Search Filters

Search Results

Found 24 results

510(k) Data Aggregation

    K Number
    K241780

    Validate with FDA (Live)

    Date Cleared
    2024-08-23

    (64 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs per ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter for Cardinal Health Nitrile Exam Gloves. While it confirms the device is substantially equivalent to legally marketed predicate devices, it does not contain information about acceptance criteria or a study design in the typical format of an AI/ML device submission.

    This type of device (nitrile exam gloves) is a physical medical device, not a software device or an AI/ML product. The "acceptance criteria" and "study" described in the input prompt are typically associated with the evaluation of performance claims for AI/ML algorithms, which involve metrics like accuracy, sensitivity, specificity, and rigorous testing on curated datasets.

    Here's what can be extracted and inferred from the provided text, acknowledging that it doesn't align with the detailed AI/ML study request:

    1. A table of acceptance criteria and the reported device performance:

    The closest information provided is the permeation resistance to chemotherapy drugs as tested per ASTM D6978. This acts as a performance standard for this specific "indication for use."

    Chemotherapy Drug and ConcentrationAcceptance Criteria (Implicit)Reported Device Performance (Minimum Breakthrough Detection Time in Minutes)
    Carmustine (BCNU) (3.3 mg/ml)(Implied to be > 30.1 minutes for safe use, but warning issued)30.1
    Cisplatin (1.0 mg/ml)(Implied to be > 240 minutes for safe use)>240
    Cyclophosphamide (Cytoxan) (20 mg/ml)(Implied to be > 240 minutes for safe use)>240
    Cytarabine (100 mg/ml)(Implied to be > 240 minutes for safe use)>240
    Dacarbazine (DTIC) (10.0 mg/ml)(Implied to be > 240 minutes for safe use)>240
    Doxorubicin Hydrochloride (2.0 mg/ml)(Implied to be > 240 minutes for safe use)>240
    Etoposide (20.0 mg/ml)(Implied to be > 240 minutes for safe use)>240
    Fluorouracil (50.0 mg/ml)(Implied to be > 240 minutes for safe use)>240
    Ifosfamide (50.0 mg/ml)(Implied to be > 240 minutes for safe use)>240
    Methotrexate (25 mg/ml)(Implied to be > 240 minutes for safe use)>240
    Mitomycin C (0.5 mg/ml)(Implied to be > 240 minutes for safe use)>240
    Mitoxantrone (2.0 mg/ml)(Implied to be > 240 minutes for safe use)>240
    Paclitaxel (Taxol) (6.0 mg/ml)(Implied to be > 240 minutes for safe use)>240
    Thiotepa (10.0 mg/ml)(Implied to be > 50.5 minutes for safe use, but warning issued)50.5
    Vincristine Sulfate (1.0 mg/ml)(Implied to be > 240 minutes for safe use)>240

    Note: The "acceptance criteria" here are implied by the testing standard ASTM D6978 and the explicit warnings for drugs with low permeation times. For most drugs, " >240 minutes" appears to be the desired performance, and the device meets this for the majority. For Carmustine and Thiotepa, where performance is lower, specific warnings are issued.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified in the provided text. The ASTM D6978 standard would dictate the number of gloves tested.
    • Data Provenance: Not specified, but generally for such testing, it would be laboratory testing conducted according to the ASTM standard. It's not "country of origin of data" in the sense of patient data, nor is it retrospective or prospective as it's a materials performance test.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This is not a diagnostic device requiring expert interpretation of medical images or data. The "ground truth" here is the physical permeation time measured by laboratory equipment following the ASTM D6978 standard.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This refers to consensus building among experts for tasks like image labeling or diagnosis, which is not relevant for glove permeation testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. The device is a physical medical glove, not an AI/ML diagnostic aid.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. The device is a physical medical glove, not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The ground truth is the measured breakthrough detection time for each chemotherapy drug, obtained through standardized laboratory testing (ASTM D6978).

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230838

    Validate with FDA (Live)

    Date Cleared
    2023-07-27

    (122 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cardinal Health™ SMARTGOWN™ Surgical Gown with ASSIST™ Instrument Pockets is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. The gowns meet the barrier protection requirements of AAMI Level 4 per ANSI/ AAMI PB70 Liquid Barrier and Performance Classification of Protective Apparel and Surgical Drapes Intended for Use in Health Care Facilities. The Cardinal Health™ SMARTGOWN™ Surgical Gown with ASSIST™ Instrument Pockets is a single use, disposable medical device provided sterile and non-sterile. Non-sterile gowns are to be sold to re-packager/re-labeler establishments for ethylene oxide (EtO) sterilization according to ISO 11135 prior to marketing to the end user

    Device Description

    Not Found

    AI/ML Overview

    This FDA 510(k) clearance letter pertains to a surgical gown, not an AI/ML device. Therefore, the requested information regarding acceptance criteria and study details for an AI/ML device is not present in the provided text.

    The document discusses the regulatory clearance of a physical medical device: Cardinal Health™ SmartGown™ EDGE Breathable Surgical Gown with ASSIST™ Instrument Pockets.

    Here's what the document does provide regarding the device and its compliance:

    • Regulation Number: 21 CFR 878.4040
    • Regulation Name: Surgical Apparel
    • Regulatory Class: Class II
    • Product Code: FYA
    • Intended Use: To be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.
    • Performance Standard: The gowns meet the barrier protection requirements of AAMI Level 4 per ANSI/AAMI PB70 Liquid Barrier and Performance Classification of Protective Apparel and Surgical Drapes Intended for Use in Health Care Facilities. This reference to AAMI Level 4 and ANSI/AAMI PB70 is the primary "acceptance criterion" mentioned, indicating the device's barrier performance.

    Since this is a traditional medical device, not an AI/ML product, the questions related to AI/ML specific criteria (sample sizes for test/training sets, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, etc.) are not applicable here.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Monoject Magellan Insulin Safety Syringes are intended for delivery of U-100 insulin. The needle stick prevention feature of the device, once activated, guards against accidental needle-sticks.

    Device Description

    The Monoject™ Magellan™ Insulin Safety Syringe consists of a piston syringe with a permanently attached needle, and safety shield designed to extend, fully cover, and permanently lock over the needle when activated by a fingertip or thumb to reduce the occurrence of accidental needle sticks. The device is provided sterile and is intended for single use only. An orange needle cap indicates for use with insulin. The barrel of the syringe is graduated for measuring a prescribed dose. The syringe is intended for use immediately after filling and is not intended to contain or store insulin for extended periods of time.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Monoject Magellan Insulin Safety Syringe). It details the device's characteristics, intended use, and comparison to a predicate device to establish substantial equivalence.

    However, this document does not contain any information about an AI/ML-driven device or its acceptance criteria and study results. The device described is a physical medical device (syringe), not a software or AI/ML product.

    Therefore, I cannot fulfill the request to describe acceptance criteria and associated study results for an AI/ML device based on the provided text. The questions posed (e.g., sample size for test set, number of experts, MRMC study, standalone performance, training set data) are relevant to the evaluation of AI/ML models, but the provided document is not about such a device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201592

    Validate with FDA (Live)

    Date Cleared
    2021-09-26

    (471 days)

    Product Code
    Regulation Number
    878.4460
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cardinal Health™ Protexis™ PI Blue with Neu-Thera™ Surgical Gloves are powder-free surgeon's gloves that are disposable devices made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

    In addition, these gloves have been tested for resistance to permeation of various chemotherapy drugs per ASTM D6978, "Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs."

    Device Description

    powder-free surgeon's gloves that are disposable devices made of synthetic rubber

    AI/ML Overview

    The provided text describes the 510(k) clearance for "Cardinal Health™ Protexis™ PI Blue with Neu-Thera™ Surgical Gloves". This document primarily focuses on the regulatory clearance for a physical medical device (surgical gloves) and its resistance to chemotherapy drug permeation. It does NOT describe an AI/ML-driven medical device, an algorithm's performance, or any associated clinical study involving human readers or expert ground truth establishment for such an algorithm.

    Therefore, I cannot fulfill your request for information related to acceptance criteria and studies for an AI device. The questions you've asked are specifically tailored for AI/ML device evaluations (e.g., sample sizes for test/training sets, experts for ground truth, MRMC studies, standalone performance).

    Based on the provided document, the "acceptance criteria" and "study" are related to the physical properties of the surgical gloves as follows:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The primary acceptance criteria and reported performance for these surgical gloves, as detailed in the "Indications for Use" section, relate to their resistance to permeation by various chemotherapy drugs. This is assessed by reporting the minimum breakthrough detection time.

    Chemotherapy DrugAcceptance Criteria (Conceptual)Reported Device Performance (min. breakthrough detection time in minutes, 0.01 µg/cm²/min)
    Carmustine (3.3 mg/mL)(Not explicitly stated as a pass/fail threshold, but higher is better)18.5
    Cisplatin (1 mg/mL)(Not explicitly stated as a pass/fail threshold, but higher is better)>240
    Cyclophosphamide (20 mg/mL)(Not explicitly stated as a pass/fail threshold, but higher is better)>240
    Doxorubicin HCl (2 mg/mL)(Not explicitly stated as a pass/fail threshold, but higher is better)>240
    Etoposide (20 mg/mL)(Not explicitly stated as a pass/fail threshold, but higher is better)>240
    5-Fluorouracil (50 mg/mL)(Not explicitly stated as a pass/fail threshold, but higher is better)>240
    Methotrexate (25 mg/mL)(Not explicitly stated as a pass/fail threshold, but higher is better)>240
    Mitomycin C (0.5 mg/mL)(Not explicitly stated as a pass/fail threshold, but higher is better)>240
    Paclitaxel (6 mg/mL)(Not explicitly stated as a pass/fail threshold, but higher is better)>240
    Thiotepa (10 mg/mL)(Not explicitly stated as a pass/fail threshold, but higher is better)24.4
    Vincristine Sulfate (1 mg/mL)(Not explicitly stated as a pass/fail threshold, but higher is better)>240

    Explanation of the Study:
    The study proving the device meets its performance claims is referenced as testing "per ASTM D6978, 'Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.'" This is a standardized laboratory test protocol aimed at evaluating the barrier properties of medical gloves against hazardous drugs.

    I cannot answer the following questions because the document pertains to a physical medical device (gloves) and not an AI/ML-driven device:

    • 2. Sample sized used for the test set and the data provenance: Not applicable to a physical glove's chemical resistance test in the context of AI.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for gloves is defined by the chemical testing method.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is not an AI.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): The ground truth for glove performance is established through standardized chemical permeation testing, not expert consensus or medical outcomes.
    • 8. The sample size for the training set: Not applicable. This is a manufactured product, not an AI model.
    • 9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document clearly indicates a regulatory submission for surgical gloves, with performance claims related to their chemical resistance. It does not contain any information about an AI/ML device or its associated clinical and technical studies related to AI model evaluation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211390

    Validate with FDA (Live)

    Date Cleared
    2021-08-28

    (115 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these gloves have been tested for resistance to permeation of various chemotherapy drugs per ASTM D6978, "Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs."

    Device Description

    Nitrile Blue Powder-Free Examination Gloves

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding "Nitrile Blue Powder-free Examination Gloves." It is a clearance letter, not a study report for an AI/ML medical device. Therefore, it does not contain the information requested in your prompt regarding acceptance criteria and performance studies of a device related to AI, image analysis, or similar technologies.

    The document only provides information about the chemical permeation resistance of the gloves to various chemotherapy drugs.

    Thus, I cannot fulfill your request for the following reasons:

    • No AI/ML Device: The device described (nitrile examination gloves) is a physical medical device, not a software or AI/ML-based device.
    • No Performance Study Data of the Type Requested: The document details the results of ASTM D6978, which is a chemical permeation test, not a performance study involving test sets, expert readers, ground truth, or MRMC analysis for an AI algorithm.
    • No Acceptance Criteria for AI/ML Performance: The "acceptance criteria" presented are minimum breakthrough times for chemotherapy drugs, not metrics like sensitivity, specificity, AUC, or other measures common for AI/ML diagnostic tools.

    Therefore, I cannot extract the information asked for in points 1-9 of your prompt from the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200824

    Validate with FDA (Live)

    Date Cleared
    2020-07-29

    (121 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cardinal Health™ SMARTGOWN™ Breathable Surgical Gown and Cardin™ SMARTGOWN™ AIR Breathable Surgical Gown are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. The gowns meet the barrier protection requirements of AAMI Level 4 per AAMI PB70:2012 Liquid Barrier and Performance Classification of Protective Apparel and Surgical Drapes Intended for Use in Health Care Facilities. The Cardinal Health™ SMARTGOWN™ Breathable Surgical Gown and Cardinal Health™ AIR Breathable Surgical Gown are single use, disposable medical devices provided sterile and non-sterile.

    Device Description

    Cardinal Health™ SMARTGOWN™ Breathable Surgical Gown and Cardinal Health™ SMARTGOWN™ AIR Breathable Surgical Gown are single use, disposable medical devices provided sterile and non-sterile.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for surgical gowns, which are physical medical devices. This type of FDA submission focuses on demonstrating substantial equivalence to a predicate device through performance testing of the physical properties and barrier protection of the gowns.

    Therefore, the concepts of acceptance criteria, study design, sample sizes, expert ground truth, adjudication methods, multi-reader multi-case studies, standalone performance, and training sets are not applicable in the context of this document. These concepts are relevant to the evaluation of AI/ML-driven software as a medical device (SaMD) or devices that incorporate such software, which rely on data analysis and interpretation.

    The document describes the performance criteria for the surgical gowns, which are based on physical barrier protection standards.

    Here's the relevant information from the document related to "acceptance criteria" and "device performance":

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Meet the barrier protection requirements of AAMI Level 4 per AAMI PB70:2012 Liquid Barrier and Performance Classification of Protective Apparel and Surgical Drapes Intended for Use in Health Care Facilities.The gowns meet the barrier protection requirements of AAMI Level 4 per AAMI PB70:2012.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not detailed in the provided document. The 510(k) summary (which is typically a separate document) or the full submission would contain details about the specific tests performed (e.g., synthetic blood penetration, hydrostatic pressure, impact penetration) and the number of samples tested to demonstrate AAMI Level 4 compliance. This involves physical testing of the gowns, not analysis of medical image data.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. Ground truth for surgical gowns is established through standardized laboratory testing protocols, not by expert medical interpretation.

    4. Adjudication Method

    Not applicable.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is relevant for diagnostic or interpretive AI/ML devices.

    6. Standalone Performance Study

    Not applicable in the context of a physical surgical gown. The "standalone performance" of the gown refers to its ability to meet the AAMI Level 4 barrier standards independently, which is what the submission attests to.

    7. Type of Ground Truth Used

    The "ground truth" for surgical gowns is defined by the objective, quantifiable results of standardized physical tests according to AAMI PB70:2012. This involves tests for fluid resistance, microbial barrier properties, etc.

    8. Sample Size for the Training Set

    Not applicable. Surgical gowns are physical devices; there is no "training set" in the AI/ML sense.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193327

    Validate with FDA (Live)

    Date Cleared
    2020-03-20

    (109 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cardinal Health™ ChemoPlus™ Full Coverage Gowns, Closed Back, are intended to protect healthcare personnel from exposure to chemotherapy drugs during preparation, handling, and administration.

    The closed back gowns are single use, disposable medical devices provided sterile and non-sterile. The non-sterile, closed back gowns are not intended for use in the operating room.

    The Cardinal Health™ ChemoPlus™ Full Coverage Gowns, Open Back, are intended to protect healthcare personnel from exposure to chemotherapy drugs during preparation, handling, and administration.

    The open back gowns are single use, disposable medical devices provided non-sterile. The open back gown is nonprotective in the back. The open back gowns are not intended for use in the operating room.

    Device Description

    The Cardinal Health™ ChemoPlus™ Full Coverage Closed Back and Open Back Gowns are classified as Class II medical devices under Regulation 21 CFR 878.4040, General & Plastic Surgery Panel, FDA product code FYA, Surgical Gown.

    The ChemoPlus™ Full Coverage Closed Back and Open Back Gowns are made of a laminate with adhesive taped seams. Both closed and open back gowns have a hook and loop closure at the back of the neck and a waist tie feature to secure the gown to the body of the user. The sleeves of the gown have knit cuffs sewn onto the end of the sleeve at the user's wrists to keep the sleeves in place on the wearer. The entire gown including the gown sleeves are made of the same material and utilize the same manufacturing processes.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Cardinal Health™ ChemoPlus™ Full Coverage Gowns, indicating that these are medical devices subject to FDA regulation. The document describes non-clinical performance testing for these gowns.

    It's important to note that this document pertains to physical medical devices (gowns), not an AI/ML powered device or a software as a medical device (SaMD). Therefore, many of the typical acceptance criteria and study components relevant to AI/ML devices (like sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details, etc.) are not applicable in this context.

    The study described is a series of non-clinical, laboratory-based performance tests conducted on the gowns themselves, adhering to specific industry standards.

    Here's the information from the document relevant to acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    Standards TypePurposeAcceptance CriteriaReported Device Performance
    Liquid Barrier
    AATCC 42Spray ImpactWater ResistancePASS
    AATCC 127Hydrostatic PressureWater ResistancePASS
    ANSI/AAMI PB70:2012 AAMI Level 3Liquid Barrier PerformanceWater Resistance (AAMI Level 3 requirements)PASS
    Chemical Permeation
    ASTM F739-12Permeation of Liquids and Gases through Protective ClothingChemical Permeation Under Continuous ContactPASS (>480 minutes breakthrough for specified chemotherapy drugs with no permeation detected at minimum detectable permeation or 0.1 µg/cm²/min)
    Material Properties
    ASTM D3776/D3776M-17Weight of Woven FabricMaterial WeightPASS
    ASTM D5034 - 9 2017Grab Tensile, Peak Stretch, and Peak Energy – NonwovensTensile StrengthPASS
    ASTM 1683-17 (2018)Standard Test Method for Failure in Sewn Seams of Woven Apparel FabricsSeam StrengthPASS
    ASTM D5587-15 (2019)Standard Test Method for Tearing Strength of Fabrics by the Trapezoid ProcedureTrapezoid TearPASS
    NWSP 160.1Resistance to Linting of Nonwoven Fabrics (Dry)ParticulatePASS
    Flammability
    16 CFR 1610Standard for the Flammability for Clothing TextilesFlammability (Meet Class I)PASS
    Biocompatibility
    ISO 10993-5:2009ISO MEM Elution CytotoxicityCytotoxicity (Non-cytotoxic)PASS
    ISO 10993 10:2010ISO Indirect Primary Skin Irritation TestIrritation (Non-irritating)PASS
    ISO 10993-10:2010Guinea Pig Maximization TestSensitization (Non-sensitizing)PASS
    Sterilization Residuals
    ISO 10993-7:2008 (R) 2012EO Sterilization Residuals (Sterile Catalog Numbers)EO and ECH ResidualsPASS

    Study Proving Acceptance Criteria:
    The study involved "Non-Clinical Performance Testing" of the Cardinal Health™ ChemoPlus™ Full Coverage Gowns (both closed and open back). This testing was conducted according to various recognized industry standards (AAMI, ASTM, AATCC, NWSP, CFR, ISO) to evaluate specific properties of the gowns. The results for all tests, as reported in the document, uniformly indicated "PASS," demonstrating that the proposed devices meet the acceptance criteria for their intended use.

    Specific details for the chemical permeation test (ASTM F739-12) include:

    • Chemotherapy drugs tested: Carmustine, Cisplatin, Cyclophosphamide, Cytarabine Hydrochloride, Dacarbazine, Daunorubicin Hydrochloride, Doxorubicin Hydrochloride, Etoposide, Fluorouracil, Ifosfamide, Methotrexate, Mitomycin C, Mitoxantrone, Paclitaxel, Thiotepa, Vincristine Sulfate.
    • Reported performance: Average standardized breakthrough time for all tested chemotherapy drugs was ">480 minutes."
    • Detection limit: No permeation was detected at either the minimum detectable permeation or 0.1 µg/cm²/min for any of the tested drugs.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The tests are non-clinical, material performance tests, where "sample size" would refer to the number of gown material specimens tested according to the specific test standard's requirements, rather than patient data. Data provenance (country, retrospective/prospective) is not relevant for this type of materials testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable. The "ground truth" for material performance criteria is established by the accepted scientific and engineering principles embedded in the referenced ASTM, ISO, AAMI, etc., standards, and the measurement equipment used for testing. It does not involve human expert consensus in the way an AI/ML medical image analysis device would.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as the evaluation is based on objective measurements against defined technical standards, not subjective human interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This is not an AI/ML device, nor does it involve human readers or diagnostic tasks.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This is not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for these tests is defined by the objective measurement protocols and acceptance limits set forth in the various referenced international and national standards (e.g., AATCC 42, ASTM F739-12, ISO 10993-5:2009). For example, chemical permeation is measured directly using analytical methods, not expert consensus or pathology.

    8. The sample size for the training set

    This is not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    This is not applicable. This is not an AI/ML device that requires a training set or ground truth for training.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192374

    Validate with FDA (Live)

    Date Cleared
    2019-12-02

    (94 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cardinal Health™ Level 3 surgical masks with Anti-Fog Foam Strip are intended to be worn by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and airborne particulates. The Cardinal Health™ Level 3 surgical masks are single use, disposable devices provided non-sterile.

    Device Description

    The Cardinal Health™ Level 3 Surgical Mask with Anti-Fog Foam Strip is a four-layer mask constructed of 1 layer of nonwoven polyester/polyethylene blend (inner facing), 1 layer of nonwoven polyolefin melt blown (filter media), 1 layer of nonwoven polyester/polyethylene blend (middle layer), and 1 layer polyolefin spunbond material (outer facing). The four layers of the mask body are collated and sonically welded around the edges to enclose the filter media. The mask is provided with nonwoven polyolefin ties attached by ultrasonic welding. A malleable nosepiece is placed within the binding for comfort and individualized fit around the wearer's nose. The Anti-Fog Foam Strip is intended to minimize fogging.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study conducted for the Cardinal Health™ Level 3 Surgical Mask with Anti-Fog Foam Strip (K192374).

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestPurposeAcceptance Criteria per ASTM F2100-11 Level 3 (AQL = 4.0%)Reported Device Performance (Test Results)Average Performance
    ASTM F1862 Synthetic BloodDetermine synthetic blood penetration resistance160 mmHgPass (31/32)N/A
    ASTM F2101 BFEDetermine the bacterial filtration efficiency≥ 98%Pass (32/32)99.7%
    ASTM F2299 PFE at 0.1 micronDetermine sub-micron particulate filtration efficiency≥ 98%Pass (32/32)99.2%
    Mil-M-36954C Delta PDetermine breathing resistance or differential pressure< 5.0 mm H2O/cm2Pass (32/32)2.8 mm H2O/cm2
    CPSC 1610 FlammabilityDetermine flammability or flame spreadClass 1Pass (32/32)Did Not Ignite

    2. Sample size used for the test set and the data provenance:

    • Sample Size: For all tests (synthetic blood penetration, BFE, PFE, Delta P, and Flammability), the sample size was 32.
    • Data Provenance: The document does not specify the country of origin or whether the data was retrospective or prospective. It only states that the tests were conducted according to established ASTM and CPSC standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable to a surgical mask. The "ground truth" for a device like a surgical mask is established by objective physical and chemical performance tests against predefined standards, not by expert interpretation or judgment of images or clinical data.

    4. Adjudication method for the test set:

    This information is not applicable. As mentioned above, the "ground truth" is based on objective measurements and predefined acceptance criteria in engineering and material science, not on expert adjudication of ambiguous cases.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable. The device is a surgical mask, not an AI-powered diagnostic or assistive technology that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable. The device is a surgical mask and does not involve any algorithms or AI for standalone performance.

    7. The type of ground truth used:

    The ground truth used for this device is based on objective performance standards and measurements as defined by:

    • ASTM F2100-11, Standard Specification for Performance of Materials Used in Medical Face Masks (which includes ASTM F1862, ASTM F2101, ASTM F2299, and Mil-M-36954C)
    • CPSC 1610 for Flammability

    These standards define the physical and material properties that classify a surgical mask for a particular performance level (Level 3 in this case).

    8. The sample size for the training set:

    This information is not applicable. The device is a physical product (surgical mask) and does not involve AI or machine learning models that require training sets.

    9. How the ground truth for the training set was established:

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182830

    Validate with FDA (Live)

    Date Cleared
    2019-04-26

    (203 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cardinal Health™ Poly-Coated Open-Back Protective Gown is intended to protect health care personnel and patients from the transfer of microorganisms, body fluids, and particulate material. Not intended for use in the operating room.

    The proposed Cardinal Health™ Poly-Coated Open-Back Protective Gown meets the barrier protection requirements of AAMI Level 4 per ANSI/AAMI PB70:2012. Liquid Barrier Performance and Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities, but has an open back which is non-protective.

    The Cardinal Health™ Poly-Coated Open-Back Protective Gowns is a single use, disposable medical device provided non-sterile.

    Device Description

    The Cardinal Health™ Poly-Coated Open-Back Protective Gown is a Class II medical device under the FDA product code of FYC, General & Plastic Surgery Panel, and Regulation 21 CFR 878.4040. The device description of the Cardinal Health™ Poly-Coated Open-Back Protective Gown is in accordance with the Guidance on Premarket Notification [510(k)] Submissions for Surgical Gowns and Surgical Drapes, issued on August 1, 1993 and Guidance for Industry and FDA Staff: Premarket Notification Requirements Concerning Gowns Intended for Use in Healthcare Settings, issued on December 9, 2015.

    The Cardinal Health™ Poly-Coated Open-Back Protective Gown consists of a laminate (blue spunbond polypropylene coated with polyethylene) throughout the body of the gown (including sleeves). The sleeves have a thumbhook cuff formed into the sleeve or have a knit cuff sewn onto sleeve for keeping the sleeves in place on the wearer. The collar has a neck removal feature and the belt ties are integrated into the body. These gowns are non-sterile, single use, disposable gowns that are packaged in pastic bags and/or dispenser boxes. These gowns are offered in blue and come in both Universal and XX-Large sizes.

    AI/ML Overview

    The document describes the non-clinical performance testing of the Cardinal Health™ Poly-Coated Open-Back Protective Gown. It evaluates the device against specific acceptance criteria and details the study methods used.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Element of ComparisonSpecification (Acceptance Criteria)Reported Device Performance (Mean (min/max))
    Basis Weight (ASTM D3776/D3776M-17)39.0 gsm min (average) 42.0 gsm target 45.0 gsm max (average)43.3 (39.9 / 46.1) gsm (Sleeve and Body material)
    Tensile Strength, CD (ASTM D5034-17)5.8 lb min (individual) 3.5 lb min (average)14.2 (12.6 / 16.1) lb
    Tear Strength, MD (ASTM D5587-15)6.7 lb min (individual) 2.8 lb min (average)6.0 (4.4 / 7.5) lb
    Flammability (CPSC, Part 1610)Class 1Class 1
    Hydrostatic Head (AATCC 127:2017)140 cmH20 target 100 cmH20 min (individual)>130 (130 / >130) cmH20 >51 (51 / >51) inH20
    Viral barrier (ASTM F1671-13)PassPass
    Liquid Barrier Performance Classification Properties (ANSI/AAMI PB70:2012 / ASTM F1671)Meets AAMI Level 4 performance requirementsAll areas tested meet Level 4 performance requirements
    BiocompatibilityNon-cytotoxic, non-irritating, non-sensitizing per ISO-10993-1Non-cytotoxic, non-irritating, non-sensitizing and met requirements per ISO-10993-1

    2. Sample size used for the test set and the data provenance

    • Liquid Barrier Performance Classification Properties: Testing was performed using 3 lots and 32 samples per lot in each critical zone area (body or sleeve, and heat sealed seam(s)).
    • Other tests (Basis Weight, Tensile Strength, Tear Strength, Hydrostatic Head): The document provides mean and min/max values, but a specific sample size for these individual tests beyond the "Finished Good Test Results" is not explicitly stated.
    • Data Provenance: The document implies that the testing was conducted internally by the manufacturer, Cardinal Health, or by a contracted lab. The data is presented as "Finished Good Test Results" for the "Subject Device: Cardinal Health™ Poly-Coated Open-Back Protective Gown." This data would be considered prospective as it's generated specifically for the device submission. The country of origin for the data is not specified, but the submission is to the U.S. FDA by a company based in Illinois, USA.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The tests described are non-clinical performance tests that use standardized methodologies (e.g., ASTM, AATCC, ISO standards) to measure physical and barrier properties. These types of tests do not typically involve human expert interpretation or "ground truth" in the same way clinical studies or diagnostic device evaluations would. The "ground truth" is established by the test method itself and the specified performance criteria. Therefore, no human experts were used to establish ground truth for this test set in the context of diagnostic interpretation.

    4. Adjudication method for the test set

    Not applicable. As noted above, these are non-clinical, objective performance tests following standardized methods, not subjective clinical assessments requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a submission for a protective gown, which is a physical medical device, not an AI-powered diagnostic tool. Therefore, an MRMC study related to human readers and AI assistance is not relevant or presented.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm or AI system.

    7. The type of ground truth used

    The ground truth for the non-clinical performance tests is based on predefined objective standards and specifications (e.g., AAMI Level 4 for barrier protection, specific tensile strength values, flammability classes). These standards define what constitutes a "pass" or "fail" for each characteristic.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not a machine learning model, and therefore there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170762

    Validate with FDA (Live)

    Date Cleared
    2017-06-09

    (88 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cardinal Health™ Non-Reinforced Surgical Gown is intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate material. In addition, this surgical gown meets the requirements of AAMI Level 3 barrier protection for a surgical gown per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities (AAMI PB70). The Cardinal Health™ Non-Reinforced Surgical Gowns are single use, disposable medical devices; provided sterile and non-sterile.

    Device Description

    The Cardinal Health™ Non-Reinforced Surgical Gown is a Class II medical device under the FDA product code of FYA, General & Plastic Surgery Panel, and Regulation 21 CFR 878.4040. The device description of the Cardinal Health™ Non-Reinforced Surgical Gown is in accordance with the Guidance on Premarket Notification [510(k)] Submissions for Surgical Gowns and Surgical Drapes, issued on August 1, 1993.

    The chest front and sleeve critical zones of the Cardinal Health™ Non-Reinforced Surgical Gowns are constructed from a blue polyolefin SMS (spunbond, meltblown, spunbond) and have been tested according to AAMI PB70:2012 and meet AAMI Level 3 barrier level protection for a surgical gown. The Cardinal Health™ Non-Reinforced Surgical Gown is a single use, disposable medical device that will be provided in a variety of sterile and nonsterile packaging configurations.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for a Cardinal Health™ Non-Reinforced Surgical Gown (K170762). This document focuses on demonstrating the substantial equivalence of the new device to a predicate device (Convertors® SMS Polyolefin Standard Gown (K020593)) through performance testing against established standards for surgical gowns. This is not a study of a device employing AI or machine learning.

    Therefore, many of the requested fields are not applicable (N/A) to this type of medical device submission. However, I can extract information related to acceptance criteria and performance testing:


    1. Table of Acceptance Criteria and Reported Device Performance

    Test/Element of ComparisonPredicate Device (K020593) Test ResultsProposed Device (Cardinal Health™ Non-Reinforced Surgical Gown) Test Results (Mean)Specification/Acceptance Criteria (for Proposed Device)
    Physical Properties
    Basis weight (oz/yd2) - ASTM D3776N/A (older standard used)1.32 oz/yd2Target Mean= 1.30; Mean Min = 1.20; Ind Min = 1.07
    Grab tensile, MD (lbs) - ASTM D5034N/A (older standard used)21.57 lbsN/A (no specific specification provided for proposed mean)
    Grab tensile, CD (lbs) - ASTM D5034N/A (older standard used)13.60 lbsTarget Mean = 15.00; Mean Min = 11.00; Ind Min = 9.50
    Trap Tear, MD (lbs) - ASTM D5587-153.2-7.4 lbs3.47 lbsTarget Mean = 4.00; Mean Min = 2.40; Ind Min = 2.00
    Trap Tear, CD (lbs) - ASTM D5587-157.0-11.8 lbs5.63 lbsN/A (no specific specification provided for proposed mean)
    Flammability (sec) - CPSC, Part 1610Class IClass IClass I
    Alcohol Repellency (rating) - NWSP>69Target Mean = 7; Mean Min = 6; Ind Min = 5
    Hydrostatic Head (cm) - AATCC-12765-92 cm89.90 cmTarget Mean = 70; Mean Min = 60; Ind Min = 53
    Water Impact (g) - AATCC-420.0-0.10 g0.10 gTarget Mean = 0.10; Mean Max = 0.30; Ind Max = 1.0
    Liquid Barrier Performance
    AAMI Level 3 barrier protectionMet equivalent AAMI Level 3Meets AAMI Level 3 barrier protection requirementsAAMI Level 3 barrier protection
    Biocompatibility
    Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity (ISO 10993-1)Met requirements and acceptableNon-cytotoxic, non-irritating, and non-sensitizingNon-cytotoxic, non-irritating, and non-sensitizing

    2. Sample size used for the test set and the data provenance
    The document states that for Liquid Barrier Performance Classification Properties, "Testing was performed using 3 lots and 32 samples per lot in each critical zone area." The critical zones tested were the body and sleeve (same fabric), the sleeve seam, and front belt attachment.
    For other physical properties (e.g., basis weight, grab tensile), test results are reported as "Test Results (3 Lots)," indicating multiple samples were taken from three different production lots. The data provenance is not explicitly stated, but it can be inferred that these are prospective tests conducted by the manufacturer (Cardinal Health) on their proposed device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    N/A. This device submission does not involve clinical data that would require expert-established ground truth. The "ground truth" for this device is based on standardized physical and barrier performance tests.

    4. Adjudication method for the test set
    N/A. This device submission does not involve clinical data or interpretations that require adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    N/A. This is a surgical gown, not an AI/ML-enabled diagnostic or interpretive device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    N/A. This is a surgical gown, not an AI/ML-enabled device.

    7. The type of ground truth used
    The "ground truth" for this device is based on objective, standardized laboratory test methods and specifications as defined by organizations like ASTM, AATCC, NWSP, CPSC, and ANSI/AAMI PB70:2012. For biological safety, it's based on the requirements of ISO 10993-1.

    8. The sample size for the training set
    N/A. This is a surgical gown, not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established
    N/A. This is a surgical gown, not an AI/ML device that requires a training set with ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 3