Search Results
Found 3 results
510(k) Data Aggregation
(370 days)
pTASWako DCP Control H
Primary Classification Name:
AFP-L3% Immunological Test System
21 CFR $866.6030
Wako uTASWako DCP Calibrator Set Wako µTASWako DCP Control L and Control H Regulation Number: 21CFR§866.6030
The µTASWako AFP-L3 Immunological Test System is an in vitro device that consists of reagents used with the uTASWako i30 Immunoanalyzer to quantitatively measure, by immunochemical techniques, AFP-L3% in human serum. The device is intended for in vitro diagnostic use as an aid in the risk assessment of patients with chronic liver disease for development of hepatocellular carcinoma (HCC) in conjunction with other laboratory findings, imaging studies and clinical assessment. Patients with elevated AFP-L3% values (≥ 10%) have been shown to be associated with an increase in the risk of developing HCC within the next 21 months and should be more intensely evaluated for evidence of HCC according to the existing HCC practice guidelines in oncology.
The uTASWako DCP Immunological Test System is an in vitro device that consists of reagents used with the µTASWako i30 Immunoanalyzer to quantitatively measure, by immunochemical techniques, DCP in human serum. The device is intended for in vitro diagnostic use as an aid in the risk assessment of patients with chronic liver disease for development of hepatocellular carcinoma (HCC) in conjunction with other laboratory findings, imaging studies, and clinical assessment.
The µTASWako i30 Immunoanalyzer is an in vitro diagnostic automated instrument intended for use to quantitatively measure analytes in clinical chemistry by immunochemical techniques. The uTASWako i30 Immunoanalyzer is indicated for use by healthcare professionals. It is intended for assays cleared or approved for use on this instrument.
The Wako uTASWako AFP-L3 Calibrator Set is designed to be used with the Wako µTASWako AFP-L3 Immunological Test System for the quantitative determination of AFP-L3% in human serum.
The Wako µTASWako AFP-L3 Control L is designed to be used as quality control material for the quantitative determination of AFP-L3% in human serum using the Wako µTASWako AFP-L3 Immunological Test System.
The Wako uTASWako AFP-L3 Control H is designed to be used as quality control material for the quantitative determination of AFP-L3% in human serum using the Wako uTASWako AFP-L3 Immunological Test System.
The Wako µTASWako DCP Calibrator Set is designed to be used with the Wako uTASWako DCP Immunological Test System for the quantitative determination of DCP in human serum.
The Wako µTASWako DCP Control L is designed to be used as a quality control material for the quantitative determination of DCP in human serum using the Wako µTASWako DCP Immunological Test System.
The Wako µTASWako DCP Control H is designed to be used as a quality control material for the quantitative determination of DCP in human serum using the Wako uTASWako DCP Immunological Test System.
The µTASWako i30 Immunoanalyzer System is a fully automated immunoassay system that can perform assays of the uTASWako AFP-L3 and µTASWako DCP Immunological Test Systems. This system automatically conducts sampling, mixing, separation, and fluorescence detection on a microfluidic chip to achieve high sensitivity and accuracy. The instrument contains an automated liquid dispenser, temperature controlled reagent container, chip station, analysis compartment, and sample rack station. The outside panel has a printer and a touch panel with a menu to order measurements and to check the availability for reagent, chip, wash solution, and pure water. A chip is used for each test and is disposable. The instrument is designed to automatically and constantly monitor the reagents, chips, dispensing system and the measurement process so that measurement results are not given when an error occurs.
The system is comprised of the following products:
uTASWako i30
uTASWako AFP-L3, Calibrator Set, Control L and Control H
uTASWako DCP, Calibrator Set, Control L and Control H
Instrument and assay accessories as per labeling
The provided text describes the performance data for the µTASWako i30 Immunoanalyzer System and its associated AFP-L3 and DCP immunological test systems. This is a medical device, and the criteria and studies described relate to analytical performance, not clinical diagnostic accuracy or reader studies with human experts.
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Characteristic | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Sensitivity (LoD) | Not explicitly stated; "distinguished from blank" implied. | AFP-L1: 0.030 ng/mLAFP-L3: 0.028 ng/mLDCP: 0.042 ng/mL |
| Linearity/Reportable Range | "Full assay linearity was demonstrated" over claimed ranges. | Total AFP: 0.3 - 1000 ng/mLAFP-L3%: 0.5 - 99.5%DCP: 0.1 - 950 ng/mL |
| High Dose Hook Effect | "No effect" of high concentration. | Total AFP (for AFP-L3 assay): No effect up to 1,272,000 ng/mLDCP (for DCP assay): No effect up to 23,000 ng/mL |
| Within-Run Precision | Total AFP & AFP-L3%: CV% within 10%DCP: CV% within 10% (≥ 1 ng/mL), within 15% (< 1 ng/mL) | Total AFP: 0.7% to 1.5%AFP-L3%: 0.3% to 5.6%DCP: 1.1% to 6.7% |
| Total Precision | Total AFP & AFP-L3%: CV% within 10%DCP: CV% within 10% (≥ 1 ng/mL), within 15% (< 1 ng/mL) | Total AFP: 1.4% to 3.1%AFP-L3%: 0.4% to 6.3%DCP: 1.3% to 7.9% |
| Reproducibility (Instrument to Instrument) | Total AFP & AFP-L3%: CV% within 10%DCP: CV% within 10% | Total AFP & AFP-L3%: 1.6% to 2.7%DCP: 4.9% to 5.6% |
| Recovery | Not explicitly stated; "evaluated the accuracy" implies acceptable recovery. | Total AFP: 97.8% to 104.9%AFP-L3%: 98.1% to 100.9%DCP: 94.0% to 111.6% |
| Interference | "No significant effect" from potential interferents. | Total AFP & AFP-L3%: No significant effect from various interferents.DCP: No significant effect from various interferents (glucose and galactose not tested for DCP). |
| HAMA Interference | "No significant effect" from HAMA interferents. | No significant effect for Total AFP, AFP-L3%, and DCP assays. |
| Method Comparison/Correlation | "Acceptable correlation" with predicate device; Specific concordance rates for clinical cut-offs. | AFP-L3%: Concordance rate of 90.4% (at 10% clinical cut-off) between µTASWako i30 and LiBASys.DCP: Concordance rate of 95.5% (at 7.5 ng/mL clinical cut-off) between µTASWako i30 and LiBASys. |
| Stability | "Demonstrated stability according to the labeled storage conditions." | Long-term stability and stability after opening for reagent, calibrator set, and controls demonstrated. 30-day stability of one-time instrument calibration also supported. |
2. Sample Size Used for the Test Set and Data Provenance
- Sensitivity (LoD): Sample size not specified, but involved distinguishing analytes from blank.
- Linearity/Assay Reportable Range: Sample size not specified.
- High Dose Hook Effect: Sample size not specified.
- Within-Run Precision: Samples with Total AFP 10-950 ng/mL, AFP-L3% 6-80%, DCP 0.2-910 ng/mL. Number of runs/replicates not specified in summary.
- Total Precision: 7 pooled human serum samples and 2 levels of controls. Samples with Total AFP 10-950 ng/mL, AFP-L3% 6-80%, DCP 0.2-910 ng/mL. Measured over 21 days.
- Reproducibility (Instrument to Instrument): 24 instruments were used. Sample details not explicitly stated, but for AFP-L3 and DCP assays.
- Recovery: Sample details not explicitly stated.
- Interference: Tested with known amounts of various interfering substances. Sample details not explicitly stated.
- HAMA Interference: Tested with two types of HAMA interferents. Sample details not explicitly stated.
- Method Comparison/Correlation (AFP-L3 and DCP):
- Test Set: 200 samples from 100 patients.
- Additionally:
- AFP-L3: 40 serum samples spiked with AFP-L1 and AFP-L3.
- DCP: 20 serum samples spiked with DCP.
- Data Provenance: Not explicitly stated, but the studies were conducted to support a US FDA 510(k) submission, suggesting a focus on samples relevant to the intended patient population, likely from clinical settings. It is retrospective in the sense that the samples are collected and then tested.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This document describes the analytical performance of an in vitro diagnostic (IVD) device. The "ground truth" for these types of studies is based on the quantitative measurement of analytes and comparison to a legally marketed predicate device (LiBASys instrument with LBA AFP-L3 and LBA DCP test systems). Therefore, human expert judgment in the diagnostic sense (e.g., radiologists interpreting images) is not applicable or described in this context. The "ground truth" for the method comparison studies would be the results obtained from the predicate device (LiBASys).
4. Adjudication Method for the Test Set
Not applicable. The studies described are analytical performance studies, not subjective diagnostic interpretations requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
Not applicable. This is an IVD immunoassay system, not an AI-assisted diagnostic imaging device that involves human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
The performance data presented is inherently "standalone" in the context of the device's function, as it evaluates the instrument and test systems' analytical capabilities directly. The µTASWako i30 Immunoanalyzer System is a fully automated immunoassay system that performs sampling, mixing, separation, and fluorescence detection. Its output (quantitative measurements of AFP-L3% and DCP) does not involve human interpretation in the same way an AI imaging algorithm would. The method comparison studies compare the new device's automated results directly against a predicate automated device.
7. The Type of Ground Truth Used
The ground truth for the analytical performance studies is largely based on:
- Reference Standards: For sensitivity and linearity, the accurate measurement of known concentrations of analytes.
- Predicate Device Results: For the method comparison/correlation studies, the results obtained from the legally marketed predicate devices (LiBASys instrument with LBA AFP-L3 and LBA DCP test systems) are used as the comparative "ground truth."
- Clinical Cut-off Values: For calculating concordance rates (e.g., 10% for AFP-L3%, 7.5 ng/mL for DCP), these are established clinical values, not derived from a de novo ground truth for the study but rather applied to the measured values.
8. The Sample Size for the Training Set
The document does not describe a "training set" in the context of machine learning or AI development. These are validation studies for an IVD device. The methods described include validation of reagents, calibrators, and system performance through various analytical tests (precision, linearity, recovery, etc.).
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no mention of a "training set" in the context of an AI/ML algorithm. The studies conducted are for analytical validation of an IVD system against established analytical performance metrics and comparison to a predicate device.
Ask a specific question about this device
(170 days)
Device Name: Wako LBA DCP, Wako DCP Calibrator Set, Wako DCP Control Set
Regulation Number: 21 CFR 866.6030
The Wako LBA DCP immunological test system is an in vitro device that consists of reagents and an automated instrument used to quantitatively measure by immunochemical techniques DCP in human serum. The device is intended for in vitro diagnostic use as an aid in the risk assessment of patients with chronic liver disease for progression to hepatocellular carcinoma in conjunction with other laboratory findings, imaging studies and clinical assessment.
The Wako LBA DCP immunological test system is an in vitro device that consists of reagents and an automated instrument used to quantitatively measure by immunochemical techniques DCP in human serum. The Wako LBA DCP assay is test kit for the quantitative determination of DCP based on a new method, LBA (Liquid-phase Binding Assay). The method uses a liquid-phase binding reaction between antigen and antibody and separates bound and free forms by column chromatography without a need for a solid phase. LBA DCP can offer fully automatic and highly precise DCP measurement by using an automated analyzer "LiBASys". This reagent consists of anti-DCP monoclonal antibodies and anti-Prothrombin monoclonal antibodies which are used as Fab' molecules and a substrate for fluorophotometric measurement.
Here's a breakdown of the acceptance criteria and study details for the Wako LBA DCP immunological test system, based on the provided text:
Acceptance Criteria and Reported Device Performance
The provided document defines the device's utility in terms of its ability to identify patients at risk of developing HCC. A specific threshold for DCP concentration (≥ 7.5 ng/mL) is used to categorize risk. The primary performance metric presented is the relative risk.
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Relative Risk of developing HCC with an elevated DCP test result (≥ 7.5 ng/mL) compared to a negative DCP test result (< 7.5 ng/mL) | 4.8 (95% C.I.: 2.8 - 8.4) |
| Risk of HCC given DCP positive (≥ 7.5 ng/mL) | 36.5% (95% C.I.: 23.5% - 49.6%) |
| Risk of HCC given DCP negative (< 7.5 ng/mL) | 7.6% (95% C.I.: 4.4% - 10.8%) |
The study demonstrates that an elevated DCP level (≥ 7.5 ng/mL) is associated with a 4.8-fold increased risk of developing HCC, with a 36.5% risk in the elevated group versus a 7.6% risk in the non-elevated group.
Study Details
-
Sample size used for the test set and the data provenance:
- Total Subjects: 441
- Subjects included in the primary analysis table: 316 (excluding "Suspicious" cases from the main risk calculation table)
- Males: 324
- Females: 117
- Data Provenance: The document states "Longitudinal data was collected," implying prospective or retrospectively analyzed longitudinal data. The country of origin is not explicitly stated in the provided text.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the given text. The ground truth is defined as "HCC" or "No HCC," but the method of establishing this diagnosis and the involvement or qualifications of experts are not described.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- This information is not provided in the given text.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- A multi-reader multi-case (MRMC) comparative effectiveness study was not performed, nor is it applicable. The Wako LBA DCP system is an in vitro diagnostic device for quantitative measurement of a biomarker, not an imaging or interpretive AI system that would assist human readers. It provides a numerical result (DCP concentration) to aid in risk assessment.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone performance study was done for the device in its intended use. The device quantitatively measures DCP, and the risk assessment is based directly on this numerical output compared to a threshold (≥ 7.5 ng/mL). The performance metrics (relative risk, risk percentages) are derived from the device's output.
-
The type of ground truth used:
- The ground truth used is "HCC" (Hepatocellular Carcinoma) or "No HCC." This would typically be established by clinical diagnosis, including imaging, biopsy/histopathology, and other clinical assessments, but the specific methods are not detailed in the provided text. It is presented as clinical outcomes data.
-
The sample size for the training set:
- The provided document describes a "Longitudinal data" study of 441 subjects used to evaluate the device's performance. It does not explicitly mention a separate "training set" for the device's development or for establishing the 7.5 ng/mL threshold. The data presented appear to be from a single evaluation cohort.
-
How the ground truth for the training set was established:
- As no separate training set is explicitly identified, the method for establishing ground truth for the "training set" is not described. For the evaluation data, the ground truth ("HCC" or "No HCC") would have been established through clinical diagnosis, as mentioned in point 6, but the specifics are absent.
Ask a specific question about this device
(42 days)
Regulation section:
- 21 CFR§866.6030, AFP-L3% immunological test system
- 21 CFR§862.1150, Calibrator
The device is classified as Class II under regulation 21 CFR 866.6030 with special controls.
The Wako AFP-L3% assay is intended as a risk assessment test for the development of hepatocellular carcinoma (HCC) in patients with chronic liver diseases (CLD). Elevated AFPL3% values (≥ 10%) have been shown to be associated with a seven-fold increase in the risk of developing HCC within the next 21 months. Patients with elevated serum AFPL3% should be more intensely evaluated for evidence of HCC according to the existing HCC practice guidelines in oncology.
The Wako AFP-L3% device consists of reagent 1 (LCA and anion 1-conjugated anti-AFP mouse monoclonal antibody), reagent 2 (horseradish peroxidase (POD)labeled anti-AFP mouse monoclonal antibody and anion 2 conjugated anti-AFP mouse monoclonal antibody, substrate 1 (4 acetamidophenol in 2-propanol) and substrate 2 (hydrogen peroxide) and a column. Reagent 1, reagent 2 and the column are ready-to-use. Elution buffers A to C, sample cups, inside and outside cuvettes are sold separately from kit.
The Wako AFP-L3 Calibrator set and Control set are sold separately. The calibrator set consisted of Calibrator 1 and 2. Calibrator 1 contains human AFP -L1 fraction and Calibrator 2 has human AFP-L3 fraction. The control set consisted of Control 1 and 2, each containing different concentrations of human AFP-L1 and L-3.
Here's an analysis of the provided 510(k) summary, extracting the requested information about acceptance criteria and the supporting study:
Acceptance Criteria and Device Performance for Wako LBA AFP-L3
1. Table of Acceptance Criteria and Reported Device Performance
The 510(k) summary provided does not explicitly define acceptance criteria in a structured table for each performance characteristic. However, it presents measured performance values and, for linearity, implicitly sets criteria within the reported ranges for slope and R². For clinical performance, the intended use statement sets a clear threshold for risk assessment.
Here's an inferred table based on the provided data:
| Performance Characteristic | Acceptance Criteria (Inferred/Stated) | Reported Device Performance |
|---|---|---|
| Analytical Performance | ||
| Within-run Precision (%CV) | Low (Control 1): AFP <1.5%, AFP-L3 <3.6%, AFP-L3% <3.3% | AFP: 1.5%, AFP-L3: 3.6%, AFP-L3%: 3.3% |
| Medium (Control 2): AFP <1.5%, AFP-L3 <3.7%, AFP-L3% <3.0% | AFP: 1.5%, AFP-L3: 3.7%, AFP-L3%: 3.0% | |
| High (Control 3): AFP <1.4%, AFP-L3 <1.6%, AFP-L3% <0.7% | AFP: 1.4%, AFP-L3: 1.6%, AFP-L3%: 0.7% | |
| Total Precision (%CV) | Low (Sample 1): AFP <5.7%, AFP-L3 <9.6%, AFP-L3% <6.4% | AFP: 5.7%, AFP-L3: 9.6%, AFP-L3%: 6.4% |
| Medium (Sample 2): AFP <3.7%, AFP-L3 <8.0%, AFP-L3% <6.7% | AFP: 3.7%, AFP-L3: 8.0%, AFP-L3%: 6.7% | |
| High (Sample 3): AFP <3.9%, AFP-L3 <5.9%, AFP-L3% <2.7% | AFP: 3.9%, AFP-L3: 5.9%, AFP-L3%: 2.7% | |
| Reproducibility (%CV) | AFP <3.9%, AFP-L3 <6.9% | AFP: 1.0-3.9%, AFP-L3: 1.5-6.9% |
| Recovery | AFP: 95.1-108.4%, AFP-L3: 95.3-118.2% | AFP: 95.1-108.4%, AFP-L3: 95.3-118.2% |
| Linearity (Total AFP) | R² >0.99 for all sites/runs | R² ranges from 0.9974 to 0.9997 |
| Linearity (AFP-L3%) | R² >0.98 for all sites/runs | R² ranges from 0.9835 to 0.9999 |
| Analytical Sensitivity (MDC) | <0.8 ng/mL (claimed) | 0.26 ng/mL |
| Functional Sensitivity | <10 ng/mL | 10 ng/mL |
| HAMA interference (% Rec.) | AFP: 94.5-100.1%, AFP-L3: 95-112.8%, AFP-L3%: 97.9-113.5% | AFP: 94.5-100.1%, AFP-L3: 95-112.8%, AFP-L3%: 97.9-113.5% |
| Clinical Performance | ||
| Risk Assessment | Elevated AFP-L3% (≥ 10%) associated with a seven-fold increase in HCC risk within 21 months | Relative Risk = 7.0 (95% CI: 4.1 to 11.9) for AFP-L3% ≥10% vs. <10% (excluding Group 3) |
| Clinical Cut-off | AFP-L3% >10% considered positive for HCC | Confirmed as threshold for risk assessment |
2. Sample Size Used for the Test Set and Data Provenance
The primary clinical study involved 494 evaluable patients from:
- 6 US clinical sites (Lahey (MA), MCV (VA), Miami (FL), Mt. Sinai (NY), UCSF (SF), UPenn (PA))
- 1 Canadian clinical site (Toronto (Canada))
The study was a double-blind, multi-site prospective study. Serum samples were collected and stored frozen, and AFP-L3% tests were performed retrospectively by Wako Chemicals USA, Inc.
For the relative risk calculation, 57 patients from Group 4 (No HCC) were excluded due to less than 21 months of follow-up, resulting in a dataset of 312 patients (39 HCC, 273 No HCC) being used for the primary relative risk calculation.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
The summary does not explicitly state the number or specific qualifications of experts used to establish the ground truth for the clinical study. However, the diagnosis of HCC for all enrolled patients was made by "at least one or a combination of the following observations," indicating clinical judgment by physician investigators:
- HCC result on a liver biopsy
- Enlarging mass by imaging (ultrasound, CT, MRI) with elevated serum total AFP
- Enlarging mass by CT or MRI in the setting of cirrhosis
- Very high serum total AFP (>400-500 ng/mL) alone
- At least 3 serial blood draws showing rising serum AFP in the setting of a liver mass
- Mass on CT scan enhancing in arterial phase and hypoattenuating in venous phase.
For evaluable subjects without HCC at enrollment, they were categorized by "the physician investigators" based on biopsy, explanted liver histology, and imaging results. This implies that the ground truth was established by the clinical team at each site.
4. Adjudication Method for the Test Set
The adjudication method for the test set is not explicitly described as a formal numerical system (e.g., 2+1, 3+1). Instead, the diagnosis of HCC was based on a combination of clinical observations and confirmatory tests, as determined by the "physician investigators." This suggests a consensus-based or standard clinical practice approach at each site, rather than a centralized, predefined adjudication panel for the study.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was done. This device is a quantitative immunoassay for a biomarker (AFP-L3%) and is not an imaging device or AI-assisted diagnostic tool that would typically involve human readers. The study focuses on the standalone performance of the test in assessing HCC risk.
6. Standalone Performance (Algorithm Only Without Human-in-the-Loop)
Yes, a standalone performance study was done. The entire clinical study described is a standalone evaluation of the Wako AFP-L3% assay. The device measures AFP-L3% in serum, and its performance is evaluated based on its correlation with subsequent HCC development (risk assessment). Human intervention in the diagnostic process occurred after the test result was obtained, not as part of the test interpretation itself. The AFP-L3% values and the 10% cutoff are directly used to stratify patient risk.
7. Type of Ground Truth Used
The ground truth for the clinical study was primarily established through clinical diagnosis based on a combination of pathology (liver biopsy, explanted liver histology), imaging results (CT, MRI, ultrasound), and serial biomarker measurements (total AFP), as determined by physician investigators and standard oncology practice guidelines. This represents a robust clinical ground truth.
8. Sample Size for the Training Set
The 510(k) summary does not specify a separate training set for the clinical performance evaluation. The clinical study described appears to be the primary validation study (test set) for the device's intended use claim. For the analytical performance (precision, linearity, etc.), specific samples (control sera, spiked samples, diluted samples) were used, but these are not referred to as a "training set" in the context of machine learning model development. Given this is an immunoassay, the "training" aspect is more about assay optimization and standard curve generation during assay development than a machine learning training phase.
9. How the Ground Truth for the Training Set Was Established
As no dedicated "training set" is identified in the context of a machine learning algorithm, this question is not directly applicable. For the analytical studies, the ground truth was inherently established by:
- Known concentrations in spiked samples.
- Reference materials (e.g., 1st International standard for AFP from WHO) for calibrator and control value assignments.
- Manufacturer-defined preparations for control sera (e.g., purified AFP-L1 and L-3 spiked into pooled serum).
Ask a specific question about this device