K Number
K212891

Validate with FDA (Live)

Device Name
TempSure System
Manufacturer
Date Cleared
2022-03-18

(189 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Small 10mm, 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids.

The Small 10mm, 18mm, 25mm, 30mm, and 60mm handpieces and the FlexSure Applicators provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

The massage device is intended to provide a temporary reduction in the appearance of cellulite.

The following surgical modes are applicable to the generator:

Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed.

Cutting: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, tumors, cosmetic repairs, development of skin flaps, skin tags and blepharoplasty.

Blended Cutting and Coagulation: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids, keratosis, verrucae, basal cell carcinoma, nevi, fistulas, epithelidma, cosmetic repairs, cysts, abscesses, and development of skin flaps.

Fulguration: basal cell carcinoma, papilloma, cyst destruction, tumors. verrucae, hemostasis.

Bipolar: pinpoint precise coagulation, pinpoint hemostasis in any, field (wet or dry), snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment and turbinate shrinkage.

Device Description

The Cynosure TempSure™ System is a radiofrequency generator with a variety of applications both aesthetic and surgical procedures. The intended action is achieved through application of radiofrequency energy to the patient which results in minimization of heat dissipation and cellular alteration. Output of energy is controlled via the guided user interface (GUI) and the foot and/or hand-switch.

An additional Small 10mm Smart Handpiece is now included for the purpose of tissue heating for selected medical conditions and the treatment of mild to moderate facial wrinkles and rhytids. The handpiece is capable of operating at maximum energy setting of 30 (28W) in the Smart Handpiece Mode. There have been no additional changes to the previous existing handpieces (10, 15, 18, 20, 25, 30 or 60mm), FlexSure Applicators, or Surgical modes (COAG, Cut, Blend, Bipolar, Fulgurate) of the system. The TempSure system is still used with the same existing electrosurgical accessories.

The TempSure™ System includes:
TempSure™ Generator
Temperature Sensing Handpieces (Small 10, 10, 15, 18, 20, 25, 30, and 60mm)
Massage Heads (25, 30, and 60mm)
FlexSure™ Applicators (Large and Medium)
IEC Power Cord
Footswitch
Disposable/Reusable Neutral Pads
Surgical Fingerswitch/Foot Controlled Handpieces
Monopolar Cables
Disposable/Reusable Electrodes, Forceps

AI/ML Overview

The TempSure System is an electrosurgical cutting and coagulation device. The acceptance criteria and the study that proves the device meets them are summarized below.

1. Table of Acceptance Criteria & Reported Device Performance:

Acceptance CriteriaReported Device Performance
Non-clinical Tests:
Tissue Heating with Small 10mm Handpiece
Able to heat and maintain temperature of treatment area in simulated clinical conditionsDemonstrated that the device is able to maintain a temperature for at least the 10-minute treatment time when used as intended under clinical conditions at various treatment areas and settings.
Electromagnetic Compatibility and Electrical Safety
Compliance with IEC 60601-1 (General requirements for basic safety and essential performance)Test reports provided in accordance with FDA Guidance demonstrate compliance.
Compliance with IEC 60601-1-2 (Electromagnetic disturbances requirements and tests)Test reports provided in accordance with FDA Guidance demonstrate compliance.
Compliance with IEC 60601-2-2 (Particular requirements for basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories)Test reports provided in accordance with FDA Guidance demonstrate compliance.

2. Sample Size for the Test Set and Data Provenance:

The document states that bench testing was conducted for the tissue heating evaluation. However, it does not specify a numerical sample size (e.g., number of handpieces tested, number of simulated treatment areas). The provenance is explicitly a "simulated clinical condition," indicating a laboratory-based, prospective evaluation rather than real-world patient data.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

Not applicable. The reported studies are non-clinical bench tests. There is no mention of expert involvement for ground truth establishment.

4. Adjudication Method for the Test Set:

Not applicable. The reported studies are non-clinical bench tests, which do not typically involve adjudication methods like those used for expert consensus on clinical data.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No. The document explicitly states: "N/A - No clinical tests submitted." Therefore, no MRMC comparative effectiveness study was done.

6. Standalone Performance:

Yes, a standalone (algorithm only without human-in-the-loop performance) was done, but in the context of a medical device rather than an algorithm. The non-clinical tests evaluated the device's technical performance (tissue heating capabilities, electrical safety) directly, without human interpretation of results in a diagnostic or treatment decision-making cycle.

7. Type of Ground Truth Used:

For the tissue heating test, the ground truth was based on the ability of the device to heat and maintain a specific temperature in "simulated clinical conditions." This would likely involve quantitative measurements of temperature. For the electrical safety and electromagnetic compatibility tests, the ground truth was regulatory compliance with established international standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2), which are objectively measurable criteria.

8. Sample Size for the Training Set:

Not applicable. The device is a radiofrequency generator with handpieces, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as no training set was used for an AI/ML algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA acronym is in a blue square, and the full name "U.S. Food & Drug Administration" is written in blue letters next to it.

March 18, 2022

Cynosure, LLC Michael King Regulatory Affairs Specialist III 5 Carlisle Road Westford, Massachusetts 01886

Re: K212891

Trade/Device Name: TempSure System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI, PBX Dated: February 17, 2022 Received: February 18, 2022

Dear Michael King:

We have reviewed your Section 510/k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) BBCDOX K212891

Device Name TempSure System

Indications for Use (Describe)

The Small 10mm, 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids.

The Small 10mm, 18mm, 25mm, 30mm, and 60mm handpieces and the FlexSure Applicators provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

The massage device is intended to provide a temporary reduction in the appearance of cellulite.

The following surgical modes are applicable to the generator:

Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed.

Cutting: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, turnors, cosmetic repairs, development of skin flaps, skin tags and blepharoplasty.

Blended Cutting and Coagulation: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids, keratosis, verrucae, basal cell carcinoma, nevi, fistulas, epithelidma, cosmetic repairs, cysts, abscesses, and development of skin flaps.

Fulguration: basal cell carcinoma, papilloma, cyst destruction, tumors. verrucae, hemostasis.

Bipolar: pinpoint precise coagulation, pinpoint hemostasis in any, field (wet or dry), snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment and turbinate shrinkage.

Type of Use ( Select one or both, as applicable )
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Cynosure, LLC

Section 5

510(k) Summary

510(k) Summary for Cynosure TempSure

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.

807.92(a)(1) Submitter Information
ApplicantCynosure, LLC
Address5 Carlisle Road, Westford MA, 01886
Phone Number781-993-2454
Fax Number978-256-6556
Establishment RegistrationNumber1222993
Contact PersonMichael King
Preparation DateSeptember 9, 2021
807.92(a)(2) Name of Device
Trade or Proprietary NameTempSure System
Common or Usual NameSurgical RF Generator
Classification NameElectrosurgical, Cutting & Coagulation & Accessories;Massager, Vacuum, Radio Frequency Induced Heat
Classification PanelGeneral & Plastic Surgery
Regulation21 CFR 878.4440
Regulatory ClassII
Product Code(s)GEI, PBX
807.92 (a)(3) Legally marketed device(s) to which equivalence is claimed
Predicate DevicesCynosure TempSure (K200241)
807.92(a)(4) Device Description
The Cynosure TempSure™ System is a radiofrequency generator witha variety of applications both aesthetic and surgical procedures. Theintended action is achieved through application of radiofrequencyenergy to the patient which results in minimization of heat dissipationand cellular alteration. Output of energy is controlled via the guideduser interface (GUI) and the foot and/or hand-switch.An additional Small 10mm Smart Handpiece is now included for thepurpose of tissue heating for selected medical conditions and thetreatment of mild to moderate facial wrinkles and rhytids. Thehandpiece is capable of operating at maximum energy setting of 30(28W) in the Smart Handpiece Mode. There have been no additionalchanges to the previous existing handpieces (10, 15, 18, 20, 25, 30 or60mm), FlexSure Applicators, or Surgical modes (COAG, Cut, Blend,
Section 5510(k) Summary
Bipolar, Fulgurate) of the system. The TempSure system is still used with the same existing electrosurgical accessories.
The TempSure™ System includes:
TempSure™ Generator Temperature Sensing Handpieces (Small 10, 10, 15, 18, 20, 25, 30, and 60mm) Massage Heads (25, 30, and 60mm) FlexSure™ Applicators (Large and Medium) IEC Power Cord Footswitch Disposable/Reusable Neutral Pads Surgical Fingerswitch/Foot Controlled Handpieces Monopolar Cables Disposable/Reusable Electrodes, Forceps
807.92(a)(5) Intended Use of the Device
The Small 10mm, 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids.The Small 10mm, 18mm, 25mm, 30mm, and 60mm handpieces and the FlexSure Applicators provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.The massage device is intended to provide a temporary reduction in the appearance of celluliteThe following surgical modes are applicable to the generator:Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed.Cutting: snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, tumors, cosmetic repairs, development of skin flaps, skin tags, and blepharoplasty.Blended Cutting and Coagulation: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids, keratosis, verrucae, basal cell carcinoma,
nevi, fistulas, epithelidma, cosmetic repairs, cysts, abscesses, anddevelopment of skin flaps.
Fulguration: basal cell carcinoma, papilloma, cyst destruction,tumors, verrucae, hemostasis.
Bipolar: pinpoint precise coagulation, pinpoint hemostasis in any,field (wet or dry), snoring, submucosal palatal shrinkage, traditionaluvulopalatoplasty (RAUP), myringotomy with effective hemorrhagecontrol, epistaxis treatment, and turbinate shrinkage
807.92(a)(6) Summary of the Technological Characteristics of the Device Compared to

{5}------------------------------------------------

{6}------------------------------------------------

Section 5

the Predicate - Please refer to the System Specification Comparison Table below

807.92(b)(1) Non-clinical tests submitted

The following non-clinical tests have been included in this 510k submission in determination of substantial equivalence between the test device and the referenced predicates.

Tissue Heating with Small 10mm Handpiece

Bench testing was conducted to show that the new Small 10mm Smart Handpiece, when used in a simulated clinical condition, was able to heat and maintain temperature of the treatment area for 10 minutes. Refer to Section 18 - Performance Testing, Bench for additional information.

Electromagnetic Compatibility and Electrical Safety

Electrical safety testing for the Cynosure TempSure was also completed to prove the safe use of the device. These test reports are provided in accordance with FDA Guidance "Premarket Notification 510k Submissions for Electrosurgical Devices for General Surgery" - Section XII - Electrical Safety and Electromagnetic Compatibility". The following test reports are available in Section 17 -Electromagnetic Compatibility and Electrical Safety.

  • IEC 60601-1, Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performances
  • IEC 60601-1-2, Medical Electrical Equipment – Part 1 -2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances requirements and tests
  • IEC 60601-2-2, Medical electrical equipment Part 2 -2: Particular requirements for the basic o safety and essential performance of high frequency surgical equipment and high frequency surgical accessories

807.92(b)(2) Clinical tests submitted - N/A - No clinical tests submitted

807.92(b)(3) Conclusions drawn from clinical and non-clinical tests submitted

{7}------------------------------------------------

Page 4 of 7 Cynosure, LLC

Section 5

510(k) Summary

The non-clinical tests demonstrate that the TempSure generator with the Small 10mm handpiece is safe and effective and performs as well or better than the legally marketed predicate device. The Tissue Heating bench test demonstrated that the device is able to maintain temperature for at least the 10minute treatment time when used as intended under clinical conditions at various treatment areas and settings. In addition, the Electrosurgical Compatibility and Electrical Safety testing shows that the device is safe to use and meets required standards. These non-clinical tests show that the TempSure generator using the Small 10mm handpiece meets design specifications as well as performance requirements.

{8}------------------------------------------------

Cynosure TempSure 510(k)
KPending

CharacteristicCynosure TempSure System (KPending)Cynosure TempSure System (K200241)
Indications for UseThe Small 10mm, 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids.The Small 10mm, 18mm, 25mm, 30mm, and 60mm handpieces and the FlexSure Applicators provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.The 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids.The 18mm, 25mm, 30mm, and 60mm handpieces and the FlexSure Applicators provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.
The massage device is intended to provide a temporary reduction in the appearance of celluliteThe massage device is intended to provide a temporary reduction in the appearance of cellulite
The following surgical modes are applicable to the generator:The following surgical modes are applicable to the generator:
Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performedCoagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed
Cutting: snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage. skin incisions, biopsy, cysts, abscesses. tumors. cosmetic repairs, development of skin flaps. skin tags and blepharoplasty.Cutting: snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage. skin incisions, biopsy, cysts, abscesses. tumors. cosmetic repairs, development of skin flaps. skin tags and blepharoplasty.
Blended Cutting and Coagulation: snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment,Blended Cutting and Coagulation: snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment,
Section 5510(k) Summary
CharacteristicCynosure TempSure System (KPending)Cynosure TempSure System (K200241)
effective hemorrhage control. epistaxis treatment,turbinate shrinkage, skin tags, papilloma keloids.keratosis, verrucae, basal cell carcinoma. nevi. fistulas.epithelidma. cosmetic repairs. cysts. abscesses. anddevelopment of skin flaps.turbinate shrinkage, skin tags, papilloma keloids.keratosis, verrucae, basal cell carcinoma. nevi. fistulas.epithelidma. cosmetic repairs. cysts. abscesses. anddevelopment of skin flaps.
Fulguration: basal cell carcinoma. papilloma. cystdestruction, tumors. verrucae, hemostasis.Fulguration: basal cell carcinoma. papilloma. cystdestruction, tumors. verrucae, hemostasis.
Bipolar: pinpoint precise coagulation. pinpointhemostasis in any, field (wet or dry). snoring.submucosal palatal shrinkage, traditionaluvulopalatoplasty (RAUP). myringotomy witheffective hemorrhage control. epistaxis treatment andturbinate shrinkageBipolar: pinpoint precise coagulation. pinpointhemostasis in any, field (wet or dry). snoring.submucosal palatal shrinkage, traditionaluvulopalatoplasty (RAUP). myringotomy witheffective hemorrhage control. epistaxis treatment andturbinate shrinkage
Rx/OTCPrescriptionPrescription
Energy TypeRadiofrequencyRadiofrequency
ModalityMonopolar, BipolarMonopolar, Bipolar
Temperature SensingTemperature-Sensitive HandpieceTemperature-Sensitive Handpiece
Temperature Response Time<1 second<1 second
Handpiece SizeSmall 10mm, 10mm, 15mm, 18mm, 20mm, 25 mm, 30mm, 60mm, Large & Medium FlexSure Applicators10mm, 15mm, 18mm, 20mm, 25 mm, 30 mm, 60mm,Large & Medium FlexSure Applicators
Massage HeadYes (25mm, 30mm, 60mm)Yes (25mm, 30mm, 60mm)
Surgical AccessoriesThe TempSure™ generator is compatible withexisting, previously cleared electrosurgicalaccessories: neutral pads, electrodes, surgicalhandpieces, cables, etc.The TempSure™ generator is compatible withexisting, previously cleared electrosurgical accessories:neutral pads, electrodes, surgical handpieces, cables,etc.
Treatment ActivationFingerswitch FootswitchFingerswitch Footswitch
Section 5510(k) Summary
CharacteristicCynosure TempSure System (KPending)Cynosure TempSure System (K200241)
esthetic OptimalTemperature39-45°C39-45°C
Patient Contacting MaterialSmall 10mm Handpiece: Gold-plated 316 stainlesssteelHandpieces: PVDF coating, Gold-plated brass, Gold-plated aluminum, Loctite (M-31CL), Polyetherimide(Ultem 1000)FlexSure: Polyimide, HydrogelMassage Heads: Polycarbonate (Sabic Lexan HP-1),DelrinHandpieces: PVDF coating, Gold-plated brass, Gold-plated aluminum, Loctite (M-31CL), Polyetherimide(Ultem 1000)FlexSure: Polyimide, HydrogelMassage Heads: Polycarbonate (Sabic Lexan HP-1),Delrin
Input Voltage110V110V
Output Waveform4.0 MHz Sin-wave CW, Fully Rectified, PartiallyRectified, and 1.7 MHz for Bipolar4.0 MHz Sin-wave CW, Fully Rectified, PartiallyRectified, and 1.7 MHz for Bipolar
ModesSurgical (Coag, Cut, Blend, Fulgurate, Bipolar)Smart Handpiece ModeSurgical (Coag, Cut, Blend, Fulgurate, Bipolar)Smart Handpiece Mode
Max Power Output300W (Surgical)120W (Wrinkles)300W (Tissue Heating)300W (Surgical)120W (Wrinkles)300W (Tissue Heating)
Safety Features• Compliant with medical electrical equipmentrequirements for safety and performance (60601-1,60601-1-2, 60601-2-2)• Small 10mm Smart Handpiece is equipped with anintegrated accelerometer to detect handpiece motion.RF will be disabled when no handpiece motion isdetected.• Compliant with medical electrical equipmentrequirements for safety and performance (60601-1,60601-1-2, 60601-2-2)
crest Factor3.1 (Surgical)1.5 (Smart Handpiece)3.1 (Surgical)1.5 (Smart Handpiece)
Dimensions22.5" x 18" x 12"22.5" x 18" x 12"

{9}------------------------------------------------

Cynosure TempSure 510(k)
KPending

{10}------------------------------------------------

Cynosure TempSure 510(k)
KPending

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.