Search Filters

Search Results

Found 36 results

510(k) Data Aggregation

    K Number
    K233537

    Validate with FDA (Live)

    Date Cleared
    2024-04-23

    (172 days)

    Product Code
    Regulation Number
    862.1355
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Libre 2 Sensor users:

    The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.

    The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.

    The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.

    Libre 2 Plus Sensor users:

    The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 2 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.

    The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.

    The System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The System can be used alone or in conjunction with these digitally connected devices for the purpose of managing diabetes.

    Libre 3 Sensor users:

    The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.

    The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.

    The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.

    Libre 3 Plus Sensor users:

    The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes in persons age 2 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.

    The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.

    The System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The System can be used alone or in conjunction with these digitally connected devices for the purpose of managing diabetes.

    Device Description

    The FreeStyle Libre 3 Continuous Glucose Monitoring System (hereinafter also referred to as 'FSL3 System') and FreeStyle Libre 2 Flash Glucose Monitoring System (hereinafter also referred to as the 'FSL2 System') are integrated continuous glucose monitoring (iCGM) systems that provide continuous glucose measurements every minute to provide glucose levels, trends, and real-time alarms capability to aid in the management of diabetes. The FSL2 and FSL3 Systems also provide configurable alarms designed to warn the user of Low Glucose, High Glucose or Signal Loss. The user may make treatment decisions based in part on the sensor glucose results provided by both Systems. The FSL2 and FSL3 Systems require a prescription and are intended for home use.

    The subject FSL3 System consists of a sensor (the FreeStyle Libre 3 Sensor (FSL3 Sensor) or FreeStyle Libre 3 Plus Sensor (FSL3 Plus Sensor)) and a primary display device (the FreeStyle Libre 3 Reader (FSL3 Reader) or the FreeStyle Libre App (FSL App) [iOS and Android] downloaded to a compatible phone). The FSL3 Reader and the FSL App do not interact with each other.

    The subject FSL2 System consists of a sensor (the FreeStyle Libre 2 Sensor (FSL2 Sensor) or FreeStyle Libre 2 Plus Sensor (FSL2 Plus Sensor)) and a primary display device (the FreeStyle Libre 2 Reader (FSL2 Reader) or the FSL App (iOS and Android) downloaded to a compatible phone). The FSL2 Reader and FSL App do not interact with each other.

    Both the FSL2 and FSL3 Systems are compatible with the Libre Data Sharing API cleared in K223537. The display device of the connected FSL2 or FSL3 Systems, which directly receives the data from the sensor, continues to serve as a primary display device for the glucose data and alarms.

    AI/ML Overview

    This document is a 510(k) summary from Abbott Diabetes Care Inc. regarding their FreeStyle Libre 3 Continuous Glucose Monitoring System and FreeStyle Libre 2 Flash Glucose Monitoring System. It outlines the device details, indications for use, comparison to predicate devices, and a summary of performance testing.


    Acceptance Criteria and Study Proving Device Meets Criteria

    The provided text focuses on the substantial equivalence of the new FreeStyle Libre 2 Plus and 3 Plus sensors and updated app features to their predicate devices, rather than establishing new acceptance criteria for the entire system's glucose monitoring accuracy. The document states that clinical performance and human factors were "established in the predicate device (K223435) and are not affected by the introduction of the FSL App in this 510(k)." This implies that the glucose accuracy and user interface performance metrics (e.g., MARD, % in Zones A+B, usability outcomes) for the core glucose monitoring function were previously accepted for the predicate devices and are assumed to hold true for the current submission due to the nature of the changes being primarily related to sensor life, age ranges, and app features.

    Therefore, the acceptance criteria and performance data discussed below are extrapolated from the types of performance studies conducted to demonstrate substantial equivalence for the modifications presented in this 510(k), particularly regarding safety and compatibility, and from the implied reliance on the predicate device's established clinical performance.

    Please note: This response does not contain specific quantitative acceptance criteria or precise performance metrics for glucose accuracy (e.g., MARD values) as these details are explicitly referred to as having been established in the predicate device (K223435) and are not re-evaluated in this submission. The focus of this 510(k) is on the changes from the predicate.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of this 510(k) (modifications to existing systems), the presented "acceptance criteria" are derived from the tests conducted to demonstrate that the changes do not adversely affect safety and effectiveness.

    CategorySpecific Test/FeatureAcceptance Criteria (Implied/Stated)Reported Device Performance/Conclusion
    SoftwareSoftware Verification and ValidationCompliant with IEC 62304 and FDA Guidance "Content of Premarket Submissions for Device Software Functions" (June 14, 2023) and "Multiple Function Device Products: Policy and Considerations" (July 29, 2020).Results of executed protocols met the acceptance criteria and therefore support that the system software is acceptable for its intended use.
    CybersecurityCybersecurity Risk ManagementCompliant with FDA guidance "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions'' (September 27, 2023). Appropriate risk mitigation for identified threats and vulnerabilities.Risk assessment performed, appropriate risk mitigation controls implemented and tested.
    Bench TestingRemoval of CT/MRI Contraindications, X-ray CautionFunctionality testing acceptance criteria met for sensor performance after exposure to CT, MRI (MR conditional), and X-ray.The test results showed all functionality testing acceptance criteria was met. (Previous mechanical, electrical, and functional testing established in the predicate device are not affected).
    App Feature: Auto-Display (FSL2)Automatic display of glucose results without user-initiated scan.Functionality: Glucose data automatically displays via BLE without user scan, while NFC scan option remains.Functionality confirmed. (Users may still perform NFC scan as an option, consistent with predicate FSL2 App behavior.)
    App Feature: Alarms EscalationIncremental volume increase for alarms (High, Low, Urgent Low, Signal Loss).Volume increases incrementally over 30 seconds until maximum volume, equal to predicate device alarm volume.Functionality confirmed: Alarms increase volume over 30 seconds to the maximum level of predicate device alarms.
    Core Glucose Monitoring PerformanceClinical Performance (e.g., Accuracy)Established in predicate device K223435. (No new evaluation specified in this 510(k) summary for these aspects).Established in the predicate device (K223437) and not affected by the introduction of the FSL App in this 510(k). (This implies that the existing performance data from K223435 is deemed sufficient for these modified devices).
    Human FactorsUsability and User Interface (new features)Established in predicate device K223435. (No new evaluation specified in this 510(k) summary for these aspects beyond software V&V).Established in the predicate device (K223437) and not affected by the introduction of the FSL App in this 510(k). (The app navigation change to "Bottom bar navigation menu" from "Side panel navigation", and the "App Stopped Alarm" for Android were likely part of software V&V and user acceptance testing, implied by overall software V&V acceptance).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not provide specific sample sizes for the detailed technical tests (Software V&V, Cybersecurity, Bench Testing for CT/MRI/X-ray). These types of tests typically involve controlled laboratory environments rather than human subject test sets in the traditional sense for clinical performance.

    For the core clinical performance (glucose accuracy, etc.), the document explicitly states these characteristics were "established in the predicate device (K223435) and are not affected." Therefore, any sample size for these fundamental performance aspects would refer to those used in the studies for the predicate device, which are not detailed in this 510(k) summary.

    Data Provenance: The document does not specify the country of origin for any testing data. The nature of the tests (Software V&V, Cybersecurity, Bench Testing) suggests they are likely conducted in a controlled environment, potentially in-house or by a certified testing facility. The "retrospective or prospective" nature is not specified, but these are typically prospective tests conducted as part of the development and verification process.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information (number of experts, qualifications, and ground truth establishment) is typically relevant for clinical studies involving expert adjudication (e.g., for diagnostic imaging devices). Given that the 510(k) summary indicates clinical performance was "established in the predicate device" and that the current submission primarily addresses technical and software improvements/modifications, there's no mention of a new clinical study involving multiple experts to establish ground truth for this specific submission.

    For the bench testing related to CT/MRI/X-ray compatibility, ground truth would be established by physical measurements and functional checks against engineering specifications, not by human experts.


    4. Adjudication Method for the Test Set

    Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies, particularly for diagnostic accuracy where multiple human readers interpret data. As explained above, this 510(k) focuses on technical and software changes, and refers back to the predicate device for clinical performance. Therefore, no information on an adjudication method for a test set is provided or expected in this context.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done, and it is not typically relevant for a CGM device. MRMC studies are primarily conducted for diagnostic imaging devices to assess how AI assistance impacts human reader performance (e.g., radiologists interpreting scans). CGM devices provide quantitative glucose measurements, and their performance is evaluated directly against a reference method (e.g., laboratory glucose measurements), not through human reader interpretation of complex images or cases.


    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, for the core glucose algorithm, a standalone performance evaluation is inherent to a CGM device. The document explicitly states:

    • Glucose Algorithm: "ADC Glucose Algorithm"
    • Location of Glucose Algorithm: "FSL2 System: Receiver", "FSL3 System: Sensor"

    This implies that the algorithm processes the sensor's raw electrical signals to calculate a glucose value and trend independently. The clinical performance of this algorithm (e.g., MARD against a reference method) would have been part of the studies for the predicate device (K223435). This 510(k) does not present new data for this, but rather relies on the predicate's established performance, stating that ADC (Abbott Diabetes Care) Glucose Algorithm is the "Same" for the subject device.


    7. The Type of Ground Truth Used

    For Continuous Glucose Monitoring (CGM) devices, the ground truth for establishing accuracy is typically obtained from laboratory reference methods for blood glucose concentration, such as:

    • YSI analysis (Yellow Springs Instrument): This is a widely accepted laboratory reference method for measuring glucose in blood or plasma samples.

    While not explicitly stated in this 510(k) summary for the ground truth method, it is the standard for CGM accuracy studies which would have been performed for the predicate device to which this submission refers for clinical performance.


    8. The Sample Size for the Training Set

    The document does not provide a sample size for the training set of the glucose algorithm. This information would typically be detailed in the original predicate device's submission which established the clinical performance. The current 510(k) refers to the "ADC Glucose Algorithm" as being the same as the predicate, indicating no retraining or significant algorithm changes that would necessitate reporting new training data.


    9. How the Ground Truth for the Training Set was Established

    Similar to the sample size, the specifics of how the ground truth was established for the training set would have been detailed in the original submission for the predicate device. For CGM algorithms, this involves comparing the sensor's raw electrical signals to paired blood glucose values obtained from a highly accurate laboratory reference method (like YSI), collected over time from a cohort of study participants under various glycemic conditions. This establishes the true glucose concentration against which the algorithm learns to convert sensor signals into accurate glucose readings.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223435

    Validate with FDA (Live)

    Date Cleared
    2023-04-13

    (150 days)

    Product Code
    Regulation Number
    862.1355
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 2 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.

    The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.

    The System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The System can be used alone or in conjunction with these digitally connected devices for the purpose of managing diabetes.

    The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes in persons age 2 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.

    The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.

    The System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The System can be used alone or in conjunction with these digitally connected devices for the purpose of managing diabetes.

    Device Description

    The FreeStyle Libre 3 Continuous Glucose Monitoring System (hereinafter also referred to as 'FSL3 System') and FreeStyle Libre 2 Flash Glucose Monitoring System (hereinafter also referred to as the 'FSL2 System') are integrated continuous glucose monitoring (iCGM) systems that provide continuous glucose measurements every minute to provide glucose levels, trends, and real-time alarms capability to aid in the management of diabetes. The FSL2 and FSL3 Systems also provide configurable alarms designed to warn the user of Low Glucose, High Glucose or Signal Loss. The user may make treatment decisions based in part on the Sensor glucose results provided by both Systems. The FSL2 and FSL3 Systems require a prescription and are intended for home use.

    The FSL3 System consists of the FreeStyle Libre 3 Sensor and either the FreeStyle Libre 3 Reader or the FreeStyle Libre 3 App (iOS and Android) downloaded to a compatible smartphone as a primary display device. The FSL3 Reader and FSL3 App do not interact with each other.

    The FSL2 System consists of the FreeStyle Libre 2 Sensor and either the FreeStyle Libre 2 Reader or the FreeStyle Libre 2 App (iOS and Android) downloaded to a compatible smartphone as a primary display device. The FSL2 Reader and FSL2 App do not interact with each other.

    Both the FSL2 and FSL3 Systems are compatible with the Libre Data Sharing API cleared in K223537. The display device of the connected FSL2 or FSL3 Systems, which directly receives the data from the Sensor, continues to serve as a primary display device for the glucose data and alarms.

    AI/ML Overview

    1. A table of acceptance criteria and the reported device performance:

    The document primarily focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed table of clinical acceptance criteria and specific numerical performance. However, it does state that "System accuracy was demonstrated to meet the iCGM special controls requirements per 21 CFR 862.1355." This regulation for Integrated Continuous Glucose Monitoring Systems outlines performance standards.

    Based on the document, the following aspects were evaluated:

    Acceptance Criteria CategoryReported Device Performance
    Accuracy (Clinical)Met iCGM special controls requirements per 21 CFR 862.1355.
    SterilitySterility Assurance Level (SAL) of 10^-6 achieved with minimum sterilization dose of 25 kGy, established by VDmax25 method and ISO 11137-1/11137-2.
    Shelf-Life, Packaging Integrity, ShippingSeal integrity, user accessibility, and device functionality met acceptance criteria after worst-case testing.
    Electrical SafetyCompliant with IEC 60601-1: 2005(r)2012, IEC 60601-1-6:2010+A1:2013, and IEC 60601-1-11:2015.
    Electromagnetic Compatibility (EMC)Compliant with IEC 60601-1-2 and IEC CISPR 11. Wireless coexistence testing confirmed functionality within acceptable limits in presence of common radiating electronic devices (AAMI TIR69, ANSI C63.27, FDA Guidance). Compliant with FCC Regulations Part 15.225 and Part 15.247, and FAA Advisory Circular RTCA DO-160.
    Mechanical EngineeringMet acceptance criteria for mechanical, electrical, and functional testing at system level and for individual Sensor Applicator components.
    BiocompatibilityDemonstrated biocompatibility per ISO10993-1 and FDA Guidance "Use of International Standard ISO 10993-1."
    Software Verification and ValidationMet acceptance criteria for executed protocols per IEC 62304 and FDA Guidance "Content of Premarket Submissions for Software Contained in Medical Devices."
    CybersecurityCybersecurity risk management documentation analyzed confidentiality, integrity, and availability of data, information, and software. Appropriate risk mitigation controls implemented and tested per October 2014 FDA Guidance.
    Human FactorsRisk analysis demonstrated design changes met usability requirements per ANSI/AAMI/IEC 62366, IEC 60601-1-6, and FDA Guidance "Applying Human Factors and Usability Engineering to Medical Devices."
    InteroperabilityIncorporated an approach for interoperability in alignment with FDA Guidance "Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices."

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the sample size for clinical test sets. For the "Clinical Performance" section, it only mentions that "System accuracy was demonstrated to meet the iCGM special controls requirements per 21 CFR 862.1355." This often implies a clinical trial or study, but specific numbers are not provided in this 510(k) summary. The provenance of the data (country of origin, retrospective/prospective) is also not detailed in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. For glucose monitoring systems, "ground truth" is typically established by laboratory reference methods (e.g., YSI glucose analyzer) rather than expert human interpretation.

    4. Adjudication method for the test set:

    This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable to the FreeStyle Libre Continuous Glucose Monitoring System. This device is a standalone diagnostic tool that provides glucose readings directly to the user (or digitally connected devices), not a system designed to assist human readers in interpreting medical images or data. Therefore, an MRMC study comparing human reader improvement with/without AI assistance would not be conducted for this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Yes, the device is inherently designed to operate in a standalone manner (algorithm only) to provide glucose readings. The "Clinical Performance" section implicitly refers to this standalone performance when stating "System accuracy was demonstrated to meet the iCGM special controls requirements per 21 CFR 862.1355." The FSL3 System's glucose algorithm resides in the Sensor, and for FSL2, it resides in the Receiver (App or Reader), generating glucose values without human real-time intervention for each reading.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    For Continuous Glucose Monitoring (CGM) systems, the ground truth for accuracy studies is typically established by laboratory reference methods, such as those using a YSI glucose analyzer, which provides highly accurate and precise blood glucose measurements. While not explicitly stated in this summary, this is standard practice for CGM device validation.

    8. The sample size for the training set:

    The document does not provide information about the training set size. This 510(k) summary focuses on the substantial equivalence and performance of the final device, not the specifics of its development or algorithm training.

    9. How the ground truth for the training set was established:

    This information is not provided in the document. Similar to the test set, it would typically involve comparison to highly accurate laboratory reference methods.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222447

    Validate with FDA (Live)

    Date Cleared
    2023-03-03

    (200 days)

    Product Code
    Regulation Number
    862.1355
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 2 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.

    The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.

    The System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The System can be used alone or in conjunction with these digitally connected devices for the purpose of managing diabetes.

    The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes in persons age 2 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.

    The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.

    The System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The System can be used alone or in conjunction with these digitally connected devices for the purpose of managing diabetes.

    Device Description

    The FreeStyle Libre 2 Flash Glucose Monitoring System (with modified Sensor tail), hereinafter also referred to as 'modified FSL2 System' and FreeStyle Libre 3 Continuous Glucose Monitoring System (with modified Sensor tail), hereinafter also referred to as 'modified FSL3 System' are integrated continuous glucose monitoring (iCGM) systems that provide continuous glucose measurements every minute to provide glucose levels, trends, and real-time alarms capability to aid in the management of diabetes. The Systems also provide configurable alarms designed to warn the user of Low Glucose or Signal Loss. The user may make treatment decisions based in part on the Sensor glucose results provided by the Systems require a prescription and are intended for home use.

    The FreeStyle Libre 2 System consists of the FreeStyle Libre 2 Sensor and either the FreeStyle Libre 2 Reader or the FreeStyle Libre 2 App downloaded to a compatible smartphone as a primary display device. The FreeStyle Libre 3 System consists of the FreeStyle Libre 3 Sensor and the FreeStyle Libre 3 App downloaded to a compatible smartphone as a primary display device.

    Both the FreeStyle Libre 2 and FreeStyle Libre 3 Systems are compatible with the Libre Data Sharing API cleared in K223537. The display device of the connected FreeStyle Libre 2 or FreeStyle Libre 3 Systems, which directly receives the data from the Sensor, continues to serve as a primary display device for the glucose data and alarms.

    AI/ML Overview

    The Abbott FreeStyle Libre 2 Flash Glucose Monitoring System and FreeStyle Libre 3 Continuous Glucose Monitoring System sought clearance for modified sensor tails to reduce vitamin C interference, extend sensor wear to 15 days, and enable compatibility with automated insulin dosing (AID) systems.

    Here's an analysis of the acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria & Reported Device Performance:

    The document doesn't explicitly list a table of "acceptance criteria" with specific numeric targets for device performance. Instead, it states that the devices meet the iCGM special controls for clinical performance set forth in 21 CFR 862.1355. This regulation broadly outlines requirements for accuracy, reliability, and safety of integrated Continuous Glucose Monitoring systems.

    However, based on the provided text, we can infer some key performance objectives and the reported outcome:

    Acceptance Criteria (Inferred from Changes & Regulations)Reported Device Performance
    Accuracy across measuring range (15-day wear)The clinical study demonstrated that the subject devices meet the iCGM special controls for clinical performance. Specific accuracy metrics (e.g., MARD) are not provided in this summary but would be part of the full submission.
    Reduced Vitamin C InterferenceAn additional clinical study evaluated the effects of high doses of vitamin C, and the results showed the devices met iCGM special controls. Labeling was updated to reflect Vitamin C interference at doses >1000 mg/day (improved from >500 mg/day for the predicate).
    Extended Sensor Wear Duration (from 14 to 15 days)The subject devices (modified FreeStyle Libre 2 and 3 Sensors) are designed for a 15-day wear duration, and performance was evaluated throughout this duration in the clinical study.
    Compatibility with Automated Insulin Dosing (AID) systemsInteroperability testing and an interoperability communication plan were provided to ensure secure and reliable data transfer with digitally connected devices, including AID systems.
    Software Functionality and SafetySoftware regression testing was conducted in accordance with established specifications and IEC 62304. Results met acceptance criteria.
    Non-clinical PerformanceNon-clinical performance testing was conducted as per internal design control process requirements, and results met acceptance criteria.
    Known Performance of Unaffected AspectsSterilization, biocompatibility, shelf life stability, packaging integrity/shipping integrity, mechanical engineering, electrical safety and electromagnetic compatibility, and human factors were established in predicate/reference devices and are not affected. Cybersecurity for FSL3 also relies on existing data.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The clinical study evaluated performance in adult (18 years and older) and pediatric (2 to 17 years) participants with diabetes. Specific numbers of participants within these age groups are not provided in this summary.
    • Data Provenance: The document does not specify the country of origin. It indicates the studies were clinical studies, implying a prospective nature.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    The document mentions that the clinical study evaluated performance with respect to reference venous plasma sample YSI measurements.

    • Ground Truth Providers: YSI measurements are obtained from laboratory instruments, which are considered the gold standard for glucose measurement. The "experts" involved would be trained lab technicians and medical professionals responsible for collecting and processing blood samples and operating the YSI instrumentation.
    • Qualifications: Not explicitly stated, but they would be certified laboratory personnel and healthcare providers responsible for patient care and sample collection within a clinical study setting.

    4. Adjudication Method for the Test Set:

    Not applicable. The ground truth was established by YSI laboratory measurements, not subjective expert review or adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    A Human-in-the-loop or MRMC study is not applicable for this device type (Continuous Glucose Monitoring System). The device provides continuous glucose values, and its performance is assessed against a laboratory reference, not human interpretation of images or other data that would involve multiple readers.

    6. Standalone Performance:

    Yes, the studies described are primarily focused on the standalone performance of the algorithm (as embedded in the device) in measuring glucose levels against a reference method (YSI venous plasma samples). The "clinical study" directly assesses how well the device's sensor and algorithm perform without direct human intervention in the glucose reading process itself, other than the placement and use of the device by participants.

    7. Type of Ground Truth Used:

    The ground truth used was venous plasma sample YSI measurements. This is a highly accurate and accepted laboratory reference method for glucose concentration.

    8. Sample Size for the Training Set:

    The document does not specify the sample size for the training set used for the device's glucose algorithm. It mentions that the "ADC Glucose Algorithm established for the predicate device" is used, implying that the core algorithm was developed and trained previously, and this submission focuses on modifications and re-validation.

    9. How the Ground Truth for the Training Set Was Established:

    The document does not specify how the ground truth for the training set of the ADC Glucose Algorithm was established. However, given that the clinical performance evaluation uses venous plasma YSI measurements as ground truth, it is highly probable that similar high-accuracy reference methods would have been employed during the development and training of the original algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223537

    Validate with FDA (Live)

    Date Cleared
    2023-02-21

    (90 days)

    Product Code
    Regulation Number
    862.1355
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.

    The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.

    The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.

    The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.

    The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.

    The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.

    Device Description

    The FreeStyle Libre 2 and FreeStyle Libre 3 are integrated continuous glucose monitoring (iCGM) Systems designed to be used alone or in conjunction with digitally connected devices. The FreeStyle Libre 2 System consists of a Sensor and either a Reader or the FreeStyle Libre 2 App downloaded to a compatible smartphone as a primary display device. The FreeStyle Libre 3 System consists of a Sensor and the FreeStyle Libre 3 App downloaded to a compatible smartphone as a primary display device. Both Systems can communicate glucose data and other information wirelessly and securely to and from these digitally connected devices as described below:

    • Wireless communication from the FreeStyle Libre 2 Sensor or FreeStyle Libre 3 Sensor directly to an interoperable receiver device, which connects with the Sensor using the near field communication (NFC) and Bluetooth Low Energy wireless interfaces provided by the Sensor
    • The FreeStyle Libre 2 App or FreeStyle Libre 3 App communicates through the cloud to another software device, such as LibreView.

    Compared to the respective predicate devices, the proposed subject devices include an additional software component, the Libre Data Sharing API. The Libre Data Sharing API is a cloud-based application programming interface (API) that enables communication of glucose data including alarms through the cloud from the FreeStyle Libre 2 System or FreeStyle Libre 3 System to authorized client software on digitally connected devices. The data transmitted by the API to authorized client software can be used for specific and permitted use cases, including nonmedical device applications, medical device data analysis, CGM secondary display alarm, active patient monitoring, and treatment decisions. Use of the Libre Data Sharing API and the CGM information it transmits is limited by the indications for use of the iCGM systems with which it is used.

    The Libre Data Sharing API does not have any command or control over the client software, nor does it allow for the client software to have any command or control over the FreeStyle Libre 2 or FreeStyle Libre 3 Systems. Additionally, glucose data and alarms from the connected iCGM system are not modified or manipulated by the Libre Data Sharing API through its transmission to the authorized client software.

    The display device of the connected FreeStyle Libre 2 or FreeStyle Libre 3 Systems, which directly receives the data from the Sensor, continues to serve as a primary display device for the glucose data and alarms. The current components of the FreeStyle Libre 2 and FreeStyle Libre 3 Systems (sensor/applicator and primary display devices) have not been modified as a result of the added the Libre Data Sharing API.

    AI/ML Overview

    This document is a 510(k) Summary for the FreeStyle Libre 2 Flash Glucose Monitoring System and FreeStyle Libre 3 Continuous Glucose Monitoring System. It describes the devices and argues for their substantial equivalence to previously cleared predicate devices.

    Here's an analysis of the acceptance criteria and study information provided in the document:

    1. Acceptance Criteria and Reported Device Performance:

    The document doesn't present a table of numerical acceptance criteria with corresponding performance statistics. Instead, it argues for substantial equivalence based on the technological characteristics and intended use being the same as the predicate devices, with the addition of the Libre Data Sharing API.

    The key acceptance criteria, implicitly, are that the devices continue to perform according to specifications and meet their technological and performance criteria, as demonstrated by verification and validation, aligning with the predicate devices' cleared performance.

    Acceptance Criteria Category (Implicit)Reported Device Performance
    Technological CharacteristicsIdentical to predicate devices (amperometric measurement, glucose oxidase chemical reaction).
    Intended Use / Indications for UseSame as predicate devices. The Libre Data Sharing API enables communication of iCGM data for specific, permitted use cases, but does not change the core indications.
    Primary Display Device FunctionUnchanged. Continues to act as the primary display and issue glucose alarms.
    Safety and EffectivenessNo impact from the Libre Data Sharing API. The API has no command/control over client software or the iCGM systems, and glucose data/alarms are not modified during transmission.
    Adherence to Special ControlsConforms to iCGM special controls per 21 CFR 862.1355.

    2. Sample Size and Data Provenance:

    The document states: "The proposed subject devices with the Libre Data Sharing API were verified and validated according to ADC's internal design control process and in accordance with the applicable special controls for integrated continuous glucose monitoring systems. The testing demonstrated that the subject devices conform to the iCGM special controls per 21 CFR 862.1355 and that they performed according to specifications and met their technological and performance criteria."

    • Sample Size: The document does not specify the sample sizes used for the test set or the training set. It refers broadly to "performance testing" and "data provided in this pre-market notification."
    • Data Provenance: The document does not explicitly state the country of origin or whether the studies were retrospective or prospective. Given it's a 510(k) submission for a medical device, it's highly likely that the underlying performance data comes from clinical studies, often multi-site and prospective, but this document does not detail them. The focus of this 510(k) is specifically on the added software component (Libre Data Sharing API) and its impact on substantial equivalence, rather than re-proving the core glucose monitoring accuracy which would have been established in previous 510(k) clearances.

    3. Number of Experts and Qualifications for Ground Truth:

    The document concerns the addition of a data sharing API to existing glucose monitoring systems. It does not mention the use of experts to establish ground truth in the context of a reader study for the new component. The primary ground truth for glucose monitoring devices is typically derived from lab reference methods (e.g., YSI analyzer for blood glucose), rather than expert adjudication of images or clinical assessments.

    4. Adjudication Method for the Test Set:

    Not applicable in the context of this submission. The submission is about an API for device data sharing, not a diagnostic imaging or clinical assessment device requiring expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC study was not conducted or mentioned. This type of study is relevant for AI-assisted diagnostic tools that impact human reader performance, typically in radiology or pathology. This submission is for a continuous glucose monitoring system and an API to share its data.

    6. Standalone (Algorithm Only) Performance:

    The document states: "The Libre Data Sharing API does not have any command or control over the client software, nor does it allow for the client software to have any command or control over the FreeStyle Libre 2 and FreeStyle Libre 3 Systems. Additionally, glucose data and alarms from the FreeStyle Libre 2 and FreeStyle Libre 3 Systems are not modified or manipulated by the Libre Data Sharing API through its secured transmission to the authorized client software."

    This implies that the core algorithm performance (i.e., glucose measurement accuracy) of the FreeStyle Libre 2 and 3 systems, which operates independently of the Libre Data Sharing API, has already been established in previous clearances (K210943, K213996, K212132). This submission focuses on the safety and effectiveness of the data sharing mechanism, not the standalone performance of the glucose measurement algorithms themselves. The API itself is primarily a data transmission layer, not a
    diagnostic algorithm.

    7. Type of Ground Truth Used:

    For the underlying glucose monitoring systems (FreeStyle Libre 2 and 3), the ground truth for performance studies is typically laboratory reference measurements (e.g., plasma glucose values obtained from a YSI glucose analyzer) against which the accuracy of the interstitial glucose readings are compared. The document does not specifically detail how the ground truth was established for the "performance testing" related to the API, as the API's function is data transmission, not direct measurement. The ground truth for the API's performance would likely focus on the integrity, security, and timeliness of data transfer.

    8. Sample Size for the Training Set:

    The document does not specify the sample size for any training set. This submission is for a device modification (adding an API), not a de novo AI/ML algorithm that typically undergoes distinct training and test phases. The core glucose algorithms were already developed and cleared.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as this submission pertains to an API for data sharing and refers to "verification and validation" for "performance testing" of the modified device, rather than the development and training of a new algorithm requiring a specific ground truth establishment for a training set. The underlying glucose measurement algorithms would have had their own ground truth established during their prior development and clearance, likely using reference lab methods.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213996

    Validate with FDA (Live)

    Date Cleared
    2022-05-26

    (157 days)

    Product Code
    Regulation Number
    862.1355
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.

    The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.

    The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.

    Device Description

    The FreeStyle Libre 3 Continuous Glucose Monitoring System (herein referred to as the 'FreeStyle Libre 3 System' or 'System') is an integrated continuous glucose monitoring system (iCGM) that provides real time continuous glucose measurements every minute to provide glucose levels, trends, and alarms. The System requires a prescription and is intended for home use. The System consists of the following components: a Sensor which transmits via Bluetooth Low Energy (BLE), and a mobile application, FreeStyle Libre 3 App, downloaded to a compatible smartphone running on Android operating system. The FreeStyle Libre 3 System provides the user with real-time glucose measurements (glucose values) accompanied by trend information (glucose arrows) and historical glucose information (glucose graph). The user may make treatment decisions based in part on the Sensor glucose results provided by the System. The System also provides fixed and configurable alarms designed to warn the user of Low Glucose, High Glucose, or Signal Loss.

    AI/ML Overview

    The document details the FreeStyle Libre 3 Continuous Glucose Monitoring System and its performance testing to support substantial equivalence to its predicate device, the FreeStyle Libre 2 Flash Glucose Monitoring System.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that a bridging clinical study was conducted, and the "combined System accuracy of the FreeStyle Libre 3 System and FreeStyle Libre 2 System met the iCGM special controls requirements per 21 CFR 862.1355." However, specific numerical acceptance criteria (e.g., MARD percentage, accuracy at different glucose ranges) and the exact reported performance of the FreeStyle Libre 3 System are not explicitly listed in a detailed table within the provided text segments. Instead, a general statement of meeting the iCGM special controls is given.

    To fill this, we must infer that the implicit acceptance criteria are those defined by the iCGM special controls (21 CFR 862.1355) for integrated continuous glucose monitoring systems. These typically include accuracy metrics like MARD (Mean Absolute Relative Difference) against a reference method, as well as performance across different glucose ranges (hypoglycemic, euglycemic, hyperglycemic). Without the actual study report, we cannot provide specific numbers for the FreeStyle Libre 3's performance or exact acceptance criteria.

    General Interpretation of iCGM Special Controls (Implicit Acceptance Criteria):
    iCGM special controls generally require demonstration of:

    • Accuracy: Sufficient accuracy across the glucose range (e.g., MARD values below a certain threshold when compared to a laboratory reference).
    • Safety: Acceptable safety profile with minimal adverse events.
    • Performance in various conditions: Performance evaluated across different patient populations, wear sites, and duration of use.
    • Alarm performance: Accuracy and reliability of hypoglycemia and hyperglycemia alarms.
    • Interoperability: Functionality with digitally connected devices.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "ADC conducted a bridging clinical study to demonstrate comparability of the performance of the FreeStyle Libre 3 System to the predicate FreeStyle Libre 2 System, cleared under K210943."

    • Sample Size for Test Set: Not explicitly stated in the provided text.
    • Data Provenance: Not explicitly stated in the provided text (e.g., country of origin, retrospective or prospective). It is implied to be prospective as it's a clinical study for FDA clearance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the text. For CGM devices, the "ground truth" (reference glucose values) is typically established using a highly accurate laboratory reference method (e.g., YSI glucose analyzer) on blood samples, not expert consensus interpreting images or clinical cases. Therefore, the concept of "experts establishing ground truth" in the way it applies to image interpretation is not directly applicable here. The ground truth would be the laboratory-measured glucose values.

    4. Adjudication Method for the Test Set

    The concept of an "adjudication method" (like 2+1 or 3+1) is primarily relevant for studies where multiple human readers interpret data (e.g., radiology images) and a consensus or tie-breaking mechanism is needed for the ground truth. For a CGM device, the ground truth is established by a quantitative, objective measurement (laboratory blood glucose). Therefore, an adjudication method in this context is not applicable/not mentioned as it's not a reader study.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. The FreeStyle Libre 3 is a standalone device that measures glucose; it does not involve human readers interpreting its output in a diagnostic context that would warrant an MRMC study. The "comparative effectiveness" mentioned refers to its performance against a predicate device and standard laboratory methods, not a comparison of human reader performance with and without AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, a standalone study was done. The entire premise of the submission is to demonstrate the performance of the FreeStyle Libre 3 System as a standalone integrated continuous glucose monitoring device. The bridging clinical study, designed to show "comparability of the performance of the FreeStyle Libre 3 System to the predicate FreeStyle Libre 2 System," inherently evaluates the algorithm's performance without direct human intervention in the glucose measurement and calculation process. The FreeStyle Libre 3 delivers "real-time continuous glucose measurements every minute" and its "System accuracy" was evaluated.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for a CGM device is typically a highly accurate laboratory reference method (e.g., YSI glucose analyzer) performed on venous blood samples taken concurrently with the CGM readings. While not explicitly stated, this is the standard for CGM accuracy studies.

    8. The Sample Size for the Training Set

    The document mentions that the "Sensor Glucose Algorithm established for the predicate device" is the same for the FreeStyle Libre 3. It also states that the FreeStyle Libre 3 contains a "software algorithm for conversion of the raw glucose measurements from the Sensor to calculate glucose results." This implies the algorithm was developed and trained prior to this submission, likely using data from previous FreeStyle Libre sensor generations.

    The sample size for the training set is not provided in this document.

    9. How the Ground Truth for the Training Set was Established

    Given that the "Sensor Glucose Algorithm" is the same as the predicate device, it can be inferred that the ground truth for training this algorithm was established similarly to the predicate device's clearance. This would typically involve pairing sensor readings with laboratory reference blood glucose measurements (e.g., YSI) taken from human subjects during previous clinical studies. The specific details of how this ground truth was established for the training set are not provided in the given text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212132

    Validate with FDA (Live)

    Date Cleared
    2022-05-26

    (322 days)

    Product Code
    Regulation Number
    862.1355
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.

    The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.

    The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.

    Device Description

    The FreeStyle Libre 3 Continuous Glucose Monitoring System (herein referred to as the 'FreeStyle Libre 3 System' or 'System') is an integrated continuous glucose monitoring system (iCGM) that provides real time continuous glucose measurements every minute to provide glucose levels, trends, and alarms. The System requires a prescription and is intended for home use. The System consists of the following components: a Sensor which transmits via Bluetooth Low Energy (BLE), and a mobile application, FreeStyle Libre 3 App, downloaded to a compatible smartphone running iOS operating system. The FreeStyle Libre 3 System provides the user with real-time glucose measurements (glucose values) accompanied by trend information (glucose arrows) and historical glucose information (glucose graph). The user may make treatment decisions based in part on the Sensor glucose results provided by the System. The System also provides fixed and configurable alarms designed to warn the user of Low Glucose, High Glucose, or Signal Loss.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the FreeStyle Libre 3 Continuous Glucose Monitoring System, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't provide a specific table of numerical acceptance criteria for clinical performance (like MARD values with defined thresholds). Instead, it states that the device "met the iCGM special controls requirements per 21 CFR 862.1355." However, it does list several performance characteristics that were "confirmed to support substantial equivalence," implying that specific acceptance criteria were met for each. Without the actual criteria for each, here's a general table based on the information provided:

    Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    SterilitySterility Assurance Level (SAL) of 10^-6Achieved SAL of 10^-6 with minimum sterilization dose of 25 kGy (established by VDmax25 method described in ISO 11137-2).
    Shelf-Life & PackagingSeal integrity, user accessibility, and device functionality metAttributes related to seal integrity, user accessibility, & device functionality (including sterile barrier system integrity) met acceptance criteria after subjecting units to worst-case sealing, sterilization, shipping, storage, handling, and transit challenges.
    Electrical SafetyCompliance to IEC 60601-1, IEC 60601-1-6, IEC 60601-1-11Demonstrated compliance to IEC 60601-1: 2005(r)2012, IEC 60601-1-6:2010+A1:2013, and IEC 60601-1-11:2015.
    ElectromagneticWithstand EMI and emissions (IEC 60601-1-2, CISPR 11)Verified system withstands EMI and emissions in compliance with IEC 60601-1-2 and IEC CISPR 11. Wireless coexistence testing confirmed functionality within acceptable limits in presence of common radiating electronic devices (FDA Guidance, AAMI TIR69, ANSI C63.27). Complied with FCC Part 15.225, Part 15.247, and FAA Advisory Circular RTCA DO-160.
    Mechanical EngineeringMechanical, electrical, and functional testing metTest results showed mechanical, electrical, and functional testing all met acceptance criteria at system and component level.
    BiocompatibilityCompliance with ISO 10993-1 and FDA GuidanceEvaluation and testing performed in accordance with ISO10993-1 and FDA Guidance.
    Software V&VMet established specifications and IEC 62304Results of executed protocols met acceptance criteria, supporting the System's embedded software is acceptable for intended use (in accordance with IEC 62304 and FDA Guidance).
    CybersecurityAnalysis of confidentiality, integrity, availability, risk mgmt.Provided cybersecurity risk management documentation (analysis of confidentiality, integrity, availability for data, info, & software per FDA Draft Guidance). Risk assessment performed for identified threats/vulnerabilities, and appropriate risk mitigation controls implemented/tested.
    Clinical performanceMet iCGM special controls requirements per 21 CFR 862.1355A bridging clinical study demonstrated comparability of performance to the predicate FreeStyle Libre 2 System, and the combined System accuracy of the FreeStyle Libre 3 and FreeStyle Libre 2 System met the iCGM special controls.
    Human FactorsMet usability requirements for intended use (ANSI/AAMI/IEC 62366)Risk analysis of design and user interface (in accordance with ANSI/AAMI/IEC 62366, IEC 60601-1-6, and FDA Guidance) demonstrated design changes met usability requirements.
    InteroperabilityAlignment with FDA Guidance "Design Considerations..."Incorporated an approach for interoperability developed in alignment with FDA Guidance.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document mentions "a bridging clinical study" but does not explicitly state the number of subjects or data points used in this study.
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective). The phrasing "bridging clinical study" generally implies a prospective study designed to bridge results from a previous study or device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. For glucose monitoring systems, ground truth is typically established using a highly accurate reference method (e.g., YSI glucose analyzer) rather than expert consensus on images.

    4. Adjudication Method for the Test Set

    This information is not provided in the document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC study was mentioned. The FreeStyle Libre 3 is a Continuous Glucose Monitoring (CGM) system, which provides direct glucose readings to the user, not an AI system that assists human "readers" in interpreting diagnostic images or data where MRMC studies are typically performed. The device is intended to replace blood glucose testing for treatment decisions, not to aid human interpretation of complex medical cases.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, a standalone performance assessment was done. The document states, "The subject device calculates glucose information identically to the predicate device, and the combined System accuracy of the FreeStyle Libre 3 System and FreeStyle Libre 2 System met the iCGM special controls requirements per 21 CFR 862.1355." This refers to the accuracy of the algorithm itself in determining glucose values from the sensor data. The "clinical performance" testing directly assesses the device's accuracy without human intervention influencing the glucose measurement.

    7. The Type of Ground Truth Used

    • Reference Glucose Methods: While not explicitly named, for CGM devices, the standard ground truth for clinical accuracy studies is typically venous blood samples analyzed by a laboratory reference method, such as a YSI glucose analyzer. The document states accuracy was demonstrated, which implies comparison to such a reference.

    8. The Sample Size for the Training Set

    • Not provided. The document describes a "bridging clinical study" for performance validation but does not detail the size of any training sets used for algorithm development. It does state that the "Sensor Glucose Algorithm established for the predicate device" is the same, implying the algorithm's core might have been trained previously for the FreeStyle Libre 2.

    9. How the Ground Truth for the Training Set was Established

    • Not provided. Given that the subject device uses the "ADC Glucose Algorithm established for the predicate device," the ground truth establishment for any original training would have likely involved comparisons to laboratory reference glucose measurements in clinical studies. However, details of such a training process or ground truth for a training set are not in this document.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210943

    Validate with FDA (Live)

    Date Cleared
    2021-11-22

    (237 days)

    Product Code
    Regulation Number
    862.1355
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.

    The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.

    The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.

    The System can be used with the FreeStyle Libre 2 Sensor (14 day) or the FreeStyle Libre 2 MediRx Sensor (10 day).

    Device Description

    The FreeStyle Libre 2 Flash Glucose Monitoring System with the FreeStyle Libre 2 App -Android (herein referred to as the 'FreeStyle Libre 2 System' or 'System') is an integrated continuous glucose monitoring system (iCGM) that provides continuous glucose measurements every minute to provide glucose levels, trends, and real time alarms capability to aid in the management of diabetes. The System requires a prescription and is intended for home use. The System consists of the following components: a Sensor which transmits yia Bluetooth Low Energy (BLE), a BLE enabled display device (Reader), and an Android or iOS mobile app (FreeStyle Libre 2 App) downloaded to a compatible smartphone. Scanning of the Sensor via Reader or App provides the user with real-time glucose measurements (glucose values) accompanied by trend information (glucose arrows) and historical glucose information (glucose graph). The user may make treatment decisions based in part on the Sensor glucose results provided by the System. The System also provides configurable alarms designed to warn the user of Low Glucose, High Glucose or Signal Loss.

    AI/ML Overview

    This 510(k) summary describes the FreeStyle Libre 2 Flash Glucose Monitoring System (with FreeStyle Libre 2 App), focusing on the addition of an Android application to the existing system. The document states that the Android app utilizes the identical algorithm and implements the same wireless interfaces with the sensor as the previously cleared Reader and iOS app. Therefore, no additional clinical data was used in this 510(k) to support a determination of substantial equivalence for clinical performance. The original clinical performance data was established in predicate devices K193371 and K211102.

    Since this 510(k) relies on the clinical performance data established in prior submissions (K193371 and K211102) and explicitly states that no new clinical performance data was generated for this specific submission (K210943) regarding the Android App, the detailed acceptance criteria and study information for clinical accuracy are not directly present within this provided text for K210943. The document focuses on non-clinical performance testing for the Android app.

    However, based on the information provided regarding the non-clinical performance of the Android App within this specific 510(k) (K210943):

    Table of Acceptance Criteria and Reported Device Performance (Non-Clinical)

    Acceptance Criteria CategoryReported Device Performance
    Software Verification and ValidationResults of executed protocols met the acceptance criteria, supporting that the Android App software is acceptable for its intended use. (Conducted in accordance with IEC 62304 and FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," May 11, 2005.)
    Human FactorsResults demonstrated that the System met usability requirements. (Conducted in accordance with ANSI/AAMI/IEC 62366, IEC 60601-1-6, and FDA Guidance "Applying Human Factors and Usability Engineering to Medical Devices," dated February 3, 2016.)
    Wireless CoexistenceTest results showed the System could tolerate interference generated by tested RF interfering devices and still meet the target performance criteria. (Consistent with AAMI TIR69 and ANSI C63.27, including in-band interference sources and other expected wireless interference sources from the intended use environment.)
    CybersecurityAppropriate risk mitigation controls have been implemented and tested. (Risk management documentation includes analysis of confidentiality, integrity, and availability for data, information, and software, with risk assessment and mitigation for identified threats and vulnerabilities.)
    Electrical Safety and Electromagnetic Compatibility (EMC)Results of the evaluation demonstrated that the System complies with electrical safety and EMC requirements. (Per IEC 60601-1:2005(r)2012 and IEC 60601-1-2:2014, respectively.)

    Due to the nature of this 510(k) being primarily for the addition of the Android app, the following information regarding clinical performance criteria, sample sizes, expert involvement, and ground truth would have been established in the predicate device studies (K193371 and K211102) and are not detailed in this document (K210943):

    • Sample size used for the test set and the data provenance: Not specified in K210943, as clinical data was referenced from predicate devices. For iCGM, clinical studies typically involve a significant number of participants (e.g., hundreds) across varied demographics to assess accuracy across the glycemic range. Data provenance would likely be from multi-center prospective clinical trials conducted in various regions.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not specified in K210943. In iCGM clinical trials, ground truth is usually established by highly precise laboratory reference methods (e.g., YSI glucose analyzer) rather than expert consensus on device readings.
    • Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for iCGM clinical accuracy studies where the "ground truth" is typically a laboratory reference method.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable for iCGM clinical accuracy studies as described. The device provides direct glucose values, not interpretations requiring human reader assistance in the same way as imaging diagnostics. It replaces blood glucose testing for treatment decisions.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Yes, the core accuracy of the algorithm is evaluated in a standalone manner against laboratory reference values in the predicate device studies. The FreeStyle Libre 2 System itself, including the sensor and its algorithm, performs autonomously to produce glucose readings.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc): For iCGM devices, the gold standard ground truth for accuracy studies is typically laboratory reference methods, such as those performed by a YSI glucose analyzer, which measures glucose concentration in blood plasma.
    • The sample size for the training set: Not specified in K210943. Training data for the glucose algorithm would have been part of the original development and validation of the FreeStyle Libre 2 system (K193371 and K211102), and would involve a large, diverse dataset of sensor readings and corresponding reference glucose values.
    • How the ground truth for the training set was established: Not specified in K210943. Similar to the test set, the ground truth for training would have been established using highly accurate laboratory reference methods (e.g., YSI analyzer) from blood samples collected concurrently with sensor readings.
    Ask a Question

    Ask a specific question about this device

    K Number
    K201761

    Validate with FDA (Live)

    Date Cleared
    2021-07-30

    (396 days)

    Product Code
    Regulation Number
    862.1355
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.

    The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.

    The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.

    Device Description

    The FreeStyle Libre 2 Flash Glucose Monitoring System with the FreeStyle Libre 2 App (herein referred to as the 'FreeStyle Libre 2 System' or 'System') is an integrated continuous glucose monitoring system (iCGM) that provides continuous glucose measurements every minute to provide glucose levels, trends and alerts. The System requires a prescription and is intended for home use. The System consists of the following components: a sensor which transmits via Bluetooth Low Energy (BLE), a BLE enabled display device (Reader), and an iOS mobile app (FreeStyle Libre 2 App) downloaded to a compatible smartphone. Scanning of the sensor via Reader or App provides the user with real-time glucose measurements (glucose values) accompanied by trend information (glucose arrows) and historical glucose information (glucose graph). The user may make treatment decisions based in part on the sensor glucose results provided by the System. The System also provides configurable alarms designed to warn the user of Low Glucose, High Glucose or Signal Loss.

    AI/ML Overview

    The provided text describes the FreeStyle Libre 2 Flash Glucose Monitoring System (with FreeStyle Libre 2 App) and its substantial equivalence to a predicate device. However, the document primarily focuses on explaining the device, its intended use, comparison with a predicate, and the general types of performance testing conducted to support substantial equivalence for the app component.

    Crucially, the acceptance criteria and detailed study results proving the device meets those criteria (specific performance metrics, sample sizes for clinical trials, ground truth establishment, expert adjudication, MRMC studies, etc.) are NOT explicitly present in this excerpt. The document states "no additional clinical data beyond that provided in K193371 was used in this 510(k) to support a determination of substantial equivalence" for clinical performance, indicating reliance on prior submissions for the predicate device. Therefore, a comprehensive answer for the requested points regarding the study that proves the device meets the acceptance criteria cannot be fully extracted from this document alone.

    However, based on the information provided, I can infer and answer parts of your request.


    Here's an attempt to answer your questions based on the provided text, and where information is missing, I will clearly state that:

    Device: FreeStyle Libre 2 Flash Glucose Monitoring System (with FreeStyle Libre 2 App)
    Predicate Device: FreeStyle Libre 2 Flash Glucose Monitoring System (K193371)

    A Table of Acceptance Criteria and Reported Device Performance:

    The document does not provide a specific table of acceptance criteria with corresponding performance metrics. It generally states that "Results of executed protocols met the acceptance criteria" for various tests, but does not quantify these criteria or the results.

    Acceptance Criteria (Inferred from general statements)Reported Device Performance (General Statement in document)
    Software Verification and Validation: Meets established specifications and IEC 62304 guidance."Results of executed protocols for the met the acceptance criteria and therefore supports that the System's embedded software is acceptable for its intended use."
    Human Factors: Meets usability requirements."Results demonstrated that the System met usability requirements."
    Wireless Coexistence: Tolerates interference from various RF interfering devices and meets target performance criteria."Test results showed the System could tolerate interference generated by these RF interfering devices and still meet the target performance criteria."
    Cybersecurity: Risk mitigation controls are implemented and tested."Appropriate risk mitigation controls have been implemented and tested."
    Electrical Safety and EMC: Complies with IEC 60601-1:2005(r)2012 and IEC 60601-1-2:2014."demonstrated that the System complies with electrical safety and EMC requirements."
    Clinical Performance (Accuracy)"The App utilizes the identical algorithm and implements the same wireless interfaces with the Sensor as used by the Reader. As the App calculates glucose information identically to the Reader, no additional clinical data beyond that provided in K193371 was used in this 510(k) to support a determination of substantial equivalence." (This implies that the clinical performance established for the predicate device is considered applicable and meets criteria, but specific numbers are not here).

    1. Sample sized used for the test set and the data provenance:

    • Sample Size for Test Set: Not specified in this document. The document states that no additional clinical data was used for this 510(k) beyond what was presented in K193371 (the predicate device's submission). The performance testing described (Software V&V, Human Factors, Wireless Coexistence, Cybersecurity, Electrical Safety/EMC) are generally system-level or component-level tests, not direct patient clinical trials with specific "test sets" in the sense of clinical data.
    • Data Provenance: Not specified for the clinical data that supports the predicate device (K193371). For the testing described in this document (Software, Human Factors, etc.), it's implied to be internal testing by Abbott Diabetes Care Inc. and conducted in laboratory environments. The document does not specify if it's retrospective or prospective.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not describe specific "ground truth" establishment for a clinical test set in this submission, as it defers to a previous 510(k) (K193371) for clinical performance data. Therefore, the number and qualifications of experts for establishing clinical ground truth are not provided here.
    • For Human Factors testing, "usability requirements" were met, implying expert evaluation or user testing, but the specifics of who established the "ground truth" or expert qualifications are not detailed.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable/Not specified. Since clinical data specifics are not provided in this submission (deferring to K193371), there's no mention of an adjudication method commonly used in clinical trials for ground truth.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a Continuous Glucose Monitoring (CGM) system, not an AI-assisted diagnostic imaging device that involves human "readers" interpreting cases. Therefore, an MRMC study is not relevant to this device type.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The document implies that standalone performance of the algorithm was assessed for the core glucose measurement. It states, "The App utilizes the identical algorithm and implements the same wireless interfaces with the Sensor as used by the Reader. As the App calculates glucose information identically to the Reader..." This suggests that the algorithm's performance in converting raw sensor measurements to glucose values was established and relied upon from the predicate device's data (K193371). No specific standalone performance metrics are provided in this document, but the fact that the algorithm is "identical" means its standalone performance is assumed to be the same as the predicate.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the clinical performance of a CGM, the ground truth typically involves lab-based reference glucose measurements (e.g., YSI analyzer or similar highly accurate methods) from blood samples taken concurrently with sensor readings. However, none of these specifics for ground truth are provided in this document, as it refers back to K193371 for clinical data.

    7. The sample size for the training set:

    • Not specified. The document does not discuss the training of the glucose algorithm. It merely confirms that the app uses the "identical algorithm" as the predicate device.

    8. How the ground truth for the training set was established:

    • Not specified. As the document does not detail the training set, it also does not elaborate on how its ground truth was established.

    In summary, this 510(k) summary focuses on demonstrating substantial equivalence for the addition of an app to an already cleared CGM system. It relies heavily on the clinical performance and established ground truth from the predicate device's prior submission (K193371) for the core glucose measurement accuracy. The testing described in this particular document pertains mostly to software validation, human factors, wireless communication, cybersecurity, and electrical safety/EMC, which are specific to the new app component and its interaction with the existing system, rather than new clinical accuracy studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193371

    Validate with FDA (Live)

    Date Cleared
    2020-06-12

    (190 days)

    Product Code
    Regulation Number
    862.1355
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.

    The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.

    The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.

    Device Description

    The FreeStyle Libre 2 Flash Glucose Monitoring System (herein referred to as the 'FreeStyle Libre 2 System or 'System') is an integrated continuous glucose monitoring system (iCGM) that provides continuous glucose measurements every minute to provide glucose levels, trends and alerts. The FreeStyle Libre 2 System consists of two primary components: a sensor that transmits via Bluetooth Low Energy (BLE), and a BLE enabled display device (Reader). User-initiated RFID scanning provides the user with real-time glucose measurements (glucose values) accompanied by trend information (glucose arrows) and historical glucose information (glucose graph). The user may use the sensor glucose results and information provided by the System in making treatment decisions. The System also provides configurable alerts and alarms designed to warn the user of Low Glucose, High Glucose or Signal Loss.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the studies conducted for the FreeStyle Libre 2 Flash Glucose Monitoring System, based on the provided document:

    Acceptance Criteria and Device Performance

    Acceptance Criteria CategorySpecific Criteria/MetricReported Device Performance (Adult)Reported Device Performance (Pediatric 6-17)Reported Device Performance (Pediatric 4-5)
    Accuracy (Overall)Percent of iCGM values within 20% of YSI reference90.2% (95% LCL: 88.7%)90.3% (95% LCL: 88.1%)85.4% (95% LCL: 80.3%) (SMBG as comparator)
    Accuracy (Point/Percent, iCGM Range < 70 mg/dL / <54 mg/dL / 54-69 mg/dL)% within 15 mg/dL89.0% (<70mg/dL), 85.9% (<54mg/dL), 89.5% (54-69mg/dL)82.7% (<70mg/dL), 71.9% (<54mg/dL), 86.4% (54-69mg/dL)N/A (SMBG used for 4-5)
    Accuracy (Point/Percent, iCGM Range 70-180 mg/dL / 181-250 mg/dL / >250 mg/dL)% within 15%75.9% (70-180mg/dL), 89.1% (181-250mg/dL), 94.0% (>250mg/dL)78.0% (70-180mg/dL), 86.0% (181-250mg/dL), 92.2% (>250mg/dL)N/A (SMBG used for 4-5)
    Accuracy (Point/Percent, YSI Range < 70 mg/dL / <54 mg/dL / 54-69 mg/dL)% within 15 mg/dL95.3% (<70mg/dL), 91.1% (<54mg/dL), 94.7% (54-69mg/dL)94.9% (<70mg/dL), 93.9% (<54mg/dL), 96.5% (54-69mg/dL)N/A (SMBG used for 4-5)
    Accuracy (Point/Percent, YSI Range 70-180 mg/dL / 181-250 mg/dL / >250 mg/dL)% within 15%76.5% (70-180mg/dL), 87.9% (181-250mg/dL), 90.9% (>250mg/dL)74.3% (70-180mg/dL), 86.6% (181-250mg/dL), 90.2% (>250mg/dL)N/A (SMBG used for 4-5)
    PrecisionMean Coefficient of Variation (Mean %CV)5.7%5.8%4.8%
    Low Glucose Alarm (e.g., 70 mg/dL)True Alarm Rate86.0%80.3%N/A
    False Alarm Rate14.0%19.7%N/A
    Correct Detection Rate89.3%93.5%N/A
    Missed Detection Rate10.7%6.5%N/A
    High Glucose Alarm (e.g., 200 mg/dL)True Alarm Rate99.2%98.0%N/A
    False Alarm Rate0.8%2.0%N/A
    Correct Detection Rate97.1%98.0%N/A
    Missed Detection Rate2.9%2.0%N/A
    Sensor Stability (MARD)Mean Absolute Relative Difference (MARD) throughout wear duration9.9% (Beginning), 8.5% (Early Middle), 8.8% (Late Middle), 9.1% (End)10.7% (Beginning), 8.0% (Early Middle), 9.7% (Late Middle), 10.2% (End)N/A
    Sensor LifeSurvival Rate by Day 1471.1%78.1%N/A
    Glucose Reading AvailabilityCapture Rate over wear durationRanges from 98.1% to 98.6%Ranges from 94.6% to 96.0%N/A
    Reportable Range40 to 400 mg/dLSupported by clinical study data (display 'LO' or 'HI' outside range)Supported by clinical study data (display 'LO' or 'HI' outside range)Supported by clinical study data (display 'LO' or 'HI' outside range)
    Ascorbic Acid InterferenceMax average sensor bias after 3x1000mg dosesLess than 20 mg/dL (after second and third 1000mg doses)Not explicitly stratified by age for this study, assumed adults.N/A

    Study Information

    1. Sample Size and Data Provenance:

      • Clinical Accuracy Studies (Study 1 & Study 2):
        • Study 1 (Adults): 146 subjects (91.1% Type 1, 8.9% Type 2 diabetes, all requiring insulin).
        • Study 2 (Pediatric): 139 subjects (98.6% Type 1, 1.4% Type 2 diabetes, all requiring insulin).
        • Age Groups in Study 2 (for YSI comparison): 77 pediatric subjects for <70 mg/dL range, 124 for 70-180 mg/dL and >180 mg/dL range. Specific breakdown for 4-5 and 6-17 years where YSI was used.
        • Age Groups in Study 2 (for SMBG comparison): 8 pediatric subjects (4-5 years old) only had SMBG as comparator.
        • Data Provenance: Prospective, multi-center studies conducted at five (adult) and four (pediatric) centers in the United States.
      • Ascorbic Acid Interference Study: 60 subjects (adults, age 18 and older) with diabetes, across 4 sites.
      • Precision Study (from Clinical Studies):
        • Adults (18+): 146 subjects, 26,791 paired readings.
        • Pediatric 4-5 years: 7 subjects, 248 paired readings.
        • Pediatric 6-17 years: 130 subjects, 10,623 paired readings.
    2. Number of Experts and Qualifications for Ground Truth:

      • The document implies that the ground truth for glucose values was established using a "laboratory glucose reference" (YSI) analyzer. While YSI analyzers are highly accurate, the document does not explicitly state the number of experts or their qualifications for establishing the ground truth. It refers to the YSI as the comparator method, which is a laboratory instrument, not human experts. For pediatric subjects aged 4-5 years, Self-Monitoring Blood Glucose (SMBG) tests were used as the comparator method.
    3. Adjudication Method for the Test Set:

      • Not applicable in the traditional sense of medical image adjudication. The "ground truth" was established by a laboratory reference method (YSI for most cases, SMBG for some pediatric cases). There's no indication of an adjudication process involving multiple human readers to resolve discrepancies in ground truth measurements. The comparison is between the device and the reference instrument.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No, an MRMC comparative effectiveness study was not done. This type of study typically involves human readers interpreting cases with and without AI assistance. The FreeStyle Libre 2 is a standalone diagnostic device that measures glucose directly, so its performance is compared against a reference standard (YSI, SMBG), not human interpretation with or without AI.
    5. Standalone (Algorithm Only) Performance:

      • Yes, the studies primarily assessed standalone performance. The device automatically provides continuous glucose measurements. The "iCGM glucose values" are the output of the algorithm. The intent is for the device to replace blood glucose testing for treatment decisions, which is a standalone use case. Human interpretation of trends and patterns is part of the overall management, but the device's accuracy itself is assessed in a standalone manner against the reference.
    6. Type of Ground Truth Used:

      • Laboratory Reference and Self-Monitoring Blood Glucose (SMBG):
        • For adult subjects and pediatric subjects aged 6-17 years, the primary ground truth was established by YSI glucose measurements. YSI (Yellow Springs Instrument) is a common laboratory method considered a highly accurate reference for glucose.
        • For pediatric subjects aged 4-5 years, the ground truth was established by SMBG (Self-Monitoring Blood Glucose) tests.
    7. Sample Size for the Training Set:

      • The document mentions "clinical evaluation" and "pivotal clinical studies" but does not explicitly state the sample size of a training set specifically for algorithm development or machine learning. The provided clinical studies (Study 1 and Study 2) describe the test sets used for performance evaluation, not necessarily for training the algorithm. CGM devices typically use complex calibration algorithms developed with extensive internal data that is not always detailed in 510(k) summaries.
    8. How the Ground Truth for the Training Set Was Established:

      • As the document does not provide details on a specific training set or its size, it also does not elaborate on how the ground truth for such a set was established. We can infer that similar high-accuracy laboratory reference methods (like YSI) would have been used during the development and calibration phases of the device, prior to the validation studies described in the submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171941

    Validate with FDA (Live)

    Date Cleared
    2017-09-27

    (91 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FreeStyle Precision Neo Blood Glucose Test Strips are for use outside the body only (in vitro diagnostic use) with the FreeStyle Libre Reader's built-in meter as components of the Flash Glucose Monitoring System to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. They are not intended to be used for testing neonatal blood samples or screening of diabetes.

    The FreeStyle Precision Neo Blood Glucose Test Strips are indicated for home (lay) user in the management of patients with diabetes. They are intended to be used by a single person and should not be shared.

    Device Description

    The FreeStyle Precision Neo Blood Glucose Test Strips are used for the quantitative measurement of glucose in capillary whole blood samples. The FreeStyle Precision Neo Blood Glucose Test Strips, in conjunction with the FreeStyle Libre Reader's built-in meter, works on the principal of amperometric technology, measuring glucose by its reaction with Glucose Dehydrogenase (GDH) in blood samples or control solutions through electrical mediation.

    The FreeStyle Precision Neo Blood Glucose Test Strips are compatible with the FreeStyle Precision Neo Blood Glucose Meter (cleared under K140371) or the FreeStyle Libre Reader's built-in meter.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device: FreeStyle Precision Neo Blood Glucose Test Strips. The submission seeks to expand the indications for use of existing test strips to include compatibility with the FreeStyle Libre Reader's built-in meter.

    Here's an analysis of the acceptance criteria and study information, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that a clinical study was performed to assess device accuracy compared to a YSI reference method, and "all performance characteristics were met." However, it does not explicitly list specific numerical acceptance criteria (e.g., accuracy percentages within certain glucose ranges) or the detailed reported performance values against those criteria. The general statement is that the device "is safe and effective for its intended use and technological characteristics when used with the FreeStyle Libre Reader's built-in meter."

    Acceptance Criteria CategorySpecific Acceptance Criteria (Not Explicitly Stated in Detail)Reported Device Performance
    Accuracy (Clinical Study)Implicit: Device accuracy compared to YSI reference. Detailed statistical metrics (e.g., % within +/-15 mg/dL or +/-15% of reference for different glucose ranges, Mean Absolute Relative Difference (MARD)) are not provided in this summary."All performance characteristics were met." (Implies satisfactory accuracy against YSI reference).
    PrecisionImplicit: Acceptable levels of variability in repeated measurements."All performance characteristics were met."
    LinearityImplicit: Accurate measurement across the specified glucose operating range (20-500 mg/dL)."All performance characteristics were met."
    Detection LimitsImplicit: Ability to accurately detect glucose at the low and high ends of the operating range."All performance characteristics were met."
    InterferenceImplicit: Robustness to common interfering substances."All performance characteristics were met."
    Environmental ConditionsImplicit: Performance within specified operating temperature (59-104°F) and humidity (10-90%)."All performance characteristics were met."
    Sample VolumeImplicit: Accurate performance with the specified sample volume (0.6 microliters)."All performance characteristics were met."
    Infection ControlImplicit: Satisfactory infection control measures/design."All performance characteristics were met."

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "a clinical study was performed" but does not specify the sample size used for this clinical study (the test set).

    Regarding data provenance:

    • Country of Origin: Not explicitly stated for the clinical study. However, the legal manufacturer establishment is Abbott Diabetes Care Ltd., located in Witney, Oxon, UK. This suggests the study might have been conducted in the UK or under their oversight, but this is an inference, not a direct statement about the study itself.
    • Retrospective or Prospective: Not explicitly stated. Clinical studies for device accuracy are typically prospective to control variables and collect fresh samples, but this is an assumption based on common practice, not directly from the text.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document states that the ground truth for the clinical study was established by comparison to a YSI reference method.

    • Number of Experts: Not applicable, as YSI is an automated laboratory analyzer, not human experts.
    • Qualifications of Experts: Not applicable. The YSI reference method is a gold standard laboratory method for glucose measurement, operated by trained laboratory personnel.

    4. Adjudication Method for the Test Set

    Not applicable, as the ground truth was established by an automated YSI reference method, not by human expert consensus requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not conducted. This type of study is more common for diagnostic imaging devices where human readers interpret images. For a blood glucose test strip, the "reader" is the device itself (the FreeStyle Libre Reader's built-in meter), and the primary assessment is standalone analytical and clinical accuracy.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Yes, a standalone performance assessment was conducted. The "performance testing" and "clinical study" described assess the FreeStyle Precision Neo Blood Glucose Test Strips in conjunction with the FreeStyle Libre Reader's built-in meter as a complete system, without human interpretation as a variable affecting the glucose measurement outcome. The accuracy is directly compared to a reference method (YSI), reflecting the device's intrinsic performance.

    7. The Type of Ground Truth Used

    The ground truth used was a YSI reference method. This is a highly accurate laboratory method commonly considered the gold standard for glucose measurement in clinical studies of blood glucose monitoring systems.

    8. The Sample Size for the Training Set

    The document does not provide information on a "training set" sample size. For in vitro diagnostic devices like blood glucose test strips, the development process might involve initial analytical verification and validation studies during device design and iterative refinement. However, the "clinical study" mentioned typically refers to the final validation conducted on a separate, independent "test set" population. The concept of a distinct 'training set' as used in machine learning models is not typically applied or reported in the same way for these types of IVDs.

    9. How the Ground Truth for the Training Set Was Established

    Since no information about a distinct "training set" is provided, how its ground truth was established is also not stated. The document focuses on the testing and validation of the device, where the YSI reference method serves as the ground truth for comparison.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 4