K Number
K151268

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-03-16

(308 days)

Product Code
Regulation Number
888.1100
Age Range
0 - 21
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

When used with a fiberoptic endoscope, SPINAUT-V can be used for observing epidural anatomy, pathology and delivery of drugs approved for epidural indications.

The SPINAUT-S is intended to be used by physicians for the illumination and visualization of tissues of the epidural space in the lumbar and sacral spine for the purpose of assisting in the diagnosis of disease.

SPINAUT-I is intended for the percutaneous introduction and placement of a video guided catheter and/or an endoscope.

Device Description

SPINAUT-V consists of a flexible catheter, steering handle and associated ports for access to the lumen. The catheter has built in steering mechanism that allows for guiding the soft tip through the epidural space and soft tissues for optimal access to the source of distress. The design allows the use of a flexible fiber optic endoscope for visual examination of the area and surrounding tissues at the distal end of the catheter. This visual examination then allows the physician to diagnose the potential causes of neural distress and pain. SPINAUT-V is a sterile, single use device. When placing SPINAUT-V into the patient, SPINAUT-I can be used instead of SPINUT-V as an introducer device.

SPINAUT-S is a surgically invasive flexible fiberoptic system that can be used either with a video guided catheter (SPINAUT-V or other video guided catheters with proper catheter size) or with an introducer device such as SPINAUT-I. The SPINIUT-S is a reusable device that is supplied non-sterile with instructions for cleaning, sterilization and re-use. The SPINAUT-S can be used with the light source and camera system.

SPINAUT-I is intended to be used to prepare or preserve a path for a video guided catheter or an endoscope. It is supplied sterile and for single use.

AI/ML Overview

The provided document describes the substantial equivalence of the SPINAUT-V, SPINAUT-S, and SPINAUT-I devices to a predicate device, Myelotec Video Guided Catheter (K980734). It does not contain information about acceptance criteria and a study that proves the device meets those criteria in the context of a typical AI/ML medical device submission.

The document is a 510(k) summary for a traditional medical device (arthroscope/catheter/introducer) and focuses on bench testing, biocompatibility, and sterilization validation to demonstrate substantial equivalence to a predicate device, rather than clinical performance based on defined acceptance criteria like sensitivity/specificity/accuracy or multi-reader studies.

Therefore, many of the requested fields cannot be extracted directly from this document as they are not applicable to the type of device and submission described here.

Here's a breakdown of what can and cannot be provided based on the input:

1. A table of acceptance criteria and the reported device performance:

Acceptance Criteria (Bench Test Type)Reported Device Performance (Summary)Device(s)
Appearance (Visual Inspection)Met pre-set criteriaSPINAUT-V, SPINAUT-I
Leakage Test (Pressure gauge)Met pre-set criteriaSPINAUT-V, SPINAUT-I
Tensile strength (ISO 10555-1)Met pre-set criteriaSPINAUT-V, SPINAUT-I
Curved strength (Carved angle measurement)Met pre-set criteriaSPINAUT-V
Fatigue testMet pre-set criteriaSPINAUT-V
X-ray impermeable testMet pre-set criteria (shown clearly)SPINAUT-V
Corrosion resistance test (ISO 10555-1)Met pre-set criteriaSPINAUT-V
Flexural Rigidity (MFDS Notification No. 2013-214)Met pre-set criteriaSPINAUT-V, SPINAUT-I
Sterility test (Korean Pharmacopeia)Met pre-set criteriaSPINAUT-V, SPINAUT-I
Direction of View (ISO 8600-3)Met pre-set criteriaSPINAUT-S
Field of View (ISO 8600-3:2013)90° (Subject Device) vs 60° (Predicate Device) - Met specifications and performed as well as predicate.SPINAUT-S
Observation Depth (Depth of Focus) (ISO 8600-1:2005, ISO 12233:2014)3 mm(f₁) ~ 20 mm(f₂) (Subject Device) vs 5 mm(f₁) ~ 7.0 mm(f₂) (Predicate Device) - Met specifications and performed as well as predicate.SPINAUT-S
Internal Pressure (Capacity to Resist Pressure) (ISO 8600-1:2005)Met pre-set criteriaSPINAUT-S
Dimensions (Total length, insertion portion, outer diameter)Met pre-set criteria (Specific values provided for comparison but implicitly met; e.g., SPINAUT-S total length 950.5 mm vs predicate 994.0 mm)SPINAUT-S
Connecting Part (Conical) Fitting (ISO 594-1, ISO 594:2)Met pre-set criteriaSPINAUT-I
Biocompatibility (ISO 10993-4, 5, 10, 11 and USP 37 <151>)BiocompatibleAll Devices (SPINAUT-V, S, I)
Sterilization Validation (ISO 11137-1, 2, 3 and ISO 11737-1, 2, 3)Met pre-set criteriaAll Devices (SPINAUT-V, S, I)
Shelf Life Validation (ASTM F1980-07)3-year shelf life validatedAll Devices (SPINAUT-V, S, I)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
The document describes bench tests and validation tests rather than clinical studies with "test sets" in the context of diagnostic accuracy. Therefore, information on sample size for image/patient data and data provenance like country of origin or retrospective/prospective is not applicable/provided. The tests cited are standard methods for device manufacturing and safety.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
Not applicable. The tests performed are engineering and material science tests, not clinical performance evaluations requiring expert ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is not an AI/ML device, and no MRMC study information is provided.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This is not an AI/ML device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
For the bench tests, the ground truth is the performance standards and specifications defined by the referenced ISO, ASTM, MFDS, and USP standards (e.g., a catheter must not leak under a certain pressure, a material must be biocompatible, a sterilization process must achieve a certain Sterility Assurance Level).
For the biocompatibility tests, the ground truth is adherence to specific biological response criteria as per ISO 10993 and USP standards.
For sterilization and shelf-life validation, the ground truth is the scientific principles and validated methodologies outlined in the relevant ISO and ASTM standards.

8. The sample size for the training set:
Not applicable. This is not an AI/ML device using training data.

9. How the ground truth for the training set was established:
Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 16, 2016

IMEDICOM Co., Ltd. % Ms. Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 2651 E Chapman Ave Ste 110 Fullerton, California 92831

Re: K151268

Trade/Device Name: SPINAUT-V. SPINAUT-S. SPINAUT-I Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: February 12, 2016 Received: February 22, 2016

Dear Ms. Priscilla Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151268

Device Name

SPINAUT-V, SPINAUT-S, and SPINAUT-I

Indications for Use (Describe)

When used with a fiberoptic endoscope, SPINAUT-V can be used for observing epidural anatomy, pathology and delivery of drugs approved for epidural indications.

The SPINAUT-S is intended to be used by physicians for the illumination of tissues of the epidural space in the lumbar and sacral spine for the purpose of assisting in the diagnosis of disease.

SPINAUT-I is intended for the percutaneous introduction and placement of a video guided catheter and/or an endossope.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

|X | Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

(K151268)

This summary of 510(K) - safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: Mar 15, 2016

1. Applicant / Submitter

IMEDICOM Co., Ltd. #612, 172, LS-ro, Gunpo-si, Gyeonggi-do, 435-824 Republic of Korea Tel:+82-31-479-1156 Fax:+82-31-479-1157

2. Submission Correspondent

LK Consulting Group USA, Inc. 2651 E Chapman Ave. Ste 110, Fullerton CA 92831 Priscilla Chung Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com

3. Device

  • Trade Name: SPINAUT-V, SPINAUT-S, and SPINAUT-I
  • Common Name: Arthroscope
  • י Classification Name: Arthroscope
  • Product Code: HRX "
  • ' Classification regulation: Class II, 21 CFR 888.1100

4. Predicate Device:

  • י Myelotec Video Guided Catheter by Myelotec., Inc (K980734)

5. Description:

SPINAUT-V consists of a flexible catheter, steering handle and associated ports for access to the lumen. The catheter has built in steering mechanism that allows for guiding the soft tip through the epidural space and soft tissues for optimal access to the source of

{4}------------------------------------------------

distress. The design allows the use of a flexible fiber optic endoscope for visual examination of the area and surrounding tissues at the distal end of the catheter. This visual examination then allows the physician to diagnose the potential causes of neural distress and pain. SPINAUT-V is a sterile, single use device. When placing SPINAUT-V into the patient, SPINAUT-I can be used instead of SPINUT-V as an introducer device.

SPINAUT-S is a surgically invasive flexible fiberoptic system that can be used either with a video guided catheter (SPINAUT-V or other video guided catheters with proper catheter size) or with an introducer device such as SPINAUT-I. The SPINIUT-S is a reusable device that is supplied non-sterile with instructions for cleaning, sterilization and re-use. The SPINAUT-S can be used with the light source and camera system.

SPINAUT-I is intended to be used to prepare or preserve a path for a video guided catheter or an endoscope. It is supplied sterile and for single use.

6. Indication for use:

When used with a fiberoptic endoscope, SPINAUT-V can be used for observing epidural anatomy, pathology and delivery of drugs approved for epidural indications.

The SPINAUT-S is intended to be used by physicians for the illumination and visualization of tissues of the epidural space in the lumbar and sacral spine for the purpose of assisting in the diagnosis of disease.

SPINAUT-I is intended for the percutaneous introduction and placement of a video guided catheter and/or an endoscope.

7. Basis for Substantial Equivalence

7.1. SPINAUT-V

SPINUAT-V is substantially equivalent to the Myelotec Video Guided Catheter (K980734) made by Myelotec, Inc.

There are no significant differences between the subject device and the predicate device. They share the following same characteristics.

  • Intended Use
  • Principle of Operation
  • Design
  • Size
  • Sterilization Method

The materials used in the subject devices might be different from the predicate device (the materials of the predicate device are unknown.); however, the biocompatibility testing results of the subject device support that the subject devices are biocompatible

{5}------------------------------------------------

and the performance testing results show that the subject device would perform as well as the predicate device. Based on the test results, we conclude that despite the difference, the subject device is substantially equivalent to the predicate device.

Subject DevicePredicate Device
Device NameSPINAUT-VMyelotec Video Guided Catheter
510(k) NumberK151268K980734
ManufacturerIMEDICOM Co., Ltd.Myelotec., Inc.
Product CodeHRXHRX
Intended UseWhen used with a fiberopticendoscope, SPINAUT-V can be usedfor observing epidural anatomy,pathology and delivery of drugsapproved for epidural indications.When used with a fiberopticendoscope, this device can be usedfor observing epidural anatomy,pathology and delivery of drugsapproved for epidural indications.
Diameter(O.D.) ofCatheter3.0mm2.7 mm, 3.0mm, 3.3mm
Lengthof Catheter300mm/ 315mm/ 330mm300mm
Infusion portDual portsDual ports
MaterialSus 304, PU, ABS, PE, PTFE
SteeringSteerableSteerable
Single useYesYes
ComponentsVideo guided catheter, 15G needle,needle cap, introducer, guide wireVideo guided catheter, 15G needle,needle cap, introducer, guide wire
BiocompatibilityBiocompatibleBiocompatible
Endoscope capableCapableCapable
Insertion PointSacral HiatusSacral Hiatus
SterilizationGamma SterilizationGamma Sterilization

{6}------------------------------------------------

7.2. SPINAUT-S

SPINUAT-V is substantially equivalent to the Myelotec Video Guided Catheter (K980734) made by Myelotec, Inc.

There are no significant differences between the subject device and the predicate device. They share the following same characteristics.

  • Intended Use
  • · Principle of Operation
  • Materials
  • · Design
  • · Sterilization Method

The field of view and the observation depth of the subject device are different from the predicate device; however, the performance testing results show that the subject device meets the specifications of those performance characteristics and also supports that it would perform as well as the predicate device. Based on the test results, we conclude that the subject device is substantially equivalent to the predicate device.

Subject DevicePredicate Device
Trade nameSPINAUT-SMyelotec Video Guided Catheter
510(k) NumberK151268K980734
ManufacturerIMEDICOM Co., Ltd.Myelotec., Inc.
Product CodeHRXHRX
Intended for useThe SPINAUT-S is intended to be used byphysicians for the illumination andvisualization of tissues of the epiduralspace in the lumbar and sacral spine forthe purpose of assisting in the diagnosis ofdisease.When used with a fiberoptic endoscope,this device can be used for observingepidural anatomy, pathology and deliveryof drugs approved for epidural indications.
PerformancePermits direct visualization of conditionsinside the lower spine with minimalimpact on the patient's musculoskeletalstructurePermits direct visualization of conditionsinside the lower spine with minimalimpact on the patient's musculoskeletalstructure
MaterialOpticalglassInorganic glassInorganic glass
TubePolyamidePolyamide
Inner pathOptical fiberOptical fiber

{7}------------------------------------------------

PerformanceField of view90°60°
Direct view0°±10%0°±10%
Observation Depth3 mm(f₁) ~ 20 mm(f₂)5 mm(f₁) ~ 7.0 mm(f₂)
SizeTotal length950.5 mm994.0 mm
Compatibility w/other devicesVideo Guided Catheter Universal camera systemsVideo Guided Catheter Universal camera systems
Chemical SafetyBiocompatibleBiocompatible
SterilizationNon-sterilizedNon-sterilized

7.3. SPINAUT-I

SPINUAT-V is substantially equivalent to the Myelotec Video Guided Catheter (K980734) made by Myelotec, Inc.

There are no significant differences between the subject device and the predicate device. They share the following same characteristics.

  • Intended Use
  • Principle of Operation
  • · Design
  • Materials
  • Sterilization Method

We have found no significant differences between the two devices which would raise different questions of safety and effectiveness.

Subject DevicePredicate Device
Trade nameSPINAUT-IMyelotec Video Guided Catheter
510(k) NumberK151268K980734
ManufacturerIMEDICOM Co., Ltd.Myelotec., Inc.
Product CodeHRXHRX

{8}------------------------------------------------

Intended for useSPINAUT-I is intended for thepercutaneous introduction andplacement of Video Guided Catheterand/or Endoscope.When used with a fiberoptic endoscope,this device can be used for observingepidural anatomy, pathology anddelivery of drugs approved for epiduralindications.
PerformanceTo be used to prepare or preserve a pathfor a lead.To be used to prepare or preserve a pathfor a lead.
MaterialBodyABSABS
TubePolyurethanePolyurethane
Compatibility w/otherdevicesVideo Guided CatheterEndoscopeVideo Guided CatheterEndoscope
Chemical SafetyBiocompatibleBiocompatible
SterilizationGamma SterilizationGamma Sterilization

8. Non-Clinical Testing

  • · Sterilization validating testing has been performed in accordance with ISO 11137-1, 2, 3 and ISO 11737-1, 2, 3, and the test results met the pre-set criteria.
  • The tests to validate the shelf life of the device through the proposed shelf life were conducted using the accelerated aging method in accordance with ASTM F1980-07 and the test results validated 3 year shelf life.
  • · Biocompatibility tests were performed in accordance with ISO 10993-4, 5, 10, 11 and USP 37 <151>, and the test results supported that the subject devices are biocompatible.
  • · The following bench tests were performed to evaluate the performance and the safety of the subject devices and the test results met the pre-set criteria.
No.TestTest Description
1AppearanceVisual Inspection
2Leakage Test(Catheter)Evaluate if the catheter leaks any water using a pressuregauge
3Tensile strength(Catheter)Test in accordance with ISO 10555-1, IntravascularCatheters - sterile and single use catheter – Part 1:General requirements.
4Curved strength(Catheter)Measure a carved angle of the catheter.
SPINAUTr. V
1999

{9}------------------------------------------------

5Fatigue test(Catheter)Evaluate the fatigue strength of the catheter
6X-ray impermeabletestEvaluate if the device can be shown on an X-ray imageclearly.
7Corrosion resistancetestTest in accordance with ISO 10555-1 IntravascularCatheters - sterile and single use catheter - Part 1:General requirements.
8Flexural Rigidity(Catheter introducer)Test in accordance with MFDS Notification No. 2013-214: Standard Specifications for Medical Device –Sterile hypodermic needles for single use - Clause 5.
9Sterility testTest in accordance with Korean Pharmacopeia (10thEdition) : General Requirements for Tests and Assays /Sterility Test

SPINAUT-S

No.TestTest Description
1Direction of ViewTest in accordance with ISO 8600-3
2Field of ViewTest in accordance with ISO 8600-3:2013
3(Depth of Focus)Observation DepthTest in accordance with ISO 8600-1:2005 and ISO12233:2014
4Internal Pressure(Capacity to ResistPressure)Test in accordance with ISO 8600-1:2005
5DimensionsMeasure the total length, insertion portion and the outerdiameter by vernier calipers.

SPINAUT-I

No.TestTest Description
1AppearanceVisual Inspection
2Leakage TestEvaluate if the catheter leaks any water using a pressuregauge
3Tensile strengthMeasure tensile strength of the device
4Flexural RigidityTest in accordance with MFDS Notification No. 2013-214: Standard Specifications for Medical Device -Sterile hypodermic needles for single use - Clause 5.
5Connecting Part(Conical) FittingTest in accordance with ISO 594-1 and ISO 594:2.• Gauging test• Liquid leakage• Air leakage• Separation force• Unscrewing torque• Ease of assembly

{10}------------------------------------------------

▪ Resistance to overriding▪ Stress cracking
6Sterility testTest in accordance with Korean Pharmacopeia(10th Edition) : General Requirements for Tests andAssays / Sterility Test

The test results supported that the subject device is substantially equivalent to the predicate devices.

9. Conclusion

Based on the similarities, we conclude that the SPINAUT-V, SPINAUT-S, and SPINAUT-I are substantially equivalent to the predicate devices.

§ 888.1100 Arthroscope.

(a)
Identification. An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.(b)
Classification. (1) Class II (performance standards).(2) Class I for the following manual arthroscopic instruments: cannulas, currettes, drill guides, forceps, gouges, graspers, knives, obturators, osteotomes, probes, punches, rasps, retractors, rongeurs, suture passers, suture knotpushers, suture punches, switching rods, and trocars. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 888.9.