K Number
K133627

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-12-23

(27 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AVAIRA (enfilcon A) SPHERE Soft (Hydrophilic) Contact Lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

AVAIRA (enfilcon A) TORIC Soft (Hydrophilic) Contact Lenses are indicated for the correction of visual ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

DAILY DISPOSABLE:

The AVAIRA (enfilcon A) Soft (hydrophilic) Contact Lenses are indicated for single-use disposable wear.

FREQUENT REPLACEMENT:

The AVAIRA (enfilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear. When prescribed for planned replacement the lens may be disinfected using a chemical or hydrogen peroxide disinfecting system.

Device Description

The Avaira (enfilcon A) soft contact lens is a daily wear silicone hydrogel contact lens that is not surface treated and is characterized by a high oxygen permeability (Dk). The lens material, enfilcon A, is composed of silicone macromers cross linked with other monomers, incorporating phthalocyanine blue as an integrated handling tint. A UV blocker is added to reduce the amount of ultraviolet light transmitted into the eye. The Avaira (enfilcon A) Soft (hydrophilic) contact lenses are a hemispherical shell. This 510(k) submission covers a process modification to add a step to the preparation of the monomer mixture. The physical properties and dimensions of the finished lenses are unchanged from predicate 510(k)s.

AI/ML Overview

The provided text describes the acceptance criteria and the study conducted for the CooperVision AVAIRA (enfilcon A) Sphere and Toric Contact Lenses.

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance:

Tested CharacteristicAcceptance Criteria (Implied: "Same as predicate" or within acceptable range)Reported Device Performance (Results)
Refractive Index1.40 (Predicate)Pass
Total ExtractablesNot explicitly stated, implied to be within acceptable rangePass
Monomer ResidualsNot explicitly stated, implied to be within acceptable rangePass
IonofluxNot explicitly stated, implied to be within acceptable rangePass
Contact AngleNot explicitly stated, implied to be within acceptable rangePass
UV TransmissionNot explicitly stated, implied to be within acceptable rangePass
ModulusNot explicitly stated, implied to be within acceptable rangePass
ElongationNot explicitly stated, implied to be within acceptable rangePass
Tensile StrengthNot explicitly stated, implied to be within acceptable rangePass
Power ConformanceNot explicitly stated, implied to be within acceptable rangePass
For Substantial Equivalence Comparison (showing alignment with predicate devices):
Indications for Use (Sphere)Correction of ametropia in aphakic and non-aphakic persons with non-diseased eyes and astigmatism ≤2.00 dioptersSame
Indications for Use (Toric)Correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic correction from -0.25 to -10.00 dioptersSame
Single UseSingle-use, disposableSame
ReplacementDaily Wear, disinfect using chemical or hydrogen peroxide disinfectant when prescribed for planned replacementSame
Material USAN Nameenfilcon ASame
Principle of OperationDesigned to fit over corneal surface of eye to provide corrective refraction for functional conditions of the eyeSame
Refractive Index1.40Same
Oxygen Permeability x 10 -11100 (cm²/sec)(ml O₂)/ml*mmHg)Same
Base Curve8.2 to 9.2 mmSame
Diameter13.5 to 15.0 mmSame
Power Range-20.00 to +20.00 DSame
Cylinder Power-0.25 to -10.00 DSame
Water Content (avg, %wt)46%46%
Monomer sonication stepNoYes

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: The document does not specify a distinct "test set" sample size for clinical studies because it explicitly states that "In-vivo clinical studies were not required for this change." The evaluation was based on physiochemical studies.
  • Data Provenance: The physiochemical studies were "conducted by CooperVision following scientific protocols." No specific country of origin or whether it was retrospective or prospective is mentioned, but it implies internal, prospective laboratory testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. Since in-vivo clinical studies were not required and the assessment was based on physiochemical properties, there was no "ground truth" established by experts in a clinical context for a test set. The ground truth was based on physical and chemical measurements against established specifications for contact lenses and comparison to predicate devices.

4. Adjudication method for the test set:

  • Not applicable. No clinical test set involving human judgment was used. The physiochemical studies would involve laboratory measurements conforming to predefined technical specifications.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This submission is for contact lenses, not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This submission is for contact lenses, not an algorithm.

7. The type of ground truth used:

  • The "ground truth" for the physiochemical studies was based on physical and chemical measurements of the lens properties (e.g., refractive index, oxygen permeability, water content, modulus, etc.) against established scientific standards and comparison to the characteristics of the predicate device (AVAIRA Sphere and Toric Soft Contact Lenses, K071736/K113759).

8. The sample size for the training set:

  • Not applicable. This submission is for a physical medical device (contact lenses), not a machine learning or AI device that requires a training set.

9. How the ground truth for the training set was established:

  • Not applicable. As above, no training set was used.

{0}------------------------------------------------

CooperVision, Inc.

K133627

510k Summary AVAIRA (enfilcon A) Sphere and Toric Contact Lenses 12/20/2013

DEC 2 3 2013

Applicant:CooperVision, Inc.6150 Stoneridge Mall Road, Suite 370Pleasanton, CA 94588 USA
Applicant Contact:Annette NelsonSenior Regulatory Affairs Specialist925-621-2453 (phone)925-621-2488 (fax)
Email:anelson@coopervision.com
Date Prepared:November 15, 2013
Device Trade Name:AVAIRA Sphere and Toric Soft Contact Lenses
Common/Usual Name:Enfilcon A Soft (Hydrophilic) Contact Lens
Classification Name:Soft (Hydrophilic) Contact Lens, Daily Wear, Disposable
Device Classification:Class II (21 CFR 886.5925); Product Codes MVN and LPL
Predicate Devices:AVAIRA Sphere and Toric Soft Contact Lenses(K071736/K113759)

Device Description:

The Avaira (enfilcon A) soft contact lens is a daily wear silicone hydrogel contact lens that is not surface treated and is characterized by a high oxygen permeability (Dk). The lens material, enfilcon A, is composed of silicone macromers cross linked with other monomers, incorporating phthalocyanine blue as an integrated handling tint. A UV blocker is added to reduce the amount of ultraviolet light transmitted into the eye. The Avaira (enfilcon A) Soft (hydrophilic) contact lenses are a hemispherical shell. This 510(k) submission covers a process modification to add a step to the preparation of the monomer mixture. The physical properties and dimensions of the finished lenses are unchanged from predicate 510(k)s.

Intended Use:

SPHERICAL:

Avaira (enfilcon A) SPHERE Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes. The

Cooper Vision, Inc. AVAIRA (enfilcon A) Contact Lens 510(k)

{1}------------------------------------------------

lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

TORIC:

Avaira (enfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

Physiochemical Studies

In-vitro studies measuring the physical, optical and chemical properties of lenses subjected to the modified manufacturing process were performed. The results show there is no significant change to the physicochemical properties of the lenses.

Tested CharacteristicsResults
Refractive Index, Total Extractables, MonomerResiduals, Ionoflux, Contact Angle, UV Transmission,Modulus, Elongation, Tensile Strength, and PowerConformancePass

Clinical Studies

In-vivo clinical studies were not required for this change, as the dimensions, formulation, lens manufacturing process, and product specifications remain the same.

Conclusion

Validity of Scientific Data

Physiochemical studies were conducted by CooperVision following scientific protocols.

Substantial Equivalence

The information presented in this submission establishes the substantial equivalence of the modified AVAIRA (enfilcon A) SPHERE and TORIC daily wear contact lenses to the predicate devices. The following table summarizes the substantial equivalence comparison information.

{2}------------------------------------------------

:

and the comments of the comments of

.

:

. . . .

Substantial Equivalence Comparison
CharacteristicAVAIRA SPHERE and TORIC(K071736/K113759)AVAIRA SPHERE and TORIC(This Submission)
Indications for UseAVAIRA SPHERECorrection of ametropia in aphakicand non-aphakic persons with non-diseased eyes and astigmatism≤2.00 dioptersSame
Indications for UseAVAIRA TORICCorrection of ametropia (myopiaor hyperopia with astigmatism) inaphakic and non-aphakic personswith non-diseased eyes in powersfrom -20.00 to +20.00 diopters andastigmatic correction from -0.25 to-10.00 dioptersSame
Single UseSingle-use, disposableSame
ReplacementDaily Wear, disinfect usingchemical or hydrogen peroxidedisinfectant when prescribed forplanned replacementSame
Material USAN Nameenfilcon ASame
Principle of OperationDesigned to fit over cornealsurface of eye to providecorrective refraction forfunctional conditions of the eyeSame
Refractive Index1.40Same
Oxygen Permeability x 10-11100 (cm²/sec)(ml O₂)/ml*mmHg)Same
Base Curve8.2 to 9.2 mmSame
Diameter13.5 to 15.0 mmSame
Power Range-20.00 to +20.00 DSame
Cylinder Power-0.25 to -10.00 DSame
Water Content (avg, %wt)46%46%
Monomer sonication stepNoYes

.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows a partial view of a seal or emblem. The visible text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the edge of the emblem. The central part of the emblem features a stylized design, possibly representing a caduceus or a similar symbol associated with health and medicine. The image is in black and white, with the text and design elements appearing in solid black against a white background.

December 23, 2013

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

CooperVision, Inc. % Ms. Annette Nelson Senior Regulatory Affairs Specialist 6150 Stoneridge Mall Road, Suite 370 Pleasanton, CA 94588

Re: K133627

Trade/Device Name: AVAIRA Sphere and Toric Soft Contact Lenses Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL Secondary Product Code: MVN Dated: November 15, 2013 Received: November 26, 2013

Dear Ms. Nelson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{4}------------------------------------------------

Page 2 - Ms. Annette Nelson

You must comply with all the Act's requirements, including, but not limited to: Registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Deborah L. Falls -S

for Malvina B. Eydelman. M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

CooperVision, Inc.

Indications for Use

510(k) Number: K133627

Device Name: AVAIRA (enfilcon A) Sphere and Toric Soft Contact Lenses

Indications For Use:

A VAIRA (enfilcon A) SPHERE Soft (Hydrophilic) Contact Lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

AVAIRA (enfilcon A) TORIC Soft (Hydrophilic) Contact Lenses are indicated for the correction of visual ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

DAILY DISPOSABLE:

The AVAIRA (enfilcon A) Soft (hydrophilic) Contact Lenses are indicated for single-use disposable wear.

FREQUENT REPLACEMENT:

The AVAIRA (enfilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear. When prescribed for planned replacement the lens may be disinfected using a chemical or hydrogen peroxide disinfecting system.

Prescription Use X ______________________________________________________________________________________________________________________________________________________________________________ (part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Jeffrey M. Brocious -S 2013.12.20 13:46:29 -05'00'

(Division Sign-Off) Division of Ophthalmic and Ear, Nose, and Throat Devices

510(k) Number: K133627

Page 1 of

K133627

Cooper Vision Inc., AVAIRA (enfilcon A) Contact Lens 510(k)

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.