Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K182287

    Validate with FDA (Live)

    Device Name
    MEDINAUT Plus
    Manufacturer
    Date Cleared
    2019-04-06

    (226 days)

    Product Code
    Regulation Number
    888.1100
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MEDINAUT Plus Kyphoplasty System is intended to be used as a conventional bone tamp for the reduction of fractures and/or creation of a void in cancellous bone in the spine (including use during balloon kyphoplasty with a PMMA-based bone cement that is cleared for use in kyphoplasty procedures), hand, tibia, radius, and calcaneus.

    Device Description

    The MEDINAUT Plus Kyphoplasty System is designed to reduce compression fracture and create a void in cancellous bone in the spine. By creating a space in the operating point, the major benefits of MEDINAUT Plus Kyphoplasty System are the reduction in pain and the increase of patient s functional abilities, which allow for the patient's return to the previous level of activity. The MEDINAUT Plus Kyphoplasty System consists of the Balloon Expander, Balloon Catheter, Cement Dispenser System, Cement Mixer System and Syringe. The Balloon Catheter consists of an inner-outer tube, Y-port and an inflatable balloon located at the distal tip. The radiopaque markers located at the distal and proximal end allow fluoroscopic visualization of the deflated balloon catheter during positioning. The Balloon Expander consists of a gauge, housing, a half nut, a plunger, a piston, a hose, a luer connector and a 3 way valve. The Balloon Expander is used for inflating the balloon by rotating the plunger clockwise. The lock mechanism maintains pressure. The Cement Dispenser System consists of Cannula, Needle (3 types: Bevel, Diamond, Trocar), Spacer, Biopsy, Biopsy Cap. Cement Filler, Cement Pusher and Cement Side Filler. The Cement Mixer System consists of Cement Mixer, Tube, Cap, Funnel and Pusher. The user can mix cement inside the cylinder by moving the mixer up and down repeatedly and apply the cement into a patient by rotating the pusher handle. The Syringe is an optional device. If there is air found in the Catheter Balloon before surgery, the user can remove the air with this syringe. All the components including Balloon Expander, Balloon Catheter, Cement Dispenser System. Cement Mixer System and Syringe are supplied sterile and are disposable.

    AI/ML Overview

    The provided text, K182287, outlines the 510(k) summary for the MEDINAUT Plus Kyphoplasty System. However, it does not describe acceptance criteria and a study proving the device meets those criteria in the context of an AI/algorithm-driven device performance evaluation (such as sensitivity, specificity, accuracy, or human reader improvement).

    Instead, this document is a regulatory submission for a physical medical device (an inflatable bone tamp for kyphoplasty) and focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing (bench tests, biocompatibility, sterilization, shelf life).

    Therefore, I cannot extract the requested information (points 1-9) as it pertains to AI/algorithm performance studies typically found in submissions for AI-enabled software as a medical device (SaMD). The concepts of "acceptance criteria for algorithm performance," "sample size for test set," "number of experts," "adjudication method," "MRMC study," "standalone performance," "ground truth," and "training set" are not applicable to the non-clinical bench testing and physical device comparisons described in this 510(k) summary.

    The document's "Non-Clinical Testing" section describes physical and material property tests:

    • Sterilization validation: Ensuring the device can be sterilized.
    • Shelf life validation: Ensuring the device maintains its properties over time.
    • Biocompatibility testing: Ensuring the materials are safe for use in the body.
    • Bench tests:
      • Visual Inspection
      • Size Measurement
      • Liquid leakage under pressure
      • Tensile force
      • Burst strength
      • Balloon fatigue
      • Inflated balloon size
      • Balloon deflation
      • EO Gas residue
      • Sterility Test
      • Balloon rated burst pressure (RBP) in constrained condition
      • Balloon rated burst pressure (RBP) in unconstrained condition
      • Freedom from leakage and damage on inflation
      • Balloon diameter to inflation pressure
      • Insertion force and withdrawal force

    These tests are to confirm the physical and mechanical integrity and safety of the device, not the performance of an AI algorithm. The conclusion states that the test results supported "substantial equivalence to the predicate devices" based on these physical characteristics and performance.

    In summary, none of the requested information regarding AI/algorithm performance studies can be provided from this document because it describes a physical medical device, not an AI/algorithm-driven one.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173541

    Validate with FDA (Live)

    Device Name
    EPINAUT
    Manufacturer
    Date Cleared
    2018-08-24

    (282 days)

    Product Code
    Regulation Number
    868.5120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For administration of anesthetic agents into the epidural space to provide epidural or caudal anesthesia.

    Device Description

    EPINAUT Spring Guide Epidural Catheter is constructed of a stainless steel continuous spring. EPINUAT Spring Guide Epidural Catheter offers two types of catheter. The catheter types, L330 & L600 have an uncoated distal tip which is flexible, smooth and rounded with coils slightly spread for maximum flexibility and lateral distribution of injectant. The other catheter types, C310 & C600 have a distal end which is entirely coated by tube for maximum smoothness. There are 16 models in EPINAUT: L33011320, L33013820, L33011315, L33013815, L60011320, L60013820, L60011315, L60013815, C31011320, C31013820, C31011315, C31013815, C60011320, C60013820, C60011315 and C60013815. Each model is characterized by different length of catheter, length and outer diameter of wire, inner diameter of adapter, length and diameter of introducer needle and needle cap.

    EPINUAT Spring Guided Epidural Catheter have enhanced tensile/ break strength and restrict longitudinal catheter stretch, while the spring coils make the catheter resistant to kinking and collapsing. Wire is also provided with the device which consists of a stainless steel and a molded plastic hub.

    The introducer needle is inserted into the epidural space through the sacral hiatus under intermittent fluoroscopy. After appropriate determination of the epidurogram and target area, the catheter with the wire in it is advanced through the introducer needle to the concerned area as determined by MRI or symptomatology. When the catheter is approached to target site, the wire is removed from the catheter and then the medication can be injected by attaching the adapter to a syringe.

    AI/ML Overview

    The medical device in question, EPINAUT, is an Anesthesia Conduction Catheter. The provided context does not include information about clinical studies involving human patients or AI-based performance metrics. The information focuses on non-clinical testing to establish substantial equivalence to a predicate device.

    Here's an analysis of the provided text in relation to the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes various non-clinical tests performed to evaluate the performance of EPINAUT and to demonstrate substantial equivalence to the predicate device, the Racz Epidural Catheter. The acceptance criteria are implicitly met if the test results "met the pre-set criteria" or "supported that the subject device is substantially equivalent." Specific numerical acceptance criteria are not explicitly stated for all tests; rather, the successful completion of the test according to a standard often serves as the acceptance.

    No.TestAcceptance Criteria (Implicit)Reported Device Performance
    1Visual Inspection (Appearance)Device appears as intended, free from defects.Test performed, results met pre-set criteria.
    2Size MeasurementDimensions conform to specifications.Test performed, results met pre-set criteria.
    3Leakage Test (Catheter)No leakage.Test performed, results met pre-set criteria.
    4Tensile strength (Catheter)Meets specified tensile strength (e.g., ISO 10555-1 Annex B for catheter, BS 6196:1989 Appendix D for catheter-hub/Stylet/Needle Stylet).Test performed, results met pre-set criteria. (Also see side-by-side testing results below).
    5Assemble compatibility (Catheter)Components fit together correctly.Test performed, results met pre-set criteria.
    6Radiopacity testDevice is visible under fluoroscopy.Test performed, results met pre-set criteria.
    7Corrosion resistance testNo significant corrosion.Test performed, results met pre-set criteria.
    8Flexural Rigidity (Catheter introducer and needle)Meets specified flexibility requirements.Test performed, results met pre-set criteria.
    9Pulling (Drawing) Test (Catheter introducer and needle)Meets specified pulling force requirements.Test performed, results met pre-set criteria.
    10Flexural Rigidity (Wire)Meets specified flexibility requirements.Test performed, results met pre-set criteria.
    11EO Gas residueResidue levels below accepted limits.Test performed, results met pre-set criteria.
    12Sterility TestSterility Assurance Level (SAL) of 10^-6 (in accordance with ISO 11135 and ISO 11737-1, 2).Test performed, results met pre-set criteria.
    13Endotoxin TestEndotoxin levels below specified limits.Test performed, results met pre-set criteria.
    14Particulate Matter TestParticulate matter levels below specified limits.Test performed, results met pre-set criteria.
    Side-by-Side Comparative Tests with Predicate Device:
    1Tensile strength (Catheter, Catheter-Hub, Stylet, Needle Stylet)Performance comparable to predicate device, meeting relevant standards (ISO 10555-1 Annex B, BS 6196:1989 Appendix D for various components, BS 6196:1989 Appendix F for Needle Stylet)."The test results supported that the subject device is substantially equivalent to the predicate devices." Implies comparable or better performance.
    2Dullness testPerformance comparable to predicate device, meeting BS 6196:1989 Appendix G."The test results supported that the subject device is substantially equivalent to the predicate devices." Implies comparable or better performance.
    3Fluid pressure of the catheter/adapter connectionPerformance comparable to predicate device, meeting ISO 10555-1 Annex G."The test results supported that the subject device is substantially equivalent to the predicate devices." Implies comparable or better performance.
    4Flow ratePerformance comparable to predicate device, meeting ISO 10555-1 Annex E."The test results supported that the subject device is substantially equivalent to the predicate devices." Implies comparable or better performance.
    5Luminal integrityPerformance comparable to predicate device, meeting ISO 10555-1 Annex C."The test results supported that the subject device is substantially equivalent to the predicate devices." Implies comparable or better performance.
    Biocompatibility:Biocompatible as per ISO 10993-4, 5, 10, 11 and USP 39 <151> for External communicating device (tissue/bone/dentin communicating), Limited exposure (<24 hours)."Biocompatibility tests were performed... and the test results supported that the subject devices are biocompatible."
    Shelf Life:Validated 3-year shelf life using accelerated aging (ASTM F1980)."The test results validated 3 year shelf life."

    2. Sample size used for the test set and the data provenance:

    The document describes non-clinical bench testing. It does not refer to "test sets" in the context of clinical data, human subjects, or AI performance evaluation. Therefore, information on sample sizes for a clinical test set or data provenance (country of origin, retrospective/prospective) is not applicable or provided. The tests mentioned adhere to international standards (e.g., ISO, ASTM, BS) which typically specify the number of samples required for each test.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable. The device is an Anesthesia Conduction Catheter, and its performance evaluation relies on engineering, material, and biocompatibility testing according to established standards, not on expert-adjudicated ground truth from clinical image interpretation or similar AI applications.

    4. Adjudication method for the test set:

    Not applicable, as there is no "test set" in the context of clinical data requiring expert adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. The EPINAUT device is an anesthesia conduction catheter, a physical medical device, not an AI software. There is no mention of AI assistance or MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used:

    For non-clinical performance tests, the "ground truth" is defined by the requirements of the relevant international standards (e.g., ISO, ASTM, BS) and the pre-set internal specifications of the manufacturer. For biocompatibility, the ground truth is established by the results of tests performed according to ISO 10993 and USP standards, demonstrating the material not to be toxic or irritating. For sterilization, the ground truth is a Sterility Assurance Level (SAL) of 10^-6 demonstrated through validated methods according to ISO 11135 and ISO 11737.

    8. The sample size for the training set:

    Not applicable. This is a physical medical device, not an AI system that requires a "training set."

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153296

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-07-29

    (259 days)

    Product Code
    Regulation Number
    888.3027
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MEDINAUT Kyphoplasty System is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine, tibia, radius, and calcaneus. This includes percutaneous vertebral augmentation. The system is to be used with cleared spinal polymethylmethacrylate (PMMA) bone cements indicated for use during percutaneous vertebral augmentation, such as kyphoplasty.

    Device Description

    The MEDINAUT Kyphoplasty System is designed to reduce compression fracture and create a void in cancellous bone in the spine, tibia, radius, and calcaneus. Bone cement insertion using cement dispensing plunger is applied for spinal use only.

    The MEDINAUT Kyphoplasty System consists of the MEDINAUT-X (Inflatable Bone Expander System) and MEDINAUT-I (Cement Dispenser System).

    The MEDINAUT-X is comprised of a Balloon Catheter and a Bone Expander Syringe. The Balloon Catheter's main components are the shaft, Y-hub and the inflatable balloon located at the distal tip. The inflatable balloon is covered with a maximum 0.03g of silicone fluid to enhance lubricity when the balloon catheter is inserted in the cannula. Radiopaque markers located at the distal and proximal end of deflated working surface allow fluoroscopic visualization of the deflated balloon catheter during positioning. The Balloon Catheter and Bone Expander Syringe are supplied sterile and are disposable.

    The MEDINAUT-I is a cement dispenser system consisting of a Needle Pipe, Needle Pin, Wire Pin, Cannula, Expander, Spacer, Cannula Expander, Cement Pusher, Cement Filler, and Guide Wire. They are supplied sterile and are disposable.

    AI/ML Overview

    This document is a 510(k) summary for the MEDINAUT Kyphoplasty System. It is an FDA submission seeking approval to market a medical device, asserting its substantial equivalence to previously cleared devices. It primarily details physical and functional attributes and does not contain detailed "acceptance criteria" or "study that proves the device meets the acceptance criteria" in the way one might expect for a diagnostic AI device.

    Instead, this document focuses on demonstrating substantial equivalence to predicate devices. This means that the device is shown to be as safe and effective as a legally marketed device (the predicate device) that does not require premarket approval.

    Here's an analysis based on the provided text, addressing your points where possible, and noting where the information is not applicable or unavailable in this type of submission:

    1. A table of acceptance criteria and the reported device performance

    This information is not present in the document as it would be for a diagnostic AI device. For a device like the MEDINAUT Kyphoplasty System, "acceptance criteria" are related to mechanical performance, biocompatibility, and sterilization standards, and "reported device performance" refers to the results of non-clinical bench tests comparing these attributes to pre-set criteria or predicate devices.

    The closest equivalent is the "Non-Clinical Testing" section and the comparison table.

    Non-Clinical Testing Summary:

    Test TypeAcceptance Criteria (Implied)Reported Device Performance
    Sterilization ValidationIn accordance with ISO 11137-1, 2, 3 and ISO 11737-1, 2, 3Test results met pre-set criteria.
    Shelf LifeValidated shelf lifeTest results validated 1-year shelf life.
    BiocompatibilityIn accordance with ISO 10993-4, 5, 10, 11 (Meets ISO 10993)Test results supported that devices are biocompatible.
    Bench Tests (Mechanical/Performance)Met pre-set criteria for:Test results met pre-set criteria for all listed tests.
      - Balloon deflation time
      - Burst pressure constrained
      - Burst pressure unconstrained
      - Fatigue strength
      - Balloon dimension (before/after)
      - Insertion and withdrawal force
      - Tensile bond strength

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable/provided in this context. This document describes a medical device (kyphoplasty system), not an AI/software device that processes data. The "test set" here refers to physical samples of the device undergoing mechanical, chemical, and biological testing, not a dataset of patient information. No patient data or provenance are mentioned because it's a non-clinical submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/provided. The "ground truth" for this type of device is established through engineering and laboratory standards (e.g., ISO standards for sterilization, biocompatibility, and mechanical properties). There are no "experts" in the sense of clinicians establishing a diagnostic ground truth for patient data. The "experts" involved would be engineers and laboratory technicians performing standardized tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/provided. Adjudication methods like 2+1 or 3+1 are used in clinical studies or evaluations of diagnostic performance where human interpretation of medical data (e.g., images) is involved. For non-clinical bench testing of a physical medical device, test results are typically validated against established quantitative standards and protocols, not through expert adjudication of uncertain cases.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable/provided. MRMC studies are specific to evaluating diagnostic systems, especially AI in medical imaging, and its impact on human reader performance. This document concerns a physical surgical device for bone fracture reduction, not a diagnostic system. There is no AI component mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable/provided. As stated, this is a physical medical device (kyphoplasty system), not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's safety and effectiveness is established through adherence to international standards (e.g., ISO 11137, ISO 11737, ISO 10993) and engineering specifications. For example:

    • Sterilization: Demonstrated bacterial inactivation based on ISO standards.
    • Biocompatibility: Demonstrated material safety based on ISO 10993.
    • Mechanical Performance: Quantitative measurements (e.g., burst pressure, tensile strength) meeting design specifications and being comparable to predicate devices.

    8. The sample size for the training set

    This information is not applicable/provided. This device is not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable/provided. As above, no training set for an AI model is involved.

    Summary of Device and Equivalence Claim:

    The document states that the MEDINAUT Kyphoplasty System (K153296) is substantially equivalent to its primary predicate device (MEDINAUT Kyphoplasty System, K133669) because it is identical in raw materials, design, and manufacturing processes. The purpose of the current 510(k) is to add more anatomical sites (tibia, radius, and calcaneus) to the indications for use. This expanded indication is supported by referencing an additional predicate device, the KyphX Inflatable Bone Tamp (K010246, K041454, K981251), which already has these broader claims and shares similar design, function, size, shape, materials, and physical specifications.

    The "study that proves the device meets the acceptance criteria" (in this context, substantial equivalence to predicates for the expanded indications) is the non-clinical testing enumerated in section 8, which includes sterilization, shelf life, biocompatibility, and various bench tests. These tests confirmed that the device performs as intended and is safe when compared to established standards and the characteristics of the predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150915

    Validate with FDA (Live)

    Device Name
    SPINAUT-P
    Manufacturer
    Date Cleared
    2016-03-18

    (347 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SPINUAT-P is indicated for ablation, and decompression of disc material to treat symptomatic patients with contained herniated discs.

    Device Description

    The SPINAUT-P is a minimally invasive RF electrode device offered in 496mm length for percutaneous disc decompression surgery using radiofrequency energy and designed for use with radiofrequency generators in RF procedure. SPINAUT-P is intended to be placed into the epidural space or posterolateral / posterocentral area in the disc to remove, ablate or coagulate disc material in the area. The electrode consists of two stainless steel shafts with straight electrode configuration design. The shaft is housed in a coating composed of Polyamide. The device is single-use gamma radiation sterilized and is compatible with most standard electrosurgical generators that provide a cable with two pin outlet (2-pin, 22mm) with 1kVp maximum electrical capacity.

    AI/ML Overview

    The provided text is a 510(k) summary for the SPINAUT-P device, an electrosurgical device for disc decompression. It details the device's technical specifications and the non-clinical tests performed to demonstrate its substantial equivalence to predicate devices. However, it does not contain any information about a clinical study involving human patients or the performance of a device against specific acceptance criteria in a clinical setting.

    The document focuses on non-clinical (bench and lab) testing to support a premarket notification for a medical device. Therefore, I cannot provide details on acceptance criteria and a study that proves the device meets those criteria in a clinical context, or information related to human reader studies.

    Here's a breakdown of what can be extracted from the document regarding non-clinical performance and testing:

    1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical)

    The document primarily focuses on demonstrating that the SPINAUT-P device meets electrical safety, biocompatibility, sterilization, and some performance characteristics comparable to predicate devices. It lists test descriptions and referenced standards, implying that the acceptance criteria are met by adhering to these standards and the established "pre-set criteria" mentioned.

    Acceptance Criteria (Implied)Reported Device PerformanceComments
    Electrical Safety
    RF leakage current ≤ 7.2dfL mA (as per predicate)≤ 7.2dfL mAExplicitly stated as met, comparable to predicates.
    Electrical conductivityElectrode is able to pass electricity.Met.
    Cable Dielectric StrengthComplies with IEC 60601-2-2 Edition 5.0 2009-02, Clause 201.8.8.3.104 and 201.8.8.3.103.Met.
    High Frequency LeakageComplies with ANSI/AAMI HF 18:2001 Electrosurgical Devices Clause 4.2.5.2.Met.
    Material & Physical Properties
    Tensile Strength of Electrode CordComplies with ANSI/AAMI HF18:2001 Clause 4.2.5.5.Met.
    Appearance (Visual Inspection)"Visual Inspection" conducted.Implied satisfactory.
    Size Measurement"Measure the size of the test article by vernier calipers."Implied satisfactory and consistent with specifications.
    Sterilization
    SterilityComplies with ISO 11137-1, 2, 3 and ISO 11737-1, 2, 3 and Korean Pharmacopeia (10th Edition).Met pre-set criteria.
    Shelf Life
    Shelf Life Validation3-year shelf life validated using ASTM F1980-07 (accelerated aging).Met pre-set criteria.
    Packaging integrity
    Package Tensile TestComplies with ASTM D882-12.Met.
    Package Peeling TestComplies with ASTM F88/F88M-09.Met.
    Package Dye Penetration TestComplies with ASTM F1929-12.Met.
    Biocompatibility
    CytotoxicityComplies with ISO 10993-5:2009/(R) 2014.Supported as biocompatible.
    Skin SensitizationComplies with ISO 10993-10 Third Edition 2010-08-01.Supported as biocompatible.
    Intracutaneous reactivityComplies with ISO 10993-10 Third Edition 2010-08-01.Supported as biocompatible.
    Acute Systemic ToxicityComplies with ISO 10993-11 Second edition 2006-08-15.Supported as biocompatible.
    PyrogenComplies with USP 37-NF32:2014, <151>.Supported as biocompatible.
    HaemolysisComplies with ISO 10993-4:2002.Supported as biocompatible.
    Performance Comparison (Porcine Spines)
    Tissue Removal PerformanceDemonstrated comparable performance to predicate device (L'disQ).The test demonstrated the device being as safe and as effective as the predicate.
    Durability of ElectrodeDemonstrated comparable performance to predicate device (L'disQ).The test demonstrated the device being as safe and as effective as the predicate.
    Ability to be introduced, directed to and extracted from the surgical site (target area)Demonstrated comparable performance to predicate device (L'disQ).The test demonstrated the device being as safe and as effective as the predicate.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document mentions tests on "porcine spines" for performance comparison, but does not specify the number of porcine spines or individual tests conducted. For other tests (sterilization, shelf life, biocompatibility, electrical safety), it refers to established standards (e.g., ISO, ASTM, ANSI/AAMI, USP) which dictate sample sizes, but these specific numbers are not presented in the summary.
    • Data Provenance: The data is from non-clinical (bench and lab) testing, not human studies. The tests were performed by or for IMEDICOM Co., Ltd. (Republic of Korea). The specific location of testing labs is not detailed beyond the company's origin.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    • This information is not applicable to the non-clinical testing described. Ground truth or expert consensus is typically associated with clinical studies involving human data, where human experts (e.g., radiologists, pathologists) provide annotations or diagnoses.

    4. Adjudication Method for the Test Set

    • This information is not applicable as the document describes non-clinical laboratory testing, not a clinical study where adjudication of human expert opinions would be necessary.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not conducted. The document focuses on demonstrating substantial equivalence through non-clinical testing, not a clinical trial comparing human readers with and without AI assistance. The device itself is an electrosurgical tool, not an AI diagnostic or assistance system.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • No, a standalone algorithm performance study was not conducted. This device is an electrosurgical accessory, not an AI algorithm.

    7. Type of Ground Truth Used

    • For the non-clinical performance comparison using porcine spines, the "ground truth" would be the objective measurements and observations of tissue removal, durability, and navigability, compared against the predicate device under controlled experimental conditions.
    • For other tests (electrical, sterility, biocompatibility), the "ground truth" is adherence to the published specifications and pass/fail criteria defined by the referenced international and national standards (e.g., ISO, ASTM, ANSI/AAMI, USP).

    8. Sample Size for the Training Set

    • This information is not applicable. The device is an electrosurgical tool, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • This information is not applicable for the same reason as point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K151268

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-03-16

    (308 days)

    Product Code
    Regulation Number
    888.1100
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used with a fiberoptic endoscope, SPINAUT-V can be used for observing epidural anatomy, pathology and delivery of drugs approved for epidural indications.

    The SPINAUT-S is intended to be used by physicians for the illumination and visualization of tissues of the epidural space in the lumbar and sacral spine for the purpose of assisting in the diagnosis of disease.

    SPINAUT-I is intended for the percutaneous introduction and placement of a video guided catheter and/or an endoscope.

    Device Description

    SPINAUT-V consists of a flexible catheter, steering handle and associated ports for access to the lumen. The catheter has built in steering mechanism that allows for guiding the soft tip through the epidural space and soft tissues for optimal access to the source of distress. The design allows the use of a flexible fiber optic endoscope for visual examination of the area and surrounding tissues at the distal end of the catheter. This visual examination then allows the physician to diagnose the potential causes of neural distress and pain. SPINAUT-V is a sterile, single use device. When placing SPINAUT-V into the patient, SPINAUT-I can be used instead of SPINUT-V as an introducer device.

    SPINAUT-S is a surgically invasive flexible fiberoptic system that can be used either with a video guided catheter (SPINAUT-V or other video guided catheters with proper catheter size) or with an introducer device such as SPINAUT-I. The SPINIUT-S is a reusable device that is supplied non-sterile with instructions for cleaning, sterilization and re-use. The SPINAUT-S can be used with the light source and camera system.

    SPINAUT-I is intended to be used to prepare or preserve a path for a video guided catheter or an endoscope. It is supplied sterile and for single use.

    AI/ML Overview

    The provided document describes the substantial equivalence of the SPINAUT-V, SPINAUT-S, and SPINAUT-I devices to a predicate device, Myelotec Video Guided Catheter (K980734). It does not contain information about acceptance criteria and a study that proves the device meets those criteria in the context of a typical AI/ML medical device submission.

    The document is a 510(k) summary for a traditional medical device (arthroscope/catheter/introducer) and focuses on bench testing, biocompatibility, and sterilization validation to demonstrate substantial equivalence to a predicate device, rather than clinical performance based on defined acceptance criteria like sensitivity/specificity/accuracy or multi-reader studies.

    Therefore, many of the requested fields cannot be extracted directly from this document as they are not applicable to the type of device and submission described here.

    Here's a breakdown of what can and cannot be provided based on the input:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Bench Test Type)Reported Device Performance (Summary)Device(s)
    Appearance (Visual Inspection)Met pre-set criteriaSPINAUT-V, SPINAUT-I
    Leakage Test (Pressure gauge)Met pre-set criteriaSPINAUT-V, SPINAUT-I
    Tensile strength (ISO 10555-1)Met pre-set criteriaSPINAUT-V, SPINAUT-I
    Curved strength (Carved angle measurement)Met pre-set criteriaSPINAUT-V
    Fatigue testMet pre-set criteriaSPINAUT-V
    X-ray impermeable testMet pre-set criteria (shown clearly)SPINAUT-V
    Corrosion resistance test (ISO 10555-1)Met pre-set criteriaSPINAUT-V
    Flexural Rigidity (MFDS Notification No. 2013-214)Met pre-set criteriaSPINAUT-V, SPINAUT-I
    Sterility test (Korean Pharmacopeia)Met pre-set criteriaSPINAUT-V, SPINAUT-I
    Direction of View (ISO 8600-3)Met pre-set criteriaSPINAUT-S
    Field of View (ISO 8600-3:2013)90° (Subject Device) vs 60° (Predicate Device) - Met specifications and performed as well as predicate.SPINAUT-S
    Observation Depth (Depth of Focus) (ISO 8600-1:2005, ISO 12233:2014)3 mm(f₁) ~ 20 mm(f₂) (Subject Device) vs 5 mm(f₁) ~ 7.0 mm(f₂) (Predicate Device) - Met specifications and performed as well as predicate.SPINAUT-S
    Internal Pressure (Capacity to Resist Pressure) (ISO 8600-1:2005)Met pre-set criteriaSPINAUT-S
    Dimensions (Total length, insertion portion, outer diameter)Met pre-set criteria (Specific values provided for comparison but implicitly met; e.g., SPINAUT-S total length 950.5 mm vs predicate 994.0 mm)SPINAUT-S
    Connecting Part (Conical) Fitting (ISO 594-1, ISO 594:2)Met pre-set criteriaSPINAUT-I
    Biocompatibility (ISO 10993-4, 5, 10, 11 and USP 37 <151>)BiocompatibleAll Devices (SPINAUT-V, S, I)
    Sterilization Validation (ISO 11137-1, 2, 3 and ISO 11737-1, 2, 3)Met pre-set criteriaAll Devices (SPINAUT-V, S, I)
    Shelf Life Validation (ASTM F1980-07)3-year shelf life validatedAll Devices (SPINAUT-V, S, I)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
    The document describes bench tests and validation tests rather than clinical studies with "test sets" in the context of diagnostic accuracy. Therefore, information on sample size for image/patient data and data provenance like country of origin or retrospective/prospective is not applicable/provided. The tests cited are standard methods for device manufacturing and safety.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
    Not applicable. The tests performed are engineering and material science tests, not clinical performance evaluations requiring expert ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This is not an AI/ML device, and no MRMC study information is provided.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
    For the bench tests, the ground truth is the performance standards and specifications defined by the referenced ISO, ASTM, MFDS, and USP standards (e.g., a catheter must not leak under a certain pressure, a material must be biocompatible, a sterilization process must achieve a certain Sterility Assurance Level).
    For the biocompatibility tests, the ground truth is adherence to specific biological response criteria as per ISO 10993 and USP standards.
    For sterilization and shelf-life validation, the ground truth is the scientific principles and validated methodologies outlined in the relevant ISO and ASTM standards.

    8. The sample size for the training set:
    Not applicable. This is not an AI/ML device using training data.

    9. How the ground truth for the training set was established:
    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150789

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-03-01

    (342 days)

    Product Code
    Regulation Number
    868.5120
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SPINAUT-E is used for delivery of drugs that have been indicated for the epidural space.

    SPINAUT-I is intended for the percutaneous introduction and placement of an epidural catheter.

    Device Description

    SPINAUT-E consists of a flexible catheter, steering handle and a port for access to the lumen. The catheter has built in steering mechanism that allows for guiding the soft tip through the epidural space and soft tissues for optimal access to the source of distress. The port facilitates the connection of syringes to deliver therapeutic agents by physicians as appropriate to their diagnosis. It is supplied sterile and it is for single use.

    There are 9 models in SPINAUT-E: E1000, E2000, E1100, E2100, E3100, S1000, S2000, and S3000. The catheter tip of the model E1000, and E3000 can be bent 50° to the left or to the right (the user can rotate the tip to adjust the direction), but the tip of the model E1100, E2100, and E 3100 can be bent either direction up to 100°. The S1000, S2000, and S3000 have an extra infusion port for drug so that the other main port can be used for other devices. Each model is characterized by the additional port, the lever, and the dimensions.

    The SPINAUT-I is intended to be used to prepare or preserve a path for an epidural catheter. When placing SPINAUT-E into the patient, SPINAUT-I can be used as a replacement of the SPINUT-E Catheter Introducer Set. It is supplied sterile and it is for single use.

    There are 3 models in SPINAUT-I: ID1000, ID2000, and ID3000 and each model is characterized by the length.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device (SPINAUT-E and SPINAUT-I anesthesia conduction catheters), not a study demonstrating AI performance. Therefore, most of the requested information regarding AI acceptance criteria, sample sizes for test/training sets, expert ground truth, MRMC studies, or standalone performance is not applicable to this document.

    However, I can extract information related to the non-clinical testing performed to demonstrate substantial equivalence to predicate devices, which serves a similar purpose of validating device performance against pre-set criteria.

    Here's a summary based on the provided text, focusing on the available non-clinical testing details:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document broadly states that "the test results met the pre-set criteria" for all tests. It does not provide specific numerical acceptance criteria or detailed numerical device performance outcomes for each test. Instead, it lists the types of tests conducted and concludes that the predefined criteria were met.

    Test TypeAcceptance Criteria (Implied)Reported Device Performance
    Sterilization ValidationMet pre-set criteria (ISO 11137-1, 2, 3; ISO 11737-1, 2, 3)Met pre-set criteria
    Shelf Life Validation (Accelerated Aging)Validated 3-year shelf life (ASTM F1980-07)Validated 3-year shelf life
    BiocompatibilityBiocompatible (ISO 10993, USP 37 & 38)Supported that devices are biocompatible
    Performance Bench Tests (SPINAUT-E)Met pre-set criteriaMet pre-set criteria
    Performance Bench Tests (SPINAUT-I)Met pre-set criteriaMet pre-set criteria

    Specific Tests listed under "Performance Bench Tests" (SPINAUT-E and SPINAUT-I):

    • Visual Inspection (Appearance)
    • Size Measurement (SPINAUT-E only for shelf life, SPINAUT-I for shelf life and performance)
    • Leakage Test (Catheter)
    • Tensile strength (Catheter)
    • Curved strength (Catheter)
    • Fatigue test (Catheter)
    • X-ray impermeable test
    • Corrosion resistance test
    • Flexural Rigidity (Catheter introducer and the needle)
    • Pulling (Drawing) Test (Catheter introducer and the needle)
    • Package Tensile Test
    • Package Peeling Test
    • Package Burst Test
    • Package Dye Penetration Test
    • Package Bubble Test
    • Sterility Test
    • Connecting Part (Conical) Fitting (SPINAUT-I only)

    2. Sample size used for the test set and the data provenance:

    • The document does not specify the exact sample sizes used for each individual non-clinical test. It only states that tests were performed.
    • Data Provenance: The tests are "non-clinical testing" and "bench tests" performed by the manufacturer, IMEDICOM Co., Ltd., which is based in the Republic of Korea. The data provenance is industrial laboratory testing, not patient-derived data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This is not a study involving expert assessment or ground truth establishment in the traditional sense of medical image analysis or clinical trials. The "ground truth" for these non-clinical tests is based on objective measurements against established engineering and medical device standards (e.g., ISO, ASTM, USP).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a study requiring adjudication of expert opinions.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This document describes a traditional medical device (catheter) and its non-clinical testing for 510(k) clearance, not an AI-powered device or an MRMC study.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the non-clinical tests is based on objective measurements against established international and national standards (e.g., ISO 11137, ISO 11737, ASTM F1980-07, ISO 10993, USP 37, USP 38) and the manufacturer's pre-set internal criteria for performance specifications (e.g., tensile strength, leakage rates, sterility).

    8. The sample size for the training set:

    • Not applicable. There is no AI training set involved.

    9. How the ground truth for the training set was established:

    • Not applicable. There is no AI training set involved.
    Ask a Question

    Ask a specific question about this device

    K Number
    K133669

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-04-04

    (126 days)

    Product Code
    Regulation Number
    888.3027
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MEDINAUT Kyphoplasty System is intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. This system is to be used with cleared spinal Polymethylmethacrylate (PMMA) bone cements indicated for use during percutaneous vertebral augmentation, such as kyphoplasty.

    Device Description

    The MEDINAUT Kyphoplasty System is designed to reduce spinal compression fracture and restore sagittal alignment. By creating a space in the vertebral body it facilitate the insertion of bone cement through the use of the cement dispensing plunger. The benefits of MEDINAUT Kyphoplasty System are the reduction in back pain and increase of patient's functional abilities, allowing a return to the previous level of activity.

    It consists of Balloon Catheter, Bone Expander Syringe, Needle Pipe, Needle Pin, Wire Pin (Troca Type, Round Type), Cannula, Expander, Spacer, Cement Pusher, Cement Filler, and Guide Wire. It is supplied sterile and for single disposable use.

    The MEDINAUT Kyphoplasty System is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. The system is substantially equivalent in design, function and intended use to the predicate devices.

    AI/ML Overview

    This device is a medical device, not an AI/ML device, so many of the requested fields are not applicable.

    Here's an analysis of the provided 510(k) summary for the MEDINAUT Kyphoplasty System, focusing on acceptance criteria and performance data:

    The MEDINAUT Kyphoplasty System is a physical medical device (an inflatable bone tamp system) used in kyphoplasty procedures. The "acceptance criteria" and "study" described are primarily bench tests comparing the device's physical properties and performance against a predicate device, rather than clinical studies or evaluations of an AI model's diagnostic accuracy.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document doesn't explicitly state "acceptance criteria" in the format of pass/fail thresholds for each test. Instead, it presents "Test Criteria" which are more akin to the methodology or conditions of the test, and then provides "Test Result" for both the MEDINAUT Kyphoplasty System and the predicate device (Kyphon Inflatable Bone Tamp). The implicit acceptance criterion for this 510(k) submission is substantial equivalence to the predicate device. This means the new device's performance should be comparable to or better than the predicate's, and not raise new questions of safety or effectiveness.

    TestTest CriteriaMEDINAUT Kyphoplasty System (Reported Performance)Kyphon Inflatable Bone Tamp (Predicate Performance)
    Tensile Bond StrengthOuter shaft and Hub with 1mm/sec, Test report #; IMT-TBST12-01IBE-15; 20.85kgfK13A; 14.76kgf
    Fatigue StrengthNo burst and leakage at 15 cycles of maximum pressure limit 400psi, Test report #; IMT-FST12-01IBE-10; PassIBE-15; PassIBE-20; PassK13A; PassK09A; PassK08A; Pass
    Balloon Deflation TimeCompare the performance (seconds), Test report:IMT-BDT12-01IBE-10 2ml; 0.50 sec.3ml; 0.76 sec.IBE-15 3ml; 0.51 sec.5ml; 1.16 sec.IBE-20 3ml; 0.47 sec.5ml; 1.12 sec.7ml; 1.29 sec.K13A 2ml; 0.27 sec.4ml; 0.62 sec.K09A 2ml; 0.28 sec.4ml; 0.62mlK08A 2ml; 0.31 sec.4ml; 0.54 sec.6ml; 0.69 sec.
    Burst Pressure Constrained400 psi at the Constrained jig for 30 seconds, Test report: IMT-BPC12-08IBE-10; 3.5ml PassIBE-15; 5.5ml PassIBE-20; 7.5ml PassK09A; 5.5ml PassK08A; 7.5ml Pass
    Burst Strength UnconstrainedMeasure the burst inflation pressure, Test report #: IMT-BST12-01IBE-10; 250psiIBE-15; 207psiIBE-20; 183psiK09A; 243psiK08A; 237psi
    Balloon Dimension before and after inflationInitial balloon length (IBE-10; 10mm, IBE-15; 15mm, IBE-20; 20mm), Test report #: IMT-IDT11-02Balloon Diameter after inflation:IBE-10, 3ml: 14.1mmIBE-15, 5ml: 17.1mmIBE-20, 7ml: 19.3mmBalloon Length after inflation:IBE-10, 3ml: 16.3mmIBE-15, 5ml: 22.2mmIBE-20, 7ml: 29.4mmBalloon Diameter after inflation:K09A, 4ml: 15.2mmK08A, 6ml: 16.8mmBalloon Length after inflation:K09A, 4ml: 20.1mmK08A, 6ml: 29.2mm
    Insertion and Withdrawal ForceMeasure the insertion and withdrawal force with 0.01m/s test velocity, Test report #: IMT-IFWF12-01Insertion Force:IBE-10; MEAN 1.24NIBE-15; MEAN 1.65NIBE-20; MEAN 2.11NWithdrawal Force:IBE-10; MEAN 1.11NIBE-15; MEAN 1.42NIBE-20; MEAN 1.79NInsertion Force:K09A; MEAN 1.08NK08A; MEAN 1.49NWithdrawal Force:K09A; MEAN 0.90NK08A; MEAN 1.26N

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the exact sample size (e.g., number of devices tested for each parameter) for the bench tests. It refers to "IBE-10," "IBE-15," "IBE-20" for the applicant's device and "K13A," "K09A," "K08A" for the predicate, which likely correspond to different sizes or models of the devices tested. Without further detail, we cannot determine the n for each specific test.

    The data provenance is from bench testing conducted by IMEDICOM Co., Ltd. (Republic of Korea) and compared against the Kyphon Inflatable Bone Tamp. This is laboratory data, not patient-derived data, and therefore the concepts of "country of origin of the data," "retrospective or prospective" (in a clinical sense) are not directly applicable.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This question is not applicable to this type of device and study. The "ground truth" for these bench tests is established by physical measurement and engineering standards, not by expert interpretation of medical images or patient outcomes. The physical properties (e.g., tensile strength, burst pressure) are objectively measurable.

    4. Adjudication Method for the Test Set

    This question is not applicable. Adjudication methods (like 2+1 or 3+1) are used for establishing ground truth in human-reviewed data sets, especially when there's variability in expert interpretations. Bench tests rely on standardized measurement protocols.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI/ML diagnostic devices where the output of the AI is compared to human readers' performance, with or without AI assistance, on a set of clinical cases. The MEDINAUT Kyphoplasty System is a physical surgical tool, not a diagnostic AI.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, this is not applicable. The device is a physical, non-AI medical tool. The concept of "standalone performance" without human intervention doesn't apply to a surgical instrument that is used by a surgeon.

    7. The Type of Ground Truth Used

    The "ground truth" for the bench tests is based on engineering measurements and specifications of the physical properties (e.g., strength, pressure resistance, dimensions, forces). The comparison is made against the performance of a legally marketed predicate device to demonstrate substantial equivalence.

    8. The Sample Size for the Training Set

    This question is not applicable. As a physical medical device, there is no "training set" in the context of machine learning or AI. The design of the device is based on engineering principles and iterative development, not data-driven machine learning training.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1