K Number
K202756

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-10-20

(29 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CLARITI 1 DAY (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eves that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuty.

The CLARITI 1 DAY TORIC (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia) in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters.

The CLARITI 1 DAY MULTIFOCAL (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical corrective ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with non-diseased eyes that may require a reading addition of +3.00 Diopters or less and may exhibit astigmatism up to 1.50 Diopters or less.

The CLARITI 1 DAY MULTIFOCAL TORIC (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical corrective ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters and require a reading addition of +3.00 Diopters or less.

Device Description

The Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker is available as sphere lens, toric lens, multifocal lens, and multifocal toric lens.

In its hydrated state, Clariti 1 day (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV Blocker when placed on the cornea acts as a refracting media to focus light rays on the retina.

Clariti 1 day (somofilcon A) Soft (hydrophilic) Contact Lenses for Daily Wear Single Use are a hydrophilic co-polymer of silicone containing monomers and hydrophilic monomers which is cross-linked with tetraethyleneglycol dimethacrylate and di-functional methacryloxypropyl-terminated poly(dimethylsiloxane).

When hydrated the lens consists of 44.0% somofilcon A and 56.0% water by weight of saline immersed in normal saline. A benzophenone UV absorbing monomer is used in the contact lens to help protect against transmission of harmful UV radiation and Clariti 1 day (somofilcon A) Soft contact lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

The average transmittance characteristics are less than 5% in the UVB range of 280 to 315nm and less than 50% in the UVA range of 316-380nm.

The lens has a hemispherical flexible shell, which covers the cornea and a portion of the adjacent sclera, with the following dimensions:
• Chord Diameter: 13.0mm to 15.5mm
• Centre Thickness: 0.03mm to 0.50mm
• Base Curve: 7.5mm to 9.30mm
• Powers: -20.00 DS to +20.00 DS
• Toric Cylinder options: -0.75, -1.25, -1.75 and -2.25
• Toric Axis options: 10° to 180° (10° steps)
• Multifocal Add:
Lens "LOW" = "low" for spectacle near ADD lens (Max +2.25 ADD)
Lens "HIGH" = "high" for spectacle near ADD lens (+2.50 ADD or greater)

The physical/optical properties of the lenses are:
• Refractive Index: 1.4003
• %Transmittance @ 590nm: 98.13
• %Transmittance @ 280-315nm: 0.71
• %Transmittance @ 316-380nm: 20.62
• Surface Character: Hydrophilic
• Water Content: 56%
• Oxygen Permeability (DK): 60 x 10-11 (cm²/sec)(ml O2/ml x mmHg) at 35°C (Fatt Method for determination of oxygen permeability)
• Specific Gravity: 1.17

AI/ML Overview

The provided document describes a 510(k) premarket notification for a contact lens, Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker. The submission asserts substantial equivalence to a predicate device and does not involve performance testing to support this claim, but rather a change from biological release to parametric release of the finished product. Therefore, the document does not describe a study testing the acceptance criteria of the device.

However, it does list several performance specifications/parameters for the device, which can be interpreted as acceptance criteria based on its equivalence to the predicate device.

Here's the information derived from the document, with explanations for aspects not explicitly stated:

1. A table of acceptance criteria and the reported device performance (based on the predicate device's performance properties):

Acceptance Criteria (Performance Specifications/Parameters)Reported Device Performance (Subject Device)Comments
Water Content (%)56Same as predicate
Refractive Index1.401Same as predicate
Oxygen Permeability (Dk @ 35°C)$60 x 10^{-11} [(cm^2/sec) x (ml O_2)/(ml x mm Hg)]$Same as predicate
Base Curve (mm)8.6Same as predicate
Diameter (mm)14.0Same as predicate
Light Transmittance (%)>95Same as predicate
Modulus (MPa)≥0.3Same as predicate
Tensile Strength (MPa)≥0.4Same as predicate
Elongation to Break (%)≥100Same as predicate
Surface TreatmentNoSame as predicate
Center Thickness (mm)Varies with powerSame as predicate
Power Range (D)-20.00 to + 20.00Same as predicate
UV Blocker Transmittance (UVB range of 280 to 315nm)< 5%Based on description of UV absorbing monomer
UV Blocker Transmittance (UVA range of 316-380nm)< 50%Based on description of UV absorbing monomer

2. Sample size used for the test set and the data provenance:

  • Not applicable. The document explicitly states: "The modification does not warrant performance testing to support substantial equivalence." This means a new performance study to test acceptance criteria was not conducted for this submission. The equivalence is based on a change in manufacturing release method (from biological to parametric).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. No new performance study was conducted.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable. No new performance study was conducted.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is a contact lens, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Not applicable. This device is a contact lens, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Not applicable. No new performance study was conducted. The "ground truth" for the device's properties is established through the physical/optical properties of the lens material and design, which are stated to be identical to the predicate device.

8. The sample size for the training set:

  • Not applicable. No new performance study was conducted.

9. How the ground truth for the training set was established:

  • Not applicable. No new performance study was conducted.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 20, 2020

CooperVision, Inc. Marie Dutton Senior Regulatory Affairs Specialist 6101 Bollinger Canyon Road, Suite 500 San Ramon, CA 94583

Re: K202756

Trade/Device Name: Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV blocker Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: September 18, 2020 Received: September 21, 2020

Dear Marie Dutton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202756

Device Name

Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker

Indications for Use (Describe)

The CLARITI 1 DAY (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eves that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuty.

The CLARITI 1 DAY TORIC (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia) in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters.

The CLARITI 1 DAY MULTIFOCAL (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical corrective ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with non-diseased eyes that may require a reading addition of +3.00 Diopters or less and may exhibit astigmatism up to 1.50 Diopters or less.

The CLARITI 1 DAY MULTIFOCAL TORIC (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical corrective ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters and require a reading addition of +3.00 Diopters or less.

Type of Use ( Select one or both, as applicable )
----------------------------------------------------------
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

I. SUBMITTER:

CooperVision, Inc. 6101 Bollinger Canyon Road, Suite 500 San Ramon, CA 94583

Contact Person:

Marie Dutton Senior Regulatory Affairs Specialist 6101 Bollinger Canyon Road, Suite 500 San Ramon, CA 94583 Phone: (925) 660-4474 E-mail: MDutton@coopervision.com

Date Prepared:

September 18, 2020

II. DEVICE:

Trade Name: Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker Common Name: Soft (hydrophilic) Contact Lens Classification Name: Lens, Contact, (Disposable) [21 CFR 886.5925 (b) (1)] Regulatory Class: II Product Code: LPL, MVN Classification Panel: Ophthalmic

III. PREDICATE DEVICE:

CooperVision's Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker, K130331

IV. DEVICE DESCRIPTION:

The device description for the subject device and predicate device is identical. The Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker is available as sphere lens, toric lens, multifocal lens, and multifocal toric lens.

In its hydrated state, Clariti 1 day (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV Blocker when placed on the cornea acts as a refracting media to focus light rays on the retina.

Clariti 1 day (somofilcon A) Soft (hydrophilic) Contact Lenses for Daily Wear Single Use are a hydrophilic co-polymer of silicone containing monomers and hydrophilic monomers which is cross-linked with tetraethyleneglycol dimethacrylate and di-functional methacryloxypropyl-terminated poly(dimethylsiloxane).

{4}------------------------------------------------

When hydrated the lens consists of 44.0% somofilcon A and 56.0% water by weight of saline immersed in normal saline. A benzophenone UV absorbing monomer is used in the contact lens to help protect against transmission of harmful UV radiation and Clariti 1 day (somofilcon A) Soft contact lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

The average transmittance characteristics are less than 5% in the UVB range of 280 to 315nm and less than 50% in the UVA range of 316-380nm.

The lens has a hemispherical flexible shell, which covers the cornea and a portion of the adjacent sclera, with the following dimensions:

• Chord Diameter:13.0mm to 15.5mm
• Centre Thickness:0.03mm to 0.50mm
• Base Curve:7.5mm to 9.30mm
• Powers:-20.00 DS to +20.00 DS
• Toric Cylinder options:-0.75, -1.25, -1.75 and -2.25
• Toric Axis options:10° to 180° (10° steps)
• Multifocal Add:

Lens "LOW" = "low" for spectacle near ADD lens (Max +2.25 ADD) Lens "HIGH" = "high" for spectacle near ADD lens (+2.50 ADD or greater)

The physical/optical properties of the lenses are:

• Refractive Index:1.4003
• %Transmittance @ 590nm:98.13
• %Transmittance @ 280-315nm:0.71
• %Transmittance @ 316-380nm:20.62
• Surface Character:Hydrophilic
• Water Content:56%
• Oxygen Permeability (DK):60 x 10-11 (cm²/sec)(ml O2/ml x mmHg) at 35°C (FattMethod for determination of oxygenpermeability)
• Specific Gravity:1.17

{5}------------------------------------------------

V. INDICATIONS FOR USE:

The indications for use statement for the subject device and predicate device is identical.

Lens DesignIndication
SphereThe CLARITI 1 DAY (somofilcon A) Soft (hydrophilic) DailyDisposable Contact Lens with UV blocker is indicated for daily wearsingle use only for the correction of refractive ametropia (myopia andhyperopia) in phakic or aphakic persons with non-diseased eyes thatmay exhibit astigmatism up to 2.00 Diopters that does not interferewith visual acuity.
ToricThe CLARITI 1 DAY TORIC (somofilcon A) Soft (hydrophilic)Daily Disposable Contact Lens with UV blocker is indicated for dailywear single use only for the optical correction of refractive ametropia(myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters.
MultifocalThe CLARITI 1 DAY MULTIFOCAL (somofilcon A) Soft(hydrophilic) Daily Disposable Contact Lens with UV blocker isindicated for daily wear single use only for the optical correction ofrefractive ametropia (myopia and hyperopia) and/or presbyopia inphakic or aphakic persons with non-diseased eyes that may require areading addition of +3.00 Diopters or less and may exhibit astigmatismup to 1.50 Diopters or less.
MultifocalToricThe CLARITI 1 DAY MULTIFOCAL TORIC (somofilcon A) Soft(hydrophilic) Daily Disposable Contact Lens with UV blocker isindicated for daily wear single use only for the optical correction ofrefractive ametropia (myopia and hyperopia) and/or presbyopia inphakic or aphakic persons with non-diseased eyes that may exhibitastigmatism up to 10.00 Diopters and require a reading addition of+3.00 Diopters or less.

{6}------------------------------------------------

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:

The technological principle for both the subject and predicate device is identical.

Technology/Material Comparison
Predicate DeviceSubject Device
Product NameClariti 1 day (somofilconA) Soft (Hydrophilic)Daily Disposable ContactLens with UV BlockerSame
Material USAN Namesomofilcon ASame
510(k) NumberK130331Current Submission
FDA Category (Group)Materials having a Dkgreater than 40 Dk units(using mmHg) and havinga Dk greater than thatexpected based on thematerials' water contentalone (Group V)Same
Manufacturing MethodCast MoldingSame
Wearing andReplacement ScheduleDaily Wear Single UseSame
Sterilization MethodMoist HeatSame
Packaging MaterialsInjection moldedpolypropylene blisterscovered by aluminum foillaminate and blister stripsare packed into printedcartonsSame
Packaging SolutionBorate Buffered SalineSolution containing0.005% w/v Poloxamer407Same
Blue Visibility TintNoSame
TintNoneSame
UV BlockerUV416Same

{7}------------------------------------------------

VII. PERFORMANCE DATA:

The performance specifications/parameters for both the subject and predicate device are identical.

Performance Specifications/Parameters Comparison
Predicate DeviceClariti 1 day(somofilcon A)K130331Subject DeviceClariti 1 day(somofilcon A)Current Submission
Water Content (%)56Same
Refractive Index1.401Same
Oxygen Permeability(Dk @ 35°C)$60 x 10^{-11} [(cm^2/sec) x (ml O_2)/(ml x mm Hg)]$Same
Base Curve (mm)8.6Same
Diameter (mm)14.0Same
Light Transmittance (%)>95Same
Modulus (MPa)≥0.3Same
Tensile Strength (MPa)≥0.4Same
Elongation to Break (%)≥100Same
Surface TreatmentNoSame
Center Thickness (mm)Varies with powerSame
Power Range (D)-20.00 to + 20.00Same

VIII. CONCLUSIONS:

This 510(k) is submitted in accordance with the May 12, 1994 Premarket Notification [510(k)] Guidance Document for Daily Wear Contact Lenses, amended June 28, 1994, for the change from a biological release to parametric release of finished product. The modification does not warrant performance testing to support substantial equivalence. Therefore, the subject device is considered substantially equivalent to the predicate device.

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.