K Number
K171862

Validate with FDA (Live)

Date Cleared
2017-09-25

(95 days)

Product Code
Regulation Number
882.5950
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GALAXY G3 Mini Microcoil Delivery System is intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations fistulae, and is also intended for arterial and venous embolizations in the peripheral vasculature.

Device Description

The GALAXY G3 Mini Microcoil Delivery Systems consist of three components, a Microcoil System, a connecting cable, and a Detachment Control Box (DCB). Each component is sold separately. As shown in Figure 1, the Microcoil System consists of a microcoil attached to a Device Positioning Unit (DPU). The Microcoil System is packaged in an introducer sheath designed to protect the coil in the packaging dispenser and to provide support for introducing the coil into the microcatheter catheter. The microcoil is the implantable segment of the device, and is detached from the Device Positioning Unit (DPU) using the Detachment Control System (Detachment Control Box and connecting cable). The microcoil is fabricated from a platinum alloy wire. The wire is wound into a primary coil which contains a polypropylene suture (SR) and then formed into a secondary shape. The secondary shape is complex. The DPU is a variable stiffness wire and has a radiopaque marker band located three (3) cm from its distal end. The Device Positioning Unit includes five (5) fluoro saver markers on the proximal section of the shaft. The markers are intended to indicate when the tip of the microcoil is approaching the tip of the microcatheter. When the distal-most marker reaches the proximal end of the Rotating Hemostatic Valve (RHV) on the microcatheter, the tip of the coil is approaching the tip of the microcatheter and fluoroscopy should be used to guide further coil insertion. The introducer sheath has three main components: an introducer tip, a translucent introducer body, and a re-sheathing tool. The EnPOWER Detachment Control Box (DCB) provides the energy necessary to allow for a thermo-mechanical detachment of the microcoil from the DPU. The connecting cable delivers the energy necessary to detach the embolic coil from the Microcoil System's detachment zone. The connecting cable is connected between the Microcoil System's hub connector on the DPU and the output connector on the DCB. The connecting cables may be one of two types: one with a remote detach button (the EnPower Control Cable) catalog no. ECB000182-00, or one without a detach button (standard connecting cable) catalog no. CCB00157-00. The EnPower Detachment Control Box works with the EnPower Control Cable and with the standard connecting cable.

AI/ML Overview

The provided document describes the development and testing of the GALAXY G3 Mini Microcoil Delivery System. Here's a breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document reports several performance tests, and for each, the result is "PASS: Samples passed the established acceptance criterion." The specific numerical acceptance criteria are generally not explicitly stated, but the passing result indicates they were met.

TestTest Method SummaryReported Device Performance
Spring ConstantMeasures the softness of the coil by recording the spring constant of the primary wind.PASS: Samples passed the established acceptance criterion
Complex ShapeVisually inspects the complex shape of the G3 Mini Microcoils.PASS: Samples passed the established acceptance criterion
ParticulateMeasures particulate count during simulated use per USP788.PASS: Samples passed the established acceptance criterion
Atraumatic BeadVisually verifies that the bead end of the coil meets the final assembly specification.PASS: Samples passed the established acceptance criterion
DPU 3 System Outer DiameterVerifies the OD is within specification to ensure microcatheter compatibility.PASS: Samples passed the established acceptance criterion
Microcatheter Tip Deflection ForceMeasures the deflection and/or stability of the microcatheter by recording the force generated at the distal tip as the DPU device is advanced to the tip.PASS: Samples passed the established acceptance criterion
Detachment Zone Tensile StrengthEvaluates the attachment strength of the detachment fiber to prevent unintentional coil detachments.PASS: Samples passed the established acceptance criterion
Stretch Resistance Fiber Tensile StrengthVerifies that the coil provides sufficient stretch resistance under tensile loading to ensure the coil can be retracted and repositioned without stretching.PASS: Samples passed the established acceptance criterion
Track Force (Delivery)Evaluates the force it takes to deliver the device through a microcatheter and into a clinically relevant model; utilizing the system Catheter Performance Simulation System (CPSS).PASS: Samples passed the established acceptance criterion
Dimensional Inspection of FDL Diameter and Coil LengthVerifies the FDL diameter and the coil length meets the specification.PASS: Samples passed the established acceptance criterion
Coil OD Verification on Final AssemblyVerifies the OD is within specification.PASS: Samples passed the established acceptance criterion
Dimensional Inspection of the Distal Fluro-saver markersVerifies that the Fluoro Saver Markers are in the correct proximal position to give a visual indication that the microcoil is approaching the distal tip of the microcatheter.PASS: Samples passed the established acceptance criterion
Coil DurabilityEvaluates the coil's ability to stay attached to the device during simulated use of six insertions and withdrawals cycled into and out of a clinically relevant aneurysm model.PASS: Samples passed the established acceptance criterion
Detachment, Coil Durability & ReliabilityEvaluates the reliability of the detachment mechanism after being cycled into and then out of a clinically relevant anatomical model six times.PASS: Samples passed the established acceptance criterion
Resheathing ReliabilityEvaluates the ability to re-insert the device into the split sheath introducer after it has been unzipped after the device has been inserted and withdrawn from a clinically relevant model.PASS: Samples passed the established acceptance criterion
Fluro saver Marker DurabilityEvaluates the ability of the Fluoro Saver Markers to stay affixed and in the correct position on the shaft after being cycled into and then out of a clinically relevant anatomical model six times.PASS: Samples passed the established acceptance criterion
Distal Outer Sheath DurabilityEvaluates the durability of the distal outer sheath during the simulated use of six insertions and withdrawals of the device into and out of a clinically relevant aneurysm model.PASS: Samples passed the established acceptance criterion
Dimensional Inspection of the IntroducerIntroducer underwent dimensional inspection per approved test method.PASS: Samples passed the established acceptance criterion
Introducer Bond StrengthVerifies that the bond strength of the introducer fuse joint.PASS: Samples passed the established acceptance criterion
Coil Transfer to MicrocatheterEvaluates the introducer sheath for allowing for insertion of the embolic coil into the microcatheter through the RHV.PASS: Samples passed the established acceptance criterion
Introducer FlushingVisually inspects to confirm flushing.PASS: Samples passed the established acceptance criterion
RadiopacityVerifies the ability to visualize the coil under fluoroscopy for physician to determine the location of the coil during use.PASS: Samples passed the established acceptance criterion
Overall Coil PerformanceEvaluates physician's satisfaction rating on performance compared to a predicate device.PASS: Samples passed the established acceptance criterion
MRI TestingDemonstrates that GALAXY G3 MINI Microcoil to be "MR-conditional" according to the specific conditions used for the assessment.PASS: Samples passed the established acceptance criterion

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the exact sample sizes used for each individual performance test (test sets). It generally refers to "samples" being tested.

The data provenance is bench testing, which implies the data was collected within a laboratory setting, likely in the US, given the submission to the FDA. It is retrospective in the sense that it's testing a finished device against predetermined criteria.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

No individual experts or their qualifications are mentioned for establishing ground truth for the bench tests. The "ground truth" for these tests is based on established engineering specifications, industry standards (e.g., USP788, ISO standards), and clinical relevance/simulated use scenarios. For the "Overall Coil Performance" test, "physician's satisfaction rating" was evaluated, implying input from medical professionals, but the number and qualifications are not specified.

4. Adjudication Method for the Test Set

Not applicable for the reported tests. The tests are primarily objective measurements against established criteria, or visual inspections by qualified personnel (not specified as "experts" in the context of adjudication).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. The document explicitly states: "A clinical study was not required as appropriate verification and validation of the GALAXY G3 Mini Microcoil Delivery System was achieved based on the similarities of the proposed device to the predicate device, and from results of bench testing."

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical medical device (microcoil delivery system), not an algorithm or AI software. Therefore, the concept of "standalone algorithm performance" does not apply. The performance tests are for the physical device itself.

7. The Type of Ground Truth Used

The ground truth for the bench tests is based on:

  • Engineering specifications and design requirements: For dimensional checks, material properties, and functional performance (e.g., spring constant, detachment force, tensile strength).
  • Industry standards: Such as USP788 for particulate matter, ISO 11607 for packaging, ISO 10993-1 for biocompatibility, and ISO 11135-1 for sterilization.
  • Simulated use in clinically relevant models: For tests like track force, coil durability, detachment durability, resheathing reliability, fluoro saver marker durability, and distal outer sheath durability.
  • Physician satisfaction rating: For "Overall Coil Performance" compared to a predicate device, which implies a subjective expert assessment.

8. The Sample Size for the Training Set

Not applicable. This is a physical medical device, not an AI or machine learning model that requires a "training set." The listed studies are verification and validation tests for the device itself.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked one behind the other, creating a sense of depth and unity.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 25, 2017

Codman & Shurtleff, Inc. Yoon Hee Beattv Sr. Regulatory Affairs Specialist 325 Paramount Drive Raynham, Massachusetts 02767

Re: K171862

Trade/Device Name: GALAXY G3 Mini Microcoil Delivery System Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: August 21, 2017 Received: August 23, 2017

Dear Ms. Beatty:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Carlos L. Pena -S

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171862

Device Name

GALAXY G3 Mini Microcoil Delivery System

Indications for Use (Describe)

The GALAXY G3 Mini Microcoil Delivery System is intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations fistulae, and is also intended for arterial and venous embolizations in the peripheral vasculature.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Codman & Shurtleff, Inc. I. Submitter 325 Paramount Drive Raynham, MA 02767

Tel: (305) 265-2919 Fax: (305) 265-6889

Contact Person: Yoon Hee Beatty Date Prepared: August 21, 2017

II. Device

Table 1. Device
Device Proprietary NameGALAXY G3 Mini Microcoil Delivery System
Common or Usual NameDevice, Neurovascular Embolization & Vascular, For PromotingEmbolization
Classification NameDevice, Neurovascular Embolization, Class II, 21 CFR 882.5950 &Vascular, For Promoting Embolization, Class II 21 CFR 870.3300
Regulatory ClassificationII
Product CodesHCG, KRD

The corresponding predicate devices that are listed in Table 2 below are applicable III. Predicate Device to the devices in this submission.

Table 2. Prior 510(k) Clearance
510(k) NumberDate ClearedNameManufacturer
Primary PredicateK15031906/12/2015Micrus Microcoil System DELTAXFTCodman & Shurtleff, Inc.
Reference PredicateK08364601/02/2009Micrus Microcoil System DELTAPLUSHMicrus EndovascularCorporation*
Reference DeviceK09397305/26/2010ORBIT GALAXY Detachable Coil SystemCodman & Shurtleff, Inc.
Reference DeviceK02159105/22/2002PROWLER SELECT Infusion CathetersCordis Neurovascular,Inc.**
On 09/27/10, Micrus Endovascular Corporation was acquired by Johnson & Johnson and now operates as a wholly-ownedsubsidiary of Codman & Shurtleff, Inc within the Johnson & Johnson family of companies. Medos International SARL (Medos) willbe the recognized legal manufacturer and Codman & Shurtleff, Inc. will be the subcontractor appointed by Medos.* This 510(k) was originally owned by Cordis Neurovascular, Inc., a subsidiary of Cordis Corporation. In 2008, CordisNeurovascular, Inc., was merged with Codman and Shurtleff, Inc. This 510(k) is presently owned by Codman & Shurtleff, Inc.

{4}------------------------------------------------

IV. Device The GALAXY G3 Mini Microcoil Delivery Systems consist of three components, a Description Microcoil System, a connecting cable, and a Detachment Control Box (DCB). Each component is sold separately. As shown in Figure 1, the Microcoil System consists of a microcoil attached to a Device Positioning Unit (DPU).

Image /page/4/Figure/4 description: The image is titled "Figure 1. Microcoil System" and shows a diagram of the microcoil system. The diagram labels the different parts of the system, including the hub connector, re-sheathing tool, introducer sheath body, introducer sheath tip, device positioning unit (DPU), and microcoil. The microcoil is located at the tip of the system and is a small, coiled wire.

The Microcoil System is packaged in an introducer sheath designed to protect the coil in the packaging dispenser and to provide support for introducing the coil into the microcatheter catheter. The microcoil is the implantable segment of the device, and is detached from the Device Positioning Unit (DPU) using the Detachment Control System (Detachment Control Box and connecting cable).

  • The microcoil is fabricated from a platinum alloy wire. The wire is wound into a primary coil which contains a polypropylene suture (SR) and then formed into a secondary shape. The secondary shape is complex.
  • o The DPU is a variable stiffness wire and has a radiopaque marker band located three (3) cm from its distal end. The Device Positioning Unit includes five (5) fluoro saver markers on the proximal section of the shaft. The markers are intended to indicate when the tip of the microcoil is approaching the tip of the microcatheter. When the distal-most marker reaches the proximal end of the Rotating Hemostatic Valve (RHV) on the microcatheter, the tip of the coil is approaching the tip of the microcatheter and fluoroscopy should be used to guide further coil insertion.
  • The introducer sheath has three main components: an introducer tip, a translucent introducer body, and a re-sheathing tool.

The EnPOWER Detachment Control Box (DCB) provides the energy necessary to allow for a thermo-mechanical detachment of the microcoil from the DPU. The connecting cable delivers the energy necessary to detach the embolic coil from the Microcoil System's detachment zone. The connecting cable is connected between the Microcoil System's hub connector on the DPU and the output connector on the DCB.

  • The connecting cables may be one of two types: one with a remote detach button (the EnPower Control Cable) catalog no. ECB000182-00, or one without a detach button (standard connecting cable) catalog no. CCB00157-00.
  • . The EnPower Detachment Control Box works with the EnPower Control Cable and with the standard connecting cable.

{5}------------------------------------------------

V. Indications for Use

The GALAXY G3 Mini Microcoil Delivery System is intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae, and is also intended for arterial and venous embolizations in the peripheral vasculature.

{6}------------------------------------------------

VI. Comparison of Technological Characteristics With Predicate Device

Endovascular coil embolization is the technological principle for both the subject and predicate devices. It is based on placing embolic coils in the neurovascular or peripheral vasculature in order to reduce or block blood flow. The subject device and predicate devices are based on the following same technological characteristics:

Table 3. Technological Characteristics of the Predicate and Proposed Devices
DescriptionPredicate Device:DELTAXSFTMicrocoil Delivery Systems(K150319)This Submission:GALAXY G3 MiniMicrocoil Delivery Systems
Indications for UseDELTAXSFT Microcoil DeliverySystems are intended forendovascular embolization ofintracranial aneurysms, otherneurovascular abnormalities such asarteriovenous malformations andarteriovenous fistulae, and are alsointended for arterial and venousembolizations in the peripheralvasculature.Same as predicateGALAXY G3 Mini Microcoil DeliverySystem is intended for endovascularembolization of intracranial aneurysms,other neurovascular abnormalities suchas arteriovenous malformations andarteriovenous fistulae, and is alsointended for arterial and venousembolizations in the peripheralvasculature.
Microcoil
Microcoil MaterialPlatinum/TungstenSame as predicate
Microcoil Primary WindTriangularCylindrical*
Microcoil Secondary ShapeHelicalComplex*
Microcoil Stretch-ResistantPGA= Polyglycolic Acid SuturePP= Polypropylene Suture**
Primary Coil Wind OuterDiameter0.0096"0.009"
Secondary Shape OuterDiameter Ranges1.5mm-4mm1.0mm - 3.0mm
Microcoil Length Ranges1cm-8cmSame as predicate
Delivery System
Delivery System TypeWire Shaft with radiopaque markerSame as predicate
Delivery System IntroducerSheathHDPE IntroducerSame as predicate
Delivery System ResheathingToolNylon 12Same as predicate
Flush PortsNoneFlush ports
Introducer Sheath Tip ShapeStraightTapered
Introducer Sheath Length120cm81cm and 120cm
Device Positioning Unit (DPU3)Delivery System Length190cm ± 5cmSame as predicate
Device Positioning UnitDiameter0.0159"Same as predicate
Fluoroscopy Saver MarkersFive Markers Located on theProximal Section of the ShaftSame as predicate
Fluoro Saver MarkerMicrocatheter Compatibility150cm LengthSame as predicate
Distal Segment of the DevicePositioning UnitDevice Positioning Unit 3 (DPU3)Same as predicate
Mechanism of DetachmentConnection to Microcoil System:Uses Connecting Cable orEnPOWER Control CableDetachment: Thermo-MechanicalSystem uses the EnPOWERDetachment Control Box (DCB)with EnPOWER Control Cable orConnecting CableSame as predicate
Sterilization and Shelf Life
Sterilization MethodE-Beam RadiationEthylene Oxide (EtO)*
Shelf Life3 yearsSame as predicate
PackagingPackaged in a plastic hoop andenclosed in a pouch.Same as predicate***
* Same shape and sterilization method as ORBIT GALAXY DCS (K093973)
** Same material as DELTAPLUSH Microcoil (K083646)
*** Same pouch material as PROWLER SELECT Infusion Catheters (K021591)

{7}------------------------------------------------

VII. Performance Data

Verification and Validation Testing

Verification and validation activities were conducted on the finished sterile device based on current standards and FDA Guidance Document; "Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices". The following performance data were provided in support of the substantial equivalence determination. The bench testing included the following tests:

Table 4. Verification and Validation Testing
TestTest Method SummaryResults
SpringConstantThe purpose of the Spring Constant test is tomeasure the softness of the coil by recording thespring constant of the primary wind.PASS:Samples passed theestablished acceptancecriterion
ComplexShapeThe purpose of the Complex Shape test is tovisually inspect the complex shape of the G3Mini Microcoils.PASS:Samples passed theestablished acceptancecriterion
ParticulateThe purpose of the Particulate test is to measureparticulate count during simulated use per therequirements of USP788.PASS:Samples passed theestablished acceptancecriterion
AtraumaticBeadThe purpose of the Atraumatic Bead test is tovisually verify that the bead end of the coilmeets the final assembly specification.PASS:Samples passed theestablished acceptancecriterion
DPU 3System OuterDiameterThe purpose of the DPU 3 System OuterDiameter test is to verify the OD is withinspecification to ensure microcathetercompatibility.PASS:Samples passed theestablished acceptancecriterion
MicrocatheterTipDeflectionForceThe purpose of the Microcatheter Stability (TipDeflection Force) test is to measure thedeflection and/or stability of the microcatheterby recording the force generated at the distal tipof the microcatheter as the DPU device isadvanced to the tip of the microcatheter.PASS:Samples passed theestablished acceptancecriterion
DetachmentZone TensileStrengthThe purpose of the Detachment Zone TensileStrength test was to evaluate the attachmentstrength of the detachment fiber to preventunintentional coil detachmentsPASS:Samples passed theestablished acceptancecriterion
StretchResistanceFiber TensileStrengthThe objective of test is to verify that the coilprovides sufficient stretch resistance undertensile loading to ensure the coil can be retractedand repositioned without stretching.PASS:Samples passed theestablished acceptancecriterion
VII.Table 4. Verification and Validation Testing, continued
PerformanceData,continuedTestTest Method SummaryResults
Track Force(Delivery)The purpose of the Track Force test was toevaluate the force it takes to deliver theproposed device through a microcatheter andinto a clinically relevant model; utilizing thesystem Catheter Performance Simulation System(CPSS).PASS:Samples passed theestablished acceptancecriterion
DimensionalInspection ofFDLDiameter andCoil LengthThe objective of test is to verify the FDLdiameter and the coil length meets thespecification.PASS:Samples passed theestablished acceptancecriterion
Coil ODVerificationon FinalAssemblyThe purpose of the Dimensional Inspection ofthe Outer Diameter of the proposed device is toverify the OD is within specification.PASS:Samples passed theestablished acceptancecriterion
DimensionalInspection ofthe DistalFluro-savermarkersThe purpose of the Dimensional Inspection ofthe Distal Fluoro Saver Marker was to verifythat the Fluoro Saver Markers are in thecorrect proximal position in order to give thephysician a visual indication that the microcoil isapproaching the distal tip of the microcatheter.PASS:Samples passed theestablished acceptancecriterion
CoilDurabilityThe purpose of the Coil Durability test was toevaluate the coil's ability to stay attached to theproposed device during simulated use of sixinsertions and withdrawals cycled into and outof a clinically relevant aneurysm model.PASS:Samples passed theestablished acceptancecriterion
Detachment,CoilDurability &ReliabilityThe purpose of the Coil Detachment Durability& Reliability test was to evaluate the reliabilityof the detachment mechanism of the proposeddevice after being cycled into and then out of aclinically relevant anatomical model six times.PASS:Samples passed theestablished acceptancecriterion
ResheathingReliabilityThe purpose of the Re-Sheathing Reliability testwas to evaluate the ability to re-insert theproposed device into the split sheath introducerafter it has been unzipped after the proposeddevice has been inserted and withdrawal from aclinically relevant model.PASS:Samples passed theestablished acceptancecriterion
Fluro saverMarkerDurabilityThe purpose of the Fluoro Saver MarkerDurability test was to evaluate the ability of theFluoro Saver Markers to stay affixed and in thecorrect position on the shaft of the proposeddevice after being cycled into and then out of aclinically relevant anatomical model six times.PASS:Samples passed theestablished acceptancecriterion

{8}------------------------------------------------

510(k) Summary, Continued

{9}------------------------------------------------

VII.
PerformanceTest
Data, (Cont.)
Distal OuterSheathDurability
Table 4. Verification and Validation Testing, continued
TestTest Method SummaryResults
Distal OuterSheathDurabilityThe purpose of the Distal Outer SheathDurability test was to evaluate the durability ofthe distal outer sheath during the simulated useof six insertions and withdrawals of theproposed device into and then out of a clinicallyrelevant aneurysm model.PASS:Samples passed theestablished acceptancecriterion
DimensionalInspection oftheIntroducerThe introducer underwent dimensionalinspection per approved test method.PASS:Samples passed theestablished acceptancecriterion
IntroducerBondStrengthThe objective of the test is to verify that thebond strength of the introducer fuse joint.PASS:Samples passed theestablished acceptancecriterion
Coil TransfertoMicrocatheterThe purpose of the test is to evaluate theintroducer sheath for allowing for insertion ofthe embolic coil into the microcatheter throughthe RHV.PASS:Samples passed theestablished acceptancecriterion
IntroducerFlushingThe purpose of the test is to visually inspect toconfirm flushing.PASS:Samples passed theestablished acceptancecriterion
RadiopacityThe purpose of the test is to verify the ability tovisualize the coil under fluoroscopy forphysician to determine the location of the coilduring use.PASS:Samples passed theestablished acceptancecriterion
Overall CoilPerformanceThe objective of the test is to evaluatephysician's satisfaction rating on performancecompared to a predicate device.PASS:Samples passed theestablished acceptancecriterion
MRI TestingThe purpose of the test is to demonstrate thatGALAXY G3 MINI Microcoil to be "MR-conditional" according to the specific conditionsused for the assessment.PASS:Samples passed theestablished acceptancecriterion

Shelf-Life Testing

Codman successfully completed accelerated aging testing to support a shelf life of 12 months. Finished sterile devices were subjected to environmental conditioning, simulated transportation and accelerated aging before being tested.

{10}------------------------------------------------

VII. Performance Packaging

Data (Cont.)

The modification to the current packaging consists of a change in pouch material due to the change in sterilization method from e-beam to ethylene oxide. Utilizing previously validated pouch material used in the same sterilization cycle (K021591), the new packaging configuration was validated in accordance with applicable recommendations in International Standard ISO 11607-1 "Packaging for Terminally Sterilized Medical Devices - Part 1: Requirements for Materials, Sterile Barrier Systems, and Packaging Systems" and ISO 11607-2 "Packaging for Terminally Sterilized Medical Devices - Part 2: Validation Requirements for Forming, Sealing, and Assembly Processes" as recognized by FDA.

Biocompatibility Testing

Biocompatibility testing was conducted in accordance with International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation of Testing Within a Risk Management Process" (ISO 10993-1:2009/2010AC), and FDA's Guidance Document entitled "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" (June 16. 2016).

The following tests were performed:

  • ISO MEM Elution
  • Hemocompatibility Tests ●
    • O In Vitro Hemolysis Direct Contact (ASTM F756)
    • O In Vitro Hemolysis - Extract
    • In Vivo Thromboresistance O
  • Chemical Characterization of Extractables (ISO 10993-18)

Sterilization

The GALAXY G3 Mini Microcoil System is ethylene oxide sterilized and was validated to assure a Sterility Assurance Level (SAL) 10-6 according to the requirements of International Standards ISO 11135-1 "Sterilization of Health Care Products Ethylene Oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices, and ISO 10993-7 "Biological Evaluation of medical devices - Part 7: Ethylene Oxide sterilization residual" as recognized by FDA.

{11}------------------------------------------------

VII. Performance Animal Study

Data (Cont.)

An animal study was not required as appropriate verification and validation of the GALAXY G3 Mini Microcoil Delivery System was achieved based on the similarities of the proposed device to the predicate device, and from results of bench testing.

Summary of Clinical Testing

A clinical study was not required as appropriate verification and validation of the GALAXY G3 Mini Microcoil Delivery System was achieved based on the similarities of the proposed device to the predicate device, and from results of bench testing.

Based upon the design, materials, function, intended use, performance, VIII. Conclusion manufacturing and sterilization process and the non-clinical testing performed by Codman, it is concluded that the GALAXY G3 Mini Microcoil Delivery Systems are substantially equivalent to the currently marketed CODMAN DELTAXSFT Microcoil Delivery Systems (K150319) and therefore, does not raise any new questions of safety or effectiveness.

9

§ 882.5950 Neurovascular embolization device.

(a)
Identification. A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.(b)
Classification. Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).