Search Filters

Search Results

Found 62 results

510(k) Data Aggregation

    K Number
    K193380

    Validate with FDA (Live)

    Date Cleared
    2020-07-20

    (228 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CERENOVUS Large Bore Catheter, with the CERENOVUS® Aspiration Tubing Set and NOUVAG Vacuson 60 aspiration pump (or equivalent aspiration pump), is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment.

    The CERENOVUS® Aspiration Tubing Set is intended to connect the CERENOVUS Large Bore Catheter to the canister of the NOUVAG Vacuson 60 Aspiration Pump (or equivalent vacuum pump) and to allow the user to control the fluid flow.

    Device Description

    The CERENOVUS Large Bore Catheter is a variable stiffness, single lumen catheter designed to be introduced over a steerable guide wire or microcatheter into the neuro vasculature. The catheter shaft is composed of a stainless steel variable pitch braid with a PTFE inner liner to facilitate movement of guide wires and other devices. The exterior of the catheter shaft is covered with polymer materials, which encapsulate the stainless steel braid construction. The catheter has a stiff proximal shaft which transitions into the flexible distal shaft to facilitate the advancement of the catheter in the anatomy. The distal end of the catheter has a radiopaque marker band to facilitate fluoroscopic visualization and has a hydrophilic coating to provide lubricity for navigation of vessels. The proximal end of the catheter has a luer fitting located on the end of the catheter hub which can be used to attach accessories for flushing and aspiration. An ID band is placed at the distal end of the hub over a strain relief. The catheter is packaged with a hemostasis valve with a side port and two peel-away introducers as accessories. The hemostasis valve with side port is used for flushing, insertion of catheters, and connection to an external aspiration system. The peel away introducer sheaths are designed to protect the distal tip of the catheter during insertion into the hemostasis valve.

    The CERENOVUS Large Bore Catheter can be connected to the NOUVAG Vacuson 60 aspiration pump (or equivalent aspiration pump) using the CERENOVUS® Aspiration Tubing Set.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device (CERENOVUS Large Bore Catheter and Tubing Set), which focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance, animal, sterilization, shelf-life, and biocompatibility testing. It does not include information about AI/ML device performance, human reader studies, or the establishment of ground truth by expert consensus. Therefore, I cannot generate a response that fulfills the requirements of the prompt regarding acceptance criteria for an AI/ML device or studies proving its performance.

    The document is purely about a physical medical device and its mechanical/physical/biological properties, not an AI/ML diagnostic or assistive technology.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191237

    Validate with FDA (Live)

    Date Cleared
    2019-11-08

    (184 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CERENOVUS Large Bore Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system. The CERENOVUS Large Bore Catheter is also indicated for use as a conduit for retrieval devices.

    Device Description

    The CERENOVUS Large Bore Catheter is a variable stiffness, single lumen catheter designed to be introduced over a steerable guide wire or microcatheter into the neuro vasculature. The catheter shaft is composed of a stainless steel variable pitch braid with a PTFE inner liner to facilitate movement of guide wires and other devices. The exterior of the catheter shaft is covered with polymer materials, which encapsulate the stainless steel braid construction. The catheter has a stiff proximal shaft which transitions into the flexible distal shaft to facilitate the advancement of the catheter in the anatomy. The distal end of the catheter has a radiopague marker band to facilitate fluoroscopic visualization and has a hydrophilic coating to provide lubricity for navigation of vessels. The proximal end of the catheter has a luer fitting located on the end of the catheter hub. An ID band is placed at the distal end of the hub over a strain relief. The catheter is packaged with a hemostasis valve with a side port and two peel-away introducers as accessories. The hemostasis valve with side port is used for flushing and insertion of catheters. The peel away introducer sheaths are designed to protect the distal tip of the catheter during insertion into the hemostasis valve.

    AI/ML Overview

    Here's the information about the acceptance criteria and the study proving the device meets them, based on the provided text:

    Note: The provided document is a 510(k) summary for a medical device (CERENOVUS Large Bore Catheter). This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving efficacy in the same way a new drug or high-risk device might. Therefore, the "study" described is primarily a series of bench tests and biocompatibility assessments, not clinical trials comparing device performance with and without AI, or studies involving human experts for ground truth establishment. Many of the specific points requested (like sample size for test set, number of experts for ground truth, adjudication method, MRMC study effect size, standalone performance, training set details) are generally not applicable or not reported in this type of document for a Class II percutaneous catheter.


    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Visual InspectionCatheter meets visual requirements described in ISO 10555-1 Section 4.4PASS: Samples met the established acceptance criteria
    Catheter IDCatheter internal diameters meet requirementsPASS: Samples met the established acceptance criteria
    Introducer IDIntroducer internal diameters meet requirementsPASS: Samples met the established acceptance criteria
    Catheter ODCatheter outer diameters meet requirementsPASS: Samples met the established acceptance criteria
    Introducer ODIntroducer outer diameters meet requirementsPASS: Samples met the established acceptance criteria
    Catheter Working LengthCatheter working length confirms to ISO10555-1 Section 3.6PASS: Samples met the established acceptance criteria
    Introducer Working LengthIntroducer working length confirms to requirementsPASS: Samples met the established acceptance criteria
    Distal Tip LengthDistal tip length of catheter meets requirementsPASS: Samples met the established acceptance criteria
    Hub Luer TaperCatheter hub luer taper fits standard luer fittings using a taper devicePASS: Samples met the established acceptance criteria
    Air Leak testingNo air leak into the hub subassemblyPASS: Samples met the established acceptance criteria
    System Liquid LeakageCatheter joint strength meets freedom from leakage (liquid during pressurization) requirements of ISO 10555-1:2013, section 4.7PASS: Samples met the established acceptance criteria
    Delamination of PTFE LinerPTFE has appropriately adhered to the inner lumen of the catheter with braid reinforcementPASS: Samples met the established acceptance criteria
    Kink (Distal & Proximal)Catheter remains stable and does not kink during usePASS: Samples met the established acceptance criteria
    Tip MovementCatheter meets tip column stiffness requirementPASS: Samples met the established acceptance criteria
    Tip Linear StiffnessDemonstrates tip flexibility relative to other devices of similar designPASS: Samples met the established acceptance criteria
    Coating Lubricity & DurabilityLubriciousness and durability of the catheter hydrophilic coatingPASS: Samples met the established acceptance criteria
    Coating LengthCatheter hydrophilic coating length meets design requirementsPASS: Samples met the established acceptance criteria
    Catheter Tensile StrengthCatheter joint strength meets requirements of Section 4.5 of ISO 10555-1PASS: Samples met the established acceptance criteria
    Introducer Separation ForceForce required to separate the peel-away introducer accessoryPASS: Samples met the established acceptance criteria
    Particle CountCoating integrity of the catheter's outer surface meets requirements for content of particulatesPASS: Samples met the established acceptance criteria
    Burst Pressure (static)Maximum hydrostatic pressure a catheter can withstandPASS: Samples met the established acceptance criteria
    In-vitro Usability StudiesEvaluate catheter usability parameters such as trackability, tip stability and visibility under fluoroscopy, durability, etc.PASS: Samples met the established acceptance criteria
    Packaging Visual InspectionProposed packaging system has no visible defects that would result in exposure of the primary package or IFUPASS: Samples met the established acceptance criteria
    Packaging Dye LeakSterile package integrity by detecting and locating defects in package seals and pinholes in the packaging materialsPASS: Samples met the established acceptance criteria
    Biocompatibility (Catheter)
    Cytotoxicity (MEM Elution)Test article extract shows no evidence of causing cell lysis or toxicity (grade < 2)PASS: Grade = 0, considered non-cytotoxic
    Sensitization (Maximization Study)Test article extracts show no evidence of causing delayed dermal contact sensitization (all erythema scores =0)PASS: Not considered a sensitizer
    Irritation (Intracutaneous Reactivity)Difference between test and control mean scores is 0.0 for saline and sesame oilPASS: Considered a non-irritant
    Acute Systemic ToxicityNo mortality or evidence of systemic toxicity from extractsPASS: Does not indicate signs of toxicity
    Pyrogenicity (Material Mediated)No single animal shows temperature rise of ≥ 0.5 °C; total rise within USP limitsPASS: Judged as nonpyrogenic
    In Vitro HemolysisHemolytic index for direct contact and extract is 0.0%PASS: Non-hemolytic
    Complement ActivationNot considered a potential activator of the complement systemPASS: Not considered a potential activator of the complement system
    In Vivo Thromboresistance Study - Jugular VeinThrombus formation is slight (Grade ≤ 2) for both test and control articlesPASS: Both test and control articles considered thromboresistant
    ISO 10993-18 Chemical CharacterizationExtractable compounds do not present a toxicological risk to patientsPASS: Extractable compounds do not present a toxicological risk
    Biocompatibility (Introducer)
    Cytotoxicity (MEM Elution)Test article extract shows no evidence of causing cell lysis or toxicity (grade < 2)PASS: Grade = 0, considered non-cytotoxic
    Sensitization (Maximization Study)Test article extracts show no evidence of causing delayed dermal contact sensitization (all erythema scores =0)PASS: Not considered a sensitizer
    Irritation (Intracutaneous Reactivity)Difference between test and control mean scores is 0.0 for saline and sesame oilPASS: Considered a non-irritant
    Acute Systemic ToxicityNo mortality or evidence of systemic toxicity from extractsPASS: Does not indicate signs of toxicity
    Pyrogenicity (Material Mediated)No single animal shows temperature rise of ≥ 0.5°C; total rise within USP limitsPASS: Judged as nonpyrogenic
    In Vitro HemolysisHemolytic index for direct contact and extract is 0.0%PASS: Non-hemolytic
    Complement ActivationNot considered a potential activator of the complement systemPASS: Not considered to be a potential activator of the complement system
    Chemical CharacterizationNo peaks of significant interest, nor any residues in the extract found above quantitation limitsPASS: No presence of extractables

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document repeatedly states "Samples met the established acceptance criteria" for bench tests. While the exact number of samples for each test is not explicitly given, it mentions, "All testing was conducted using sampling methods as required by Codman & Shurtleff, Inc. Design Control procedures." For biocompatibility, some tests specify the number of animals (e.g., mice for systemic toxicity, rabbits for pyrogenicity, 3 for "In Vivo Thromboresistance Study - Jugular Vein").
    • Data Provenance: The studies are bench testing (in-vitro physical and mechanical tests) and biocompatibility testing (in-vitro and some in-vivo animal tests).
      • Country of Origin: Not specified, but likely performed by or for Codman & Shurtleff, Inc. (Raynham, MA, USA).
      • Retrospective or Prospective: These were prospective tests conducted specifically for this 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • Not Applicable. The device is a physical medical instrument (catheter), not an AI or diagnostic device that relies on expert interpretation for ground truth. "Ground truth" in this context is established by objective measurements based on engineering specifications and international standards (e.g., ISO, USP).

    4. Adjudication Method for the Test Set

    • Not Applicable. As above, this is not a study requiring human adjudication for diagnostic ground truth. The acceptance is based on objective measurements and compliance with specified standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No. This is not an AI-assisted diagnostic device, so an MRMC study comparing human readers with and without AI assistance was not performed and is not relevant to this submission.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • No. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    • Objective Measurements and Compliance with International Standards:
      • Bench Testing: Engineering specifications (e.g., dimensions, strength, flow rates), and compliance with international standards such as ISO 10555-1 and related test methods.
      • Sterilization: ISO 11135-1 and requirements for Ethylene Oxide residuals per EN ISO 10993-7 for SAL of 10⁻⁶.
      • Biocompatibility: ISO 10993-1, specific parts of ISO 10993 for various tests (e.g., cytotoxicity, sensitization, irritation), and USP limits for pyrogenicity.

    8. The Sample Size for the Training Set

    • Not Applicable. There is no "training set" as this is not an AI/machine learning device. The "training" in this context refers to the development and manufacturing process based on established engineering principles.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As above, no training set in the AI sense. The design and manufacturing processes are validated against established engineering and quality control standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173192

    Validate with FDA (Live)

    Date Cleared
    2017-11-08

    (37 days)

    Product Code
    Regulation Number
    882.1620
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CereLink ICP Sensor Basic Kit (82-6850); CereLink ICP Sensor Metal Skull (82-6851); CereLink ICP Sensor Plastic Skull (82-6852)

    Indicated when direct ICP monitoring is required. The kit is indicated for use in both subdural and intraparenchymal pressure monitoring applications only.

    CereLink ICP Sensor Ventricular Catheter Kit (82-6854)

    Indicated when direct intraventricular pressure monitoring is required. The kit is indicated for use in ICP monitoring and cerebrospinal fluid (CSF) drainage applications

    Device Description

    The CereLink ICP Sensor Kits are used to monitor intracranial pressure (ICP) through either a stand-alone probe, or a probe coupled with an External Ventricular Drainage (EVD) catheter. The probe, also known as the CereLink ICP Sensor is intended to be used in conjunction with all of Codman's neuromonitoring devices: the Codman ICP Express Monitor (product code 82-6634) and the DirectLink ICP Module (product code 82-6828). The ICP Express and DirectLink are intended for use in ICUs. The CereLink ICP Sensor converts the pressure sensor to a voltage signal. The monitor provides power to the sensor, interprets the voltage signal from the sensor, and displays the corresponding pressure measurements taken by the sensor during a patient's treatment and during patient transport. There is no change to the currently marketed Codman ICP Express or DirectLink as a result of the probe modifications described in this submission.

    The CereLink ICP Sensor contains a small, thin pressure sensor used to measure the intracranial pressure. The sensing element uses a strain gauge located at the tip of the probe. The sensing element is protected by a titanium housing and is exposed to the environment via a silicone membrane. The sensor is connected via wires to a plastic connector housing, and the wires are snaked through a nylon catheter. The connector housing includes a compensation/calibration passive circuit on a Printed Circuit Board (PCB). Additionally, the CereLink ICP Sensor's connector housing includes a new memory PCB board. When the CereLink ICP Sensor is used with either the ICP Express or DirectLink, it functions identically to the cleared predicate Codman Microsensors. Additionally, the connector housing has an electrical connector to attach to any of the monitoring devices.

    The CereLink ICP Sensor Kits include components needed to facilitate the surgical implantation of the Cerelink ICP sensor. The components that will be included with the proposed CereLink ICP Sensor Kits are currently cleared devices, and are identical to the components currently packaged within the predicate Codman Microsensor Kits (i.e. there are no changes being made to the kit components, only the ICP sensor is being modified). Each component and their function are described in the Description section of the Instructions for Use for each kit.

    AI/ML Overview

    The provided text is a 510(k) Summary for the CereLink ICP Sensor Kits, describing the device and its substantial equivalence to a predicate device. It is not an AI/ML device, and therefore does not contain information on acceptance criteria for algorithm performance, sample sizes for test/training sets, expert ground truth establishment, MRMC studies, or standalone algorithm performance.

    The document focuses on demonstrating that the modified device (CereLink ICP Sensor Kits) is substantially equivalent to a previously cleared device (Codman Microsensor Kits) by showing that it has:

    • The same indications for use and intended use.
    • The same fundamental scientific technology and basic design.
    • Incorporates the same materials for the implantable portion.
    • Uses the same packaging and sterilization methods.

    The changes primarily involve minor differences in the plastic connector housing (e.g., shape, addition of PCB and memory PCB, pad printing replacing a paper label) and updated labeling.

    Instead, the document details performance testing for a medical device (intracranial pressure sensor) related to its physical and functional attributes, not AI/ML algorithm performance.

    Here's a breakdown of the information that is present in the document, which primarily focuses on traditional medical device testing and comparisons for regulatory submission:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document includes a "Summary of Testing" table (page 8) that lists various performance tests, relevant standards, and the general "Result" for the subject device. However, it does not provide specific numerical acceptance criteria or reported performance values in a typical table format that would be expected for AI/ML performance metrics (e.g., accuracy, sensitivity, specificity with numerical thresholds).

    Instead, the results are qualitative and confirm that the device "met the established acceptance criteria and is therefore substantially equivalent to the predicate" or "Pass".

    Test CategoryStandards/CriteriaReported Performance/Result
    Bench Testing (including MRI Compatibility)EN 62366:2008 (Usability engineering), ASTM F2052-15 (MRI displacement force), ASTM F2182-11a (MRI RF heating), ASTM F2119-07 (MRI image artifacts), ASTM F2213-06 (MRI torque), ASTM F2503-13 (MRI marking)"Pass - subject device design met the established acceptance criteria and is therefore substantially equivalent to the predicate" (page 8)
    Electrical Safety and Electromagnetic Compatibility TestingIEC 60601-1 (General requirements), IEC 60601-1-2 (Electromagnetic disturbances), IEC 60601-1-6 (Usability), IEC 60601-1-9 (Environmentally conscious design)"Pass - subject device design met the established acceptance criteria and is therefore substantially equivalent to the predicate" (page 8, also detailed on page 9: "The CereLink ICP Sensor design is compliant for Electrical Safety and EMC per IEC 60601-1 2nd and 3th editions.")
    SterilizationEN ISO 11135:2014 (Ethylene Oxide Sterilization), EN ISO 10993-7:2008/AC2009 (Ethylene Oxide Sterilization Residuals), EN ISO 11737-2:2009 (Sterility tests), ANSI ST72:2011 (Bacterial endotoxins)"Pass - subject device design met the established acceptance criteria and is therefore substantially equivalent to the predicate" (page 8, also detailed on page 10: "validated to ensure a sterility assurance level (SAL) of 10^-6", "ethylene oxide residuals can be reduced to an acceptable level", "successfully adopted into Codman's existing sterilization cycle.")
    Shelf Life - Functionality TestingEN ISO 11607-1:2009 (Packaging - Materials), EN ISO 11607-2:2006 (Packaging - Validation)"Pass - subject device design met the established acceptance criteria and is therefore equivalent to the predicate" (page 8, also detailed on page 10: "The shelf life for the proposed CereLink ICP Sensor Kits will be 2 years (same as predicate Codman Microsensor Kits).")
    BiocompatibilityEN ISO 10993-1:2009/AC2010 (Biological Evaluation of Medical Devices), FDA's Guidance document, "Use of International Standard ISO 10993-1. Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process " issued on June 16, 2016."Non-cytotoxic, Non-sensitizing, Non-irritating, Non-toxic, Non-pyrogenic, Non-mutagenic, Non-clastogenic, Non-toxic, and Met USP 37 <661> limits where applicable" (page 9, also detailed on page 10: "test results... demonstrates that the proposed Cerelink ICP Sensor Kits are biocompatible.")
    Functional Pressure RangeNot explicitly stated as a test in the table, but listed as a "Device Characteristic" in comparison table.-50mmHg to 250mmHg (Same as predicate) (page 7)
    Functional Over Pressure Range Without DamageNot explicitly stated as a test in the table, but listed as a "Device Characteristic" in comparison table.-700mmHg to 1250mmHg (Same as predicate) (page 7)
    Zero DriftNot explicitly stated as a test in the table, but listed as a "Device Characteristic" in comparison table.No greater than 5mmHg over 30 days (Same as predicate) (page 7)

    2. Sample Size Used for the Test Set and Data Provenance:

    This information is not applicable to this document as it describes a physical medical device, not an AI/ML algorithm. The "test set" here refers to the actual physical devices subjected to bench testing, sterilization validations, and biocompatibility assessments, rather than a dataset of images or patient records. No information on data provenance (country, retrospective/prospective) is relevant or provided beyond the general understanding that testing was conducted by or for the manufacturer (Codman & Shurtleff, Inc. in Raynham, Massachusetts).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    This is not applicable to a physical device submission like this. Ground truth in this context would be physical measurements, chemical analyses, and adherence to engineering and biological safety standards, not expert consensus on interpretations.

    4. Adjudication Method for the Test Set:

    Not applicable. Adjudication methods (e.g., 2+1, 3+1) are for human interpretation of data, typically in studies involving subjective assessments or labeling of complex medical images, which is not what this document addresses.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC study was done, as this is not an AI/ML device that assists human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Performance:

    Not applicable. This device is an intracranial pressure sensor, whose performance is measured physically (e.g., pressure readings, electrical characteristics, material safety), not as an algorithm.

    7. The Type of Ground Truth Used:

    The "ground truth" for this device's performance is established through:

    • Adherence to recognized international and national standards (e.g., ISO, ASTM, IEC) for medical devices.
    • Bench testing to verify physical and functional characteristics (e.g., MRI compatibility, electrical safety, pressure range).
    • Sterilization validation confirming a specific sterility assurance level.
    • Biocompatibility testing to ensure no adverse biological reactions.
    • Comparison to the predicate device's established performance and characteristics, demonstrating "substantial equivalence."

    8. The Sample Size for the Training Set:

    Not applicable. There is no concept of a "training set" for this type of physical medical device in the context of this submission. The "training" here would be the design, engineering, and manufacturing processes.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. No training set as described for AI/ML algorithms.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172537

    Validate with FDA (Live)

    Date Cleared
    2017-10-16

    (55 days)

    Product Code
    Regulation Number
    882.5550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Use of the CODMAN EDS 3 CSF External Drainage System (EDS 3) is indicated for draining cerebrospinal fluid (CSF) from the cerebral ventricles or the lumbar subarachnoid space as a means of reducing intracranial pressure and CSF volume when the insertion of a permanent, internal shunt is not indicated.

    Use the External Drainage System Collection Bag with the CODMAN EDS 3 External Drainage System to measure and collect cerebrospinal fluid (CSF).

    Device Description

    The Codman EDS 3 CSF External Drainage System (Codman EDS 3 System) is designed to drain cerebral spinal fluid (CSF) at a controlled rate based on differential pressure between the device and the patient. Collecting CSF from the patient is performed in efforts to reduce elevated intracranial pressure (ICP) post trauma. The EDS 3 device is comprised of four (4) main parts: ventricular catheter, patient drain line, base frame drip chamber assembly, and a collection bag. Note: The collection bag is sold as part of the Codman EDS 3 System, as well as sold separately.

    The principle of operation of the proposed Codman EDS 3 system is identical to the currently marketed Codman EDS 3 system. The ventricular catheter is placed into one of the ventricles in the brain or in the subarachnoid space and is then connected to the patient drainage line. CSF flows from the brain or lumbar region through the patient line and enters into the 100 mL graduated drip chamber assembly, where it is collected over a period of time to calculate a flow rate. The drip chamber assembly can then be raised or lowered along the base frame, thereby adjusting the differential pressure to achieve the appropriate flow rate. Once the drip chamber height is set, the collected CSF is then drained into the attached 700 mL collection bag.

    The Codman EDS 3 CSF External Drainage System Collection Bag is a sterile, vented 700 mL capacity bag that is graduated in 50 mL increments for accurate measurement. A microbial-retentive atmospheric vent facilitates CSF flow into the bag. One bag is provided with the system and replacement bags are sold separately as an accessory to the drain.

    The EDS 3 system is a complete, disposable unit that is provided sterile and is available with or without a ventricular catheter.

    AI/ML Overview

    Here is a summary of the acceptance criteria and study information for the CODMAN EDS 3 CSF External Drainage System and Collection Bag, based on the provided document:

    1. Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    IV Pole Clamp Design VerificationThe new IV Pole Clamp must hold at least the same amount of weight as the predicate device.Pass - proposed device design met the acceptance criteria and is therefore substantially equivalent to the predicate device.
    Collection Bag Design VerificationThe new Collection Bag must connect to the EDS 3 system, be leak-free, and exhibit at least the same tensile strength as the predicate device.Pass - proposed device design met the acceptance criteria and is therefore substantially equivalent to the predicate device.
    Drainage Time Design VerificationThe drainage time for the new system must meet design input specifications and be at least as fast as the predicate device.Pass - proposed device design met the acceptance criteria and is therefore substantially equivalent to the predicate device.
    Design Validation / Simulated UseThe finished product must meet all user inputs and needs and inputs of the predicate device.Pass - proposed device design met the acceptance criteria and is therefore substantially equivalent to the predicate device.
    Sterilization Assurance Level (SAL)Achieve a SAL of 10-6 in accordance with ISO 11135-1:2014.The sterilization cycle has been validated to ensure a sterility assurance level (SAL) of 10-6.
    Shelf LifeDevice and packaging remain functional and maintain sterility for a specified period (3 years for EDS3 System and Collection Bags).Aging studies established that the device and packaging remain functional and maintain sterility for up to 3 years.
    BiocompatibilityMaterials for non-patient contacting components are acceptable for intended use and meet ISO 10993-1:2009.The proposed materials were determined to be acceptable for their intended use and the final device continues to meet the requirements of ISO 10993-1:2009.

    2. Sample Size for Test Set and Data Provenance

    The document describes only bench testing for design verification and validation. It does not explicitly state a 'test set' in the context of patient data or clinical trials. The evaluation was primarily based on comparing the performance of the modified device components against the predicate device in a laboratory setting. No data provenance in terms of country of origin or retrospective/prospective study type is applicable here, as no human data was used.

    3. Number of Experts and Qualifications for Ground Truth

    Not applicable. The study involved bench testing and engineering verification, not a clinical study requiring expert consensus for ground truth establishment.

    4. Adjudication Method for Test Set

    Not applicable. There was no clinical test set requiring adjudication. The verification and validation activities were based on standardized engineering and performance testing methods.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. The submission relies on bench testing demonstrating substantial equivalence to a predicate device, not on comparative clinical performance with human readers.

    6. Standalone Algorithm Performance

    Not applicable. This device is a medical drainage system, not an AI or algorithm-based device.

    7. Type of Ground Truth Used

    The "ground truth" for this submission was established through engineering design specifications, industry standards (e.g., ISO 11135-1:2014, ISO 10993-1:2009), and performance of the legally marketed predicate device. The modified components' performance was compared against these established benchmarks.

    8. Sample Size for the Training Set

    Not applicable. There was no machine learning or AI component, thus no training set was used.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there was no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172022

    Validate with FDA (Live)

    Date Cleared
    2017-10-03

    (90 days)

    Product Code
    Regulation Number
    882.5550
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Codman HAKIM Precision Valve System is an implantable device that provides constant intraventricular pressure and drainage of cerebral spinal fluid (CSF) for the management of hydrocephalus.

    The Codman HAKIM Programmable Valve System is an implantable device that provides constant intraventricular pressure and drainages of cerebral spinal fluid (CSF) for the management of hydrocephalus.

    The Codman HOLTER Lumboperitoneal (LP) Shunt is indicated for shunting cerebrospinal fluid when the lumboperitoneal route is the procedure of choice in the treatment of communicating hydrocephalus.

    The Codman HOLTER Atrial Catheters are indicated for use to shunt cerebrospinal fluid, when shunting of cerebrospinal fluid to the atrium is the procedure of choice in the treatment of hydrocephalus.

    The Codman HOLTER Ventricular Catheters are indicated for use to gain access to the ventricles for diagnostic purposes and in the treatment of hydrocephalus.

    The Codman Medos Ventricular Catheter is indicated for use in the treatment of hydrocephalus as a component of a shunt system when draining or shunting of cerebrospinal fluid (CSF) is indicated.

    The UNI-SHUNT System is indicated for use as a one-piece ventriculo-peritoneal shunt system for the palliative treatment of hydrocephalus. No other use is recommended.

    The Codman BACTISEAL Catheters are indicated for use in the treatment of hydrocephalus as a component of a shunt system when draining or shunting of cerebrospinal fluid (CSF) is indicated.

    Device Description

    Codman HAKIM Precision and Programmable Valves:
    The Codman HAKIM Precision and Programmable Valves are implantable devices that provide constant intraventricular pressure and drainage of cerebral spinal fluid (CSF) for the management of hydrocephalus. Both the Codman HAKIM Precision and Programmable Valves are pressure regulating valves which maintain intraventricular pressure at a constant level. The Codman HAKIM Precision valves are fixed pressure valves and are available in 5 different opening pressure ranges. The Codman HAKIM Programmable Valves, not having fixed pressures, permit non-invasive adjustment of the valve opening pressure. The Codman HAKIM Programmable Valves can be adjusted to 18 different opening pressure settings.

    Codman HOLTER Catheters:
    The HOLTER Catheter, is a barium-impregnated silicone rubber open-ended catheter. Two stainless steel Type "A" Fixation and Joining Connectors are included with each catheter to use in rejoining the catheter if it has been cut for lengthening or revision. The HOLTER Catheter, Salmon Design, is a barium-impregnated silicone rubber catheter. Four longitudinal slits (90° apart) near the closed distal tip of the catheter are for drainage of cerebrospinal fluid. Two stainless steel Type "A" Connectors are included with each catheter to use in rejoining the catheter if it has been cut for lengthening or revision.

    Codman Medos Ventricular Catheter:
    The Medos Ventricular Catheter is made from barium-impregnated silicone tubing. The catheter is 140 mm in length and is supplied with 24 inlet holes, 3 rows of 8 holes, at the proximal end. The catheter, with stainless steel stylet and right angle adapter, is supplied sterile.

    UNI-SHUNT® System:
    The UNI-SHUNT® system with Reservoir incorporates a double dome access port to facilitate injections and aspirations of CSF samples. It is a continuous length of barium-impregnated silicone tubing with an access reservoir made of self-sealing silicone which can be punctured with a 25 gauge or smaller Huber type needle.

    Codman BACTISEAL Catheters:
    The BACTISEAL Catheters are made of radiopaque (barium-impregnated) silicone tubing and are supplied sterile. BACTISEAL Catheters are subjected to a treatment process by which the silicone tubing is impregnated with rifampin and clindamycin hydrochloride. The catheter is 14 cm in length and is supplied with 24 inlet holes (3 rows of 8 holes) at the proximal end. Depth marks have been added to the catheter (one dot at 5 cm and two dots at 10 cm). A stainless-steel stylet and right angle adapter are packaged with the ventricular catheter. The peritoneal catheter has a beveled tip at one end and the other end of the catheter has a flat tip. The catheter is 120 cm long and may be trimmed to the proper length.

    AI/ML Overview

    This document, K172022, primarily discusses the substantial equivalence of updated Codman Hydrocephalus Shunt Systems to previously cleared predicate devices. The changes noted are related to labeling updates for MRI compatibility and compliance with ISO 7197 standards for pressure flow characteristics. The performance data provided is entirely from bench testing; there are no human studies (clinical or multi-reader multi-case) described for proving device performance.

    Therefore, many of the requested elements regarding human studies (like MRMC studies, expert ground truth adjudication, and clinical study sample sizes) are not applicable to this specific submission as presented in the document.

    Here's the breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria and reported device performance are based on bench testing for various aspects of the device, primarily related to MRI safety and conformity to specific ISO and ASTM standards.

    StandardAcceptance CriteriaReported Device Performance
    ISO 7197Characterize the pressure flow characteristics of the device per ISO 7197.Pass – Pressure and Flow characteristic results added to the product IFUs.
    ASTM F647Characterize the long-term stability of the implantable shunt assemblies per ASTM F647. Mean difference of Post stability – pre-stability testing (P-Q post bursting pressure) must not be greater than 10 mmH2O at each pressure setting tested and must comply with PQ characteristic specifications in the IFU.Pass – average difference less than 10mm H2O and results comply with characteristics provided in the IFU.
    ASTM F2119Image Artifact information was collected for the devices in both 1.5T and 3.0T MR environments. For each device, scans were made in three planes (sagittal, coronal, and axial) for using both gradient and spin echo sequences. Our labeling lists the worst-case image artifact for gradient echo sequencing. (Implied acceptance: artifacts are within acceptable limits for safe imaging and adequate information for labeling).Pass – results added to MR Information in the product IFUs.
    ASTM F2182The acceptance criterion for this test was to characterize the implanted device increase in temperature after 15 minutes of continuous scanning (in both 1.5 T and 3.0 T MR environments). (Implied acceptance: temperature increase is within safe limits for human use).Pass - results added to MR Information in the product IFUs.
    ASTM F2052This test assessed if the amount of magnetically induced force on the device is less than or equal to the force on the device due to gravity. The magnetically induced force for the devices was considered to meet the acceptance criteria in both 1.5T and 3.0T MR environment, thus supporting the MR Conditional claim. The maximum acceptable spatial gradient was determined on the basis of the component with the largest deflection, and is listed in our labeling. (Implied acceptance: magnetically induced force is not clinically significant such that it dislodges the device or causes harm).Pass - results added to MR Information in the product IFUs.
    ASTM F2213This test assessed if the amount of magnetically induced torque on the device is less than or equal to the gravitational torque. The magnetically induced torque for the devices was considered to meet the acceptance criteria in both 1.5T and 3.0T MR environments, thus supporting the MR Conditional claim. (Implied acceptance: magnetically induced torque is not clinically significant such that it dislodges the device or causes harm, particularly with rotation).Pass - results added to MR Information in the product IFUs.

    Study Details for Device Performance

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not specified in terms of number of devices tested. The document refers to "the devices" being tested.
      • Data Provenance: The studies are bench tests, meaning they were conducted in a laboratory setting. There is no information about country of origin for the data or whether it was retrospective or prospective.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable. Ground truth, in this context, would typically refer to clinical diagnosis or expert annotations for an AI/diagnostic device. This submission describes physical device testing against engineering standards. The "ground truth" is defined by the technical specifications outlined in the ASTM and ISO standards and the performance of the device against those specifications. No human experts were involved in establishing the "ground truth" for the bench tests.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not Applicable. As this is bench testing against specified engineering criteria, there is no need for human adjudication of test results.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. The document explicitly states "No clinical studies were required". This device is a hydrocephalus shunt system, not an AI, diagnostic imaging, or reader-assisted device.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Not Applicable. This is a physical medical device (shunt system), not a software algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for these bench tests are the established engineering and safety specifications defined by the relevant ASTM and ISO standards (e.g., maximum allowable temperature increase, force/torque limits, pressure flow characteristics).
    7. The sample size for the training set:

      • Not Applicable. This is a physical medical device, not a machine learning model, so there is no concept of a "training set."
    8. How the ground truth for the training set was established:

      • Not Applicable. See point 7.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171862

    Validate with FDA (Live)

    Date Cleared
    2017-09-25

    (95 days)

    Product Code
    Regulation Number
    882.5950
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GALAXY G3 Mini Microcoil Delivery System is intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations fistulae, and is also intended for arterial and venous embolizations in the peripheral vasculature.

    Device Description

    The GALAXY G3 Mini Microcoil Delivery Systems consist of three components, a Microcoil System, a connecting cable, and a Detachment Control Box (DCB). Each component is sold separately. As shown in Figure 1, the Microcoil System consists of a microcoil attached to a Device Positioning Unit (DPU). The Microcoil System is packaged in an introducer sheath designed to protect the coil in the packaging dispenser and to provide support for introducing the coil into the microcatheter catheter. The microcoil is the implantable segment of the device, and is detached from the Device Positioning Unit (DPU) using the Detachment Control System (Detachment Control Box and connecting cable). The microcoil is fabricated from a platinum alloy wire. The wire is wound into a primary coil which contains a polypropylene suture (SR) and then formed into a secondary shape. The secondary shape is complex. The DPU is a variable stiffness wire and has a radiopaque marker band located three (3) cm from its distal end. The Device Positioning Unit includes five (5) fluoro saver markers on the proximal section of the shaft. The markers are intended to indicate when the tip of the microcoil is approaching the tip of the microcatheter. When the distal-most marker reaches the proximal end of the Rotating Hemostatic Valve (RHV) on the microcatheter, the tip of the coil is approaching the tip of the microcatheter and fluoroscopy should be used to guide further coil insertion. The introducer sheath has three main components: an introducer tip, a translucent introducer body, and a re-sheathing tool. The EnPOWER Detachment Control Box (DCB) provides the energy necessary to allow for a thermo-mechanical detachment of the microcoil from the DPU. The connecting cable delivers the energy necessary to detach the embolic coil from the Microcoil System's detachment zone. The connecting cable is connected between the Microcoil System's hub connector on the DPU and the output connector on the DCB. The connecting cables may be one of two types: one with a remote detach button (the EnPower Control Cable) catalog no. ECB000182-00, or one without a detach button (standard connecting cable) catalog no. CCB00157-00. The EnPower Detachment Control Box works with the EnPower Control Cable and with the standard connecting cable.

    AI/ML Overview

    The provided document describes the development and testing of the GALAXY G3 Mini Microcoil Delivery System. Here's a breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document reports several performance tests, and for each, the result is "PASS: Samples passed the established acceptance criterion." The specific numerical acceptance criteria are generally not explicitly stated, but the passing result indicates they were met.

    TestTest Method SummaryReported Device Performance
    Spring ConstantMeasures the softness of the coil by recording the spring constant of the primary wind.PASS: Samples passed the established acceptance criterion
    Complex ShapeVisually inspects the complex shape of the G3 Mini Microcoils.PASS: Samples passed the established acceptance criterion
    ParticulateMeasures particulate count during simulated use per USP788.PASS: Samples passed the established acceptance criterion
    Atraumatic BeadVisually verifies that the bead end of the coil meets the final assembly specification.PASS: Samples passed the established acceptance criterion
    DPU 3 System Outer DiameterVerifies the OD is within specification to ensure microcatheter compatibility.PASS: Samples passed the established acceptance criterion
    Microcatheter Tip Deflection ForceMeasures the deflection and/or stability of the microcatheter by recording the force generated at the distal tip as the DPU device is advanced to the tip.PASS: Samples passed the established acceptance criterion
    Detachment Zone Tensile StrengthEvaluates the attachment strength of the detachment fiber to prevent unintentional coil detachments.PASS: Samples passed the established acceptance criterion
    Stretch Resistance Fiber Tensile StrengthVerifies that the coil provides sufficient stretch resistance under tensile loading to ensure the coil can be retracted and repositioned without stretching.PASS: Samples passed the established acceptance criterion
    Track Force (Delivery)Evaluates the force it takes to deliver the device through a microcatheter and into a clinically relevant model; utilizing the system Catheter Performance Simulation System (CPSS).PASS: Samples passed the established acceptance criterion
    Dimensional Inspection of FDL Diameter and Coil LengthVerifies the FDL diameter and the coil length meets the specification.PASS: Samples passed the established acceptance criterion
    Coil OD Verification on Final AssemblyVerifies the OD is within specification.PASS: Samples passed the established acceptance criterion
    Dimensional Inspection of the Distal Fluro-saver markersVerifies that the Fluoro Saver Markers are in the correct proximal position to give a visual indication that the microcoil is approaching the distal tip of the microcatheter.PASS: Samples passed the established acceptance criterion
    Coil DurabilityEvaluates the coil's ability to stay attached to the device during simulated use of six insertions and withdrawals cycled into and out of a clinically relevant aneurysm model.PASS: Samples passed the established acceptance criterion
    Detachment, Coil Durability & ReliabilityEvaluates the reliability of the detachment mechanism after being cycled into and then out of a clinically relevant anatomical model six times.PASS: Samples passed the established acceptance criterion
    Resheathing ReliabilityEvaluates the ability to re-insert the device into the split sheath introducer after it has been unzipped after the device has been inserted and withdrawn from a clinically relevant model.PASS: Samples passed the established acceptance criterion
    Fluro saver Marker DurabilityEvaluates the ability of the Fluoro Saver Markers to stay affixed and in the correct position on the shaft after being cycled into and then out of a clinically relevant anatomical model six times.PASS: Samples passed the established acceptance criterion
    Distal Outer Sheath DurabilityEvaluates the durability of the distal outer sheath during the simulated use of six insertions and withdrawals of the device into and out of a clinically relevant aneurysm model.PASS: Samples passed the established acceptance criterion
    Dimensional Inspection of the IntroducerIntroducer underwent dimensional inspection per approved test method.PASS: Samples passed the established acceptance criterion
    Introducer Bond StrengthVerifies that the bond strength of the introducer fuse joint.PASS: Samples passed the established acceptance criterion
    Coil Transfer to MicrocatheterEvaluates the introducer sheath for allowing for insertion of the embolic coil into the microcatheter through the RHV.PASS: Samples passed the established acceptance criterion
    Introducer FlushingVisually inspects to confirm flushing.PASS: Samples passed the established acceptance criterion
    RadiopacityVerifies the ability to visualize the coil under fluoroscopy for physician to determine the location of the coil during use.PASS: Samples passed the established acceptance criterion
    Overall Coil PerformanceEvaluates physician's satisfaction rating on performance compared to a predicate device.PASS: Samples passed the established acceptance criterion
    MRI TestingDemonstrates that GALAXY G3 MINI Microcoil to be "MR-conditional" according to the specific conditions used for the assessment.PASS: Samples passed the established acceptance criterion

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the exact sample sizes used for each individual performance test (test sets). It generally refers to "samples" being tested.

    The data provenance is bench testing, which implies the data was collected within a laboratory setting, likely in the US, given the submission to the FDA. It is retrospective in the sense that it's testing a finished device against predetermined criteria.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    No individual experts or their qualifications are mentioned for establishing ground truth for the bench tests. The "ground truth" for these tests is based on established engineering specifications, industry standards (e.g., USP788, ISO standards), and clinical relevance/simulated use scenarios. For the "Overall Coil Performance" test, "physician's satisfaction rating" was evaluated, implying input from medical professionals, but the number and qualifications are not specified.

    4. Adjudication Method for the Test Set

    Not applicable for the reported tests. The tests are primarily objective measurements against established criteria, or visual inspections by qualified personnel (not specified as "experts" in the context of adjudication).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "A clinical study was not required as appropriate verification and validation of the GALAXY G3 Mini Microcoil Delivery System was achieved based on the similarities of the proposed device to the predicate device, and from results of bench testing."

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical device (microcoil delivery system), not an algorithm or AI software. Therefore, the concept of "standalone algorithm performance" does not apply. The performance tests are for the physical device itself.

    7. The Type of Ground Truth Used

    The ground truth for the bench tests is based on:

    • Engineering specifications and design requirements: For dimensional checks, material properties, and functional performance (e.g., spring constant, detachment force, tensile strength).
    • Industry standards: Such as USP788 for particulate matter, ISO 11607 for packaging, ISO 10993-1 for biocompatibility, and ISO 11135-1 for sterilization.
    • Simulated use in clinically relevant models: For tests like track force, coil durability, detachment durability, resheathing reliability, fluoro saver marker durability, and distal outer sheath durability.
    • Physician satisfaction rating: For "Overall Coil Performance" compared to a predicate device, which implies a subjective expert assessment.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device, not an AI or machine learning model that requires a "training set." The listed studies are verification and validation tests for the device itself.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MICRUSFRAME, DELTAFILL, and DELTAXSFT Microcoil Delivery Systems are intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae, and are also intended for arterial and venous embolizations in the peripheral vasculature.

    The GALAXY G3 FILL Microcoil Delivery System is intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations and is also intended for arterial and venous embolizations in the peripheral vasculature.

    The GALAXY G3 XSFT Microcoil Delivery System is intended for endovascular embolization of intracranial aneurysms.

    Device Description

    The MICRUSFRAME, DELTAFILL, DELTAXSFT, GALAXY G3 FILL, GALAXY G3 XSFT Microcoil Delivery Systems consist of three components, a Microcoil System, a connecting cable, and a Detachment Control Box (DCB). Each component is sold separately. As shown in Figure 1, the Microcoil System consists of a microcoil attached to a Device Positioning Unit (DPU). The Microcoil System is packaged in an introducer sheath designed to protect the coil in the packaging dispenser and to provide support for introducing the coil into the microcatheter catheter. The microcoil is the implantable segment of the device, and is detached from the Device Positioning Unit (DPU) using the Detachment Control System (Detachment Control Box and connecting cable). The devices in this submission include minor design changes only to the Device Positioning Unit's introducer sheath (introducer). There are no modifications to components or materials of the micro-coil or the ENPOWER Detachment Control System. Minor dimensional and design modifications to the introducer will help improve deliverability of the micro-coils.

    AI/ML Overview

    The document describes the MICRUSFRAME, DELTAFILL, DELTAXSFT, GALAXY G3 FILL, and GALAXY G3 XSFT Microcoil Delivery Systems. These devices are intended for endovascular embolization of intracranial aneurysms, various neurovascular abnormalities, and arterial and venous embolizations in the peripheral vasculature. The submission is for minor design changes to the introducer sheath component of the delivery system's device positioning unit.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document provides a table of verification and validation testing, which includes the tests performed, a summary of the test methods, and the results. The acceptance criterion for all tests was "Pass," meaning the samples met the established criteria.

    TestTest Method SummaryReported Device Performance (Results)Acceptance Criteria (Implied)
    Visual InspectionVisual inspection of the test units to check for cosmetic defects to ensure the units are prepared for verification testing as per established test method.Pass (All units passed visual inspection)All units pass visual inspection
    Tracking Force (delivery)The purpose of the Track Force test was to evaluate the force it takes to deliver the proposed device through a microcatheter and into a clinically relevant model; utilizing the system Catheter Performance Simulation System (CPSS). Test samples were delivered through a compatible microcatheter to verify track forces per approved test method.Pass (Samples passed the established acceptance criterion)Samples pass the established acceptance criterion
    Re-sheathing ReliabilityThe purpose of the Re-Sheathing Reliability test was to evaluate the ability to re-insert the proposed device into the split sheath introducer after it has been unzipped after the proposed device has been inserted and withdrawal from a clinically relevant model. The introducer sheath underwent 1 re-sheathing cycle to verify reliability per approved test method.Pass (Samples passed the established acceptance criterion)Samples pass the established acceptance criterion
    Dimensional InspectionThe Introducer underwent dimensional inspection per approved test method.Pass (Samples passed the established acceptance criterion)Samples pass the established acceptance criterion
    Particulate TestingThe full assembly underwent particulate testing per approved test method. Simulated use consisted of pushing forward to the tip of the microcatheter and then pulling back 8" and repeating five times.Pass (Samples passed the established acceptance criterion)Samples pass the established acceptance criterion
    Introducer Fuse Joint TestingThe Introducer underwent tensile strength testing per approved test method.Pass (Samples passed the established acceptance criterion)Samples pass the established acceptance criterion
    Acute In-Vivo Animal StudyAn in-vivo model allowed the assessment of the acute performance of the test article to deliver an embolic coil to the target parent vessel in swine.Demonstrated acceptable overall performance in all attributes evaluated.Acceptable overall performance in all attributes.
    Biocompatibility TestingA limited subset of the recommended biocompatibility tests, including in vitro cytotoxicity and in vitro hemolysis were successfully conducted on the modified introducer. In addition, chemical characterization of extractables of the Introducers manufactured with the current heat shrink polymer and the Introducers manufactured with a new heat shrink polymer were successfully conducted per ISO 10993-18.Successfully conducted (in vitro cytotoxicity, in vitro hemolysis, chemical characterization of extractables).Successful completion of required biocompatibility tests.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Test Set Sample Size: The document repeatedly states "Samples passédes the established acceptance criterion" but does not specify the numerical sample sizes used for each individual bench test (Visual Inspection, Tracking Force, Re-sheathing Reliability, Dimensional Inspection, Particulate Testing, Introducer Fuse Joint Testing). It mentions "statistical sampling methods" were used for all testing as required by Codman & Shurtleff, Inc. Design Control procedures. For the animal study, it states "an acute in-vivo animal study" was performed in "swine," but the number of animals or tests performed is not specified.
    • Data Provenance: The document does not specify the country of origin for the data generated from the verification and validation testing. The studies are described as prospective testing conducted in a laboratory setting for the bench tests, and an acute in-vivo animal study for the performance assessment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • This information is not applicable to the type of testing described. The tests are engineering and biological performance evaluations of a medical device, not diagnostic evaluations requiring expert interpretation of images or patient data to establish ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This information is not applicable. The tests performed are objective measurements and evaluations of device performance characteristics, not subjective assessments requiring adjudication by multiple experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This device is a microcoil delivery system, not an AI or imaging-related diagnostic tool that would typically involve human readers or AI assistance in interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This device is a physical medical instrument, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the bench tests was established through objective engineering specifications and validated test methods. For example, for "Tracking Force," the ground truth is a specific force measurement threshold. For "Biocompatibility," the ground truth is adherence to internationally recognized standards (ISO 10993-1, FDA guidance). For the animal study, the ground truth was "acceptable overall performance" in delivering the embolic coil, implying pre-defined success criteria for deployment and functional integrity within the in-vivo model.

    8. The sample size for the training set:

    • Not applicable. This device is a physical medical instrument, not a machine learning algorithm that requires a training set. The device itself is "trained" through prior design iterations and predicate device experience, and then validated through non-clinical testing.

    9. How the ground truth for the training set was established:

    • Not applicable. As stated above, this is not a machine learning algorithm. The "ground truth" for the design and manufacturing of the device stems from established engineering principles, material science, regulatory standards (e.g., ISO, FDA guidance), and the performance history of predicate devices.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171653

    Validate with FDA (Live)

    Date Cleared
    2017-07-03

    (28 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The YOGA Microcatheter is intended for use in peripheral, coronary, and neuro vasculature for the intravascular introduction of interventional/diagnostic devices.

    Device Description

    The YOGA Microcatheter is a variable stiffness, single lumen catheter designed to access small, tortuous vasculature. The catheter shaft is composed of a variable pitch stainless steel braid with a PTFE inner liner to facilitate movement of guide wires and other devices. The exterior of the catheter shaft is covered with a polymer material, which encapsulates the stainless steel braid construction. The distal end of the catheter has a radiopaque marker band and has a hydrophilic coating to provide lubricity for navigation of vessels. The proximal end of the catheter has a hub and an ID band is placed at the distal end of the hub over a strain relief. A steam shaping mandrel is provided in the package. The two new YOGA Microcatheters are provided in back up (XB) and extra backup (XXB) configurations, i.e. a stiffer distal end to provide additional support.

    AI/ML Overview

    The provided text is for a 510(k) premarket notification for a medical device called the "YOGA Microcatheter". This document focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than proving the device meets specific performance acceptance criteria through the kind of study described in the prompt.

    Therefore, the document does not contain the detailed information necessary to answer the prompt's questions. The prompt is designed to elicit information about a study that assesses AI/algorithm performance (e.g., accuracy, sensitivity, specificity, or improvement in human reader performance), which typically involves a test set, ground truth experts, and statistical analysis against predefined acceptance criteria. This 510(k) submission, however, details the physical and functional characteristics of a medical catheter and demonstrates its similarity to a previously approved device through bench testing, not a diagnostic study involving AI or image analysis.

    Here's why the information isn't present:

    • Device Type: The YOGA Microcatheter is a physical medical device (a catheter), not a software algorithm, diagnostic imaging tool, or AI system. Its performance is evaluated through physical and functional characteristics (e.g., stiffness, trackability, biocompatibility, sterilization) rather than diagnostic accuracy metrics.
    • Regulatory Pathway: A 510(k) submission seeks to establish "substantial equivalence" to a predicate device. This often involves demonstrating similar technological characteristics and performance through bench testing, not necessarily conducting large-scale clinical trials or diagnostic accuracy studies like those performed for AI/CAD devices.
    • Nature of Testing: The "Performance Data" section describes "Bench Test Summary" including Linear Stiffness, Lateral Stiffness, and Track Testing. These are engineering tests, not studies involving human readers, ground truth consensus for imaging data, or AI performance metrics.

    Therefore, I cannot provide the requested table or answer the specific questions about acceptance criteria for an AI/algorithm, sample sizes for test/training sets, expert qualifications, or MRMC studies, as this information is not present in the provided 510(k) document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162563

    Validate with FDA (Live)

    Date Cleared
    2017-01-05

    (113 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The YOGA Microcatheter is intended for use in the peripheral, coronary, and neuro vasculature for the intravascular introduction of interventional/diagnostic devices.

    Device Description

    The YOGA Microcatheter is a variable stiffness, end to end braided single lumen catheter designed to access small, tortuous vasculature. The microcatheter has an outer hydrophilic coating that provides lubricity during navigation of vessels. The lubricious PTFE lined inner lumen is designed to facilitate movement of guide wires and other devices. A radiopaque marker band is provided at the catheter tip to aid fluoroscopic visualization. A luer fitting located on the proximal end of the catheter hub is used to attach accessories. A steam shaping mandrel is provided in the package.

    AI/ML Overview

    The provided document describes the substantial equivalence determination for the YOGA Microcatheter (K162563) based on its comparison to a predicate device, the ENVOY DA Guiding Catheter (K140080), and other reference devices. The focus of the provided text is on demonstrating the device's functional integrity and biological compatibility through various performance tests, rather than clinical efficacy as would be seen with an AI/ML device.

    Here's an analysis of the "acceptance criteria and the study that proves the device meets the acceptance criteria" based on the provided text, structured to address your specific points:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the YOGA Microcatheter are implicitly defined by the "PASS" results for each of the performance tests listed. The device is deemed to meet these criteria if its performance matches or is equivalent to established standards or those of the predicate device.

    Test TitleAcceptance Criteria (Implicit)Reported Device Performance
    Visual InspectionSamples meet established acceptance criteria (e.g., no defects)PASS
    Catheter IDCatheter ID verified to pass requirementsPASS
    Catheter ODCatheter OD measured within established specificationsPASS
    Catheter Working LengthCatheter working length measured within established specificationsPASS
    Distal Tip LengthDistal Tip Length measured within established specificationsPASS
    Catheter Tensile StrengthWithstands specified tensile forcesPASS
    Hub Luer TaperComplies with ISO 594-1 and ISO 594-2PASS
    Air Leak testingNo air leakage observedPASS
    Flow Rate (static)Complies with ISO 10555-1 specified flow ratesPASS
    System Liquid LeakageNo liquid leakage under specified pressurePASS
    Flow Rate (dynamic)Meets dynamic flow rate specifications at 100/300 psiPASS
    Burst Pressure (static)Withstands specified burst pressurePASS
    Lumen flushMeets particulate count requirements per USP<788>PASS
    Coating IntegrityMeets particulate count requirements per USP<788> after simulated usePASS
    Delamination of PTFE LinerNo delamination of PTFE liner observedPASS
    Aseptic RemovalFacilitates aseptic removal per procedurePASS
    Steam ShapingSuccessful shaping per Instructions for UsePASS
    Linear Stiffness TestMeets linear stiffness specificationsPASS
    Lateral Stiffness TestMeets lateral stiffness specificationsPASS
    Track TestingForce required to track through anatomical model within limitsPASS
    Radiopacity (Animal Model)Distal end of catheter visible under fluoroscopyPASS
    Biocompatibility (various tests)Complies with ISO 10993 series and FDA Bluebook Memorandum G95-1PASS
    Sterilization Assurance Level (SAL)Achieves SAL of 10^-6PASS
    Shelf-Life (3 years)Demonstrated shelf-life equivalent to predicate devicesPASS

    Study Details

    The provided document describes physical and biological performance testing for a medical device (microcatheter), not an AI/ML device. Therefore, many of your specific questions regarding AI/ML studies (like ground truth, experts, MRMC, standalone performance) are not applicable to this submission. However, I will answer the relevant points based on the provided text.

    1. Sample size used for the test set and the data provenance:

      • Sample Size: The document states that "All testing was conducted using sampling methods as required by Codman & Shurtleff, Inc. Design Control procedures." However, specific sample sizes for each test are not provided in this document.
      • Data Provenance: The tests are "bench" (in vitro) and "animal" (in vivo) tests conducted by the manufacturer, Codman & Shurtleff, Inc. The document does not specify the country of origin for the animal studies, but the manufacturer is based in Raynham, Massachusetts, USA. The studies are prospective as they were conducted to support the 510(k) submission.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable as the document describes performance testing for a physical medical device, not an AI/ML algorithm requiring expert ground truth for classification or diagnosis. The "ground truth" for these tests is based on established engineering and biological standards (e.g., ISO, ASTM, USP).
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This is not applicable. Adjudication methods are typically relevant for human review of AI/ML outputs or clinical endpoints with subjective interpretation. For the physical and biological tests described, the determination of "PASS" or "FAIL" would be based on objective measurements against pre-defined acceptance criteria, not an adjudication process involving multiple human reviewers.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This is not applicable. This document describes the clearance of a physical medical device and does not involve AI or human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This is not applicable. The device is a microcatheter, a physical instrument, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The "ground truth" for the device's performance is based on established engineering standards, international standards (ISO, ASTM, USP), and validated internal design control procedures. For example, tensile strength is measured against engineering specifications, biocompatibility against ISO 10993 standards, and flow rates against ISO 10555-1. For radiopacity, the ground truth is the visual confirmation by a qualified individual (presumably a radiologist or veterinarian during the animal study) that the catheter is visible under fluoroscopy.
    7. The sample size for the training set:

      • This is not applicable. There is no "training set" as this is not an AI/ML device.
    8. How the ground truth for the training set was established:

      • This is not applicable. There is no "training set" for this physical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K162437

    Validate with FDA (Live)

    Date Cleared
    2016-10-28

    (58 days)

    Product Code
    Regulation Number
    882.5550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Use of the Codman EDS3 CSF External Drainage System is indicated for draining cerebrospinal fluid (CSF) from the cerebral ventricles or the lumbar subarachnoid space as a means of reducing intracranial pressure and CSF volume when the insertion of a permanent, internal shunt is not indicated.

    Device Description

    The Codman EDS3 CSF External Drainage System (Codman EDS3 System) is designed to collect cerebral spinal fluid (CSF) from the patient at a controlled rate based on differential pressure between the device and the patient. Collecting CSF from the patient is performed in efforts to reduce elevated intracranial pressure (ICP) post trauma. The EDS3 device is comprised of four (4) main parts: ventricular catheter, patient drain line, base frame burette tube assembly and a collection bag. The ventricular catheter is placed into one of the ventricles in the brain or in the subarachnoid space and is then connected to the patient drainage line. CSF flows from the brain through the patient line and enters into the 100 mL graduated burette tube assembly, where it is collected over a period of time to calculate a flow rate. The burette tube assembly can then be raised or lowered along the base frame, thereby adjusting the differential pressure to achieve the appropriate flow rate. Once the burette tube height is set, the collected CSF is then drained into the attached 700 ml collection bag. The EDS3 System is a complete, disposable unit that is provided sterile and is available with or without a ventricular catheter.

    AI/ML Overview

    The provided document describes the Codman EDS3 CSF External Drainage System (K162437), which is essentially an updated version of a previously cleared device (K061568). The primary change in the subject device is the use of a different adhesive (Loctite 3924 instead of Loctite 3341) for bonding tubing lines. Therefore, the acceptance criteria and studies focus on demonstrating that this change does not negatively impact the device's safety and effectiveness.

    Here's the breakdown of the information requested:

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestTest Method / PurposeAcceptance CriteriaReported Device Performance
    Visual InspectionPIC-CL205063 (Ensure finished product meets design specifications)Pass Visual StandardsPass
    Pressurized Leak and FlowTM-TM095 (Ensure leak-free product assembly)Pass Leak and Flow TestPass
    Tensile and Torque TestingTM-TM227, TM-TM100234 (Ensure product meets functional specifications)Pass Tensile and Torque Strength SpecificationPass
    Shelf LifeEnsure the product remains functional after sterilization and 3 years of aging.Meet functional criteria after 2x EtO sterilization and 3 years of agingPass
    BiocompatibilityISO 10993 (Evaluate the Loctite 3924 material to ensure that it is a safe substitute for Loctite 3341)Pass Cytotoxicity Testing, Pass Irritation and Sensitization Testing, No evidence of systemic Toxicity, Pass Dermal Irritation Test, No evidence of Pyrogenic response, No evidence of potential to cause Hemolysis, Pass Physiochemical attributes reviewPass

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes used for each bench test (e.g., number of devices tested for tensile strength or shelf life). It's a summary document, and typically, detailed test reports would contain this information.

    Regarding data provenance:

    • Country of origin: Not specified, but likely the US given the submission to the FDA.
    • Retrospective or prospective: The bench tests described are prospective in nature, as they involve testing newly manufactured devices with the new adhesive according to established protocols.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This question is not directly applicable in the context of this device and study. The "ground truth" for this type of device (a medical drainage system) is not established by expert consensus or interpretations in the same way it would be for an AI diagnostic algorithm. Instead, it's established by pre-defined performance specifications (acceptance criteria) based on engineering, material science, and regulatory standards. The "experts" involve the engineers and scientists who designed the tests and reviewed the results, but they are not adjudicating observational data.

    4. Adjudication Method for the Test Set:

    Not applicable. The tests are objective and based on measurable parameters (e.g., leak detection, force measurements, chemical analysis). There's no human "adjudication" of the test results in the way it's done for interpreting medical images. The results are directly compared to the pre-established numerical or qualitative "Pass/Fail" acceptance criteria.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices (especially those involving image interpretation by human readers, potentially augmented by AI) to assess the impact on human performance. The Codman EDS3 CSF External Drainage System is a physical medical device, not a diagnostic tool where human reader performance is a direct output.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    No, a standalone (algorithm-only) performance study was not done. This device is not an algorithm; it's a physical medical device. The "performance" refers to its physical and functional attributes.

    7. The Type of Ground Truth Used:

    The ground truth for this study is based on engineering specifications, material science standards (e.g., ISO 10993 for biocompatibility), and established medical device performance requirements. The acceptance criteria are essentially the "ground truth" against which the device's performance is measured. These are not pathology results, expert consensus on disease, or patient outcomes data in this context.

    8. The Sample Size for the Training Set:

    Not applicable. This device is a physical medical device, not an AI, machine learning, or software algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 7