Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K223292

    Validate with FDA (Live)

    Device Name
    TempSure System
    Manufacturer
    Date Cleared
    2022-11-23

    (28 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Small 10mm, 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids.

    The Small 10mm, 18mm, 25mm, and 60mm handpieces and the FlexSure™ Applicators provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

    The massage device is intended to provide a temporary reduction in the appearance of cellulite

    The following surgical modes are applicable to the generator:

    Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed.

    Cutting: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, tumors, cosmetic repairs, development of skin flaps, skin tags and blepharoplasty.

    Blended Cutting and Coagulation: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids, keratosis, verrucae, basal cell carcinoma, nevi, fistulas, epithelioma, cosmetic repairs, cysts, abscesses, and development of skin flaps.

    Fulguration: basal cell carcinoma, papilloma, cyst destruction, tumors, verrucae, hemostasis.

    Bipolar: pinpoint precise coagulation, pinpoint hemostasis in any field (wet or dry), snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment and turbinate shrinkage.

    Device Description

    The Cynosure TempSure™ System is a radiofrequency generator with a variety of applications both aesthetic and surgical procedures. The intended action is achieved through application of radiofrequency energy to the patient which results in minimization of heat dissipation and cellular alteration. Output of energy is controlled via the guided user interface (GUI) and the foot and/or hand-switch.
    There have been no changes to the TempSure Device or handpieces (Small 10, 10, 15, 18, 20, 25, 30 or 60mm), FlexSure Applicators, or Surgical modes (COAG, Cut, Blend, Bipolar, Fulgurate) of the system. The TempSure system is still used with the same existing electrosurgical accessories.

    The TempSure™ System includes:
    • TempSure™ Generator
    • Temperature Sensing Handpieces (Small 10, 10, 15, 18, 20, 25, 30, and 60mm)
    • Massage Heads (25, 30, and 60mm)
    • FlexSure™ Applicators (Large and Medium)
    • IEC Power Cord
    • Footswitch
    • Disposable/Reusable Neutral Pads
    • Surgical Fingerswitch/Foot Controlled Handpieces
    • Monopolar Cables
    • Disposable/Reusable Electrodes, Forceps

    AI/ML Overview

    The provided document (K223292) is a 510(k) premarket notification for the Cynosure TempSure System. This type of submission asserts substantial equivalence to a legally marketed predicate device, rather than providing new clinical data to establish safety and effectiveness from scratch. As such, the document explicitly states:

    • "807.92(b)(1) Non-clinical tests submitted– N/A – No non-clinical tests submitted"
    • "807.92(b)(2) Clinical tests submitted – N/A – No clinical tests submitted"
    • "807.92(b)(3) Conclusions drawn from clinical and non-clinical tests submitted– N/A"

    Therefore, this document does not contain information about acceptance criteria or a study proving the device meets those criteria, because no new non-clinical or clinical studies were submitted for this 510(k) notification.

    The submission establishes substantial equivalence by demonstrating that the device has the same intended use and similar technological characteristics as its predicate device (Cynosure TempSure System, K212891), with no changes to the device or handpieces, only an update to labeling with an additional contraindication.

    Because no studies were submitted, the following requested information cannot be extracted from this document:

    1. A table of acceptance criteria and the reported device performance
    2. Sample size used for the test set and the data provenance
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    4. Adjudication method for the test set
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, and its effect size
    6. If a standalone performance (algorithm only) was done
    7. The type of ground truth used
    8. The sample size for the training set
    9. How the ground truth for the training set was established
    Ask a Question

    Ask a specific question about this device

    K Number
    K210226

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-03-30

    (61 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    755mm:
    The PicoSure Workstation is indicated for tattoo and benign pigmented lesions removal including but not limited to: Nevus of Ota, Hori macules (nevus of Hori), and Melasma. The PicoSure Workstation with the 2mm and 6mm hand pieces and the Focus Array are indicated for the treatment of acne scars and wrinkles in Skin Types I – IV.

    532nm:
    The PicoSure 532-nm delivery system is indicated for tattoo removal and benign pigmented lesions removal in Skin Types I-III.

    1064nm:
    The PicoSure 1064-nm delivery system is indicated for tattoo and benign pigmented lesions removal.

    Device Description

    The PicoSure Workstation is a high-powered, Alexandrite system that delivers laser energy in the 755-nm nominal wavelength. The system offers treatment through a variety of spot sizes, fluences and repetition rates. Laser activation is by footswitch. In addition to the 755nm handpiece, optional 532nm Laser Delivery System and/or 1064nm Laser Delivery System can replace the 755nm handpiece at the distal end of the articulated arm. These Delivery Systems convert the 755nm laser energy into a 532nm wavelength or a 1064 nm wavelength and are available in multiple spot sizes.

    AI/ML Overview

    The provided document is a 510(k) summary for the PicoSure Workstation, a laser surgical instrument. It details the device's indications for use and compares it to predicate devices. However, it does not contain the specific information required to complete a table of acceptance criteria and reported device performance directly in the format you requested, especially regarding quantitative performance metrics, sample sizes for test sets, and details about expert involvement in establishing ground truth.

    The document indicates that to support the expanded indications for use (Nevus of Ota, Hori macules, and Melasma), a literature search was conducted. This means the clinical evidence is based on existing published studies rather than a new, dedicated clinical trial performed for this 510(k) submission.

    Here's an attempt to answer your questions based on the provided text, with clear indications where information is not available from this document:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative acceptance criteria (e.g., minimum sensitivity, specificity, or specific improvement percentages) directly associated with this 510(k) submission. Instead, it relies on demonstrating substantial equivalence to a predicate device and supporting new indications through literature review. The reported "performance" is qualitative, focusing on whether the existing literature showed "acceptable clinical effectiveness and safety" for the new indications.

    Criteria (Hypothesized based on text)Reported Device Performance (from literature review)
    Clinical EffectivenessSix identified articles reported acceptable clinical effectiveness for Nevus of Ota, Hori macules (nevus of Hori), and Melasma using the PicoSure at 755nm.
    Safety ProfileSix identified articles reported acceptable safety. No serious adverse events were reported. Side effects were primarily transient and included common acceptable side effects for laser treatments as described in the product labeling.
    Substantial EquivalenceThe device is deemed substantially equivalent to the PicoSure Workstation (K173199) and RevLite Q-Switched Nd:YAG Laser System (K133254) which were legally marketed predicate devices, with no changes to the technological characteristics, only an expansion of indications based on clinical literature.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample size: Not explicitly stated as a single test set size. The document mentions "at least ten individuals" in each of the six identified articles. Thus, the total number of individuals across all six studies is at least 60. Individual study sizes are not detailed.
    • Data provenance: The document states "peer reviewed, published articles." No specific countries of origin are mentioned.
    • Retrospective or prospective: The articles were identified as "prospective studies."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Not Applicable / Not Available. Since the evidence is based on published clinical studies, the "ground truth" and its establishment would be inherent to the methodologies of those individual studies. The 510(k) summary does not provide details on how ground truth was established within these external studies or the qualifications of any experts involved in those studies.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not Applicable / Not Available. This information would pertain to the specific methodologies of the clinical studies identified in the literature search. The 510(k) summary does not provide these details.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This document is for a laser workstation, not an AI-assisted diagnostic device. Therefore, no MRMC study or AI assistance effect size is relevant or mentioned.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a hardware device (laser workstation), not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not Applicable / Not Available. The "ground truth" would relate to the diagnosis of the benign pigmented lesions (Nevus of Ota, Hori macules, Melasma) and the success of their removal or clinical improvement. The specific method used to establish these in the surveyed literature is not detailed in this 510(k) summary. It likely involved clinical diagnosis by dermatologists and visual assessment of treatment outcomes.

    8. The sample size for the training set

    • Not Applicable. There is no "training set" in the context of this device. The device's operation is based on its physical properties and established laser physics, not on a machine learning model trained on data. The clinical literature reviewed supported the indications for use, not a training process for the device itself.

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200241

    Validate with FDA (Live)

    Device Name
    TempSure System
    Manufacturer
    Date Cleared
    2020-03-25

    (54 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 10mm, 15mm, and 20mm Smart Handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids.

    The 18mm, 25mm, 30mm, 60mm Smart Handpieces and FlexSure™ applicators provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

    The massage device is intended to provide a temporary reduction in the appearance of cellulite

    The following surgical modes are applicable to the generator.

    Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed

    Cutting: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage. skin incisions, biopsy, cysts, abscesses. tumors. cosmetic repairs, development of' skin flaps. skin tags and blepharoplasty.

    Blended Cutting and Coagulation: submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids. keratosis, verrucae, basal cell carcinoma. nevi. fistulas. epithelidma. cosmetic repairs. cysts. abscesses. and development of skin flaps.

    Fulguration: basal cell carcinoma. papilloma. cyst destruction, tumors. verrucae, hemostasis.

    Bipolar: pinpoint precise coagulation. pinpoint hemostasis in any, field (wet or dry). snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control. epistaxis treatment and turbinate shrinkage

    Device Description

    The Cynosure TempSure™ System is a radiofrequency generator with a variety of applications both aesthetic and surgical procedures. The intended action is achieved through application of radiofrequency energy to the patient which results in minimization of heat dissipation and cellular alteration. Output of energy is controlled via the guided user interface (GUI) and the foot and/or hand-switch.

    The FlexSure TM Applicators are now included for the purpose of tissue heating for selected medical conditions. The applicator comes in 2 sizes (large and medium) and has the same maximum output power of 300W in the Smart Handpiece Mode. The FlexSure applicator utilizes the larger disposable neutral pad. There have been no changes to the Smart Handpiece (10, 15, 18, 20, 25, 30 or 60mm) or Surgical modes (COAG, Cut, Blend, Bipolar, Fulgurate) of the system. The TempSure system is still used with the same existing electrosurgical accessories.

    The TempSure™ System includes:
    TempSure™ Generator
    Temperature Sensing Handpieces (10, 15, 18, 20, 25, 30, and 60mm)
    Temperature Sensing single use applicator (FlexSure)
    Massage Heads (25, 30, and 60mm)
    IEC Power Cord
    Footswitch
    Disposable/Reusable Neutral Pads
    Surgical Fingerswitch/Foot Controlled Handpieces
    Monopolar Cables
    Disposable/Reusable Electrodes, Forceps

    New Additions to the Tempsure System include:
    • FlexSure™ Large Applicator
    • FlexSure™ Medium Applicator

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Cynosure TempSure™ System. This submission focuses on establishing substantial equivalence to a legally marketed predicate device (Cynosure TempSure™ K190678), with new additions being the FlexSure™ Applicators. The "study" proving the device meets acceptance criteria primarily relies on non-clinical bench testing.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided document:

    Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria in the way one might expect for a clinical study with specific performance metrics (e.g., sensitivity, specificity for an AI diagnostic device). Instead, the "acceptance criteria" here are implicitly tied to demonstrating substantial equivalence to the predicate device, especially regarding safety and effectiveness.

    The key performance aspect tested for the new addition (FlexSure™ Applicator) is its ability to heat and maintain temperature of the treatment area for 10 minutes in a simulated clinical condition.

    Table of "Acceptance Criteria" and Reported Device Performance (as inferred from the document for the newly added feature):

    Acceptance Criteria (Inferred for New Feature)Reported Device Performance
    Ability of FlexSure™ Applicator to heat and maintain temperature of the treatment area for at least 10 minutes."The Tissue Heating bench test demonstrated that the device is able to maintain temperature for at least the 10-minute treatment time when used as intended under clinical conditions at various treatment areas and settings."
    Compliance with Electromagnetic Compatibility (EMC) standards (e.g., IEC 60601-1-2)."Electromagnetic Compatibility and Electrical Safety testing shows that the device is safe to use and meets required standards." (Specific standards listed: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2)
    Compliance with Electrical Safety standards (e.g., IEC 60601-1, IEC 60601-2-2)."Electrical safety testing for the Cynosure TempSure was also completed to prove the safe use of the device... and meets required standards."

    Study Details:

    This submission is a 510(k) Premarket Notification, which primarily relies on demonstrating substantial equivalence to a predicate device, rather than novel clinical efficacy data. The "study" in this context refers to the non-clinical testing performed to support this equivalence for the new components.

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not explicitly stated as a number of "samples" in the traditional sense of patient data. For the Tissue Heating with FlexSure™ Applicator bench test, it involved testing the applicator with a "large neutral pad" in a "simulated clinical condition." The specifics of how many times this was tested or over how many "simulated" areas are not provided.
      • Data Provenance: The data originates from bench testing conducted by Cynosure, LLC. The location of the testing is not specified, but the applicant (Cynosure) is based in Westford, Massachusetts, USA. The data is retrospective in the sense that it's reported from completed tests, not ongoing prospective patient studies.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable in the context of this 510(k) submission. The ground truth for the performance of this electrosurgical device is established through engineering and physics principles, measured physical characteristics (temperature, electrical safety parameters), and compliance with international standards (IEC). There are no human experts establishing a "ground truth" for the device's diagnostic performance, as it is a treatment and surgical device, not a diagnostic AI system.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical trials or studies where human readers (e.g., radiologists) interpret medical images or data, and their interpretations need to be reconciled to form a ground truth. This is a non-clinical bench test of an electrosurgical device.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI diagnostic aids where human interpretation is involved. This device is an electrosurgical system. The document explicitly states: "No clinical tests submitted."
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • N/A (not applicable in the AI sense). This device is a physical electrosurgical system. Its "standalone" performance means its inherent operational characteristics (e.g., temperature output, electrical safety) independent of a human operator's actions beyond basic control. The bench tests evaluated the device itself. It does not refer to an AI algorithm functioning in isolation.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth for the non-clinical tests was based on physical measurements and engineering standards.
        • For Tissue Heating: The "ground truth" was the measured temperature achieved and maintained over time in the simulated tissue, compared against the design specification of "at least 10 minutes."
        • For Electrical Safety and EMC: The "ground truth" was compliance with specified international standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2).
    7. The sample size for the training set:

      • Not applicable. This submission is for a physical medical device, not an AI/Machine Learning algorithm that requires a "training set."
    8. How the ground truth for the training set was established:

      • Not applicable. As there is no training set for an AI algorithm, this question is irrelevant to this device submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K193426

    Validate with FDA (Live)

    Device Name
    Elite iQ
    Manufacturer
    Date Cleared
    2020-03-09

    (90 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    755mm:
    The Elite iQ Laser System is indicated for stable long-term or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured at 6, 9 or 12 months after the completion of a treatment regime. It is used for skin types (Fitzpatrick I-VI) including tanned skin. It is also indicated for the treatment of vascular lesions, benign pigmented lesions, and wrinkles.

    1064nm:
    The Elite iQ Laser System is intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangiectasia, rosacea, venous lakes, leg veins and poikiloderma of Civatte; and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis. The laser is also intended for the treatment of benign pigmented lesions such as, but not limited to, lentigines (age spots), solar lentigines (sunspots), cafe au lait macules, sebortheic keratoses, nevi, chloasma, verrucae, skin tags, keratosis and plaques.

    The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.

    Additionally, the laser is indicated for the treatment of pseudo folliculitis barbae (PFB) and for stable long-term or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

    The Skintel Reader is intended as an objective measurement tool for examining skin melanin content for determining and setting a test spot starting fluence.

    Device Description

    The Elite iQ™ workstation is a dual wavelength system that delivers laser energy in both the Nd:YAG (1064-nm) and Alexandrite (755-nm) wavelengths. Through various spot sizes, fluences and repetition rates, the system offers hair removal treatment and aesthetic treatments across all skin types. An Alexandrite standalone workstation is also available for purchase. The Elite iQ delivers the laser energy through a lens-coupled optical fiber with a wide range of interchangeable, quick-release laser handpieces with electronic spot recognition. The Elite iQ also includes the Skintel® Melanin Reader for objective measurement of the melanin content of skin.

    The locking casters allow this stand-alone laser system to be secured in place, as well as to be conveniently moved or transported. When not in use, handpiece components and other system components can be stowed in the storage area located in the side drawer. All system connections, such as the foot switch and remote interlock connections, are located on the rear of the laser. This includes all applicable device labels. User-selectable controls and functions are located on the front of the laser. Elite iQ software features include a Windows® operating system, LCD touch screen and a state-of-the-art graphic user interface.

    AI/ML Overview

    The provided FDA 510(k) summary for the Cynosure Elite iQ laser system primarily focuses on demonstrating substantial equivalence to predicate devices through non-clinical performance and safety testing. It does not involve an AI/ML-driven device, clinical studies to evaluate diagnostic performance (like those for AI-driven image analysis), or human-in-the-loop performance studies. Therefore, many of the requested categories (e.g., sample size for test/training sets, expert ground truth adjudication for AI, MRMC studies) are not applicable to this submission.

    The "Performance Testing - Bench" section describes a validation study related to the "Hair Removal Treatment Guidance Mode" (which provides starting settings based on user input, not an AI algorithm) for safety. This is the closest aspect to "acceptance criteria and study that proves the device meets acceptance criteria" within the provided text, but it's not an AI/ML context.

    Here's a breakdown based on the provided document, acknowledging the non-AI/ML context:


    Device: Cynosure Elite iQ Laser System

    Purpose of Submission (K193426): To demonstrate substantial equivalence to legally marketed predicate devices for indications related to hair reduction, vascular lesions, pigmented lesions, and wrinkle treatment, and for the Skintel Reader which measures skin melanin content.

    1. A table of acceptance criteria and the reported device performance:

    The document states that "Performance testing showed no complications from using the recommended settings, thus confirming the treatment parameters set forth in the Hair Removal Treatment Guidance Mode are a safe starting point for practitioners performing test spots." It also mentions "Results showed no undesirable or unexpected AEs from test spots."

    While explicit numerical acceptance criteria (e.g., "no more than X% of patients experience AE") are not presented in this summary section, the implicit acceptance criterion for the "Hair Removal Treatment Guidance Mode" validation was that the recommended settings would lead to no undesirable or unexpected adverse events (AEs) when applied as test spots.

    Acceptance Criteria (Implicit for Hair Removal Treatment Guidance Mode)Reported Device Performance
    Recommended test spot settings are a safe starting point.No undesirable or unexpected AEs from test spots.

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated as a number of subjects. The document refers to "The subjects chosen were at the high end of the recommended skin type range."
    • Data Provenance: The validation study was performed by E. Victor Ross MD, affiliated with Scripps Green Hospital in San Diego, CA (USA). It appears to be a prospective validation study, as it describes patients being "evaluated 48-72 hours post treatment."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: One expert is explicitly mentioned: E. Victor Ross MD.
    • Qualifications of Experts: Affiliated with Scripps Green Hospital in San Diego, CA. (Specific domain experience like "dermatologist with X years experience" is not detailed in this summary).

    4. Adjudication method for the test set:

    Not applicable in the context of this safety validation. The expert (E. Victor Ross MD) performed the study and evaluated the outcomes. There's no mention of multiple reviewers or an adjudication process for what would typically be an AI ground truth dataset in an AI/ML context.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a laser system and a skin melanin reader; it does not involve AI assistance for human readers or interpretation of medical images.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable in the context of AI/ML. The "Hair Removal Treatment Guidance Mode" provides recommendations, but those are based on pre-programmed settings and user inputs, not an AI algorithm performing alone.

    7. The type of ground truth used:

    For the "Hair Removal Treatment Guidance Mode" validation, the "ground truth" was the observed absence of undesirable or unexpected adverse events (AEs) on patients following treatment with the system's recommended settings. This is akin to outcomes data related to safety.

    8. The sample size for the training set:

    • Training Set Sample Size: Not applicable. The "Hair Removal Treatment Guidance Mode" settings were "created based on previous user experience with the Elite+ Laser System," rather than being derived from a formal training dataset for an AI model.

    9. How the ground truth for the training set was established:

    Not applicable, as there was no formal "training set" in the context of an AI/ML algorithm. The settings were derived from "previous user experience" with an earlier product (Elite+ Laser System), suggesting historical clinical observations and expert consensus from practical application over time.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1