510(k) Data Aggregation

    K Number
    K081394

    Validate with FDA (Live)

    Date Cleared
    2008-09-15

    (119 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PTFE Peelable Introducer is intended for use in the percutaneous insertion of pacing leads or catheters in the venous system.

    Device Description

    The Enpath Medical PTFE Peelable Introducer kit consists of a disposable needle, a disposable syringe, a guidewire, and a peelable introducer set consisting of a dilator and sheath with integrated proximal handles. The Introducer is available in various lengths in 3.5F through 7F (micro) and 7F through 16.5F (macro). The dilator is designed to be delivered over a 0.018" guidewire (micro) up to a 0.038" guidewire (macro). The dilator and hub is constructed of high density polyethylene. The outer sheath and handle is constructed of PTFE and TPX. The inner dilator and outer sheath are radiopaque so that they are visible under fluoroscopy. The hub of the dilator has a spin lock design that connects to the luer of the sheath when rotated. The Percutaneous Introducer sheath has a "tear-away" feature which allows the user to remove the sheath without removing the inserted catheter or pacing lead. The kit is packaged in a tray and placed into a poly-Tyvek pouch and sealed.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the PTFE Peelable Introducer:

    This document is a 510(k) summary for a medical device seeking substantial equivalence to a predicate device. It focuses on non-clinical testing for equivalence rather than establishing new clinical effectiveness or performance criteria for a novel device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Device functions in a safe and effective mannerTest results verify that the device performs per specification requirements and is equivalent to the predicate device.
    Device performs per specification requirementsTest results confirm the device performs as intended.
    Device is equivalent to the predicate device without creating additional risk to the patient or userSubstantially equivalent to the currently marketed Enpath Medical Percutaneous Introducers based on a comparison of the indications for use and the technological characteristics.
    Functional and dimensional testing of the PTFE Peelable Introducer meets requirementsPerformance testing included functional and dimensional testing. (Specific criteria not detailed, but states it met them).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not specified. The document states "Performance Testing: The performance testing for this device included testing to verify that the device continues to functions in a safe and effective manner." This implies a physical testing or engineering study, not a clinical trial with human subjects.
    • Data Provenance: The testing was "non-clinical." This means it was likely laboratory or bench testing performed by the manufacturer, Enpath Medical, probably in the US where the company is based. It is retrospective in the sense that it's based on internal validation against predefined specifications rather than a prospective trial with patient enrollment.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    • Not Applicable. As this was non-clinical performance testing, there was no "ground truth" to be established by clinical experts in the traditional sense. The "ground truth" was likely the engineering specifications and performance characteristics of the predicate device.

    4. Adjudication Method for the Test Set

    • Not Applicable. There was no human reading or interpretation of results that would require an adjudication method. The testing involved objective measurements against specifications.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. The document explicitly states: "Summary of Clinical Testing: No clinical evaluations of this product have been performed." Therefore, no MRMC study was conducted.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Not Applicable. This is a physical medical device (an introducer kit), not an AI algorithm or software. Therefore, the concept of "standalone performance" for an algorithm does not apply. The performance assessed was the physical device itself.

    7. Type of Ground Truth Used

    • Engineering Specifications / Predicate Device Characteristics. For the non-clinical testing, the ground truth was the established engineering specifications for the device and the performance characteristics of the predicate device (Enpath Medical, Inc. Percutaneous Introducer K000824). The goal was to show the new device met its own specifications and performed equivalently to the predicate.

    8. Sample Size for the Training Set

    • Not Applicable. There was no "training set" as this device is not an AI/ML algorithm. The testing involved physical device samples.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As there was no training set, this question is irrelevant. The "ground truth" for the non-clinical testing (not training) was based on established engineering standards and comparison to a legally marketed predicate device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K072248

    Validate with FDA (Live)

    Date Cleared
    2007-11-09

    (88 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ViaPeel™ Peelable Introducer is intended for use in the percutaneous insertion of pacing leads or catheters in the venous system.

    Device Description

    The ViaPeel Peelable Introducer will be available in 3.5F through 7F (micro), and 7F (macro) in lengths of 5cm or 10cm. The sheath handles are color coded for identification of French size. The locking configuration has been modified on the sheath and dilator to an annular snap fit. The dilator will lock into the hub without the use of a rotating collar allowing the user to lock and secure the dilator to the sheath.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the ViaPeel™ Peelable Introducer, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Implicit)Reported Device Performance
    Functional PerformanceDevice functions per specification requirements."Test results verify that the device performs per specification requirements..."
    Equivalence to PredicateDevice is equivalent to the predicate device without creating additional risk to the patient or user."...equivalent to the predicate device without creating additional risk to the patient or user."
    Mechanical/DimensionalAppropriate dimensions and locking mechanism function (annular snap fit on sheath and dilator, dilator locks into hub)."The performance testing included the device specifications, functional and dimensional testing of the ViaPeel Peelable Introducer..." (Implied successful adherence to these designs)
    Material BiocompatibilityComponent materials are biocompatible."The material in the components used to create the ViaPeel Peelable Introducer has been demonstrated to be biocompatible through biocompatibility testing."

    2. Sample Size Used for the Test Set and Data Provenance

    The summary does not provide details on the specific sample sizes used for the non-clinical performance testing. It also does not specify the data provenance (e.g., country of origin, retrospective or prospective) for any of the non-clinical tests. Since no clinical studies were performed, there is no clinical test set.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Since no clinical studies were performed, there was no ground truth established by experts for a clinical test set. The non-clinical performance testing would have relied on engineering and quality assurance standards, not expert medical consensus on patient outcomes.

    4. Adjudication Method for the Test Set

    As there was no clinical test set involving human judgment, there was no adjudication method described. The non-clinical tests would involve repeatable measurements and objective criteria rather than expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was performed. The summary explicitly states: "No clinical evaluations of this product have been performed." Therefore, there is no information on how much human readers might improve with or without AI assistance, as AI is not part of this device.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No standalone study was performed as this device is a physical medical instrument (introducer) and does not involve an algorithm or AI.

    7. Type of Ground Truth Used

    For the non-clinical performance testing, the "ground truth" was established by engineering specifications, functional requirements, and material biocompatibility standards. These are objective, measurable criteria rather than expert consensus, pathology, or outcomes data in the typical sense for a diagnostic or AI device.

    8. Sample Size for the Training Set

    Since no AI or algorithm is involved, there was no training set for this device. The development process would have involved design, prototyping, and testing of physical units.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there was no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K071574

    Validate with FDA (Live)

    Date Cleared
    2007-07-05

    (27 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stiffer Coaxial Micro-Introducer Kit is indicated for use to introduce up to a 0.038 inch guide wire or catheter into the vascular systems following a small 21 gauge needle stick.

    Device Description

    The Coaxial Micro-Introducer Kit contains a 21 gauge disposable aspiration and injection needle, a 0.018 inch diameter guidewire, and a coaxial micro-introducer set consisting of a dilator and sheath available in 4F, 4.5F, or 5F.

    The modified device contains the same kit configuration as above except that the dilator and sheath are available in 4F or 5F. The inner dilator includes a stainless steel stiffening hypotube. The inner and outer dimensions of the dilator remain the same as the currently marketed Coaxial Micro-Introducer Kit.

    AI/ML Overview

    The Enpath Medical, Inc. Stiffer Coaxial Micro-Introducer Set (K071574) did not provide traditional acceptance criteria with quantitative performance metrics in the provided document. This submission is for a medical device that introduces a guidewire/catheter, not an AI/ML powered diagnostic device, so the typical AI/ML powered diagnostic study information is not applicable.

    Instead, the submission focuses on demonstrating substantial equivalence to a predicate device through performance testing and biocompatibility testing.

    Here's an interpretation based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Implicitly based on predicate device)Reported Device Performance
    Device SpecificationsConforms to intended design and dimensions."Test results verify that the device performs per specification requirements"
    FunctionalityDevice functions as intended for introducing guidewires/catheters."testing to verify that the device functions in a safe and effective manner."
    Dimensional EquivalenceInner/outer dimensions of dilator remain the same as the predicate device."The inner and outer dimensions of the dilator remain the same as the currently marketed Coaxial Micro-Introducer Kit."
    SafetyNo new risks to the patient or user compared to the predicate device."equivalent to the predicate device without creating additional risk to the patient or user."
    EffectivenessPerforms as intended for its indicated use."The testing performed confirms that the Stiffer Coaxial Micro-Introducer Kit will perform as intended."
    BiocompatibilityMaterials used are biocompatible."The material in the components used to create the Stiffer Coaxial Micro-Introducer Kit has been demonstrated to be biocompatible through biocompatibility testing."

    2. Sample Size Used for the Test Set and Data Provenance

    • The document does not specify sample sizes for the performance or biocompatibility testing. These tests are typically conducted on a small number of devices or material samples, as appropriate for non-clinical testing of a physical medical device.
    • Data Provenance: The data is generated from non-clinical testing conducted by the manufacturer, Enpath Medical, Inc., presumably at their facilities. There is no mention of country of origin of data in a geographic sense, as it is laboratory testing. It is inherently prospective in the sense that the new device was manufactured and then tested.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    • This information is not applicable to this type of device submission. Ground truth, in the context of diagnostic devices, usually refers to clinical diagnoses or pathology reports. For this introducer set, the "ground truth" is adherence to engineering specifications and performance standards through physical and material testing. The experts involved would be engineers and material scientists performing the tests and comparing results to established standards or predicate device performance. Their qualifications are not detailed in this summary.

    4. Adjudication Method for the Test Set

    • This is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies, particularly for diagnostic devices, to resolve disagreements among human readers or expert reviewers. For non-clinical performance and biocompatibility testing of a physical device, results are typically objective measurements or observations against predefined criteria, not subject to subjective adjudication in this manner.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation of images or data, often comparing AI-assisted performance against unaided human performance. The Stiffer Coaxial Micro-Introducer Kit is a physical device, and its safety and effectiveness are assessed through non-clinical performance and biocompatibility testing, not through MRMC studies.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • No, a standalone (algorithm only) performance study was not done. This concept is entirely irrelevant for a physical medical device like an introducer set, which does not involve an algorithm or artificial intelligence.

    7. The Type of Ground Truth Used

    • The "ground truth" for this device's evaluation is primarily:
      • Engineering specifications and design requirements: The device's physical dimensions, materials, and functional characteristics are designed to meet specific criteria.
      • Performance standards: The device is tested to ensure it performs its intended function (e.g., smoothly introducing a guidewire/catheter, maintaining integrity under stress).
      • Biocompatibility standards: Materials are tested to ensure they are safe for contact with the body.
      • Predicate device characteristics: The new device's performance, dimensions, and materials are compared to the legally marketed predicate device (Enpath Medical, Inc. Coaxial Micro-Introducer Set K990705) to demonstrate substantial equivalence.

    8. The Sample Size for the Training Set

    • This information is not applicable. "Training set" refers to data used to train an AI/ML algorithm. This device is a physical product and does not involve AI/ML.

    9. How the Ground Truth for the Training Set Was Established

    • This information is not applicable, as there is no AI/ML training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K063182

    Validate with FDA (Live)

    Date Cleared
    2006-12-21

    (63 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Enpath ViaSeal Valved Peelable Introducer is indicated for use in the percutaneous insertion of pacing leads or catheters in the venous system.

    Device Description

    The Enpath ViaSeal Valved Peelable Introducer is a small diameter tubular shaped device with integrated proximal handles. The ViaSeal Valved Peelable Introducer is designed to provide a relatively atraumatic method for implanting catheters and pacemaker leads into the venous system while providing hemostatic sealing to venous pressures. The ViaSeal Valved Peelable Introducer has a "peel-away" feature common to the predicate device. This feature allows the user to remove the introducer without removing the inserted catheter or pacing lead.

    The ViaSeal Valved Peelable Introducer is packaged in a sterile 5-pack convenience kit containing a ViaSeal Valved Peelable Introducer, a thin-wall needle, a disposable syringe, and a flexible guidewire. The kit is packaged in a PETG tray with a Tyvek lid, and a sealed Tyvek/LDPE pouch.

    Introducer sizes of the sheath and dilator range from 7 French to 10.5 French. The materials and construction are the same for all French sizes. Enpath ViaSeal Valved Peelable Introducer kits will be packaged and ETO sterilized for one time use and sealed in a Tyvek/LDPE pouch.

    AI/ML Overview

    The provided text is a 510(k) summary for the Enpath Medical ViaSeal Valved Peelable Introducer. This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed clinical trial data with specific acceptance criteria in the manner typically seen for novel therapeutic devices or diagnostic AI systems.

    Therefore, many of the specific questions about acceptance criteria, detailed study design, and AI model performance metrics cannot be directly answered from this document.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a formal table of specific acceptance criteria with quantifiable metrics such as sensitivity, specificity, accuracy, or a detailed breakdown of device performance against these. Instead, it makes a general statement about performance.

    Acceptance Criteria CategoryReported Device Performance
    Functional EquivalencePerforms per specification requirements and is equivalent to the predicate device.
    SafetyDoes not create additional risk to the patient or user.
    HemostasisPerforms hemostasis to venous pressures.
    DimensionalTested for dimensional specifications.
    RegulatorySubstantially equivalent to predicate devices (FlowGuard Valved Peelable Introducer, K040150, and Thomas Medical Products Inc. SafeSheath MSP Introducer Kit, K003731).

    2. Sample size used for the test set and the data provenance

    • Test set sample size: Not applicable/Not mentioned. The study was primarily non-clinical "performance testing."
    • Data provenance: Not applicable/Not mentioned for clinical data. The performance testing would have been conducted internally by the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This was not a study requiring expert-established ground truth for diagnostic or interpretative tasks. It was performance testing of a medical device.

    4. Adjudication method for the test set

    Not applicable. This was not a study requiring adjudication of expert interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a non-AI medical device (a catheter introducer). No MRMC study was conducted, and no AI component is involved.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a non-AI medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable in the context of diagnostic performance. For a medical device like this, "ground truth" would relate to engineering specifications, material properties, and functional performance benchmarks (e.g., ability to peel, maintain hemostasis, dimensions). These would be established through engineering and quality control standards.

    8. The sample size for the training set

    Not applicable. This is a non-AI medical device; no training set for an algorithm was used.

    9. How the ground truth for the training set was established

    Not applicable. No training set for an algorithm was used.


    Summary of the Study that Proves the Device Meets Acceptance Criteria:

    The "study" refers to a series of non-clinical performance tests conducted by Enpath Medical Inc. to demonstrate the device's safety and effectiveness and its substantial equivalence to predicate devices.

    • Study Title: "Summary of Studies" (as per the 510(k) document)
    • Study Type: Non-clinical (in vitro/bench testing)
    • Objective: To verify that the device functions in a safe and effective manner and performs per specification requirements, equivalent to predicate devices, without creating additional risk.
    • Tests Performed:
      • Device specifications verification
      • Dimensional testing
      • Functional testing
      • Hemostasis testing to venous pressures
    • Results: "Test results verify that the device performs per specification requirements and is equivalent to the predicate device without creating additional risk to the patient or user."
    • Key Finding for Approval: The device was deemed "substantially equivalent" to the Enpath Medical FlowGuard Valved Peelable Introducer (K040150) and the Thomas Medical Products Inc. SafeSheath MSP Introducer Kit (K003731) based on these non-clinical tests.
    • Clinical Evaluations: Explicitly stated: "No clinical evaluations for this submission have been conducted."
    Ask a Question

    Ask a specific question about this device

    K Number
    K061119

    Validate with FDA (Live)

    Date Cleared
    2006-05-18

    (27 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Enpath Medical Steerable Sheath is intended to be used to facilitate the placement of interventional devices into the peripheral and coronary systems.

    Device Description

    The Enpath Steerable Sheath is a guide catheter (delivery sheath) which has a distal tip that deflects to create a variable curve shape. The deflection is controlled through a handle mechanism with the ability to lock at any point along the catheter tip curve travel. The handle incorporates a luer fitting or a hemostasis valve.

    AI/ML Overview

    The provided document is a 510(k) Summary for the Enpath Medical Steerable Sheath. It describes a Special 510(k) Submission, meaning it's for a modification to a previously cleared device (K043489). Therefore, the study focuses on demonstrating substantial equivalence to the predicate device, rather than a de novo evaluation of efficacy against acceptance criteria in the same way a novel device might.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a formal table of explicit acceptance criteria with numerical targets. Instead, it states that "Test results verify that the device performs per specification requirements." The "specification requirements" are not detailed in this summary.

    Acceptance Criteria (Implied)Reported Device Performance
    Device functions safely and effectivelyTest results verify that the device performs per specification requirements.
    Articulation functions correctlyFunctional testing of the articulation was performed and passed.
    Deflection radius meets specificationsDeflection radius testing was performed and passed.
    Deflection force meets specificationsDeflection force testing was performed and passed.
    Hemostasis of valve portion meets specificationsHemostasis of the valve portion testing was performed and passed.
    Deflection cycle endurance meets specificationsDeflection cycle testing was performed and passed.
    Other applicable device specifications are met"Other testing as applicable to this device" was performed and passed.
    Device is equivalent to the predicate device (K043489) without creating additional risk."Test results verify that the device performs per specification requirements and is equivalent to the predicate device without creating additional risk to the patient or user."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not specify the sample size for the test set for any of the performance tests (articulation, deflection radius, deflection force, hemostasis, deflection cycle testing, etc.). It only mentions that "The performance testing for this device included testing..."
    • Data Provenance: Not explicitly stated, but given this is a U.S. FDA submission for a U.S.-based company (Enpath Medical Inc., Minneapolis, MN), it can be inferred that the testing and data likely originated from the U.S. The testing appears to be prospective as it's described as "performance testing" to "verify that the device functions."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This was an engineering performance study for a medical device modification, not a study involving human interpretation of data where expert consensus would establish ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. As noted above, this was an engineering performance study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a Steerable Sheath, a physical medical device, not an AI or imaging diagnostic tool. Therefore, MRMC studies and AI assistance metrics are not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    A standalone performance evaluation of the device itself (not an algorithm) was performed. The document states: "The performance testing for this device included testing to verify that the device functions in a safe and effective manner." This refers to laboratory-based functional and mechanical testing.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance testing would be the established engineering and design specifications for the device's functional and mechanical properties. For example, the specified range for deflection radius, the maximum deflection force, and the number of cycles the sheath must withstand. These specifications are determined during the device's design and development phase to ensure it functions as intended and safely.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI-based device, so there is no concept of a "training set" in the context of this submission. The training set for the manufacturing process (e.g., quality control samples) is not discussed here.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set."