Search Results
Found 3 results
510(k) Data Aggregation
(99 days)
The Flexi-Stat SpO2 Sensor is indicated for continuous, non-invasive functional arterial oxygen saturation and pulse rate monitoring in patients weighing ≥ 30 kg.
The Flexi-Stat(tm) SpO2 Sensor is an electro-optical sensor that functions without skin penetration, electrical contact, or heat transfer. The sensor uses optical means to determine the light absorption of functional arterial hemoglobin by being connected between the patient and the oximeter. The sensor contains three optical components: two light emitting diodes (LED) that serve as light sources and one photodiode that acts as a light detector. The optical components are housed in adhesive film. The sensor cable is terminated in a DB-9 style connector, with an adapter cable for Datexcompatible models.
Here's a summary of the acceptance criteria and study information for the Elekon Flexi-Stat SpO2 Sensor based on the provided 510(k) submission:
Acceptance Criteria and Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Clinically equivalent to predicate devices (Nellcor, BCI, Datex, Envitec sensors) in terms of safety and effectiveness. | "Performance testing was conducted during clinical hypoxia studies conducted in an independent laboratory. In these studies, accuracy of the Flexi-Stat was compared to arterial blood samples analyzed on a laboratory co-oximeter and found to be equivalent to predicate device accuracy claims." |
Biocompatibility | Conformed to established industry standards. |
Electrical Safety | Conformed to established industry standards. |
EMC Testing | Conformed to established industry standards. |
Note: The document does not provide specific numerical targets for SpO2 accuracy (e.g., standard deviation or root mean square error) within a defined range. Instead, it refers to equivalence to predicate device accuracy claims.
Study Details:
-
Sample size used for the test set and the data provenance:
- Sample Size: Not explicitly stated. The document mentions "clinical hypoxia studies" without specifying the number of subjects or data points.
- Data Provenance: The studies were conducted in an "independent laboratory." The country of origin is not specified but given the submitter is "Elekon Industries, USA, Inc." it's likely they were conducted in the USA or a region with comparable regulatory standards. The study appears to be prospective, involving induced hypoxia, rather than retrospective.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/Not specified for SpO2 accuracy. The ground truth for SpO2 accuracy was established by "arterial blood samples analyzed on a laboratory co-oximeter." There is no mention of experts interpreting these co-oximeter results; rather, the co-oximeter itself provides the objective measurement.
-
Adjudication method for the test set:
- Not applicable. The ground truth was based on objective laboratory measurements (co-oximetry) of arterial blood samples, not subjective assessments requiring expert adjudication.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is an SpO2 sensor, which provides a direct physiological measurement, not an AI-assisted diagnostic tool that requires human interpretation.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- This is a standalone device in the sense that its performance, as an accessory to a pulse oximeter, is evaluated based on its direct measurements compared to a gold standard (co-oximetry), without requiring a human-in-the-loop to interpret its output for the accuracy study. The sensor outputs a numerical SpO2 value.
-
The type of ground truth used:
- Objective measurement (Laboratory Co-Oximetry): Arterial blood samples were analyzed on a laboratory co-oximeter to provide the gold standard for arterial oxygen saturation.
-
The sample size for the training set:
- Not applicable. As a medical device (sensor) that measures a physiological parameter, there is no "training set" in the machine learning sense for the core SpO2 measurement algorithm itself, which is based on established electro-optical principles. The performance testing focuses on validation.
-
How the ground truth for the training set was established:
- Not applicable, as there is no training set in the context of this device's SpO2 measurement function.
Ask a specific question about this device
(77 days)
The Flexi-Stat SpO2 Sensor is indicated for continuous, non-invasive functional arterial oxygen saturation and pulse rate monitoring. The Flexi-Stat SpO2 Sensor is indicated for use in patients weighing >30 kg.
The Flexi-Stat(tm) SpO2 Sensor is an electro-optical sensor that functions without skin penetration, electrical contact, or heat transfer. The sensor uses optical means to determine the light absorption of functional arterial hemoglobin by being connected between the patient and the oximeter. The sensor contains three optical components: two light emitting diodes (LED) that serve as light sources and one photodiode that acts as a light detector. The optical components are housed in a durable finger clip housing. The sensor cable is terminated in a DB-9 style connector, with an adapter cable for Datex-compatible models.
Here's a breakdown of the acceptance criteria and study information for the Elekon Flexi-Stat SpO2 Sensor, based on the provided text:
Acceptance Criteria and Device Performance
Acceptance Criteria Category | Specific Criteria/Outcome Demonstrated | Reported Device Performance |
---|---|---|
Accuracy (SpO2) | Equivalent to predicate device accuracy claims. | "found to be equivalent to predicate device accuracy claims." |
Biocompatibility | Conformance with established industry standards. | Performed; conformance demonstrated. |
Electrical Safety | Conformance with established industry standards. | Performed; conformance demonstrated. |
EMC (Electromagnetic Compatibility) | Conformance with established industry standards. | Performed; conformance demonstrated. |
Intended Use | Continuous, non-invasive functional arterial oxygen saturation and pulse rate monitoring. For patients weighing >30 kg. | Device is indicated for this use. |
Principle of Operation | Uses the same theory and principle of operation as predicate device. | Confirmed. |
Design Characteristics | Equivalent to predicate devices in terms of safety and effectiveness. | Confirmed by testing and accuracy claims. |
Study Details
2. Sample size used for the test set and the data provenance:
- Sample Size: Not explicitly stated. The text mentions "clinical hypoxia studies" without specifying the number of subjects.
- Data Provenance: Clinical hypoxia studies conducted in an independent research lab. The country of origin is not specified, but the submission is from Elekon Industries, USA, Inc.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: Not specified.
- Qualifications of Experts: Not specified. However, the ground truth was "arterial blood samples analyzed on a laboratory co-oximeter," implying that experts in laboratory co-oximetry would have been involved in analyzing these samples.
4. Adjudication method for the test set:
- Not applicable. The ground truth was established through direct comparison with arterial blood samples analyzed on a co-oximeter, not through expert consensus requiring adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is an SpO2 sensor, not an AI-assisted diagnostic tool for which MRMC studies are typically performed. The study focused on the sensor's accuracy compared to a reference standard.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, in essence. The performance testing described is standalone for the device. The Flexi-Stat SpO2 sensor's readings were compared directly to the ground truth from arterial blood samples without human interpretation being part of the device's measurement process, though human operators would connect and read the oximeter display. The focus was on the sensor's accuracy.
7. The type of ground truth used:
- Reference Standard: Arterial blood samples analyzed on a laboratory co-oximeter. This is considered a gold standard for blood oxygen saturation measurement.
8. The sample size for the training set:
- Not applicable / Not specified. This device is a physiological sensor, not a machine learning algorithm that typically requires a distinct training set. The performance testing validates its function.
9. How the ground truth for the training set was established:
- Not applicable. As a non-AI physiological sensor, there isn't a "training set" in the machine learning sense. The ground truth for performance evaluation was established by comparing sensor readings to concurrently drawn and analyzed arterial blood samples using a laboratory co-oximeter.
Ask a specific question about this device
(42 days)
The Flexi-Stat SpO2 Sensor is indicated for continuous, non-invasive functional arterial oxygen saturation and pulse rate monitoring in adult, pediatric, and neonatal patients.
The Flexi-Stat(tm) SpO2 Sensor is an electro-optical sensor that functions without skin penetration, electrical contact, or heat transfer. The sensor uses optical means to determine the light absorption of functional arterial hemoglobin by being connected between the patient and the oximeter. The sensor contains three optical components: two light emitting diodes (LED) that serve as light sources and one photodiode that acts as a light detector. The optical components are housed in a foam and Velcro wrap. The sensor cable is terminated in a DB-9 style connector, with an adapter cable for Datex-compatible models.
This submission describes the Flexi-Stat SpO2 Sensor, an accessory to a pulse oximeter for continuous, non-invasive functional arterial oxygen saturation and pulse rate monitoring. The summary provided primarily focuses on the substantial equivalence of the Flexi-Stat SpO2 Sensor to existing predicate devices, rather than detailed acceptance criteria and supporting studies for a novel or significantly modified device. Therefore, some information requested, particularly regarding specific numerical acceptance criteria and a detailed study design as might be seen for a new algorithmic device, is not explicitly present in the provided text.
Here's the breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance
Specific numerical acceptance criteria (e.g., accuracy ranges for SpO2 measurements) are not explicitly stated in the provided text for the Flexi-Stat SpO2 Sensor itself. Instead, the document states: "The Flexi-Stat was compared to arterial blood samples analyzed on a laboratory co-oximeter and found to be equivalent to predicate device accuracy claims." This implies that the acceptance criteria for the Flexi-Stat SpO2 Sensor were to meet or be equivalent to the accuracy claims of its predicate devices.
Acceptance Criteria Category | Acceptance Criteria (Implicit) | Reported Device Performance |
---|---|---|
Accuracy (SpO2) | Equivalent to predicate device accuracy claims | "found to be equivalent to predicate device accuracy claims" based on comparison to arterial blood samples analyzed on a laboratory co-oximeter. |
Biocompatibility | Conformance with established industry standards | "Biocompatibility... was also performed to demonstrate conformance with established industry standards." |
Electrical Safety | Conformance with established industry standards | "electrical safety... was also performed to demonstrate conformance with established industry standards." |
EMC (Electromagnetic Compatibility) | Conformance with established industry standards | "EMC testing was also performed to demonstrate conformance with established industry standards." |
Principle of Operation | Same as predicate device | "The Flexi-Stat SpO2 Sensor uses the same theory and principle of operation as the predicate device." |
Design Characteristics | Equivalent to predicate device in terms of safety and effectiveness | "Design characteristics are equivalent in terms of safety and effectiveness, as demonstrated by product testing and accuracy claims." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: This information is not explicitly stated in the provided text. The document mentions "clinical hypoxia studies" but does not give the number of subjects.
- Data Provenance: The studies were described as "clinical hypoxia studies conducted in an independent research lab." No specific country of origin is mentioned, but the company is US-based. It is a prospective study as clinical hypoxia studies involve actively inducing hypoxia in subjects under controlled conditions.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This information is not applicable to this type of device and study. The ground truth was established using an objective measurement from a laboratory co-oximeter and arterial blood samples, not through expert consensus on interpretations.
4. Adjudication Method for the Test Set
This information is not applicable. The ground truth was based on objective laboratory measurements (co-oximetry of arterial blood samples), not on expert interpretation requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for devices involving human interpretation (e.g., radiology AI), not for objective physiological measurement devices like an SpO2 sensor where the output is a direct numerical reading.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done
Yes, the accuracy assessment described ("The Flexi-Stat was compared to arterial blood samples analyzed on a laboratory co-oximeter") represents a standalone performance evaluation of the device's measurement capabilities. There is no human intervention mentioned in the process of generating the SpO2 values from the Flexi-Stat sensor and comparing them to the ground truth.
7. The Type of Ground Truth Used
The ground truth used was objective physiological measurement:
- "arterial blood samples analyzed on a laboratory co-oximeter"
8. The Sample Size for the Training Set
This information is not applicable as the Flexi-Stat SpO2 Sensor is not a machine learning or AI-based device that typically requires a "training set" in the conventional sense. Its function is based on electro-optical principles, not learned patterns from data.
9. How the Ground Truth for the Training Set was Established
This information is not applicable for the reasons mentioned above.
Ask a specific question about this device
Page 1 of 1