Search Results
Found 11 results
510(k) Data Aggregation
(193 days)
The ARK Levetiracetam II Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of levetiracetam in human serum or plasma on automated clinical chemistry analyzers. Levetiracetam concentrations can be used as an aid in management of patients treated with levetiracetam.
The ARK Levetiracetam II Assay is a homogeneous immunoassay based on competition between drug in the specimen and levetiracetam labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for binding to the antibody reagent. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly related to the drug concentration. Active enzyme converts the coenzyme nicotinamide adenine dinucleotide (NAD) to NADH that is measured spectrophotometrically as a rate of change in absorbance. Endogenous serum G6PDH does not interfere with the results because the coenzyme NAD functions only with the bacterial enzyme used in the assay.
The ARK Levetiracetam II Assay consists of reagents R1 anti-levetiracetam monoclonal antibody with substrate and R2 levetiracetam labeled with bacterial G6PDH enzyme.
The provided text describes the performance of a diagnostic assay (ARK Levetiracetam II Assay), not an AI/ML-enabled medical device. Therefore, many of the requested criteria related to AI/ML evaluation (such as MRMC studies, training set details, expert ground truth establishment for AI) are not applicable.
However, I can extract the relevant acceptance criteria and performance data for this in-vitro diagnostic device:
Device Name: ARK Levetiracetam II Assay
Regulatory Class: Class II
Product Code: ORI
Intended Use: Quantitative determination of levetiracetam in human serum or plasma on automated clinical chemistry analyzers, as an aid in management of patients treated with levetiracetam.
Here's the information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Characteristic | Acceptance Criteria (Implicit from study design/CLSI guidelines) | Reported Device Performance (ARK Levetiracetam II Assay) |
|---|---|---|
| Limit of Quantitation (LoQ) | ≤20% CV precision and ±15% recovery | 2.0 µg/mL (at 2.80% CV and 95.0% recovery) |
| Measurement Range | Not explicitly stated as acceptance criterion, but established. | 2.0 - 100.0 µg/mL |
| Recovery | ±10% of the expected sample concentration | All tested concentrations (2.0-100.0 µg/mL) showed %Recovery within ±10% (e.g., 95.0% to 102.6%) |
| Linearity | Percent difference (Deviation) ±10% between predicted and observed results | All tested concentrations (2.0-100.0 µg/mL) showed %Deviation within ±10% (e.g., -6.1% to 9.5%) |
| Precision (Total CV) | <10% Total CV | For all control and human serum samples, Total CV ranged from 1.6% to 2.9% |
| Interfering Substances | ≤10% error (relative to serum control mean result) | All tested interferents showed ≤10% error (e.g., 91.0% to 102.7% recovery) |
| Metabolites (Cross-reactivity) | ≤10% error (relative to serum control mean result) | ucb L057 showed 0.0% cross-reactivity and ≤10% interference (0.8% and 0.1% for 15 and 50 µg/mL Levetiracetam, respectively) |
| Drug Interference (Other Anti-Epileptic/Coadministered Drugs) | ≤10% error (relative to serum control mean result) | All tested drugs (except brivaracetam) showed ≤10% error. |
| Sample Stability | Not explicitly stated (implied sufficient stability for clinical use) | Stable for 48 hours at 22°C, 40 days at 2-8°C, and after 3 freeze/thaw cycles. |
| Product Stability (Shelf-life) | Not explicitly stated (implied sufficient shelf-life) | 18 months when stored unopened at 2-8°C. |
| On-Board Stability (Reagents) | Not explicitly stated (implied sufficient stability) | Stable up to 96 days. |
| Calibration Curve Stability | Not explicitly stated (implied sufficient stability) | Effective up to at least 28 days. |
2. Sample Size Used for the Test Set and Data Provenance
- LoQ: 40 replicates (8 replicates x 5 runs) for each of 3 concentrations (pooled human serum supplemented with levetiracetam).
- Recovery: 6 replicates (3 replicates x 2 analytical runs) for each concentration (human serum negative for levetiracetam, spiked with drug).
- Linearity: 6 replicates (3 replicates x 2 analytical runs) for each dilution (human serum, spiked with drug and diluted).
- Method Comparison: 104 samples (levetiracetam concentrations 3.4 ug/mL to 98.3 ug/mL). No specific provenance (e.g., country of origin) is mentioned, but the samples are clinical human samples or quality control materials. The study is a prospective analytical study comparing two assays.
- Precision: 160 replicates per sample/control level (quadruplicate twice a day for 20 non-consecutive days) for tri-level controls and three human serum samples.
- Interfering Substances: 6 replicates (3 replicates x 2 analytical runs) for each interfering substance level in two known levetiracetam concentrations (human serum).
- Metabolites/Drug Interference: Not explicitly stated, but similar to interfering substances: "high concentration of each compound was spiked into normal human serum with known levels of levetiracetam."
The data provenance is from analytical studies conducted by the manufacturer, likely in a laboratory setting, using human serum/plasma samples/materials. The document does not specify country of origin for the samples/data, beyond "human serum/plasma". These are prospective analytical studies designed to characterize the device's performance.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
This section is not applicable as the document describes an in-vitro diagnostic assay for quantitative determination of a drug. The "ground truth" for such an assay is established by the known concentrations of calibrators, controls, and spiked samples, and comparison to a legally marketed predicate device using analytical methods (e.g., spectrophotometry). There are no human "experts" establishing a "ground truth" in the sense of image interpretation or clinical diagnosis.
4. Adjudication Method for the Test Set
Not applicable. This is an in-vitro diagnostic device providing quantitative measurements. There is no qualitative assessment or interpretation by multiple readers that would require an adjudication method.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No. This is an in-vitro diagnostic assay, not an AI/ML medical device where human readers interact with AI.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, in essence. The performance studies (LoQ, Recovery, Linearity, Precision, Interference) demonstrate the "standalone" performance of the assay itself, as an automated clinical chemistry analyzer performs the measurements. There is no human-in-the-loop variable in the measurement process of an immunoassay. The output is a quantitative value, not a diagnostic interpretation that a human would then use as "assistance."
7. The Type of Ground Truth Used
The ground truth used for performance evaluation is primarily:
- Known concentrations: For LoQ, Recovery, Linearity, Interfering Substances, Metabolites, and Drug Interference studies, concentrations of levetiracetam and potential interferents are precisely measured, prepared, or spiked into matrices.
- Reference Method/Predicate Device: For Method Comparison, the predicate ARK Levetiracetam Assay performed on the Roche/Hitachi 917 serves as the comparative "reference" for evaluating the substantial equivalence of the new assay. This is a common practice for IVD assays.
8. The Sample Size for the Training Set
Not applicable. This is a reagent-based immunoassay, not an AI/ML algorithm that requires a training set in the typical sense for machine learning. The "development" or "training" of such a diagnostic involves chemical formulation, antibody development, and optimization of reaction conditions, not data-driven model training.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As explained above, there is no AI/ML training set in this context.
Ask a specific question about this device
(166 days)
The ARK Methotrexate II Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of methotrexate in human serum or plasma on automated clinical chemistry analyzers. The measurements obtained are used in monitoring levels of methotrexate to help ensure appropriate therapy. Specimens from patients who have received glucarpidase (carboxypeptidase G2) as a high dose methotrexate rescue therapy should not be tested with the ARK Methotrexate II Assay.
The ARK Methotrexate II Assay is a homogeneous immunoassay based on competition between drug in the specimen and methotrexate labeled with the recombinant enzyme glucose-6phosphate dehydrogenase (rG6PDH) for binding to the antibody reagent. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly related to the drug concentration. Active enzyme converts the coenzyme nicotinamide adenine dinucleotide (NAD) to NADH that is measured spectrophotometrically as a rate of change in absorbance. Endogenous serum G6PDH does not interfere with the results because the coenyzme NAD functions only with the bacterial enzyme (rG6PDH) used in the assay. The ARK Methotrexate II Assay consists of reagents R1 anti-methotrexate monoclonal antibody with substrate and R2 methotrexate labeled with recombinant G6PDH enzyme. The test system includes the ARK Methotrexate II Calibrator, ARK Methotrexate II Control, and ARK Methotrexate II Dilution Buffer.
The ARK Methotrexate II Assay is a homogeneous enzyme immunoassay for the quantitative determination of methotrexate in human serum or plasma.
Here's an analysis of the acceptance criteria and the study proving the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria | Device Performance (Reported) |
|---|---|
| Limit of Quantitation (LoQ) | Established as 0.030 umol/L. Acceptable inter-assay precision (<0.010 SD) and recovery (±0.010 umol/L) observed at 0.030 umol/L. |
| Measurement Range | 0.030 - 1.300 umol/L. |
| Recovery | All concentrations tested showed ±10% recovery of the expected sample concentration. |
| Linearity | A linear relationship was demonstrated between 0.030 and 1.300 umol/L with a maximum deviation of 8.73% from linearity. |
| Clinical Accuracy (vs. LC-MS/MS) | Slope of 1.03, intercept of 0.00, and Pearson correlation (r²) of 0.98. Mean bias of 0.01 with 95% Limits of Agreement from -0.11 to 0.14. |
| Method Comparison (vs. Predicate Device) | Slope = 0.98, y-intercept = -0.02, and correlation coefficient (r²) = 0.97. |
| Precision (Total CV) | ≤10% for all tested levels of controls and human serum samples. |
| Interference by Endogenous Substances | Elevated concentrations of common endogenous substances (albumin, bilirubin, cholesterol, etc.) did not interfere significantly (<±7.48% interference). |
| Analytical Specificity (7-OH-MTX) | Not substantially affected by 7-Hydroxymethotrexate (8.72% interference at 0.050 µmol/L MTX, 0.58% at 0.500 µmol/L MTX). |
| Analytical Specificity (DAMPA) | Significant interference with DAMPA (up to 57.50% cross-reactivity at 0.040 µmol/L DAMPA). Device should not be used during glucarpidase rescue therapy. |
| Analytical Specificity (Other Compounds) | Methotrexate-selective antibody did not cross-react with various potentially co-administered drugs and folate derivatives (all within ±10% interference). |
| Sample Stability | Stable for at least 14 days refrigerated (2-8 °C), 14 days at room temperature (25 °C), 15 months frozen (-20 °C), and after 3 freeze/thaw cycles. |
| Product Stability | Shelf-life stability claim of up to 18 months when stored unopened at 2-8°C. |
| On-Board Stability | Reagents stable up to 100 days on-board the instrument. |
| Calibration Curve Stability | Effective up to at least 100 days. |
2. Sample Size Used for the Test Set and Data Provenance:
- Clinical Accuracy (vs. LC-MS/MS): 90 patient samples.
- Method Comparison (vs. Predicate Device): 123 patient samples.
- LoQ, Recovery, Linearity, Precision, Interference, Analytical Specificity: These studies involved spiked human serum samples, pooled human serum, and controls, rather than a specific number of unique patient samples for the test set. For LoQ, 40 replicates were tested. For Recovery, 6 replicates per sample were tested. For Linearity, 6 replicates per sample were tested. For Precision, 160 replicates were tested for each of the 6 control levels and 6 human serum levels. For Interference and Analytical Specificity, 6 replicates per sample were tested.
- Data Provenance: "Leftover specimens were obtained from persons receiving high-dose methotrexate therapy" for the method comparison studies. The document does not specify the country of origin of the data or whether the studies were retrospective or prospective, though "leftover specimens" suggests a retrospective collection.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- The ground truth for the clinical accuracy study was established by an LC-MS/MS reference method. This is an analytical laboratory technique, not based on expert consensus. Therefore, no "experts" in the traditional sense (e.g., radiologists) were used for ground truth establishment. The performance of LC-MS/MS as a reference method typically implies its results are considered the gold standard.
4. Adjudication Method for the Test Set:
- Not applicable. The ground truth for the clinical accuracy and method comparison studies was established using objective analytical methods (LC-MS/MS and the predicate device), not through expert review requiring an adjudication method.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is an in vitro diagnostic (IVD) assay for quantitative determination of a substance in human samples. It does not involve human readers interpreting medical images or data.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, the performance data presented (Limit of Quantitation, Measurement Range, Recovery, Linearity, Clinical Accuracy, Method Comparison, Precision, Interference, Analytical Specificity) reflects the standalone performance of the ARK Methotrexate II Assay (an automated immunoassay) without human intervention in the result generation. The device itself is designed to provide quantitative measurements.
7. The Type of Ground Truth Used:
- Clinical Accuracy: Liquid Chromatography with Tandem Mass Spectrometry (LC-MS/MS) was used as the reference method (gold standard) for ground truth.
- Method Comparison: The legally marketed predicate device, ARK™ Methotrexate Assay (K111904), was used to compare results.
- Other performance metrics (LoQ, Recovery, Linearity, Precision, Interference, Analytical Specificity): Ground truth was established by preparing samples with known concentrations of methotrexate or interfering substances, typically by spiking human serum with certified reference materials.
8. The Sample Size for the Training Set:
- Not applicable. This device is an immunoassay (laboratory test kit), not a machine learning or AI algorithm that requires a "training set" in the conventional sense of computational models. The development and optimization of the assay would involve various experiments and reagent formulations, but not a distinct "training set" like for AI models.
9. How the Ground Truth for the Training Set Was Established:
- Not applicable, as there is no "training set" in the context of an AI/ML algorithm for this type of medical device. The "ground truth" for the development of the assay components would involve standard analytical chemistry practices, such as using reference materials and established analytical methods to verify the concentration and purity of substances used in the reagents.
Ask a specific question about this device
(147 days)
The ARK Hydrocodone Assay is an immunoassay intended for the qualitative detection and/or semi-quantitative estimation of hydrocodone and its metabolites in human urine at a cutoff of 300 ng/mL. The semi-quantitative mode is for the purpose of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method, such as Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS), or (2) permitting laboratories to establish quality control procedures.
The ARK Hydrocodone Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed positive analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug test result, particularly when the preliminary test result is positive.
The ARK Hydrocodone Assay is supplied as a liquid ready-to-use homogeneous enzyme immunoassay. The assay is based on competition between hydrocodone in the specimen and hydrocodone labeled with recombinant glucose-o-phosphate dehydrogenase (rG6PDH) for antibody binding sites. As the latter binds antibody, enzyme activity decreases. In the presence of hydrocodone from the specimen, enzyme activity increases and is directly related to the hydrocodone concentration. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH in the presence of glucose-6-phosphate (G6P), resulting in an absorbance change that is measured spectrophotometrically. Endogenous G6PDH does not interfere because the coenzyme NAD functions only with the bacterial enzyme used in the assay.
The ARK Hydrocodone Assay consists of reagents R1 anti-hydrocodone monoclonal rabbit antibodies with substrate and R2 hydrocodone derivative labeled with bacterial recombinant G6PDH enzyme.
The provided document is a 510(k) summary for the ARK Hydrocodone Assay, an immunoassay for the qualitative detection and/or semi-quantitative estimation of hydrocodone and its metabolites in human urine.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria & Reported Device Performance:
The document doesn't explicitly define formal "acceptance criteria" in a separate table with numerical thresholds for accuracy, precision, etc. Instead, it presents various performance characteristics that collectively demonstrate the device's suitability for its stated intended use and substantial equivalence to a predicate device. The implied acceptance criteria are that the device performs reliably and similarly to the predicate in key analytical measures.
Below is a table summarizing the reported device performance, which implicitly functions as the evidence meeting the "acceptance criteria" for analytical validation.
| Performance Characteristic | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Precision (Qualitative) | Consistent classification (Positive/Negative) at specified concentrations relative to the 300 ng/mL cutoff, especially near the cutoff. | 0 ng/mL: 160 Negative (N=160)75 ng/mL: 160 Negative (N=160)150 ng/mL: 160 Negative (N=160)225 ng/mL (-25% Cutoff): 160 Negative (N=160)300 ng/mL (Cutoff): 50 Negative / 110 Positive (N=160)375 ng/mL (+25% Cutoff): 160 Positive (N=160)450 ng/mL (+50% Cutoff): 160 Positive (N=160)525 ng/mL (+75% Cutoff): 160 Positive (N=160)600 ng/mL (+100% Cutoff): 160 Positive (N=160) |
| Precision (Semi-quantitative) | Consistent mean recovery and classification at specified concentrations relative to the 300 ng/mL cutoff. | 0 ng/mL: Mean 0, 160 Negative75 ng/mL: Mean 78, 160 Negative150 ng/mL: Mean 142, 160 Negative225 ng/mL: Mean 229, 160 Negative300 ng/mL (Cutoff): Mean 314, 24 Neg / 136 Pos375 ng/mL: Mean 388, 160 Positive450 ng/mL: Mean 459, 160 Positive525 ng/mL: Mean 539, 160 Positive600 ng/mL: Mean 620, 160 Positive |
| Analytical Recovery | Percent recovery close to 100% across the assay range. | Ranges from 86% to 103% (e.g., 80 ng/mL expected yielded 99% recovery, 720 ng/mL yielded 86% recovery, 800 ng/mL yielded 92% recovery). |
| Analytical Specificity (Cross-reactivity to Metabolites) | Cross-reactivity for hydrocodone should be high, while for metabolites it should be understood and ideally lower if they are not the primary target the assay is trying to quantify. | Hydrocodone: 103%Hydromorphone: 100%Hydromorphone-3β-Glucuronide: 0.7%Norhydrocodone: 13.2%Dihydrocodeine: <0.3% (at >100,000 ng/mL tested) |
| Analytical Specificity (Cross-reactivity to Structurally Related/Unrelated Compounds) | No significant cross-reactivity with other common opiate compounds or structurally unrelated substances should lead to false positives at specified concentrations. | All 30 tested structurally related or unrelated opiate compounds (e.g., Morphine, Codeine, Fentanyl, Oxycodone, Methadone, Buprenorphine, Heroin) showed NEGATIVE results at concentrations up to 100,000 ng/mL, with cross-reactivity generally <0.3% (or <0.6% for some at 50,000 ng/mL). |
| Interference (Endogenous Substances) | No interference (maintaining correct POS/NEG classification) from high concentrations of common endogenous substances. | All 17 tested endogenous substances (e.g., Acetaminophen, Creatinine, Glucose, Hemoglobin, Urea) showed no interference, maintaining NEG for 225 ng/mL hydrocodone and POS for 375 ng/mL hydrocodone. |
| Interference (Boric Acid) | No interference from boric acid. | Boric Acid (1% w/v) showed no interference, maintaining NEG for 225 ng/mL hydrocodone and NEG for 375 ng/mL hydrocodone (this second result seems like a typo in the table if it should be POS, but is recorded as NEG). Correction needed for interpretation of the Boric Acid result for 375 ng/mL hydrocodone. Given the other interference studies, it is likely it should have been POS, or implies an interfering effect with boric acid at the higher concentration. This would require clarification. |
| Specific Gravity Interference | No interference across a physiological range of urine specific gravity. | No interference observed from urine samples with specific gravity values ranging from 1.000 to 1.030, maintaining NEG for 225 ng/mL hydrocodone and POS for 375 ng/mL hydrocodone. |
| pH Interference | No interference across a physiological range of urine pH. | No interference observed from urine samples with pH values from 3.0 to 11.0, maintaining NEG for 225 ng/mL hydrocodone and POS for 375 ng/mL hydrocodone. |
| Method Comparison (Overall Concordance with LC-MS/MS) | High overall concordance with the confirmatory method (LC-MS/MS). | Overall concordance: 92.5% |
| Method Comparison (Qualitative Performance) | High sensitivity and specificity, especially around the cutoff. Discordant results should be explainable. | True Negatives: 134 samples (<150 ng/mL by LC-MS/MS) correctly identified as Negative. True Positives: 9 samples (300-450 ng/mL by LC-MS/MS) and 66 samples (>450 ng/mL by LC-MS/MS) correctly identified as Positive. Discordant Results: 8 samples (<150 ng/mL) and 8 samples (150-299 ng/mL) were assay Positive but LC-MS/MS Negative. 1 sample (300-450 ng/mL) was assay Negative but LC-MS/MS Positive. Discordances are explained as cross-reactivity with hydromorphone. |
2. Sample Sizes Used for the Test Set and Data Provenance:
- Precision Studies: N=160 for both qualitative and semi-quantitative precision (quadruplicate assays twice a day for 20 days for 8 concentration levels).
- Analytical Recovery: 5 replicates for each of the 11 concentration points across the assay range.
- Analytical Specificity (Cross-reactivity): Concentrations tested for various compounds (e.g., 100,000 ng/mL, 50,000 ng/mL). The exact number of individual tests/replicates is not specified but implied testing at least one sample per compound.
- Interference (Endogenous Substances, Boric Acid, Specific Gravity, pH): Not explicitly stated, but for Specific Gravity and pH, "N=3" is mentioned for each condition, suggesting triplicate testing.
- Method Comparison (Clinical Specimens): 226 unaltered clinical urine specimens.
- Data Provenance: The document states that the clinical urine specimens for method comparison were "unaltered clinical urine specimens that are not individually identifiable." It does not specify the country of origin or if they were retrospectively or prospectively collected. The analytical studies (precision, recovery, specificity, interference) appear to be laboratory-based studies using spiked drug-free human urine.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
This information is not provided in the document. The ground truth for the clinical specimens in the method comparison study was established by LC-MS/MS (Liquid Chromatography/tandem Mass Spectrometry), which is stated as the "preferred confirmatory method." LC-MS/MS is an analytical chemistry technique, not a human expert interpretation.
4. Adjudication Method for the Test Set:
This information is not applicable/provided. The ground truth was established by LC-MS/MS, an objective analytical method. Therefore, there was no need for human expert adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an immunoassay (a laboratory diagnostic test), not an imaging AI or a device that assists human readers in interpreting complex data. The performance study compares the device's analytical results to a gold standard analytical method (LC-MS/MS), not a human's performance.
6. If a Standalone (Algorithm Only Without Human-In-The-Loop Performance) was done:
The device itself is a "standalone" test in the sense that it provides a direct analytical result (qualitative or semi-quantitative) without human interpretation of raw data for diagnosis. The study presented (precision, analytical recovery, specificity, method comparison) evaluates the algorithm's (assay's) performance independent of direct human-in-the-loop diagnostic assistance. While a human operates the instrument and interprets the final result, the assay's core function is automated chemical analysis.
7. The Type of Ground Truth Used:
The primary ground truth used for the method comparison study was LC-MS/MS (Liquid Chromatography/tandem Mass Spectrometry) results. This is considered a highly specific and sensitive "confirmatory method" in analytical chemistry, often serving as a gold standard for drug quantification. For the other analytical studies (precision, recovery, specificity, interference), the ground truth was the known concentrations of spiked analytes into drug-free urine.
8. The Sample Size for the Training Set:
This information is not applicable/provided. This is an immunoassay, not a machine learning/AI model that requires a "training set" in the conventional sense. The "training" for such a device occurs during its development and optimization of the reagents and assay parameters by the manufacturer, rather than through a dataset in a machine learning paradigm.
9. How the Ground Truth for the Training Set Was Established:
As mentioned above, this concept doesn't directly apply here. The "ground truth" during the development and optimization of the assay would have been established through controlled laboratory experiments using known concentrations of hydrocodone and related substances, verified perhaps by independent analytical methods (e.g., HPLC, LC-MS/MS) and chemical purity assessments. The goal is to optimize the assay's sensitivity, specificity, and precision to meet performance targets.
Ask a specific question about this device
(469 days)
The ARK Lacosamide Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of lacosamide in human serum on automated clinical chemistry analyzers. The measurements obtained are used in monitoring levels of lacosamide to help ensure appropriate therapy.
The ARK Lacosamide Assay is a homogeneous enzyme immunoassay based on competition between drug in the specimen and lacosamide labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for binding to the antibody reagent. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly related to the drug concentration. Active enzyme converts the coenzyme nicotinamide adenine dinucleotide (NAD) to NADH that is measured spectrophotometrically as a rate of change in absorbance. Endogenous serum G6PDH does not interfere with the results because the coenzyme NAD functions only with the bacterial enzyme used in the assay.
The ARK Lacosamide Assay consists of reagents R1 anti-lacosamide polyclonal antibody with substrate and R2 lacosamide labeled with bacterial G6PDH enzyme.
The provided text describes the ARK Lacosamide Assay, a homogeneous enzyme immunoassay for quantitative determination of lacosamide in human serum. This device is intended for monitoring lacosamide levels to ensure appropriate therapy. The substantial equivalence is demonstrated through comparative testing against a predicate device (ARKTM Topiramate Assay, K083799) and various performance characteristic studies.
Here's a breakdown of the acceptance criteria and study details:
1. Table of Acceptance Criteria and Reported Device Performance:
| Performance Characteristic | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Limit of Quantitation (LoQ) | Acceptable inter-assay precision (<20% CV) and recovery (±15%) | 0.40 ug/mL, with CVs ranging from 3.4% to 5.0% and recoveries from 83.2% to 94.1% for tested concentrations. |
| Measurement Range | N/A (Range established) | 0.40 - 24.00 µg/mL |
| Recovery | Within ±10% of expected sample concentration | Recoveries for various concentrations from 0.40 to 20.00 µg/mL were between 90.4% and 105.8%. |
| Linearity | Percent difference ±10% (for >1.00 µg/mL) or ≤0.20 µg/mL (for ≤1.00 µg/mL) between 1st and 2nd order regressed values. | Linear relationship demonstrated between 0.40 and 25.00 µg/mL (y = 0.9998x - 0.0170). Differences within acceptable limits. |
| Precision (Total CV) | ≤10% total CV | For controls and human serum samples, total CVs ranged from 3.9% to 4.5%. |
| Interfering Substances | Measurement of lacosamide resulted in ≤10% error | All tested interfering substances resulted in ≤10% error (recoveries ranging from 95.8% to 103.5%). |
| Specificity (O-Desmethyl Metabolite) | Not clinically significant (< 3.0% crossreactivity) | Crossreactivity of O-desmethyl lacosamide metabolite was not clinically significant. |
| Crossreactivity (Other Drugs) | Recoveries within 10% of the expected level | Recoveries for 77 compounds ranged from 90.9% to 109.5%. |
| Sample Stability | N/A (Stability demonstrated) | Stable for at least 48 hours at room temperature, 28 days refrigerated, and 34 months frozen. Stable after 3 freeze/thaw cycles. |
| Shelf-life Stability | N/A (Stability demonstrated) | Up to 18 months when stored unopened at 2-8°C. |
| On-Board Stability | N/A (Stability demonstrated) | Up to 60 days on-board the instrument. |
| Calibration Curve Stability | N/A (Stability demonstrated) | Effective up to at least 14 days. |
| Method Comparison (vs. LC-MS/MS) | N/A (Correlation established) | Slope: 1.01 (0.99 to 1.04), y-intercept: 0.03 (-0.10 to 0.15), r2: 0.98 (0.98 to 0.99) |
2. Sample size used for the test set and the data provenance:
- LoQ: 40 replicates (8 replicates x 5 runs) for each of 3 concentrations. Data provenance is implied to be laboratory-generated (pooled human serum supplemented with lacosamide).
- Measurement Range: Not a directly tested "test set" in terms of patient samples; rather, it refers to the range characterized by other studies like recovery and linearity.
- Recovery: Not explicitly stated as a separate "test set" size for recovery studies, but involved adding concentrated lacosamide to pooled human serum. "Two analytical runs of three replicates of each sample were assayed."
- Linearity: Dilutions of a 30.00 µg/mL lacosamide serum sample. "Two analytical runs of three replicates of each sample were assayed."
- Method Comparison: 150 unaltered, human serum specimens. The data provenance is described as "human serum specimens" which implies retrospective clinical samples, but the country of origin is not specified. They are "not individually identifiable."
- Precision: 160 replicates for each of 6 samples/controls (quadruplicate twice a day for 20 days). Implied laboratory-generated (tri-level controls and pooled human serum with lacosamide).
- Interfering Substances: For each interfering substance, lacosamide was present at 2 concentrations (2.0 and 15.0 µg/mL). "Two analytical runs of three replicates of each sample (6 replicates total) were assayed." Implied laboratory-generated.
- Specificity (O-Desmethyl Metabolite): Assayed lacosamide at 2 concentrations (2.00 and 15.00 µg/mL) in the absence and presence of the metabolite at 2 concentrations (5.0 and 30.0 µg/mL). Implied laboratory-generated.
- Crossreactivity (Other Drugs): For each of 77 compounds, lacosamide was present at 2 concentrations (2.00 and 15.00 µg/mL). Implied laboratory-generated.
- Sample Stability: Not a statistical "test set" but based on supporting data; implies samples stored under different conditions.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This device is an in vitro diagnostic assay that measures a chemical compound (lacosamide concentration). The "ground truth" for method comparison studies is established by a reference method, not by expert interpretation.
- In the method comparison study, the ARK Lacosamide Assay results were compared against LC-MS/MS (Liquid Chromatography-Mass Spectrometry/Mass Spectrometry). LC-MS/MS is a highly accurate and sensitive analytical chemistry technique, often considered a "gold standard" for quantifying small molecules like drugs in biological matrices.
- Therefore, no human experts were used to establish the ground truth; instead, an established analytical method served as the reference.
4. Adjudication method for the test set:
- Not applicable as the ground truth is established by a quantitative analytical method (LC-MS/MS), not by human interpretation or consensus.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is an in vitro diagnostic assay, not an AI-powered diagnostic imaging or interpretation tool that assists human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, the performance characteristics described (LoQ, Measurement Range, Recovery, Linearity, Precision, Interfering Substances, Specificity, Crossreactivity, Stability) are all standalone algorithm (assay) performance studies. The method comparison study also evaluated the standalone performance of the ARK Lacosamide Assay against LC-MS/MS.
- The device explicitly states it is "intended for the quantitative determination of lacosamide in human serum on automated clinical chemistry analyzers," indicating standalone operation with human oversight rather than human-in-the-loop interpretation of complex data.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the method comparison study, the ground truth was "results from LC-MS/MS," which is a reference analytical method.
- For other performance studies (LoQ, recovery, linearity, precision, interference, specificity, cross-reactivity), the ground truth was based on the known concentrations of lacosamide and interfering substances added to human serum samples.
8. The sample size for the training set:
- This is an in vitro diagnostic assay, not a machine learning or AI model that requires a "training set" in the conventional sense. The development of the assay's reagents and methodologies involves analytical chemistry and biochemical principles rather than statistical training on data.
9. How the ground truth for the training set was established:
- Not applicable for the reasons stated in point 8. The "ground truth" for the development of this assay would be the accurate chemical characterization of lacosamide and its antibodies, and the optimization of the enzymatic reaction to accurately reflect lacosamide concentration.
Ask a specific question about this device
(30 days)
The ARK Fentanyl II Assay is an immunoassay intended for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use in laboratories with automated clinical chemistry analyzers. This in vitro diagnostic device is for prescription use only.
The ARK Fentanyl II Assay provides only a preliminary analytical result. A more specific alternative chemical method must be used in order to obtain a confirmed positive analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug test result, particularly when the preliminary test result is positive.
The ARK Fentanyl II Assay is a homogeneous enzyme immunoassay technique used for the analysis of a specific compound in human urine. The assay is based on competition between drug in the specimen and drug labeled with recombinant glucose-6-phosphate dehydrogenase (rG6PDH) for antibody binding sites. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly related to the drug concentration. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH in the presence of glucose-6-phosphate (G6P), resulting in an absorbance change that is measured spectrophotometrically. Endogenous serum G6PDH does not interfere because the coenzyme NAD functions only with the bacterial enzyme used in the assay.
The ARK Fentanyl II Assay consists of reagents R1 anti-fentanyl monoclonal antibodies with substrate and R2 fentanyl derivative labeled with bacterial recombinant G6PDH enzyme.
1. Acceptance Criteria and Reported Device Performance for ARKTM Fentanyl II Assay
The provided text details the performance characteristics of the ARKTM Fentanyl II Assay, which is an immunoassay for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The study evaluates precision, analytical specificity (cross-reactivity with structurally similar compounds and lack of interference from structurally unrelated compounds and endogenous substances), and method comparison with LC-MS/MS.
Here's a table summarizing the acceptance criteria (implied by the data presentation for each test) and the reported device performance:
| Performance Characteristic | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Precision (Qualitative) | Samples below the cutoff (e.g., -100%, -75%, -50%, -25% of cutoff) should consistently test Negative. | - 0.00 ng/mL (-100%): 160 Negative (out of 160) - 0.25 ng/mL (-75%): 160 Negative (out of 160) - 0.50 ng/mL (-50%): 160 Negative (out of 160) - 0.75 ng/mL (-25%): 160 Negative (out of 160) - 1.00 ng/mL (Cutoff): 84 Negative; 76 Positive (out of 160) - 1.25 ng/mL (+25%): 160 Positive (out of 160) - 1.50 ng/mL (+50%): 160 Positive (out of 160) - 1.75 ng/mL (+75%): 160 Positive (out of 160) - 2.00 ng/mL (+100%): 160 Positive (out of 160) |
| Samples above the cutoff (e.g., +25%, +50%, +75%, +100% of cutoff) should consistently test Positive. | ||
| Samples at the cutoff should show a mix of Negative and Positive results, indicating proper cutoff discrimination. | ||
| Analytical Specificity | Norfentanyl (major metabolite) should show some cross-reactivity. Other fentanyl structural analogs/metabolites may show varying degrees of cross-reactivity. Structurally similar opioids and functional analogs should be negative at high concentrations. | - Norfentanyl: 7% cross-reactivity at 15 ng/mL. - Acetyl fentanyl, Isobutyryl fentanyl: 90.91% cross-reactivity at 1.1 ng/mL. - Furanyl fentanyl, Para-fluoro fentanyl: 66.67% cross-reactivity at 1.5 ng/mL. - Other similar compounds (e.g., Carfentanil, Sufentanil, Alfentanil): show very low to <0.001% cross-reactivity at high concentrations. - Opioids/structurally similar compounds (e.g., Morphine, Codeine, Hydrocodone): All tested negative at 100 µg/mL. |
| Interference | Structurally unrelated compounds (e.g., common medications, illicit drugs metabolites) and endogenous substances (e.g., creatinine, albumin, pH, specific gravity) should not produce false positive or false negative results when fentanyl is present/absent at concentrations near the cutoff. Boric acid is specified as a potential interferent and should be acknowledged. | - Structurally Unrelated Compounds: No false results observed for 30+ compounds (e.g., Acetaminophen, Amphetamine, Ibuprofen) at high concentrations (up to 1000 µg/mL) when tested with 0.5 ng/mL and 1.5 ng/mL fentanyl. - Endogenous Substances: No interference observed for 14 endogenous substances (e.g., Acetone, Bilirubin, Glucose) at high concentrations (up to 4000 mg/dL) when tested with 0.5 ng/mL and 1.5 ng/mL fentanyl. - Specific Gravity & pH: No interference observed for specific gravity (1.002-1.030 g/mL) and pH (3.0-11.0) when tested with 0.5 ng/mL and 1.5 ng/mL fentanyl. - Boric Acid: Tested at 1% w/v, indicated negative result for both 0.5 ng/mL and 1.5 ng/mL fentanyl spiked samples. However, labeling includes a limitation: "Do not use Boric Acid as a preservative." |
| Method Comparison | High agreement (concordance) with a confirmatory method (LC-MS/MS) for both positive and negative samples, particularly for samples significantly above and below the cutoff. Limited discordant results are expected near the cutoff and should be explainable (e.g., cross-reactivity). | - Total Samples: 147 clinical urine specimens. - Concordance: - LC-MS/MS < 0.5 ng/mL: 50 Negative, 1 Positive (1 discordant) - LC-MS/MS 0.5-0.9 ng/mL: 2 Negative, 21 Positive (2 discordant negatives, 21 concordant positives) - LC-MS/MS 1.0-1.5 ng/mL: 0 Negative, 11 Positive (11 concordant positives) - LC-MS/MS > 1.5 ng/mL: 0 Negative, 62 Positive (62 concordant positives) - Discordant Results Explanation: The 1 discordant positive at < 0.5 ng/mL (LC-MS/MS) and 21 discordant positives between 0.5-0.9 ng/mL (LC-MS/MS) were analyzed. All these discordant immunoassay positives were found to have detectable Norfentanyl (main metabolite) by LC-MS/MS, explaining the immunoassay's positive result due to cross-reactivity. The 2 discordant negatives near cutoff (0.5-0.9 ng/mL) are not further explained but are within expected variability for qualitative assays near cutoff. Overall good agreement. |
Study Details:
-
Sample size used for the test set and the data provenance:
- Precision Study: N=160 for each of 9 concentration levels (total 1440 measurements). The samples were drug-free human urine supplemented with fentanyl. Data provenance is not explicitly stated as country of origin, but described as "drug-free, negative human urine", implying laboratory-prepared samples.
- Analytical Specificity (Cross-reactivity/Interference): Samples were drug-free, negative human urine spiked with various compounds and fentanyl. Specific sample sizes for each compound tested are not provided, but generally involve multiple replicates for dose-response evaluation or testing with fentanyl-spiked urine.
- Method Comparison: A total of 147 unaltered clinical urine specimens were used. The data provenance is not explicitly mentioned as country of origin, but it is stated these were "unaltered clinical urine specimens that are not individually identifiable." This suggests a retrospective collection of clinical samples.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The ground truth for the method comparison study was established by LC-MS/MS (Liquid Chromatography/tandem Mass Spectrometry), which is described as the "preferred confirmatory method" and was performed by a "licensed reference laboratory." This is an objective, analytical method, not reliant on human expert interpretation for the quantitative values. Therefore, individual human "experts" in the traditional sense (e.g., radiologists) were not used to establish the ground truth in this context. The expertise lies in the certified analytical process and the laboratory standards.
-
Adjudication method for the test set:
- No adjudication method (e.g., 2+1, 3+1) was applicable or mentioned, as the ground truth was established by quantitative analytical methods (LC-MS/MS), not by human expert review requiring consensus.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This is an in vitro diagnostic immunoassay and not an AI-powered image analysis or medical device that involves human readers/interpreters in its primary function. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance was not conducted and is irrelevant to this device type. The device provides a direct analytical result.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, the performance presented is a standalone performance of the immunoassay system. It measures the algorithm's (immunoassay's) ability to detect fentanyl in urine samples independently of human intervention for result interpretation. The results are generated by an "automated clinical chemistry analyzer."
-
The type of ground truth used:
- For the method comparison study, the ground truth was quantitative analytical data from LC-MS/MS (Liquid Chromatography/tandem Mass Spectrometry). This is a highly specific and sensitive confirmatory laboratory method.
-
The sample size for the training set:
- This document describes a premarket notification (510(k)) for an in vitro diagnostic device (immunoassay). For such devices, internal development and optimization (which can be considered analogous to "training") occur during the assay design and reagent formulation phases. However, the FDA 510(k) submission typically focuses on the validation (test set) of the finalized device rather than detailed disclosure of the "training set" in the context of machine learning. The document does not specify a sample size for a training set in the way it would for a machine learning algorithm. The "training" here would involve iterative testing and refinement of antibody characteristics, enzyme conjugates, and reagent formulations using proprietary in-house samples until the desired performance (specificity, sensitivity, dynamic range) is achieved.
-
How the ground truth for the training set was established:
- As mentioned above, the concept of a "training set" with established ground truth in the machine learning sense is not directly applicable to this traditional immunoassay. The development process would involve:
- Synthesizing or acquiring known concentrations of fentanyl and its metabolites.
- Evaluating various antibody candidates and enzyme conjugates against these known concentrations.
- Spiking drug-free human urine to create samples with precisely known concentrations of analytes.
- Using highly accurate analytical methods (like GC/MS or LC-MS/MS) internally to confirm the concentrations of prepared samples or to guide dose-response adjustments during reagent formulation.
- Ultimately, the ground truth for any "training" or optimization would be based on known chemical concentrations of the target analyte and potential interferents, confirmed by reference analytical methods.
- As mentioned above, the concept of a "training set" with established ground truth in the machine learning sense is not directly applicable to this traditional immunoassay. The development process would involve:
Ask a specific question about this device
(110 days)
The ARK Tramadol Assay is an immunoassay intended for the qualitative and/or semiquantitative determination of tramadol in human urine at a cutoff concentration of 100 ng/mL. The assay is intended for use in laboratories with automated clinical chemistry analyzers. This in vitro diagnostic device is for prescription use only.
The semiquantitative mode is for the purpose of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method, such as Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS), or (2) permitting laboratories to establish quality control procedures.
The ARK Tramadol Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed positive analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug test result, particularly when the preliminary test result is positive.
The ARK Tramadol Assay is a homogeneous enzyme immunoassay technique used for the analysis of tramadol in human urine. The assay is based on competition between drug in the specimen and drug labeled with recombinant glucose-6-phosphate dehydrogenase (rG6PDH) for antibody binding sites. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly related to the drug concentration. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH in the presence of glucose-6-phosphate (G6P), resulting in an absorbance change that is measured spectrophotometrically. Endogenous G6PDH does not interfere because the coenzyme NAD functions only with the bacterial enzyme used in the assay.
The ARK Tramadol Assay consists of reagents R1 anti-tramadol rabbit polyclonal antibody with substrate and R2 tramadol derivative labeled with bacterial recombinant G6PDH enzyme.
The provided text describes the performance characteristics of the ARK Tramadol Assay, an immunoassay for detecting tramadol in human urine. Here's a breakdown of the acceptance criteria and study details:
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't explicitly state "acceptance criteria" in a bulleted or numbered list with corresponding "reported performance." However, the performance parameters evaluated and the results presented implicitly define the acceptance criteria for the device to be considered substantially equivalent to the predicate. Based on the provided data, the implicit acceptance criteria and the device's performance are:
| Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|
| Precision (Qualitative): Reliable classification of samples as negative below -25% cutoff, positive above +25% cutoff, and expected indeterminate results around the cutoff. | At 0.0, 25.0, 50.0, and 75.0 ng/mL (below -25% cutoff), all 160 determinations were Negative. At 125.0, 150.0, 175.0, and 200.0 ng/mL (above +25% cutoff), all 160 determinations were Positive. At the 100 ng/mL Cutoff, results were 83 Negative/77 Positive (demonstrating expected variability around the cutoff). |
| Precision (Semiquantitative): Accurate mean concentration values for different spiked levels and reliable classification (negative below -25% cutoff, positive above +25% cutoff). | Mean concentrations closely matched nominal values (e.g., 29.2 ng/mL for 25.0 ng/mL samples, 189.0 ng/mL for 200.0 ng/mL samples). At 0.0, 25.0, 50.0, and 75.0 ng/mL, all 160 determinations were Negative. At 125.0, 150.0, 175.0, and 200.0 ng/mL, all 160 determinations were Positive. At the 100 ng/mL Cutoff, results were 97 Negative/63 Positive (demonstrating expected variability around the cutoff). |
| Analytical Recovery: Percentage recovery for various tramadol concentrations should be within an acceptable range (e.g., likely 80-120% or 90-110%, though not explicitly stated). | Recovery percentages ranged from 90.4% to 109.1% across concentrations from 50.0 to 1000.0 ng/mL, indicating good analytical recovery. |
| Analytical Specificity (Tramadol Metabolites): Known metabolites should exhibit a specific level of cross-reactivity, allowing detection without excessive interference from non-target compounds at the cutoff. | O-Desmethyltramadol: 16.67% cross-reactivity (positive at 600 ng/mL for 100 ng/mL cutoff). N-Desmethyltramadol: 66.67% cross-reactivity (positive at 150 ng/mL for 100 ng/mL cutoff). This demonstrates the assay's ability to react with key metabolites. |
| Analytical Specificity (Structurally-Related Compounds): Structurally related compounds (not metabolites) should not cause false positives at relevant concentrations, demonstrating assay specificity. | All listed structurally related compounds (e.g., 6-Acetyl Morphine, Amphetamine, Morphine, Fentanyl, etc.) were negative at the high concentrations tested (e.g., 10-500 µg/mL), confirming minimal to no cross-reactivity and high specificity for tramadol. |
| Interference (Structurally Unrelated Compounds): High concentrations of common medications/substances should not cause false positive or false negative results with tramadol present at ±25% of the cutoff. | Over 100 structurally unrelated compounds (e.g., Acetaminophen, Alprazolam, Caffeine, Cocaine, Ibuprofen, etc.) at very high concentrations (up to 500,000 ng/mL) did not yield false results (i.e., remained Negative at 75 ng/mL tramadol and Positive at 125 ng/mL tramadol). This indicates robust resistance to interference. |
| Interference (Endogenous Substances): High concentrations of common endogenous urine components should not cause false positive or false negative results. | All listed endogenous substances (e.g., Acetone, Bilirubin, Creatinine, Glucose, Hemoglobin, Urea, etc.) at physiological or high concentrations did not cause interference, maintaining expected negative or positive results at ±25% of the cutoff. |
| Interference (Specific Gravity and pH): Assay performance should be stable across a wide range of specific gravity and pH values expected in human urine. | No interference was observed across specific gravity values from 1.000 to 1.030 and pH values from 3.0 to 11.0, demonstrating robustness to variations in urine characteristics. |
| Interference (Boric Acid): The assay should either not be interfered with by common preservatives, or any interference should be clearly identified and communicated. | 1% w/v Boric Acid did interfere by causing a negative result at 125 ng/mL in the qualitative mode (expected positive). This is explicitly noted, and a warning is provided: "Boric acid interferes with results from this device. Do not test samples that have boric acid as a preservative." This demonstrates that the interference was identified and addressed with a clear instruction. |
| Method Comparison with Confirmatory Method (LC-MS/MS): A high degree of concordance with the gold standard confirmatory method (LC-MS/MS) for samples far from the cutoff, and clear explanation of discordant results near the cutoff, demonstrating clinical utility. | Concordant Results: - 50 samples with LC-MS/MS < 50 ng/mL were Negative. - 56 samples with LC-MS/MS > 150 ng/mL were Positive. - 4 samples with LC-MS/MS 100-150 ng/mL were Positive. Discordant Results: - 5 samples with LC-MS/MS 50-99 ng/mL were shown as Positive by the ARK Immunoassay but were Negative by LC-MS/MS according to the cutoff. The cause of discordance for these 5 samples was identified as the presence of O-desmethyltramadol, which cross-reacts with the assay. This provides a clear explanation for expected discrepancies. |
| Calibration Curve Stability: Calibration should remain stable for a reasonable period. | A stored calibration curve was effective for at least 30 days. |
2. Sample Size Used for the Test Set and the Data Provenance
- Precision Studies: 160 determinations for each of 9 concentration levels (0.0 to 200.0 ng/mL) for both qualitative and semiquantitative modes. This totals 160 * 9 = 1440 individual measurements across 20 days. The samples were
- Data Provenance: Drug-free, negative human urine that was spiked with tramadol to create the test concentrations. The geographic origin of the human urine or whether it was retrospective/prospective is not specified beyond "human urine."
- Analytical Recovery: Not explicitly stated, but "N=6" is mentioned for the mean concentration calculation at each of the 11 tested levels. So, 6 samples for each of 11 concentrations (50.0 to 1000.0 ng/mL) were used. These were also dilutions from drug-free, negative human urine spiked with tramadol.
- Analytical Specificity (Metabolites and Structurally Related Compounds): Not explicitly stated, but compounds were "spiked into drug-free, negative human urine." The number of samples per compound is not specified, but each compound was tested.
- Interference (Structurally Unrelated Compounds & Endogenous Substances): Not explicitly stated for each concentration or substance, but involved "High concentrations of the following structurally unrelated compounds/endogenous substances were added into urine spiked with tramadol (± 25% of the cutoff concentration) and tested."
- Interference (Specific Gravity, pH, Boric Acid): Samples were spiked urine (± 25% of cutoff) manipulated for specific gravity/pH or with boric acid added. The number of samples is not explicitly stated per condition.
- Method Comparison: A total of 115 unaltered clinical human urine specimens.
- Data Provenance: "clinical human urine specimens that are not individually identifiable." No country of origin is specified, but "unaltered clinical human urine specimens" suggests real-world samples. Whether retrospective or prospective is not explicitly stated.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
- This is an in vitro diagnostic (IVD) device for chemical analysis, not an imaging or clinical interpretation device. Therefore, the concept of "experts" establishing ground truth in the way a radiologist would for an image is not applicable.
- For the method comparison study, the ground truth was established by a "licensed reference laboratory" using Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS), which are considered the gold standard confirmatory methods for drug testing. The qualifications of the personnel performing these confirmatory tests are inherent to the "licensed reference laboratory" performing a gold-standard chemical analysis, rather than subjective expert opinion.
4. Adjudication Method for the Test Set
- Not applicable as this is a chemical assay with objective measurements (spectrophotometric absorbance changes) and comparison to a definitive analytical method (LC-MS/MS) for ground truth. Discrepancies in the method comparison were analytically explained (e.g., cross-reactivity with O-desmethyltramadol) rather than requiring human adjudication of subjective interpretations.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study design is typically used for medical imaging devices or other diagnostic tools where human interpretation plays a significant role. The ARK Tramadol Assay is an automated immunoassay for chemical analysis, not a device requiring human "readers" in the context of an MRMC study.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
- Yes, this entire study represents a standalone performance evaluation. The ARK Tramadol Assay is an automated immunoassay performed on "automated clinical chemistry analyzers." Its performance is directly compared to an established analytical method (LC-MS/MS), without "human-in-the-loop" interpretation as a primary output of the device itself. The results are quantitative (in semiquantitative mode) or qualitative (positive/negative), derived directly from the instrument's measurements.
7. The Type of Ground Truth Used
- The ground truth used was confirmatory analytical methods, specifically Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS), which are considered the gold standard for drug detection and quantification in biological samples.
8. The Sample Size for the Training Set
- The document does not provide information about a "training set" in the context of machine learning or AI. This device is an immunoassay, whose performance is based on established biochemical principles and reagents, not a machine learning algorithm that is "trained" on data. The studies described are method development and validation studies to establish the assay's analytical performance.
9. How the Ground Truth for the Training Set Was Established
- As noted in point 8, there is no "training set" in the context of this immunoassay. The concept of "ground truth for a training set" is therefore not applicable. The assay's design and reagents are based on antibody-antigen binding principles rather than data-driven machine learning.
Ask a specific question about this device
(51 days)
The ARK EDDP Assay is an immunoassay intended for the qualitative and/or semiquantitative determination of EDDP in human urine at cutoff concentrations of 100 ng/mL and 300 ng/mL. The assay is intended for use in laboratories with automated clinical chemistry analyzers. This in vitro diagnostic device is for prescription use only.
The semiquantitative mode is for the purpose of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method, such as Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS), or (2) permitting laboratories to establish quality control procedures.
The ARK EDDP Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed positive analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug test result, particularly when the preliminary test result is positive.
The ARK EDDP Assay is a homogeneous enzyme immunoassay technique used for the analysis of EDDP in human urine. The assay is based on competition between EDDP in the specimen and EDDP labeled with recombinant glucose-6-phosphate dehydrogenase (rG6PDH) for antibody binding sites. As the latter binds antibody, enzyme activity decreases. In the presence of EDDP from the specimen, enzyme activity increases and is directly related to the EDDP concentration. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH in the presence of glucose-6-phosphate (G6P), resulting in an absorbance change that is measured spectrophotometrically. Endogenous G6PDH does not interfere because the coenzyme NAD functions only with the bacterial enzyme used in the assay.
The ARK EDDP Assay consists of reagents R1 anti-EDDP rabbit antibody with substrate and R2 EDDP derivative labeled with bacterial recombinant G6PDH enzyme.
Here's a breakdown of the acceptance criteria and study details for the ARK EDDP Assay, based on the provided FDA 510(k) summary:
This document describes a medical device, the ARK EDDP Assay, which is an immunoassay for the qualitative and/or semi-quantitative determination of EDDP (a methadone metabolite) in human urine. The study presented is a non-clinical performance evaluation, as indicated by "Brief Description of Nonclinical and Clinical Data" and the nature of the tests (Precision, Analytical Recovery, Analytical Specificity, Interference, etc.). No clinical study involving human patients or human readers in an MRMC setting is described.
1. A table of acceptance criteria and the reported device performance
The document doesn't explicitly state "acceptance criteria" as a single, consolidated table with specific pass/fail values for each performance characteristic. Instead, performance is evaluated against expected behaviors for an in vitro diagnostic (IVD) device of this type. The inferred acceptance criteria are that the device should demonstrate acceptable precision, accurate analytical recovery, appropriate cross-reactivity and interference profiles, and good agreement with a gold standard (GC/MS) in method comparison for qualitative and semi-quantitative results.
| Performance Characteristic | Inferred Acceptance Criteria (Implicit) | Reported Device Performance and Discussion |
|---|---|---|
| Precision | Consistent qualitative and semi-quantitative results, particularly around the cutoff concentrations. | 100 ng/mL Cutoff: |
| - Qualitative (100 ng/mL) | All samples ≤ 75 ng/mL should be negative; all samples ≥ 125 ng/mL should be positive. Samples at 100 ng/mL (cutoff) may show mixed results. | At 0.0, 25.0, 50.0, and 75.0 ng/mL, all 160 results were Negative. At 125.0, 150.0, 175.0, and 200.0 ng/mL, all 160 results were Positive. At the 100.0 ng/mL cutoff, 123 were Negative and 37 were Positive, which is expected for a cutoff concentration. |
| - Semiquantitative (100 ng/mL) | Mean measured concentrations should be close to theoretical values. Expected mixed qualitative results at cutoff. | Mean concentrations were close to theoretical (e.g., 98.1 ng/mL for 100 ng/mL theoretical). At 100.0 ng/mL cutoff, 114 were Negative and 46 were Positive, consistent with a cutoff. |
| - Qualitative (300 ng/mL) | All samples ≤ 225 ng/mL should be negative; all samples ≥ 375 ng/mL should be positive. Samples at 300 ng/mL (cutoff) may show mixed results. | At 0.0, 75.0, 150.0, and 225.0 ng/mL, all 160 results were Negative. At 375.0, 450.0, 525.0, and 600.0 ng/mL, all 160 results were Positive. At the 300.0 ng/mL cutoff, 57 were Negative and 103 were Positive, which is expected. |
| - Semiquantitative (300 ng/mL) | Mean measured concentrations should be close to theoretical values. Expected mixed qualitative results at cutoff. | Mean concentrations were close to theoretical (e.g., 298.8 ng/mL for 300 ng/mL theoretical). At 300.0 ng/mL cutoff, 85 were Negative and 75 were Positive, consistent with a cutoff. |
| Analytical Recovery | Measured concentration should be within an acceptable percentage range of the theoretical concentration. | Recovery percentages ranged from 94.6% to 107.9% across EDDP concentrations from 50.0 to 1000.0 ng/mL, indicating good analytical recovery. |
| Analytical Specificity | Minimal to no cross-reactivity with structurally related and unrelated compounds, ensuring specificity for EDDP. | Structurally Related Compounds: EDDP showed 100% cross-reactivity (as expected). Methadone and EMDP showed very low cross-reactivity (<0.005% to <0.03%). Several other structurally related compounds (Chlorpromazine, Diphenhydramine, Methylphenidate, Doxylamine) showed low cross-reactivity (<0.1% to <0.3%). This indicates good specificity for EDDP while acknowledging minor cross-reactivity with very high concentrations of certain related compounds. |
| Interference | Endogenous substances, specific gravity variations, and pH variations should not cause false positive or false negative results. | Structurally Unrelated Compounds: Numerous compounds were tested at high concentrations (e.g., 100,000 ng/mL, 500,000 ng/mL). For both 100 ng/mL and 300 ng/mL cutoffs, none of the listed substances yielded a false result (i.e., they were negative at -25% cutoff EDDP concentration and positive at +25% cutoff EDDP concentration), indicating no significant interference from these compounds even at very high concentrations. |
| Endogenous Substances: High concentrations of common endogenous substances (e.g., Acetone, Ascorbic Acid, Bilirubin, Creatinine, Ethanol, Glucose, Hemoglobin, etc.) were tested. No interference was observed for either the 100 ng/mL or 300 ng/mL cutoff, meaning they did not cause false results. | ||
| Specific Gravity: Urine samples across the physiologic range (1.002 to 1.030) did not show interference for either cutoff, meaning results remained correct (negative at -25% cutoff EDDP, positive at +25% cutoff EDDP). | ||
| pH: Urine samples across the physiologic pH range (3.0 to 11.0) did not show interference for either cutoff, meaning results remained correct (negative at -25% cutoff EDDP, positive at +25% cutoff EDDP). | ||
| Method Comparison | High concordance with the GC/MS gold standard, particularly for samples well above or below the cutoff. Minor discrepancies near the cutoff are acceptable. | 100 ng/mL Cutoff: 104 out of 109 samples showed perfect agreement with GC/MS (40 Low Negative, 60 High Positive). 5 samples were Near Cutoff Negative by GC/MS but Negative by immunoassay (expected). 4 samples were Near Cutoff Positive by GC/MS but Positive by immunoassay (expected). This shows high concordance. |
| 300 ng/mL Cutoff: 104 out of 109 samples showed perfect agreement with GC/MS (49 Low Negative, 52 High Positive). 4 samples were Near Cutoff Negative by GC/MS but Negative by immunoassay (expected). 3 samples were Near Cutoff Positive by GC/MS but Positive by immunoassay (expected). One discordant result: Sample 51 was Immunoassay Positive but 294 ng/mL by GC/MS (which is slightly below the 300 ng/mL cutoff). This is a single false positive near the cutoff, which can occur for immunoassay methods. | ||
| Calibration Curve Stability | Calibration should remain stable for a reasonable period. | A stored calibration curve was effective up to at least 15 days, which is an acceptable stability period for a laboratory assay. |
2. Sample sizes used for the test set and the data provenance
- Precision Studies:
- Target Concentrations: 9 levels for each cutoff (0.0 to 200.0 ng/mL for 100 ng/mL cutoff; 0.0 to 600.0 ng/mL for 300 ng/mL cutoff).
- Replicates: Quaduplicate twice a day for 20 days.
- Total Sample Size per Cutoff (for precision): 9 levels * 4 replicates * 2 (times/day) * 20 days = 1440 individual readings for the qualitative precision tables. (Note: The table summary states N=160 for each concentration level, implying 160 results for each of the 9 spiked levels.)
- Data Provenance: Drug-free, negative human urine (spiked with EDDP). The country of origin is not specified but is likely the U.S. as it's an FDA submission. This is a prospective study (laboratory testing).
- Analytical Recovery:
- Target Concentrations: 12 levels (50.0 to 1000.0 ng/mL).
- Replicates: N=6 for mean concentration calculation.
- Total Sample Size: 12 levels * 6 replicates = 72 individual readings.
- Data Provenance: Drug-free, negative human urine (spiked with EDDP). Prospective laboratory testing.
- Analytical Specificity (Cross-reactivity):
- Number of Compounds Tested: 7 structurally related compounds.
- Sample Size: Each compound tested at minimum concentration for positive result (or highest tested) at both cutoff levels. Implies multiple tests per compound, but specific N is not given per compound; typically, this would be triplicate.
- Data Provenance: Drug-free, negative human urine (spiked with compounds). Prospective laboratory testing.
- Interference (Structurally Unrelated, Endogenous Substances, Specific Gravity, pH):
- Number of Compounds/Conditions Tested: Numerous (tens of compounds + SG and pH conditions).
- Sample Size: Each compound/condition tested at two EDDP levels (±25% of cutoff). Implies multiple tests per condition, but specific N is not given per condition; typically, this would be triplicate.
- Data Provenance: Drug-free, negative human urine (spiked with EDDP and interferent). Prospective laboratory testing.
- Method Comparison:
- Sample Size: 109 unaltered clinical human urine specimens.
- Data Provenance: "unaltered clinical human urine specimens that are not individually identifiable." Country of origin not specified, but usually implies local (US) clinical samples. Described as "clinical human urine specimens," suggesting these were collected from actual patients, making this a retrospective evaluation of real-world samples.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The ground truth for this diagnostic device is established by Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS), which are highly accurate analytical chemistry techniques considered the gold standard for drug confirmation in urine. These methods are performed in laboratories by trained laboratory professionals (analytical chemists, toxicologists). The document states the GC/MS confirmatory method was performed by a "licensed reference laboratory."
- Number of experts: Not explicitly stated, as GC/MS analysis is a scientific measurement, not typically an "expert reader" interpretation process like image analysis. The "experts" are the qualified laboratory personnel operating and interpreting the GC/MS results.
- Qualifications of experts: Implied to be qualified laboratory professionals with expertise in analytical chemistry and toxicology, as they work in a "licensed reference laboratory." Specific certifications or experience years are not given.
4. Adjudication method for the test set
Not applicable. The ground truth is established by a definitive analytical method (GC/MS). There is no "adjudication" in the sense of reconciling disagreements among multiple human readers for an image-based or qualitative assessment. The GC/MS provides a quantitative concentration, which is then used to determine positive/negative relative to the cutoffs. Any "discordance" (like the one noted in the method comparison for Sample 51) is a direct comparison between the immunoassay result and the GC/MS result, not a result of human reader discrepancy requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) assay designed to provide an automated analytical result (qualitative or semi-quantitative concentration of EDDP). It is not an imaging AI device that assists human readers in interpreting medical images. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, the primary performance evaluation is standalone. The ARK EDDP Assay is an automated immunoassay performed on clinical chemistry analyzers. The results (qualitative positive/negative or semi-quantitative concentration) are generated by the assay system (reagents + analyzer) without human interpretive input for the result itself. The results are then reported to a clinician for medical interpretation in conjunction with other clinical information. The precision, analytical recovery, specificity, interference, and method comparison studies all represent the "algorithm only" or "device only" performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth used is definitive analytical method (Gas Chromatography/Mass Spectrometry - GC/MS). For in vitro diagnostic assays measuring specific chemical compounds, GC/MS is considered the most accurate and reliable "truth" method.
8. The sample size for the training set
This document does not specify a separate "training set" sample size. For an immunoassay like the ARK EDDP Assay, the "training" (development and optimization) phase would involve numerous laboratory experiments to select reagents, optimize assay conditions, and establish initial performance characteristics. This is distinct from machine learning model training sets. The studies presented (precision, analytical recovery, specificity, interference, method comparison) are performance validation studies for the final device, not datasets used to train an algorithm.
9. How the ground truth for the training set was established
Not applicable in the context of an immunoassay. The "ground truth" for developing an immunoassay would be established by controlled spiking experiments using certified reference materials of the analyte (EDDP) in a known matrix (drug-free human urine), and confirming those concentrations using methods like GC/MS or LC-MS/MS, or by gravimetric/volumetric preparation. The goal is to ensure the assay accurately measures the target compound.
Ask a specific question about this device
(110 days)
The ARK Fentanyl Assay is an immunoasay intended for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use in laboratories with automated clinical chemistry analyzers. This in vitro diagnostic device is for prescription use only.
The ARK Fentanyl Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug test result, particularly when the prelininary test result is positive.
The ARK Fentanyl Assay is a homogeneous enzyme immunoassay technique used for the analysis of a specific compound in human urine. The assay is based on competition between drug in the specimen and drug labeled with recombinant glucose-6-phosphate dehydrogenase (rG6PDH) for antibody binding sites. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly proportional to the drug concentration. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH in the presence of glucose-6-phosphate (G6P), resulting in an absorbance change that is measured spectrophotometrically. Endogenous G6PDH does not interfere because the coenzyme NAD functions only with the bacterial enzyme used in the assay.
The ARK Fentanyl Assay consists of reagents R1 anti-fentanyl polyclonal antibody with substrate and R2 fentanyl derivative labeled with bacterial recombinant G6PDH enzyme.
The provided text describes the performance characteristics of the ARK Fentanyl Assay, an immunoassay for the qualitative detection of fentanyl in human urine. Here's a breakdown of the requested information based on the document:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" in a typical quantitative pass/fail table format for all parameters. Instead, it presents various performance studies with their results, implying that these results met the internal criteria for the manufacturer to claim substantial equivalence. The precision table comes closest to showing performance relative to a cutoff.
| Performance Characteristic | Acceptance Criteria (Implied/Direct) | Reported Device Performance |
|---|---|---|
| Precision | Clarity on positive/negative results around cutoff (1.0 ng/mL) | At Cutoff (1.0 ng/mL): 97 Negative, 63 Positive results out of 160. At +25% Cutoff (1.25 ng/mL) and above: 160/160 (100%) Positive. At -25% Cutoff (0.75 ng/mL) and below: 160/160 (100%) Negative. |
| Analytical Specificity (Cross-reactivity) | Detection of fentanyl and its major metabolite (Norfentanyl); minimal cross-reactivity with other substances. | Norfentanyl: 10% cross-reactivity (at 2.5 ng/mL). Other fentanyl analogs showed varying cross-reactivity (e.g., Acetyl fentanyl: 83.33% at 1.2 ng/mL). Other opioids, structurally similar, and functional analogs tested negative at high concentrations (e.g., Morphine: 100 µg/mL). |
| Interference (Structurally Unrelated Compounds) | No false results in presence of common drugs/substances. | No false negative or false positive results observed for 36 tested compounds at high concentrations (e.g., Acetaminophen 500 µg/mL, Ibuprofen 500 µg/mL) when spiked into urine at ±50% of the cutoff concentration. |
| Interference (Endogenous Substances) | No interference from common endogenous substances in urine. | No interference observed for 14 tested endogenous substances (e.g., Acetone 1000 mg/dL, Glucose 3000 mg/dL) at high concentrations when spiked into urine at ±50% of the cutoff concentration. |
| Interference (Specific Gravity & pH) | No interference across physiological range. | No interference observed for specific gravity (1.001 to 1.030) and pH (3.0 to 11.0) when tested at ±50% of the cutoff concentration. |
| Interference (Boric Acid) | No interference. | No interference observed with 1% w/v boric acid when tested at ±50% of the cutoff concentration. |
| Method Comparison (Concordance with LC-MS/MS) | High agreement with confirmatory method, especially at and away from cutoff. | Total samples: 150. High Positive (>1.5 ng/mL LC-MS/MS): 64/64 (100%) Positive by ARK. Low Negative (<0.5 ng/mL LC-MS/MS): 50/51 (98%) Negative by ARK (1 discordant positive). Near Cutoff Negative (0.5-0.9 ng/mL LC-MS/MS): 20 positive, 3 negative by ARK (23 total). Near Cutoff Positive (1.0-1.5 ng/mL LC-MS/MS): 12 positive, 0 negative by ARK (12 total). |
| Calibration Curve Stability | Maintain effectiveness for a specified period. | Effective up to at least 15 days. |
2. Sample size used for the test set and the data provenance
- Precision Test Set: Not explicitly stated as a separate "test set" but 160 analyses were performed for each of 9 fentanyl concentration levels, totaling 1440 analyses.
- Analytical Specificity / Interference Test Set: Concentrations of various compounds were spiked into drug-free, negative human urine. The number of individual urine samples used is not specified, but the testing was performed on Beckman Coulter AU680.
- Method Comparison Test Set: A total of 150 unaltered clinical urine specimens.
- Data Provenance: The document states "human urine" was used for testing, and for the method comparison, it specifies "unaltered clinical urine specimens that are not individually identifiable." No specific country of origin is mentioned, but the context implies it was likely conducted in the US, given the FDA submission. The studies appear to be prospective as they were specifically designed and conducted to evaluate the device's performance for this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This section is not applicable as the device is an in-vitro diagnostic (IVD) immunoassay for detecting a chemical (fentanyl) in urine, not an image-based AI device requiring expert interpretation of medical images. The "ground truth" for the method comparison was established by a laboratory-based confirmatory method, LC-MS/MS (Liquid Chromatography/tandem Mass Spectrometry).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable for this type of IVD device. Adjudication methods like "2+1" or "3+1" are typically used in medical imaging studies where human readers provide interpretations that need to be reconciled to establish a consensus ground truth. Here, the ground truth is an analytical measurement (LC-MS/MS).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device is an automated immunoassay, which does not involve human readers for interpretation, nor is it an AI-assisted diagnostic tool. Its performance is compared against a gold-standard laboratory method (LC-MS/MS).
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the studies presented (Precision, Analytical Specificity, Interference, and Method Comparison) represent the standalone performance of the ARK Fentanyl Assay, as it is an automated immunoassay performed on clinical chemistry analyzers, without human intervention for result interpretation beyond the machine's output. The device itself is the "algorithm only" in this context.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth for the method comparison study was established by LC-MS/MS (Liquid Chromatography/tandem Mass Spectrometry), which is considered a definitive confirmatory chemical method for fentanyl detection.
8. The sample size for the training set
This information is not provided in the document. The document describes performance validation studies for the device, but it does not detail the internal development or "training" process (if any, in the context of an immunoassay optimization rather than machine learning). Immunoassays are biochemically developed systems, not typically "trained" on data sets in the same way an AI algorithm is.
9. How the ground truth for the training set was established
As the document does not describe a "training set" in the context of machine learning, this information is not applicable. The development of an immunoassay involves extensive biochemical research, optimization, and validation using well-characterized samples, rather than a data-driven "training" process with a ground truth dataset in the AI sense. The calibrators and controls used are prepared by volumetric dilution of high purity fentanyl traceable to HPLC into drug-free human urine.
Ask a specific question about this device
(261 days)
The ARK™ Methotrexate Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of methotrexate in human serum or plasma on automated clinical chemistry analyzers. The measurements obtained are used in monitoring levels of methotrexate to help ensure appropriate therapy.
Specimens from patients who have received glucarpidase (carboxypeptidase G2) as a high dose methotrexate rescue therapy should not be tested with the ARK Methotrexate Assay.
The ARK Methotrexate Assay is a homogeneous immunoassay based on competition between drug in the specimen and Methotrexate labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for binding to the antibody reagent. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly proportional to the drug concentration. Active enzyme converts the coenzyme nicotinamide adenine dinucleotide (NAD) to NADH that is measured spectrophotometrically as a rate of change in absorbance. Endogenous serum G6PDH does not interfere with the results because the coenzyme NAD functions only with the bacterial enzyme used in the assay.
The ARK Methotrexate Assay consists of reagents R1 anti-Methotrexate polyclonal antibody with substrate and R2 Methotrexate labeled with bacterial G6PDH enzyme. The ARK Methotrexate Calibrator consists of a six-level set to calibrate the assay, and the ARK Methotrexate Control consists of a six-level set used for quality control of the assay (tri-level calibration range set and tri-level high range set). The ARK Methotrexate Dilution Buffer is equivalent to zero calibrator (Calibrator A).
Here's an analysis of the provided text to extract the acceptance criteria and study information:
Acceptance Criteria and Device Performance for ARK™ Methotrexate Assay
The provided document describes the acceptance criteria and the study that proves the ARK™ Methotrexate Assay, when used on the Beckman Coulter AU680 analyzer, meets these criteria. The study aims to demonstrate substantial equivalence to the assay's performance on the predicate device, the Roche/Hitachi 917 analyzer.
1. Table of Acceptance Criteria and Reported Device Performance
The document states: "The pre-determined Pass/Fail acceptance criteria were met for all of the above performance studies." However, the specific numerical acceptance criteria or performance values for each characteristic are not provided in the given text. It only lists the types of performance characteristics evaluated.
| Performance Characteristic | Acceptance Criteria (Not Detailed) | Reported Device Performance |
|---|---|---|
| Limit of Blank | Undisclosed Pass/Fail Criteria | Met Acceptance Criteria |
| Limit of Detection | Undisclosed Pass/Fail Criteria | Met Acceptance Criteria |
| Limit of Quantitation | Undisclosed Pass/Fail Criteria | Met Acceptance Criteria |
| Recovery | Undisclosed Pass/Fail Criteria | Met Acceptance Criteria |
| Linearity | Undisclosed Pass/Fail Criteria | Met Acceptance Criteria |
| Accuracy (Method Comparison) | Undisclosed Pass/Fail Criteria | Met Acceptance Criteria |
| Precision | Undisclosed Pass/Fail Criteria | Met Acceptance Criteria |
| Specificity | Undisclosed Pass/Fail Criteria | Met Acceptance Criteria |
| Cross-reactivity | Undisclosed Pass/Fail Criteria | Met Acceptance Criteria |
| Carry-over | Undisclosed Pass/Fail Criteria | Met Acceptance Criteria |
| Dilution Recovery | Undisclosed Pass/Fail Criteria | Met Acceptance Criteria |
| On-board Stability | Undisclosed Pass/Fail Criteria | Met Acceptance Criteria |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample sizes used for the test set for any of the performance studies.
The document refers to "human serum or plasma" as the sample type. The country of origin of the data is not explicitly stated. The study is a "validation and verification" activity, implying it was conducted specifically for this submission, likely making it a prospective study for the purpose of device validation.
3. Number of Experts and Qualifications for Ground Truth
The concept of "experts" and their qualifications for establishing ground truth, as typically seen in image analysis or diagnostic interpretation, is not applicable to this type of in vitro diagnostic device (immunoassay) study. The ground truth for this device's performance characteristics is established through analytical methods and reference standards, not expert interpretation.
4. Adjudication Method for the Test Set
The concept of an "adjudication method" (e.g., 2+1, 3+1) is not applicable to this type of in vitro diagnostic device (immunoassay) study. Performance is assessed through quantitative measurements against established analytical criteria, not through expert consensus on diagnostic outcomes.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study is not applicable to this type of in vitro diagnostic device (immunoassay). This study design is typically used for imaging or diagnostic interpretation tasks where human readers' performance is being evaluated, often with and without AI assistance. This device is an automated assay, not an AI for human interpretive tasks.
6. Standalone (Algorithm Only) Performance Study
The study described is a standalone performance study for the algorithm (assay system) on the Beckman Coulter AU680 analyzer. The validation and verification activities listed (Limit of Blank, Limit of Detection, Accuracy, Precision, etc.) directly evaluate the performance of the device itself, without human interpretation in the loop. The purpose is to demonstrate that the assay system alone performs comparably to its predicate.
7. Type of Ground Truth Used
For an immunoassay like the ARK™ Methotrexate Assay, the "ground truth" for evaluating its performance characteristics (analytical accuracy, precision, linearity, etc.) would be established through:
- Reference materials/standards: Calibrators and controls with precisely known concentrations of methotrexate.
- Reference methods: Comparison with established, validated analytical methods (e.g., LC-MS/MS, or the predicate device's performance which was previously validated) for samples with unknown concentrations.
- Spiking studies: Adding known amounts of analyte to a sample and measuring recovery to assess accuracy.
The document implicitly refers to these as part of the "analytical performance studies" and "validation and verification activities."
8. Sample Size for the Training Set
The document does not mention a training set or its sample size. This is expected as the ARK™ Methotrexate Assay is a homogeneous enzyme immunoassay kit, not a machine learning or AI model that requires a "training set" in the conventional sense. The "training" for such a system would involve validating its components and establishing its operational parameters on the target analyzer.
9. How Ground Truth for the Training Set Was Established
Since there is no "training set" in the context of machine learning, this question is not applicable. The assay is a chemical and biological system where parameters are set based on chemical principles and reagent characteristics, and then validated through performance studies.
Ask a specific question about this device
(294 days)
The ARK Voriconazole II Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of voriconazole in human serum on automated clinical chemistry analyzers. The measurements obtained are used in monitoring levels of voriconazole to help ensure appropriate therapy. The assay should only be used in conjunction with information available from clinical evaluations and other diagnostic procedures.
ARK Voriconazole II Calibrator is intended for use in calibration of the ARK Voriconazole II Assay.
ARK Voriconazole II Control is an assayed quality control material intended for use in quality control of the ARK Voriconazole II Assay.
The ARK Voriconazole II Assay Test System consists of the ARK Voriconazole II Assay, the ARK Voriconazole II Calibrator, and the ARK Voriconazole II Control.
The ARK Voriconazole II Assay consists of:
- Reagent R1: rabbit polyclonal antibodies to voriconazole, glucose-6-phosphate, nicotinamide adenine dinucleotide, bovine serum albumin, sodium azide, and stabilizers.
- Reagent R2: voriconazole labeled with bacterial G6PDH buffer, bovine serum albumin, sodium azide, and stabilizers.
The ARK Voriconazole II Calibrator has six levels and consists of voriconazole, buffer, bovine serum albumin, and sodium azide.
The ARK Voriconazole II Control has three levels and consists of voriconazole, buffer, bovine serum albumin, and sodium azide.
Here's a breakdown of the acceptance criteria and the study details for the ARK Voriconazole II Assay Test System, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally implied by the special controls and the types of studies performed, aiming for performance characteristics that make the device fit for its intended use. Specific numerical acceptance criteria were not explicitly stated in all sections (e.g., for precision, rather than a universal requirement, the reported values are provided). However, for interference studies, a clear threshold was defined.
| Performance Characteristic | Acceptance Criteria (Implied/Explicit) | Reported Device Performance |
|---|---|---|
| Precision (Internal) | Not explicitly stated (demonstrate precision) | Total CV% for controls: LOW (4.9), MID (4.3), HIGH (4.5). Total CV% for human serum: LOW (4.6), MID (4.3), HIGH (4.2). Second internal study Total CV%: LOW (5.9), MID (5.1), HIGH (6.3). |
| Precision (Multi-Site) | Not explicitly stated (demonstrate precision) | Reproducibility CV% for controls: LOW (6.0), MID (5.5), HIGH (6.5). Reproducibility CV% for human serum: LOW (5.4), MID (5.2), HIGH (5.5). |
| Linearity/Reportable Range | Claimed measuring range (0.5 to 14.0 ug/mL) supported by regression. | Y = 1.0209 X - 0.0416, R²: 0.9995. Supports range 0.5 to 14.0 ug/mL. |
| Analytical Recovery | Not explicitly stated (assess recovery). | Ranged from 90.0% to 104.9%. |
| Traceability | Calibrators traceable to certified standard. | Calibrators traceable to USP Reference Standard. |
| Detection Limit | LoB, LoD, LoQ determined per CLSI EP17-A2. LoQ bias ≤15% and within-lab precision ≤10%. | LoB = 0.003 ug/mL. LoD = 0.05 ug/mL. LoQ = 0.5 ng/mL (0.0005 ug/mL). Note: The text states 0.5 ng/mL but the linear range is 0.5 ug/mL, implying a discrepancy or typo in the LoQ value in the document. Assuming it meant 0.5 ug/mL for consistency with the linear range. |
| Analytical Specificity (Endogenous Substances) | Interference considered significant if analytical recovery is outside of ± 10% of initial value. | All reported percentage recoveries for endogenous substances at 1.0 µg/mL and 5.0 µg/mL voriconazole were within 90-110%. |
| Cross-reactivity (N-oxide Voriconazole) | Data from studies performed to evaluate cross-reactivity. | Non-significant (≤ 3.0%) cross-reactivity observed. |
| Potentially Co-Administered Medications | Data from interference studies. Interference considered significant if analytical recovery is outside of ± 10% of initial value. | No significant interference observed. All reported percentage recoveries were within the 90-110% range. |
| Method Comparison (Accuracy) | Data demonstrating accuracy; comparator method not subject to bias. | Passing Bablok regression slope 0.98-0.99 (95% CI covering 1), intercept 0.05-0.08 (95% CI covering 0), R² 0.95-0.97. |
2. Sample Size Used for the Test Set and Data Provenance
- Precision (Internal Study 1): 160 replicates for each of 3 control levels and 3 human serum pools. Likely internal ARK data.
- Precision (Internal Study 2): 120 replicates for each of 3 patient sample pools. Likely internal ARK data.
- Precision (Multi-Site Study): 120 replicates for each of 3 control levels and 3 human serum pools across 3 sites (360 total per control/pool level). Data from ARK and two external sites.
- Linearity: 11 levels of samples, tested in two runs with three replicates per run. Samples were prepared by dilution of pure voriconazole in pooled human serum.
- Analytical Recovery: Serum samples prepared by gravimetric and volumetric addition of pure voriconazole to human serum. Number of samples not explicitly stated but covers 7 concentration points.
- Detection Limit Studies:
- LoB: 20 blank patient specimens tested.
- LoD/LoQ: A low concentration pooled serum sample.
- Analytical Specificity (Endogenous Substances): Samples with known levels of voriconazole (1 and 5 ug/mL) spiked with interferents. Each sample tested in two runs with three replicates per run.
- Cross-reactivity: N-oxide-voriconazole at 5.0 ug/mL and 10.0 µg/mL levels, tested in absence or presence of voriconazole.
- Potentially Co-Administered Medications: Samples with known levels of voriconazole (1 and 5 ug/mL) spiked with 58 different medications.
- Method Comparison Study: 165 serum specimens. Samples represented a diverse population of in-hospital patients. Tested at 3 study sites. Provenance is prospective in the sense that they are patient samples collected for comparison, likely from the hospitals where the sites were located. Origin (country) not specified.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Not Applicable. This device is an in-vitro diagnostic test for measuring an analyte (voriconazole concentration) in human serum. The "ground truth" for these studies is established through either:
- Known gravimetric/volumetric preparation of standards (for linearity, recovery, detection limits, interference, cross-reactivity).
- Comparison with a reference method (LC-MS/MS for the method comparison study).
The concept of "experts" to establish ground truth as typically found in image analysis or clinical diagnosis contexts (e.g., radiologists, pathologists) does not apply here.
4. Adjudication Method for the Test Set
- Not Applicable. As explained above, ground truth is established through analytical means or comparison to a reference method, not through human expert adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
- Not Applicable. This is an in-vitro diagnostic device for quantitative measurement, not an AI-assisted diagnostic imaging or interpretation device that involves human "readers."
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Yes, this is a standalone device validation. The ARK Voriconazole II Assay Test System is an automated clinical chemistry analyzer-based test. The performance characteristics described are for the device (the assay and analyzer) itself, without human intervention in the measurement process once the sample is loaded. The "algorithm" here is the chemical reaction and spectrophotometric measurement programmed into the analyzer.
7. The Type of Ground Truth Used
- Known concentrations: For precision, linearity, analytical recovery, detection limits, endogenous substance interference, cross-reactivity, and co-administered medication interference studies, the "ground truth" was either:
- Gravimetrically/volumetrically prepared samples with known concentrations of voriconazole or interferents.
- Pooled human serum samples with established mean concentrations.
- Reference method (LC-MS/MS): For the method comparison study, the validated LC-MS/MS method served as the reference standard for establishing the ground truth concentrations in patient samples.
8. The Sample Size for the Training Set
- Not Applicable. This is an in-vitro diagnostic assay based on a homogeneous enzyme immunoassay principle, not a machine learning or AI algorithm that requires a "training set" in the conventional sense. The "training" of the device involves calibration using the provided ARK Voriconazole II Calibrators.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. As above, it's not an AI/ML training set. The "ground truth" for the calibrators (which serve a similar function to training in some contexts, by enabling the device to accurately convert signal into concentration) is established through internal procedures and traceability to a certified USP Reference Standard. The text states: "The ARK Voriconazole II Calibrators are traceable to a certified USP Reference Standard." and "Concentrations of the ARK Voriconazole II Calibrators and Controls are assigned through internal procedures that were reviewed and found to be acceptable."
Ask a specific question about this device