Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K082727
    Date Cleared
    2009-02-06

    (142 days)

    Product Code
    Regulation Number
    884.6100
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K991273

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Follicle Aspiration Set, Reduced Single Lumen is intended for flushing and/or aspiration of oocytes from ovarian follicles

    Device Description

    The Follicle Aspiration Set, Reduced Single Lumen device is a modification of the current Vitrolife Follicle Aspiration Set (K991273). The Follicle Aspiration Set, Reduced Single has the same intended use as the predicate device technological characteristics related to safety and effectiveness. A thinner type of needle end is being introduced, and the last 40-60 mm close to the tip of a standard needle is reduced from an inner diameter of 1.2 mm to an inner diameter of 0.6 mm.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Vitrolife Follicle Aspiration Set, Reduced Single Lumen.

    Important Note: The provided document is a 510(k) Premarket Notification Summary, which focuses on demonstrating substantial equivalence to a predicate device. It primarily details the safety and effectiveness testing required to prove this equivalence, rather than setting and meeting formal "acceptance criteria" in the way a clinical trial for a novel device would. Therefore, the "acceptance criteria" here are inferred from the safety tests performed.


    Acceptance Criteria and Device Performance

    Acceptance Criteria (Inferred from Safety Concerns)Reported Device Performance
    Resistance to breakage of the reduced needle part during procedure. (Main safety concern due to thinner diameter)The needle has undergone testing according to ISO 9626 with respect to resistance of tubing to breakage and passed. The needle also passed an extended angle test. It's noted that the wall of the reduced part is at least 50% thicker than the wall of the standard part, contributing to its strength.
    Maintenance of intended use and technological characteristics related to safety and effectiveness (compared to predicate)The device has the same intended use and technological characteristics related to safety and effectiveness as the predicate device (Vitrolife Follicle Aspiration Set - K991273). The change is a thinner needle end.

    Study Details

    Given the nature of the 510(k) submission and the device (a modification of an existing medical instrument), the "study" is primarily focused on bench testing to demonstrate the safety of the design change. It's not a clinical study with human participants or a traditional AI/software efficacy study.

    1. Sample sizes used for the test set and data provenance:
    * Test Set Sample Size: Not explicitly stated as a number of individual needles. The testing is described as occurring "according to ISO 9626" and also an "extended angle test." ISO standards typically define sample sizes for such tests, but the specific number used by Vitrolife is not reported here.
    * Data Provenance: Not applicable in the context of a clinical study. This is a report on internal bench testing conducted by the manufacturer, Vitrolife Sweden AB.

    2. Number of experts used to establish the ground truth for the test set and qualifications of those experts:
    * This is not applicable as the "ground truth" for mechanical component testing is defined by engineering standards (e.g., ISO 9626) and passing specific physical tests, not by expert consensus on clinical data.

    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
    * Not applicable. Adjudication methods like 2+1 are used for interpreting clinical data, not for bench testing. The passing or failing of these mechanical tests is objective.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    * No. This is a physical medical device (a needle), not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study is irrelevant.

    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
    * No. This is a physical medical device, not an algorithm.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    * The "ground truth" for the safety tests is based on engineering standards and specifications (specifically ISO 9626 for resistance to breakage and an "extended angle test"). The device passing these prescribed mechanical tests constitutes the "ground truth" for demonstrating its structural integrity.

    7. The sample size for the training set:
    * Not applicable. This device is a modified physical instrument, not an AI model, so there is no "training set."

    8. How the ground truth for the training set was established:
    * Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K991273
    Date Cleared
    1999-08-18

    (127 days)

    Product Code
    Regulation Number
    884.6100
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K991273

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Swemed Follicle Aspiration Sets are intended to be used to obtain gametes from the body. The Swemed Follicle Aspiration Sets are indicated for use in flushing and/or aspiration of oocytes from ovarian follicles.

    Device Description

    The Swemed Follicle Aspiration Sets are designed with either a single lumen which is capable for aspiration from the follicle or a double lumen which is capable of aspiration and flushing of sterile medium through the outer lumen up to the follicle. The needle has a beveled tip provided with shallow grooves to improve ultrasound image enhancement. The needle are supplied with a protective tube on the tip, which must be removed before use. The needles have inner diameters of 0.8 - 1.2 mm and outer diameters of 1.47 - 1.65 mm, which approximates a 16 and 17 gauge needle, respectively. The needle length ranges from 25 to 35 cm. The set is packed in an inner and outer wrapping, and steam sterilized.

    AI/ML Overview

    The provided text focuses on the 510(k) summary for the Swemed Follicle Aspiration Set, demonstrating its substantial equivalence to a predicate device. This type of submission relies on comparisons to existing legally marketed devices rather than extensive new clinical studies with detailed acceptance criteria and performance metrics. Therefore, many of the requested details about acceptance criteria, specific studies, ground truth establishment, and expert involvement are not present in this document.

    Here's a breakdown of the information that can be extracted and a clear indication of what is not available based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    This information is not provided in the document. The 510(k) submission for the Swemed Follicle Aspiration Set is based on demonstrating substantial equivalence to a predicate device (Cook OB/GYN Ovum Pick-Up Aspiration Needles). Substantial equivalence means the device is as safe and effective as a legally marketed predicate device, not that it met specific quantitative performance criteria in a de novo study with acceptance thresholds. The document states:

    "The Swemed Follicle Aspiration Sets and the Cook Ovum Pick-Up Needles share the exact same intended use: to obtain oocytes from the body and the exact same indication for use: flushing and/or aspiration oocytes from ovarian follicles. Except for four minor features, the Swemed Follicle Aspiration Sets and the Cook Ovum Pick-Up Needles are nearly identical. ... These minor differences do not affect the safety or effectiveness of the device."

    This implies that the "acceptance criteria" were met by demonstrating that the device's technological characteristics and intended/indications for use are sufficiently similar to a device already deemed safe and effective, and that the minor differences do not negatively impact safety or effectiveness. No specific performance metrics (e.g., aspiration efficiency, follicle retrieval rate) are provided or compared.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided. The document describes the device's features and its proposed equivalence to a predicate device, but it does not detail any specific clinical testing or studies with a "test set" in the context of typical AI/diagnostic device validation. There is no mention of data provenance (e.g., country of origin, retrospective/prospective).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided. As no specific test set or study is detailed, there's no mention of experts establishing a ground truth.

    4. Adjudication Method for the Test Set

    This information is not provided.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of an MRMC study and, crucially, this device is a medical instrument (needles), not an AI-powered diagnostic tool. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not applicable as the device is a medical instrument, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    This information is not provided. Since the submission relies on substantial equivalence and not a de novo performance study, there's no detailed discussion of ground truth. The "ground truth" implicitly references the established safety and effectiveness of the predicate device.

    8. The Sample Size for the Training Set

    This information is not provided. As this is not an AI/algorithmic device, there would be no "training set."

    9. How the Ground Truth for the Training Set was Established

    This information is not provided and is not applicable for this type of device and submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1