K Number
K250884

Validate with FDA (Live)

Date Cleared
2025-12-02

(253 days)

Product Code
Regulation Number
880.5200
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SP Syringe is intended to be used only for the flushing of indwelling vascular access devices.

Device Description

BD PosiFlush™ SP Syringe is a three-piece, single use syringe with a Luer connector that is compatible with ISO 80369-7:2021 small-bore connectors for liquids and gases in healthcare applications- Part 7: Connectors for intravascular or hypodermic applications. The syringe is prefilled with 0.9% sodium chloride injection, USP and sealed with a tip cap. PosiFlush SP Syringe is individually packaged inside a flow wrap. The moist heat sterilization process renders the syringe a Sterility Assurance Level of 10⁻⁶ for the fluid path and saline solution.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - K250884

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.02

December 2, 2025

Becton, Dickinson and Company
Anamika Tiwari
Staff Regulatory Affairs Specialist
1 Becton Drive
Franklin Lakes, New Jersey 07417

Re: K250884
Trade/Device Name: 0.9 % Sodium Chloride Injection, USP, BD PosiFlush™ SP Syringe
Regulation Number: 21 CFR 880.5200
Regulation Name: Intravascular Catheter
Regulatory Class: Class II
Product Code: NGT
Dated: November 4, 2025
Received: November 5, 2025

Dear Anamika Tiwari:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K250884 - Anamika Tiwari Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250884 - Anamika Tiwari Page 3

Sincerely,

ALLAN GUAN -S

For Bifeng Qian, M.D., Ph.D.
Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K250884

Device Name: 0.9 % Sodium Chloride Injection, USP, BD PosiFlush™ SP Syringe

Indications for Use (Describe):
The 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SP Syringe is intended to be used only for the flushing of indwelling vascular access devices.

Type of Use (Select one or both, as applicable):
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number"

FORM FDA 3881 (8/23) Page 1 of 1

Page 5

0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SP Saline Flush Syringe 510(k) Summary (21 CFR §807.92)

K250884

Submitter Information
Submitter Name:Becton, Dickinson, and Company
Submitter Address:1 Becton Drive, Franklin Lakes, NJ 07417
Contact Person:Anamika Awadhesh TiwariStaff Regulatory Affairs Specialist
Email Address:Anamika.tiwari@bd.com
Phone Number:201.847.5005
Fax Number:201.847.5307
Date of Preparation:December 1st, 2025
Subject Device
Trade Name:0.9% Sodium Chloride Injection, USP, BD PosiFlush SP Syringe
Model Number:306546
Common Name:0.9% Sodium Chloride Injection Flush Syringe
Regulation Number:21 CFR 880.5200
Classification Name:Saline, Vascular Access Flush
Regulatory Class:Class II
Product Code:NGT
Classification Panel:General Hospital
Predicate Device
510(k) Number:K161552
Trade Name:0.9 % Sodium Chloride Injection, USP, BD PosiFlush SP Syringe
Common Name:0.9% Sodium Chloride Injection Flush Syringe
Regulation Number:21 CFR 880.5200
Classification Name:Saline, Vascular Access Flush
Regulatory Class:Class II
Product Code:NGT
Classification Panel:General Hospital

Page 6

Page 2 of 10

Device DescriptionBD PosiFlush™ SP Syringe is a three-piece, single use syringe with a Luer connector that is compatible with ISO 80369-7:2021 small-bore connectors for liquids and gases in healthcare applications- Part 7: Connectors for intravascular or hypodermic applications. The syringe is prefilled with 0.9% sodium chloride injection, USP and sealed with a tip cap. PosiFlush SP Syringe is individually packaged inside a flow wrap. The moist heat sterilization process renders the syringe a Sterility Assurance Level of 10⁻⁶ for the fluid path and saline solution.
Indications for UseThe 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SP Syringe is intended to be used only for the flushing of indwelling vascular access devices.
Technological CharacteristicsThe following table provides a comparison between the subject and predicate devices.

Page 7

Page 3 of 10

AttributeSubject Device(BD PosiFlush™ SP Syringe)K250884Predicate Device(BD PosiFlush™ SP Syringe (K161552)Comparison
Intended Use/ Indications for UseThe 0.9% Sodium Chloride Injection, USP, BD PosiFlush SP Syringe is intended to be used only for the flushing of indwelling vascular access devices.The 0.9% Sodium Chloride Injection, USP, BD PosiFlush SP Syringe is intended to be used only for the flushing of indwelling vascular access devices. Catalog Number 306547 10 mL BD PosiFlush SP Syringes are generally compatible for use with syringe pumps.Similar(306547 is not in scope of this 510k)
Operating PrincipleThe BD PosiFlush SP Syringe is a three- piece, sterile, single use syringe with a 6% (Luer) connector pre- filled with 0.9% Sodium Chloride Injection, USP, and sealed with a tip cap.The BD PosiFlush SP Syringe is a three-piece, sterile, single use syringe with a 6% (Luer) connector pre- filled with 0.9% Sodium Chloride Injection, USP, and sealed with a tip cap.Identical
Syringe Configuration10 mL3, 5, and 10mLDifferent
Single Use?YesYesIdentical
Sterile?Sterile Fluid PathSterile Fluid PathIdentical
Use in Sterile Field?NoNoIdentical
Content of syringe packageOne pre-filled syringe per flow wrap packOne pre-filled syringe per flow wrap packIdentical
Device Components• Barrel• 0.9% NaCl USP solution• Plunger Rod• Tip Cap• Stopper• Barrel• 0.9% NaCl USP solution• Plunger Rod• Tip Cap• StopperIdentical

Page 8

Page 4 of 10

AttributeSubject Device(BD PosiFlush™ SP Syringe)K250884Predicate Device(BD PosiFlush™ SP Syringe (K161552)Comparison
Barrel MaterialPolypropylenePolypropyleneSimilar(Subject device material is assessed as per ISO 10993-1)
Barrel LubricantSiliconeSiliconeIdentical
Plunger Rod MaterialPolypropylenePolypropyleneIdentical
Stopper MaterialStyrene-Butadiene RubberStyrene-Butadiene RubberIdentical
Stopper LubricantSiliconeSiliconeIdentical
Tip Cap MaterialPolypropylenePolypropyleneSimilar(Subject device material is assessed as per ISO 10993-1)
Tip Cap ColorantWhiteWhiteIdentical
Packaging Configuration• Flow Wrap• Shelf Carton• Case Carton• Flow Wrap• Shelf Carton• Case CartonIdentical
Primary Packaging MaterialFlow wrapFlow wrapIdentical
Mode of SterilizationMoist HeatMoist HeatIdentical
SAL10⁻⁶10⁻⁶Identical
Shelf Life1 year3 yearsDifferent
Biocompatibility testing• Cytotoxicity• Sensitization• Intracutaneous Reactivity• Material-mediated Pyrogenicity• Acute Systemic Toxicity• Cytotoxicity• Hemolysis• Acute systematic toxicity• Sub Chronic Toxicity• Intracutaneous Reactivity• Ocular IrritationDifferent

Page 9

Page 5 of 10

AttributeSubject Device(BD PosiFlush™ SP Syringe)K250884Predicate Device(BD PosiFlush™ SP Syringe (K161552)Comparison
Hemocompatibility• Partial Thromboplastin Time• Sensitization• Pyrogenicity• Genotoxicity• Chemical extractable analysis
Performance testing• Tip Cap Removal Torque• Reflux• Tip Cap Leakage• Leakage past Stopper• Break loose force• Breakout Force• Sustaining Force• Retaining Ring Force• Expelled Volume• NaCl Chemistry Testing• Particulate Analysis• Appearance and Solution Clarity• Barrel Transparency• Distribution Test• Container Closure Integrity• Break loose force• Breakout force• Sustaining force• Stopper Separation• Leakage• Pump Force• Distribution Test• Dead Space• Syringe Induced RefluxDifferent

Page 10

Page 6 of 10

DiscussionThe subject device and predicate device are different with respect to the following item:1. The barrel and tip cap resin of both subject and predicate device is different.2. The subject devices configuration includes the 10ml, when compared to the 3ml, 5ml, and 10mL configurations of the predicate device.The different technological characteristics between the subject and predicate device are evaluated in performance testing, biocompatibility tests, and distribution tests demonstrating that there are no new or different questions of safety and effectiveness.
Non-Clinical TestingBD has performed the following performance tests in accordance with 21 CFR §820.30 to demonstrate that the PosiFlush SP Syringe performs equivalent to the predicate devices.The following tests were performed on the subject device to an internal specification or a Standard:

Page 11

Page 7 of 10

TestPurposeAcceptance CriteriaResult
Performance/Design Verification Tests
Tip Cap Removal TorqueMeasure the torque required to remove the tip cap from the syringeTip Cap can be twisted off as per BD validated forcePass
RefluxMeasure the potential blood reflux into the catheter from PosiFlush SP syringeReflux greater than BD validated internal valuePass
Tip Cap LeakageEvaluate the resistance to leakage between the barrel Luer and the tip cap for pre- filled syringesNo evidence of Tip Cap leakagePass
Leakage past StopperEvaluate the resistance to leakage between the barrel and stopper ribsNo leakage of solution past the stopper ribsPass
Break loose forceMeasure the initial maximum force required to move the plunger rod/stopper in the syringe barrelForce to move plunger rod/stopper is less than the BD validated forcePass
Breakout ForceMeasure the maximum force required to move the plunger rod/stopper in the syringe barrel a short time after initially moving the plunger rod/stopperPass
Sustaining ForceMeasure the average force required to move the plunger rod/stopper in the syringe barrelPass
Retaining Ring ForceMeasure the force necessary to remove the plunger rod/stopper assembly from the syringe barrel after assemblyForce to move plunger rod/stopper is greater than the BD validated forcePass
Expelled VolumeMeasure the volume of saline solution expelled from the syringeUSP43-NF38 <697>Container Content for InjectionsPass
NaCl AssayMeasure the NaCl concentration saline solution in syringe samples0.9% NaCl Injection USP MonographPass
pHMeasure the pH of saline solution in syringe samples0.9% NaCl Injection USP Monograph and USP43-NF38 <791> pHPass

Page 12

Page 8 of 10

TestPurposeAcceptance CriteriaResult
IronMeasure the amount of iron in saline solution0.9% NaCl Injection USP Monograph and USP43-NF38 <241> IronPass
Heavy MetalsMeasure the heavy metals such as Cadmium, Arsenic, Cobalt, Vanadium, Copper, Lead, Nickel, Lithium, Antimony and Mercury in saline solutionUSP43-NF38 <232> Elemental Impurities - LimitsPass
NaCl Solution weight lossMeasure the weight loss of saline solution in pre-filled saline syringesUSP43-NF38 <671> Containers Performance TestingPass
UV AnalysisMeasure the UV absorbance between 220-360 nm for pre- filled saline syringes< 0.4 AUPass
Bacterial EndotoxinDetermine the amount of endotoxin in saline pre-filled syringes0.9% NaCl Injection USP Monograph and USP43-NF38 <85> Bacterial Endotoxins TestPass
SterilityTo verify 10⁻⁶ SAL in the fluid path.Pass
Surface SterilityTo verify 10⁻⁶ SAL on all surface of syringe and inside surface of primary packagingSterile; No growth in mediaPass
Particulate AnalysisMeasure the number of particulates in saline pre-filled syringes≥ 10 μm < 3,000≥ 25 μm < 300Pass
Appearance and Solution Clarity/ Barrel TransparencyExamine appearance, solution clarity and barrel transparency in pre-filled saline syringesSolution and components are clearPass
Distribution TestEvaluates the product and packaging integrity after distribution.No unit packaging open seal (end or fin) cuts, tears or holes greater than BD validated internal value.No fluid between specified stopper ribsContainer Closure Integrity maintainedPassPassPass

Page 13

Page 9 of 10

Biocompatibility

TestStandard/MethodAcceptance CriteriaResult
CytotoxicityISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityGrade ≤ 2Pass
SensitizationISO 10993-10:2021 Biological Evaluation of Medical Devices - Part 10: Tests for Skin SensitizationNon-SensitizerPass
Irritation or Intracutaneous ActivityISO 10993-23:2021 Biological Evaluation of Medical Devices - Part 23: Tests for IrritationFinal Test Sample Score ≤ 1Pass
Acute Systemic ToxicityISO 10993-11:2017 Biological evaluation of medical devices — Part 11: Tests for systemic toxicityNo significantly greater biological reaction than the controlPass
Material Mediated PyrogenicityISO 10993-11:2017 Biological evaluation of medical devices — Part 11: Tests for systemic toxicityUSP43-NF38 <151>Pyrogen Test (USP Rabbit Test)No temperature rise ≥ 0.5°CPass
HemocompatabilityISO 10993-4:2017 Biological evaluation of medical devices Part 4: Selection of tests for interactions with bloodASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials≤ 5% hemolysisPass
Particulate AnalysisMeasure the number of particulates in saline pre-filled syringes≥ 10 μm < 3,000≥ 25 μm < 300Pass

The subject device met all the predetermined acceptance criteria for the above listed performance, packaging and biocompatibility tests.

Page 14

Page 10 of 10

Clinical TestingNot applicable.
ConclusionThe conclusion drawn from the nonclinical test demonstrates that the subject device 0.9% Sodium Chloride Injection, USP BD PosiFlush SP Syringe is as safe, as effective, and performs as well as or better than the legally marketed predicate device K161552.

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).