative determination of total testosterone (bound and unbound) in human serum and plasma using the ADVIA Centaur XP system.

Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females, hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenoqenital syndromes.

ADVIA Centaur TSTII Master Curve Material

The ADVIA Centaur Testosterone II (TSTII) Master Curve Material is for in vitro diagnostic use in the verification of calibration and reportable range of the ADVIA Centaur Testosterone II (TSTII) assay.

ADVIA Centaur SHBG

The ADVIA Centaur SHBG assay is an in vitro diagnostic immunoassay for the quantitative determination of sex hormone-binding globulin (SHBG) in human serum and plasma using the ADVIA Centaur XP system.

The ADVIA Centaur SHBG assay is intended for use as an aid in the diagnosis of androgen disorders.

ADVIA Centaur SHBG Calibrator

For in vitro diagnostic use in calibrating the ADVIA Centaur SHBG assay on the ADVIA Centaur systems.

ADVIA Centaur SHBG Master Curve Material

The ADVIA Centaur SHBG Master Curve Material is for in vitro diagnostic use in the verification of calibration and reportable range of the ADVIA Centaur SHBG assay.

{7}------------------------------------------------

6. Device Description

Table 6. Summary of Ingredients of the ADVIA Centaur TSTII Assay Components

ComponentVolumeIngredients
ADVIA Centaur TSTII Primary Reagent ReadyPack (included in assay kit)
ADVIA Centaur TSTII LiteReagent10.0 mL/packAcridinium ester-labeled hapten (36 $μ$ g/mL) inbuffered saline with preservatives
ADVIA Centaur TSTIISolid Phase Reagent17.0 mL/packStreptavidin-coated latex particles (0.33 g/L) inbuffered saline with preservatives
ADVIA Centaur TSTII Ancillary Reagent ReadyPack (included in assay kit)
ADVIA Centaur TSTIIReleasing Agent10.0 mL/packSteroid releasing agent (0.4 $μ$ g/mL) and biotinylatedsheep monoclonal anti-testosterone antibody (27 $μ$ g/L) in buffered saline and preservatives
ADVIA Centaur TSTII Calibrator (included in assay kit)
ADVIA Centaur TSTIILow and High Calibrators2.0 mL/vialAfter reconstitution, low or high levels of USP-gradeestradiol, testosterone, cortisol, and progesteronespiked in charcoal-stripped, defibrinated humanplasma with sodium azide (0.1%) and preservatives
ADVIA Centaur TSTII Master Curve Material (sold separately)
ADVIA Centaur TSTIIMCM11.0 mL/vialLyophilized human plasma with sodium azide (0.1%after reconstitution) and preservatives
ADVIA Centaur TSTIIMCM2-71.0 mL/vialVarious levels of testosterone in lyophilized humanplasma with sodium azide (0.1% after reconstitution)and preservatives

Table 7. Summary of Ingredients of the ADVIA Centaur SHBG Assay Components

ComponentVolumeIngredients
ADVIA Centaur SHBG Primary Reagent ReadyPack (included in assay kit)
ADVIA Centaur SHBGLite Reagent3.0 mL/packMouse monoclonal anti-SHBG antibody (~130 ng/mL)labeled with acridinium ester in HEPES-bufferedsaline with bovine serum albumin, mouse serum,sodium azide (< 0.1%), surfactant, and preservatives
ADVIA Centaur SHBGSolid Phase Reagent11.0 mL/packStreptavidin coupled to latex paramagnetic particles(~150 $\mu$ g/mL) in HEPES buffered saline with bovineserum albumin, mouse serum, sodium azide (< 0.1%),surfactant, and preservatives
ADVIA Centaur SHBGAncillary Well Reagent3.0 mL/packBiotinylated monoclonal mouse anti-SHBG antibody(~6 $\mu$ g/mL) in HEPES buffered saline with bovineserum albumin, mouse serum, sodium azide (< 0.1%),surfactant, and preservatives
ADVIA Centaur SHBG Calibrator (sold separately)
ADVIA Centaur SHBGLow and High Calibrators2.0 mL/vialAfter reconstitution, low or high levels of SHBG inequine serum, detergents and preservatives

{8}------------------------------------------------

ComponentVolumeIngredients
ADVIA Centaur SHBG Master Curve Material (sold separately)
ADVIA Centaur SHBGMCM1-51.0 mL/vialVarious levels of SHBG in equine serum withdetergents and preservatives

Purpose of the Submission 7.

The purpose of this submission is to submit a new device (ADVIA Centaur TSTII assay) and to submit a modification to the ADVIA Centaur SHBG assay due to the re-standardization to the WHO 2nd International Standard for SHBG (08/266).

Comparison of Predicate Device and Modified Device 8.

ItemADVIA Centaur Testosterone II(TSTII) Assay(Candidate Device)Elecsys Testosterone II Assay(Predicate Device)
Intended UseFor in vitro diagnostic use in thequantitative determination of totaltestosterone (bound and unbound) inserum using the ADVIA Centaur andADVIA Centaur XP Systems.Immunoassay for the in vitroquantitative determination oftestosterone in human serum andplasma. Theelectrochemiluminescenceimmunoassay "ECLIA" is intendedfor use on Elecsys and cobas eimmunoassay analyzers.
Indicationsfor UseMeasurements of testosterone are usedin the diagnosis and treatment ofdisorders involving the male sexhormones (androgens), includingprimary and secondary hypogonadism,delayed or precocious puberty,impotence in males and, in femaleshirsutism (excessive hair) and virilization(masculinization) due to tumors,polycystic ovaries, and adrenogenitalsyndromes.Same
MethodologyChemiluminescenceElectrochemiluminescence
Assay ProtocolCompetitive immunoassaySame
Traceability/StandardizationID-LC-MS/MS (CDC HoSt TestosteroneReference Measurement Procedure)ID-GC/MS
Specimen TypeHuman serum and plasmaSame
Sample Volume20 µLSame
Lower Limit ofMeasuring RangeLoQLoD
Measuring Range7.0-1500 ng/dL2.5-1500 ng/dL
Calibration2-point calibrationSame

Table 8. Comparison of ADVIA Centaur TSTII Assay to Predicate

{9}------------------------------------------------

ItemADVIA Centaur Testosterone II(TSTII) MCM(Candidate Device)ADVIA Centaur Testosterone(TSTO) MCM(Predicate Device)
Intended UseThe ADVIA Centaur Testosterone II(TSTII) Master Curve Material is for invitro diagnostic use in the verification ofcalibration and reportable range of theADVIA Centaur Testosterone II (TSTII)assay.The ADVIA Centaur Testosterone(TSTO) Master Curve Material is forin vitro diagnostic use in theverification of calibration andreportable range of the ADVIACentaur Testosterone assay.
ApproximateTarget Values(Actual NumberVaries by Lot)Level 1: 0.00 ng/dLLevel 2: 50.0 ng/dLLevel 3: 100.0 ng/dLLevel 4: 500.0 ng/dLLevel 5: 750.0 ng/dLLevel 6: 1000.0 ng/dLLevel 7: 1600.0 ng/dLSame
MatrixLyophilized human plasmaSame
Value AssignmentProcessFull curve testingTwo point nested testing

Table 9. Comparison of ADVIA Centaur TSTII Master Curve Material to Predicate

Table 10. Comparison of ADVIA Centaur SHBG Assay to Predicate

ltemADVIA Centaur SHBG Assay:WHO 08/266 200 IS(Candidate Device)ADVIA Centaur SHBG Assay:WHO 95/560 1st IS(Predicate Device)
Intended UseThe ADVIA Centaur SHBG assay is anin vitro diagnostic immunoassay for thequantitative determination of sexhormone-binding qlobulin (SHBG) inhuman serum and plasma using theADVIA Centaur XP system.Same
Indicationsfor UseThe ADVIA Centaur SHBG assay isintended for use as an aid in thediagnosis of androgen disorders.Same
MethodologyChemiluminescenceSame
Assay ProtocolSandwich immunoassaySame
Traceability/StandardizationStandardized to WHO 200 International Traceable to WHO 1* InternationalStandard (08/266)Standard (95/560)
Specimen TypeHuman serum and plasmaSame
Sample Volume10 µLSame
Measuring Range1.60–180 nmol/LSame
Calibration2-point calibrationSame

{10}------------------------------------------------

ItemADVIA Centaur SHBG Calibrator:WHO 08/266 2nd IS(Candidate Device)ADVIA Centaur SHBG Calibrator:WHO 95/560 1st IS(Predicate Device)
Intended UseFor in vitro diagnostic use in calibratingthe ADVIA Centaur SHBG assay on theADVIA Centaur systems.Same
Target ValuesLow Calibrator: 10.0 nmol/LHigh Calibrator: 140.0 nmol/LSame
MatrixEquine serumSame
TraceabilityWHO 2nd International StandardWHO 1st International Standard
Value AssignmentProcessFull curve testingTwo point nested testing

Table 11. Comparison of ADVIA Centaur SHBG Calibrator to Predicate

Table 12. Comparison of ADVIA Centaur SHBG Master Curve Material to Predicate
---------------------------------------------------------------------------------------
ItemADVIA Centaur SHBG MCM:WHO 08/266 2nd IS(Candidate Device)ADVIA Centaur SHBG MCM:WHO 95/560 1st IS(Predicate Device)
Intended UseThe ADVIA Centaur SHBG MasterCurve Material is for in vitro diagnosticuse in the verification of calibration andreportable range of the ADVIA CentaurSHBG assay.Same
ApproximateTarget Values(Actual NumberVaries by Lot)Level 1: 0.1 nmol/LLevel 2: 10.0 nmol/LLevel 3: 50.0 nmol/LLevel 4: 100.0 nmol/LLevel 5: 200.0 nmol/LSame
MatrixEquine serumSame
TraceabilityWHO 2nd International StandardWHO 1st International Standard
Value AssignmentProcessTwo point nested testingSame

Standard/Guidance Document References 9.

The following recognized standards from Clinical Laboratory Standards Institute (CLSI) were used as a basis of the study procedures described in this submission:

  • Evaluation of Precision Performance of Quantitative Measurement Methods; l Approved Guideline - Second Edition (CLSI EP05-A2, 2004; Recognition No. 7-110)
  • 트 Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline (CLSI EP06-A, 2003; Recognition No. 7-193)
  • Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition (CLSI EP07-A2, 2005; Recognition No. 7-127)
  • Measurement Procedure Comparison And Bias Estimation Using Patient Samples --Third Edition (CLSI EP9-A3, 2013; Recognition Number 7-245)
  • Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; l Approved Guideline -- Second Edition (CLSI EP17-A2, 2013: Recognition No. 7-233)

{11}------------------------------------------------

  • Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition (CLSI EP28-A3c - formerly C28-A3c, 2010; Recognition No. 7-224)
  • Medical devices - Application of risk management to medical devices (ANSI/AAMI/ISO 14971:2007/(R)2010; Recognition No. 5-70)

10. Performance Characteristics: ADVIA Centaur TSTII

10.1 Precision

A 20-day precision study was performed according to CLSI EP5-A2. Samples included 5 serum patient pools, three levels of controls and one female patient serum specimen. Samples were assayed twice a day in replicates of 2, for 20 days (n = 80 replicates per sample). The following results were obtained:

SampleMean(ng/mL)RepeatabilityWithin-Lab
SD(ng/mL)%CVSD(ng/mL)%CV
Female Patient Serum Specimen10.580.908.51.3312.6
Serum Patient Pool 126.451.415.32.077.8
Serum Patient Pool 281.782.493.04.295.3
Serum Patient Pool 3311.9712.374.016.545.3
Serum Patient Pool 4768.7423.813.141.715.4
Serum Patient Pool 51127.0754.154.977.676.9
Control 1226.347.113.110.194.5
Control 2578.1919.283.334.536.0
Control 3735.7224.613.344.436.0

10.2 Linearity

A linearity study was performed using the modified device according to CLSI EP06-A using 11 serially diluted samples spanning the assay range. The mean was taken from each sample tested in duplicate. As presented below, the bias from the linear fit estimate was <10%.

SampleExpectedDose(ng/dL)ObservedDose(ng/dL)WeightedLinear FitEstimateDeviationfromLinear Fit(ng/dL)DeviationfromLinear Fit(%)
A1653.771653.771521.25132.528.71
B1242.011155.851142.4313.431.18
C830.24834.69763.6171.099.31
D418.48402.81384.7818.034.68
E212.60198.74195.373.361.72
F109.66103.94100.673.273.25
G58.1949.4853.32-3.84-7.19
H32.4627.0429.64-2.59-8.75
I19.5915.8017.80-2.00-11.25
J13.1510.5311.88-1.35-11.35
K6.726.725.960.7612.71

{12}------------------------------------------------

The weighted linear regression equation is presented below.

Observed = 0.92(Expected) - 0.22 na/dL

10.3 Dilution Recovery

Five serum samples in the range of 1520.24–1715.58 ng/dL of testosterone were diluted 1:2 with ADVIA Centaur Multi-Diluent 3, and assayed for recovery. The recoveries ranged from 93.9 to 114.2% with a mean of 106.7%.

SampleDilution FactorObserved Dose(ng/dL)Expected Dose(ng/dL)% Recovery
1Neat1581.781684.0693.9%
1:2790.89842.03
2Neat1757.541636.03107.4%
1:2878.77818.01
3Neat1730.521520.25113.8%
1:2865.26760.12
4Neat1825.771598.79114.2%
1:2912.89799.39
5Neat1782.641715.58103.9%
1:2891.32857.79
Mean106.7%

Method Comparison 10.4

A method comparison study was performed by comparing the ADVIA Centaur TSTII assay to the CDC HoSt Testosterone ID-LC-MS/MS method with 128 serum samples distributed over the assay range (8.61 to 1394.00 ng/dL). The analysis was performed using Passing-Bablok regression. The regression equation from the analysis is presented below.

ADVIA Centaur TSTII = 0.97(ID-LC-MS/MS) + 1.94 ng/dL (r = 0.98)

An additional method comparison study was performed against the Dimension Vista Total Testosterone assay (K151529) with 117 serum samples from adults and 61 samples from pediatrics) ranging from 8.00 ng/dL to 825.00 ng/dL. An analysis was also performed with the 61 samples from pediatric subjects (27 females age 7 to 18 and 34 males age 9 to 18) ranging from 8.00 to 732.00 ng/dL. The analyses were performed using Weighted Deming regression. The regression equations from the analyses are presented below.

Adult and pediatric specimens:

ADVIA Centaur TSTII = 1.01(Dimension Vista TST) + 0.99 ng/dL (r = 0.99)

Pediatric specimens:

ADVIA Centaur TSTII = 1.04/Dimension Vista TST) + 0.22 ng/dL (r = 0.99)

10.5 Matrix Comparison

The ADVIA Centaur TSTII assay was evaluated using different specimen matrices and tube collection types. A specimen collection study was performed using a minimum of 50 matched specimens drawn in different tube types including serum red top, serum separator tube, plasma separator tube, dipotassium EDTA, lithium heparin, and sodium heparin. No significant difference between tube types was observed. The following results were obtained:

{13}------------------------------------------------

Tube TypesSlopeIntercept (ng/dL)r
Serum vs. serum separator tube0.983.760.991
Serum vs. plasma separator tube0.943.560.993
Serum vs. dipotassium EDTA0.97-1.610.994
Serum vs. lithium heparin0.954.500.996
Serum vs. sodium heparin0.942.850.996

10.6 Reference Intervals

Reference intervals for the ADVIA Centaur TSTII assay were established for pediatrics and adults according to CLSI EP28-A3c.

Reference intervals for the ADVIA Centaur SHBG assay for adults according to CLSI EP28-АЗс.

The adult population included samples from apparently healthy normal cycling women, postmenopausal women, and normal males. Based on a central 95% interval, the following reference intervals were established.

Testosterone

GroupnMedian(ng/dL)ReferenceValues(ng/dL)
Males Age <50 (21-49)119388.24123.06 – 813.86
Males Age ≥50 (50-89)137374.7286.98 – 780.10
Females Pre-Menopause (Age 21-60)16722.269.01 - 47.94
Females Post-Menopause (Age 45-89)12715.18< 7.00 - 45.62

SHBG

GroupnMedian(nmol/L)ReferenceValues(nmol/L)
Males Age <50 (21-49)12234.9714.55 - 94.64
Males Age ≥50 (50-89)13752.4921.63 - 113.13
Females Pre-Menopause (Age 21-60)16758.2110.84 - >180.00
Females Post-Menopause (Age 45-89)13461.4423.15 - 159.07

FAI

GroupnMedianReferenceValues
Males Age <50 (21-49)11939.1411.14 - 81.01
Males Age ≥50 (50-89)13727.178.99 - 46.51
Females Pre-Menopause (Age 21-60)1671.280.27 - 7.64
Females Post-Menopause (Age 45-89)1270.870.17 - 4.15

Reference intervals characterized by age and Tanner stage were established for the ADVIA Centaur TSTII assay for a pediatric population in accordance with the CLSI guideline EP28-АЗс.

{14}------------------------------------------------

For groups of 120 or more subjects, the lower and upper reference limits were estimated as the 2.5" and the 97.5" percentiles of the distribution of test results using a non-parametric approach in accordance with the recommendation in CLSI quideline EP28-A3.

For groups of 40 to 119 subjects, the lower and upper reference limits were estimated as the 2.5" and the 97.5" percentiles of the distribution of test results using a robust measure of location and spread, as developed by Horn and Pesce.

For groups of less than 40 subjects, the lower and upper reference limits were estimated as the 5th and the 95th percentiles of the distribution of test results.

Samples were collected prospectively from apparently healthy pediatric subjects (good endocrinological health) using predefined inclusion criteria.

GroupnMedian(ng/dL)ReferenceValues(ng/dL)
Males Age 2-1040< 7.00< 7.00 - 29.44
Males Age 112011.89< 7.00 - 353.00
Males Age 122570.30< 7.00 - 562.42
Males Age 1321120.378.12 - 582.87
Males Age 1431284.9020.35 - 777.38
Males Age 1522334.86127.21 - 849.36
Males Age 16-2144456.26113.19 - 882.10
Males Tanner Stage 136< 7.00< 7.00 - 47.43
Males Tanner Stage 23811.79< 7.00 - 174.45
Males Tanner Stage 344250.4410.54 - 802.75
Males Tanner Stage 444356.1263.89 - 736.20
Males Tanner Stage 542458.2455.69 - 897.40
Females Age 2-1040< 7.00< 7.00 - 117.76
Females Age 11-1513016.03< 7.00 - 38.92
Females Age 16-213625.2015.06 - 42.41
Females Tanner Stage 144< 7.00< 7.00 - 89.56
Females Tanner Stage 24212.81< 7.00 - 38.29
Females Tanner Stage 34216.71< 7.00 - 33.86
Females Tanner Stage 44019.82< 7.00 - 38.72
Females Tanner Stage 54222.7910.83 - 50.08

10.7 Detection Limit

The limit of blank (LoB), limit of detection (LoD), and the limit of quantitation (LoQ) were determined as described in CLSI protocol EP17-A2. The ADVIA Centaur TSTII assay has an LoB of 2.50 ng/dL, an LoD of 5.00 ng/dL, and an LoQ of 7.00 ng/dL.

The LoB is defined as the highest measurement result that is likely to be observed for a blank sample. The LoD is defined as the lowest concentration of testosterone that can be detected with 95% probability. The LoQ is defined as the lowest concentration of testosterone that can be detected at a total CV of 20%.

{15}------------------------------------------------

10.8 Interference

Interference studies were performed according to CLSI EP07-A2. Two sample pools were tested. One sample pool had approximately 30 ng/dL testosterone. The second sample pool had approximately 300 ng/dL testosterone. These sample pools were spiked with potential interferents. Control samples were prepared by spiking sample pools with the appropriate diluent at the same volume as the interfering substance stock. For substances spiked at doses that caused >10% interference, serial measurements were taken and analyzed by linear regression. Results are presented below.

Interferent~30 ng/dL Testosterone~300 ng/dL Testosterone
DoseWithoutInterferent(ng/dL)DoseWithInterferent(ng/dL)%InterferenceDoseWithoutInterferent(ng/dL)DoseWithInterferent(ng/dL)%Interference
Hemoglobin(500 mg/dL)32.1634.025.8240.66239.81-0.4
Triglycerides(1000 mg/dL)31.2331.611.2233.78213.84-8.5
Conjugated Bilirubin(15 mg/dL)20.2121.054.2269.61258.61-4.1
Unconjugated Bilirubin(20 mg/dL)20.8122.819.6258.77261.973.2
Cholesterol(500 mg/dL)21.9123.326.4263.43238.58-9.4
Rheumatoid Factor(200 IU/mL)24.4625.594.6271.95273.980.7

The concentration of biotin that causes 10% interference, as determined by linear regression, is 30 ng/mL. Specimens that contain 40 ng/mL of biotin demonstrate a 14.6% change in results.

Samples with conjugated bilirubin concentrations > 15 mg/dL will cause erroneous results. Specimens that contain 20 mg/dL of conjugated billirubin demonstrate a -14.4% change in results.

Samples with unconjugated bilirubin concentrations > 20 mg/dL will cause erroneous results. Specimens that contain 30 mg/dL of unconjugated bilirubin demonstrate a 20.1% change in results.

10.9 Heterophile Interference

Heterophile interference testing was performed by spiking 19 HAMA-positive serum samples with ~30 and ~300 ng/dL USP testosterone. The Expected Dose was calculated by adding the "spiked" testosterone concentration to the endogenous (neat) testosterone concentration of the sample. The percent recoveries were determined by dividing the Observed Dose by the Expected Dose. The mean recovery for all samples spiked with ~30 ng/dL USP testosterone was 98.69%. The mean recovery for all samples spiked with ~300 ng/dL USP testosterone was 94.55%.

10.10 Cross-Reactivity

Cross-reactivity studies were performed using two sample pools of approximately 0 ng/dL and 300 ng/dL testosterone. These sample pools were spiked with potential cross-reactants.

{16}------------------------------------------------

Control samples were prepared by spiking sample pools with the appropriate diluent at the same volume as the interfering substance stock. Results are presented below.

Potential~0 nq/dL Testosterone~300 ng/dL Testosterone
Cross-ReactingSubstanceDoseWithoutSubstance(ng/dL)DoseWithSubstance(ng/dL)%Cross-ReactivityDoseWithoutSubstance(ng/dL)DoseWithSubstance(ng/dL)%Cross-Reactivity
5-Androstene-3ß, 17ß-diol(10,0000)0.00124.120.124315.34147.87-0.167
Androstenedione(10,000)1.98141.691.397329.36240.73-0.886
Androsterone(100,000)2.184.990.003320.65294.93-0.026
Canrenone(18,000)2.182.250.000320.65332.500.066
Canrenone(50,000)2.182.820.001320.65323.090.005
Canrenone(100,000)2.182.700.001320.65329.040.008
Corticosterone(100,000)2.181.720.000320.65334.390.014
Cortisol(100,000)2.182.680.000320.65342.270.022
Cyproterone(10,000)2.1815.010.128320.65326.090.054
Danazol(100,000)2.18122.240.120320.65199.51-0.121
11-Deoxycortisol(100,000)2.184.110.002320.65322.850.002
Dexamethasone(100,000)2.182.000.000320.65326.690.006
DHEA(100,000)2.187.860.006320.65278.20-0.042
DHEAs(5,000,000)2.1873.880.001320.65278.70-0.001
5a-Dihydrotestosterone(10,000)2.1877.560.754320.65179.68-1.410
17β-Estradiol(100,000)0.0921.120.021355.5288.27-0.267
Estrone(10,000)2.182.410.002320.65319.76-0.009
Ethisterone(100,000)0.00445.990.446271.00678.950.408
113-Hydroxytestosterone(10,000)0.001546.8515.469271.001612.3413.413
11-Keto-testosterone(100,000)0.001765.931.766271.001854.411.583
Nandrolone decanoate(100,000)N/A*N/A*N/A*N/A*N/A*N/A*
Norgestrel(100,000)0.00145.010.145271.00197.62-0.073

{17}------------------------------------------------

Potential~0 ng/dL Testosterone~300 ng/dL Testosterone
Cross-ReactingSubstanceDoseWithoutSubstance(ng/dL)DoseWithSubstance(ng/dL)%Cross-ReactivityDoseWithoutSubstance(ng/dL)DoseWithSubstance(ng/dL)%Cross-Reactivity
Oxymetholone(10,000)2.182.940.008320.65263.08-0.576
Prednisolone(100,000)0.000.000.000271.00364.170.093
Prednisone(100,000)0.000.000.000271.00282.170.011
Progesterone(100,000)2.182.450.000320.65335.040.014
Spironolactone(8,000)2.182.590.005320.65329.060.105
Spironolactone(100,000)2.182.090.000320.65331.680.011
Testosterone propionate(10,000)0.003.540.035271.00565.562.946
7α-Thiomethyl spironolactone(40,000)2.182.450.001320.65320.960.001
7α-Thiomethyl spironolactone(100,000)2.182.400.000320.65318.27-0.002
  • Crossreactivity for this compound could not be calculated. This compound tested above the measuring interval.

Traceability and Value Assignment 10.11

The ADVIA Centaur TSTII assay is standardized using internal standards made from USPgrade testosterone, which are traceable to isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS). The ID-LC-MS/MS is traceable to the primary testosterone standard National Measurement Institute (NMI) M914.

The calibrator values used for two-point calibration of the ADVIA Centaur TSTII assay are value assigned using standards traceable to the ID-LC-MS/MS method.

The ADVIA Centaur TSTII Master Curve Material assigned values are lot specific of target values.

10.12 Stability

The ADVIA Centaur TSTII Reagents, Calibrators and Master Curve Materials are stable until the date printed on the box label when stored at 2-8°C.

The onboard stability of the ADVIA Centaur TSTII Reagents is 18 days with a calibration interval of 28 days.

{18}------------------------------------------------

Performance Characteristics: ADVIA Centaur SHBG 11.

11.1 Precision

A 5-day precision study was performed using the modified device to verify existing claims. Samples included 3 patient serum specimens, 2 patient pools and three levels of controls. Samples were assayed once a day in replicates of 5, for 5 days (n = 25 replicates per sample). The following results were obtained:

SampleMean(nmol/L)RepeatabilityWithin-Lab
SD(nmol/L)%CVSD(nmol/L)%CV
Patient specimen 170.711.582.21.892.7
Patient specimen 2115.953.052.63.843.3
Patient specimen 3145.594.393.06.444.4
Patient pool 110.690.211.90.262.4
Patient pool 239.890.721.80.822.1
Control 122.490.401.80.431.9
Control 244.950.962.10.992.2
Control 339.540.932.31.002.5

11.2 Linearity

A linearity study was performed using the modified device according to CLSI EP06-A using 11 serially diluted samples spanning the assay range. The mean was taken from each sample tested in duplicate. As presented below, the bias from the linear fit estimate was ≤15%.

SampleExpectedDose(nmol/L)ObservedDose(nmol/L)WeightedLinear FitEstimateDeviationfromLinear Fit(nmol/L)DeviationfromLinear Fit(%)
A198.79195.95195.830.120.06
B183.50183.50180.792.711.50
C159.03161.07156.724.362.78
D119.27117.44117.60-0.16-0.13
E59.6459.4358.920.510.87
F29.8229.2829.58-0.31-1.04
G14.9114.7714.91-0.15-0.99
H7.457.517.58-0.07-0.97
I3.733.923.910.010.30
J1.862.042.08-0.04-2.03
K0.931.161.160.00-0.10

The weighted linear regression equation is presented below.

Observed = 0.984(Expected) + 0.245 ng/dL

{19}------------------------------------------------

11.3 Dilution Recovery

Five serum samples in the range of 161.14-175.13 nmol/L of SHBG were diluted 1:2 with ADVIA Centaur Multi-Diluent 1, and assayed for recovery. The recoveries ranged from 87 to 101% with a mean of 92%.

SampleOnboard Dilution Dose(nmol/L)Manual Dilution Dose(nmol/L)% Recovery
1175.13173.59101%
2164.64188.3787%
3161.14172.7793%
4162.89181.7890%
5164.34179.991%
Mean92%

11.4 Method Comparison

A method comparison study was performed by comparing the ADVIA Centaur SHBG assay to the Elecsys SHBG assay with 174 serum samples distributed over the assay range. The analysis was performed using Weighted Deming regression equation from the analysis is presented below.

ADVIA Centaur SHBG = 0.99(Elecsys) - 0.11 nmol/L (r = 0.99)

11.5 Matrix Comparison

The re-standardization of the ADVIA Centaur SHBG assay did not influence sample matrices.

11.6 Reference Intervals

See Section 10.6 (Reference Intervals) for ADVIA Centaur TSTII assay.

11.7 Detection Limit

The limit of blank (LoB). limit of detection (LoD). and the limit of quantitation (LoQ) were determined as described in CLSI protocol EP17-A2. The ADVIA Centaur SHBG assay has an LoB of 1.2 nmol/L, an LoD of 1.6 nmol/L, and an LoQ of 1.8 nmol/L.

The LoB is defined as the highest measurement result that is likely to be observed for a blank sample. The LoD is defined as the lowest concentration of SHBG that can be detected with 95% probability. The LoQ is defined as the lowest concentration of SHBG that can be detected at a total CV of 20%.

11.8 Interference

The re-standardization of the ADVIA Centaur SHBG assay did not influence interference. This information is described in K091867.

11.9 Cross-Reactivity

The re-standardization of the ADVIA Centaur SHBG assay did not influence cross-reactivity. This information is described in K091867.

11.10 Traceability and Value Assignment

The ADVIA Centaur SHBG assay is standardized to World Health Organization (WHO) 200 International Standard for SHBG, NIBSC code 08/266.

{20}------------------------------------------------

The calibrator values used for two-point calibration of the ADVIA Centaur SHBG assay are value assigned using standards prepared from SHBG standardized against gravimetrically prepared dilutions of the 200 International Reference Preparation (WHO 08/266) from NIBSC.

The ADVIA Centaur SHBG Master Curve Material assigned values are lot specific of target values.

11.11 Stability

The re-standardization of the ADVIA Centaur SHBG assay did not influence stability. This information is described in K091867.

12. Conclusions

The new ADVIA Centaur Testosterone II (TSTII) assay (Reagents and Calibrators) is substantially equivalent in principle and performance to the currently-marketed predicate device, the Elecsys Testosterone II assay, cleared under 510(k) K093421.

The ADVIA Centaur TSTII Master Curve Material is substantially equivalent in principle and performance to the currently-marketed predicate device, the ADVIA Centaur Testosterone (TSTO) Master Curve Material, cleared under 510(k) K140505.

The modified ADVIA Centaur SHBG assay (Reagents, Calibrators, Master Curve Material) is substantially equivalent in principle and performance to the currently-marketed predicate device, the ADVIA Centaur SHBG assay, cleared under 510(k) K091867.

§ 862.1680 Testosterone test system.

(a)
Identification. A testosterone test system is a device intended to measure testosterone (a male sex hormone) in serum, plasma, and urine. Measurement of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.(b)
Classification. Class I.