K Number
K171941

Validate with FDA (Live)

Date Cleared
2017-09-27

(91 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FreeStyle Precision Neo Blood Glucose Test Strips are for use outside the body only (in vitro diagnostic use) with the FreeStyle Libre Reader's built-in meter as components of the Flash Glucose Monitoring System to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. They are not intended to be used for testing neonatal blood samples or screening of diabetes.

The FreeStyle Precision Neo Blood Glucose Test Strips are indicated for home (lay) user in the management of patients with diabetes. They are intended to be used by a single person and should not be shared.

Device Description

The FreeStyle Precision Neo Blood Glucose Test Strips are used for the quantitative measurement of glucose in capillary whole blood samples. The FreeStyle Precision Neo Blood Glucose Test Strips, in conjunction with the FreeStyle Libre Reader's built-in meter, works on the principal of amperometric technology, measuring glucose by its reaction with Glucose Dehydrogenase (GDH) in blood samples or control solutions through electrical mediation.

The FreeStyle Precision Neo Blood Glucose Test Strips are compatible with the FreeStyle Precision Neo Blood Glucose Meter (cleared under K140371) or the FreeStyle Libre Reader's built-in meter.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device: FreeStyle Precision Neo Blood Glucose Test Strips. The submission seeks to expand the indications for use of existing test strips to include compatibility with the FreeStyle Libre Reader's built-in meter.

Here's an analysis of the acceptance criteria and study information, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document states that a clinical study was performed to assess device accuracy compared to a YSI reference method, and "all performance characteristics were met." However, it does not explicitly list specific numerical acceptance criteria (e.g., accuracy percentages within certain glucose ranges) or the detailed reported performance values against those criteria. The general statement is that the device "is safe and effective for its intended use and technological characteristics when used with the FreeStyle Libre Reader's built-in meter."

Acceptance Criteria CategorySpecific Acceptance Criteria (Not Explicitly Stated in Detail)Reported Device Performance
Accuracy (Clinical Study)Implicit: Device accuracy compared to YSI reference. Detailed statistical metrics (e.g., % within +/-15 mg/dL or +/-15% of reference for different glucose ranges, Mean Absolute Relative Difference (MARD)) are not provided in this summary."All performance characteristics were met." (Implies satisfactory accuracy against YSI reference).
PrecisionImplicit: Acceptable levels of variability in repeated measurements."All performance characteristics were met."
LinearityImplicit: Accurate measurement across the specified glucose operating range (20-500 mg/dL)."All performance characteristics were met."
Detection LimitsImplicit: Ability to accurately detect glucose at the low and high ends of the operating range."All performance characteristics were met."
InterferenceImplicit: Robustness to common interfering substances."All performance characteristics were met."
Environmental ConditionsImplicit: Performance within specified operating temperature (59-104°F) and humidity (10-90%)."All performance characteristics were met."
Sample VolumeImplicit: Accurate performance with the specified sample volume (0.6 microliters)."All performance characteristics were met."
Infection ControlImplicit: Satisfactory infection control measures/design."All performance characteristics were met."

2. Sample Size Used for the Test Set and Data Provenance

The document mentions "a clinical study was performed" but does not specify the sample size used for this clinical study (the test set).

Regarding data provenance:

  • Country of Origin: Not explicitly stated for the clinical study. However, the legal manufacturer establishment is Abbott Diabetes Care Ltd., located in Witney, Oxon, UK. This suggests the study might have been conducted in the UK or under their oversight, but this is an inference, not a direct statement about the study itself.
  • Retrospective or Prospective: Not explicitly stated. Clinical studies for device accuracy are typically prospective to control variables and collect fresh samples, but this is an assumption based on common practice, not directly from the text.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document states that the ground truth for the clinical study was established by comparison to a YSI reference method.

  • Number of Experts: Not applicable, as YSI is an automated laboratory analyzer, not human experts.
  • Qualifications of Experts: Not applicable. The YSI reference method is a gold standard laboratory method for glucose measurement, operated by trained laboratory personnel.

4. Adjudication Method for the Test Set

Not applicable, as the ground truth was established by an automated YSI reference method, not by human expert consensus requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not conducted. This type of study is more common for diagnostic imaging devices where human readers interpret images. For a blood glucose test strip, the "reader" is the device itself (the FreeStyle Libre Reader's built-in meter), and the primary assessment is standalone analytical and clinical accuracy.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Yes, a standalone performance assessment was conducted. The "performance testing" and "clinical study" described assess the FreeStyle Precision Neo Blood Glucose Test Strips in conjunction with the FreeStyle Libre Reader's built-in meter as a complete system, without human interpretation as a variable affecting the glucose measurement outcome. The accuracy is directly compared to a reference method (YSI), reflecting the device's intrinsic performance.

7. The Type of Ground Truth Used

The ground truth used was a YSI reference method. This is a highly accurate laboratory method commonly considered the gold standard for glucose measurement in clinical studies of blood glucose monitoring systems.

8. The Sample Size for the Training Set

The document does not provide information on a "training set" sample size. For in vitro diagnostic devices like blood glucose test strips, the development process might involve initial analytical verification and validation studies during device design and iterative refinement. However, the "clinical study" mentioned typically refers to the final validation conducted on a separate, independent "test set" population. The concept of a distinct 'training set' as used in machine learning models is not typically applied or reported in the same way for these types of IVDs.

9. How the Ground Truth for the Training Set Was Established

Since no information about a distinct "training set" is provided, how its ground truth was established is also not stated. The document focuses on the testing and validation of the device, where the YSI reference method serves as the ground truth for comparison.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem that features a stylized representation of three human profiles facing to the right, with flowing lines extending from the bottom of the profiles.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 27, 2017

ABBOTT DIABETES CARE INC. STEPHANIE YUN SR. REGULATORY AFFAIRS SPECIALIST 1360 SOUTH LOOP ROAD ALAMEDA, CA 94502

Re: K171941

Trade/Device Name: FreeStyle Precision Neo Blood Glucose Test Strips Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW Dated: August 31, 2017 Received: September 5, 2017

Dear Stephanie Yun:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171941

Device Name

FreeStyle Precision Neo Blood Glucose Test Strips

Indications for Use (Describe)

The FreeStyle Precision Neo Blood Glucose Test Strips are for use outside the body only (in vitro diagnostic use) with the FreeStyle Libre Reader's built-in meter as components of the Flash Glucose Monitoring System to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. They are not intended to be used for testing neonatal blood samples or screening of diabetes.

The FreeStyle Precision Neo Blood Glucose Test Strips are indicated for home (lay) user in the management of patients with diabetes. They are intended to be used by a single person and should not be shared.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a blue, stylized letter "a" on the left, followed by the text "Abbott" in bold, black letters. Below "Abbott" is the text "Diabetes Care" in a smaller, non-bold font, also in black.

K171941 510(k) Summary

According to the requirements per 21 CFR §807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Company:Abbott Laboratories
Division:Abbott Diabetes Care, Inc.
Street Address:1360 South Loop Road
City, State Zip:Alameda, CA 94502
Telephone No:510-749-5400
Fax No:510-864-4791
Contact Person:Stephanie YunTel No. 510-239-2689Fax No. 510-864-4791stephanie.yun@abbott.com
Proprietary Name:FreeStyle Precision Neo Blood Glucose Test Strips
Common Name:Glucose Test System
ClassificationName:System, Test, Blood Glucose, Over The Counter, Class II (21 CFR §862.1345)Product codes: NBW
Predicate Device:FreeStyle Precision Neo Blood Glucose Monitoring System (K140371)
LegalManufacturer:Establishment:Abbott Diabetes Care Ltd.Range RoadWitney, OxonOX29 OYL, UK
U.S. ContactEstablishment:Abbott Diabetes Care Inc.1360 South Loop RoadAlameda, CA 94502

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a blue, stylized "a" on the left, followed by the text "Abbott" in bold, black letters on the top line. Below "Abbott" is the text "Diabetes Care" in a smaller, non-bold font, also in black.

Intended Use:

The FreeStyle Precision Neo Blood Glucose Test Strips are for use outside the body only (in vitro diagnostic use) with the FreeStyle Libre Reader's built-in meter as components of the FreeStyle Libre Flash Glucose Monitoring System to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. They are not intended to be used for testing neonatal blood samples or for the diagnosis or screening of diabetes.

The FreeStyle Precision Neo Blood Glucose Test Strips are indicated for home (lay) user in the management of patients with diabetes. They are intended to be used by a single person and should not be shared.

Description of the Device:

The FreeStyle Precision Neo Blood Glucose Test Strips are used for the quantitative measurement of glucose in capillary whole blood samples. The FreeStyle Precision Neo Blood Glucose Test Strips, in conjunction with the FreeStyle Libre Reader's built-in meter, works on the principal of amperometric technology, measuring glucose by its reaction with Glucose Dehydrogenase (GDH) in blood samples or control solutions through electrical mediation.

The FreeStyle Precision Neo Blood Glucose Test Strips are compatible with the FreeStyle Precision Neo Blood Glucose Meter (cleared under K140371) or the FreeStyle Libre Reader's built-in meter.

Principles of Operation:

The FreeStyle Precision Neo Blood Glucose Test Strips (in conjunction with a compatible blood glucose meter) utilizes amperometric technology to quantitatively measure the glucose concentration in capillary whole blood samples from the fingertip and in MediSense Glucose & Ketone Control Solutions.

The FreeStyle Precision Neo Blood Glucose Test Strips measures glucose electrically when used in conjunction with a compatible blood glucose meter. The glucose biosensor is capable of recognising the glucose present in whole blood or control solutions by virtue of the glucose specificity of the enzyme GDH present on the glucose test strip. The electrons liberated by this reaction are transferred via a co-factor and mediator to the meter where they are read as a small electrical current. The size of the current is directly proportional to the level of the glucose in the applied sample.

The apply blood symbol is displayed for the user to apply blood to the test strip until the meter begins the test. The meter detects trigger current from the test strip when enough blood has covered the strip electrodes and the test countdown will start. When the countdown is complete a test result is displayed on the meter screen. The unit of measure displayed on the meter screen is fixed in mg/dL and cannot be modified by the user.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a blue, stylized "a" on the left, followed by the text "Abbott" in bold, black letters on the top line. Below "Abbott" is the text "Diabetes Care" in a smaller, non-bold font, also in black.

Substantial Equivalence:

The FreeStyle Precision Neo Blood Glucose Test Strips are substantially equivalent to the predicate test strips, which were cleared by the Agency on September 30, 2014, to market under K140371: FreeStyle Precision Neo Blood Glucose: Monitoring System and Test Strips.

The purpose of this submission is to expand the indications for use of the test strips to now include compatibility with the FreeStyle Libre Reader's built-in meter, in addition to the existing FreeStyle Precision Neo Blood Glucose Meter. There is no change to the fundamental technology of the test strip cleared under K140371.

Summary of Clinical and Performance Testing:

Clinical and performance testing was conducted with the FreeStyle Precision Neo Blood Glucose Test Strips and FreeStyle Libre Reader's built-in meter. Performance testing included analysis of precision, linearity and detection limits, interference, environmental conditions, sample volume and infection control studies. A clinical study was performed to assess the device accuracy compared to YSI reference.

Conclusions from Clinical and Performance Evaluations:

The results obtained from clinical and performance studies show that all performance characteristics were met. The results demonstrate that the FreeStyle Precision Neo Blood Glucose Test Strips are safe and effective for its intended use and technological characteristics when used with the FreeStyle Libre Reader's built-in meter, and therefore, substantially equivalent to the predicate device (K140371).

Comparison to Predicate Device:

The similarities and differences between the FreeStyle Precision Neo Blood Glucose Test Strips and the predicate test strips (K140371) are highlighted in Tables 1 and 2 below.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a stylized blue "a" on the left, followed by the words "Abbott" in bold black font and "Diabetes Care" in a regular black font on the right. The logo is simple and modern, with a focus on the company's name and area of expertise.

Proposed DevicePredicate Device
FreeStyle Precision NeoFreeStyle Precision Neo Blood
PRODUCT NAMEBlood Glucose Test StripsGlucose Monitoring System(K140371)
CHARACTERISTICS
FundamentalTechnologyThe FreeStyle Precision NeoBlood Glucose Test Strips, inconjunction with bloodglucose meter, utilizesamperometric technology toquantitatively measure theglucose concentration inwhole blood samples.Same
Principles of OperationAmperometrySame
Glucose OperatingRange20-500 mg/dLSame
ChemistryGDH-NADSame
Glucose Sample volume0.6 microlitersSame
Glucose Assay Time5 secondsSame
Coding (Calibration)No coding requiredSame
Sample ApplicationTop or end fillSame
Operating Temperature59° - 104°FSame
Operating Humidity10% - 90%, non-condensingSame
Product ClassificationCodeNBWNBW, LFR
Glucose Sample TypesFresh capillary whole bloodfrom the fingerSame
Glucose HematocritRange15-65%Same
Altitude10,000 feetSame

Table 1: Similarities between the Proposed and Predicate Device

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a stylized blue "a" on the left, followed by the words "Abbott" in bold black font and "Diabetes Care" in a regular black font on the right. The logo is simple and modern, with a focus on the company's name and area of expertise.

Proposed DevicePredicate Device
PRODUCT NAMEFreeStyle Precision Neo Blood Glucose Test StripsFreeStyle Precision Neo Blood Glucose Monitoring System(K140371)
CHARACTERISTICS
Indications for Use(for Test Strip)The FreeStyle Precision Neo Blood Glucose Test Strips are for use outside the body only (in vitro diagnostic use) with the FreeStyle Libre Reader's built-in meter as components of the FreeStyle Libre Flash Glucose Monitoring System to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. They are not intended to be used for testing neonatal blood samples or for the diagnosis or screening of diabetes.The FreeStyle Precision Neo Blood Glucose Test Strips are indicated for home (lay) user in the management of patients with diabetes. They are intended to be used by a single person and should not be shared.The FreeStyle Precision Neo Blood Glucose Test Strips are for use outside the body only (in vitro diagnostic use) with the FreeStyle Precision Neo Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. They are not intended to be used for testing neonatal blood samples or for the diagnosis or screening of diabetes.The FreeStyle Precision Neo Blood Glucose Test Strips are indicated for home (lay) user in the management of patients with diabetes. They are intended to be used by a single person and should not be shared.

Table 2: Differences between the Proposed and Predicate Device

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.