K Number
K231091

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-07-10

(84 days)

Product Code
Regulation Number
870.1425
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AcQMap High Resolution Imaging and Mapping System is intended for use in patients for whom electrophysiology procedures have been prescribed.

When used with the AcQMap Catheters, the AcQMap High Resolution Imaging System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber and displaying electrical impulses as either charge density-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone.

AND

When used with the specified Patient Electrodes, the AcQMap High Resolution Imaging System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart.

OR

When used with conventional electrophysiology catheters, the AcQMap High Resolution Imaging and Mapping System provides information about the electrical activity of the heart and about catheter location during the procedure.

Device Description

The AcQMap High Resolution Imaging and Mapping System (hereafter referred to as the AcQMap System) operates outside of the sterile field and consists of the AcQMap Console, the AcQMap Workstation and the AcQMap Auxiliary Interface Box.

The AcQMap System is a diagnostic recording system. This computer-based system is intended for use in the Electrophysiology (EP) Lab, and it is capable of imaging, navigation and mapping of the atrial chambers of the heart.

The AcQMap System hardware consists of three (3) functional subsystems:

  • . Ultrasound imaging,
  • ECG and EGM recording; and
  • Impedance based electrode Localization.

The AcQMap System is used in conjunction with the AcQMap 3D Imaging and Mapping Catheter (hereafter referred to as the AcQMap Catheter). The AcQMap System provides:

  • . 3-D cardiac chamber reconstruction – Contact and non-contact (ultrasound),
  • 3-D position of the AcQMap Catheter and conventional electrophysiology catheters,
  • . Cardiac electrical activity as waveform traces,
  • . Contact LAT and voltage amplitude maps
  • Remapping of the chamber at any time during the procedure; and
  • . Dynamic, three-dimensional, charge density maps overlaid on the cardiac chamber reconstruction to show chamber-wide electrical activation.

The AcQMap System is intended to create a surface reconstruction of the cardiac chamber as well as an electrical map of the substrate. The surface reconstruction and electrical map are then used by physicians to identify the source(s) of the arrhythmia.

In addition, the AcQMap System allows physicians to perform traditional contact mapping activities, including establishing a coordinate system, localizing conventional electrophysiology catheters relative to one (1) another within the coordinate system, recording contact electrograms, and initiating a procedure without the AcQMap Catheter present. Based on the information captured in the contact electrograms, the physician may decide to treat an arrythmia without deploying the AcQMap Catheter.

There are no modifications to the hardware of the AcQMap System nor its accessories. The modifications to the AcQMap System are specific to the software, which include:

  • The addition of the feature, Dynamic Referencing. Dynamic Referencing actively monitors impedance-based stability events like catheter drift or shift. Users will be offered guided notifications when shifts or drifts are automatically detected. Dynamic Referencing combines the capabilities of a Virtual Position Reference (VPR) and a Physical Position Reference (PPR), allowing them to be used together simultaneously. It is designed to be used instead of VPR or PPR alone. VPR and PPR remain available as options and
  • Updated elements of the user interface to make it easier for the user to conduct a variety of actions within the software.
AI/ML Overview

The provided text focuses on the substantial equivalence of the AcQMap High Resolution Imaging and Mapping System (K231091) to its predicate device (K222209), with an emphasis on changes to software and user interface. It states that "no further clinical testing is required" for the subject device and refers to clinical testing performed for the original AcQMap System (K170948).

Therefore, the acceptance criteria and study information will be derived from the description of the original AcQMap System's clinical study (DDRAMATIC-SVT) referenced in the document.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state quantitative "acceptance criteria" and "reported device performance" in a table format for the subject device (K231091) or for the original device (K170948) from its clinical study. Instead, it makes a general statement that the results of the DDRAMATIC-SVT study "demonstrated that the AcQMap System is substantially equivalent to the predicate device."

For the subject device (K231091), the modifications are described as improving "ease-of-use of some system capabilities" and having "no changes to the current mapping data." The conclusion is that "the subject device meets its established performance specifications necessary for performance during its intended use."

Without specific metrics or thresholds, a direct table of acceptance criteria and performance cannot be created from this text.

2. Sample size used for the test set and the data provenance

  • Sample Size: 84 patients
  • Data Provenance: Prospective, non-randomized, open-label study conducted at eight (8) clinical sites outside the U.S.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not specify the number of experts or their qualifications for establishing the ground truth in the DDRAMATIC-SVT study. It only mentions that the system displays electrical impulses and maps for physicians to "identify the source(s) of the arrhythmia."

4. Adjudication method for the test set

The document does not mention any adjudication method (e.g., 2+1, 3+1, none) for the test set.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document describes the clinical study (DDRAMATIC-SVT) as a non-randomized, open-label study to establish substantial equivalence. It does not indicate that it was a multi-reader multi-case (MRMC) comparative effectiveness study with or without AI assistance, nor does it provide any effect size concerning human reader improvement. The focus is on the device's diagnostic capabilities, not on direct physician performance improvement with the device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The document describes the AcQMap System as "intended to be used by physicians" to reconstruct chambers and display electrical impulses and maps to identify arrhythmia sources. This implies a human-in-the-loop performance context. The text does not mention any standalone algorithm-only performance assessment.

7. The type of ground truth used

The text describes the AcQMap System's purpose as helping physicians "identify the source(s) of the arrhythmia" using "charge density-based or voltage-based maps." This strongly suggests that the ground truth for the clinical study would be derived from electrophysiological measurements and clinical consensus regarding the arrhythmia source, but it does not explicitly state "expert consensus," "pathology," or "outcomes data." However, given the context of electrophysiology procedures, expert interpretation of electrophysiological data would be the most likely form of ground truth.

8. The sample size for the training set

The document focuses on the clinical validation study (DDRAMATIC-SVT) for the original device (K170948), which served as the test set for demonstrating substantial equivalence. It does not provide information about a separate "training set" used for developing the software features or maps of the AcQMap System.

9. How the ground truth for the training set was established

As no training set is explicitly mentioned, how its ground truth was established is not described in the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 10, 2023

Acutus Medical Inc Ryan Kenney Director, Regulatory Affairs 2210 Faraday Ave, Suite 100 Carlsbad, California 92008

Re: K231091

Trade/Device Name: AcQMap High Resolution Imaging and Mapping System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DOK, ITX, IYO Dated: April 17, 2023 Received: April 17, 2023

Dear Ryan Kenney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Aneesh S. Deoras -S

Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231091

Device Name

AcQMap High Resolution Imaging and Mapping System

Indications for Use (Describe)

The AcQMap High Resolution Imaging and Mapping System is intended for use in patients for whom electrophysiology procedures have been prescribed.

When used with the AcQMap Catheters, the AcQMap High Resolution Imaging System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber and displaying electrical impulses as either charge density-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone.

AND

When used with the specified Patient Electrodes, the AcQMap High Resolution Imaging System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart.

OR

When used with conventional electrophysiology catheters, the AcQMap High Resolution Imaging and Mapping System provides information about the electrical activity of the heart and about catheter location during the procedure.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Acutus Medical. The logo is composed of the word "ACUTUS" in a bold, sans-serif font, with a small orange dot above the letter "A". Below the word "ACUTUS" is the word "MEDICAL" in a smaller, sans-serif font. The logo is simple and modern, and the orange dot adds a touch of color.

Applicant/Correspondent:Applicant/Correspondent Contact:
Acutus Medical, Inc.Ryan Kenney
2210 Faraday Ave., Suite 100Director, Regulatory Affairs
Carlsbad, CA 92008 USAEmail: ryan.kenney@acutus.com
Phone: +1 (442) 232-6080
Fax: +1 (442) 232-6081
Date Prepared:July 6, 2023
Device Classification Name:Computer, Diagnostic, Programmable
510(k) Number:K231091
Device Name:AcQMap High Resolution Imaging and Mapping System
Regulation Number:870.1425
Classification Product Code:DQK
Subsequent Product Codes:ITX, IYO
Predicate Device:AcQMap High Resolution Imaging and Mapping System (K222209)

Device Description:

The AcQMap High Resolution Imaging and Mapping System (hereafter referred to as the AcQMap System) operates outside of the sterile field and consists of the AcQMap Console, the AcQMap Workstation and the AcQMap Auxiliary Interface Box.

The AcQMap System is a diagnostic recording system. This computer-based system is intended for use in the Electrophysiology (EP) Lab, and it is capable of imaging, navigation and mapping of the atrial chambers of the heart.

The AcQMap System hardware consists of three (3) functional subsystems:

  • . Ultrasound imaging,
  • ECG and EGM recording; and
  • Impedance based electrode Localization.

The AcQMap System is used in conjunction with the AcQMap 3D Imaging and Mapping Catheter (hereafter referred to as the AcQMap Catheter). The AcQMap System provides:

  • . 3-D cardiac chamber reconstruction – Contact and non-contact (ultrasound),
  • 3-D position of the AcQMap Catheter and conventional electrophysiology catheters,
  • . Cardiac electrical activity as waveform traces,
  • . Contact LAT and voltage amplitude maps
  • Remapping of the chamber at any time during the procedure; and
  • . Dynamic, three-dimensional, charge density maps overlaid on the cardiac chamber reconstruction to show chamber-wide electrical activation.

The AcQMap System is intended to create a surface reconstruction of the cardiac chamber as well as an electrical map of the substrate. The surface reconstruction and electrical map are then used by physicians to identify the source(s) of the arrhythmia.

{4}------------------------------------------------

In addition, the AcQMap System allows physicians to perform traditional contact mapping activities, including establishing a coordinate system, localizing conventional electrophysiology catheters relative to one (1) another within the coordinate system, recording contact electrograms, and initiating a procedure without the AcQMap Catheter present. Based on the information captured in the contact electrograms, the physician may decide to treat an arrythmia without deploying the AcQMap Catheter.

There are no modifications to the hardware of the AcQMap System nor its accessories. The modifications to the AcQMap System are specific to the software, which include:

  • The addition of the feature, Dynamic Referencing. Dynamic Referencing actively monitors impedance-based stability events like catheter drift or shift. Users will be offered guided notifications when shifts or drifts are automatically detected. Dynamic Referencing combines the capabilities of a Virtual Position Reference (VPR) and a Physical Position Reference (PPR), allowing them to be used together simultaneously. It is designed to be used instead of VPR or PPR alone. VPR and PPR remain available as options and
  • Updated elements of the user interface to make it easier for the user to conduct a variety of actions within the software.

Indications for Use Statement:

The AcQMap System is intended for use in patients for whom electrophysiology procedures have been prescribed.

When used with the AcQMap Catheters, the AcQMap System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either charge density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone.

AND

When used with the specified Patient Electrodes, the AcQMap System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart.

OR

When used with conventional electrophysiology catheters, the AcQMap System provides information about the electrical activity of the heart and about catheter location during the procedure.

{5}------------------------------------------------

UTUS® M D

Comparison of Technological Characteristics:

Table 1: Comparison of Technological Characteristics
CharacteristicsAcQMap High Resolution Imaging and Mapping SystemPredicate Device(K222209)AcQMap High Resolution Imaging and Mapping SystemSubject Device(K231091)
RegulationNumber870.1425Same as K222209
ClassificationProduct CodeDQKSame as K222209
SubsequentProduct CodesITX, IYOSame as K222209
Indications for UseThe AcQMap System is intended for use in patients for whomelectrophysiology procedures have been prescribed.When used with the AcQMap Catheters, the AcQMap Systemis intended to be used to reconstruct the selected chamberfrom ultrasound data for purposes of visualizing the chamberanatomy and displaying electrical impulses as either chargedensity-based or voltage-based maps of complex arrhythmiasthat may be difficult to identify using conventional mappingsystems alone.ANDWhen used with the specified Patient Electrodes, the AcQMapSystem is intended to display the position of AcQMapCatheters and conventional electrophysiology (EP) catheters inthe heart.ORWhen used with conventional electrophysiology catheters, theAcoMan System provides information about the electricalSame as K222209
Table 1: Comparison of Technological Characteristics
CharacteristicsAcQMap High Resolution Imaging and Mapping SystemPredicate Device(K222209)AcQMap High Resolution Imaging and Mapping SystemSubject Device(K231091)
activity of the heart and about catheter location during the procedure.
Patient Anatomy• Intracardiac StructuresSame as K222209
Testing to SupportSubstantialEquivalence• Software Verification/Validation• Electromagnetic and Electrical Safety• Verification Testing• Accuracy Testing• Animal TestingSame as K222209
System SafetyStandards• IEC 60601-1:2005 /A1:2012• IEC 60601-1-2:2014• IEC 60601-2-25:2015• IEC 60601-2-37:2015• IEC 62366-1: 2015Same as K222209
Physical Characteristics
SystemComponents• Console• Workstation• Workstation Cable• Auxiliary Interface Box• ECG Input Cable• Ampere Ablation Catheter Adapter Cable• Ampere RF Generator Adapter Cable• ECG Output Cable• Ablation Reference Cable• Ablation Electrogram Cable• ECG w/Snaps Cable• ECG POST Cable• 2mm Pin Jumper Set• Patient Electrode Kit• Ampere Generator AdapterSame as K222209
Table 1: Comparison of Technological Characteristics
CharacteristicsAcQMap High Resolution Imaging and Mapping System Predicate Device (K222209)AcQMap High Resolution Imaging and Mapping System Subject Device (K231091)
SmartAblate Generator Adapter Maestro Generator Adapter Adapter Cable-Short - Ablation Adapter Adapter Cable-Long - Ablation Adapter SmartAblate Adapter Cable - Catheter Maestro Adapter Cable - Catheter Maestro Adapter Cable - Generator Carto Force Adapter Cable DiamondTemp Generator Adapter DiamondTemp Adapter Cable - Generator
Visual/Mapping Characteristics3-D cardiac chamber reconstructions - Contact and non-contact (ultrasound) 3-D position of the AcQMap Catheter and conventional electrophysiology catheters Cardiac electrical activity as waveform traces Contact LAT and voltage amplitude maps Remapping of the chamber at any time during the procedure Dynamic, 3-D, Charge Density maps overlaid on the cardiac chamber reconstruction to show chamber-wide electrical activationSame as K222209
Visualization Device/CatheterAcQMap Catheter (electrodes & transducers) or Conventional electrophysiology cathetersSame as K222209
Physical Characteristics – Console/Amplifier Comparison
Dimensions99 cm L x 58 cm W x 76 cm DSame as K222209
Weight Maximum80 kgSame as K222209
Power Requirement100-127 VAC, 50/60 Hz, 220-230 VAC, 50 HzSame as K222209
Input Current4.6 ASame as K222209
Table 1: Comparison of Technological Characteristics
CharacteristicsAcQMap High Resolution Imaging and Mapping SystemPredicate Device(K222209)AcQMap High Resolution Imaging and Mapping SystemSubject Device(K231091)
Fuse Protection250 V, 6.3A, two high breaking capacity fusesSame as K222209
System Specifications
Safety InformationIEC 60601-1, Class I, Type Defibrillator Protected CF, continuous operation, no sterilization, equipment not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen or nitrous oxideSame as K222209
Ingress ProtectionThe Console is rated IP20Same as K222209
Functional and Performance Characteristics
Ultrasound OutputFrequency: 10 MHz ± 400 kHzMaximum Voltage: 50 V p-pMaximum Power: 1 W PeakSame as K222209
UltrasoundPerformanceSingle Operating ModeThermal Index: < 1.0Mechanical Index: < 1.0Same as K222209
LocalizationOutputFrequency: Variable 15 kHz to 50 kHzMaximum Current: 1.2 mASame as K222209
ECG & EGM InputBandwidth: 0.05 Hz to 500 HzResolution: ± 1 uVTiming Accuracy: ± 1.6 microsecondSame as K222209
Front Panel Connections
AcQMap CatheterCustom, Black, Defibrillator Protected Type CFSame as K222209
ECG Input12-pin, Latching, Red, Defibrillator Protected Type BFSame as K222209
ECG Output14-pin, Latching, BlueSame as K222209
Auxiliary InterfaceBoxCustom, Green, Defibrillator Protected Type CFSame as K222209
AcQRef IntroducerSheath orElectricalReferenceCatheter1, 2mm Female, Yellow, Defibrillator Protected Type CFSame as K222209
Table 1: Comparison of Technological Characteristics
CharacteristicsAcQMap High Resolution Imaging and Mapping SystemPredicate Device(K222209)AcQMap High Resolution Imaging and Mapping SystemSubject Device(K231091)
LocalizationReferenceElectrodes6, 2-pin, Square, Multi-color, Defibrillator Protected Type BFSame as K222209
Patient ReferenceElectrode1, 2-pin, Square, Blue, Defibrillator Protected Type BFSame as K222209
AblationGenerator10-pin, Latching, GreySame as K222209
Ablation Catheter10-pin, Latching, Grey, Defibrillator Protected Type CFSame as K222209
AblationReference1, 2mm, Female, Black, Defibrillator Protected Type BFSame as K222209
AblationElectrogramInterface1, 13-pin, Latching, WhiteSame as K222209

{6}------------------------------------------------

Cutus. M D

{7}------------------------------------------------

JTUS. M D

K231091

{8}------------------------------------------------

US® M D

{9}------------------------------------------------

TTUS® M D

{10}------------------------------------------------

Substantial Equivalence:

The AcQMap System is intended to reconstruct the physiology and display the anatomic and electrical data in the same way. These modification are to improve the ease-of-use of some system capabilities. The updates to the mapping software includes:

  • The addition of a corresponding mV scale in the color bar of charge density amplitude maps;
  • User-initiated system checks for common connection errors such as cable disconnections and patch placement locations; and
  • An improvement to the generation of the anatomic surface from ultrasound points.

These modification are to improve the ease-of-use of some system capabilities. There are no changes to the current mapping data. There are no differences between the subject device and the predicate(s) with respect to indications and intended use. Any differences in the technological characteristics between the devices do not raise any different questions of safety or effectiveness. Thus, the subject device is substantially equivalent to the predicate device.

Performance Data:

All necessary bench testing was conducted on the subject device to support a determination of substantial equivalence to the predicate device. The necessary clinical testing was completed for the original AcQMap System (K170948) and is incorporated by reference. No further clinical testing is required to support the subject device.

Non-Clinical Testing Summary:

The necessary bench testing was performed on the subject device to ensure that it conforms to the design specifications and to support a determination of substantial equivalence to the predicate device.

The following bench testing was repeated for the subject device:

  • Software Verification and Validation .
    The balance of testing is incorporated by reference to the original AcQMap System (K170948), includes the following:

  • Transportation Testing

  • AcQMap High Resolution Imaging and Mapping Verification Testing ●

  • System Accuracy Testing

  • Electromagnetic Compatibility and Electrical Safety Testing

  • AcQMap Catheter Validation Testing-Animal Study

  • Accuracy Validation Testing Animal Study

  • Clinical Simulation (Reliability)

  • Map Accuracy Evaluation

  • In-vitro Localization Accuracy Study

The subject device was tested to verify that it meets the established performance specifications. The collective results of the testing demonstrate that the design of the subject device meets its established performance specifications necessary for performance during its intended use.

{11}------------------------------------------------

The collective results of the non-clinical testing, either repeated for the subject device or incorporated by reference to the original AcQMap System (K170948), demonstrate that the materials chosen, the manufacturing processes, and design of the subject device meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the subject device does not raise different questions of safety or effectiveness when compared to the predicate device.

Clinical Testing Summary:

As discussed above, no further clinical testing is required to support the subject device. The necessary clinical testing was completed for the original AcQMap System (K170948) and is incorporated by reference. That study, entitled, "Dipole Density Right (and Left) Atrial Mapping and Assessment of Therapy in Complex Supraventricular Tachycardia, (DDRAMATIC-SVT)" was a prospective, nonrandomized, open-label study conducted at eight (8) clinical sites outside the U.S. The results for eightyfour (84) patients demonstrated that the AcQMap System is substantially equivalent to the predicate device.

Conclusion:

Extensive non-clinical performance testing, either repeated for the subject device or incorporated by reference to the original AcQMap System (K170948), was conducted to evaluate the overall performance of the device. The clinical validation of the original AcQMap System (K170948) is applicable to the subject device. Based on the performance testing and the technological characteristics, it can be concluded that the subject device is substantially equivalent to the predicate device.

§ 870.1425 Programmable diagnostic computer.

(a)
Identification. A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.(b)
Classification. Class II (performance standards).