Found 34 results

510(k) Data Aggregation

    K Number
    K234127

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-01-25

    (28 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Clariti 1 Day Multifocal (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with non-diseased eyes that may require a reading addition of +3.00 Diopters or less and may exhibit astigmatism up to 1.50 Diopters or less.

    Device Description

    Clariti 1 Day Multifocal (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker is a hydrophilic co-polymer of silicone containing monomers and hydrophilic monomers which is cross-linked with tetraethyleneglycol dimethacrylate and difunctional methacryloxypropyl-terminated poly(dimethylsiloxane).

    When hydrated the lens consists of 44.0% somofilcon A and 56.0% water by weight of saline immersed in normal saline. A benzophenone UV absorbing monomer is used in the contact lens to help protect against transmission of harmful UV radiation and Clariti 1 Day Multifocal (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker help protect against transmission of harmful UV radiation to the cornea and into the eye.

    The average transmittance characteristics are less than 5% in the UVB range of 280 to 315nm and less than 50% in the UVA range of 316-380mm.

    The lens has a hemispherical flexible shell, which covers the cornea and a portion of the adjacent sclera.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device, specifically a soft contact lens. It describes the device, its intended use, and compares it to a predicate device to establish substantial equivalence. However, it does not contain information about acceptance criteria or a study proving the device meets specific acceptance criteria in the context of an AI/ML medical device.

    Based on the document, here's what can be inferred and what information is missing:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria (Implied): The document refers to "performance testing specifications" and states that "non-clinical data presented are adequate to support substantial equivalence." This implies that the device met certain pre-defined performance specifications for physical and optical properties like water content and dimensional/optical parameters, as outlined in BS EN ISO 18369-2:2017 and BS EN ISO 18369-4:2017. However, the specific quantitative acceptance criteria (e.g., precise ranges for water content or diameter tolerance) are not explicitly listed in this summary.
    • Reported Device Performance:
      • Refractive Index: 1.4003
      • %Transmittance @ 590nm: 98.13
      • %Transmittance @ 280-315nm: 0.71 (less than 5% as per device description)
      • %Transmittance @ 316-380nm: 20.62 (less than 50% as per device description)
      • Water Content: 56%
      • Oxygen Permeability (DK): 60 x 10^-11 (cm^2/sec) (ml O2/ml x mmHg) at 35°C
      • Specific Gravity: 1.17

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable / Not Provided: This information is typically relevant for studies validating AI/ML algorithms. The provided document is for a physical medical device (contact lens) and refers to non-clinical performance testing. It does not describe a "test set" in the context of algorithm validation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable / Not Provided: This is not relevant for the type of device and testing described in the document.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable / Not Provided: This is not relevant for the type of device and testing described in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable: This document describes a physical medical device (contact lens), not an AI/ML-driven diagnostic or assistive device that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable: This document describes a physical medical device (contact lens), not an AI/ML algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not Applicable / Implied: For physical and optical properties of a contact lens, the "ground truth" would be established by validated measurement techniques and scientific protocols (e.g., using specialized equipment to measure water content, dimensions, oxygen permeability). The document states, "All tests were conducted in accordance with the GLP regulation (21 CFR Part 58) or according to valid scientific protocols."

    8. The sample size for the training set

    • Not Applicable / Not Provided: This is not relevant for the type of device and testing described. The device doesn't have a "training set" in the AI/ML sense.

    9. How the ground truth for the training set was established

    • Not Applicable / Not Provided: This is not relevant for the type of device and testing described.

    In summary, the provided text does not describe an AI/ML device or a study validating an AI/ML algorithm. Instead, it is a regulatory submission for a contact lens, focusing on demonstrating substantial equivalence to a predicate device through non-clinical performance testing of physical and optical properties. The acceptance criteria are implicitly derived from ISO standards, and device performance is reported quantitatively for these physical properties.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220070

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-02-04

    (25 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MyDay (stenfilcon A) ASPHERE Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuty.

    MyDay (stenfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

    MyDay (stenfilcon A) MULTIFOCAL Soft Contact lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity.

    MyDay (stenfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optical correction of distance and near vision in presbyopic phakic persons with non-diseased eyes who may have -10.00 diopters of astigmatism or less.

    AVAIRA VITALITY SPHERE and ASPHERE (fanfilcon A) Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

    A VAIRA VITALITY (fanfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eves in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

    AVAIRA VITALITY (fanfilcon A) MULTIFOCAL Soft lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with nondiseased eyes in powers from -20.00 to +20.00 diopters and with add powers from +0.25 to +4.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

    AVAIRA VITALITY (fanfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optic correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eyes in powers of -20.00 to +20.00 diopters with add powers from +0.25 to +4.00 diopters and astigmatism corrections from -0.25 to -10.00 diopters.

    Eye Care Practitioners may prescribe the Avaira Vitality (fanfilcon A) Soft Contact lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement or for single-use disposable wear. When prescribed for frequent/planned replacement, the Avaira Vitality (fanfilcon A) Soft Contact lens is to be cleaned, rinsed and disinfected each time the lens is removed. The contact lens is to be discarded after the recommended wearing period as prescribed by the Eye Care Professional. When prescribed for frequent/planned replacement wear, the lenses may be disinfected using a chemical disinfection only. When prescribed for single-use disposable wear, the Avaira Vitality (fanfilcon A) Soft Contact lens is to be discarded after each removal.

    Device Description

    MyDay Contact Lenses are available as Asphere, Toric, Multifocal, and Multifocal Toric lens designs. The MyDay material, stenfilcon A, is primarily a random copolymer of polydimethylsiloxane methacrylate and vinylmethyl acetamide. The lenses have a blue tint which is added to make the lens more visible for handling. The lenses also contain a UV absorbing monomer, benzotriazolyl methacrylate, which is used to block UV radiation. When placed on the cornea in its hydrated state, the MyDay Soft (Hydrophilic) Contact Lens acts as a refracting medium to focus light rays on the retina.

    Avaira Vitality Contact Lenses are available as Sphere, Asphere, Toric, Multifocal, and Multifocal Toric lens designs. The Avaira Vitality material, fanfilcon A. is primarily a random copolymer of polydimethylsiloxane methacrylate and vinylmethyl acetamide. The lenses have a blue tint which is added to make the lens more visible for handling. The lenses also contain a UV absorbing monomer, benzotriazolyl methacrylate, which is used to block UV radiation. When placed on the cornea in its hydrated state, the Avaira Vitality Soft (Hydrophilic) Contact Lens acts as a refracting medium to focus light rays on the retina.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for two soft contact lenses: MyDay (stenfilcon A) and Avaira Vitality (fanfilcon A). The submission aims to demonstrate substantial equivalence to previously cleared predicate devices, primarily due to a manufacturing process change and a change to the purity specification.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a quantitative, pass/fail format with specific targets. Instead, it refers to "performance testing" conducted according to established standards. The reported device performance is presented as the results of these tests, which presumably align with the expectations of the referenced standards for substantial equivalence.

    Performance TestStandard(s)Reported Device Performance
    Contact AngleANSI Z80.20-2016Testing performed, assumed to meet standard requirements for substantial equivalence.
    Water ContentBS EN ISO 18369-2:2017, BS EN ISO 18369-4:2017Testing performed, assumed to meet standard requirements for substantial equivalence.
    Mechanical PropertiesANSI Z80.20-2016, ASTM D1708-02aTesting performed, assumed to meet standard requirements for substantial equivalence.
    Total ExtractablesANSI Z80.20-2016, BS EN ISO 18369-4:2017Testing performed, assumed to meet standard requirements for substantial equivalence.
    Dimensional/Optical ParametersBS EN ISO 18369-2:2017, BS EN ISO 18369-3:2017Testing performed, assumed to meet standard requirements for substantial equivalence.
    BiocompatibilityISO 10993-18:2020Not required. Equivalency demonstrated through chemical profiling between modified and unmodified lenses.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that "non-clinical studies were provided in support of the substantial equivalence determination." However, it does not specify the sample sizes used for the test set for any of the performance tests.

    Regarding data provenance, the studies were "conducted in accordance with the GLP regulation (21 CFR Part 58) or according to valid scientific protocols." This indicates that the data is from prospective studies designed to meet regulatory standards. The country of origin of the data is not specified, but given it's a US FDA submission, it's typically assumed to be generated in compliance with US regulatory frameworks.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to the type of device and study described. The performance tests for contact lenses (contact angle, water content, mechanical properties, etc.) involve objective measurements rather than subjective expert interpretation for ground truth establishment.

    4. Adjudication Method for the Test Set

    This information is not applicable as the studies are focused on objective physical and chemical properties of the contact lenses, which do not involve subjective interpretation or adjudication by multiple readers to establish ground truth.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret images, and AI might assist in that interpretation. For contact lenses, the evaluation focuses on physical, chemical, and biological properties, not human interpretation of medical images.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable as the device is a physical medical device (contact lens), not an AI algorithm. Therefore, there is no "algorithm only" performance to evaluate.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance tests on contact lenses is established through objective, standardized measurements according to the specified ANSI, ISO, and ASTM standards. For example, water content is measured using a defined gravimetric method, and mechanical properties are determined through tensile testing. For biocompatibility, chemical profiling (against ISO 10993-18:2020) was used to establish equivalence to the unmodified lenses (which presumably had established biocompatibility).

    8. The Sample Size for the Training Set

    This information is not applicable. The device is not an AI/ML algorithm that requires a training set. The studies described are traditional non-clinical performance and material characterization tests.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202756

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-10-20

    (29 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CLARITI 1 DAY (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eves that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuty.

    The CLARITI 1 DAY TORIC (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia) in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters.

    The CLARITI 1 DAY MULTIFOCAL (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical corrective ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with non-diseased eyes that may require a reading addition of +3.00 Diopters or less and may exhibit astigmatism up to 1.50 Diopters or less.

    The CLARITI 1 DAY MULTIFOCAL TORIC (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical corrective ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters and require a reading addition of +3.00 Diopters or less.

    Device Description

    The Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker is available as sphere lens, toric lens, multifocal lens, and multifocal toric lens.

    In its hydrated state, Clariti 1 day (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV Blocker when placed on the cornea acts as a refracting media to focus light rays on the retina.

    Clariti 1 day (somofilcon A) Soft (hydrophilic) Contact Lenses for Daily Wear Single Use are a hydrophilic co-polymer of silicone containing monomers and hydrophilic monomers which is cross-linked with tetraethyleneglycol dimethacrylate and di-functional methacryloxypropyl-terminated poly(dimethylsiloxane).

    When hydrated the lens consists of 44.0% somofilcon A and 56.0% water by weight of saline immersed in normal saline. A benzophenone UV absorbing monomer is used in the contact lens to help protect against transmission of harmful UV radiation and Clariti 1 day (somofilcon A) Soft contact lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

    The average transmittance characteristics are less than 5% in the UVB range of 280 to 315nm and less than 50% in the UVA range of 316-380nm.

    The lens has a hemispherical flexible shell, which covers the cornea and a portion of the adjacent sclera, with the following dimensions:
    • Chord Diameter: 13.0mm to 15.5mm
    • Centre Thickness: 0.03mm to 0.50mm
    • Base Curve: 7.5mm to 9.30mm
    • Powers: -20.00 DS to +20.00 DS
    • Toric Cylinder options: -0.75, -1.25, -1.75 and -2.25
    • Toric Axis options: 10° to 180° (10° steps)
    • Multifocal Add:
    Lens "LOW" = "low" for spectacle near ADD lens (Max +2.25 ADD)
    Lens "HIGH" = "high" for spectacle near ADD lens (+2.50 ADD or greater)

    The physical/optical properties of the lenses are:
    • Refractive Index: 1.4003
    • %Transmittance @ 590nm: 98.13
    • %Transmittance @ 280-315nm: 0.71
    • %Transmittance @ 316-380nm: 20.62
    • Surface Character: Hydrophilic
    • Water Content: 56%
    • Oxygen Permeability (DK): 60 x 10-11 (cm²/sec)(ml O2/ml x mmHg) at 35°C (Fatt Method for determination of oxygen permeability)
    • Specific Gravity: 1.17

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for a contact lens, Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker. The submission asserts substantial equivalence to a predicate device and does not involve performance testing to support this claim, but rather a change from biological release to parametric release of the finished product. Therefore, the document does not describe a study testing the acceptance criteria of the device.

    However, it does list several performance specifications/parameters for the device, which can be interpreted as acceptance criteria based on its equivalence to the predicate device.

    Here's the information derived from the document, with explanations for aspects not explicitly stated:

    1. A table of acceptance criteria and the reported device performance (based on the predicate device's performance properties):

    Acceptance Criteria (Performance Specifications/Parameters)Reported Device Performance (Subject Device)Comments
    Water Content (%)56Same as predicate
    Refractive Index1.401Same as predicate
    Oxygen Permeability (Dk @ 35°C)$60 x 10^{-11} [(cm^2/sec) x (ml O_2)/(ml x mm Hg)]$Same as predicate
    Base Curve (mm)8.6Same as predicate
    Diameter (mm)14.0Same as predicate
    Light Transmittance (%)>95Same as predicate
    Modulus (MPa)≥0.3Same as predicate
    Tensile Strength (MPa)≥0.4Same as predicate
    Elongation to Break (%)≥100Same as predicate
    Surface TreatmentNoSame as predicate
    Center Thickness (mm)Varies with powerSame as predicate
    Power Range (D)-20.00 to + 20.00Same as predicate
    UV Blocker Transmittance (UVB range of 280 to 315nm)< 5%Based on description of UV absorbing monomer
    UV Blocker Transmittance (UVA range of 316-380nm)< 50%Based on description of UV absorbing monomer

    2. Sample size used for the test set and the data provenance:

    • Not applicable. The document explicitly states: "The modification does not warrant performance testing to support substantial equivalence." This means a new performance study to test acceptance criteria was not conducted for this submission. The equivalence is based on a change in manufacturing release method (from biological to parametric).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. No new performance study was conducted.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. No new performance study was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a contact lens, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This device is a contact lens, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable. No new performance study was conducted. The "ground truth" for the device's properties is established through the physical/optical properties of the lens material and design, which are stated to be identical to the predicate device.

    8. The sample size for the training set:

    • Not applicable. No new performance study was conducted.

    9. How the ground truth for the training set was established:

    • Not applicable. No new performance study was conducted.
    Ask a Question

    Ask a specific question about this device

    K Number
    K191763

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-03-23

    (266 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MyDay (stenfilcon A) ASPHERE Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuty.

    MyDay (stenfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

    MyDay (stenfilcon A) MULTIFOCAL Soft Contact lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity.

    MyDay (stenfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optical correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eyes who may have -10.00 diopters of astigmatism or less.

    Device Description

    MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens is available as an Asphere, Toric, Multifocal, and Multifocal Toric lens designs.

    The MyDay material, stenfilcon A, is primarily a random copolymer of polydimethylsiloxane methacrylate and vinylmethyl acetamide. The UV blocker used is a benzotriazolyl methacrylate. The lenses have a blue tint which is added to make the lens more visible for handling. The lenses also contain a UV absorbing monomer which is used to block UV radiation.

    When placed on the cornea in its hydrated state, the MyDay Soft (Hydrophilic) Contact Lens acts as a refracting medium to focus light rays on the retina.

    AI/ML Overview

    This document is an FDA 510(k) clearance letter for a contact lens, not a medical device that relies on AI or complex algorithms requiring extensive performance studies with human readers or AI-specific acceptance criteria. Therefore, most of the requested information regarding acceptance criteria, study design for AI models (e.g., sample size for test/training sets, ground truth establishment, expert adjudication, MRMC studies) is not applicable to this submission.

    However, I can extract the relevant performance data and the conclusion regarding substantial equivalence for this specific device.

    Device Name: MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens (with e-PL in the packaging solution)

    Predicated Device: MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens (K131378)

    Acceptance Criteria and Reported Device Performance (Non-Clinical/Bench Testing):

    The submission aimed to demonstrate substantial equivalence to a predicate device, with the primary difference being the addition of 100 ppm Epsilon-Polylysine (e-PL) to the blister packing solution as a preservative. The acceptance criteria are implicitly met by demonstrating that the physical, mechanical, optical, and toxicological properties of the new device are equivalent to the predicate device and comply with relevant standards.

    Here's a table summarizing the types of performance testing done and the overall conclusion:

    Acceptance Criteria Category (Implicitly Met by Testing)Reported Device Performance (Conclusion)
    Physicochemical Properties:
    - Water content (per ISO18369-2:2012, ISO 18369-4:2006)Equivalent to predicate device
    - Refractive index (per ISO 18369-2:2012, ISO 16369-4:2006)Equivalent to predicate device
    - Light transmittance (per ISO 18369-2:2012, ISO 18369-3:2006)Equivalent to predicate device
    - Non-polymeric residuals in lens and packaging solution (per ISO 18369-4:2006)Equivalent to predicate device
    - Total extractable (per ANSI Z80.20-2010, ISO 18369-4:2006)Equivalent to predicate device
    - Contact angle (per ANSI Z80.20-2010)Equivalent to predicate device
    - Mechanical properties (per ANSI Z80.20-2010, ASTM D1708-02a)Equivalent to predicate device
    Preservative Efficacy:
    - Preservative efficacy testing (PET) (per ISO 14730:2014)Demonstrated function of e-PL as a preservative
    - Analytical testing (uptake and release) (per ISO 11986:2010)Demonstrated function of e-PL as a preservative
    - In-vitro microbiological testing (lens handling)Demonstrated function of e-PL as a preservative
    Biocompatibility Testing:
    - Cytotoxicity (per ISO 10993-5:2009)All biocompatibility tests passed, demonstrating equivalence.
    - Systemic Toxicity (per ISO 10993-11:2006)
    - Sensitization (per ISO 10993-10:2010)
    - Irritation (per ISO 10993-10:2010)
    - Irritation (22 Day) (per ISO 9394:2012)

    Regarding the other requested information (which is not directly applicable to this non-AI device submission):

    1. Sample size used for the test set and the data provenance: Not explicitly stated as a distinct "test set" in the context of an AI model. The bench testing data would involve samples of the manufactured contact lenses. Data provenance is implied to be from laboratory testing conducted by the manufacturer.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a contact lens's physical properties is established by objective measurements based on specified standards, not expert consensus on interpretations.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. Evaluation is done by standardized laboratory testing.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI or imaging device.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI or algorithmic device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Ground truth was established through adherence to recognized international and industry standards (ISO, ANSI, ASTM) for measuring physicochemical properties and conducting biocompatibility tests (e.g., specific gravity, refractive index, water content, cytotoxicity).
    7. The sample size for the training set: Not applicable. This device does not involve a training set as it's not an AI/machine learning product.
    8. How the ground truth for the training set was established: Not applicable.

    Overall Conclusion from the Submission:

    "Conclusive evidence was provided to demonstrate that the subject device lens material is equivalent to the currently marketed predicate device lens material through the statistical analysis of the physical/mechanical/optical properties of the lens. Based on the performance testing and the fact that the subject device has the same manufacturing process as the marketed predicate device lens, clinical performance data was not required to be submitted in this 510(k). The performance testing demonstrates that the subject device is as safe, as effective, and performs as well as the predicate device."

    Ask a Question

    Ask a specific question about this device

    K Number
    K190965

    Validate with FDA (Live)

    Device Name
    MyDay
    Manufacturer
    Date Cleared
    2019-04-29

    (17 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MyDay (stenfilcon A) ASPHERE Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuty.

    MyDay (stenfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

    MyDay (stenfilcon A) MULTIFOCAL Soft Contact lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity.

    MyDay (stenfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optical correction of distance and near vision in presbyopic phakic persons with non-diseased eyes who may have - 10.00 diopters of astigmatism or less.

    Device Description

    MyDay Contact Lenses are available as an Asphere, Toric, Multifocal, and Multifocal Toric lens designs.

    The MyDay material, stenfilcon A, is primarily a random copolymer of polydimethylsiloxane methacrylate and vinylmethyl acetamide. The UV blocker used is a benzotriazolyl methacrylate. The lenses have a blue tint which is added to make the lens more visible for handling. The lenses also contain a UV absorbing monomer which is used to block UV radiation.

    When placed on the cornea in its hydrated state, the MyDay Soft (Hydrophilic) Contact Lens acts as a refracting medium to focus light rays on the retina.

    AI/ML Overview

    The provided text is a 510(k) summary for the MyDay (stenfilcon A) Soft (hydrophilic) Daily Disposable Contact Lens. It outlines testing performed and conclusions regarding its substantial equivalence to a predicate device. However, it does not contain information about acceptance criteria and a study proving a device meets those criteria in the context of AI/ML or diagnostic performance.

    The document mainly focuses on:

    • Device Description: Material, physical parameters, and optical properties of the contact lens.
    • Performance Testing (Non-clinical): A battery of tests performed according to ANSI, ISO, and ASTM standards to characterize physical/mechanical/optical properties (e.g., contact angle, light transmittance, water content, mechanical properties, refractive index, total extractables, non-polymeric residuals).
    • Biocompatibility Testing: Extractable residual analysis to compare current and modified stenfilcon A formulations.
    • Clinical Testing: A statement that clinical performance data was not required because the subject device and predicate device are equivalent in technical characteristics, manufacturing, and sterilization processes, and the non-clinical performance testing demonstrates equivalence.

    Therefore, I cannot provide the requested information because the document describes a contact lens and its regulatory submission, not an AI/ML-driven device with diagnostic performance criteria and a study to prove it meets them.

    If this were a document about an AI/ML device, the requested information would typically include:

    1. Table of Acceptance Criteria and Reported Device Performance: This would list metrics like sensitivity, specificity, AUC, recall, precision, F1-score, accuracy, or other relevant AI performance indicators, along with the predefined acceptance thresholds and the actual performance achieved by the device in a validation study.
    2. Sample Size for Test Set and Data Provenance: Details on the number of cases/samples in the independent test set used to evaluate the AI, and information about where this data came from (e.g., specific hospitals, countries, whether it was collected retrospectively or prospectively).
    3. Number of Experts and Qualifications: How many human experts reviewed the test data to establish the ground truth, and their professional qualifications (e.g., board-certified radiologists, pathologists, clinical specialists, with specified years of experience).
    4. Adjudication Method: The process used to reconcile disagreements among experts when establishing ground truth (e.g., a "2+1" method where two experts agree, or a third expert resolves disagreements; a "3+1" method; or if no adjudication was used).
    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Whether a study was performed to compare human performance with and without AI assistance, and if so, the effect size (e.g., how much the readers' sensitivity or specificity improved, or reduction in reading time).
    6. Standalone Performance: Whether the AI algorithm's performance was evaluated independently without human intervention.
    7. Type of Ground Truth: How the definitive diagnosis or reference standard was established for the test set (e.g., consensus of multiple experts, pathological confirmation, long-term clinical outcome data, surgical findings).
    8. Sample Size for Training Set: The number of cases/samples used to train the AI model.
    9. How Ground Truth for Training Set was Established: The method used to label or annotate the data used for training the AI model (often similar to the test set ground truth establishment but might involve different scales or processes).
    Ask a Question

    Ask a specific question about this device

    K Number
    K181920

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-12-11

    (146 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CLARITI 1 DAY (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuity.

    The CLARITI 1 DAY TORIC (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia) in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters.

    The CLARITI 1 DAY MULTIFOCAL (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with non-diseased eyes that may require a reading addition of +3.00 Diopters or less and may exhibit astigmatism up to 1.50 Diopters or less.

    The CLARITI 1 DAY MULTIFOCAL TORIC (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters and require a reading addition of +3.00 Diopters or less.

    The Clarti 1 Day (somofilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear single use only. The lenses are to be discarded upon removal; therefore, no cleaning or disinfection is required.

    Device Description

    The Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker is available as sphere lens, toric lens, multifocal lens, and multifocal toric lens.

    In its hydrated state, Clariti 1 day (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV Blocker when placed on the cornea acts as a refracting media to focus light rays on the retina.

    Clariti 1 day (somofilcon A) Soft (hydrophilic) Contact Lenses for Daily Wear Single Use are a hydrophilic co-polymer of silicone containing monomers and hydrophilic monomers which is cross-linked with tetraethyleneglycol dimethacrylate and di-functional methacryloxypropyl-terminated poly(dimethylsiloxane).

    When hydrated the lens consists of 44.0% somofilcon A and 56.0% water by weight of saline immersed in normal saline. A benzophenone UV absorbing monomer is used in the contact lens to help protect against transmission of harmful UV radiation and Clariti 1 day (somofilcon A) Soft contact lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

    The average transmittance characteristics are less than 5% in the UVB range of 280 to 315nm and less than 50% in the UVA range of 316-380nm.

    The lens has a hemispherical flexible shell, which covers the cornea and a portion of the adjacent sclera, with the following dimensions:

    • Chord Diameter: 13.0mm to 15.5mm
    • Centre Thickness: 0.03mm to 0.50mm
    • Base Curve: 7.5mm to 9.30mm
    • Powers: -20.00 DS to +20.00 DS
    • Toric Cylinder options: -0.75, -1.25, -1.75 and -2.25
    • Toric Axis options: 10° to 180° (10° steps)
    • Multifocal Add: Lens "LOW" = "low" for spectacle near ADD lens (Max +2.25 ADD) Lens "HIGH" = "high" for spectacle near ADD lens (+2.50 ADD or greater)
    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance & Device Description

    The document states that the performance specifications/parameters for both the subject and predicate device are identical. Therefore, the acceptance criteria are implicitly those listed under "Performance Specifications/Parameters Comparison" in the Predicate Device column, and the reported device performance for the subject device is stated as "Same" for each parameter.

    Performance Specifications/ParametersPredicate Device (Clariti 1 day, K130331)Subject Device (Clariti 1 day, K181920)
    Water Content (%)56Same
    Refractive Index1.401Same
    Oxygen Permeability (Dk @ 35°C)60 x 10^-11 [(cm²/sec) x (ml O2)/(ml x mm Hg)]Same
    Base Curve (mm)8.6Same
    Diameter (mm)14.0Same
    Light Transmittance (%)>95Same
    Modulus (MPa)≥0.3Same
    Tensile Strength (MPa)≥0.4Same
    Elongation to Break (%)≥100Same
    Surface TreatmentNoSame
    Center Thickness (mm)Varies with powerSame
    Power Range (D)-20.00 to +20.00Same
    Additional Device Description Information (not acceptance criteria but important characteristics):
    Chord Diameter:13.0mm to 15.5mmSame (implied, as device description is identical)
    Centre Thickness:0.03mm to 0.50mmSame (implied)
    Base Curve:7.5mm to 9.30mmSame (implied)
    Powers:-20.00 DS to +20.00 DSSame (implied)
    Toric Cylinder options:-0.75, -1.25, -1.75 and -2.25Same (implied)
    Toric Axis options:10° to 180° (10° steps)Same (implied)
    Multifocal Add:"LOW" (Max +2.25 ADD), "HIGH" (+2.50 ADD or greater)Same (implied)
    %Transmittance @ 590nm:98.13Same (implied)
    %Transmittance @ 280-315nm:0.71Same (implied)
    %Transmittance @ 316-380nm:20.62Same (implied)
    Surface Character:HydrophilicSame (implied)
    Specific Gravity:1.17Same (implied)

    Key Differences Between Subject and Predicate Device:
    There were two technological differences between the subject device (K181920) and the predicate device (K130331) that required performance data:

    1. Addition of 4,4'-Diallyoxyl benzophenone (DAB): The subject device includes both UV416 and DAB as UV blockers, whereas the predicate only had UV416.
    2. Change from borate to phosphate buffered packaging solution: The subject device uses Phosphate Buffered Saline Solution containing 0.020% w/v Poloxamer 407, while the predicate used Borate Buffered Saline Solution containing 0.005% w/v Poloxamer 407.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that "Conclusive evidence was provided to demonstrate that the subject device lens material is equivalent to the currently marketed predicate device lens material through the statistical analysis of the physiochemical properties, especially water content, oxygen permeability, modulus, and toughness of the lens."

    For the performance testing (bench testing), the document mentions that "Each test was conducted according to the ANSI, ISO, and/or ASTM standard indicated." However, specific sample sizes for the test set are not provided in the given text.

    The data provenance is not explicitly stated as country of origin, but the testing was conducted in accordance with international standards (ISO, ANSI, ASTM) and FDA guidance documents (e.g., Use of International Standard ISO 10993-1, Premarket Notification [510(k)] Guidance Document for Daily Wear Contact Lenses). This implies that standard laboratory testing was performed, likely at the manufacturer's or a contract research organization's facilities. The studies appear to be prospective in nature, as they were conducted to support the substantial equivalence claim for the subject device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to this submission. The "ground truth" in this context refers to the defined physical, chemical, and optical properties of the contact lenses, as measured by standardized bench tests, and biocompatibility evaluations. These measurements do not rely on expert interpretation in the way that image-based diagnostic devices might.

    4. Adjudication Method for the Test Set

    This is not applicable. The performance testing involves direct physical, chemical, and optical measurements against established standards, not subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices where human readers interpret medical images. This submission is for a medical device (contact lens) where performance is assessed through bench and biocompatibility testing, not human interpretation of data. The document explicitly states: "clinical performance data was not required to be submitted in this 510(k)."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable. This is not an algorithm or AI-powered device. The device's performance is assessed through its physical characteristics and biological safety.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance criteria was established through:

    • Bench Testing: Measurements of physical, chemical, and optical properties (e.g., water content, oxygen permeability, refractive index, mechanical properties, light transmittance, specific gravity, contact angle, total extractables, non-polymeric residuals) using established ANSI, ISO, and ASTM standards.
    • Biocompatibility Testing: Evaluation according to ISO 10993 series for cytotoxicity, irritation, systemic toxicity, and sensitization.
    • Comparison to Predicate: The key ground truth was the established performance of the legally marketed predicate device (Clariti 1 day, K130331), demonstrating that the subject device's performance was "Same" or equivalent across all measured parameters despite minor material and packaging solution changes.

    8. The Sample Size for the Training Set

    This is not applicable. This submission is for a physical medical device (contact lens), not an AI algorithm. Therefore, there is no "training set" in the context of machine learning.

    9. How the Ground Truth for the Training Set was Established

    This is not applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160803

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-07-13

    (112 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sphere/Asphere: AVAIRA VITALITY SPHERE and ASPHERE (fanfilcon A) Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

    Toric: AVAIRA VITALITY (fanfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and non- aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

    Multifocal: AVAIRA VITALITY (fanfilcon A) MULTIFOCAL Soft lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with add powers from +0.25 to +4.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

    Multifocal Toric: AVAIRA VITALITY (fanfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optic correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eyes in powers of -20.00 to +20.00 diopters with add powers from +0.25 to +4.00 diopters and astigmatism corrections from -0.25 to -10.00 diopters.

    Eye Care Practitioners may prescribe the Avaira Vitality (fanfilcon A) Soft Contact lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement or for single-use disposable wear.

    When prescribed for frequent/planned replacement, the Avaira Vitality (fanfilcon A) Soft Contact lens is to be cleaned, rinsed and disinfected each time the lens is removed. The contact lens is to be discarded after the recommended wearing period as prescribed by the Eye Care Professional. When prescribed for frequent/ planned replacement wear, the lenses may be disinfected using a chemical disinfection only.

    When prescribed for single-use disposable wear, the Avaira Vitality (fanfilcon A) Soft Contact lens is to be discarded after each removal.

    Device Description

    The AVAIRA VITALITY (fanfilcon A) Contact Lens visibility tinted with UV blocker is available as a sphere/asphere lens, toric lens, multifocal lens, and multifocal toric lens. The lenses are made of a silicone hydrogel material which is not surface treated and is characterized by high oxygen permeability (Dk). The AVAIRA VITALITY (fanfilcon A)) Contact Lens is tinted blue using Reactive Blue #246 to make the lens more visible for handling. A Norbloc UV blocker is used to reduce the amount of ultraviolet light transmitted into the eye. The AVAIRA VITALITY (fanfilcon A) lens is supplied sterile, packaged in a buffered saline solution. The composition of the lens is 45% fanfilcon A and 55% water by weight when hydrated and stored in buffered saline solution.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device (contact lenses), not an algorithm-driven AI/ML device. Therefore, the detailed requirements regarding AI/ML device acceptance criteria, study design, expert involvement, and ground truth establishment are not applicable.

    However, I can extract and structure the information related to the device's assessment for its intended purpose, which is proving substantial equivalence to predicate devices for contact lenses.

    Here's an attempt to adapt the requested structure as much as possible, reinterpreting the questions for a non-AI/ML medical device context:

    Device: Avaira Vitality (fanfilcon A) Soft (hydrophilic) Contact Lenses


    1. Table of Acceptance Criteria and the Reported Device Performance

    The concept of "acceptance criteria" for this contact lens is substantial equivalence to legally marketed predicate devices, meaning it is as safe and effective. The performance is assessed through non-clinical (material properties, toxicology, biocompatibility) and clinical testing.

    Acceptance Criteria (Reinterpreted for Contact Lenses)Reported Device Performance (Summary)
    Non-clinical:
    Material Properties Comparison (Predicate 1: CooperVision AVAIRA (enfilcon A))Water Content: 55% (vs 46% for predicate) Refractive Index @ 20℃: 1.40 (vs 1.40) Specific Gravity g/mL: 1.026 (vs 1.06) Oxygen Permeability (Dk): 90 (vs 100) Base Curve, mm: 8.4 (vs 8.5) Diameter, mm: 14.2 (vs 14.2)
    Material Properties Comparison (Predicate 2: CooperVision MyDay (stenfilcon A))Water Content: 55% (vs 54% for predicate) Refractive Index @ 20℃: 1.40 (vs 1.40) Specific Gravity g/mL: 1.026 (vs 1.033) Oxygen Permeability (Dk): 90 (vs 80) Base Curve, mm: 8.4 (vs 8.4) Diameter, mm: 14.2 (vs 14.2)
    In-vitro & In-vivo Preclinical Toxicology & Biocompatibility (General)Lens material and/or extracts are non-toxic, non-irritating, and non-sensitizing. Lens physical and material properties are consistent with currently marketed lenses.
    Clinical:
    Safety (Slit lamp findings, Adverse Event rates)Found to be substantially equivalent to the Control contact lens for safety.
    Efficacy (Visual Acuity Comparisons)Found to be substantially equivalent to the Control contact lens for efficacy.
    Secondary Measures (Average lens wearing times, Subjective lens comfort assessed by frequency or symptoms)Not explicitly detailed if these met specific numeric "acceptance criteria," but they were assessed as secondary outcomes.
    Overall Substantial EquivalenceThe device is as safe and effective as the predicate device when used as directed.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size for Clinical Study (Test Set): 90 subjects (60 subjects for Test lenses, 30 subjects for Control lenses).
    • Data Provenance: A multi-center clinical study (exact geographic location not specified, but typically US-based for FDA submissions). The study was conducted retrospectively to collect the data from the clinical trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • This is not applicable in the context of a contact lens study seeking substantial equivalence. The "ground truth" for safety and efficacy in this clinical trial would be established by the clinical findings (e.g., slit lamp examinations, visual acuity measurements) as assessed by the participating clinical investigators/ophthalmic professionals, not by a multi-expert panel establishing a "ground truth" for diagnostic accuracy as for an AI/ML algorithm.

    4. Adjudication Method for the Test Set

    • Not explicitly described as an "adjudication method" in the AI/ML sense. The clinical study was a "randomized, masked, concurrent control study." This implies that the assessment of primary and secondary outcomes was done by investigators blinded to the lens type where possible, and comparisons were made between the test and control groups.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • Not applicable. This is not an AI/ML device, and no MRMC study or human-in-the-loop performance evaluation was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. This is a contact lens, not an algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    • In the context of this contact lens study, the "ground truth" for determining safety and efficacy was based on clinical outcomes data (e.g., slit lamp findings, adverse event rates, visual acuity measurements) observed in the study subjects. This data was collected by the clinical investigators involved in the multi-center trial.

    8. The Sample Size for the Training Set

    • Not applicable. This is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. This is not an AI/ML device that requires a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K133627

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-12-23

    (27 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AVAIRA (enfilcon A) SPHERE Soft (Hydrophilic) Contact Lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

    AVAIRA (enfilcon A) TORIC Soft (Hydrophilic) Contact Lenses are indicated for the correction of visual ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

    DAILY DISPOSABLE:

    The AVAIRA (enfilcon A) Soft (hydrophilic) Contact Lenses are indicated for single-use disposable wear.

    FREQUENT REPLACEMENT:

    The AVAIRA (enfilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear. When prescribed for planned replacement the lens may be disinfected using a chemical or hydrogen peroxide disinfecting system.

    Device Description

    The Avaira (enfilcon A) soft contact lens is a daily wear silicone hydrogel contact lens that is not surface treated and is characterized by a high oxygen permeability (Dk). The lens material, enfilcon A, is composed of silicone macromers cross linked with other monomers, incorporating phthalocyanine blue as an integrated handling tint. A UV blocker is added to reduce the amount of ultraviolet light transmitted into the eye. The Avaira (enfilcon A) Soft (hydrophilic) contact lenses are a hemispherical shell. This 510(k) submission covers a process modification to add a step to the preparation of the monomer mixture. The physical properties and dimensions of the finished lenses are unchanged from predicate 510(k)s.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study conducted for the CooperVision AVAIRA (enfilcon A) Sphere and Toric Contact Lenses.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance:

    Tested CharacteristicAcceptance Criteria (Implied: "Same as predicate" or within acceptable range)Reported Device Performance (Results)
    Refractive Index1.40 (Predicate)Pass
    Total ExtractablesNot explicitly stated, implied to be within acceptable rangePass
    Monomer ResidualsNot explicitly stated, implied to be within acceptable rangePass
    IonofluxNot explicitly stated, implied to be within acceptable rangePass
    Contact AngleNot explicitly stated, implied to be within acceptable rangePass
    UV TransmissionNot explicitly stated, implied to be within acceptable rangePass
    ModulusNot explicitly stated, implied to be within acceptable rangePass
    ElongationNot explicitly stated, implied to be within acceptable rangePass
    Tensile StrengthNot explicitly stated, implied to be within acceptable rangePass
    Power ConformanceNot explicitly stated, implied to be within acceptable rangePass
    For Substantial Equivalence Comparison (showing alignment with predicate devices):
    Indications for Use (Sphere)Correction of ametropia in aphakic and non-aphakic persons with non-diseased eyes and astigmatism ≤2.00 dioptersSame
    Indications for Use (Toric)Correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic correction from -0.25 to -10.00 dioptersSame
    Single UseSingle-use, disposableSame
    ReplacementDaily Wear, disinfect using chemical or hydrogen peroxide disinfectant when prescribed for planned replacementSame
    Material USAN Nameenfilcon ASame
    Principle of OperationDesigned to fit over corneal surface of eye to provide corrective refraction for functional conditions of the eyeSame
    Refractive Index1.40Same
    Oxygen Permeability x 10 -11100 (cm²/sec)(ml O₂)/ml*mmHg)Same
    Base Curve8.2 to 9.2 mmSame
    Diameter13.5 to 15.0 mmSame
    Power Range-20.00 to +20.00 DSame
    Cylinder Power-0.25 to -10.00 DSame
    Water Content (avg, %wt)46%46%
    Monomer sonication stepNoYes

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: The document does not specify a distinct "test set" sample size for clinical studies because it explicitly states that "In-vivo clinical studies were not required for this change." The evaluation was based on physiochemical studies.
    • Data Provenance: The physiochemical studies were "conducted by CooperVision following scientific protocols." No specific country of origin or whether it was retrospective or prospective is mentioned, but it implies internal, prospective laboratory testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Since in-vivo clinical studies were not required and the assessment was based on physiochemical properties, there was no "ground truth" established by experts in a clinical context for a test set. The ground truth was based on physical and chemical measurements against established specifications for contact lenses and comparison to predicate devices.

    4. Adjudication method for the test set:

    • Not applicable. No clinical test set involving human judgment was used. The physiochemical studies would involve laboratory measurements conforming to predefined technical specifications.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This submission is for contact lenses, not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This submission is for contact lenses, not an algorithm.

    7. The type of ground truth used:

    • The "ground truth" for the physiochemical studies was based on physical and chemical measurements of the lens properties (e.g., refractive index, oxygen permeability, water content, modulus, etc.) against established scientific standards and comparison to the characteristics of the predicate device (AVAIRA Sphere and Toric Soft Contact Lenses, K071736/K113759).

    8. The sample size for the training set:

    • Not applicable. This submission is for a physical medical device (contact lenses), not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. As above, no training set was used.
    Ask a Question

    Ask a specific question about this device

    K Number
    K131378

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-08-30

    (108 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SUS (stenfilcon A) ASPHERE Soft Contact Lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00D to +20.00D diopters. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity.
    SUS (stenfilcon A) TORIC Soft Contact Lenses are indicated for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.
    SUS (stenfilcon A) MULTIFOCAL Soft Contact lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity.
    SUS (stenfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optical correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eyes who may have -10.00 diopters of astigmatism or less.

    Device Description

    The SUS (stenfilcon A) Contact Lens Visibility Tinted with UV Blocker is available as an aspherical lens, toric lens, multifocal lens, and multifocal toric lens. The lenses are made of a silicone hydrogel material which is not surface treated and is characterized by high oxygen permeability (Dk). The SUS (stenfilcon A) Contact Lens is tinted blue using Reactive Blue #246 to make the lens more visible for handling. A Norbloc UV Blocker is used to reduce the amount of ultraviolet light transmitted into the eye. The SUS (stenfilcon A) lens is supplied sterile, packaged in a buffered saline solution. The composition of the lens is 46% stenfilcon A and 54% water by weight when hydrated and stored in buffered saline solution.

    AI/ML Overview

    The acceptance criteria for the CooperVision SUS (stenfilcon A) Soft Contact Lens and the study proving it meets these criteria are detailed below:

    1. Acceptance Criteria and Reported Device Performance

    The device's performance was evaluated by comparing its material and physical properties to two predicate devices (CooperVision AVAIRA and CooperVision CLEARSIGHT 1 DAY), and through a clinical study for safety and efficacy.

    Material Comparison (vs. CooperVision AVAIRA (enfilcon A))

    Material ComparisonPredicate Device (CooperVision AVAIRA (enfilcon A))Subject Device (CooperVision SUS (stenfilcon A))Performance
    Material USAN NameEnfilcon AStenfilcon ADifferent materials, but both silicone hydrogel, deemed acceptable.
    FDA CategorySilicone HydrogelSilicone HydrogelSame classification.
    Manufacturing MethodMoldedMoldedSame.
    SterilizationMoist HeatMoist HeatSame.
    PackagingBlisterBlisterSame.
    Visibility TintPhthalocyanine BlueReactive Blue #246Different tint compounds, but both for visibility.

    Parameter Comparison (vs. CooperVision AVAIRA (enfilcon A))

    Parameter ComparisonPredicate Device (Avaira (enfilcon A)) - MeasuredPredicate Device (Avaira (enfilcon A)) - LabeledSubject Device (SUS (stenfilcon A)) - MeasuredSubject Device (SUS (stenfilcon A)) - LabeledPerformance
    Water Content, %45%46%54.3%54%Different, but within expected range for contact lenses.
    Refractive Index @ 20°C1.4021.401.40111.40Similar.
    Oxygen Permeability (Dk)99.610079.280Permeability is lower but still within general silicone hydrogel range.
    Base Curve, mm8.458.58.338.4Similar.
    Diameter, mm14.2614.214.1214.2Similar.

    Material Comparison (vs. CooperVision CLEARSIGHT 1 DAY (ocufilcon B))

    Material ComparisonPredicate Device (CooperVision CLEARSIGHT 1 DAY (ocufilcon B))Subject Device (CooperVision SUS (stenfilcon A))Performance
    Material USAN NameOcufilcon BStenfilcon ADifferent materials.
    FDA CategoryGroup IVSilicone HydrogelDifferent categories, subject device is silicone hydrogel.
    Manufacturing MethodMoldedMoldedSame.
    SterilizationMoist HeatMoist HeatSame.
    PackagingBlisterBlisterSame.
    Visibility TintVat Blue 6Reactive Blue #246Different tint compounds.

    Parameter Comparison (vs. CooperVision CLEARSIGHT 1 DAY (ocufilcon B))

    Parameter ComparisonPredicate Device (CLEARSIGHT 1 DAY (ocufilcon B)) - MeasuredPredicate Device (CLEARSIGHT 1 DAY (ocufilcon B)) - LabeledSubject Device (SUS (stenfilcon A)) - MeasuredSubject Device (SUS (stenfilcon A)) - LabeledPerformance
    Water Content, %51.0452%54.3%54%Similar or slightly higher.
    Refractive Index @ 20°C1.4151.411.40111.40Similar.
    Oxygen Permeability (Dk)16.1716.879.280Significantly higher (a benefit for silicone hydrogel).
    Base Curve, mm8.858.78.338.4Similar.
    Diameter, mm14.1214.214.1214.2Same.

    Acceptance Criteria for Clinical Study and Reported Performance:

    The primary acceptance criteria for the clinical study were the evaluation of biomicroscopy and adverse event rates, and visual acuity comparisons, with secondary measures including wearing times and subjective comfort. The reported performance was:

    • "The Test contact lens was found to be substantially equivalent to the Control contact lens for safety and efficacy."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Study (Test Set): 90 male and female subjects.
      • 60 subjects were dispensed the Test lenses (SUS (stenfilcon A)).
      • 30 subjects were dispensed the Control lenses (unspecified control contact lens, likely one of the predicate devices or a similar established lens).
    • Data Provenance: Not explicitly stated regarding country of origin, but the study was conducted by CooperVision, Inc., a US-based company, and submitted to the US FDA. The study was prospective in nature through a clinical trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. The clinical study details do not specify the number or qualifications of experts involved in establishing ground truth (e.g., ophthalmologists or optometrists assessing biomicroscopy findings).

    4. Adjudication Method for the Test Set

    The adjudication method for the clinical study is not explicitly detailed in the provided text. It mentions "The primary outcome measures were biomicroscopy and adverse event rates along with lens visual acuity comparisons" without specifying how disagreements or interpretations among clinical evaluators were resolved.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The clinical study described is for evaluating the safety and efficacy of the contact lens itself, not a comparison of human readers with and without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. This is a medical device (contact lens), not an AI algorithm. Therefore, no standalone algorithm performance was assessed.

    7. The Type of Ground Truth Used

    For the non-clinical testing, the "ground truth" was established based on biocompatibility and toxicology standards (GLP regulation - 21 CFR Part 56) and established scientific protocols for physical and material property measurements.

    For the clinical study, the "ground truth" for efficacy and safety was based on:

    • Clinical observations from biomicroscopy.
    • Adverse event reporting.
    • Visual acuity comparisons.
    • Subjective assessments of comfort and wearing times.
      This could be considered a form of expert clinical assessment/observation and patient-reported outcomes.

    8. The Sample Size for the Training Set

    Not applicable. This is a medical device (contact lens) and not an AI or machine learning model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123202

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-05-30

    (230 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Proclear Multifocal Toric (omaficon B) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes that are myopic or hyperopic which, possess astigmatism to -5.75 diopters or less, and are presbyopic.

    Proclear Multifocal XR (omafilcon B) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes that are myopic or hyperopic and are presbyopic. The lenses may be worn by persons who exhibit astigmatism of 0.75 diopters or less that does not interfere with visual acuity.

    Proclear Toric XR (omafilcon B) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with nondiseased eyes that are myopic or hyperopic which possess astigmatism corrections to -5.75 diopters.

    Proclear Sphere and Asphere: (omafilcon B) Soft (hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in non-aphakic persons with non- diseased eyes that are myopic or hyperopic and exhibit astigmatism of 2.00D or less that does not interfere with visual acuity.

    Proclear. Toric: (omafilcon B) Soft (hydrophilic) Contact Lenses are indicated for daily wear for-the-correction-of-visual-acuity-in-non-aphakic-persons-with-non-diseased-eves-thatare myopic or hyperopic .- The lens may be worn by persons who - have astigmatism of 5.00D or less.

    Proclear Multifocal: (omafilcon B) Soft (hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, and astigmatism) and presbyopia in non-aphakic persons with non-diseased eyes. The lens may be worn by persons who have astigmatism of 2.00D or less that does not interfere with visual acuity.

    Proclear (omafilcon B) Soft (hydrophilic) Contact Lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only).

    Daily wear replacement schedules may vary from patient and should be decided by eye care practitioners in consultation with their patients.

    Device Description

    The Proclear lens is composed of polymer of 2-hydroxy-ethylmethacrylate and 2-metacryloloyoxyethyl phosphorylcholine cross linked with ethylene glycol dimethacrylate. The lenses are tinted blue from edge to edge for visibility purposes. The Proclear (omafilcon B) Soft (hydrophilic) contact lenses are a hemispherical shell. The design and toxicological properties of the devices with the modified formula to increase the water content to 62% are unchanged from predicate 510ks.

    Proclear Toric and Proclear Toric XR (omafilcon B) Soft (Hydrophilic) Contact Lenses are back surface toric.

    Proclear Multifocal and Proclear Multifocal XR (omafilcon B) Soft(Hydrophilic) Contact Lenses are available as a multifocal lens with an aspherical front surface and spherical back surface for the correction of visual acuity in presbyopic persons who are myopic or hyperopic. The multifocal lens has two multifocal zones, as well as the edge shape being optimized to provide comfort without sacrificing tensile strength.

    Proclear Multifocal Toric (omafilcon B) Soft (Hydrophilic) Contact Lens front surface is aspherical, with the anterior surface having a toric generated surface for the purpose of correcting vision in an eye that is astigmatic.

    Proclear Sphere and Asphere: (omofilcon B) Soft (hydrophilic) Contact Lenses. The sphere lenses have spherical optical zone and asphere lens have an aspherical optical zone. This aspheric optical zone design (front surface) controls and limits the amount of longitudinal spherical aberration generated by the lens across the power range.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the CooperVision Proclear (omafilcon B) Soft (Hydrophilic) Contact Lenses, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Tested CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    Total Extractables, Water Content, Dk, Light Transmission, Refractive Index (per ISO 18369-4:2006 Ophthalmic Optics – Contact Lenses)Not explicitly statedPass
    CytotoxicityNo toxicity issuesNo toxicity issues
    Ocular IrritationNo toxicity issuesNo toxicity issues
    Systemic ToxicityNo toxicity issuesNo toxicity issues

    Note: The document states that the physicochemical properties were tested and "indicate no significant change" from predicate devices. The acceptance criteria for these tests are implicitly that they meet the standards for contact lens materials as per ISO 18369-4:2006, and that the results are comparable to the predicate devices. The "Pass" result signifies compliance with these implicit criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    • For Physicochemical Studies: Not explicitly stated. The studies were "conducted by CooperVision following scientific protocols."
    • For Biocompatibility Studies: Not explicitly stated. The studies were "conducted by contract laboratories under Good Manufacturing Practice regulations." Data provenance (country of origin) is not mentioned. These studies were retrospective in the sense that they were conducted on already manufactured materials.
    • No clinical studies were conducted. Therefore, there is no test set of human subjects or associated data provenance for clinical performance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not applicable as no clinical studies were conducted involving human subjects to establish ground truth for clinical performance.
    • For physicochemical and biocompatibility tests, "ground truth" is established by the standardized methods themselves and the expertise of the laboratory personnel performing the tests, though their specific qualifications or numbers are not mentioned.

    4. Adjudication Method for the Test Set

    • Not applicable as no clinical studies were conducted that would require an adjudication method for human reader/expert review.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    • Not applicable. This device is a contact lens, not an AI-assisted diagnostic or therapeutic tool. No MRMC comparative effectiveness study was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. This device is a contact lens. No standalone algorithm performance was assessed.

    7. The Type of Ground Truth Used

    • For Physicochemical Studies: The ground truth is established by standardized laboratory measurements and accepted scientific principles as defined in ISO 18369-4:2006, compared against pre-defined specifications and predicate device characteristics.
    • For Biocompatibility Studies: The ground truth is established by recognized toxicity endpoints and established scientific guidelines for cytotoxicity, ocular irritation, and systemic toxicity, which determine the safety of the material in biological systems.

    8. The Sample Size for the Training Set

    • Not applicable, as this is not an AI/machine learning device. The concept of a "training set" in this context is irrelevant. For the material properties, the "training" analogous to material science would involve R&D and historical data on similar materials, but no specific "training set" in the computational sense is mentioned.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as this is not an AI/machine learning device and thus has no "training set" in that sense. The specifications for the material and its properties were likely established through R&D, industry standards, and regulatory requirements based on the predicate devices.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 4