K Number
K200197

Validate with FDA (Live)

Date Cleared
2020-02-26

(30 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ARK Fentanyl II Assay is an immunoassay intended for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use in laboratories with automated clinical chemistry analyzers. This in vitro diagnostic device is for prescription use only.

The ARK Fentanyl II Assay provides only a preliminary analytical result. A more specific alternative chemical method must be used in order to obtain a confirmed positive analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug test result, particularly when the preliminary test result is positive.

Device Description

The ARK Fentanyl II Assay is a homogeneous enzyme immunoassay technique used for the analysis of a specific compound in human urine. The assay is based on competition between drug in the specimen and drug labeled with recombinant glucose-6-phosphate dehydrogenase (rG6PDH) for antibody binding sites. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly related to the drug concentration. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH in the presence of glucose-6-phosphate (G6P), resulting in an absorbance change that is measured spectrophotometrically. Endogenous serum G6PDH does not interfere because the coenzyme NAD functions only with the bacterial enzyme used in the assay.

The ARK Fentanyl II Assay consists of reagents R1 anti-fentanyl monoclonal antibodies with substrate and R2 fentanyl derivative labeled with bacterial recombinant G6PDH enzyme.

AI/ML Overview

1. Acceptance Criteria and Reported Device Performance for ARKTM Fentanyl II Assay

The provided text details the performance characteristics of the ARKTM Fentanyl II Assay, which is an immunoassay for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The study evaluates precision, analytical specificity (cross-reactivity with structurally similar compounds and lack of interference from structurally unrelated compounds and endogenous substances), and method comparison with LC-MS/MS.

Here's a table summarizing the acceptance criteria (implied by the data presentation for each test) and the reported device performance:

Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
Precision (Qualitative)Samples below the cutoff (e.g., -100%, -75%, -50%, -25% of cutoff) should consistently test Negative.- 0.00 ng/mL (-100%): 160 Negative (out of 160) - 0.25 ng/mL (-75%): 160 Negative (out of 160) - 0.50 ng/mL (-50%): 160 Negative (out of 160) - 0.75 ng/mL (-25%): 160 Negative (out of 160) - 1.00 ng/mL (Cutoff): 84 Negative; 76 Positive (out of 160) - 1.25 ng/mL (+25%): 160 Positive (out of 160) - 1.50 ng/mL (+50%): 160 Positive (out of 160) - 1.75 ng/mL (+75%): 160 Positive (out of 160) - 2.00 ng/mL (+100%): 160 Positive (out of 160)
Samples above the cutoff (e.g., +25%, +50%, +75%, +100% of cutoff) should consistently test Positive.
Samples at the cutoff should show a mix of Negative and Positive results, indicating proper cutoff discrimination.
Analytical SpecificityNorfentanyl (major metabolite) should show some cross-reactivity. Other fentanyl structural analogs/metabolites may show varying degrees of cross-reactivity. Structurally similar opioids and functional analogs should be negative at high concentrations.- Norfentanyl: 7% cross-reactivity at 15 ng/mL. - Acetyl fentanyl, Isobutyryl fentanyl: 90.91% cross-reactivity at 1.1 ng/mL. - Furanyl fentanyl, Para-fluoro fentanyl: 66.67% cross-reactivity at 1.5 ng/mL. - Other similar compounds (e.g., Carfentanil, Sufentanil, Alfentanil): show very low to <0.001% cross-reactivity at high concentrations. - Opioids/structurally similar compounds (e.g., Morphine, Codeine, Hydrocodone): All tested negative at 100 µg/mL.
InterferenceStructurally unrelated compounds (e.g., common medications, illicit drugs metabolites) and endogenous substances (e.g., creatinine, albumin, pH, specific gravity) should not produce false positive or false negative results when fentanyl is present/absent at concentrations near the cutoff. Boric acid is specified as a potential interferent and should be acknowledged.- Structurally Unrelated Compounds: No false results observed for 30+ compounds (e.g., Acetaminophen, Amphetamine, Ibuprofen) at high concentrations (up to 1000 µg/mL) when tested with 0.5 ng/mL and 1.5 ng/mL fentanyl. - Endogenous Substances: No interference observed for 14 endogenous substances (e.g., Acetone, Bilirubin, Glucose) at high concentrations (up to 4000 mg/dL) when tested with 0.5 ng/mL and 1.5 ng/mL fentanyl. - Specific Gravity & pH: No interference observed for specific gravity (1.002-1.030 g/mL) and pH (3.0-11.0) when tested with 0.5 ng/mL and 1.5 ng/mL fentanyl. - Boric Acid: Tested at 1% w/v, indicated negative result for both 0.5 ng/mL and 1.5 ng/mL fentanyl spiked samples. However, labeling includes a limitation: "Do not use Boric Acid as a preservative."
Method ComparisonHigh agreement (concordance) with a confirmatory method (LC-MS/MS) for both positive and negative samples, particularly for samples significantly above and below the cutoff. Limited discordant results are expected near the cutoff and should be explainable (e.g., cross-reactivity).- Total Samples: 147 clinical urine specimens. - Concordance: - LC-MS/MS < 0.5 ng/mL: 50 Negative, 1 Positive (1 discordant) - LC-MS/MS 0.5-0.9 ng/mL: 2 Negative, 21 Positive (2 discordant negatives, 21 concordant positives) - LC-MS/MS 1.0-1.5 ng/mL: 0 Negative, 11 Positive (11 concordant positives) - LC-MS/MS > 1.5 ng/mL: 0 Negative, 62 Positive (62 concordant positives) - Discordant Results Explanation: The 1 discordant positive at < 0.5 ng/mL (LC-MS/MS) and 21 discordant positives between 0.5-0.9 ng/mL (LC-MS/MS) were analyzed. All these discordant immunoassay positives were found to have detectable Norfentanyl (main metabolite) by LC-MS/MS, explaining the immunoassay's positive result due to cross-reactivity. The 2 discordant negatives near cutoff (0.5-0.9 ng/mL) are not further explained but are within expected variability for qualitative assays near cutoff. Overall good agreement.

Study Details:

  1. Sample size used for the test set and the data provenance:

    • Precision Study: N=160 for each of 9 concentration levels (total 1440 measurements). The samples were drug-free human urine supplemented with fentanyl. Data provenance is not explicitly stated as country of origin, but described as "drug-free, negative human urine", implying laboratory-prepared samples.
    • Analytical Specificity (Cross-reactivity/Interference): Samples were drug-free, negative human urine spiked with various compounds and fentanyl. Specific sample sizes for each compound tested are not provided, but generally involve multiple replicates for dose-response evaluation or testing with fentanyl-spiked urine.
    • Method Comparison: A total of 147 unaltered clinical urine specimens were used. The data provenance is not explicitly mentioned as country of origin, but it is stated these were "unaltered clinical urine specimens that are not individually identifiable." This suggests a retrospective collection of clinical samples.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The ground truth for the method comparison study was established by LC-MS/MS (Liquid Chromatography/tandem Mass Spectrometry), which is described as the "preferred confirmatory method" and was performed by a "licensed reference laboratory." This is an objective, analytical method, not reliant on human expert interpretation for the quantitative values. Therefore, individual human "experts" in the traditional sense (e.g., radiologists) were not used to establish the ground truth in this context. The expertise lies in the certified analytical process and the laboratory standards.
  3. Adjudication method for the test set:

    • No adjudication method (e.g., 2+1, 3+1) was applicable or mentioned, as the ground truth was established by quantitative analytical methods (LC-MS/MS), not by human expert review requiring consensus.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is an in vitro diagnostic immunoassay and not an AI-powered image analysis or medical device that involves human readers/interpreters in its primary function. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance was not conducted and is irrelevant to this device type. The device provides a direct analytical result.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, the performance presented is a standalone performance of the immunoassay system. It measures the algorithm's (immunoassay's) ability to detect fentanyl in urine samples independently of human intervention for result interpretation. The results are generated by an "automated clinical chemistry analyzer."
  6. The type of ground truth used:

    • For the method comparison study, the ground truth was quantitative analytical data from LC-MS/MS (Liquid Chromatography/tandem Mass Spectrometry). This is a highly specific and sensitive confirmatory laboratory method.
  7. The sample size for the training set:

    • This document describes a premarket notification (510(k)) for an in vitro diagnostic device (immunoassay). For such devices, internal development and optimization (which can be considered analogous to "training") occur during the assay design and reagent formulation phases. However, the FDA 510(k) submission typically focuses on the validation (test set) of the finalized device rather than detailed disclosure of the "training set" in the context of machine learning. The document does not specify a sample size for a training set in the way it would for a machine learning algorithm. The "training" here would involve iterative testing and refinement of antibody characteristics, enzyme conjugates, and reagent formulations using proprietary in-house samples until the desired performance (specificity, sensitivity, dynamic range) is achieved.
  8. How the ground truth for the training set was established:

    • As mentioned above, the concept of a "training set" with established ground truth in the machine learning sense is not directly applicable to this traditional immunoassay. The development process would involve:
      • Synthesizing or acquiring known concentrations of fentanyl and its metabolites.
      • Evaluating various antibody candidates and enzyme conjugates against these known concentrations.
      • Spiking drug-free human urine to create samples with precisely known concentrations of analytes.
      • Using highly accurate analytical methods (like GC/MS or LC-MS/MS) internally to confirm the concentrations of prepared samples or to guide dose-response adjustments during reagent formulation.
    • Ultimately, the ground truth for any "training" or optimization would be based on known chemical concentrations of the target analyte and potential interferents, confirmed by reference analytical methods.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue box, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

February 26, 2020

ARK Diagnostics, Inc. Thomas Houts, Ph.D. Director, Quality, Regulatory and Planning 48089 Fremont Boulevard Fremont, CA 94538

Re: K200197

Trade/Device Name: ARKTM Fentanyl II Assay Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate Test System Regulatory Class: Class II Product Code: DJG Dated: January 24, 2020 Received: January 27, 2020

Dear Thomas Houts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

  1. for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Marianela Perez-Torres, M.T., Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200197

Device Name ARKTM Fentanyl II Assay

Indications for Use (Describe)

The ARK Fentanyl II Assay is an immunoassay intended for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use in laboratories with automated clinical chemistry analyzers. This in vitro diagnostic device is for prescription use only.

The ARK Fentanyl II Assay provides only a preliminary analytical result. A more specific alternative chemical method must be used in order to obtain a confirmed positive analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug test result, particularly when the preliminary test result is positive.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

This 510(k) Summary of Safety and Effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act of 1990 and 21 CFR 807.92.

The assigned 510(k) number is K200197.

807.92 (a)(1): Name:ARK Diagnostics, Inc.
Address:48089 Fremont BlvdFremont, CA 94538 USA
Owner Operator Number:10027663
Establishment Registration:3005755244
Phone:(510) 270-6270
FAX:(510) 270-6298
Contact:Thomas Houts, Ph.D. – (510) 270-6296Director, Quality, Regulatory and Planning

Date Prepared: February 25, 2020

807.92 (a)(2): Device name – trade name and common name, and classification

Trade Name:ARKTM Fentanyl II Assay
Common Name:Homogeneous Enzyme Immunoassay, Opiate Test System

Classification:

Product CodeClassificationRegulation SectionPanel
DJGClass II21 CFR 862.3650Opiate Test SystemToxicology(91)

807.92 (a)(3): Identification of the legally marketed predicate device

ARK™ Fentanyl Assay - K180427

{4}------------------------------------------------

807.92 (a)(4): Device Description

The ARK Fentanyl II Assay is a homogeneous enzyme immunoassay technique used for the analysis of a specific compound in human urine. The assay is based on competition between drug in the specimen and drug labeled with recombinant glucose-6-phosphate dehydrogenase (rG6PDH) for antibody binding sites. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly related to the drug concentration. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH in the presence of glucose-6-phosphate (G6P), resulting in an absorbance change that is measured spectrophotometrically. Endogenous serum G6PDH does not interfere because the coenzyme NAD functions only with the bacterial enzyme used in the assay.

The ARK Fentanyl II Assay consists of reagents R1 anti-fentanyl monoclonal antibodies with substrate and R2 fentanyl derivative labeled with bacterial recombinant G6PDH enzyme.

807.92 (a)(5): Intended Use / Indications for Use

ARK Fentanyl II Assay

The ARK Fentanyl II Assay is an immunoassay intended for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use in laboratories with automated clinical chemistry analyzers. This in vitro diagnostic device is for prescription use only.

The ARK Fentanyl II Assay provides only preliminary analytical result. A more specific alternative chemical method must be used in order to obtain a confirmed positive analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug test result, particularly when the preliminary test result is positive.

{5}------------------------------------------------

807.92 (a)(6): Technological Similarities and Differences to the Predicate

SUBSTANTIAL EQUIVALENCE COMPARATIVE TABLE

Comparison between the ARK™ Fentanyl Assay and the ARK™ Fentanyl II Assay

CharacteristicPredicate DeviceARK™ Fentanyl Assay (K180427)Candidate DeviceARK™ Fentanyl II Assay
Similarities
Test SystemHomogenous enzyme immunoassay(EIA)Same
Intended UseFor the qualitative detection of fentanylin human urine at a cutoff concentrationof 1.0 ng/mLSame
Sample MatrixHuman urineSame
DetectionAbsorbance change measuredspectrophotometrically at 340 nm.Same
User EnvironmentClinical laboratories; Prescription useonlySame
MassSpectrometryConfirmationRequired to confirm preliminarypositive analytical resultsSame
Platform RequiredAutomated clinical chemistryanalyzerSame
Reagents FormLiquid - Ready to useSame
Reagent MaterialsTwo (2) reagent system:Antibody/substrate reagent andenzyme labeled conjugateSodium azide preservativeSame
Storage2-8°C until expiration dateSame
Measured AnalyteFentanylSame
Cutoff Level1.0 ng/mLSame
CharacteristicPredicate DeviceARKTM Fentanyl Assay (K180427)Candidate DeviceARKTM Fentanyl II Assay
Differences
AntibodyRabbit polyclonal antibodies tofentanylRabbit monoclonal antibodies tofentanyl

{6}------------------------------------------------

807.92 (b)(1) and 807.92 (b)(2): Brief Description of Nonclinical and Clinical Data

The following performance characteristics were obtained on the Beckman Coulter AU680® automated clinical chemistry analyzer.

Precision

Precision studies were performed using CLSI EP05-A3 as a guideline. Drug-free, negative human urine was supplemented with fentanyl (0.00 to 2.00 ng/mL). Each level was assayed in quadruplicate twice a day for 20 days (N=160). Results are summarized in the table below.

Human Urine(ng/mL)Relative %Cutoff# of ResultsQualitativePrecisionResults
0.00-100160160 Negative
0.25-75160160 Negative
0.50-50160160 Negative
0.75-25160160 Negative
1.00Cutoff16084 Negative;76 Positive
1.25+25160160 Positive
1.50+50160160 Positive
1.75+75160160 Positive
2.00+100160160 Positive

Analytical Specificity

All compounds tested were added to drug-free, negative human urine.

The cross-reactivity of the following metabolites and structural analogs of fentanyl was evaluated by spiking these compounds into drug-free, negative human urine and evaluated by dose-response to determine the approximate equivalence to the 1.0 ng/mL fentanyl cutoff. These concentrations were used to determine the percent cross-reactivity according to the formula:

% Cross-reactivity = (Cutoff concentration / Concentration approximately equivalent to the 1.0 ng/mL cutoff) X 100

For the compounds Alfentanil and Remifentanil that did not produce a positive result, the highest concentration tested was used to calculate percent cross-reactivity.

{7}------------------------------------------------

Cross-reactivity

Norfentanyl (Major Metabolite)

CompoundConcentrationTested(ng/mL)PercentCrossreactivity(%)
Norfentanyl (Major Metabolite)157

Other Metabolites and Structural Analogs of Fentanyl

CompoundConcentrationApproximatelyEquivalent tothe Cutoff(ng/mL)PercentCrossreactivity(%)
Acetyl fentanyl1.190.91
Isobutyryl fentanyl1.190.91
ω-1-Hydroxyfentanyl1.283.33
Acrylfentanyl1.376.90
Butyryl fentanyl1.471.43
Furanyl fentanyl1.566.67
Para-fluoro fentanyl1.566.67
Ocfentanil1.662.50
4-Fluoro-isobutyryl fentanyl1.952.63
Para-fluorobutyryl fentanyl (p-FBF)1.952.63
Valeryl fentanyl2.343.48
β-hydroxyfentanyl9.510.53
Acetyl norfentanyl12.18.26
(±) β-hydroxythiofentanyl32.73.06
(±)-3-cis-methyl fentanyl144.10.69
Carfentanil448.20.22
Despropionyl fentanyl (4-ANPP)471.80.21
Sufentanil2,3620.04
Remifentanil10,000<0.01
Norcarfentanil38,1960.003
Alfentanil100,000<0.001

{8}------------------------------------------------

The following opioids, structurally similar compounds, and functional analogs were negative at the concentrations tested in the ARK Fentanyl II Assay.

CompoundConcentrationTested(µg/mL)CompoundConcentrationTested(µg/mL)
6-Acetyl morphine100Naltrexone100
Buprenorphine100Norbuprenorphine100
Buprenorphine glucuronide100Norcodeine100
Codeine100Normeperidine100
Dextromethorphan100Normorphine100
Dihydrocodeine100Noroxycodone100
EDDP100Oxycodone100
EMDP100Oxymorphone100
Heroin100Pentazocine (Talwin)100
Hydrocodone100Pipamperone90
Hydromorphone100Quinine100
9-Hydroxyrisperidone100Quinidine100
Labetalol100Risperidone100
Levorphanol100Tapentadol100
M-Chlorophenylpiperazine(m-CPP)100Thioridazine100
Meperidine100Tilidine100
Methadone100Tramadol100
Morphine100Tramadol-O-Desmethyl100
Morphine-3-glucuronide100Tramadol-N-Desmethyl100
Naloxone100Trazodone100

{9}------------------------------------------------

Interference – Structurally Unrelated Compounds

High concentrations of the following structurally unrelated compounds were added into fentanylspiked urine (± 50% of the cutoff concentration). The substances listed below did not yield a false result relative to the cutoff.

CompoundConcentrationTested(µg/mL)0.5 ng/mL(-50% Cutoff)1.5 ng/mL(+50% Cutoff)
Acetaminophen500NegativePositive
Acetylsalicylic acid1000NegativePositive
Albuterol100NegativePositive
Amitriptyline100NegativePositive
Amobarbital100NegativePositive
Amphetamine100NegativePositive
Benzoylecgonine100NegativePositive
Bupropion100NegativePositive
Caffeine100NegativePositive
Carbamazepine100NegativePositive
Chlorpromazine100NegativePositive
Clomipramine100NegativePositive
Cyclobenzaprine100NegativePositive
Desipramine100NegativePositive
Doxepin100NegativePositive
Ecgonine100NegativePositive
Ephedrine100NegativePositive
Fluoxetine100NegativePositive
Fluphenazine100NegativePositive
Ibuprofen500NegativePositive
Imipramine100NegativePositive
Ketamine100NegativePositive
Lidocaine100NegativePositive
Maprotiline100NegativePositive
Methapyrilene100NegativePositive
Methaqualone100NegativePositive
Metronidazole300NegativePositive
Nicotine100NegativePositive
Norketamine100NegativePositive
Nortriptyline60NegativePositive
Oxazepam100NegativePositive
Phencyclidine100NegativePositive
Phenobarbital100NegativePositive
Propoxyphene100NegativePositive
Ranitidine100NegativePositive
CompoundConcentrationTested(µg/mL)0.5 ng/mL(-50% Cutoff)1.5 ng/mL(+50% Cutoff)
Secobarbital100NegativePositive
Valproic acid250NegativePositive
Venlafaxine100NegativePositive

{10}------------------------------------------------

Interference – Endogenous Substances

Interference studies were performed using CLSI EP07-A3 as a guideline. High concentrations of the following endogenous substances were added into fentanyl-spiked urine (± 50% of the cutoff concentration). No interference was observed when tested with the ARK Fentanyl II Assay.

CompoundConcentrationTested(mg/dL)0.5 ng/mL(-50% Cutoff)1.5 ng/mL(+50% Cutoff)
Acetone1000NegativePositive
Ascorbic Acid560NegativePositive
Bilirubin2NegativePositive
Creatinine500NegativePositive
Ethanol1000NegativePositive
Galactose10NegativePositive
Gamma Globulin500NegativePositive
Glucose3000NegativePositive
Hemoglobin500NegativePositive
Human Albumin500NegativePositive
Oxalic Acid100NegativePositive
Riboflavin7.5NegativePositive
NaCl4000NegativePositive
Urea2000NegativePositive

Interference – Specific Gravity and pH

Urine samples with specific gravity values from 1.002 to 1.030 g/mL and pH values ranging from 3.0 to 11.0 were tested in the presence of the two levels of fentanyl at ± 50% of the cutoff concentration. No interference was observed when tested with the ARK Fentanyl II Assay.

Interference – Boric Acid

One percent (1%) w/v of boric acid was tested into fentanyl-spiked urine (± 50% of the cutoff concentration). Results are provided in the table below.

CompoundConcentrationTested0.5 ng/mL(-50% Cutoff)1.5 ng/mL(+50% Cutoff)
Boric Acid1% w/vNegativeNegative

{11}------------------------------------------------

The device labeling includes the following limitation: "Do not use Boric Acid as a preservative."

Method Comparison

A total of one hundred forty seven (147) unaltered clinical urine specimens that are not individually identifiable were analyzed for fentanyl with the ARK Fentanyl II Assay and by LC-MS/MS. The LC-MS/MS confirmatory method was performed by a licensed reference laboratory and used a fentanyl cutoff of 0.2 ng/mL.

Specimens were tested with the ARK Fentanyl II Assay in single replicates on a Beckman Coulter AU680 analyzer and compared to results obtained by LC-MS/MS. Groups of up to 31 specimens were assayed per run. Each run was verified by assaying the bi-level ARK Fentanyl Controls (0.5 ng/mL and 1.5 ng/mL) as quality control samples.

Results are summarized as follows:

ARK ResultLow NegativeLess than50% belowthe CutoffNear CutoffNegativeBetween 50%below theCutoff andthe CutoffNear CutoffPositiveBetween theCutoff and50% abovethe CutoffHigh PositiveGreater than50% abovethe Cutoff
(< 0.5 ng/mLby LC-MS/MS)(0.5 - 0.9ng/mL by LC-MS/MS)(1.0 – 1.5ng/mL byLC-MS/MS)(> 1.5 ng/mLby LC-MS/MS)
Positive1 *211162
Negative50200

Discordant Results

*Norfentanyl was detected in this discordant sample (Sample ID #052) and contributed to the positive result obtained with the ARK Fentanyl Assay for this sample.

Sample IDNumberARK ImmunoassayResultFentanyl(ng/mL by LC-MS/MS)Norfentanyl(ng/mL by LC-MS/MS)
052*Positive0.47.6
065Positive0:55.2
058Positive0:57.9
069Positive0:531.2
060Positive0:5425.4
056Positive0.63.7
072Positive0.613.8
062Positive0.614.5
074Positive0.614.6
0રરPositive0.616.9

{12}------------------------------------------------

Sample IDNumberARK ImmunoassayResultFentanyl(ng/mL by LC-MS/MS)Norfentanyl(ng/mL by LC-MS/MS)
071Positive0.619.0
070Positive0.6161.7
051Positive0.72.1
066Positive0.73.1
064Positive0.815.9
073Positive0.845.8
063Positive0.92.2
061Positive0.96.5
057Positive0.912.3
053Positive0.914.0
059Positive0.962.6
054Positive0.963.4

Traceability and Value Assignment

ARK Fentanyl Calibrators and Controls are prepared by volumetric dilution of high purity fentanyl (certified solution traceable to HPLC) into non-sterile, processed human urine free of fentanyl. Testing is performed with the ARK Fentanyl II Assay on the Beckman Coulter AU680 automated clinical chemistry analyzer, calibrated with the ARK Fentanyl Calibrator.

807.92 (b)(3): Conclusions from Nonclinical Testing

As summarized above, the ARK Fentanyl II Assay system was shown to be substantially equivalent to the legally marketed predicate device, ARK Fentanyl Assay system: K180427.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).