510(k) Data Aggregation

    K Number
    K140347

    Validate with FDA (Live)

    Date Cleared
    2014-08-20

    (190 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ANKYLOS® C/X Implants of 8 mm in length or longer are for single-stage or two-stage surgical procedures and cemented, removable or screw retained restorations. The ANKYLOS® C/X Implants may be used for immediate placement and function on single tooth and/or multiple tooth applications when adequate primary stability is achievable, with appropriate occlusal loading, in order to restore chewing function. Multiple tooth applications may be splinted.

    ANKYLOS® C/X Implants of 6.6 mm in length are for two-stage surgical procedures and cemented, removable or screw retained restorations. The ANKYLOS® C/X Implants may be used for immediate placement on single tooth and/or multiple tooth applications when adequate primary stability is achievable, with appropriate occlusal loading, in order to restore chewing function. Multiple tooth applications may be splinted.

    Device Description

    The ANKYLOS® C/X Implant 6.6 mm represents a line extension of the currently marketed ANKYLOS® C/X Dental Implant System. It is an endosseous dental implant with a length of 6.6 mm and an internal tapered implant-abutment connection. The ANKYLOS® C/X Implant 6.6 mm is machined from Commercially Pure (CP) Grade 2 Titanium (conforming to ASTM F67 -Standard Specification for Unalloyed Titanium, for Surgical Implant Applications). The ANKYLOS C/X 6.6 mm implants are provided in the diameters 3.5 mm, 4.5 mm, and 5.5 mm.

    AI/ML Overview

    This looks like a 510(k) premarket notification for a dental implant system. Based on the provided text, here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The application states that the new device, the ANKYLOS® C/X Implant 6.6 mm, is a line extension of an already cleared device and has largely similar characteristics to its predicate device (K083805 - ANKYLOS® C/X Dental Implant System). The primary acceptance criteria for this type of submission are typically demonstrating substantial equivalence to a legally marketed predicate device, especially in terms of safety and effectiveness.

    Here's a table based on the provided "Similarities and Differences" section and the described non-clinical performance data to outline the criteria and the device's performance against them:

    Acceptance Criteria (based on predicate device)Reported Device Performance (ANKYLOS® C/X Implant 6.6 mm)
    Material Composition: Commercially Pure Titanium (ASTM F67)Commercially Pure (CP) Grade 2 Titanium (conforming to ASTM F67) - Meets
    Implant-Abutment Connection: TaperedTapered - Meets
    Implant Design / Thread: Thread designThread design - Meets
    Delivery: SterileSterile - Meets
    Fatigue Resistance: Equivalent to predicate device (K083805) per ISO 14801Fatigue testing conducted per ISO 14801; results support substantial equivalence. - Meets
    Surface Area & Bone-to-Implant Contact: Calculated by CAD data, comparable to predicate deviceCalculated by CAD data; results support substantial equivalence. - Meets
    Indications for Use: (Specific to 8mm+ implants) Single/two-stage, cemented/removable/screw-retained, immediate placement with adequate primary stability, single/multiple tooth, splintable.For 8mm+ implants: Identical indications as predicate. For 6.6mm implants: Two-stage surgical procedures, cemented/removable/screw-retained, immediate placement with adequate primary stability, single/multiple tooth, splintable. (Slight difference for 6.6mm in being limited to two-stage, but still "substantially equivalent" in context of safety and effectiveness based on provided data). - Meets
    Functional Safety (FMEA): No unacceptable risksFailure Mode, Effects Analysis (FMEA) performed. No unacceptable risks identified. - Meets

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: The document does not specify a numerical sample size for the "test set" in terms of how many implants were used for fatigue testing or CAD analysis. It simply states "Fatigue testing was conducted on the subject ANKYLOS® C/X Implant 6.6 mm in comparison to the predicate device" and "surface area and bone to implant contact have been calculated by CAD data for the ANKYLOS® C/X Implant 6.6 mm and the predicate device, respectively." For a 510(k) submission, especially for a line extension, the focus is often on representative samples demonstrating conformance to standards rather than large statistical cohorts typical of clinical trials.
    • Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. Given the nature of mechanical testing (fatigue, CAD), it's highly likely this was laboratory-based, prospective testing rather than clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This type of submission focuses on non-clinical performance (mechanical testing and CAD analysis). Therefore, no external experts or clinicians were explicitly used to establish ground truth for a "test set" in the way one might for a diagnostic AI device. The ground truth is established by an international standard (ISO 14801) for fatigue testing and engineering principles for CAD analysis, performed by the manufacturer's qualified personnel.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • None. As explained above, this is a non-clinical, engineering-focused evaluation based on established standards and calculations rather than subjective assessment requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This device is a dental implant, which is a physical medical device, not a diagnostic AI system. Therefore, an MRMC comparative effectiveness study involving human readers or AI assistance is not applicable and was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This is a physical dental implant, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The ground truth for the non-clinical performance data (fatigue, surface area, bone-to-implant contact) is derived from engineering standards (ISO 14801) and CAD calculations based on the device's design specifications for the implant and comparison to the predicate. The "ground truth" for the FMEA is based on risk analysis methodologies.

    8. The sample size for the training set

    • Not applicable. This device is a physical dental implant. There is no "training set" in the context of machine learning or AI. The design and manufacturing processes are refined through engineering principles, material science, and testing, not through training data in the AI sense.

    9. How the ground truth for the training set was established

    • Not applicable. As there is no training set for an AI/algorithm, there is no ground truth established for one. The "truth" about the device's functionality is established through its design, material properties, and adherence to performance standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K130862

    Validate with FDA (Live)

    Date Cleared
    2013-07-23

    (117 days)

    Product Code
    Regulation Number
    872.4850
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cavitron® Prophy-Jet® Air Polishing Prophylaxis System is designed for use in general prophylaxis and periodontal treatments and other areas of operative dentistry. It can also be used for:

    • . Stain removal,
    • Prophylaxis of orthodontic patients, .
    • Preparing tooth surfaces prior to bonding and sealant procedures, and ●
    • Implant debridement .
    Device Description

    The Cavitron® Prophy-Jet® Air Polishing Prophylaxis System is an air polishing unit that delivers an air/water/powder mixture at the air polishing insert tip that polishes the tooth enamel without direct contact of the device. The system is used with water-soluble prophy powders such as Cavitron® PROPHY-JET® Sodium Bicarbonate Prophy Powder (K970342) or Cavitron® JET-Fresh® Aluminum Trihydroxide Prophy Powder (K014188) (for patients with low sodium requirements). Updated design features on the Cavitron® Prophy-Jet® Air Polishing Prophylaxis System include a Tap-On™ Wireless Foot Pedal, prophy mode auto cycles, an illuminated diagnostic display, automated purge function, Cavitron® JET-Mate™ detachable sterilizable handpiece (K023697), and a swivel handpiece cable with more precise lavage control. Upon installation, the Cavitron® Prophy-Jet® Air Polishing Prophylaxis System unit is connected to external air and water lines, as well as an AC power cord.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Cavitron® Prophy-Jet® Air Polishing Prophylaxis System, based on the provided text:

    Important Note: The provided document is a 510(k) Summary, which typically focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed, novel clinical trial data. Therefore, many of the requested elements (especially those related to human reader studies, specific ground truth establishment for a test set, and detailed statistical results) are not available in this type of document. The study described here is primarily non-clinical performance testing.

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the compliance with regulations and successful performance in the specified tests, which aim to ensure the device is safe and effective and substantially equivalent to predicate devices. The reported device performance is that it met these criteria.

    Acceptance Criterion (Implied)Reported Device Performance
    Electrical Safety & Electromagnetic Compatibility (EMC)Complies with IEC 60601-1-2 and IEC 60601-1.
    Software ValidationComplies with FDA Guidelines for the Content of Premarket Submissions for Software Contained in Medical Devices.
    Handpiece Operation (including Lavage Control)Met Internal specifications and testing.
    Autocycle OperationMet Internal specifications and testing.
    Indicator Lights & User Interface FunctionalityMet Internal specifications and testing.
    Radio Frequency Distance for Foot PedalMet Internal specifications and testing.
    Stain RemovalMet Internal specification and third-party testing.
    Enamel AbrasionMet Internal specification and third-party testing.
    Overall Function and SafetyPerformance tests support substantial equivalence.

    Study Details

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size (Test Set): Not explicitly stated for each test (e.g., number of samples for stain removal or abrasion tests). The document refers to "internal specification and testing" and "third-party testing" for these aspects. For electrical and software compliance, the "sample" is typically the device itself or its software.
    • Data Provenance: The tests include "Internal specification and testing" and "third-party testing." The country of origin for the data is not specified, but the applicant is DENTSPLY International, located in York, PA, USA. The studies are prospective in the sense that they were conducted specifically for this submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • This information is not provided in the document. For non-clinical performance tests like stain removal and abrasion, ground truth might be established through standardized methods, quantitative measurements, or comparison to established benchmarks, rather than expert consensus on a test set in the way it might be for an imaging AI device.

    4. Adjudication Method for the Test Set:

    • Not applicable/Not provided. This type of detail is typically relevant for human review of data, such as in clinical imaging studies. For the described non-clinical performance tests, adjudication methods are not relevant.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No. The document explicitly states: "No human clinical data has been provided to support substantial equivalence." Therefore, an MRMC study was not performed.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Was Done:

    • Not applicable in the AI sense. The device is a physical instrument, not an AI algorithm. The performance evaluation focuses on the device's physical capabilities (stain removal, abrasion, electrical safety, etc.) and software functionality as an integrated system, which is inherently "standalone" in its operation relative to human decision-making. No "algorithm only" performance would be separately evaluated in this context.

    7. The Type of Ground Truth Used:

    • Standardized Test Methods/Quantitative Measurements/Predicate Device Comparison: For non-clinical tests, ground truth is established by standardized testing protocols (e.g., for electrical safety, EMC), internal specifications, and performance against accepted benchmarks (e.g., stain removal effectiveness compared to industry standards or predicate device performance, enamel abrasion within acceptable limits). The core of this submission is showing substantial equivalence to predicate devices, implying their existing performance provides a "ground truth" for acceptable levels.

    8. The Sample Size for the Training Set:

    • Not applicable. This device is a physical instrument, not an AI/ML algorithm that requires a training set in the conventional sense. The "training" for such a device involves engineering design, prototyping, and iterative testing, not data-driven model training.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. As above, there's no "training set" in the context of an AI algorithm. The design and development process for a physical medical device relies on engineering principles, material science, previous device knowledge (predicates), and internal testing to ensure it meets its intended specifications and safety standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K123615

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-02-22

    (91 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IRIS - High Strength Glass Ceramic is a system of precrystallized and/or fully crystallized zirconia reinforced lithium silicate glass ceramics and veneering ceramics processed through CAD-CAM techniques for the fabrication of: Single unit dental restorations, for example all-ceramic crowns, inlays, onlays and veneers. IRIS can be used as a substructure which is then veneered with IRIS veneering ceramic or can be used for full-contour application (without veneering) as well.

    Device Description

    IRIS - High Strength Glass Ceramic is a system of glass ceramic dental materials based on zirconia reinforced lithium silicate chemistries and feldspathic veneering ceramics and accessories optimized to be compatible with IRIS glass ceramics. The system consists of: A. IRIS Flex: High Strength Glass Ceramic (HSGC) CAD CAM blanks provided in the fully crystallized state, for machining and subsequent placement. B. IRIS CAD: High Strength Glass Ceramic (HSGC) CAD CAM blanks provided in the partially crystallized state, for machining, crystallization firing, and subsequent placement. C. Veneering ceramics provided in powder/paste forms and are intended for optional esthetic characterization of the CAD CAM restorations. IRIS materials may be veneered with a compatible dental veneering ceramic or can be used as anatomically shaped full-contour restorations without veneering.

    AI/ML Overview

    The provided text describes the "IRIS - High Strength Glass Ceramic" device and its substantial equivalence to a predicate device, rather than a clinical study evaluating its performance against specific acceptance criteria. Therefore, several requested sections of the response, such as sample sizes, expert qualifications, and adjudication methods, cannot be directly extracted as they relate to clinical effectiveness studies, which were explicitly stated as "Not applicable. No human clinical testing was conducted to support the substantial equivalence of the IRIS - High Strength Glass Ceramic."

    However, the document does detail material properties testing, which serves as the "study" proving the device meets material-related acceptance criteria.

    Here's the information that can be extracted and a clear indication of what is not applicable based on the provided text:

    Acceptance Criteria and Device Performance

    The acceptance criteria for the IRIS - High Strength Glass Ceramic are primarily based on ISO 6872 standards and comparability to the predicate device, IPS e.max CAD (K051705). The reported performance indicates the device meets these criteria.

    Acceptance Criteria CategorySpecific Criteria (Implicitly ISO 6872)Reported Device PerformanceStudy Proving Acceptance
    Mechanical PropertiesFlexural strength meets ISO 6872.IRIS Flex materials can be raised to >300 MPa through an optional glaze firing process. Measurements showed comparable mechanical (flexural strength) to predicate and met ISO 6872.Physical Properties Testing (Non-Clinical)
    Chemical StabilitySolubility meets ISO 6872.Measurements showed comparable chemical stability (solubility) to predicate and met ISO 6872.Physical Properties Testing (Non-Clinical)
    BiocompatibilityMeets ISO 10993-1:2009.Characterized for cytotoxicity, intracutaneous reactivity/irritation, systemic toxicity, and genotoxicity. Results support biocompatibility.Toxicological Testing (ISO 10993-1:2009)
    Long-term StabilityComparable to predicate device.Verified by simulated life testing (thermal cycling, mechanical loading).Physical Properties Testing (Non-Clinical)

    2. Sample size used for the test set and the data provenance

    Not Applicable for clinical performance. The document explicitly states: "No human clinical testing was conducted to support the substantial equivalence of the IRIS - High Strength Glass Ceramic." The "test set" for the non-clinical material property evaluations would refer to laboratory samples, but specific sample sizes are not provided in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not Applicable. No human clinical testing was performed, so there was no need for experts to establish ground truth in a clinical context. The "ground truth" for material properties is established by adherence to recognized international standards (e.g., ISO 6872, ISO 10993-1).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not Applicable. Adjudication methods are relevant for human interpretation in clinical studies, which were not conducted. Validation for non-clinical testing relies on standard laboratory protocols and accredited testing facilities.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not Applicable. No human clinical testing or multi-reader studies were conducted. This device is a dental material, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not Applicable. This is a dental material, not an algorithm, so the concept of standalone algorithmic performance is irrelevant. The performance is assessed based on physical and chemical properties.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical studies:

    • Mechanical and Chemical Properties: The ground truth is defined by international standards (ISO 6872) for dental ceramics. Comparability to a legally marketed predicate device (IPS e.max CAD) also serves as a ground truth baseline.
    • Biocompatibility: The ground truth is established by the accepted methods and criteria within international standard ISO 10993-1:2009 (Biological evaluation of medical devices – Part 1: Evaluation and Testing).

    8. The sample size for the training set

    Not Applicable. No "training set" in the context of machine learning or clinical trials was involved. If referring to laboratory samples used for initial development and optimization, that information is not provided.

    9. How the ground truth for the training set was established

    Not Applicable. As no "training set" for an algorithm or clinical trial was used, this question is not applicable. The development of the material would follow established scientific and engineering principles for material science, with properties characterized against known material benchmarks.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103244

    Validate with FDA (Live)

    Date Cleared
    2011-02-14

    (104 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    QMix™ 2in1 Endodontic Irrigating Solution is a premixed dual-action device that cleanses and disinfects the root canal system after endodontic instrumentation.

    Device Description

    QMix™ 2in1 Endodontic Irrigating Solution is a premixed dual-action device that cleanses and disinfects the root canal system by removing the smear layer and killing bacteria after endodontic instrumentation.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the QMix™ 2in1 Endodontic Irrigating Solution:

    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the provided text, the device is a "Cleanser, Root Canal" and its indication for use is to "cleanse and disinfect the root canal system after endodontic instrumentation." The core of the submission relies on demonstrating substantial equivalence to a predicate device.

    Acceptance CriterionReported Device Performance
    Safety- All components have been used in legally marketed devices and found safe for dental use.
    - Biocompatibility data provided.
    Effectiveness/Performance- Efficacy demonstrated via non-clinical in vitro and ex vivo studies.
    - Device "cleanses and disinfects the root canal system by removing the smear layer and killing bacteria after endodontic instrumentation," similar to the predicate device.
    Substantial Equivalence- Believed to be substantially equivalent to the cleared and marketed BioPure MTAD root canal cleanser based on prior use of components, performance, and biocompatibility data.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not applicable. No clinical studies were conducted, meaning no test set of patient data was used.
    • Data Provenance: Not applicable. The studies mentioned are non-clinical (in vitro and ex vivo).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not applicable. No clinical studies were performed requiring expert ground truth establishment for a test set.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. No clinical studies with a test set were conducted.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study: No. The document explicitly states: "No clinical studies were conducted on this device." Therefore, no MRMC study comparing human readers with and without AI assistance was performed.
    • Effect Size: Not applicable.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Standalone Study: Yes, in a sense, but not for an "algorithm" in the context of an AI device. The efficacy of the chemical solution itself was demonstrated through non-clinical in vitro and ex vivo studies. These studies evaluate the direct performance of the device (the irrigating solution) without human interpretation of images or data in a diagnostic context.

    7. Type of Ground Truth Used

    • Type of Ground Truth: For the non-clinical in vitro and ex vivo studies, the "ground truth" would be established by scientific methods appropriate for laboratory assays, such as:
      • Microbiological assays: To quantify bacteria killing.
      • Electron microscopy or other imaging techniques: To assess smear layer removal.
      • Chemical analysis: To confirm component presence, stability, and reaction kinetics.

    8. Sample Size for the Training Set

    • Sample Size: Not applicable. This device is a chemical solution, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning. The term "training set" is typically used for data used to train an AI model.

    9. How the Ground Truth for the Training Set Was Established

    • How Ground Truth Was Established: Not applicable, as there is no training set for an AI/ML model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K093221

    Validate with FDA (Live)

    Date Cleared
    2009-10-30

    (17 days)

    Product Code
    Regulation Number
    872.3750
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Pre-Cemented Orthodontic Bracket System Extension II is indicated for use in bonding orthodontic appliances for orthodontic treatment.

    Device Description

    The Pre-Cemented Orthodontic Bracket System Extension II is comprised of precemented ceramic and metal orthodontic brackets, transfer tray and adhesives.

    AI/ML Overview

    The provided text is a 510(k) summary for the "Pre-Cemented Orthodontic Bracket System Extension II". It describes the device, its intended use, and its substantial equivalence to previously cleared devices.

    Based on the information provided, here's an analysis of the acceptance criteria and study information:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Safety and EffectivenessThe submission asserts that the device is safe and effective based on: - Prior use of components in legally marketed devices. - Performance data provided (though specific metrics are not detailed in this summary). - Previous biocompatibility data. - The claim that shear bond strength testing, Failure Mode Effect Analysis (FMEA), and biocompatibility testing were not necessary due to no change in the super mesh base from predicate devices.
    Substantial EquivalenceThe device is deemed substantially equivalent to the predicate devices: - DENTSPLY International, Inc., Pre-Cemented Orthodontic Bracket System (K061252) - DENTSPLY International, Inc., Pre-Cemented Orthodontic Bracket System Extension (K081291)
    Technological CharacteristicsThe device represents an addition to K081291, and all its components have been used in legally marketed devices or found safe for dental use. No change in the super mesh base for metal and ceramic brackets.

    Study Information

    1. Sample size used for the test set and the data provenance:
      The provided 510(k) summary does not detail any specific clinical or performance study with a test set. It relies on the substantial equivalence to predicate devices and the prior use of components. It states that "performance data provided" but does not specify what this data entails or for which tests it was collected. Therefore, information regarding sample size, country of origin, or whether it was retrospective/prospective is not available.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      Not applicable. No specific "test set" requiring ground truth established by experts is described in this summary. The submission is based on substantial equivalence and component history rather than a new performance study requiring expert adjudication of a test set.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
      Not applicable. As no specific "test set" is detailed, no adjudication method is mentioned.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      Not applicable. This device is a physical orthodontic bracket system, not an AI-assisted diagnostic or imaging tool. Therefore, an MRMC study related to human readers and AI assistance is not relevant or mentioned.

    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
      Not applicable. This is a physical medical device, not an algorithm.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
      Not explicitly detailed. The "ground truth" for the device's safety and effectiveness appears to be derived from the established safety and performance of its predicate devices and the historical use of its components. There is no mention of specific new ground truth generation through e.g., pathology or new outcomes data specific to this device extension. The statement "performance data provided" is vague.

    7. The sample size for the training set:
      Not applicable. As this is a physical device and not an AI/machine learning algorithm, there is no concept of a "training set."

    8. How the ground truth for the training set was established:
      Not applicable. As there is no training set mentioned, this question is not relevant.

    In summary, this 510(k) submission primarily leverages the concept of substantial equivalence to previously cleared devices. It does not present new, extensive clinical study data or a detailed performance study with a separate test set, training set, or expert adjudication as would be seen for a novel diagnostic device or AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K083805

    Validate with FDA (Live)

    Date Cleared
    2009-05-22

    (151 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ANKYLOS® C/X Dental Implant System is for single-stage or two-stage surgical procedures and cemented or screw retained restorations. The ANKYLOS® C/X Dental Implant System is intended for immediate placement and function on single tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function. Multiple tooth applications may be splinted with a bar.

    Device Description

    The new ANK YLOS® C/X Dental Implant System introduces a new configuration of the current ANKYLOS® plus dental implant line. The new line extension includes the: ANKYLOS® C/X Implant, ANK YLOS C/ Regular Abutments, ANKYLOS /X Regular Abutments, ANKYLOS® Regular C/X Gingiva Former, and the ANK YLOS® Titanium Alloy Screws (Membrane Screw and Fixation Screw). The ANKYLOS® C/X Dental Implant was modified to integrate an indexation into the taper connection to provide a precise position of the abutment. With the new index, the relocation of the abutment is possible without using a transfer-kev for the impressions taking.

    AI/ML Overview

    The provided text describes a 510(k) summary for the ANKYLOS® C/X Dental Implant System, focusing on its substantial equivalence to predicate devices. However, the document does not contain the detailed study information typically associated with acceptance criteria, sample sizes, ground truth establishment, or expert evaluations as requested in the prompt.

    The document states:

    • "The results regarding the fatigue tests of the ANKYLOS® C/X Implant and the predicate devices were comparable. Therefore, it concluded that the ANKYLOS® C/X Implant performs as intended."

    This is the only mention of performance data. It refers to a "fatigue test" and a "comparable" result. This does not provide specific acceptance criteria, numerical performance metrics, or details about the study design beyond stating "fatigue tests."

    Therefore, I cannot populate the table or answer most of the requested questions based on the provided text.

    Here's a breakdown of what can and cannot be answered from the provided text:

    1. Table of acceptance criteria and reported device performance:

    Acceptance CriteriaReported Device Performance
    Not specified"comparable" to predicate devices in "fatigue tests" (specific metrics not provided)

    2. Sample size used for the test set and data provenance:

    • Cannot be determined from the provided text. The text only mentions "fatigue tests" without any details on the number of devices tested or the origin of the data.

    3. Number of experts used to establish the ground truth for the test set and their qualifications:

    • Cannot be determined from the provided text. This type of information is not present. The evaluation focuses on non-clinical performance (fatigue tests), not expert interpretation.

    4. Adjudication method for the test set:

    • Cannot be determined from the provided text. This is not relevant for the type of non-clinical testing described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and its effect size:

    • No. An MRMC study was not done. The document describes a non-clinical performance evaluation (fatigue tests) of a dental implant system.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • N/A. This question is typically relevant for AI/software devices. The device is a physical dental implant system, so "standalone" performance in this context would refer to its physical function, which was assessed via "fatigue tests."

    7. The type of ground truth used:

    • For the "fatigue tests," the "ground truth" would be the engineering standards for fatigue life or fracture resistance of dental implants. However, the specific standards or methods for establishing this "ground truth" are not detailed beyond the mention of "fatigue tests."

    8. The sample size for the training set:

    • N/A. This question is typically relevant for AI/machine learning models. The device is a physical dental implant system and would not have a "training set" in this context.

    9. How the ground truth for the training set was established:

    • N/A. (See point 8)
    Ask a Question

    Ask a specific question about this device

    K Number
    K082198

    Validate with FDA (Live)

    Device Name
    DURASHIELD PLUS
    Date Cleared
    2008-10-29

    (86 days)

    Product Code
    Regulation Number
    872.3260
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DURASHIELD® PLUS is indicated for the relief of dental hypersensitivity where dentin and cementum are exposed.

    Device Description

    DURASHIELD® PLUS is a topically applied, flavored cavity varnish containing sodium fluoride and NovaMin. The varnish is a viscous liquid and is insoluble in water and saliva. NovaMin is a bio-available calcium phosphosilicate that releases calcium and phosphorous when exposed to moisture.

    AI/ML Overview

    The provided text (K082198) is a 510(k) premarket notification for a dental product called DURASHIELD® PLUS. This type of regulatory submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving efficacy through clinical trials with specified acceptance criteria and performance data.

    Therefore, the document does not contain the information requested in the prompt regarding acceptance criteria, study details, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, or training set details.

    The submission primarily discusses:

    • Device Name: DURASHIELD® PLUS
    • Predicate Device: Sci-Pharm DFV Varnish
    • Description: A topically applied, flavored cavity varnish containing sodium fluoride and NovaMin.
    • Intended Use: Relief of dental hypersensitivity where dentin and cementum are exposed.
    • Technological Characteristics: Similar to predicate device, with the addition of NovaMin for remineralization.
    • Non-Clinical Performance Data: Components of DURASHIELD® PLUS passed biocompatibility testing; no additional biocompatibility tests were conducted for the final product.
    • Conclusion: Substantially equivalent to the predicate device.

    For medical devices, particularly those seeking 510(k) clearance, the focus is on demonstrating that the new device is as safe and effective as a legally marketed predicate device, not necessarily on meeting specific performance metrics from a de novo study. The FDA's letter confirms substantial equivalence, allowing the device to be marketed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072730

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-03-27

    (183 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FRIADENT® CERCON® Abutment is indicated for the support of exceptionally esthetic, single crowns in the anterior region of the maxilla or mandible region. Also ideal for XiVE®, FRIALIT® or FRIALIT®-2 cases involving very thin soft tissue.

    The ANK YLOS® CERCON® Abutment is indicated for the support of highly esthetic single tooth crowns in the anterior region of the maxilla and mandible, including where the mucosa is very thin, in combination with implants of the ANKYLOS® implant system.

    Device Description

    The CERCON® ABUTMENTS are part of the FRIADENT® prosthetics.

    The CERCON® ABUTMENTS are available as straight and angled abutments with different gingival margins and in multiple diameters and shades.

    AI/ML Overview

    This document does not contain the information required to answer your request. The provided text is a 510(k) summary for dental abutments, which outlines the device description, intended use, and substantial equivalence to predicate devices, but does not describe any studies with acceptance criteria or performance metrics in the way you've outlined.

    This submission is for a medical device (Cercon® Abutments) and, based on the provided text, the FDA classified it as substantially equivalent to existing predicate devices. This typically means that the device's technological characteristics and intended use are similar enough to already-approved devices that extensive new clinical studies with defined acceptance criteria and performance reporting might not have been necessary for the 510(k) clearance process itself. The summary states that "We believe that the prior use of the components of the CERCON® ABUTMENTS in legally marketed devices, the provided, and the biocompatibility history support the safety and effectiveness of the CERCON® ABUTMENTS for the indicated uses." This suggests reliance on existing data for similar components, rather than new performance studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072989

    Validate with FDA (Live)

    Device Name
    AXIS SYSTEM
    Date Cleared
    2008-01-28

    (97 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Axis System is an air-powered dental handpiece system indicated for use in general dentistry.

    Device Description

    The Axis System consists of air-powered dental handpieces with an electronic accessory to limit the speed.

    AI/ML Overview

    This 510(k) summary for the Axis System, a dental handpiece, does not contain the detailed information necessary to answer your request about acceptance criteria and study results.

    Here's why:

    • Device Type: The Axis System is a dental handpiece, which is a mechanical device. The concept of "acceptance criteria" and "device performance" in the context of an AI/software device (which your questions are geared towards, e.g., sensitivity, specificity, ground truth, MRMC studies) is not directly applicable to this type of medical device.
    • Information Provided: The summary focuses on established safety and effectiveness through the prior use of its components and existing predicate devices (Rapidd Highspeed Dental Handpiece and eStylus Electric Motor System). It states, "All of the components found in the Axis System have been used in legally marketed devices and/or were found safe for dental use." It also mentions "performance data provided" and "biocompatibility information provided," but the specifics of what that performance data entails or what the acceptance criteria were for a mechanical device are not detailed within this summary.
    • Regulatory Pathway: This is a Class I device (Dental Handpiece and Accessories), which generally has a less stringent regulatory pathway compared to devices that diagnose or interpret medical images using AI. The 510(k) process for such a device primarily establishes substantial equivalence to a predicate device, focusing on similar indications for use and technological characteristics rather than requiring extensive clinical performance studies to meet AI-specific metrics.

    Therefore, I cannot provide the requested table or elaborations on AI-specific study details because the provided document does not contain information relevant to an AI/software-as-a-medical-device (SaMD) product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K073117

    Validate with FDA (Live)

    Date Cleared
    2007-12-19

    (44 days)

    Product Code
    Regulation Number
    872.3260
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Treating and preventing dentinal hypersensitivity. Can be used under direct restorations, under indirect restorations, and on cervical sensitivity.

    Device Description

    The DENTIN DESENSITIZER AGENT is a chair-side agent for treating and preventing dentinal hypersensitivity.

    AI/ML Overview

    The provided K073117 document is a 510(k) summary for a "DENTIN DESENSITIZER AGENT." This document primarily focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and biocompatibility.

    It does not contain information about acceptance criteria or a study that rigorously proves the device meets specific performance criteria using a test set, ground truth, or expert review in the way that would typically be described for a diagnostic AI or imaging device.

    Therefore, I cannot populate the requested table and detailed study information as the document does not provide this type of data.

    Here's why and what information is present:

    • Type of Device: The DENTIN DESENSITIZER AGENT is a chemical agent for treating tooth sensitivity, not a software algorithm, an AI device, or an imaging modality.
    • Approval Pathway: This is a 510(k) submission, which means the device is being approved based on "substantial equivalence" to legally marketed predicate devices, not necessarily on demonstrating new performance metrics against a defined acceptance criterion via a clinical study with a control group and robust statistical analysis for novel claims.
    • Focus of the Submission: The submission highlights:
      • Components: All components have been used in legally marketed devices or found safe for dental use.
      • Biocompatibility Testing: The device passed agar overlay cytotoxicity, intracutaneous irritation, and systemic toxicity tests. This demonstrates safety, but not clinical effectiveness in terms of reducing sensitivity in a quantifiable manner against specific criteria.
      • Predicate Devices: The device is compared to Gluma Desensitizer (K962812) and Quell Desensitizer (K010957). The implication is that since these predicates are accepted for treating dentinal hypersensitivity, and the new device shares similar characteristics and has passed safety tests, it is also safe and effective for the same indication.

    Conclusion:

    Based on the provided text, there is no information to fill in the requested table and answer questions 1-9 regarding acceptance criteria, device performance, sample sizes for test/training sets, data provenance, expert qualifications, ground truth, adjudication methods, or MRMC studies. These details are typically found in submissions for devices that make novel performance claims or rely on AI/software for diagnostic or interpretive functions, which is not the case for this chemical desensitizer.