Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K231256
    Manufacturer
    Date Cleared
    2024-01-03

    (247 days)

    Product Code
    Regulation Number
    880.5450
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K231256

    Trade/Device Name: Carecube Negative Pressure Isolation Chamber Regulation Number: 21 CFR 880.5450
    Classification: | Class II |
    | Regulation Number: | 21 CFR 880.5450
    |
    | Regulation
    Number | 880.5450
    | 880.5450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Carecube is a patient isolation unit (PIU) designed for the temporary isolation of patients within a hospital setting to prevent particulate (biological) cross-contamination between user and patient, while enclosing the contaminated patient from the external environment. Device should only be used in a hospital setting. This is for temporary housing of a patient prior to transfer to an appropriate hospital destination. Transfer to a more permanent hospital setting should occur as soon as possible. The Carecube is designed with features that enable low-moderate complexity medical interventions. This includes the following procedures: blood draw, medication administration, palpating abdomen, cardiac auscultation, and connection to IV line/monitoring cables.

    Device Description

    The Carecube Negative Pressure Isolation Chamber is a modular, negative pressure, isolation patient care system with both reusable and disposable components that can rapidly augment capacity to care for airborne infectious diseases and other biological threats for common patient care protocols with features built into the unit to allow ease of execution. The device is intended to be used indoors, in a hospital setting, as a temporary isolation of patients with suspected or confirmed diagnosis of infectious disease to prevent Healthcare Practitioners (HCP) exposure to pathogenic biological airborne particulates and is designed with features that enable low-moderate complexity medical interventions.

    AI/ML Overview

    This document describes the Carecube Negative Pressure Isolation Chamber, a patient isolation unit (PIU), and its assessment for substantial equivalence to a predicate device. This device is not an AI/ML powered device, so an analysis of acceptance criteria relative to an AI/ML device is not applicable.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is not an AI/ML device, the acceptance criteria are based on various performance and safety tests rather than metrics like sensitivity, specificity, or AUC. The document lists the following performance tests and their results:

    Performance CriterionStandard/Test MethodProposed Device Result
    Biocompatibility
    CytotoxicityISO 10993-5Non-cytotoxic
    SensitizationISO 10993-10Non-sensitizing
    Intracutaneous ReactivityISO 10993-23Non-irritant
    Acute Systemic ToxicityISO 10993-11Non-toxic
    Volatile Organic Compounds; ParticulatesToxicological Risk AssessmentNon-toxic
    EMC and Electrical Safety
    IEC 60601-1Pass
    IEC 60601-1-2Pass
    IEC 60601-1-6Pass
    Usability / Human Factors
    IEC 62366-1Pass
    Packaging Testing
    ASTM D4169Pass
    ASTM D4332Pass
    Shelf-life Testing
    ASTM F198012 months
    Functional Performance Testing
    Canopy RobustnessInternalPass
    Airflow and Smoke TestingInternalPass
    Differential PressureInternalPass
    Tagged particle ClearanceInternalPass
    Tagged particle ContainmentInternalPass
    NoiseInternalPass
    Viral PenetrationASTM F1671Pass
    Cleaning ResiliencyInternalPass
    Cleaning ValidationAAMI TIR30, AAMI ST98Pass
    Disinfection ValidationAAMI TIR30, AAMI ST98Pass
    Flammability16 CFR Part 1610.7Pass
    Repeat Use TestingInternalPass
    Usage Life Testing
    System Usage LifeInternalPass
    Filter Usage LifeInternalPass
    Clinical Testing
    Human Use ComfortInternalPass

    2. Sample Size Used for the Test Set and Data Provenance

    The clinical study conducted for the Carecube Negative Pressure Isolation Chamber was a prospective, single-center, non-blinded study.

    • Sample Size: A minimum of ten (10) subjects were included.
    • Data Provenance: The study used normal, healthy volunteers aged 21-47 years old, comprising a cross-section of males and females. This indicates prospective data collection from a single, unspecified center (presumably in the US given FDA submission).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Since this is not an AI/ML device, the concept of "ground truth" as established by experts (e.g., radiologists) for a test set is not directly applicable. The clinical study focused on patient safety and comfort, assessed through routine monitoring of vital signs, STAID evaluation (for anxiety), and environmental monitoring (temperature, humidity, CO2, O2). The determination of "Pass" likely relies on established medical norms for vital signs and comfort levels, and environmental safety standards, interpreted by the study's clinical team, rather than a panel of independent expert adjudicators for a "ground truth" diagnosis.

    4. Adjudication Method

    Not applicable for this type of device and study. The study assessed physiological parameters and environmental conditions, not diagnostic outcomes requiring expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is not an AI-powered device, and the study was designed to assess safety and comfort, not diagnostic performance with or without AI assistance.

    6. Standalone Performance Study

    Not applicable. This device is not an algorithm; it is a physical patient isolation unit. The performance studies detailed are for the physical device's characteristics and its interaction with a human patient (comfort, safety).

    7. Type of Ground Truth Used

    As mentioned above, the "ground truth" in the clinical study related to:

    • Physiological measurements: Vital signs (blood pressure, body temperature, pulse, oxygen saturation), which have objective measurements.
    • Patient-reported outcomes: Level of anxiety using the STAID evaluation, which is a standardized psychological assessment.
    • Environmental measurements: Temperature, relative humidity, carbon dioxide (CO2), and atmospheric oxygen (O2) within the Carecube, which are objective measurements.

    These measurements were used to determine if the device provided a safe and comfortable environment, not to establish a diagnostic ground truth.

    8. Sample Size for the Training Set

    Not applicable. This device does not use an AI/ML algorithm, so there is no training set as understood in the context of AI.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163223
    Date Cleared
    2017-11-02

    (351 days)

    Product Code
    Regulation Number
    880.5450
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K163223

    Trade/Device Name: ORCA™ - Operational Rescue Containment Apparatus Regulation Number: 21 CFR 880.5450
    (PIU) Regulation Name: Patient care reverse isolation chamber Classification Name: Class II (21 CFR 880.5450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Model ORCA-2016CN is a portable and ambulatory Patient Isolation Unit (PIU) designed to prevent chemical and particulate (biological and radiological) cross-contamination between an enclosed contaminated patient and the external environment during evacuation and transport activities.

    Device Description

    The Operational Rescue Containment Apparatus (ORCATM) Patient Isolation Unit, Model ORCA-2016CN, is a tubular, flexible, Patient Isolation Unit (PIU) for use during transport and evacuation of chemical and particulate (biological and radiological) contaminated patients who have been medically stabilized. Each unit employs clear windows to allow visual monitoring of the patient. The unit incorporates multiple glove arms for minimal medical intervention. The PIU's have a reinforced base mat with integrated carry straps and tethers to enable lifting and attachment to a Stokes litter for hoisting operations or a North Atlantic Treaty Organization (NATO) litter for alternative transportation. Chamber airflow through the isolator and across the patient is provided by the blower from a Powered Air-Purifying Respirator (PAPR). Internal contaminated air in the PIU is vented out the exhaust ports which are fitted with two Chemical, Biological, Radiological, Nuclear (CBRN) filter cartridges, thus protecting the external environment from a contaminated patient.

    AI/ML Overview

    The provided text describes the ISOVAC Products, LLC ORCA™ - Operational Rescue Containment Apparatus (K163223), a patient isolation unit. However, the document does not contain information relevant to all parts of the request. Specifically, it does not detail a study involving an algorithm or artificial intelligence, a common scenario for multi-reader multi-case (MRMC) comparative effectiveness studies. Therefore, some sections of your request cannot be fully addressed from the provided text.

    Here's an analysis of the acceptance criteria and performance data available in the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" for each performance test in a formal table format with numerical targets. Instead, it describes tests performed and their successful outcomes, implying that successful completion of these tests serves as the acceptance criteria.

    Acceptance Criteria (Implied)Reported Device Performance
    Operational Testing (Hoisting and Transport): Device must withstand hoisting and transport in extreme conditions (e.g., helicopter rotor wash).Successfully conducted by the U.S. Coast Guard (USCG) on an MH-60 helicopter. Hoisted multiple times at various altitudes (25/50/75/100 ft.) with a trail line to stabilize the Stokes litter. The MH-60 generates 90+ knots of downward rotor wash. USCG deemed this sufficient testing to validate the "proof of concept."
    Chemical Agent Permeation: Exceed objective requirement for containment efficiency (%) for aerosols at a maximum 0.3 micron particle size against HD (distilled mustard) and GD (soman).PIU system permeation testing performed in principle to IEST-RP-CC001.3 and MIL-STD-282 Method 102.9.1. Demonstrated a containment efficiency (%) for both aerosols at a maximum 0.3 micron particle size that exceeded the objective requirement.
    Liquid Penetration / Pathogen Passage: No detectable bacteriophage (0 PFU/mL) in assay titer.Testing performed using ASTM F1671/F1671M-13 (Resistance of Materials Used in Protective Clothing to Penetration by Blood-Borne Pathogens Using Phi-X174 Bacteriophage Penetration). No detectable bacteriophage in the assay titer (i.e., 0 PFU/mL) was considered passing. The Bag Seam, Sleeve Seam, and Window Seam materials passed liquid penetration testing.
    Human Use Comfort: Participant's vital signs stable, and no significant problems reported during extended use in varying environments.A test participant (TP) was contained in the PIU for four hours across six temperature and RH% environments (65-90°F and 40-90% at 4 mph wind speed). Internal PIU temperature, RH%, CO2, O2 levels, patient vitals, and comfort levels were monitored. TP's vital signs were stable with no indication of heat stress or health issue. TP reported no significant problems.
    Weight Lift Capacity: Handles and strapping must hold a maximum specified test load.Tested under ISOVAC standard protocol procedure. Demonstrated the PIU handle strapping can hold the maximum test load of 480lb or a load of 137% of the stated maximum.

    2. Sample Size Used for the Test Set and Data Provenance

    • Operational Testing: The document mentions "multiple times" for hoisting. It doesn't specify a numerical sample size for the tests on the PIU during helicopter operations.
      • Data Provenance: Conducted by the U.S. Coast Guard, suggesting a domestic (USA) origin. The nature of the test (actual helicopter operations) implies prospective testing for the specific device, though it's evaluating a physical product's performance rather than data analysis.
    • Chemical Agent Permeation, Liquid Penetration / Pathogen Passage, Human Use Comfort, Weight Lift Testing: These were performed by a "third-party testing party under contract with the USGC." No specific numerical sample sizes are given for the number of units tested or the number of samples for materials.
      • Data Provenance: The location of the third-party testing party is not specified, but the USGC contract suggests it's likely within the US. These tests are prospective.
    • Human Use Comfort: "a test participant (TP)" was used, indicating a sample size of 1 for the comfort study.
      • Data Provenance: Not explicitly stated, but likely domestic (USA) given the context. Prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable as the document describes performance testing of a physical medical device (Patient Isolation Unit), not an AI or algorithm-based device that would require expert consensus for establishing ground truth data. The "ground truth" for these tests is based on objective measurements (e.g., lack of permeation, stable vitals, weight holding capacity) and successful operational use as determined by the USCG.

    4. Adjudication Method for the Test Set

    This is not applicable for the type of device and testing described. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies involving interpretation of medical data by multiple experts to establish a "true" label, which is not the nature of the tests conducted for this physical device. The USCG "deemed...sufficient testing" for the operational tests, indicating their internal expert judgment as an adjudicator, but not in the sense of a medical image or diagnostic interpretation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices where human readers interpret medical cases with and without AI assistance. The ORCA™ is a physical containment apparatus.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone (algorithm only) performance study was not done. The ORCA™ is a physical medical device and does not involve an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the tests described is primarily based on:

    • Objective performance measurements: For chemical permeation, liquid penetration, vital signs, and weight capacity. These are quantifiable outcomes.
    • Operational success/validation: For hoisting and transport, determined by the U.S. Coast Guard's assessment of suitability for use in their aircraft.

    8. The Sample Size for the Training Set

    Not Applicable. The ORCA™ is a physical medical device, not an AI or algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not Applicable. As mentioned above, there is no AI or algorithm with a training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K080117
    Manufacturer
    Date Cleared
    2009-01-30

    (379 days)

    Product Code
    Regulation Number
    880.5450
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K080117

    Trade/Device Name: Patient Isolation Unit, Model G019-1000 Regulation Number: 21 CFR 880.5450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Gentex Patient Isolation Unit (PIU) is a temporary, single-use, portable structure that is designed to fully contain and transport a patient and prevent particulate crosscontamination between the patient, the external environment and the caregivers. The PIU is designed to enable medical care-giving to the patient via integrated medical bulkheads interfaced with end-user supplied medical products and equipment. The PIU is intended to be used as a means to temporarily isolate a contaminated patient , outside or within patient care facilities. The PIU can also be used to transport those patients within vehicles, including aircraft, via a NATO-type or similar litter/gumey.

    Device Description

    The Gentex Patient Isolation Unit (PIU) is a temporary, single-use, portable structure that is designed to fully contain and transport a patient and prevent particulate crosscontamination between the patient, the external environment and the caregivers. The PIU is designed to enable medical care-giving to the patient via integrated medical bulkheads interfaced with end-user supplied medical products and equipment.

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a Patient Isolation Unit, Model G019-1000. It does not contain information about acceptance criteria or a study proving the device meets them in the context of AI/ML performance. The letter simply states that the FDA has determined the device is substantially equivalent to legally marketed predicate devices.

    Therefore, I cannot provide the requested information from the given text.

    The numbered information you asked for (acceptance criteria, sample sizes, expert qualifications, adjudication, MRMC study, standalone performance, ground truth types, training set size, and ground truth establishment for training set) is not present in this regulatory document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K052798
    Date Cleared
    2006-06-06

    (246 days)

    Product Code
    Regulation Number
    880.5450
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Common Name: Patient Isolation Unit Classification Name; Patient care reverse isolation chamber (21 CFR 880.5450
    Bellaire, Texas 77401

    Re: K052798

    Trade/Device Name: CAPSULSTM Patient Isolation Unit Regulation Number: 880.5450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CAPSULS™ (Containment and Protection System Utilizing Life Support) is a portable Patient Isolation Unit (PIU), which prevents particulate (biological and radiological) cross-contamination between the patient and the external environment; and with features that enable medical intervention to the patient via end-user supplied medical equipment. The CAPSULS is intended to be used for:

    • . The transport and isolation of patients on aircraft, ambulances, ships, and any vehicle capable of safely transporting a patient on a standard litter,
    • . The temporary isolation, with or without transport, of patients within hospitals or other medical facilities.
    Device Description

    The CAPSULS (Containment And Protection System Utilizing Life Support) Patient Isolation Unit Model 2004 is a clear, tubular, flexible, patient isolation unit (PIU) for use during transport and evacuation. Each unit employs a clear envelope (either polyurethane [PUR] or polyvinylchloride [PVC]) to allow visual monitoring of the patient; the unit also has multiple glove arms and tubular access ports to facilitate medical intervention.

    The PIUs have a reinforced base mat with integrated handholds and tethers to enable lifting and attachment to a standard litter, and are supported by external ribs and end stanchions. Chamber airflow to the patient is provided by a battery-powered airpurifying respirator, which HEPA-filters contaminated air. The unit is operated under negative pressure mode during transport of a contaminated patient, and is operated in positive pressure mode during transport of a patient through a contaminated environment.

    AI/ML Overview

    The provided text describes a medical device, the CAPSULS™ Patient Isolation Unit, and its 510(k) summary for FDA clearance. However, the document does not contain the detailed acceptance criteria and study information typically associated with AI/ML device performance evaluations (e.g., sensitivity, specificity, AUC, sample sizes for training/test sets, ground truth establishment by experts, MRMC studies).

    The CAPSULS™ is a physical medical device (patient isolation unit), not an AI/ML diagnostic or therapeutic software. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are framed in terms of its physical function and equivalence to a predicate device, rather than statistical performance metrics of an AI model.

    Given the nature of the device, I will extract and present the information as it relates to the device's functional performance and regulatory clearance process from the provided text.


    Here's the information based on the provided FDA 510(k) summary for the CAPSULS™ Patient Isolation Unit:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of quantitative acceptance criteria with corresponding performance values in the way one would for an AI/ML algorithm (e.g., specific sensitivity or specificity targets). Instead, the "acceptance criteria" can be inferred from the device's intended use and the "performance data" provided, focusing on its ability to isolate and prevent cross-contamination, which was verified through government testing.

    Acceptance Criterion (Inferred from Intended Use)Reported Device Performance and Verification
    Prevention of Particulate Cross-Contamination (Biological & Radiological)Verified by applicable system-level testing conducted by the US government at the West Desert Test Center (WDTC) of the US Army Dugway (Utah) Proving Ground. This testing explicitly "verified the isolation efficacy of the CAPSULS Patient Isolation Unit under both positive and negative pressure operational modes."
    Ability to Maintain Positive/Negative Pressure Operational Modes"Additional testing to measure differential pressure in both operating modes has been successfully conducted." The device incorporates a blower and NIOSH-certified P-100 HEPA filter cartridges (under 42 CFR Part 84) from a powered air purifying respirator (PAPR) system.
    Structural Integrity / Leakage"Pressure/leak testing is conducted on all CAPSULS units prior to shipment." (This is a manufacturing quality control, but also an implicit performance criterion).
    Equivalence to Predicate Device for Intended Use and Principles of Operation"The indications for use and principles of operation of the CAPSULS Patient Isolation Unit are essentially identical to those of the predicate device" (Aircraft Transport Isolator, K790657). The 510(k) notification concluded that the device is "substantially equivalent to the legally marketed predicate device."
    Functionality with Medical Intervention via User-Supplied EquipmentThe device has "multiple glove arms and tubular access ports to facilitate medical intervention." (This is a design feature supporting the intended use rather than a quantified performance metric from a study).
    Portability and Ease of Use (vs. predicate)Stated that "the CAPSULS device is more versatile, lightweight, and easier to use" due to technological advances, although no specific comparative study metrics are provided for these qualitative aspects.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document indicates that "applicable system level testing was conducted on the CAPSULS Patient Isolation Unit." It does not specify a numerical sample size for the test set (e.g., number of individual units tested, or number of test scenarios). It refers to the testing conducted on the unit itself.
    • Data Provenance: The primary performance testing was conducted by the US government at the West Desert Test Center (WDTC) of the US Army Dugway (Utah) Proving Ground. This would be considered prospective testing specifically for this device's validation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable in the context of this device. The "ground truth" for the CAPSULS unit's performance (e.g., isolation efficacy, pressure maintenance) would be established by scientific and engineering measurements in a test environment, not by expert consensus on clinical data or images.

    4. Adjudication Method for the Test Set

    Not applicable. This concept (e.g., 2+1, 3+1) is relevant for clinical expert review of data, typically in AI/ML performance studies, not for the functional testing of a physical isolation unit.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device, and no MRMC study was mentioned or would be relevant for this type of product.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the device's performance was established through physical testing and measurements (e.g., direct measurement of isolation efficacy, differential pressure, leak testing) rather than pathology, expert consensus on clinical outcomes, or observational data. The HEPA filter component's performance (P-100 certification) received NIOSH certification, a specific regulatory standard.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device, not an AI/ML model that requires a training set. The term "training set" is not relevant here.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022260
    Date Cleared
    2002-08-29

    (62 days)

    Product Code
    Regulation Number
    880.5450
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Name: Failsafe Isolation Systems Transport Model 07 and Portable Isolation Model 77 Regulation Number: 880.5450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For Transporting and Isolating a TB patient or a suspected TB patient

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Failsafe Isolation Systems Transport Model 07 and Portable Isolation Model 77). It is a letter from the FDA confirming substantial equivalence to a predicate device. This type of document does not contain details about specific acceptance criteria, study designs, sample sizes, or ground truth establishment for the device itself.

    It focuses on regulatory approval based on equivalence to existing devices, not performance criteria from a clinical trial or algorithm validation study. Therefore, I cannot extract the requested information from the provided text.

    To answer your questions, I would need a different type of document, such as a clinical study report, a device validation report, or a detailed technical submission describing the performance evaluation.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1