Search Results
Found 18 results
510(k) Data Aggregation
(174 days)
The ZeniCore™ Automatic Biopsy Instrument and optional Co-Axial Introducer Needle is a single-use device indicated for obtaining core samples of soft tissue masses via percutaneous needle biopsy. It is not intended for use in bone.
The ZeniCore™ Automatic Biopsy Instrument and optional, coaxial Introducer needle is a single-use medical device intended for obtaining core soft tissue samples via percutaneous needle biopsy. It is not intended for use in bone lesions. The ZeniCore™ Automatic Biopsy Instrument is a sterile, single patient use device comprised of a spring-loaded biopsy needle fitted into a plastic handle permitting single handed specimen collection than can be used with an optional corresponding coaxial Introducer Needle. The position of the needle may be visualized by imaging techniques such as X-ray, fluoroscopy, and ultrasound. The needle has an echogenic tip with numerically ordered centimeter markings to facilitate precise depth placement. The throw length of the needle when collecting a specimen is 25mm. There are five available needle gauge (G) sizes (12G, 14G, 16G, 18G and 20G) which are identified by a color-coded actuator button. The device is sold either standalone, or as a kit, with an optional, corresponding coaxial needle (cleared under K980004).
The optional Co-Axial Introducer Needle system includes an outer cannula and an inner stylet which terminate in a trocar tip and serves to support acquisition of tissue samples. Each Co-Axial Introducer Needle has an adjustable spring clip that serves as a locking mechanism to secure the position of the outer co-axial needle relative to the skin which ensures the biopsy needle does not accidentally slide forward or backward during the procedure, thereby maintaining the accuracy and safety of the biopsy. The Luer cap of the stylet is color coded to match the corresponding biopsy instrument it is sized to fit.
N/A
Ask a specific question about this device
(31 days)
The CLEANER™ Vac Thrombectomy System is indicated for the removal of fresh, soft thrombi and emboli from the vessels of the peripheral venous vasculature, and for the infusion of physician-specified fluids, including thrombolytics. The CLEANER™ Vac Thrombectomy System is not intended for use in the pulmonary vasculature for treating pulmonary embolism.
The CLEANER™ Vac Aspiration Canister is indicated as a vacuum source for the CLEANER™ Vac Thrombectomy System.
The CLEANER™ Vac Thrombectomy System is a single use device intended for the removal of fresh, soft emboli and thrombi and for the infusion of physician selected fluids through the side-port of the aspiration catheter.
The disposable system consists of:
- The CLEANER™ Vac Aspiration Catheter with Handpiece, and
- The CLEANER™ Vac Aspiration Canister (external vacuum reservoir with pump).
The Aspiration Catheter with handpiece includes a Dilator and may be placed over-the-wire to navigate the device to the target site. Once in the target site, to complete the system, the Aspiration Canister is connected to the handpiece. The dilator is removed, and the device is activated by the user to aspirate soft emboli and thrombi. The clot is aspirated from the distal portion of the device through the handpiece and then collected in the aspiration canister reservoir.
Additionally, the device may be used for infusion of thrombolytics and/or contrast media. Once thrombus resolution is achieved, the device is removed from the patient and discarded.
N/A
Ask a specific question about this device
(266 days)
The VariFuse Adjustable Infusion Catheter is indicated for the administration of fluids, including thrombolytic agents and contrast media, into the peripheral and pulmonary artery vasculature.
The VariFuse™ Adjustable disposable Infusion Catheter System consists of:
- An adjustable infusion catheter is comprised of the outer and inner catheters pre-assembled coaxially and inseparable, set at the maximum infusion length (i.e. 50 cm) with a Touhy Borst Adapter for shipping.
- The inner infusion catheter is designed with infusion holes spaced unevenly along the infusion length of 50 cm, a valve at the distal tip, and a radiopaque marker at the distal tip.
- The outer infusion catheter is designed to set the infusion length by covering the infusion holes and contains a radiopaque marker at the distal tip.
- A 1cc Luer Lock Syringe is included in the packaging for flushing
N/A
Ask a specific question about this device
(21 days)
The V•Stick™ Vascular Access Set is intended for use in the introduction and placement of guidewires and/or catheters.
The V•Stick™ Vascular Access Set consists of a 4F or 5F coaxial introducer set (with a 3F standard or 3F stiff dilator), a 21ga entry needle (with an echogenic or non-echogenic tip), and a 0.018" x 40cm Nitinol core and with Stainless Steel coil vascular access guidewire with an articulatable tip.
N/A
Ask a specific question about this device
(7 days)
The Option™ELITE Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava (IVC) in the following conditions:
· Pulmonary thromboembolism when anticoagulants are contraindicated
• Failure of anticoagulant therapy in thromboembolic disease
· Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced
· Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated
The Option™ELITE Filter may be removed according to the instructions supplied in the Section IX, entitled "Optional Procedure for Filter Retrieval" in patients who no longer require a filter can only be performed by the jugular approach.
The Option™ELITE Vena Cava Filter 100cm System is indicated for the prevention of recurrent pulmonary embolism (PE) via placement in the vena cava in the following conditions:
· Pulmonary thromboembolism when anticoagulants are contraindicated
· Failure of anticoagulant therapy in thromboembolic disease
· Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced
· Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated
The Option™ELITE Filter may be removed according to the instructions supplied in the Section IX, entional Procedure for Filter Retrieval" in patients who no longer require a filter in patients who no longer require a filter. Retrieval of the filter can only be performed by the jugular approach.
The Option™ELITE Vena Cava Filter is designed for the prevention of recurrent pulmonary embolism via percutaneous delivery in the inferior vena cava (IVC).
The Option™ELITE Vena Cava Filter 100cm System is designed for IVC filter insertion, delivery, deployment and placement via the popliteal and antecubital approach.
The self-centering Option™ELITE Filter is laser cut from nickel - titanium alloy (Nitinol) tubing. The Option™ELITE Filter consists of shape memory Nitinol struts emanating from a central location and is designed for optimal clot capture. Retention anchors (retention hooks) are located at the caudal portion of the filter. These anchors are intended for filter fixation to the vessel wall. The Option™ELITE Filter is intended to be used in caval diameters up to 30mm. A retrieval hook is centrally located at the cranial extremity.
The constrained Option™ELITE Filter is flexible and expands to the internal diameter of the IVC upon deployment. The Option™ELITE Filter imparts an outward radial force on the luminal surface of the vena cava to ensure proper positioning and stability. The Option™ ELITE Filter is designed to prevent pulmonary embolism while maintaining caval patency through central filtration.
The introduction kit is comprised of a filter housed in a filter cartridge. Catheter Sheath Introducer (5F ID). Angiographic Vessel Dilator with an open end, and a Pusher with deployment marker.
The Angiographic Vessel Dilator has side holes and 2 radiopaque markers, separated by 32mm (between the marker bands), that provide linear measurement of the inferior vena cava and assists in angiographic visualization when radiopaque contrast is delivered. The pusher advances the filter through the Catheter Sheath Introducer up to the deployment marker, and is then used to fix the filter in place during uncovering. The location of the distal end of the Catheter Sheath Introducer can be controlled by rotating the entire device to position the Catheter Sheath Introducer in the center of the vena cava. The Filter Cartridge houses the Option™ELITE Filter. The body of the Cartridge has text and colored arrows printed on it that identify assembly orientation, femoral is printed in green and jugular is printed in blue. The arrow of the desired access site will point into the Catheter Sheath Introducer hub. The Angiographic Vessel Dilator is designed to provide angiographic visualization and linear measurement of the vasculature when used in conjunction with the delivery of radiopaque contrast media to the vena cava.
This document describes a 510(k) premarket notification for the Option™ELITE Vena Cava Filter System and Option™ELITE Vena Cava Filter 100cm System. The purpose of the submission is to incorporate results from the PRESERVE study summary data into the Instructions for Use (IFU).
Here's an analysis of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document does not include a specific table of acceptance criteria or reported device performance metrics in the traditional sense of a performance study. This submission is for incorporating existing clinical study results (PRESERVE study) into the device's labeling, rather than presenting new performance data for a modified device. The document explicitly states:
"No new operating principles have been introduced with the subject device. The subject device operates using the identical fundamental scientific technology as the predicate device; therefore, no performance testing was necessary nor was any performed. The clinical testing was previously provided to FDA under the Investigational Device Exemption and the PRESERVE study."
Because no new performance testing was conducted for this specific 510(k) submission, there are no new device performance metrics to report against an acceptance criterion in this document. The existing performance would be based on the prior clearance of the predicate devices and the PRESERVE study.
2. Sample Size Used for the Test Set and Data Provenance
The document refers to the "PRESERVE study." However, it does not provide details on the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective) for this study within the provided text. To obtain this information, one would need to consult the original PRESERVE study documentation.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The provided document does not contain information on the number of experts used to establish ground truth or their qualifications for the PRESERVE study's test set.
4. Adjudication Method for the Test Set
The provided document does not specify any adjudication method (e.g., 2+1, 3+1, none) for the test set of the PRESERVE study.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
An MRMC comparative effectiveness study is typically used for AI-assisted diagnostic devices to assess the impact of AI on human reader performance. Since the device is a Vena Cava Filter System, it is a physical medical device, not a diagnostic imaging AI algorithm. Therefore, no MRMC comparative effectiveness study was done for this device.
6. Standalone (Algorithm Only) Performance
The device is a physical Vena Cava Filter System, not an algorithm or software-only device. Therefore, the concept of "standalone performance" for an algorithm does not apply to this device.
7. Type of Ground Truth Used
Given that this is a physical medical device (vena cava filter) and the PRESERVE study would be a clinical trial, the "ground truth" would likely be derived from patient outcomes, imaging data, and clinical assessments related to the filter's effectiveness in preventing pulmonary embolism and its safety profile. However, the document does not explicitly state the type of ground truth used for the PRESERVE study.
8. Sample Size for the Training Set
The concept of a "training set" typically applies to machine learning algorithms. Since the device is a physical medical device and not an AI/ML product, no training set was used in the context of algorithm development. Clinical trials, like PRESERVE, gather data for validation and effectiveness, not for training an algorithm.
9. How Ground Truth for the Training Set Was Established
As noted above, no training set was used in the context of an algorithm for this physical device. Therefore, how ground truth for a training set was established is not applicable.
In summary, the provided FDA 510(k) clearance letter and summary primarily focus on demonstrating substantial equivalence to a predicate device and incorporating pre-existing clinical study (PRESERVE study) data into the device's labeling. It does not present new performance data or detail the methodology of the PRESERVE study in the way one might expect for a new device's initial clearance or an AI/ML product. To get specific details about the PRESERVE study (sample size, ground truth, expert involvement, etc.), one would need to refer to its full study report or publications.
Ask a specific question about this device
(98 days)
The TLAB® Transvenous Liver Biopsy System is intended to be used for percutaneous transjugular and transfemoral venous liver access during diagnostic and interventional procedures.
The TLAB® Transvenous Liver Biopsy System is a single use, disposable, sterile device. The TLAB® Transvenous Liver Biopsy System consists of the following components: 18-Gauge (18Ga) Flexcore Biopsy Needle, 7 French (7Fr) Introducer Sheath with a curved metal stiffener, 5 French (5Fr) Straight Catheter, 5 French (5Fr) Curved Catheter, Tissue Removal Swabs, Bending Tool (used for transfemoral access).
Here's a breakdown of the acceptance criteria and study information for the TLAB® Transvenous Liver Biopsy System, based on the provided text:
This document describes a medical device (TLAB® Transvenous Liver Biopsy System) not an AI/ML device, therefore, sections related to AI/ML specific criteria (such as MRMC studies, standalone algorithm performance, training set details) are not applicable.
Acceptance Criteria and Reported Device Performance
Device Trade Name: TLAB® Transvenous Liver Biopsy System (TF-18C)
Regulation Number: 21 CFR 870.1340 - Catheter Introducer
Regulatory Class: Class II
Product Code: DYB
The device demonstrated substantial equivalence to predicate devices (TLAB® Transjugular Liver Access Set K022634 and Traveler Portal Vein Access Set K213638) based on a comparison of technological characteristics, indications for use, materials, packaging, principle of operation, design features, sterilization process, and in-vitro performance testing.
| Acceptance Criteria Category | Specific Criteria / Test Performed | Reported Device Performance / Outcome |
|---|---|---|
| Performance Testing | Dimensional | Met acceptance criteria |
| Visual | Met acceptance criteria | |
| Radiopacity | Met acceptance criteria | |
| Corrosion Resistance | Met acceptance criteria | |
| Simulated Use | Met acceptance criteria | |
| Tensile Strength | Met acceptance criteria | |
| Torque Strength | Met acceptance criteria | |
| Liquid Leakage | Met acceptance criteria | |
| Luer Functional Testing (ISO 80369-7) | Met acceptance criteria | |
| Catheter Functional Testing (ISO 10555-1) | Met acceptance criteria | |
| Resistance to Breakage (ISO 9626) | Met acceptance criteria | |
| Shipping Qualification Testing | Met acceptance criteria | |
| Design Validation | Met acceptance criteria | |
| Summative Usability | Met acceptance criteria | |
| Biocompatibility (ISO 10993-1:2020) | Cytotoxicity (ISO 10993-5) | Previously performed, applicable to new device; met criteria |
| Sensitization (ISO 10993-23) | Previously performed, applicable to new device; met criteria | |
| Irritation or Intracutaneous Reactivity (ISO 10993-10) | Previously performed, applicable to new device; met criteria | |
| Material Mediated Pyrogen (ISO 10993-11) | Previously performed, applicable to new device; met criteria | |
| Acute Systemic Toxicity (ISO 10993-11) | Previously performed, applicable to new device; met criteria | |
| Hemocompatibility (ISO 10993-4): ASTM Hemolysis Assay, Direct and Extract Methods (ISO) | Previously performed, applicable to new device; met criteria | |
| Hemocompatibility (ISO 10993-4): Complement Activation Assay, SC5b-9 Method with Comparison Article (ISO) | Performed anew; met criteria | |
| Hemocompatibility (ISO 10993-4): Partial Thromboplastin Time (PTT) Assay with Comparison Article (ISO) | Performed anew; met criteria | |
| Hemocompatibility (ISO 10993-4): Heparinized Platelet and Leukocyte Count Assay with Comparison Article (ISO) | Performed anew; met criteria | |
| Sterilization | Minimum SAL 10-6, EtO | Met acceptance criteria |
| Shelf-Life | 3 years | Met acceptance criteria (based on predicate) |
Study Details
-
Sample size used for the test set and the data provenance:
- The document primarily describes bench-top testing (in-vitro performance) and biocompatibility testing. Specific sample sizes for each non-clinical test are not explicitly stated in the provided text, but it mentions that "A series of testing was conducted".
- Data Provenance: All studies were performed following approved protocols under Good Laboratory Practices (GLP) in compliance with FDA GLP, 21 CFR Part 58. This indicates the testing was conducted in a controlled laboratory environment. No specific country of origin or retrospective/prospective nature for these non-clinical tests is given beyond GLP compliance.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This question is not applicable as the described studies are non-clinical, bench-top performance and biocompatibility tests. They do not involve human interpretation or subjective clinical "ground truth" that would require expert consensus. The acceptance criteria themselves serve as the 'ground truth' for these engineering and materials tests.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- This question is not applicable as the described studies are non-clinical, bench-top performance and biocompatibility tests. Adjudication methods are typically used in clinical studies or studies involving subjective human interpretation (e.g., image reading) where disagreement among experts needs resolution.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This question is not applicable as this document describes a conventional medical device (a biopsy system), not an AI/ML powered device. Therefore, no MRMC comparative effectiveness study was performed in the context of AI assistance.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- This question is not applicable as this document describes a conventional medical device (a biopsy system), not an AI/ML powered device. There is no algorithm to test in a standalone manner.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For Performance Testing: The "ground truth" is defined by the acceptance criteria established in accordance with protocols based on guidance and industry standards. These include established engineering and material science standards (e.g., ISO for luer, catheter, and breakage testing) that define acceptable performance.
- For Biocompatibility Testing: The "ground truth" is defined by the internationally recognized standards for biocompatibility (ISO 10993 series) and FDA Guidance, which specify test methods and acceptable limits for biological responses.
-
The sample size for the training set:
- This question is not applicable as this document describes a conventional medical device (a biopsy system), not an AI/ML powered device. There is no machine learning model that requires a training set.
-
How the ground truth for the training set was established:
- This question is not applicable as this document describes a conventional medical device (a biopsy system), not an AI/ML powered device.
Ask a specific question about this device
(107 days)
The Cleaner™ Plus 18F Thrombectomy System is indicated for mechanical de-clotting, aspiration, and controlled and selective infusion of physician-specified fluids, including thrombolytics, in the peripheral venous vasculature.
The Cleaner Plus™ 18F Thrombectomy System is a single use device used to provide thrombectomy in the peripheral venous vasculature. The device provides additional features, such as aspiration and over-the-wire device placement.
The disposable system consists of: (1) the Aspiration Catheter & Dilator, (2) the Handpiece that includes an aspiration control and an integrated Maceration Wire, and a Peel-Away Introducer and (3) the external vacuum reservoir with pump known as the Aspiration Canister.
The system is inserted percutaneously into the vessel using an introduction catheter, the system Aspiration Catheter and Dilator may be placed over-the-wire to the site of thrombus. The device macerates intra-lumen and wall adherent thrombus with a maceration wire. The macerated thrombus is removed from the vessel using an aspiration system. The aspiration of the clot can be performed simultaneous or independently.
Additionally, the device may be used for infusion of thrombolytics and/or contrast media. Once thrombus resolution is achieved, the device is removed from the patient and discarded.
The provided text describes information about the Cleaner™ Plus 18F Thrombectomy System, a medical device, and the basis for its FDA clearance through a 510(k) premarket notification (K233909). The information primarily focuses on establishing substantial equivalence to a predicate device (K211798), rather than detailing a study that proves the device meets specific acceptance criteria in the context of an AI/ML medical device.
Therefore, many of the requested items related to AI/ML device study design (e.g., sample size for test sets, expert ground truth establishment, MRMC studies, standalone algorithm performance, training set details) are not applicable or cannot be extracted from this document, as it concerns a mechanical thrombectomy system, not an AI/ML-driven device.
However, I can provide the available information regarding the device's performance assessment and the basis for its clearance.
Acceptance Criteria and Device Performance (Non-AI/ML Medical Device)
The document does not present a table of numerical "acceptance criteria" in the same way one might find for an AI/ML model's performance metrics (e.g., AUC, sensitivity, specificity thresholds). Instead, the acceptance criteria for this medical device are qualitative, based on established industry standards, guidance, and the successful outcome of various performance and safety tests demonstrating substantial equivalence to a predicate device.
The "reported device performance" refers to the successful completion and meeting of predefined parameters for each test, indicating the device performs as intended and is as safe and effective as the predicate.
1. Table of Acceptance Criteria and Reported Device Performance
As this is a mechanical device, not an AI/ML system, the "acceptance criteria" are the successful completion of various tests and the demonstration of substantial equivalence. The "reported device performance" refers to the outcomes of these tests.
| Test Category / Characteristic | Acceptance Criteria (Qualitative) | Reported Device Performance (Outcome) |
|---|---|---|
| Substantial Equivalence | Demonstrates comparable indications for use, mechanism of action, materials, technological characteristics, principle of operation, design features, and sterilization process to the predicate device, introducing no additional clinical risk. | Established based on in vitro performance testing, similarities in indications for use, materials, technological characteristics, principle of operation, design features, and sterilization process with the predicate device. Confirmed by in-vitro and in-vivo tests. Conclusion: Cleaner Plus 18F Thrombectomy System is substantially equivalent to the predicate device and introduces no additional clinical risk. |
| In-Vitro Performance Testing | Meet acceptance criteria based on requirements outlined in guidance and industry standards for each specific test. | Cleaner Plus 18F Thrombectomy System Performance: Visual, Dimensional, Simulated Use, Leak Testing.Handpiece Performance: Aspiration and Maceration Performance, Tensile and Torque, Kink Testing, Catheter Torque.Canister Performance: Simulated Use Maceration and Aspiration Performance.Maceration Wire Performance: Bond Strength, Resistance to Corrosion, Shape Retention, Flexing and Fracture, Kink Radius, Fatigue, Endurance, Torque, Tensile, Dimensional, Visual.Handpiece and Catheter Performance: Peel-Away Introducer Removal Force, Maceration Wire RPM, Wire to Coupler Tensile, Peel-Away Introducer Functionality, Tip Collapse.Other Tests: Particulates, Canister Performance (Visual, Functional, Tensile - joint strength), Luer Compliance, Pump Performance, IEC Compliance, Software Testing, Shipping Qualification. All tests met acceptance criteria. |
| Biocompatibility | Compliant with ISO 10993-1:2020 for an external communicating, circulating blood device with limited duration (<24hrs). | A Biological Risk Assessment was performed. Biocompatibility testing was appropriately leveraged from the predicate device (K211798) and a comparator device (K232679), performed following approved protocols under GLP and in compliance with FDA GLP, 21 CFR Part 58. |
| Animal Studies (In-vivo Safety & Performance) | Demonstrate substantially equivalent safety and performance outcomes, with no new questions of safety or effectiveness. Show no histological signs of thromboembolism, ruptures, hemorrhage, perforation, or dissection in treated vessels. | First animal study: Assessed safety and performance on vascular endothelium (vessel patency, clot burden, pulmonary embolism). No histological signs of thromboembolism, ruptures, hemorrhage, perforation, or dissection noted. All endpoints met.Second animal study: Assessed safety and performance of the aspiration aspect (vessel injury, patency, gross thrombogenicity). No histological signs of thromboembolism, ruptures, hemorrhage, perforation, or dissection noted. All endpoints met. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not explicitly stated numerically as a "test set" in the context of an AI/ML model for classification or prediction. The "test set" here refers to the actual physical devices (or components) used for the various in-vitro and in-vivo performance and safety tests. The specific number of devices/samples tested for each individual performance test (e.g., tensile strength, kink testing) is not provided in this summary.
- Data Provenance:
- In-vitro Performance Testing: Conducted by the manufacturer (Argon Medical Devices, Inc.)
- Biocompatibility Testing: Leveraging data from predicate device (K211798) and a comparator device (K232679). These studies were performed under Good Laboratory Practices (GLP) in compliance with FDA GLP, 21 CFR Part 58.
- Animal Studies: Conducted as part of the submission to demonstrate safety and performance. (Specific country/institution not mentioned, implied to be part of the submission package leveraged by the manufacturer).
- Retrospective or Prospective: All tests described are prospective, conducted specifically for this submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Not applicable for this type of device. The "ground truth" for a mechanical device like this is established through engineering and scientific measurement, and observable biological/physical outcomes, rather than expert interpretation of images or clinical data. Device performance is measured against established physical and safety parameters.
4. Adjudication Method for the Test Set
- Not applicable. Adjudication methods (e.g., 2+1, 3+1) are common in scenarios where human interpretation of complex data (like medical images) is being used to establish ground truth or evaluate diagnostic performance. For a mechanical device, performance is evaluated against objective, measurable criteria and laboratory test results.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
- No. This type of study is relevant for AI/ML devices where reader performance (e.g., radiologists, pathologists) is being evaluated with and without AI assistance. This document describes a mechanical device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not applicable. This relates to AI/ML algorithm performance.
7. The Type of Ground Truth Used
- Objective Engineering & Biological Measurements: The ground truth for this device is based on:
- Physical/Dimensional conformity: Verified dimensions and visual appearance.
- Functional performance: Successful operation during simulated use, aspiration and maceration efficiency, tensile strength, torque, kink resistance, etc.
- Biocompatibility: Established through ISO standards and GLP-compliant testing results (e.g., absence of cytotoxicity, sensitization, irritation).
- Biological safety and effectiveness (animal studies): Observable clinical and histological outcomes in vivo (e.g., vessel patency, absence of thromboembolism, rupture, hemorrhage).
- Compliance with industry standards: Meeting parameters set forth in standards like IEC 60601.
8. The Sample Size for the Training Set
- Not applicable. This document describes a mechanical device, not an AI/ML model that requires a "training set."
9. How the Ground Truth for the Training Set Was Established
- Not applicable. As there is no AI/ML training set, this question is not relevant.
Ask a specific question about this device
(140 days)
The Cleaner™ Pro Thrombectomy System is indicated for the removal of fresh, soft thrombi and emboli from the vessels of the peripheral venous vasculature, and for the infusion of physician-specified fluids, including thrombolytics. The Cleaner™ Pro Thrombectomy System is not intended for use in the pulmonary vasculature for treating of pulmonary embolism.
The Cleaner Pro Thrombectomy System is a single use device in the removal of fresh, soft emboli and thrombi and for the infusion of physician-selected fluids through the side-port of the aspiration catheter.
The disposable system consists of: (1) the Aspiration Catheter with Dilator, (2) the Handpiece with Flushing Adapter, and (3) the external vacuum reservoir with pump known as the Aspiration Canister.
The Aspiration Catheter with Dilator may be placed over-the-wire to navigate the device to the target site. Once in the target site, to complete the system, the Aspiration Canister is connected to the handpiece. The dilator is removed, and the device is activated by the user to aspirate soft emboli and thrombi. The clot is aspirated from the distal portion of the device through the handpiece and then collected in the aspiration canister reservoir.
Additionally, the device may be used for infusion of thrombolytics and/or contrast media. Once thrombus resolution is achieved, the device is removed from the patient and discarded.
The provided text is a 510(k) Summary for the Cleaner™ Pro Thrombectomy System. This document focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than proving the device meets specific acceptance criteria based on clinical outcomes or standalone AI algorithm performance.
Therefore, many of the requested details about acceptance criteria, test sets, ground truth establishment, expert adjudication, MRMC studies, and training sets are not present in this type of regulatory submission. This document describes the bench-top and animal performance testing to support the claim of substantial equivalence for a medical device, which is different from studies demonstrating efficacy of an AI algorithm or performance directly tied to a specific set of clinical acceptance criteria.
However, I can extract the information that is available:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not present a table of specific acceptance criteria with quantitative performance metrics for a clinical outcome study. Instead, it lists performance testing conducted to demonstrate substantial equivalence to the predicate devices. The acceptance criteria for these tests are generally that the device performs comparably to the predicate or meets established engineering/design specifications for safety and functionality.
| Test Category | Specific Tests Performed (examples) | Acceptance Criteria (Implied) | Reported Performance (Implied) |
|---|---|---|---|
| In-Vitro (Bench-top) Performance Testing | Catheter - Dimensional, Visual, Radiopacity, Hemostasis Valve Leak, System Leak, Burst, Kink Radius, Tensile Break, Torsional Break, Aspiration Tip Collapse. Dilator - Dimensional, Leak, Burst, Functional, Visual, Tensile Break, Radiopacity. Aspiration Canister - Visual, Functional, Tubing Tensile Break, Weld Strength. Handpiece - Functional, Simulate Use. System - Functional, Vacuum Decay, Noise Level, Integrity, Visual, Simulated Use. Shipping Qualification, IEC-60601 Compliance, Software Validation. | The device must meet predefined engineering specifications and perform comparably to predicate devices for these physical and functional characteristics, ensuring safety and intended function. | "The below were shown to meet the acceptance criteria that were determined to demonstrate substantial equivalence." (General statement that tests were successful) |
| Biocompatibility Testing | Cytotoxicity (ISO 10993-5), Sensitization (ISO 10993-10), Irritation or Intracutaneous Reactivity (ISO 10993-10), Material Mediated Pyrogen (ISO 10993-11), Acute Systemic Toxicity (ISO 10993-11), Hemocompatibility (ISO10993-4) - In-vitro Blood Assay Complement Activation, SC5b-9 Heparinized Platelet and Leucocyte Counts, Partial Thromboplastin Time (PTT), ASTM Hemolysis Assay (Direct and Extract Methods). | Compliance with ISO 10993 standards for the specified biological endpoints, ensuring the device is biocompatible for its intended use. | "Biocompatibility is established for the Cleaner Pro Thrombectomy System according to ISO 10993-1:2020 as an external communicating, circulating blood with limited duration <24hrs. All studies were performed following the approved protocol under Good Laboratory Practices (GLP) in compliance to FDA GLP, 21 CFR Part 58." (Overall claim of meeting biocompatibility requirements) |
| Animal Study | Evaluation for vessel injury, patency, and gross thrombogenicity in the target vasculature. | Demonstrating substantially equivalent safety and performance, specifically no vessel injury (ruptures, hemorrhage, perforation, dissection) or histological signs of thromboembolism. | "There were no histological signs of thromboembolism noted in the treated vessels, there were no ruptures, hemorrhage, perforation, or dissection noted histologically in the treated veins. All the end points of this study were met." |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set (Animal Study): The document states "An animal study was performed..." but does not specify the sample size (number of animals or procedures).
- Data Provenance: The animal study data provenance is "animal study," implying a prospective, controlled study within a laboratory setting. No country of origin for the animals or study location is explicitly stated, but it would typically be conducted under US regulations (GLP).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This document does not describe the use of human experts to establish ground truth for a test set in the context of an AI algorithm or image analysis. The animal study's "ground truth" would be established by veterinary pathologists or other trained personnel performing histological analysis and gross observation, but the number and specific qualifications are not mentioned.
4. Adjudication Method for the Test Set
Not applicable, as the document doesn't describe ground truth establishment by multiple human readers for a test set in the context of an AI/image analysis setting. The animal study results would likely be adjudicated if multiple observers interpreted the findings, but no such method is detailed.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, an MRMC comparative effectiveness study was not conducted or described. This type of study is typically done to assess the impact of an AI system on human reader performance, which is not the focus of this 510(k) submission for a mechanical thrombectomy device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
Not applicable. This device is a mechanical thrombectomy system, not an AI algorithm. Its performance is inherent in its physical and mechanical operation.
7. The Type of Ground Truth Used
For the animal study, the ground truth was based on post-procedural histological analysis of treated vessels and gross observation for signs of injury or thrombosis. For bench-top testing, the ground truth is against engineering specifications and physical measurements.
8. The Sample Size for the Training Set
Not applicable. This is a mechanical device, not an AI algorithm, so there is no "training set."
9. How the Ground Truth for the Training Set Was Established
Not applicable. As above, there is no training set.
Ask a specific question about this device
(57 days)
The 10F Sheath and Dilator Set is indicated for introduction of therapeutic devices into the vasculature, excluding coronary and neuro vasculature.
The 10F Sheath and Dilator Set is designed for single use and intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature. The 10F Sheath and Dilator Set consists of a 10F reinforced Introducer Sheath and a matching 10F radiopaque Dilator.
The 10F Introducer sheath is a coil reinforced sheath with a Platinum Iridium radiopaque marker band. The sheath has a hemostasis valve and a side port with stopcock. The 10F dilator is made of radiopaque, non-reinforced material with an atraumatic tip to minimize blood loss. The device does not include any coating.
The provided document describes a 510(k) premarket notification for the "10F Sheath and Dilator Set" and establishes its substantial equivalence to a predicate device based on in vitro performance testing and technological characteristics. It does not include information about an AI-powered device or a study proving that an AI device meets acceptance criteria. Thus, I cannot populate the requested table or answer the questions related to AI device performance.
However, I can extract the acceptance criteria and the study performed for the 10F Sheath and Dilator Set (a non-AI medical device).
1. Table of Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Radiopacity | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
| Visual | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
| Dimensional | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
| Simulative Use | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
| Leakage (Sheath and Dilator) | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
| Kink Test (System) | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
| Tensile (Sheath and Dilator) | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
| Luer testing | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
| Shipping Qualification | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
| Biocompatibility | According to ISO 10993-1:2020 for an external communicating, circulating blood device with limited duration <24hrs. | |
| - Cytotoxicity | (Implicit standard per ISO 10993-5) | Performed following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58. |
| - Sensitization | (Implicit standard per ISO 10993-10) | Performed following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58. |
| - Irritation or Intracutaneous Reactivity | (Implicit standard per ISO 10993-23) | Performed following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58. |
| - Material Mediated Pyrogenicity | (Implicit standard per ISO 10993-11) | Performed following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58. |
| - Acute Systemic Toxicity | (Implicit standard per ISO 10993-11) | Performed following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58. |
| - Hemocompatibility (In-vitro Blood Loop Assay) | (Implicit standard per ISO 10993-4) | Performed with comparison article following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58. |
| - Hemocompatibility (Complement Activation Assay, SC5b-9 Method) | (Implicit standard per ISO 10993-4) | Performed with comparison article following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58. |
| - Hemocompatibility (Heparinized Platelet and Leukocyte Count Assay) | (Implicit standard per ISO 10993-4) | Performed with comparison article following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58. |
| - Hemocompatibility (Partial Thromboplastin Time (PTT) Assay) | (Implicit standard per ISO 10993-4) | Performed with comparison article following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58. |
| - Hemocompatibility (ASTM Hemolysis Assay, Direct and Extract Methods) | (Implicit standard per ISO 10993-4) | Performed following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58. |
Study Information for 10F Sheath and Dilator Set:
Since this document pertains to a traditional medical device (sheath and dilator set) and not an AI/software device, the following points regarding AI studies are not applicable.
- Sample size used for the test set and the data provenance: Not applicable to this type of device. The tests mentioned are bench-top (in vitro) and biocompatibility tests on device components. No "test set" in the context of diagnostic performance, images, or patient data is described.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for diagnostic purposes is not involved.
- Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI device.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI device.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for the device's performance is established by meeting predefined engineering specifications, industry standards (e.g., ISO 10993), and regulatory requirements for physical and biological properties.
- The sample size for the training set: Not applicable. This device does not have a "training set" in the context of machine learning.
- How the ground truth for the training set was established: Not applicable.
Summary of the study conducted for the 10F Sheath and Dilator Set:
The study involved a series of non-clinical (bench-top) performance tests and biocompatibility tests. The non-clinical tests included:
- Radiopacity
- Visual inspection
- Dimensional analysis
- Simulative Use
- Leakage (Sheath and Dilator)
- Kink Test (System)
- Tensile (Sheath and Dilator)
- Luer testing
- Shipping Qualification
The biocompatibility testing was conducted according to ISO 10993-1:2020 for an external communicating, circulating blood device with limited duration (<24 hours). These tests included Cytotoxicity, Sensitization, Irritation or Intracutaneous Reactivity, Material Mediated Pyrogenicity, Acute Systemic Toxicity, and Hemocompatibility assays (In-vitro Blood Loop, Complement Activation, Heparinized Platelet and Leukocyte Count, Partial Thromboplastin Time (PTT), and ASTM Hemolysis). All biocompatibility studies were performed under Good Laboratory Practices (GLP) in compliance with FDA GLP, 21 CFR Part 58.
The results of these tests demonstrated that the Argon Medical Devices, Inc. 10F Sheath and Dilator Set met the acceptance criteria and was substantially equivalent to the predicate device (Flexor® Check-Flo® Introducer Set) in terms of safety and performance for its intended use.
Ask a specific question about this device
(24 days)
Single-Loop Snare Retrieval Kit and Triple-Loop Snare Retrieval Kit are intended for the percutaneous removal of retrievable inferior vena cava (IVC) filters that are no longer medically required, via jugular approach.
The Single-Loop Snare Retrieval Kit and Triple-Loop Snare Retrieval Kit are single use devices. The disposable system consists of: [A] 9F (ID) Inner Sheath, [B] 11F (ID) Outer Sheath, [C] 8F (OD) Dilator, [D] Hemostasis Valve with Sideport, [E] High Pressure Stopcock, [F] 7F [2.4mm] (OD) x 76cm Snare Catheter with Tuohy-Borst Y-Port Adapter, 20mm x 93cm Single-Loop Snare (fully expanded) or 30mm x 93cm Triple-Loop Snare (fully expanded) with Torque Handle. The snares have radiopaque loops and are preloaded in the snare catheter. The snare catheter, inner sheath, and outer sheath have a radiopaque marker band at the distal tip for enhanced fluoroscopic visualization.
The provided text describes a 510(k) premarket notification for a medical device, specifically retrieval kits for inferior vena cava (IVC) filters. The submission focuses on a modification to an existing device.
Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a 510(k) submission for a modification to an existing device, the "acceptance criteria" are not explicitly stated as numerical targets in a table. Instead, the acceptance criteria are implicit in meeting the performance standards of the predicate device and ensuring the modification doesn't negatively impact safety or effectiveness. The reported device performance is that all tests met their respective acceptance criteria.
| Acceptance Criteria (Implicit from Predicate & Safety) | Reported Device Performance |
|---|---|
| Material Properties (Outer Sheath Modification): | |
| Outer Sheath - Visual | Met acceptance criteria |
| Outer Sheath - Dimensional | Met acceptance criteria |
| Outer Sheath - Tensile Strength | Met acceptance criteria |
| Outer Sheath - Liquid Leakage | Met acceptance criteria |
| Outer Sheath - Simulated Use | Met acceptance criteria |
| Outer Sheath - Delamination | Met acceptance criteria |
| Biocompatibility (for modified material): | |
| Cytotoxicity (ISO 10993-5) | Met acceptance criteria |
| Sensitization (ISO 10993-10) | Met acceptance criteria |
| Irritation or Intracutaneous Reactivity (ISO 10993-23) | Met acceptance criteria |
| Material Mediated Pyrogen (ISO 10993-11) | Met acceptance criteria |
| Acute Systemic Toxicity (ISO 10993-11) | Met acceptance criteria |
| Hemocompatibility (ISO 10993-4): | Met acceptance criteria |
| - ASTM Hemolysis, Direct and Extract | Met acceptance criteria |
| - Complement Activation, SC5b-9 | Met acceptance criteria |
| - Partial Thromboplastin (PTT) | Met acceptance criteria |
| - Platelet and Leukocyte Count | Met acceptance criteria |
| - In vitro Blood Loop | Met acceptance criteria |
| Leveraged Non-Clinical Testing (from predicate K191758): | |
| Visual/Dimensional (Inner Sheath, Dilator, Delivery Catheter, Single-Loop Snare & Triple-Loop Snare) | Met acceptance criteria |
| Leak Test (Inner Sheath, Dilator and Delivery Catheter) | Met acceptance criteria |
| Burst Test | Met acceptance criteria |
| Pull Test/Tensile Strength (Inner Sheath, Dilator, Delivery Catheter, Single-Loop Snare & Triple-Loop Snare) | Met acceptance criteria |
| Corrosion Resistance (Single-Loop Snare & Triple-Loop Snare) | Met acceptance criteria |
| Torque Response (Snare Assembly) | Met acceptance criteria |
| Radiopacity | Met acceptance criteria |
| Luer Testing | Met acceptance criteria |
| Contrast Medium Injection | Met acceptance criteria |
| High Pressure Stopcock Testing | Met acceptance criteria |
| Flexural Modulus and Tip Flexibility Testing | Met acceptance criteria |
| Radial Force Testing | Met acceptance criteria |
| Design Validation Testing and Summative Usability Testing | Met acceptance criteria |
The document explicitly states: "Test results demonstrate that all acceptance criteria were met; therefore, the device meets the established product specifications."
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the exact sample sizes used for each individual non-clinical test. It only lists the types of tests performed. The data provenance is not explicitly mentioned in terms of country of origin, but it is implied to be from the manufacturer's internal testing. The tests are non-clinical (bench testing, biocompatibility) rather than human subject studies, so the terms "retrospective or prospective" are not applicable in this context.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable to this submission. The tests performed are non-clinical (mechanical, materials, biocompatibility) and do not involve expert interpretation or ground truth establishment in the way clinical studies with AI algorithms would. The "ground truth" for these tests is based on objective measurements against established engineering and biocompatibility standards (e.g., ISO and ASTM standards).
4. Adjudication Method for the Test Set
This is not applicable as the tests are non-clinical and do not involve human interpretation or adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This submission is for a physical medical device (retrieval kit) with a material modification, not an AI-powered diagnostic or interpretive device. The document explicitly states: "Clinical testing was not required for the determination of substantial equivalence."
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
A standalone performance study was not done. This is not an AI device.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the non-clinical tests is based on:
- Engineering specifications and standards: For mechanical and dimensional tests (e.g., tensile strength, leak test, burst test, radiopacity, Luer testing, flexural modulus, radial force).
- Biocompatibility standards: For biocompatibility tests (e.g., ISO 10993 series for cytotoxicity, sensitization, irritation, pyrogenicity, systemic toxicity, hemocompatibility).
- Simulated use conditions: For tests like simulated use and design validation.
8. The Sample Size for the Training Set
This is not applicable as this is not an AI device and therefore has no "training set."
9. How the Ground Truth for the Training Set was Established
This is not applicable as this is not an AI device and therefore has no "training set" or associated ground truth establishment process.
Ask a specific question about this device