Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K201570

    Validate with FDA (Live)

    Date Cleared
    2020-09-11

    (92 days)

    Product Code
    Regulation Number
    864.7695
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PF4 Enhanced assay is designed as a solid phase enzyme linked immunosorbent assay (ELISA). The product is intended to be used as an in vitro diagnostic kit by hematology, coagulation, or other pathology laboratories to assist in screening patient samples, 3.2% sodium citrate (plasma) or without anticoagulant (serum), for the presence of heparin associated commonly found in patients with heparin induced thrombocytopenia (HIT) or thrombosis.

    Device Description

    PF4 Enhanced Assay is an Enzyme Linked Immunosorbent Assay (ELISA). The PF4 Enhanced ELISA is intended to detect antibodies in human serum or plasma that react with Platelet Factor 4 (PF4) when it is complexed to heparin or other polyanionic compounds. The PF4 Enhanced kit contains all of the reagents necessary to perform the assay.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and supporting studies for the Immucor PF4 Enhanced assay, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/MetricAcceptance CriteriaReported Device Performance
    Positive Serum ControlAverage OD value ≥ 1.800 OD for each vial tested. Duplicate test well OD values ≤ 20% of the mean of the two values.All acceptance criteria were met for three validation batches, consistently delivering product meeting established acceptance criteria.
    Comparison of Serum & PlasmaAcceptable percent agreement between qualitative results using serum or plasma. Positive agreement in OD values with a slope > 0.9 (up to 4.0 OD).94.8% agreement (164/173) observed initially, increasing to 95.4% (164/172) after investigating and removing one incorrect fresh serum result. Comparison of OD values showed positive agreement with a slope > 0.9.
    Stopping Solution Lot-to-LotNew Stopping Solution (ESS) performs equivalently to the predicate Stopping Solution (3M NaOH).All acceptance criteria were met, demonstrating the new ESS is acceptable.
    Stopping Solution Process ValidationAll well-to-well variations meet acceptance criteria. All finished product QC panel samples (21 positive, 24 negative) yield expected qualitative results with the new ESS. All assays meet IFU QC requirements.All wells met acceptance criteria for well-to-well variation. All 21 positive and 24 negative samples met their expected qualitative result across three lots of new ESS and one lot of 3M NaOH. All assays met the QC requirements in the IFU.
    Shelf LifeStability data supports a 24-month shelf life.All acceptance criteria for stability studies were met, supporting a 24-month shelf life.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Positive Serum Control Raw Material Change: 30 vials tested (total number of samples from 3 validation batches is not explicitly stated beyond this).
    • Comparison of Serum and Plasma Sample Types: 174 paired (serum vs. plasma) fresh samples.
    • Data Provenance for Serum/Plasma Comparison: Florida Hospital conducted the study. The data was retrospective as Genetic Testing Institute (the prior company) used the transcribed raw data for analysis. The country of origin is implicitly the U.S.
    • Stopping Solution Lot-to-Lot Comparison: 3 lots of new Stopping Solution (ESS) and 3M NaOH (predicate). Number of actual test samples per lot not specified beyond implied sufficient testing to meet acceptance criteria.
    • Stopping Solution Process Validation: 3 lots of new Stopping Solution (ESS) and 1 lot of 3M NaOH (predicate). Tested with a finished product QC panel of 21 positive and 24 negative samples (total 45 samples per lot/solution type).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the number of experts or their qualifications for establishing ground truth. For the Comparison of Serum and Plasma Sample Types, it mentions "Florida Hospital conducted a study" and "Genetic Testing Institute used the transcribed raw data to perform the analysis." This suggests clinical outcomes or internal lab testing might have served as ground truth, rather than expert consensus on individual cases. For other studies, "established acceptance criteria" and "expected qualitative result" imply pre-defined truths based on known characteristics of control materials or validated samples.

    4. Adjudication Method for the Test Set

    The document does not mention an explicit adjudication method (e.g., 2+1, 3+1). For the Comparison of Serum and Plasma Sample Types, it notes an "investigation of the ninth discrepant sample," where "additional testing for the plasma and serum sample both matched that of the fresh plasma sample," suggesting a form of re-evaluation or retesting for discrepant results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. The device is an in vitro diagnostic (ELISA) kit.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Yes, the studies presented are all standalone performance evaluations of the assay components and the assay as a whole, without human-in-the-loop performance being a specific metric. The device itself is an assay kit.

    7. The Type of Ground Truth Used

    • Positive Serum Control Raw Material Change: Ground truth was based on pre-defined expected optical density (OD) values and variation limits for the control material.
    • Comparison of Serum and Plasma Sample Types: Ground truth for sample positivity/negativity would have been established by the clinical laboratory (Florida Hospital) using the predicate device or a clinical diagnosis, which the new method was compared against. It's implied this was based on the "qualitative results obtained."
    • Stopping Solution Lot-to-Lot Comparison: Ground truth was based on the performance of the predicate Stopping Solution (3M NaOH) and established qualitative results for known samples.
    • Stopping Solution Process Validation: Ground truth was based on the expected positive/negative qualitative results of a "finished product QC panel" and the performance of the predicate 3M NaOH Stopping Solution.

    8. The Sample Size for the Training Set

    The document does not specify a training set for algorithm development, as the device is an ELISA kit and not an AI/ML algorithm that typically requires a separate training set. The studies described are validation and performance testing of the kit's components and overall function.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no mention of an algorithm training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201311

    Validate with FDA (Live)

    Device Name
    PF4 IgG assay
    Date Cleared
    2020-06-18

    (31 days)

    Product Code
    Regulation Number
    864.7695
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PF4 IgG assay is a qualitative screening assay for the detection of heparin associated IgG antibodies in human serum or plasma. The presence of heparin-associated antibodies are commonly found in patients with Heparin Induced Thrombocytopenia (HIT).

    Device Description

    PF4 IgG assay is an enzyme linked immunosorbent assay (ELISA). The PF4 IgG ELISA is intended to detect IgG antibodies in human serum or plasma that react with Platelet Factor 4 (PF4) when it is complexed to heparin or other polyanionic compounds. The PF4 IgG kit contains all of the reagents necessary to perform the assay.

    AI/ML Overview

    The Immucor GTI Diagnostics, Inc. PF4 IgG assay is a qualitative screening assay for the detection of heparin-associated IgG antibodies in human serum or plasma, commonly found in patients with Heparin Induced Thrombocytopenia (HIT). The device is a Class II medical device, product code LCO.

    This 510(k) submission is for modifications to a previously cleared device (K071781) and includes: a change in the Positive Serum Control raw material, the addition of plasma as a sample type, and a change in the Stopping Solution.

    1. Table of acceptance criteria and reported device performance:

    Feature/CharacteristicAcceptance Criteria (Predicate)Reported Device Performance (Candidate)Comments
    Modification: Positive Serum Control Raw Material Change
    Process ValidationProduct meets established acceptance criteria.All acceptance criteria were met. Product consistently delivers product meeting established acceptance criteria.Validation batches of PF4 Positive Serum Control manufactured and tested.
    Precision Study100% agreement within run and between run. All pre-defined Acceptance Criteria met.100% positive, demonstrating 100% agreement within run and between run. All pre-defined Acceptance Criteria met.Tested by 3 operators, 5 assays each, using a single PF4 IgG kit lot and 3 validation lots of Positive Serum Control.
    Modification: Addition of Plasma as a Sample Type
    Agreement between Serum and PlasmaNot explicitly stated for predicate, but implies equivalent results for both sample types.98.85% Agreement (172/174) between qualitative results of serum and plasma.Internal study.
    Correlation between OD values (Plasma vs. Serum)Not explicitly stated for predicate, but implies strong correlation.R² = 99.3%. Slope = 0.9476, Intercept = -0.009914 (very good agreement).Linear regression analysis of OD values.
    Equivalence of ACD and Sodium Citrate Plasma to SerumEquivalent results.All samples tested gave equivalent results to serum samples.Internal study.
    Modification: Stopping Solution Change
    Lot to Lot Comparison (New vs. Predicate Stopping Solution)New Stopping Solution performs equivalently, meeting established acceptance criteria.All acceptance criteria were met, demonstrating equivalence.Three lots of new Stopping Solution (ESS) compared to predicate (3M NaOH).
    Process Validation (New Stopping Solution)Well-to-well variation meets acceptance criteria. Qualitative results meet expected results for QC panel.All wells met acceptance criteria for well-to-well variation. All 21 positive and 24 negative samples met expected qualitative results on all three lots of new Stopping Solution and one lot of 3M NaOH.Three lots of new Stopping Solution (ESS) tested with PF4 IgG assay.
    Shelf LifeNot explicitly detailed but assumed to be acceptable.24 months. All acceptance criteria were met for stability studies.Supported by stability studies.

    2. Sample size used for the test set and the data provenance:

    • Positive Serum Control Raw Material Change (Precision Study):
      • No specific "test set" sample size for patient samples is provided. The study focused on the performance of the new positive serum control material within the assay.
      • The study used a single lot of the PF4 IgG kit and three validation lots of the Positive Serum Control.
      • Data provenance is internal, generated as part of process validation.
    • Addition of Plasma as a Sample Type:
      • Test set size: 174 frozen serum and plasma samples.
      • Data provenance: Samples obtained from Florida Hospital for an internal study. Retrospective, as samples were frozen.
      • Sub-study for ACD/Sodium Citrate vs. Serum: Blood collected from 24 individuals. Origin not specified beyond "internal study." Retrospective.
    • Stopping Solution Change:
      • Test set size: No patient samples explicitly mentioned in the summary for the Stopping Solution evaluation itself. The testing involved comparing the performance of the new Stopping Solution with the predicate using various assay components and a QC panel.
      • QC Panel: 21 positive samples and 24 negative samples.
      • Data provenance: Internal, generated as part of lot-to-lot comparison and process validation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    The document does not provide information on the number or qualifications of experts used to establish ground truth for the test sets. For the plasma sample type study, it states 174 frozen serum and plasma samples were used, and for the ACD/Sodium Citrate comparison, blood was collected from 24 individuals "known to be PF4: heparin antibody negative." How this "known" status was established (e.g., by experts, by a predicate device, by clinical diagnosis) is not detailed. The ground truth for the QC panel used in the stopping solution study is based on "expected qualitative result," implying pre-characterized samples, not expert consensus at the time of the study.

    4. Adjudication method for the test set:

    The document does not describe any adjudication method for establishing ground truth for the test sets. The studies appear to rely on the inherent reactivity of the samples (e.g., "known to be PF4: heparin antibody negative," "high level reacting anti:PF4 heparin antibody sample," "expected qualitative result" for QC panels).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This device is an in vitro diagnostic (IVD) assay, specifically an ELISA. It is not an AI-assisted diagnostic tool that involves human readers interpreting images or data. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The device is an IVD assay, not an algorithm. Its performance is evaluated biochemically based on optical density measurements, which are then interpreted qualitatively (positive/negative) according to the assay's cut-offs. The studies described are standalone performance evaluations of the modified assay components and sample compatibility.

    7. The type of ground truth used:

    • For the plasma sample type comparison, the ground truth appears to be based on the qualitative results obtained from the serum sample, as the goal was to demonstrate equivalence between serum and plasma samples. For the sub-study on ACD/Sodium Citrate, it refers to individuals "known to be PF4: heparin antibody negative" and "high level reacting anti:PF4 heparin antibody sample," indicating pre-characterized samples.
    • For the stopping solution change, the ground truth for the QC panel was based on "expected qualitative result," implying pre-defined and characterized positive and negative samples.
    • The primary ground truth for the assay itself (predicate and candidate) would typically be established against either clinical diagnosis of HIT (outcome data) or a well-established reference method, but this is not explicitly detailed for these specific modification studies. The studies focus on demonstrating that the modifications do not adversely affect the device's performance compared to its previous state.

    8. The sample size for the training set:

    The document does not describe a training set as this is not an algorithm requiring machine learning. The studies are performance and validation studies for changes to an IVD assay.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1