Intended Use

755mm:
The PicoSure Workstation is indicated for tattoo and benign pigmented lesions removal including but not limited to: Nevus of Ota, Hori macules (nevus of Hori), and Melasma. The PicoSure Workstation with the 2mm and 6mm hand pieces and the Focus Array are indicated for the treatment of acne scars and wrinkles in Skin Types I – IV.

532nm:
The PicoSure 532-nm delivery system is indicated for tattoo removal and benign pigmented lesions removal in Skin Types I-III.

1064nm:
The PicoSure 1064-nm delivery system is indicated for tattoo and benign pigmented lesions removal.

Device Description

The PicoSure Workstation is a high-powered, Alexandrite system that delivers laser energy in the 755-nm nominal wavelength. The system offers treatment through a variety of spot sizes, fluences and repetition rates. Laser activation is by footswitch. In addition to the 755nm handpiece, optional 532nm Laser Delivery System and/or 1064nm Laser Delivery System can replace the 755nm handpiece at the distal end of the articulated arm. These Delivery Systems convert the 755nm laser energy into a 532nm wavelength or a 1064 nm wavelength and are available in multiple spot sizes.

AI/ML Overview

The provided document is a 510(k) summary for the PicoSure Workstation, a laser surgical instrument. It details the device's indications for use and compares it to predicate devices. However, it does not contain the specific information required to complete a table of acceptance criteria and reported device performance directly in the format you requested, especially regarding quantitative performance metrics, sample sizes for test sets, and details about expert involvement in establishing ground truth.

The document indicates that to support the expanded indications for use (Nevus of Ota, Hori macules, and Melasma), a literature search was conducted. This means the clinical evidence is based on existing published studies rather than a new, dedicated clinical trial performed for this 510(k) submission.

Here's an attempt to answer your questions based on the provided text, with clear indications where information is not available from this document:

1. A table of acceptance criteria and the reported device performance

The document does not specify quantitative acceptance criteria (e.g., minimum sensitivity, specificity, or specific improvement percentages) directly associated with this 510(k) submission. Instead, it relies on demonstrating substantial equivalence to a predicate device and supporting new indications through literature review. The reported "performance" is qualitative, focusing on whether the existing literature showed "acceptable clinical effectiveness and safety" for the new indications.

Criteria (Hypothesized based on text)Reported Device Performance (from literature review)
Clinical EffectivenessSix identified articles reported acceptable clinical effectiveness for Nevus of Ota, Hori macules (nevus of Hori), and Melasma using the PicoSure at 755nm.
Safety ProfileSix identified articles reported acceptable safety. No serious adverse events were reported. Side effects were primarily transient and included common acceptable side effects for laser treatments as described in the product labeling.
Substantial EquivalenceThe device is deemed substantially equivalent to the PicoSure Workstation (K173199) and RevLite Q-Switched Nd:YAG Laser System (K133254) which were legally marketed predicate devices, with no changes to the technological characteristics, only an expansion of indications based on clinical literature.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample size: Not explicitly stated as a single test set size. The document mentions "at least ten individuals" in each of the six identified articles. Thus, the total number of individuals across all six studies is at least 60. Individual study sizes are not detailed.
  • Data provenance: The document states "peer reviewed, published articles." No specific countries of origin are mentioned.
  • Retrospective or prospective: The articles were identified as "prospective studies."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Not Applicable / Not Available. Since the evidence is based on published clinical studies, the "ground truth" and its establishment would be inherent to the methodologies of those individual studies. The 510(k) summary does not provide details on how ground truth was established within these external studies or the qualifications of any experts involved in those studies.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not Applicable / Not Available. This information would pertain to the specific methodologies of the clinical studies identified in the literature search. The 510(k) summary does not provide these details.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This document is for a laser workstation, not an AI-assisted diagnostic device. Therefore, no MRMC study or AI assistance effect size is relevant or mentioned.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is a hardware device (laser workstation), not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Not Applicable / Not Available. The "ground truth" would relate to the diagnosis of the benign pigmented lesions (Nevus of Ota, Hori macules, Melasma) and the success of their removal or clinical improvement. The specific method used to establish these in the surveyed literature is not detailed in this 510(k) summary. It likely involved clinical diagnosis by dermatologists and visual assessment of treatment outcomes.

8. The sample size for the training set

  • Not Applicable. There is no "training set" in the context of this device. The device's operation is based on its physical properties and established laser physics, not on a machine learning model trained on data. The clinical literature reviewed supported the indications for use, not a training process for the device itself.

9. How the ground truth for the training set was established

  • Not Applicable. As there is no training set, this question is not relevant.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 30, 2021

Cynosure LLC Michael King Regulatory Affairs Specialist III 5 Carlisle Road Westford, Massachusetts 01886

Re: K210226

Trade/Device Name: PicoSure Workstation Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: January 26, 2021 Received: January 28, 2021

Dear Michael King:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) KPending

Device Name PicoSure™ Workstation

Indications for Use (Describe)

755mm:

The PicoSure Workstation is indicated for tattoo and benign pigmented lesions removal including but not limited to: Nevus of Ota, Hori macules (nevus of Hori), and Melasma. The PicoSure Workstation with the 2mm and 6mm hand pieces and the Focus Array are indicated for the treatment of acne scars and wrinkles in Skin Types I – IV.

532nm:

The PicoSure 532-nm delivery system is indicated for tattoo removal and benign pigmented lesions removal in Skin Types I-III.

1064nm:

The PicoSure 1064-nm delivery system is indicated for tattoo and benign pigmented lesions removal.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K210226 – 510(k) Summary for Cynosure PicoSure

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.

807.92(a)(1) Submitter Information
ApplicantCynosure, LLC
Address5 Carlisle Road, Westford MA, 01886
Phone Number781-993-2454
Fax Number978-256-6556
Establishment RegistrationNumber1222993
Contact PersonMichael King
Preparation DateMarch 22, 2021
807.92(a)(2) Name of Device
Trade or Proprietary NamePicoSure™ Workstation
Common or Usual NameLaser Workstation
Classification NamePowered Laser Surgical Instrument
Classification PanelGeneral & Plastic Surgery
Regulation21 CFR 878.4810
Regulatory ClassII
Product Code(s)GEX
807.92 (a)(3) Legally marketed device(s) to which equivalence is claimed
Predicate DevicePicoSure Workstation (K173199)
Reference DeviceRevLite Q-Switched Nd:YAG Laser System (K133254)
807.92(a)(4) Device Description
The PicoSure Workstation is a high-powered, Alexandritesystem that delivers laser energy in the 755-nm nominalwavelength. The system offers treatment through a variety ofspot sizes, fluences and repetition rates. Laser activation is byfootswitch. In addition to the 755nm handpiece, optional 532nmLaser Delivery System and/or 1064nm Laser Delivery Systemcan replace the 755nm handpiece at the distal end of thearticulated arm. These Delivery Systems convert the 755nmlaser energy into a 532nm wavelength or a 1064 nm wavelengthand are available in multiple spot sizes.
807.92(a)(5) Intended Use of the Device755nm:The PicoSure Workstation is indicated for tattoo and benignpigmented lesions removal including but not limited to: Nevusof Ota, Hori macules (nevus of Hori), and Melasma. ThePicoSure Workstation with the 2mm and 6mm hand pieces andthe Focus Array are indicated for the treatment of acne scars andwrinkles in Skin Types I - IV.
532nm:The PicoSure 532-nm delivery system is indicated for tattooremoval and benign pigmented lesions removal in Skin Types I-III.
1064nm:The PicoSure 1064-nm delivery system is indicated for tattooand benign pigmented lesions removal.
807.92(a)(6) Summary of the Technological Characteristics of the Device Compared tothe PredicateThere have been no changes to the technological characteristicsof the device compared to the predicate PicoSure Workstation(K173199). A review of clinical literature was conducted tosupport the addition of specific indications nevus of Ota, Horimacules (nevus of Hori), and melasma to the generalindications for use of the PicoSure Workstation.Please refer to the Laser System Specification comparison tablebelow.RevLite (K133254) is being included as a reference device forthis submission. PicoSure was originally cleared underK121346, using RevLite as a predicate device.

{4}------------------------------------------------

807.92(b)(1) Non-clinical tests submitted- N/A - No non-clinical tests submitted

807.92(b)(2) Clinical tests submitted - A literature search was conducted to identify peer reviewed, published articles in which the Picosure device was used to treat specific benign pigmented lesions nevus of Ota, Hori macules (nevus of Hori), and melasma. Six articles were identified that reported prospective studies that were controlled or randomized and that treated at least ten individuals with the target pigmented lesion using the PicoSure at 755nm. These studies reported acceptable clinical effectiveness and safety.

There were no serious adverse events reported in the publications. Side effects reported in the studies were primarily transient and included common acceptable side effects for laser treatments as described in the product labeling.

{5}------------------------------------------------

807.92(b)(3) Conclusions drawn from clinical and non-clinical tests submitted – The clinical data submitted demonstrates that the PicoSure device utilizing the 755nm wavelength is safe and effective for the treatment of specific benign pigmented lesions nevus of Ota, Hori macules (nevus of Hori), and melasma. We therefore conclude that the subject device is substantially equivalent to the PicoSure Workstation predicate device cleared in K173199.

{6}------------------------------------------------

Indications for Use

PicoSure WorkstationPicoSure WorkstationRevLite Q-Switched Nd: YAG Laser System
(K210226)(K173199)(K133254)
755mm:The PicoSure Workstation is indicatedfor tattoo and benign pigmented lesionsremoval including but not limited to:Nevus of Ota, Hori macules (nevus ofHori), and Melasma. The PicoSureWorkstation with the 2mm and 6mmhand pieces and the Focus Array areindicated for the treatment of acne scarsand wrinkles in Skin Types I - IV.532mm:The PicoSure 532-nm delivery system isindicated for tattoo removal and benignpigmented lesions removal in SkinTypes I- III.1064nm:The PicoSure 1064-nm delivery systemis indicated for tattoo and benignpigmented lesions removal.755 nm:The PicoSure Workstation is indicatedfor tattoo and benign pigmented lesionsremoval. The PicoSure Workstationwith the 2mm and 6mm handpieces andthe Focus Array are indicated forthe treatment of acne scars and wrinklesin Skin Types I-IV.532 nm:The PicoSure 532 nm Laser DeliverySystem is indicated for tattoo removaland benign pigmented lesion removal inSkin Types I-III.1064 nm:The PicoSure 1064 nm Laser DeliverySystem is indicated for tattoo and benignpigmented lesions removal.1064 nm:Tattoo Removal (dark ink: blue and black) DermalPigmented Lesions; including but not limited to; Nevus ofOta, Lentigines, Nevi, Melasma and Café-au-laitRemoval or lightening of hair with or without adjuvantpreparation.Skin Resurfacing for Acne Scars and WrinklesBenign cutaneous lesions; including but not limited to:striae and scars (excludes the 650nm wavelength)Reduction of red pigmentation in hypertrophic and keloidscars where vascularity is an integral part of the scar(excludes the 650nm wavelength)532nm:(nominal delivered energy of 585nm and 650nm with theOptional Multilite Dye Laser Handpiece)Tattoo removal (light ink: red, sky blue, green) Vascularlesions including but not limited to: port wine birthmarks,telangiectasias, spider angioma, cherry angioma, spiderneviEpidermal Pigmented lesions; including, but not limited to:café-au-lait birthmarks, solar lentigines, senile lentigines,Becker's nevi, Freckles, Nevus spilus, seborrheic keratosisSkin Resurfacing for Acne Scars and WrinklesBenign cutaneous lesions; including but not limited: striaeand scars (excludes the 650nm wavelength)Reduction of red pigmentation in hypertrophic and keloidscars where vascularity is an integral part of the scar(excludes the 650nm wavelength)

{7}------------------------------------------------

Laser System Specifications

DescriptionPicoSure Workstation(K210226)PicoSure Workstation(K173199)RevLite Q-Switched Nd: YAGLaser System (K133254)
Laser TypeNd:YVO4FrequencyDoubledNd:YVO4AlexandriteNd:YVO4FrequencyDoubledNd:YVO4AlexandriteNd:YAG
Wavelength (nm)1064 nm532 nm755 nm1064 nm532 nm755 nm1064nm532nm
MaximumAverage Fluence(J/cm²)3.6 J/cm²1.5 J/cm²6.37 J/cm²3.6 J/cm²1.5 J/cm²6.37 J/cm²8 J/cm²
Repetition Rate(Hz)1, 2.5, 5,101, 2.5, 5, 10Single Shot,1, 2.5, 5, 101, 2.5, 5,101, 2.5, 5, 10Single Shot,1, 2.5, 5, 10Single shot, 1, 2, 5, 10 Hz
Pulse Width450- 900 ps450- 900 ps450- 900 ps450- 900 ps7 – 20 ns
Spot Sizes (mm)Fixed 1.4 -4.0 mmFixed 1.5 –3.5 mmZoom 2-6mmFixed 6, 8,10 mmFixed 1.4 –4.0 mmFixed 1.5 –3.5 mmZoom 2-6mmFixed 6, 8,10 mm2-8.5mm range with 0.1mmincrements
Rx/OTCPrescriptionPrescriptionPrescription
PatientContactingMaterial316 Stainless Steel316 Stainless Steel304 Stainless SteelAnodized Aluminum
Electrical Power200-240 V~, 4.5 kVA, 50/60 Hz, SinglePhase200-240 V~, 4.5 kVA, 50/60 Hz, SinglePhaseAC 230 V, 50/60 Hz
Dimensions43 in (109 cm) Hx 22in (56cm) x 42in(107cm)Note: Height with arm extended 62in(158cm)43 in (109 cm) H x 22in (56cm) x 42in(107cm)Note: Height with arm extended 62in(158cm)31.8" (H) x 12" (W) x 28.5" (D)
Weight375 lb (171 kg)375 lb (171 kg)131 lbs

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.