K Number
K201089

Validate with FDA (Live)

Date Cleared
2021-08-05

(469 days)

Product Code
Regulation Number
862.3350
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ARK Lacosamide Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of lacosamide in human serum on automated clinical chemistry analyzers. The measurements obtained are used in monitoring levels of lacosamide to help ensure appropriate therapy.

Device Description

The ARK Lacosamide Assay is a homogeneous enzyme immunoassay based on competition between drug in the specimen and lacosamide labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for binding to the antibody reagent. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly related to the drug concentration. Active enzyme converts the coenzyme nicotinamide adenine dinucleotide (NAD) to NADH that is measured spectrophotometrically as a rate of change in absorbance. Endogenous serum G6PDH does not interfere with the results because the coenzyme NAD functions only with the bacterial enzyme used in the assay.

The ARK Lacosamide Assay consists of reagents R1 anti-lacosamide polyclonal antibody with substrate and R2 lacosamide labeled with bacterial G6PDH enzyme.

AI/ML Overview

The provided text describes the ARK Lacosamide Assay, a homogeneous enzyme immunoassay for quantitative determination of lacosamide in human serum. This device is intended for monitoring lacosamide levels to ensure appropriate therapy. The substantial equivalence is demonstrated through comparative testing against a predicate device (ARKTM Topiramate Assay, K083799) and various performance characteristic studies.

Here's a breakdown of the acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance:

Performance CharacteristicAcceptance CriteriaReported Device Performance
Limit of Quantitation (LoQ)Acceptable inter-assay precision (<20% CV) and recovery (±15%)0.40 ug/mL, with CVs ranging from 3.4% to 5.0% and recoveries from 83.2% to 94.1% for tested concentrations.
Measurement RangeN/A (Range established)0.40 - 24.00 µg/mL
RecoveryWithin ±10% of expected sample concentrationRecoveries for various concentrations from 0.40 to 20.00 µg/mL were between 90.4% and 105.8%.
LinearityPercent difference ±10% (for >1.00 µg/mL) or ≤0.20 µg/mL (for ≤1.00 µg/mL) between 1st and 2nd order regressed values.Linear relationship demonstrated between 0.40 and 25.00 µg/mL (y = 0.9998x - 0.0170). Differences within acceptable limits.
Precision (Total CV)≤10% total CVFor controls and human serum samples, total CVs ranged from 3.9% to 4.5%.
Interfering SubstancesMeasurement of lacosamide resulted in ≤10% errorAll tested interfering substances resulted in ≤10% error (recoveries ranging from 95.8% to 103.5%).
Specificity (O-Desmethyl Metabolite)Not clinically significant (< 3.0% crossreactivity)Crossreactivity of O-desmethyl lacosamide metabolite was not clinically significant.
Crossreactivity (Other Drugs)Recoveries within 10% of the expected levelRecoveries for 77 compounds ranged from 90.9% to 109.5%.
Sample StabilityN/A (Stability demonstrated)Stable for at least 48 hours at room temperature, 28 days refrigerated, and 34 months frozen. Stable after 3 freeze/thaw cycles.
Shelf-life StabilityN/A (Stability demonstrated)Up to 18 months when stored unopened at 2-8°C.
On-Board StabilityN/A (Stability demonstrated)Up to 60 days on-board the instrument.
Calibration Curve StabilityN/A (Stability demonstrated)Effective up to at least 14 days.
Method Comparison (vs. LC-MS/MS)N/A (Correlation established)Slope: 1.01 (0.99 to 1.04), y-intercept: 0.03 (-0.10 to 0.15), r2: 0.98 (0.98 to 0.99)

2. Sample size used for the test set and the data provenance:

  • LoQ: 40 replicates (8 replicates x 5 runs) for each of 3 concentrations. Data provenance is implied to be laboratory-generated (pooled human serum supplemented with lacosamide).
  • Measurement Range: Not a directly tested "test set" in terms of patient samples; rather, it refers to the range characterized by other studies like recovery and linearity.
  • Recovery: Not explicitly stated as a separate "test set" size for recovery studies, but involved adding concentrated lacosamide to pooled human serum. "Two analytical runs of three replicates of each sample were assayed."
  • Linearity: Dilutions of a 30.00 µg/mL lacosamide serum sample. "Two analytical runs of three replicates of each sample were assayed."
  • Method Comparison: 150 unaltered, human serum specimens. The data provenance is described as "human serum specimens" which implies retrospective clinical samples, but the country of origin is not specified. They are "not individually identifiable."
  • Precision: 160 replicates for each of 6 samples/controls (quadruplicate twice a day for 20 days). Implied laboratory-generated (tri-level controls and pooled human serum with lacosamide).
  • Interfering Substances: For each interfering substance, lacosamide was present at 2 concentrations (2.0 and 15.0 µg/mL). "Two analytical runs of three replicates of each sample (6 replicates total) were assayed." Implied laboratory-generated.
  • Specificity (O-Desmethyl Metabolite): Assayed lacosamide at 2 concentrations (2.00 and 15.00 µg/mL) in the absence and presence of the metabolite at 2 concentrations (5.0 and 30.0 µg/mL). Implied laboratory-generated.
  • Crossreactivity (Other Drugs): For each of 77 compounds, lacosamide was present at 2 concentrations (2.00 and 15.00 µg/mL). Implied laboratory-generated.
  • Sample Stability: Not a statistical "test set" but based on supporting data; implies samples stored under different conditions.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This device is an in vitro diagnostic assay that measures a chemical compound (lacosamide concentration). The "ground truth" for method comparison studies is established by a reference method, not by expert interpretation.
  • In the method comparison study, the ARK Lacosamide Assay results were compared against LC-MS/MS (Liquid Chromatography-Mass Spectrometry/Mass Spectrometry). LC-MS/MS is a highly accurate and sensitive analytical chemistry technique, often considered a "gold standard" for quantifying small molecules like drugs in biological matrices.
  • Therefore, no human experts were used to establish the ground truth; instead, an established analytical method served as the reference.

4. Adjudication method for the test set:

  • Not applicable as the ground truth is established by a quantitative analytical method (LC-MS/MS), not by human interpretation or consensus.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is an in vitro diagnostic assay, not an AI-powered diagnostic imaging or interpretation tool that assists human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Yes, the performance characteristics described (LoQ, Measurement Range, Recovery, Linearity, Precision, Interfering Substances, Specificity, Crossreactivity, Stability) are all standalone algorithm (assay) performance studies. The method comparison study also evaluated the standalone performance of the ARK Lacosamide Assay against LC-MS/MS.
  • The device explicitly states it is "intended for the quantitative determination of lacosamide in human serum on automated clinical chemistry analyzers," indicating standalone operation with human oversight rather than human-in-the-loop interpretation of complex data.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For the method comparison study, the ground truth was "results from LC-MS/MS," which is a reference analytical method.
  • For other performance studies (LoQ, recovery, linearity, precision, interference, specificity, cross-reactivity), the ground truth was based on the known concentrations of lacosamide and interfering substances added to human serum samples.

8. The sample size for the training set:

  • This is an in vitro diagnostic assay, not a machine learning or AI model that requires a "training set" in the conventional sense. The development of the assay's reagents and methodologies involves analytical chemistry and biochemical principles rather than statistical training on data.

9. How the ground truth for the training set was established:

  • Not applicable for the reasons stated in point 8. The "ground truth" for the development of this assay would be the accurate chemical characterization of lacosamide and its antibodies, and the optimization of the enzymatic reaction to accurately reflect lacosamide concentration.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the rest of the name.

August 5, 2021

ARK Diagnostics, Inc. Thomas Houts Director, Quality Regulatory and Planning 48089 Fremont Boulevard Fremont, CA 94538

Re: K201089

Trade/Device Name: ARK Lacosamide Assay Regulation Number: 21 CFR 862.3350 Regulation Name: Diphenylhydantoin test system Regulatory Class: Class II Product Code: NWM Dated: October 28, 2020 Received: November 19, 2020

Dear Thomas Houts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Marianela Perez-Torres, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201089

Device Name ARK Lacosamide Assay

Indications for Use (Describe)

ARK Lacosamide Assay:

The ARK Lacosamide Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of lacosamide in human serum on automated clinical chemistry analyzers. The measurements obtained are used in monitoring levels of lacosamide to help ensure appropriate therapy.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5: 510(k) SUMMARY

This 510(k) Summary of Safety and Effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act of 1990 and 21 CFR 807.92.

The assigned 510(k) number is K201089.

807.92 (a)(1): Name:ARK Diagnostics, Inc.
Address:48089 Fremont BlvdFremont, CA 94538 USA
Owner Operator Number:10027663
Establishment Registration:3005755244
Phone:(510) 270-6270
FAX:(510) 270-6298
Contact:Thomas Houts, Ph.D.Director, Quality, Regulatory and PlanningEmail: tom@arkt-dm.comDirect phone: 510-270-6296

Date Prepared: April 22, 2020

807.92 (a)(2): Device name - trade name and common name, and classification

Trade Name:ARK Lacosamide Assay
-----------------------------------

Common Name: Homogeneous Enzyme Immunoassay

Classification:

Product CodeClassificationRegulation SectionPanel
NWMClass II862.3350Diphenylhydantoin Test SystemToxicology(91)

{4}------------------------------------------------

807.92 (a)(3): Identification of the legally marketed predicate device

Predicate Device Name: ARKTM Topiramate Assay Predicate 510(k) Number: K083799

807.92 (a)(4): Device Description

The ARK Lacosamide Assay is a homogeneous enzyme immunoassay based on competition between drug in the specimen and lacosamide labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for binding to the antibody reagent. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly related to the drug concentration. Active enzyme converts the coenzyme nicotinamide adenine dinucleotide (NAD) to NADH that is measured spectrophotometrically as a rate of change in absorbance. Endogenous serum G6PDH does not interfere with the results because the coenzyme NAD functions only with the bacterial enzyme used in the assay.

The ARK Lacosamide Assay consists of reagents R1 anti-lacosamide polyclonal antibody with substrate and R2 lacosamide labeled with bacterial G6PDH enzyme.

Summary and Explanation of Test

Lacosamide (Vimpat®, UCB, Inc.) [(R)-2-acetamido-N-benzyl-3-methoxypropionamide] is indicated for adjunctive therapy of partial-onset seizures in patients ≥17 years.

807.92 (a)(5): Intended Use / Indications for Use

ARK Lacosamide Assay

The ARK Lacosamide Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of lacosamide in human serum on automated clinical chemistry analyzers. The measurements obtained are used in monitoring levels of lacosamide to help ensure appropriate therapy.

{5}------------------------------------------------

807.92 (a)(6): Technological Similarities and Differences to the Predicate

SUBSTANTIAL EQUIVALENCE COMPARATIVE TABLES

Predicate Device Candidate Device Characteristic ARKTM Topiramate Assay (K083799) ARK™ Lacosamide Assay The ARK Topiramate The ARK Lacosamide is Assay is Assay a a homogeneous enzyme immunoassay intended homogeneous enzyme immunoassay Intended Use the quantitative determination intended for the quantitative determination of for of topiramate in human serum or plasma on lacosamide in human serum on automated automated clinical chemistry analyzers. clinical chemistry analyzers. The results obtained are used in the diagnosis The measurements obtained are used in Indications for and treatment of topiramate overdose and in monitoring levels of lacosamide to help monitoring levels of topiramate to help Use ensure appropriate therapy. ensure appropriate therapy. Sample Matrix Human serum or plasma Human serum Two (2) reagent system: Two (2) reagent system: Anti-topiramate Antibody/Substrate Reagent Anti-lacosamide Antibody/Substrate Reagent (R1) containing rabbit polyclonal antibodies (R1) containing rabbit polyclonal antibodies glucose-6-phosphate, topiramate、 lacosamide、 glucose-6-phosphate, to to nicotinamide adenine dinucleotide, boyine Reagent nicotinamide adenine dinucleotide, bovine Components serum albumin, preservatives, and stabilizers serum albumin, sodium azide, and stabilizers Enzyme Reagent (R2) containing topiramate Enzyme Reagent (R2) containing lacosamide epitope labeled with bacterial G6PDH, labeled with bacterial G6PDH. buffer. bovine serum albumin, sodium azide, and buffer, bovine serum albumin, preservatives, and stabilizers stabilizers Methodology Homogeneous Enzyme Immunoassay (EIA) Same Platform Same Automated Clinical Chemistry Analyzer Required User Clinical Laboratories; Prescription Use Only Same Environment Liquid - Ready to use Reagents Form Same 2-8° C until expiration date Storage Same Analyte Topiramate Lacosamide

Comparison between the ARK™ Lacosamide Assay and the ARK™ Topiramate Assay

{6}------------------------------------------------

807.92 (b)(1) and 807.92 (b)(2): Brief Description of Nonclinical and Clinical Data

The following performance characteristics were obtained on the Beckman Coulter AU680® automated clinical chemistry analyzer.

Limit of Quantitation (LoQ)

The LoQ for the ARK Lacosamide Assay was determined to be 0.40 ug/mL, and may depend on analyzer-specific performance. The LoQ was determined according to CLSI EP17-A2 and is defined as the lowest concentration for which acceptable inter-assay precision (<20% CV) and recovery (±15%) is observed. Pooled human serum was supplemented with lacosamide to give concentrations of 0.30, 0.40 and 0.50 ug/mL. Eight (8) replicates of each sample were tested in each of five (5) runs to give a minimum of 40 replicates of each LoQ sample tested.

Nominal(ug/mL)Mean(ug/mL)SD% CV% RecoveryN
0.300.250.0135.083.240
0.400.350.0123.488.140
0.500.470.0163.594.140

Measurement Range

The measurement range of the ARK Lacosamide Assay is 0.40 - 24.00 ug/mL. Specimens containing lacosamide in higher concentrations (>24.00 µg/mL) may be assayed by dilution of the specimen into the measurement range for a quantitative result or otherwise reported as detected above the measurement range. Multiply the assay result by the dilution factor to obtain the concentration of lacosamide in the undiluted specimen.

Recovery

Analytical recovery throughout the measurement range was performed by adding concentrated lacosamide drug into human serum negative for lacosamide. A certified stock concentrate of highly pure lacosamide was added volumetrically to human serum negative for lacosamide, representing drug concentrations across the assay range. Two analytical runs of three replicates of each sample were assaved on an automated clinical chemistry analyzer. The results of the six replicates of each sample were averaged and compared to the target concentration and percent recovery calculated. Recovery at all samples tested was within ±10% of the expected sample concentration.

TheoreticalConcentration(µg/mL)Mean RecoveredConcentration(µg/mL)PercentRecovery (%)
0.400.3690.4
0.500.4793.3
1.001.04104.2

{7}------------------------------------------------

TheoreticalConcentration(µg/mL)Mean RecoveredConcentration(µg/mL)PercentRecovery (%)
3.003.07102.3
6.006.15102.6
9.008.9299.1
15.0014.4296.1
20.0021.15105.8

Linearity

Linearity studies were performed as suggested in CLSI EP06-A. A 30.00 ug/mL lacosamide serum sample was prepared and dilutions were made proportionally with human serum negative for lacosamide. Two analytical runs of three replicates of each sample were assayed on an automated clinical chemistry analyzer. The results of the six replicates of each sample were averaged. The data were analyzed using linear regression as well as non-linear fitted polynomial regression. Linearity at specific dilutions was considered acceptable if the percent difference was ±10% between the predicted 1st and 2nd order regressed values at concentrations >1.00 µg/mL, or ≤0.20 µg/mL at concentrations ≤1.00 µg/mL. A linear relationship was demonstrated between 0.40 and 25.00 µg/mL (y = 0.9998x - 0.0170).

Nominal(µg/mL)MeasuredResults(µg/mL)1st OrderPredictedResults2nd OrderPredictedResultsDifference
0.000.00-0.02-0.08NA
0.400.360.380.33-0.05 µg/mL
1.501.551.481.45-2.0 %
3.002.952.982.980.0 %
6.005.835.986.020.7 %
9.008.918.989.050.7 %
12.0012.0111.9812.050.6 %
15.0015.0214.9815.040.4 %
18.0018.1117.9818.010.2 %
21.0021.4120.9820.97-0.1 %
25.0024.5524.9824.87-0.4 %

{8}------------------------------------------------

Method Comparison

Method comparison studies were performed using CLSI EP09-A3 as a guideline. Method comparison was conducted with 150 unaltered, human serum specimens that are not individually identifiable. Results from the ARK Lacosamide Assay were compared with results from LC-MS/MS. The lacosamide concentrations ranged from 0.65 µg/mL to 23.50 µg/mL. Results of the Passing-Bablok regression analysis are shown below (with 95% confidence limits).

Slope1.01(0.99 to 1.04)
y-intercept0.03(-0.10 to 0.15)
Correlation Coefficient (r2)0.98(0.98 to 0.99)
Number of Samples150

Image /page/8/Figure/3 description: This image is a scatter plot comparing ARK Lacosamide Assay and LC-MS/MS measurements, both in micrograms per milliliter. The x-axis represents LC-MS/MS values ranging from 0 to 25, while the y-axis represents ARK Lacosamide Assay values in the same range. A cluster of data points forms a strong positive correlation, closely aligned with a blue trend line, indicating agreement between the two measurement methods. The data suggests a linear relationship between the two assays.

{9}------------------------------------------------

Precision

Precision was determined as described in CLSI EP05-A3. Tri-level controls and three samples of lacosamide in pooled human serum were used in the study. Data were collected on a single analyzer over twenty (20) non-consecutive days. Two (2) calibrations were performed according to requirements for quality control. Each level was assayed in quadruplicate twice a day for 20 days. Each of the runs per day was separated by at least two hours. The within run, between day, total SD, and percent CVs were calculated. Results are shown below. Acceptance criteria: ≤10% total CV.

SampleNMean(µg/mL)Within RunBetween DayTotal
SDCV(%)SDCV(%)SDCV(%)
ARK Lacosamide Control
LOW1601.550.0493.10.0493.10.0704.5
MID1607.130.2022.80.2042.90.2874.0
HIGH16014.940.4503.00.4453.00.6644.4
Human Serum
LOW1601.490.0453.00.0372.50.0583.9
MID1607.100.1752.50.2173.10.2834.0
HIGH16015.180.4563.00.4322.80.6574.3

Interfering Substances

Interference studies were conducted using CLSI EP07-A3 as a guideline. Clinically high concentrations of the following potentially interfering substances in serum with known levels of lacosamide (2.0 and 15.0 µg/mL) were evaluated. Two analytical runs of three replicates of each sample (6 replicates total) were assayed using the ARK Lacosamide Assay, along with a serum control of lacosamide. The mean results of lacosamide were calculated and the percentage recovery relative to the serum control mean result was determined. Measurement of lacosamide resulted in ≤10% error in the presence of interfering substances at the levels tested.

Percentage Recovery (%)
InterferingSubstanceInterferentConcentration2.0 µg/mLLacosamide15.0 µg/mLLacosamide
Albumin12 g/dL99.8101.7
Bilirubin - conjugated70 mg/dL97.396.5
Bilirubin - unconjugated70 mg/dL101.198.3
Cholesterol620 mg/dL95.8100.1
Gamma-Globulin12 g/dL103.598.5

{10}------------------------------------------------

Percentage Recovery (%)
InterferingSubstanceInterferentConcentration2.0 µg/mLLacosamide15.0 µg/mLLacosamide
Hemoglobin1000 mg/dL101.0101.6
Rheumatoid Factor1000 IU/mL97.396.8
Triglycerides1000 mg/dL97.996.2
Uric Acid30 mg/dL102.596.6

Specificity

Lacosamide is eliminated primarily from the systemic circulation by renal excretion and biotransformation. After oral and intravenous administration, approximately 95% of lacosamide administered is recovered in the urine and less than 0.5% in the feces. The major compounds excreted are unchanged lacosamide (approximately 40% of the dose), its O-desmethyl metabolite (approximately 30%), and a structurally unknown polar fraction (~20%). The plasma circulating levels of the major human metabolite, O-desmethyl-lacosamide, is approximately 10% of that of lacosamide. This metabolite has no known pharmacological activity.

The crossreactivity of O-desmethyl lacosamide metabolite (5.0 ug/mL or 30.0 ug/mL) in the ARK Lacosamide Assay was not clinically significant (< 3.0% crossreactivity). Lacosamide (2.00 ug/mL or 15.00 ug/mL in human serum) was tested in the absence or presence of metabolite at higher than expected concentrations of metabolite.

O-DesmethylLacosamide(µg/mL)Measured Lacosamide in Absence/Presence ofMetabolite (µg/mL)
Lacosamide(2.00 µg/mL)Lacosamide(15.00 µg/mL)
MetaboliteAbsentMetabolitePresentMetaboliteAbsentMetabolitePresent
5.02.182.23Not Tested
30.0Not Tested15.5116.40

Crossreactivity

The compounds listed below did not interfere with the ARK Lacosamide Assay when tested in the presence of lacosamide (2.00 µg/mL and 15.00 µg/mL). Levels tested were at or above maximum physiological or pharmacological concentrations. Lacosamide concentrations of samples containing interferent were compared to the lacosamide level in a normal serum control. No significant interference was observed with other anti-epileptic or co-administered drugs at the concentrations tested. Recoveries ranged from 90.9% to 109.5%, within 10% of the expected level.

{11}------------------------------------------------

#CompoundConcentrationTested(µg/mL)Percentage Recovery (%)
Lacosamide(2.0 µg/mL)Lacosamide(15.0 µg/mL)
1Acetaminophen200100.9100.1
2Acetazolamide100100.5100.9
3Acetylsalicylic acid1000102.4102.1
4Amikacin10096.599.7
5Amitriptyline2096.995.6
6Amoxapine1097.499.5
7Amphotericin B10098.699.7
8Ampicillin10097.7100.1
9Ascorbic acid100102.8104.0
10Baclofen10098.8100.3
11Buproprion10100.099.0
12Caffeine100102.4102.7
13Carbamazepine10099.0101.8
14Chloramphenicol250102.198.6
15Chlorpromazine1096.397.3
16Citalopram1097.596.0
17Clobazam10098.996.0
18Clonazepam1094.999.2
19Cyclosporin A40100.298.5
20Diazepam2097.898.2
21Digoxin1095.997.3
22Doxepin1096.999.2
23Erythromycin20099.6100.6
24Ethanol4000 (0.4%)98.698.6
25Ethotoin100103.9102.0
26Ethosuximide250103.298.4
27Felbamate250102.6100.7
28Fluoxetine1098.198.3
29Furosemide100100.7101.0
30Gabapentin20097.5102.3
31Gentamicin10098.298.5
32Haloperidol1098.8102.4
33Heparin200 U/mL98.698.9
34Ibuprofen50095.2100.9
35Imipramine1097.899.7
36Kanamycin A20096.699.1
37Lamotrigine40097.8100.0
38Levetiracetam40098.198.5
39Lidocaine10097.998.9
40Lincomycin1000105.0104.4
41Mephenytoin100100.5102.9
#CompoundConcentrationTested(µg/mL)Percentage Recovery (%)
Lacosamide(2.0 µg/mL)Lacosamide(15.0 µg/mL)
42Mesoridazine10100.998.7
43Methicillin250102.1100.1
44Naproxen600104.299.9
45Neomycin100098.997.4
46Niacin100102.499.2
47Nitrazepam2098.296.3
48Nortriptyline2099.295.9
49Olanzapine10101.999.2
50Oxcarbazepine100104.8101.1
51Paroxetine1099.099.9
522-phenyl-2-ethyl-malonamide (PEMA)1000104.5105.4
53Penicillin V100100.195.9
54Perphenazine100108.2102.7
55Phenobarbital20097.690.9
56Phenytoin200108.391.8
57Pregabalin20096.892.6
58Primidone100103.095.2
59Procainamide100103.298.7
60Prochloroperazine10100.497.3
61Ranitidine100102.6106.4
62Rifampin100103.299.1
63Risperidone10100.2103.1
64Sertraline100102.9103.7
65Spectinomycin10096.499.8
66Stiripentol100103.4101.8
67Sulfamethoxazole400103.1105.2
68Theophylline200100.6106.6
69Thioridazine10101.5103.0
70Tobramycin100100.5101.9
71Tiagabine200103.2103.4
72Topiramate250101.0100.2
73Trimethoprim40101.6102.1
74Valproic Acid600103.8103.9
75Vancomycin25099.1101.2
76Vigabatrin150100.4100.4
77Zonisamide40099.996.6

{12}------------------------------------------------

Sample Stability

Serum specimens were shown to be stable for at least forty-eight (48) hours at room temperature (22 °C), twenty-eight (28) days when refrigerated (2-8 °C), frozen (-20 °C) for at least 34 months, and after three (3) successive freeze/thaw cycles based on supporting data.

{13}------------------------------------------------

Stability

Accelerated stability studies and real time stability studies support a shelf-life stability claim of up to 18 months for the ARK Lacosamide Reagents when stored unopened at 2-8°C.

On-Board Stability

Reagents were stable up to 60 days when stored on-board the instrument based on supporting data.

Calibration Curve Stability

A stored calibration curve was effective up to at least 14 days based on supporting data. Calibration curve stability may depend on individual laboratory performance.

Traceability

ARK Lacosamide Calibrators are prepared by volumetric dilution of high purity lacosamide (certified solution traceable to HPLC) into a synthetic proteinaceous matrix free of lacosamide.

807.92 (b)(3): Conclusions from Nonclinical Testing

As summarized above, the ARK Lacosamide Assay is substantially equivalent to the legally marketed predicate device K083799.

§ 862.3350 Diphenylhydantoin test system.

(a)
Identification. A diphenylhydantoin test system is a device intended to measure diphenylhydantoin, an antiepileptic drug, in human specimens. Measurements obtained by this device are used in the diagnosis and treatment of diphenylhydantoin overdose and in monitoring levels of diphenylhydantoin to ensure appropriate therapy.(b)
Classification. Class II.