Search Results
Found 101 results
510(k) Data Aggregation
(254 days)
Ask a specific question about this device
(245 days)
Ask a specific question about this device
(71 days)
Ask a specific question about this device
(256 days)
• BD Vacutainer® Plasma Separator Tubes (PST™)
The BD Vacutainer® Plasma Separator Tubes (PST™) are evacuated, sterile, single-use, in vitro diagnostic medical devices. They are intended to be used by healthcare professionals for the collection, containment, and transport of human venous blood specimens, and the subsequent separation and storage of plasma for in vitro diagnostic testing.
BD Vacutainer® Plasma Separator Tubes (PST™) are used for testing chemistry analytes in plasma samples.
• BD Vacutainer® Sodium Heparin Blood Collection Tubes
The BD Vacutainer® Sodium Heparin Blood Collection Tube is a sterile, single-use, in vitro diagnostic medical device for the collection, containment, transport of human venous blood specimens, and the subsequent separation of plasma by centrifugation, for in vitro diagnostic testing. It is used in settings where a venous blood specimen is collected by a healthcare professional.
The BD Vacutainer® Sodium Heparin Blood Collection Tube is used for testing lithium.
BD Vacutainer® Plasma Separator Tubes (PST™) and BD Vacutainer® Sodium Heparin Blood Collection Tubes are sterile Blood Collection Tubes which use a controlled vacuum to pull a specific volume of blood into the tube. Each BD Vacutainer® Plasma Separator Tube (PST™) and BD Vacutainer® Sodium Heparin Blood Collection Tube consists of 1) a plastic tube manufactured from PET (polyethylene terephthalate), 2) a BD Hemogard™ cap assembly or Conventional rubber stopper with lubricant, and 3) a spray-dried anti-coagulated formulated from lithium or sodium heparin. The BD Vacutainer® Plasma Separator Tubes (PST™) additionally include an inert polymer separator gel.
N/A
Ask a specific question about this device
(75 days)
BD Surgiphor™ Antimicrobial Irrigation System is intended to mechanically loosen and remove debris, and foreign materials, including microorganisms, from wounds.
The subject BD Surgiphor™ Antimicrobial Irrigation System is a terminally sterilized 450 mL aqueous solution for irrigation and debridement of wounds. The device includes one screw cap and one bottle of Surgiphor solution (0.5% povidone-iodine (PVP-I)), which is used to loosen and remove wound debris via manual lavage. The mechanical action of fluid moving across the wound provides the mechanism of action and aids in the loosening and removal of debris, and foreign materials, including microorganisms, from wounds. The PVP-I in the Surgiphor solution serves as a preservative to ensure that no microbial growth occurs in the solution after the bottle is open.
N/A
Ask a specific question about this device
(253 days)
The 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SP Syringe is intended to be used only for the flushing of indwelling vascular access devices.
BD PosiFlush™ SP Syringe is a three-piece, single use syringe with a Luer connector that is compatible with ISO 80369-7:2021 small-bore connectors for liquids and gases in healthcare applications- Part 7: Connectors for intravascular or hypodermic applications. The syringe is prefilled with 0.9% sodium chloride injection, USP and sealed with a tip cap. PosiFlush SP Syringe is individually packaged inside a flow wrap. The moist heat sterilization process renders the syringe a Sterility Assurance Level of 10⁻⁶ for the fluid path and saline solution.
N/A
Ask a specific question about this device
(266 days)
BD Enteric Bacterial Panel for BD COR™ System
BD Enteric Bacterial Panel for BD COR™ System is an automated in vitro diagnostic test for the direct qualitative detection and differentiation of enteric bacterial pathogens. BD Enteric Bacterial Panel for BD COR™ System detects nucleic acids from:
- Salmonella spp.
- Campylobacter spp. (C. jejuni and C. coli)
- Shigella spp. / Enteroinvasive Escherichia coli (EIEC)
- Shiga toxin 1 (stx1) / Shiga toxin 2 (stx2) genes (found in Shiga toxin-producing Escherichia coli [STEC]) as well as Shigella dysenteriae, which can possess a Shiga toxin gene (stx) that is identical to the stx1 gene of STEC.
Testing is performed utilizing the BD Fecal Collection and Transport Kit. Unpreserved soft to diarrheal stool specimens from symptomatic patients with suspected acute gastroenteritis, enteritis or colitis are probed using the flocked swab and material is transferred to the BD Fecal Collection and Transport tube containing modified Cary-Blair medium. The test is performed utilizing real-time polymerase chain reaction (PCR) for the amplification of specific sequences of target DNA. The test utilizes fluorogenic sequence-specific hybridization probes for detection of the amplified DNA.
This test is intended for use, in conjunction with clinical presentation, laboratory findings, and epidemiological information, as an aid in the differential diagnosis of Salmonella, Shigella/EIEC, Campylobacter, and Shiga toxin-producing E. coli (STEC). Results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out co-infection with other organisms that are not detected by this test and may not be the sole or definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
BD Enteric Bacterial Panel plus for BD COR™ System
BD Enteric Bacterial Panel plus for BD COR™ System is an automated in vitro diagnostic test for the direct qualitative detection and differentiation of enteric bacterial pathogens. BD Enteric Bacterial Panel plus for BD COR™ System detects nucleic acids from:
- Salmonella spp.
- Campylobacter spp. (C. jejuni and C. coli)
- Shigella spp. / Enteroinvasive Escherichia coli (EIEC)
- Shiga toxin 1 (stx1) / Shiga toxin 2 (stx2) genes (found in Shiga toxin-producing Escherichia coli [STEC]) as well as Shigella dysenteriae, which can possess a Shiga toxin gene (stx) that is identical to the stx1 gene of STEC.
- Plesiomonas shigelloides
- Vibrio spp. (V. vulnificus, V. parahaemolyticus, and V. cholerae)
- Enterotoxigenic Escherichia coli (ETEC) heat-labile enterotoxin (LT) / heat-stable enterotoxin (ST) genes
- Yersinia enterocolitica
Testing is performed utilizing the BD Fecal Collection and Transport Kit. Unpreserved soft to diarrheal stool specimens from symptomatic patients with suspected acute gastroenteritis, enteritis or colitis are probed using the flocked swab and material is transferred to the BD Fecal Collection and Transport tube containing modified Cary-Blair medium. The test is performed utilizing real-time polymerase chain reaction (PCR) for the amplification of specific sequences of target DNA. The test utilizes fluorogenic sequence-specific hybridization probes for detection of the amplified DNA.
This test is intended for use, in conjunction with clinical presentation, laboratory findings, and epidemiological information, as an aid in the differential diagnosis of Salmonella, Shigella/EIEC, Campylobacter, Shiga toxin-producing E. coli (STEC), Plesiomonas shigelloides, Vibrio spp. (V. vulnificus, V. parahaemolyticus, and V. cholerae), Enterotoxigenic Escherichia coli (ETEC) LT/ST, and Yersinia enterocolitica. Results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out co-infection with other organisms that are not detected by this test and may not be the sole or definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
Enteric Bacterial Panel Diluent for BD COR™ System
The Enteric Bacterial Panel Diluent for BD COR™ System is intended to be used in clinical settings according to instructions provided for aliquoting into Molecular Aliquot Tubes by the BD COR™ System. The Enteric Bacterial Panel Diluent for BD COR™ System is only for use with BD Fecal Collection and Transport Kit specimens tested on BD COR™ Systems.
The BD Enteric Bacterial Panel for BD COR™ System (BD EBP for BD COR™ System) simultaneously detects pathogens responsible for gastroenteritis due to Salmonella spp., Campylobacter spp. (C. jejuni and C. coli), Shigella spp./EIEC, stx/stx1/stx2 found in Shiga toxin-producing E. coli and in Shigella dysenteriae, and an internal Sample Processing Control. The BD Enteric Bacterial Panel plus for BD COR™ Systems (BD EBP plus for BD COR™ System) additionally detects Plesiomonas shigelloides, Vibrio (V. vulnificus, V. parahaemolyticus, and V. cholerae), Enterotoxigenic E. coli (ETEC) LT/ST, Yersinia enterocolitica and an internal Sample Processing Control with a second master mix. The assays automate the testing process and minimize operator intervention from the time the sample is placed onto the BD COR™ System until results are available.
The BD COR™ System is comprised of a single BD COR™ PX System attached to a BD COR™ MX Analyzer as described in K210585 and K224653. Once the specimens are loaded, the BD COR™ PX System performs the necessary pre-analytical steps such as vortexing, aliquoting into a diluent filled molecular aliquot tube, sorting/grouping of the secondary samples for testing by assay, pre-warming and cooling of the sample (where required), and transport of the sample into the BD COR™ MX Analyzer, where extraction, amplification and detection take place.
Additionally, the steps of ordering tests on the instrument for specific samples will be managed directly by the user interaction with the Laboratory Information System (LIS), which communicates directly with the instrument.
Once the clinical specimens are received in the laboratory and loaded into the transport racks, the user will not be required to directly handle the specimen again prior to result reporting and removal from the system.
N/A
Ask a specific question about this device
(245 days)
The BD PhaSeal™ Optima system is an airtight and leakproof closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The BD PhaSeal™ Optima system also prevents microbial ingress for up to 168 hours.
BD PhaSeal™ Optima Closed System Transfer Devices (CSTD) are sterile, single use closed system drug transfer devices intended for the reconstitution and transfer of antineoplastic or other hazardous drugs in the healthcare setting. The BD PhaSeal™ Optima system is comprised of four devices—Protector, Injector, Connector, and Infusion Adapter.
The closed transfer of liquid drugs takes place through a double membrane utilizing self-sealing elastomeric membranes that are tightly fitted together through the collet-style fitting on each of the BD PhaSeal™ Optima system devices. During use, the single lumen cannula of the Injector perforates the double membranes for the transfer of liquids. When the cannula is retracted, the membranes seal to prevent the transfer of environmental contaminants into the system and/or escape of drug or vapor concentrations outside the system, thereby minimizing the individual and environmental exposure to drug vapor, aerosols, leaks and spills. The BD PhaSeal™ Optima system prevents microbial ingress for up to 168 hours. Performance of the self-sealing membrane has been substantiated for up to 10 penetrations.
The BD PhaSeal™ Optima Connecting Set (C83-O) is a sterile, single patient use sterile device that enables the administration of non-hazardous and hazardous parenteral drugs when used with the devices that have the compatible mating component - the BD PhaSeal™ Optima Spike Set and/or Connector. The BD PhaSeal™ Optima Connecting Set will be utilized by healthcare workers who administer parenteral hazardous drugs.
The BD PhaSeal™ Optima Spike Set (C180-O) is a sterile, single patient use sterile bag access device that enables the preparation and administration of non-hazardous and hazardous parenteral drugs when used with devices that have the compatible mating component - the BD PhaSeal™ Optima Injector and/or Connecting Set. The BD PhaSeal™ Optima Spike Set will be utilized by healthcare workers who prepare and administer parenteral hazardous drugs.
N/A
Ask a specific question about this device
(66 days)
The BD Phoenix Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacterales and Non-Enterobacterales and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.
This premarket notification is for the BD Phoenix Automated Microbiology System with Eravacycline at a concentration of 0.125-2 µg/mL. Testing is indicated for Enterobacterales as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC).
The BD Phoenix Automated Microbiology System - GN Eravacycline (0.125-2 µg/mL) has demonstrated acceptable performance with the following organisms:
Enterobacterales (Citrobacter amalonaticus, Citrobacter braakii, Citrobacter farmeri, Citrobacter freundii, Citrobacter koseri, Citrobacter youngae, Enterobacter cloacae, Escherichia coli, Klebsiella aerogenes, Klebsiella oxytoca, and Klebsiella pneumoniae)
This submission is for addition of Eravacycline (0.125-2 µg/mL) to the BD Phoenix™ ID/AST or AST only panels. The ID portion of the ID/AST combination panel was not subject to review in this submission.
The Phoenix AST method is a broth-based microdilution test. The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. The ID/AST combination panel includes an ID side (51 wells) with dried substrates for bacterial identification and an AST side (85 wells). The AST panel contains a wide range of two-fold doubling dilution concentrations of antimicrobial agents and growth and fluorescent controls at appropriate well locations. The AST panel does not include wells for isolate identification.
The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. The organism to be tested must be a pure culture and be preliminarily identified as Gram-positive or Gram-negative. Colonies are then suspended in ID broth and equated to a 0.5 McFarland suspension using a nephelometer device. A further dilution is made into AST broth (a cation-adjusted formulation of Mueller-Hinton broth containing 0.010% Tween 80), to which the redox-buffered oxidation-reduction AST indicator solution is added producing a blue color in the wells. The concentration of organisms in the final AST broth suspension is approximately 5 X 10⁵ CFU/mL.
The Phoenix AST Broth is poured into the inoculation port of the AST panel and the inoculum flows into the panel, filling panel wells. Polyethylene caps are applied to seal the inoculation ports. An air admittance port is located in the panel lid to ensure adequate oxygen tension in the panel for the duration of the test. Inoculated panels are barcode scanned and loaded into the BD Phoenix Automated Microbiology System instrument where panels are continuously incubated at 35 °C ± 1 °C.
Continuous measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. The instrument takes readings every 20 minutes. Organisms growing in the presence of a given antimicrobial agent reduce the indicator (changing it to a pink color). This signals organism growth and resistance to that antimicrobial agent. Organisms killed or inhibited by the antimicrobial agent do not cause reduction of the indicator and therefore do not produce a color change. The Phoenix instrument reads and records the results of the antimicrobial tests contained in the panel and interprets the reactions (based on the organism identification) to give a minimal inhibitory concentration (MIC) value and category interpretations (susceptible, intermediate, resistant, or not susceptible). AST results are available within 16 hours.
This is an auto read result; no manual readings are possible with this system.
Additional comments concerning specific organism/antimicrobial combinations are provided from the software-driven expert system (BDXpert), using rules derived from CLSI documentation and/or the FDA-approved drug labeling.
Here's a breakdown of the acceptance criteria and the study details for the BD Phoenix Automated Microbiology System - GN Eravacycline (0.125-2 µg/mL), based on the provided FDA 510(k) clearance letter:
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Metric | Acceptance Criteria (FDA Limits) | Reported Device Performance (Combined Clinical and Challenge Isolates - Manual Inoculation) | Notes |
|---|---|---|---|
| Reproducibility | > 95% (± 1 dilution) agreement | Manual PhoenixSpec™ Nephelometer: 100% (324/324) | Achieved |
| Phoenix™ AP Instrument: 100% (324/324) | Achieved | ||
| Overall Essential Agreement (EA) | > 90% | 97.8% (850/869) | Achieved |
| Evaluable Essential Agreement (EA) | > 90% | 95.3% (384/403) | Achieved |
| Overall Category Agreement (CA) | > 90% | 97.1% (844/869) | Achieved |
| Adjusted Major Error Rate (Maj) | ≤ 3% | 0% (0/798) | Achieved (All 4 original major errors were within essential agreement) |
| Adjusted Very Major Error Rate (Vmj) | ≤ 1.5% | 15.5% (11/71) | Not Achieved - This is where the device had specific limitations and required additional labeling for confirmatory testing. |
| Adjusted Major Error Rate (Maj) - Challenge Isolates (Manual) | ≤ 3% | 0% (0/54) | Achieved (All 2 original major errors were within essential agreement) |
| Adjusted Very Major Error Rate (Vmj) - Challenge Isolates (Manual) | ≤ 1.5% | 3.3% (1/30) | Not Achieved - Addressed with limitations in the product insert. |
| Adjusted Major Error Rate (Maj) - Challenge Isolates (Phoenix AP) | ≤ 3% | 0% (0/58) | Achieved (All 5 original major errors were within essential agreement) |
| Adjusted Very Major Error Rate (Vmj) - Challenge Isolates (Phoenix AP) | ≤ 1.5% | 0% (0/22) | Achieved (The 1 original very major error was within essential agreement) |
| Trending (MIC Values) | Difference between % higher vs. lower readings ≤ 30% or not statistically significant | Observed trending toward lower MIC values for: * Citrobacter freundii (-45%) * Citrobacter koseri (-71%) * Escherichia coli (-74%) | Not Achieved (for certain organisms) - Addressed with specific footnotes in the performance table of the device labeling. |
| Growth Failure Rate | Not explicitly stated in acceptance criteria, but 0% is good. | 0% | Achieved |
| Quality Control (QC) | > 95% of tests performed in acceptable range | Acceptable for greater than 95% of tests performed using both inoculation methods. | Achieved |
Note on VMEs: A very major error (VME) occurs when a resistant isolate is categorized as susceptible by the device. This is a critical error as it can lead to inappropriate treatment. The FDA's acceptable rate for adjusted VMEs is typically ≤ 1.5%. The device exceeded this threshold for combined clinical and challenge isolates with manual inoculation, and for challenge isolates with manual inoculation, which required specific cautionary statements and recommendations for confirmatory testing in the product labeling.
2. Sample Size Used for the Test Set and Data Provenance
- Clinical Isolates: 785 isolates (626 fresh, 159 stock)
- Provenance: "three U.S. sites" (presumably clinical laboratories in the US). The exact country of origin for the isolates themselves is not specified beyond "U.S. sites."
- Retrospective/Prospective: Not explicitly stated, but "fresh" and "stock" isolates suggest a mix, likely collected over time (retrospective component for stock, and potentially prospective for fresh if collected specifically for the study, or recent retrospective).
- Challenge Isolates: 84 isolates (stock isolates with known resistance mechanisms).
- Provenance: "Additional stock challenge isolates were tested at each study site." (three U.S. sites, as mentioned for clinical testing).
- Retrospective/Prospective: Retrospective (stock isolates).
- Reproducibility Isolates: 12 on-scale isolates.
- Quality Control Isolates: E. coli ATCC 25922 and P. aeruginosa ATCC 27853 (standard ATCC strains).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The ground truth for the test set (both clinical and challenge isolates) was established using the CLSI frozen broth microdilution reference panel, prepared according to CLSI M07 guidelines. This is a standardized laboratory method, not reliant on individual human experts in the same way, for example, a radiology image interpretation study would be. Therefore, the concept of "number of experts" and their "qualifications" doesn't directly apply in this context. The "expert" in this case is the CLSI standard method itself, which is developed by committees of microbiology experts.
4. Adjudication Method for the Test Set
Not applicable in the traditional sense of multiple human readers or a consensus process. The reference method (CLSI frozen broth microdilution) serves as the "gold standard" or ground truth. Discrepancies between the device and the reference method were analyzed for Essential Agreement (EA) and Category Agreement (CA). Further analysis and "adjustments" for major and very major errors were based on whether the MIC values of the errors fell within essential agreement (i.e., within one doubling dilution of the reference method).
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically used for diagnostic devices that involve human interpretation (e.g., radiologists reading images) to compare the performance of human readers with and without AI assistance. The BD Phoenix system is an automated platform for antimicrobial susceptibility testing, where human interpretation of results is minimal once the system produces MIC values and interpretations.
6. Standalone Performance Study
Yes, a standalone performance study was done. The entire study described, which compares the BD Phoenix Automated Microbiology System's results directly against the CLSI frozen broth microdilution reference method, represents the standalone performance of the algorithm/device without human-in-the-loop performance influencing the primary results. The system automatically reads and interprets the results.
7. Type of Ground Truth Used
The type of ground truth used was a reference method, specifically the CLSI frozen broth microdilution reference panel prepared according to CLSI M07 guidelines. This is considered the "gold standard" for antimicrobial susceptibility testing.
8. Sample Size for the Training Set
The document does not explicitly mention a separate training set or its sample size. This is common for predicate-based 510(k) submissions, where the focus is on demonstrating substantial equivalence to a legally marketed predicate device, and the "training" of the internal device algorithms might have occurred during its initial development or earlier predicate clearances. The performance data presented here are for validation and comparison against the reference method.
9. How the Ground Truth for the Training Set Was Established
As no specific training set is mentioned, the method for establishing its ground truth is also not described. If the device uses machine learning, its initial development and training would typically involve large datasets with ground truth established by expert-reviewed reference methods. However, this clearance focuses on the validation of a specific addition (Eravacycline) to an already established system.
Ask a specific question about this device
(90 days)
The BD Plastipak™ Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection.
The BD Plastipak™ Syringe is a three-piece, single use, sterile, hypodermic syringe with a 6% (Luer) male connector in 20 mL and 50 mL eccentric luer slip tip configurations. The BD Plastipak™ Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection. The syringe assembly consists of a lubricated polypropylene barrel with a graduated scale, a lubricated synthetic rubber stopper and a polypropylene plunger rod. The plunger rod is pulled back to aspirate fluids or depressed to inject or expel fluids. The syringe barrel incorporates a male 6% (Luer) connector which is connectable to a compatible female 6% (Luer) connector. The BD Plastipak™ Syringe is provided sterile by Ethylene Oxide Gas (ETO) sterilization method.
The provided text is a 510(k) Clearance Letter for a medical device (BD Plastipak™ Syringe). It details the device's characteristics, intended use, and comparison to a predicate device. However, it does not describe an AI/ML-driven medical device or a study involving human readers or expert consensus for ground truth establishment.
The document discusses bench performance testing and biocompatibility tests for a physical device (syringe), not a software or AI-based diagnostic tool. Therefore, many of the requested criteria (like sample size for test/training set, number of experts, adjudication method, MRMC study, standalone performance, ground truth types) are not applicable to this specific submission.
Despite the irrelevance of some questions to the provided document, I will structure the answer based on the questions asked, indicating "Not Applicable" or providing the information that is present in the document.
Here's an analysis of the provided 510(k) clearance letter in the context of the requested information about acceptance criteria and study data:
This 510(k) clearance letter pertains to a physical medical device, the BD Plastipak™ Syringe, not an AI/ML-driven diagnostic or image analysis tool. As such, many of the typical acceptance criteria and study methodologies applicable to AI models (e.g., ground truth established by experts, MRMC studies, training/test set sizes for algorithms, human reader improvement with AI assistance) are not relevant or described in this document.
The "study" referenced in the document primarily consists of non-clinical performance and biocompatibility testing to demonstrate the substantial equivalence of the new syringe (with a changed barrel resin) to a previously cleared predicate syringe.
1. Table of Acceptance Criteria and Reported Device Performance
The document states, "The subject device met all the predetermined acceptance criteria for the above listed performance and biocompatibility tests." However, it does not explicitly list the quantitative acceptance criteria or the specific numerical performance results for each test. It only lists the tests performed and the standards they adhere to.
| Test Category | Test | Purpose (as described) | Acceptance Criteria (Not Explicitly Stated Quantitatively) | Reported Performance (Not Explicitly Stated Quantitatively) |
|---|---|---|---|---|
| Performance Testing | Breakout Force | ISO 7886-1:2017- Performance evaluation of force to operate the piston. | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) |
| Sustaining Force | ISO 7886-1:2017- Evaluation of force to operate the piston. | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
| Leakage Past Stopper | ISO 7886-1:2017- Evaluation of Freedom from air and liquid leakage past plunger stopper. | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
| Volumetric Accuracy | ISO 7886-1:2017- Evaluation of Volumetric Accuracy | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
| Dead Space | ISO 7886-1:2017- Evaluation of residual volume | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
| Luer Leakage | ISO 80369-7:2021 - Evaluation of the luer fittings for leakage | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
| Stress Cracking | ISO 80369-7:2021- Evaluation of the luer fittings for stress cracking | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
| Resistance to separation from axial load | ISO 80369-7:2021- Evaluation of the luer fitting for separation when subjected to axial force. | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
| Biocompatibility Testing | Cytotoxicity | ISO 10993-5:2009- Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) |
| Sensitization | ISO 10993-10:2021- Biological Evaluation of Medical Devices - Part 10: Tests for Skin Sensitization | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
| Intracutaneous Reactivity | ISO 10993-23:2021- Biological evaluation of medical devices- Test for Irritation | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
| Acute Systemic Toxicity | ISO 10993-11:2017- Biological evaluation of medical devices — Part 11: Tests for systemic toxicity | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
| Material Mediated Pyrogenicity | ISO 10993-11:2017- Biological evaluation of medical devices — Part 11: Tests for systemic toxicity USP43-NF38 <151>Pyrogen Test (USP Rabbit Test) | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
| Hemocompatibility | ISO 10993-4:2017- Biological evaluation of medical devices Part 4: Selection of tests for interactions with blood ASTM F756-17- Standard Practice for Hemolytic Properties of Materials | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not specified in the provided document. The tests are "bench performance testing" on various syringe units.
- Data Provenance: Not specified, but generally, bench testing for physical devices is conducted in a controlled lab environment by the manufacturer. It is non-clinical.
- Retrospective or Prospective: Not applicable for this type of physical device testing.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Not Applicable. Ground truth for this physical device testing is established through standardized laboratory test methods and measurements against international or internal specifications, not by human experts interpreting clinical data.
4. Adjudication Method for the Test Set
- Not Applicable. (See point 3)
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This is not an AI/ML device. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was conducted or is relevant.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
- No. This is not an AI/ML device. Therefore, no standalone algorithm performance was assessed. The performance tests are for the physical syringe itself.
7. The Type of Ground Truth Used
- The "ground truth" for this device's performance is based on measurements against established engineering specifications and international standards (e.g., ISO, ASTM, USP) for physical and material properties (e.g., force, leakage, volume accuracy, biocompatibility reactions). It is not based on expert consensus, pathology, or outcomes data in the clinical sense.
8. The Sample Size for the Training Set
- Not Applicable. This is not an AI/ML device. There is no concept of a "training set" in the context of the reported non-clinical bench testing for a physical syringe.
9. How the Ground Truth for the Training Set was Established
- Not Applicable. (See point 8)
Ask a specific question about this device
Page 1 of 11