510(k) Data Aggregation

    K Number
    K193496

    Validate with FDA (Live)

    Device Name
    CharmFil
    Manufacturer
    Date Cleared
    2020-05-06

    (141 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CharmFil is indicated for following restorative applications.

    1. Class I, II, V restorations of posterior teeth
    2. Class III, IV, V restorations of anterior teeth
    3. Cervical cavities or defects involving root surfaces
    Device Description

    This device is light curing dental for dental surgery treatment and type 1, class 2, group 1 according to ISO 4049 classification. It is made in the form of paste and filled in syringe. When you use the product, press out the content in the syringe by screw. It is used to restore a cavity as taking out from it as much as you need with an instrument. It is dental filling material for restoring the cavities.

    AI/ML Overview

    The acceptance criteria and the study proving the device meets them are outlined below based on the provided text for the CharmFil device.

    1. Table of Acceptance Criteria and Reported Device Performance

    No.Test ItemAcceptance Criteria (Requirements)Reported Device Performance (Result)
    Physical and chemical property (Polymerization time) / ISO 4049
    1ColorThe exposed and unexposed halves of each of the specimens and the unirradiated specimen for any color differenceSuitable
    2Appearance testShall be homogeneous and co substances are observedSuitable
    3Quantity testNo less than ±5% that claimed valueSuitable
    4Sensitivity to ambient lightAfter exposure for 60sec, it should be not any charge of consistencySuitable
    5Polymerization depthAfter exposure for 1 minute, Opaque shade is more than 1mm, and other shades is over 1.5mmSuitable
    6Flexural strength+80 MPaSuitable
    7Water solubility- 40 µg/mm²Suitable
    8Water sorption-7.5 µg/mm²Suitable
    9Radio-opacityShall be equal to or greater than that of the same thickness of aluminum and no less than 0.5mmSuitable
    10Color stabilityThe exposed and unexposed halves of each of the specimens and the unirradiated specimen for any color differenceSuitable
    Elution substance test
    11Character testIt should be homogeneous and appropriate. Without any foreign substance, impurity, floating substance.Suitable
    12pHpH difference ≤ 1.5Suitable
    13Heavy metal (Lead content)It should be less ticker than comparison liquid. (-100ppm)Suitable
    14Permanganic acid calcium reduction nature substanceDifference of consumption of permanganic acid calcium≤ 2.0 mlSuitable
    15Evaporation residueDifference of residues ≤ 1.0 mgSuitable
    16Ultraviolet rays absorption spectrumMax. absorbance numerical value ≤ 0.1Suitable
    Biological safety test (Biocompatibility, irritation)
    17Cell toxicity test01 (NoneMild)Suitable
    18Short period whole body toxicity test (Oral route)During the period of examination, there should not be odd or dead on it.Suitable
    19Oral route impetus testCompare it with contrast product, stimulus response should not appear on mouth mucousSuitable
    20Endo-toxin testAccording to the method of examination, it should be appropriate.Suitable

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes for each performance test. The nature of the tests (physical and chemical properties, elution substance, biological safety) suggests that material samples of the CharmFil device were tested in a laboratory setting. The data provenance is not specified beyond indicating that Dentkist, Inc. is located in Gunpo-si, Gyeonggi-do, Korea, implying that the testing likely occurred in South Korea. The study appears to be retrospective, as it's a submission for premarket notification based on completed tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This device is a dental restorative material, and the acceptance criteria are based on objective physical, chemical, and biological safety standards (ISO 4049). There is no mention of expert readers establishing ground truth in the context of diagnostic interpretation.

    4. Adjudication Method for the Test Set

    Not applicable, as the acceptance criteria are objective and assessed against established international standards (ISO 4049) and biological safety guidelines. There is no qualitative assessment that would require adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. This type of study is typically used for diagnostic devices that involve human interpretation of images or other data. The CharmFil device is a dental restorative material, and its performance is evaluated based on material properties and biological safety.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in essence. The performance testing described is "standalone" in that it assesses the intrinsic properties and safety of the material itself, without direct human interaction as part of the performance evaluation (beyond conducting the tests).

    7. The Type of Ground Truth Used

    The "ground truth" for the performance tests consists of established industry standards (e.g., ISO 4049 for physical and chemical properties) and biological safety criteria. These are objective, quantitative or qualitative thresholds that the device's characteristics must meet.

    8. The Sample Size for the Training Set

    Not applicable. As this device is a dental restorative material and its performance is assessed against standardized material properties and biological safety, there is no "training set" in the context of machine learning or diagnostic algorithm development.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152766

    Validate with FDA (Live)

    Device Name
    CharmFlex
    Manufacturer
    Date Cleared
    2016-03-08

    (166 days)

    Product Code
    Regulation Number
    872.3660
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • Impression material in a dual phase impression technique
    • Precise duplication of models
    • Capturing multiple unit impressions
    Device Description

    CharmFlex is impression material. It has 5 types: Putty, Regular, Light, Heavy, Bite.

    1. CharmFlex®Putty/ CharmFlex®Putty Green/ CharmFlex®Putty Soft CharmFlex®Putty consists of various Putty-bodies type and dental addition silicone impression material made by polymerization as a mixture of Base and Catalyst. In dental treatment, an impression material used to record oral tissue anatomy. CharmFlex®Putty has viscosity of putty consistency classification according to type 0 ISO 4823 classification. Packaging consists of two putty systems which are supplied by in two jars, the one is base and another is catalyst. This product is used for two step impression and the material for the first impression is for recording oral tissue anatomy. Base and catalyst of impression are packed separatelv
    2. CharmFlex®Heavy cartridge type/ CharmFlex®Heavy tube type CharmFlex® Heavy series are hydrophilic vinyl polysiloxane impression materials type of Heavybodies. It is used for the one-step impressions teeth or individual tooth in the mouth. CharmFlex®Heavy as the additional polymerization silicone type is rubber impression materials that registers oral tissue anatomy. It is used with Light-body for impression taking. Packages are cartridge type which is supplied by from of two syringes. This product is used for the first or second impression taking of whole teeth or individual tooth in mouth before the final impression taking. It is silicone based dental rubber impression materials.
    3. CharmFlex®Regular/ CharmFlex®Denture/ CharmFlex®Light LV/ CharmFlex®Light XLV/ CharmFlex®Light Premium CharmFlex® Light series are hydrophilic vinyl polysiloxane impression material type of LightLV, Regular and LightXLV. It is used for the one-step or two-step impression taking of teeth or individual tooth in the mouth. CharmFlex® Light XLV/Regular series as the additional polymerization silicone type are rubber impression materials that records oral tissue anatomy. Packages are cartridge type which is supplied by from of two syringes. This product is used for the first or second impression taking of teeth or individual tooth in mouth before the final impression taking. It is silicone based dental rubber impression materials.
    4. CharmFlex®Bite/ CharmFlex®Bite Clear/ CharmFlex®Bite Fast CharmFlex Bite series are bite registration impression material to measure of the occlusal surface. impression of the teeth for a three-dimension position of the mandible in relation to the maxilla. It has a short polymerization time and a high final hardness. Making it suitable for bite registration.
    AI/ML Overview

    This document is a 510(k) Summary for the medical device "CharmFlex®", a dental impression material. It seeks to prove substantial equivalence to existing predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document presents performance data for three categories of CharmFlex® products against predicate devices. The acceptance criteria aren't explicitly stated as numerical targets in a separate section, but are implicitly defined by performing the tests against established ISO standards (ISO 4823 for Dentistry-Elastomeric impression materials) and comparing the results to those of legally marketed predicate devices. The "reported device performance" are the values listed for the "Subject Device" in the comparison tables.

    Here's a consolidated table, noting that specific numerical acceptance criteria (e.g., "must be < 5%") are not explicitly given as such, but rather the performance is demonstrated to be comparable to or within the range of the predicate device.

    Test ItemAcceptance Criteria (Implied by Predicate/Standard)CharmFlex® Heavy/Regular/Light Performance (Subject Device)CharmFlex® Putty Performance (Subject Device)CharmFlex® Bite Performance (Subject Device)
    Bench Tests
    AppearanceComplies with general aesthetic and structural integrity expectations for impression materials (not explicitly quantified but passed).PassPassPass
    WeightComplies with general weight expectations for impression materials (not explicitly quantified but passed).PassPassPass
    ComponentComplies with general component specifications for impression materials (not explicitly quantified but passed).PassPassPass
    Working timeISO 4823/Comparable to predicate (e.g., 1'30" - 2'30" for Light/Medium/Heavy, 1.30min - 2min for Putty, 30-45sec for Bite)Heavy cartridge: 1'30"; Heavy tube: 2'; Regular/Denture/Light: 2'~2'30"1'~1' 30" (26 °C)20~35sec (23 °C)
    Minimum strengthISO 4823/Comparable to predicate (e.g., ~50-65 for Light/Medium/Heavy, ~62-75 for Putty, ~88-89 for Bite)55~65; Heavy (62±2)65~8589±2
    HardnessISO 4823/Comparable to predicateNot explicitly listed, but implied by Mechanical StrengthNot explicitly listed, but implied by Mechanical StrengthNot explicitly listed, but implied by Mechanical Strength
    Linear dimensional changeISO 4823/Comparable to predicate (e.g., 22-30 μm for impression materials, 1.5-2 μm for bite registration)20 μm30 μm1.5 μm
    ConsistencyISO 4823/Comparable to predicate (e.g., 33-48mm for impression materials, 26-30mm for putty/bite)48mm; Heavy (34.25mm)25.83mm32mm
    Detail reproductionISO 4823/No specific numerical value provided, but presumably meets the standard for impression materials.PassPassPass
    Compatibility with gypsumISO 4823/Comparable to predicate (e.g., 20-30 μm)20 μm30 μm1.5 μm
    Elastic recoveryISO 4823/No specific numerical value provided, but presumably meets the standard for impression materials.PassPassPass
    Strain-in-compressionISO 4823/Comparable to predicate (e.g., 4.0-5.0% for impression materials, 3.0-4.0% for putty, 1.5-2.5% for bite)4.0~5.0%3.0~4.0%1.5~2.5%
    Flow propertiesComparable to predicate (e.g., Low/High viscosity for impression, High viscosity for putty/bite)Low viscosity; Heavy (high viscosity)High viscosityHigh viscosity
    ViscosityComparable to predicate (e.g., 30-48mm for impression, 25-32mm for putty/bite)48mm; Heavy (34.25mm)25.83mm32mm
    WettabilityComparable to predicate (e.g., High/Low)High; Heavy (Low)LowLow
    Setting timeComparable to predicate (e.g., 2'-4' for impression, 2.22'-4' for putty, 30-60sec for bite)2'~4'2.22'~4'30~60sec
    Working humidityComparable to predicate (50%±10)50±10%50%±1050%±10
    SafetyDemonstrated to be safe through biocompatibility testing.SafeSafeSafe
    Keeping qualitiesCooler and dry storage conditions comparable to predicate (18-24 °C/64-75 °F)Cool and dry place (1824 °C/6475 °F)Cool and dry place (1824 °C/6475 °F)Cool and dry place (1824 °C/6475 °F)
    Curve of the shrinkageNo data for both subject and predicate devices, so equivalence is claimed on lack of difference. (Not an acceptance criterion per se, but a reported comparison).No dataNo dataBrittle feature

    2. Sample Size Used for the Test Set and Data Provenance:

    The document describes bench testing for performance data and biocompatibility testing. However, it does not specify the sample size used for any of these tests.

    The data provenance is from Dentkist, Inc., which is located in Gunpo-si, Gyeonggi-do, Korea. The study described is a retrospective analysis of the device's performance against established standards and predicate devices (as implied by a 510(k) submission primarily comparing new device data to existing ones). There is no mention of prospective clinical studies with human participants that would generate new patient data.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. For bench test data, ground truth is typically established by the test methods and instruments specified in the ISO standards, not by expert consensus in the typical sense of clinical image interpretation.


    4. Adjudication Method for the Test Set:

    This information is not applicable as the document describes bench testing and biocompatibility assessments, not a study requiring adjudication of expert interpretations of data (e.g., medical images).


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. The device is an impression material, and the evaluation is based on physical and chemical properties and biocompatibility, not on human interpretation of data (e.g., diagnostic images).


    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    Not applicable. This device is not an algorithm or AI software. It is a physical dental impression material. The described tests are "standalone" in the sense that they are independent evaluations of the material's properties according to established standards.


    7. Type of Ground Truth Used:

    The "ground truth" for the performance data (e.g., mechanical strength, dimensional accuracy) is based on International Standards (ISO 4823) and potentially ASTM 1980 for shelf-life, as well as previous clearances (predicate devices) for comparison. For biocompatibility, the ground truth is established by the methods outlined in ISO 10993-5 and ISO 10993-10. In essence, the ground truth is defined by established scientific and regulatory standards and the performance of legally marketed equivalent devices.


    8. Sample Size for the Training Set:

    Not applicable. This is not an AI/ML device that requires a training set. The "training set" concept does not apply to the evaluation of this physical medical device.


    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this type of device.