Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K231091

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-07-10

    (84 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AcQMap High Resolution Imaging and Mapping System is intended for use in patients for whom electrophysiology procedures have been prescribed.

    When used with the AcQMap Catheters, the AcQMap High Resolution Imaging System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber and displaying electrical impulses as either charge density-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone.

    AND

    When used with the specified Patient Electrodes, the AcQMap High Resolution Imaging System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart.

    OR

    When used with conventional electrophysiology catheters, the AcQMap High Resolution Imaging and Mapping System provides information about the electrical activity of the heart and about catheter location during the procedure.

    Device Description

    The AcQMap High Resolution Imaging and Mapping System (hereafter referred to as the AcQMap System) operates outside of the sterile field and consists of the AcQMap Console, the AcQMap Workstation and the AcQMap Auxiliary Interface Box.

    The AcQMap System is a diagnostic recording system. This computer-based system is intended for use in the Electrophysiology (EP) Lab, and it is capable of imaging, navigation and mapping of the atrial chambers of the heart.

    The AcQMap System hardware consists of three (3) functional subsystems:

    • . Ultrasound imaging,
    • ECG and EGM recording; and
    • Impedance based electrode Localization.

    The AcQMap System is used in conjunction with the AcQMap 3D Imaging and Mapping Catheter (hereafter referred to as the AcQMap Catheter). The AcQMap System provides:

    • . 3-D cardiac chamber reconstruction – Contact and non-contact (ultrasound),
    • 3-D position of the AcQMap Catheter and conventional electrophysiology catheters,
    • . Cardiac electrical activity as waveform traces,
    • . Contact LAT and voltage amplitude maps
    • Remapping of the chamber at any time during the procedure; and
    • . Dynamic, three-dimensional, charge density maps overlaid on the cardiac chamber reconstruction to show chamber-wide electrical activation.

    The AcQMap System is intended to create a surface reconstruction of the cardiac chamber as well as an electrical map of the substrate. The surface reconstruction and electrical map are then used by physicians to identify the source(s) of the arrhythmia.

    In addition, the AcQMap System allows physicians to perform traditional contact mapping activities, including establishing a coordinate system, localizing conventional electrophysiology catheters relative to one (1) another within the coordinate system, recording contact electrograms, and initiating a procedure without the AcQMap Catheter present. Based on the information captured in the contact electrograms, the physician may decide to treat an arrythmia without deploying the AcQMap Catheter.

    There are no modifications to the hardware of the AcQMap System nor its accessories. The modifications to the AcQMap System are specific to the software, which include:

    • The addition of the feature, Dynamic Referencing. Dynamic Referencing actively monitors impedance-based stability events like catheter drift or shift. Users will be offered guided notifications when shifts or drifts are automatically detected. Dynamic Referencing combines the capabilities of a Virtual Position Reference (VPR) and a Physical Position Reference (PPR), allowing them to be used together simultaneously. It is designed to be used instead of VPR or PPR alone. VPR and PPR remain available as options and
    • Updated elements of the user interface to make it easier for the user to conduct a variety of actions within the software.
    AI/ML Overview

    The provided text focuses on the substantial equivalence of the AcQMap High Resolution Imaging and Mapping System (K231091) to its predicate device (K222209), with an emphasis on changes to software and user interface. It states that "no further clinical testing is required" for the subject device and refers to clinical testing performed for the original AcQMap System (K170948).

    Therefore, the acceptance criteria and study information will be derived from the description of the original AcQMap System's clinical study (DDRAMATIC-SVT) referenced in the document.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative "acceptance criteria" and "reported device performance" in a table format for the subject device (K231091) or for the original device (K170948) from its clinical study. Instead, it makes a general statement that the results of the DDRAMATIC-SVT study "demonstrated that the AcQMap System is substantially equivalent to the predicate device."

    For the subject device (K231091), the modifications are described as improving "ease-of-use of some system capabilities" and having "no changes to the current mapping data." The conclusion is that "the subject device meets its established performance specifications necessary for performance during its intended use."

    Without specific metrics or thresholds, a direct table of acceptance criteria and performance cannot be created from this text.

    2. Sample size used for the test set and the data provenance

    • Sample Size: 84 patients
    • Data Provenance: Prospective, non-randomized, open-label study conducted at eight (8) clinical sites outside the U.S.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number of experts or their qualifications for establishing the ground truth in the DDRAMATIC-SVT study. It only mentions that the system displays electrical impulses and maps for physicians to "identify the source(s) of the arrhythmia."

    4. Adjudication method for the test set

    The document does not mention any adjudication method (e.g., 2+1, 3+1, none) for the test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document describes the clinical study (DDRAMATIC-SVT) as a non-randomized, open-label study to establish substantial equivalence. It does not indicate that it was a multi-reader multi-case (MRMC) comparative effectiveness study with or without AI assistance, nor does it provide any effect size concerning human reader improvement. The focus is on the device's diagnostic capabilities, not on direct physician performance improvement with the device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document describes the AcQMap System as "intended to be used by physicians" to reconstruct chambers and display electrical impulses and maps to identify arrhythmia sources. This implies a human-in-the-loop performance context. The text does not mention any standalone algorithm-only performance assessment.

    7. The type of ground truth used

    The text describes the AcQMap System's purpose as helping physicians "identify the source(s) of the arrhythmia" using "charge density-based or voltage-based maps." This strongly suggests that the ground truth for the clinical study would be derived from electrophysiological measurements and clinical consensus regarding the arrhythmia source, but it does not explicitly state "expert consensus," "pathology," or "outcomes data." However, given the context of electrophysiology procedures, expert interpretation of electrophysiological data would be the most likely form of ground truth.

    8. The sample size for the training set

    The document focuses on the clinical validation study (DDRAMATIC-SVT) for the original device (K170948), which served as the test set for demonstrating substantial equivalence. It does not provide information about a separate "training set" used for developing the software features or maps of the AcQMap System.

    9. How the ground truth for the training set was established

    As no training set is explicitly mentioned, how its ground truth was established is not described in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222209

    Validate with FDA (Live)

    Date Cleared
    2022-10-17

    (84 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AcQMap High Resolution Imaging and Mapping System is intended for use in patients for whom electrophysiology procedures have been prescribed. When used with the AcQMap Catheters, the AcQMap High Resolution Imaging and Mapping System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either charge density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone. AND When used with the specified Patient Electrodes, the AcQMap High Resolution Imaging System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart. OR When used with conventional electrophysiology catheters, the AcQMap High Resolution Imaging System provides information about the electrical activity of the heart and about catheter location during the procedure.

    Device Description

    The AcQMap High Resolution Imaging and Mapping System operates outside of the sterile field and consists of the AcQMap Console, the AcQMap Workstation and the AcQMap Auxiliary Interface Box. The AcQMap High Resolution Imaging and Mapping System is a diagnostic recording system. This computer-based system is intended for use in the Electrophysiology (EP) Lab, and it is capable of imaging, navigation, and mapping of the atrial chambers of the heart. The AcQMap High Resolution Imaging and Mapping System hardware consists of three functional subsystems: Ultrasound imaging, ECG and EGM recording; and Impedance based electrode Localization. The AcQMap High Resolution Imaging and Mapping System is used in conjunction with the associated AcQMap 3D Imaging and Mapping Catheter models 900003 (cleared under K201341). The AcQMap High Resolution Imaging and Mapping System provides: 3-D cardiac chamber reconstruction Contact and non-contact (ultrasound), Three-dimensional position of the AcQMap Catheter and conventional electrophysiology catheters, Cardiac electrical activity as waveform traces, Contact LAT and voltage amplitude maps Remapping of the chamber at any time during the procedure; and Dynamic, three-dimensional, charge density maps overlaid on the cardiac chamber reconstruction to show chamber-wide electrical activation. The AcQMap High Resolution Imaging and Mapping System is intended to create a surface reconstruction of the cardiac chamber as well as an electrical map of the substrate. The surface reconstruction and electrical map are then used by physicians to identify the source(s) of the arrhythmia. Additionally, the AcQMap High Resolution Imaging and Mapping System allows physicians to perform traditional contact mapping activities, including establishing a coordinate system, localizing conventional electrophysiology catheters relative to one another within the coordinate system, recording contact electrograms, and initiating a procedure without the AcQMap Catheter present. Based on the information captured in the contact electrograms, the physician may decide to treat an arrythmia without deploying the AcQMap Catheter. The modifications to the AcQMap High Resolution Imaging System include the addition of new accessories to facilitate connectivity between the cleared AcQMap High Resolution Imaging and Mapping System and DiamondTemp™ Ablation Generator.

    AI/ML Overview

    Below is an analysis of the provided text regarding the AcmMap High Resolution Imaging and Mapping System:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) submission does not specify numerical acceptance criteria for the device's performance regarding imaging, mapping accuracy, or specific clinical outcomes. Instead, the submission relies on demonstrating substantial equivalence to a predicate device (AcQMap High Resolution Imaging and Mapping System, K220784) and the original version of the system (K170948).

    Therefore, the "acceptance criteria" are implicitly met by demonstrating that the modified device performs similarly to the predicate/original device and meets established non-clinical performance specifications.

    Acceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance
    Non-clinical Performance: Meets design specifications, including:The modified device met all established performance specifications. The collective results of nonclinical testing demonstrate that the materials, manufacturing processes, and design meet established specifications for consistent performance.
    - Third party system CompatibilitySuccessfully demonstrated.
    - System RF Attenuation characterizationSuccessfully demonstrated.
    - Therapeutic Waveform Fidelity AssessmentSuccessfully demonstrated.
    - Safety TestingSuccessfully demonstrated.
    - Packaging TestingSuccessfully demonstrated.
    - Inspection and Labeling ReviewSuccessfully demonstrated.
    - Common Mode Choke (CMC) performance verification TestingSuccessfully demonstrated.
    - Transportation Testing (incorporated by reference)Successfully demonstrated.
    - AcQMap High Resolution Imaging and Mapping Verification Testing (incorporated by reference)Successfully demonstrated.
    - System Accuracy Testing (incorporated by reference)Successfully demonstrated.
    - Electromagnetic Compatibility and Electrical Safety Testing (incorporated by reference)Successfully demonstrated.
    - Map Accuracy Evaluation (incorporated by reference)Successfully demonstrated.
    - In-vitro Localization Accuracy Study (incorporated by reference)Successfully demonstrated.
    Clinical Performance: Substantially equivalent to predicate device (K170948) regarding intended use and clinical outcomes for complex arrhythmias.The original AcQMap High Resolution Imaging and Mapping System (K170948) demonstrated substantial equivalence in a clinical study with 84 patients, which is incorporated by reference. The modified device's indications for use are identical, and modifications do not raise new safety or effectiveness concerns.

    Note: The document explicitly states "The collective results of the nonclinical testing... demonstrate that the proposed device does not raise different questions of safety or effectiveness when compared to the predicate device." This implies the performance is at least equivalent to the predicate.

    2. Sample Size Used for the Test Set and Data Provenance

    • Nonclinical Testing Test Set (Modified Device): Not explicitly stated as a number of devices. The description implies testing was conducted on "the modified AcQMap High Resolution Imaging and Mapping System" to verify specifications. It's likely one or a few units were tested for each specific bench test (e.g., RF attenuation, waveform fidelity).
    • Clinical Testing Test Set (Original Device, K170948, incorporated by reference): "84 patients".
    • Data Provenance (Clinical Study): "prospective, non-randomized, open-label study conducted at eight clinical sites outside the U.S." This indicates the data is prospective and from international sites.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The provided document does not specify the number or qualifications of experts for establishing ground truth.

    • For the nonclinical bench testing, "ground truth" would be engineering specifications and measurements.
    • For the clinical study (DDRAMATIC-SVT), it's a study on "Assesment of Therapy In Complex Supraventricular Tachycardia," suggesting clinical outcomes and electrophysiological assessments would be the standard. The direct role of external experts in establishing "ground truth" as a separate process from the study conduct itself is not detailed. Standard clinical trials rely on physician diagnoses and assessments, but the specific process for ground truth establishment by an independent panel is not mentioned.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the test set, either for nonclinical testing or for the clinical study incorporated by reference. Clinical trials usually involve endpoint adjudication, but the details are not included here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study is not mentioned. The clinical study (DDRAMATIC-SVT) was a prospective, non-randomized, open-label study designed to assess the performance of the original AcQMap System (K170948) and demonstrate its substantial equivalence. It does not appear to be a study comparing human readers with and without AI assistance to measure an effect size.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    The document primarily focuses on the system as a diagnostic recording system, capable of imaging, navigation, and mapping for use by physicians to identify sources of arrhythmia. The outputs (3-D reconstructions, electrical maps, catheter positions, waveforms) are intended to assist physicians.

    While there are "System Accuracy Testing" and "Map Accuracy Evaluation" as part of the non-clinical testing, these likely assess the technical accuracy of the algorithms in generating mapping data against defined inputs, rather than an "algorithm only" clinical performance study in the sense of making diagnostic decisions without human interpretation. The system is designed to provide information to the clinician, not to make a standalone diagnosis.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    • Nonclinical Testing: Engineering specifications, physical measurements, and established safety standards (IEC 60601-1, IEC 60601-1-2, IEC 62366-1, etc.) would serve as ground truth for these tests.
    • Clinical Study (DDRAMATIC-SVT): The study title "Assessment of Therapy In Complex Supraventricular Tachycardia" suggests the ground truth would involve clinical outcomes related to the diagnosis and treatment of arrhythmias, likely including electrophysiological assessments and patient responses to therapy. The specifics of how "ground truth" was established (e.g., confirmed arrhythmia diagnosis by an independent panel, successful ablation outcomes) are not detailed in this excerpt.

    8. The Sample Size for the Training Set

    The document does not mention a training set sample size. The device is described as "Programable diagnostic computer and Ultrasonic pulsed echo imaging system," but the application is for a modification to an existing system (adding connectivity accessories). The previous versions of the system (K220784 predicate and K170948 original) would have undergone validation, but whether a machine learning model with a distinct "training set" was used for the core mapping functionalities is not specified in this document. Given the nature of previous predicate devices for mapping systems, it's less likely to be a deep learning model requiring a large training set in the modern AI sense, and more likely based on signal processing and physics-based reconstruction algorithms.

    9. How the Ground Truth for the Training Set Was Established

    Since no explicit training set is mentioned for an AI/ML component, the method for establishing its ground truth is not described. For traditional signal processing and reconstruction algorithms, "ground truth" during development often comes from simulated data, phantom studies, and expert-validated real-world data used for algorithm design and initial tuning. These details are not in the provided document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210685

    Validate with FDA (Live)

    Date Cleared
    2021-04-06

    (29 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AcQCross™ Qx Integrated Transseptal DilatorNeedle is indicated to puncture the interatrial septum to gain access to the left side of the heart whereby various cardiovascular catheters are introduced.

    Device Description

    AcQCross™ Qx combines the conventional vessel dilator and transseptal needle into a single device (Figure 1). AcQCross™ Ox consists of an elongated shaft with a tapered tip and central lumen to track over a guidewire. The lumen of AcQCross™ Qx is fitted with a hollow stainless steel transseptal needle (Figure 2). Both the shaft and needle are connected to the proximal handle of AcQCross™ QX. The lumen of the needle will allow for guidewires up to 0.032" in diameter. The needle is affixed to a spring-tensioned actuator in the handle of AcQCross™ QX that prevents needle extension until the operator purposely advances the needle via a slider button located on the outer surface of the handle. The proximal handle is fitted with a Luer connector to gain access to the central lumen of the needle. The handle is also fitted with an electrical connector that allows for monitoring intracardiac electrograms (EGMs) from the needle while in the heart utilizing the EGM adapter cable, and/or allows for the application of radiofrequency (RF) current from an electrosurgical generator to facilitate the septal puncture utilizing the ES adapter cable. AcQCross™ Qx is for single-use only and is provided sterile.

    AI/ML Overview

    This document is a 510(k) summary for the AcQCross™ Qx Integrated Transseptal Dilator/Needle, a Class II medical device. It focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study proving performance against specific acceptance criteria. Therefore, several of the requested sections cannot be fully populated as this type of information is generally not included in a 510(k) summary that relies on demonstrating equivalence to a previously cleared device.

    However, I can extract the relevant information regarding performance testing and how substantial equivalence was established.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state numerical acceptance criteria for the device performance tests. Instead, it relies on demonstrating that the AcQCross™ Qx performs "as intended" and that "minor differences do not potentially impact the safety and effectiveness" compared to the predicate device. The performance tests were conducted in accordance with ISO standards and additional bench testing was performed on the modified standalone device.

    Test CategoryRelevant Standard(s) / Description of TestReported Device Performance
    BiocompatibilityAAMI/ANSI/ISO 10993-1:2009 - Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process."The results demonstrate that AcQCross™ Qx meets the requirements of ISO 10993-1 and is biocompatible." (Leveraged from K170373 and K171081. Additional cytotoxicity and hemolysis testing due to a new shaft supplier was performed.)
    SterilizationISO 11135: 2014 - Sterilization of health-care products - Ethylene Oxide - Requirements for development, validation and routine control of a sterilization process for medical devices."AcQCross™ Qx is subjected to the identical ethylene oxide (EO) sterilization process as the predicate device... to meet a sterility assurance level (SAL) of 10-6." (Leveraged from K170373 and K171081. Minor packaging change adopted per AAMI TIR28, requiring no further process validation.)
    Electrical Safety and EMCANSI/AAMI IEC 60601-1:2005, IEC 60601-1-2, IEC 60601-2-2.Testing leveraged from K170373, K171081, and K193509. (No specific performance data detailed in this summary, but implies compliance with standards.)
    Bench Testing (General)ISO 11070: 2014 --- Sterile single-use intravascular introducers, dilators and guidewires. ISO 10555-1: 2013 --- Sterile, single-use intravascular catheters --- Part 1: General Requirements."AcQCross™ Qx performs as intended." (Leveraged from K170373, K171081, and K193509.)
    Bench Testing (Specific to modified standalone AcQCross™ Qx)Curve retention(No specific quantitative results are provided in the summary, but the conclusion states "The non-clinical tests demonstrated that the device is as safe and effective as the predicate device.")
    Pushability
    Needle Actuation
    Electrical Continuity
    Shaft to handle tensile
    Bench Testing (with compatible sheaths)Needle Actuation"AcQCross™ performs as intended, and the addition of compatible sheath configurations do not potentially impact the safety or effectiveness of the device as compared to the predicate device."
    Visual inspection
    Aspiration/Flushing
    Electrical Continuity

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document is a 510(k) summary for a medical device that relies on substantial equivalence. It does not contain information about the sample sizes used for the performance tests. The data provenance (country of origin, retrospective/prospective) is also not specified, which is typical for this type of regulatory submission focusing on engineering and bench testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable to the type of testing described in this document. The performance tests are engineering-based (biocompatibility, sterilization, electrical safety, bench testing) and do not involve "ground truth" derived from expert medical interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods are typically used in clinical studies involving interpretation of data by multiple experts, which is not the case for the engineering performance tests described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device (transseptal dilator/needle) is a mechanical medical device, not an AI-powered diagnostic tool. Therefore, MRMC studies and "human reader improvement with AI" are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable as the device is a physical medical instrument, not an algorithm. The phrase "standalone device" in the document refers to the AcQCross™ Qx being cleared for use by itself (with compatible sheaths) rather than only as part of specific introducer sets.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Ground truth, in the context of clinical or diagnostic studies, is not applicable here. The "truth" for the performance tests (e.g., biocompatibility showing no toxicity, sterilization achieving SAL, mechanical forces meeting specifications) is established through standardized laboratory testing and engineering validation.

    8. The sample size for the training set

    This information is not applicable. The device is a mechanical tool; it does not involve training data sets or machine learning algorithms.

    9. How the ground truth for the training set was established

    This information is not applicable as there is no training set for this type of medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1