510(k) Data Aggregation

    K Number
    K152740

    Validate with FDA (Live)

    Date Cleared
    2015-11-16

    (54 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Glidewire is designed to direct a catheter to the desired anatomical location in the peripheral vasculature during diagnostic or interventional procedures. This device is not intended for neurovascular or coronary interventions.

    Device Description

    The Radifocus Glidewire is a guide wire which is designed to fit inside a percutaneous catheter for the purpose of directing a catheter through the blood vessel. It is provided sterile and is intended for single use only. It consists of a Nickel Titanium alloy core wire; a polyurethane coating (containing tungsten) and a hydrophilic polymer coating are applied to the entire wire. There are two shaft configurations: standard and stiff. There are two distal tip shapes: straight and angled, and two types of flexible part lengths of the tip: 3 and 5cm. The Radifocus Glidewire is packaged in a plastic holder that is contained within an individual package. A guide wire inserter is included within the individual package to assist with the insertion of the guide wire into a catheter.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the "Radifocus Glidewire". This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials for a novel device.

    Therefore, the concept of "acceptance criteria" and a "study that proves the device meets the acceptance criteria" in the context of an AI/Software as a Medical Device (SaMD), as implied by the detailed questions about ground truth, expert adjudication, MRMC studies, and standalone performance, does not directly apply to this document.

    The document describes performance testing for a physical medical device (a guide wire) to ensure it meets established engineering and material standards and demonstrates substantial equivalence to a predicate device. It does not involve AI or software development that would require an AI-specific validation study as outlined in your prompt.

    Here's a breakdown of why the requested information cannot be extracted from this document, and what is provided:

    Why the requested information for AI/SaMD is not applicable/available:

    • No AI or Software Component: The Radifocus Glidewire is a physical guide wire. There is no mention of any AI algorithm, software, or digital health component.
    • No "Acceptance Criteria" for AI Performance: The "acceptance criteria" discussed in the document relate to physical properties and performance tests of the guide wire (e.g., surface, radiodetectability, fracture test, tensile force, torque strength, biocompatibility, sterilization). These are not AI performance metrics like sensitivity, specificity, AUC, or F-measure.
    • No "Ground Truth" for AI/Imaging Data: Since there's no AI interpreting images or patient data, there's no need for establishing ground truth through expert consensus, pathology, or outcomes data, nor for training/test sets sourced from specific countries or collection methodologies (retrospective/prospective).
    • No "Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study": This type of study is relevant for evaluating human reader performance with and without AI assistance in interpreting diagnostic images. It's not applicable to a physical guide wire.
    • No "Standalone (Algorithm Only) Performance": This refers to the performance of an AI algorithm in isolation. It's not relevant here.
    • No "Sample Sizes for Training/Test Sets" or "Adjudication Methods" for AI: These concepts are tied to machine learning model development and validation, which is absent from this submission.

    What the document does provide regarding "acceptance criteria" and "proof":

    The document details the non-clinical tests performed to demonstrate the device's performance and substantial equivalence to the predicate device.

    1. A table of acceptance criteria and the reported device performance:

      The document provides tables of performance tests based on ISO standards and FDA guidance documents, stating that all samples tested met the standard applicable to each test, and the subject device complies with the acceptance criteria established based on the predicate device and/or the FDA guidance documents. It doesn't provide specific numerical values for each test result, but rather a statement of compliance.

      TestStandard / Acceptance Criteria (Implicitly Met)Reported Device Performance
      ISO 11070:2014-based Tests:
      SurfaceSection 4.3 of ISO 11070:2014Met the standard applicable to the test. No new issues of safety and effectiveness were raised.
      RadiodetectabilitySection 4.5 of ISO 11070:2014 & ASTM F640-12Met the standard applicable to the test. Only non-aged sample was tested.
      Fracture TestSection 8.4 of ISO 11070:2014Met the standard applicable to the test. No new issues of safety and effectiveness were raised.
      Flexing TestSection 8.5 of ISO 11070:2014Met the standard applicable to the test. No new issues of safety and effectiveness were raised.
      Peak Tensile Force of Guidewire (ISO)Section 8.6 of ISO 11070:2014Met the standard applicable to the test. No new issues of safety and effectiveness were raised.
      FDA Guidance & In-house Standard-based Tests:
      Peak Tensile Force of Guidewire (FDA/In-house)3.a of Coronary and Cerebrovascular Guidewire Guidance, January 1995 & In-house StandardComplies with the acceptance criteria established based on the predicate device and/or the FDA guidance documents.
      Torque Strength3.b of Coronary and Cerebrovascular Guidewire Guidance, January 1995 & In-house StandardComplies with the acceptance criteria established based on the predicate device and/or the FDA guidance documents.
      Torqueability3.c of Coronary and Cerebrovascular Guidewire Guidance, January 1995 & In-house StandardComplies with the acceptance criteria established based on the predicate device and/or the FDA guidance documents.
      Tip Flexibility3.d of Coronary and Cerebrovascular Guidewire Guidance, January 1995 & In-house StandardComplies with the acceptance criteria established based on the predicate device and/or the FDA guidance documents.
      Sliding Resistance/Coating Integrity3.e of Coronary and Cerebrovascular Guidewire Guidance, January 1995 & In-house StandardComplies with the acceptance criteria established based on the predicate device and/or the FDA guidance documents.
      Particulate EvaluationVIII.A.13 of Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 2010; USP<788>; In-house StandardComplies with the acceptance criteria established based on the predicate device and/or the FDA guidance documents.
      Product DimensionIn-house StandardComplies with the acceptance criteria established based on the predicate device and/or the FDA guidance documents.
      BiocompatibilityISO 10993-1:2009 (Externally Communicating Devices, Circulating blood, Limited Contact < 24 hrs)Materials are the same as the predicate device (K863138), which has a demonstrated history of safe and effective use. Concluded as biocompatible for its intended use.
      SterilizationISO 11135:2014 for Ethylene Oxide sterilization (min. SAL of 10-6)Validated in accordance with ISO 11135:2014 to provide SAL of 10-6.
      Shelf-LifeNot explicitly stated standard, but tests performed on non-aged and accelerated aged samplesPerformance ensured throughout the labeled shelf life (24 months).
    2. Sample sizes used for the test set and the data provenance:
      The document does not specify exact sample sizes for each test. It mentions "samples" and "accelerated aged samples" but no specific numbers. Data provenance is implied to be internal testing by the manufacturer (Terumo Corporation) based on recognized standards (ISO, ASTM, FDA guidance) and in-house standards. This is for physical device performance, not clinical data or imaging data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      Not applicable, as there is no "ground truth" related to AI or diagnostic interpretation. The "truth" here is compliance with engineering specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
      Not applicable. Adjudication methods are used in consensus building for ground truth in AI/imaging studies.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      Not applicable. This is for evaluating diagnostic performance of AI-assisted human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      Not applicable, as there is no AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
      Not applicable. The "ground truth" for this device is the established engineering and material standards and the performance of the predicate device.

    8. The sample size for the training set:
      Not applicable. There is no training set as no AI model is being developed.

    9. How the ground truth for the training set was established:
      Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150232

    Validate with FDA (Live)

    Date Cleared
    2015-10-19

    (259 days)

    Product Code
    Regulation Number
    870.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Radifocus Optitorque Angiographic Catheter is indicated for use in cardiac and vascular procedures. It is designed to deliver radiopaque media, guide wires, catheters, and therapeutic agents to selected sites in the vascular system. The different shapes are designed to selectively engage arteries from access sites such as the femoral, radial, and brachial artery.

    Device Description

    The Radifocus Optitorque Angiographic Catheter is comprised of a two-layer construction featuring stainless steel mesh sandwiched between layers of polyurethane and polyamide elastomers. The shaft inner layer and outer layer contain barium sulfate for visibility and contrast under fluoroscopy. A soft tip is attached to the distal portion of some 4 Fr and all 5 Fr and 6 Fr catheters except those with a pigtail design; a soft tube is attached to the distal portion of 5 Fr and 6 Fr catheters with a pigtail design; the 4 Fr products are available with or without a soft tip. Constructed of flexible, supple polyurethane that is permanently welded to the catheter shaft, the soft tip and soft tube are designed to minimize trauma to the vessel wall. There is no change to the design of the device as a result of this submission.

    The device is offered in lengths of 65-120 cm. French sizes and shaft inner diameters are as follows:

    French SizeShaft Inner Diameter
    41.05mm
    51.22mm
    61.32mm

    It is a disposable device intended for single use only. This device is individually packaged and sterilized by ethylene oxide gas.

    AI/ML Overview

    The provided 510(k) summary describes a medical device, the Radifocus Optitorque Angiographic Catheter, and its performance testing. However, it does not include information about AI/ML device performance, human reader studies, or ground truth establishment relevant to AI/ML. The entire document focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance and biocompatibility testing for a material change.

    Therefore, many of the requested items related to AI/ML device performance, sample sizes for test/training sets, expert qualifications, and ground truth establishment cannot be extracted from this document.

    Here's the information that can be extracted, along with explanations for the missing items:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific Test/StandardReported Device Performance/Results
    ISO StandardsRadio-detectability (ISO 10555-1:2013 Section 4.2, ASTM F640-12)"All samples tested met the standard applicable to each test."
    Surface (ISO 10555-1:2013 Section 4.4)"All samples tested met the standard applicable to each test."
    Peak tensile force (ISO 10555-1:2013 Section 4.6)"All samples tested met the standard applicable to each test."
    Freedom from leakage (ISO 10555-1:2013 Section 4.7)"All samples tested met the standard applicable to each test."
    Power injection (ISO 10555-1:2013 Section 4.10, Annex F and G)"All samples tested met the standard applicable to each test."
    Distal tip (ISO 10555-1:2013 Section 4.12)"All samples tested met the standard applicable to each test."
    Flow rate (ISO 10555-1:2013 Section 6.3)"All samples tested met the standard applicable to each test."
    Burst pressure (ISO 10555-1:2013 Section 6.3)"All samples tested met the standard applicable to each test."
    Internal StandardsShaft stiffness (Stiffness of braided-area)"Meets acceptance criteria"
    Torque transmission"Meets acceptance criteria"
    Torque strength"Meets acceptance criteria"
    Tensile strength (Shaft)"Meets acceptance criteria"
    Bending strength"Meets acceptance criteria"
    Cleanliness"Meets acceptance criteria"
    Strength of hub junction"Meets acceptance criteria"
    Tensile Strength of soft tip"Meets acceptance criteria"
    Product dimensions (I.D.; O.D.; Tip I.D.; Tip O.D.; Length)"Meets acceptance criteria"
    Embolic cube testing"Meets acceptance criteria"
    BiocompatibilityCytotoxicity (non-aged and accelerated aged (3 years))Results of the testing demonstrate that the device remains biocompatible throughout its shelf life. A full list of specific acceptance criteria is not provided, but the overall conclusion indicates meeting standards.
    Sensitization (non-aged)Results of the testing demonstrate that the device remains biocompatible throughout its shelf life.
    Intracutaneous Reactivity (non-aged)Results of the testing demonstrate that the device remains biocompatible throughout its shelf life.
    Systemic Toxicity (acute) (non-aged)Results of the testing demonstrate that the device remains biocompatible throughout its shelf life.
    Pyrogenicity (non-aged)Certified to be non-pyrogenic in the unopened and undamaged package. Performed in accordance with ISO 10993-11 (2006), ISO 10993-12 (2012), and USP 35, NF 30, 2012 <151> Pyrogen Test (USP Rabbit Test).
    Hemolysis (non-aged and accelerated aged (3 years))Results of the testing demonstrate that the device remains biocompatible throughout its shelf life.
    Thrombogenicity (non-aged)Results of the testing demonstrate that the device remains biocompatible throughout its shelf life.
    Complement Activation (Immunology) (non-aged)Results of the testing demonstrate that the device remains biocompatible throughout its shelf life.
    Chemical Extraction (non-aged and accelerated aged (3 years))Results of the testing demonstrate that the device remains biocompatible throughout its shelf life.
    SterilizationSterility Assurance Level (SAL)Validated to provide a SAL of 10^-6 per ISO 11135:2014.
    Ethylene Oxide (EO) ResidualsEO will not exceed 4 mg per device and Ethylene Chlorohydrin (ECH) will not exceed 9 mg per device, meeting requirements for limited exposure devices per ISO 10993-7.
    Risk AnalysisISO 14971 Compliance"A Product Risk Analysis was conducted... and it was determined that any new risks were adequately captured and mitigated."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This document describes non-clinical performance and biocompatibility testing for a physical medical device. It does not refer to "test sets" in the context of AI/ML models or clinical data. The testing was performed on "samples" of the device, both non-aged and accelerated aged (3 years). The exact number of samples for each test is not specified, only that "All samples tested met the standard applicable to each test."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable to the information provided. The document concerns a physical medical device and its non-clinical performance, not an AI/ML device that requires expert-established ground truth from clinical images or data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable to the information provided. The document concerns a physical medical device and its non-clinical performance, not an AI/ML device that requires adjudication of expert labels.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such study was conducted or reported in this document. This filing is for a conventional medical device, not an AI/ML diagnostic aid. The document explicitly states: "This 510(k) does not include data from clinical tests."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device described is not an algorithm or AI/ML product.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical performance tests, the "ground truth" or acceptance criteria were derived from recognized consensus standards (ISO, ASTM, USP) and the manufacturer's own internal standards for device functionality and safety. For biocompatibility, it was based on ISO 10993 standards and FDA G95-1 guidance.

    8. The sample size for the training set

    This is not applicable. There is no AI/ML model or "training set" relevant to this device.

    9. How the ground truth for the training set was established

    This is not applicable. There is no AI/ML model or "training set" relevant to this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130520

    Validate with FDA (Live)

    Date Cleared
    2013-03-13

    (13 days)

    Product Code
    Regulation Number
    870.4350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Capiox FX Hollow Fiber Oxygenator and Arterial Filter is intended to be used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery.

    The integrated arterial filter is intended to filtrate non-biologic particles and emboli and to facilitate air bubble removal from the blood flowing through the cardiopulmonary bypass circuit.

    The integrated heat exchanger is used to warm or cool blood and/or perfusion fluid as it flows through the device.

    The hardshell reservoir is used to store blood during extra-corporeal circulation from the venous line and the cardiotomy line. The reservoir contains a venous section that is comprised of a filter and defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air bubble removal. The 3-liter and 4-liter reservoirs may be used for Vacuum Assisted Drainage procedures and Post Operative Chest Drainage Procedures.

    The Capiox FX15 is for use with patients when the required blood flow rate will not exceed 5.0 L/min. when used with a 4 Liter Reservoir; and when the required blood flow rate will not exceed 4.0 L/min. when used with a 3 Liter Reservoir.

    The Capiox FX25 is for use with patients when the required blood flow rate will not exceed 7.0 L/min.

    The Capiox FX Oxygenator/Reservoir/Arterial Filter assemblies can be used in procedures lasting up to 6 hours.

    Device Description

    The modified Capiox® FX15 and FX25 Oxygenator utilizes porous fiber technology to facilitate the transfer of gases between a blood-phase environment and a gas-phase environment for the intent of satisfying the gas exchange needs of a patient during cardiopulmonary bypass surgery. A fiber bundle offers the porous membrane surface to sufficiently permit the movement of gases through the walls of the hollow fibers via diffusion.

    The modified Capiox® FX15 and FX25 device has an integrated heat exchanger that is comprised of stainless steel encased in a polycarbonate housing. The stainless steel acts as a heat transfer material that permits heat that is generated from a temperature controlled external water bath to transverse across the walls of the stainless steel to effect the necessary temperature change upon circulating blood.

    With respect to the filtration of arterial blood, the modified Capiox® FX15 and FX25 Oxygenator/Reservoir relies upon mechanical entrapment of particulates and emboli within the filter mesh as a means to remove those particulates from the blood.

    The subject of this Special 510(k) is a modification being made to the detachable Hardshell Reservoir. The design of the Hardshell Reservoir component remains identical to the design of the original reservoir that was cleared by FDA with K071494 -- except that a positive pressure relief valve will be included on the lid of the reservoir.

    The materials that are used in the construction of the Capiox® FX15 and FX25 Oxygenator/Reservoir, but are not limited to, nylon, polycarbonate, stainless steel, polyvinyl chloride, polyurethane, polyester, polypropylene, polyethylene terephthalate, polyethylene and X-Coating™. The positive pressure relief valve that is included with the modified reservoir is constructed of polycarbonate and nylon.

    AI/ML Overview

    The provided text describes a 510(k) summary for modifications to the Capiox® FX15 and FX25 Oxygenator/Reservoir. This document is a regulatory submission for medical devices, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed scientific study with specific acceptance criteria, experimental results, and statistical analyses typically found in research papers.

    Therefore, many of the requested elements (like sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, and detailed ground truth data for training sets) are not applicable or extractable from this type of regulatory document.

    Here's an attempt to answer the questions based only on the provided text, indicating where information is not available:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes "Performance Evaluations" as a demonstration of "substantial equivalence" to the predicate device, implying that the modified device should perform comparably to the original. Explicit numerical acceptance criteria are not detailed, but the evaluations are qualitative in nature to confirm the safety and effectiveness of the modification (the pressure relief valve).

    Acceptance Criteria (Implied)Reported Device Performance
    Pressure Relief Valve-to-Reservoir Interface TestingThe modification ensures the relief valve functions correctly at its interface with the reservoir.
    Relief Valve performance following application of vacuum to the reservoirThe relief valve performs as intended when vacuum is applied to the reservoir.
    Assessment of reservoir pressure during simulated bypass procedure at flow rates of 1, 2, 3, 4 and 5 liters per minuteThe relief valve effectively manages reservoir pressure during simulated bypass procedures at various flow rates (1-5 L/min), preventing excessive pressure accumulation.
    Sterilization AssessmentSterilization conditions have been validated to provide a Sterility Assurance Level (SAL) of 10^-6. Ethylene oxide residues will not exceed maximum residue limits. Biocompatibility studies conducted per ISO 10993 found blood-contacting materials to be biocompatible.
    Drop TestingNot explicitly detailed, but implied to show the device maintains integrity after drops.
    Vibration TestingNot explicitly detailed, but implied to show the device maintains integrity after vibration.
    Overall Performance EquivalenceThe modified Capiox® FX15 and FX25 device exhibits equivalent performance to the unmodified predicate device, with "no changes made to the finished device that would alter the performance of the device" beyond the addition of the relief valve.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample Size (Test Set): Not specified. The performance evaluations are described as "in-vitro," implying laboratory testing, but the number of units tested is not provided.
    • Data Provenance: The studies are described as "in-vitro performance evaluations," which are laboratory tests. The manufacturer is Terumo Corporation (Ashitaka Factory) in Fujinomiya City, Shizuoka Pref., Japan, suggesting the testing likely occurred in-house or by a contracted lab in Japan. The nature of the tests (e.g., pressure, sterilization) would be considered prospective for the specific purpose of this submission.
    • Retrospective or Prospective: Prospective, as these evaluations were conducted to support the 510(k) submission for the device modification.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • Number of Experts: Not applicable. These were in-vitro engineering and performance tests, not clinical studies requiring expert interpretation of patient data to establish "ground truth."
    • Qualifications of Experts: Not applicable.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Adjudication Method: Not applicable. This refers to consensus methods for expert interpretation of data, which is not relevant for in-vitro engineering tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No, an MRMC study was not done. This device is an oxygenator/reservoir, not an AI-powered diagnostic tool, so such a study would not be relevant.
    • Effect Size: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Standalone Performance: Not applicable. This device is not an algorithm or AI system; it's a medical hardware device. The performance evaluations assess the device's physical and functional properties, which is its "standalone" performance in a laboratory setting.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Type of Ground Truth: For the in-vitro performance evaluations, the "ground truth" would be established engineering specifications, validated test methods, and industry standards for medical device performance (e.g., pressure measurements, sterility levels, biocompatibility standards).

    8. The sample size for the training set:

    • Sample Size (Training Set): Not applicable. This is not an AI/machine learning device that requires a training set. The device's design and materials are based on established engineering principles and prior validated devices.

    9. How the ground truth for the training set was established:

    • Ground Truth for Training Set: Not applicable, as there is no training set for this type of device. The "ground truth" for the device's design and manufacturing is based on established medical device standards and the performance of its predicate device.