Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K993803
    Date Cleared
    2000-02-06

    (89 days)

    Product Code
    Regulation Number
    892.1000
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K922798, K965068, K961092, K942609, K983110

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Imaging of: The Whole Body (including head, abdomen, breast, heart, pelvis, joints, neck, TMJ, spine, blood vessels, limbs and extremities). [Application terms include MRCP (MR Cholangiopancreatography), MR Cisternography, MR Urography, MR Myelography, MR Fluoroscopy, SAS (Surface Anatomy Scan), Dynamic Scan, Cine Imaging, and Cardiac tagging.] Fluid Visualization 2D/3D Imaging MR Angiography/MR Vascular Imaging Blood Oxygenation Level Dependent (BOLD) Imaging Diffusion Imaging The EPI Plus optional package adds the following to the above indications: Perfusion Imaging

    Device Description

    This submission consists of a new Magnetic Resonance Imaging system which is based primarily on the cleared EXCELARTTM (K990620) with the following new functions added to the standard system. • Pianissimo: Gradient acoustic reduction system • QD Whole Body coil • QD Head coil • QD C/T/L Spine Array coil • 18.1" LCD flat screen monitor • Increase of SAR limit from

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a Magnetic Resonance Diagnostic Device Accessory (MRI system), specifically the EXCELART™ with Pianissimo. The submission is focused on demonstrating substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria for a novel AI algorithm or diagnostic tool.

    Therefore, many of the requested categories (e.g., sample sizes for test/training sets, ground truth establishment, expert qualifications, MRMC studies, standalone performance) are not applicable or not explicitly detailed in this type of submission for this kind of device. The focus here is on the safety and performance characteristics of the MRI system itself and its accessories.

    Here's a breakdown based on the provided text, indicating what information is available and what is not:


    1. Table of Acceptance Criteria and Reported Device Performance

    For an MRI system, performance is generally evaluated against safety parameters and imaging capabilities rather than diagnostic accuracy metrics. The document specifies safety parameters and imaging volumes.

    Acceptance Criteria CategoryAcceptance Criteria (from text)Reported Device Performance (from text)
    Safety Parameters
    Maximum static field strength- (Implied to meet predicate's 1.5T)1.5T
    Rate of change of magnetic field- (Implied to meet predicate's ~19.5 T/sec.)19.35 T/sec. (Pianissimo) vs 19.5 T/sec (EXCELART™)
    Maximum radio frequency power (SAR)
    Ask a Question

    Ask a specific question about this device

    K Number
    K993574
    Date Cleared
    2000-01-18

    (89 days)

    Product Code
    Regulation Number
    892.1000
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K962933, K983110, K962138, K946244/A1, K933018/S1

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Imaging of: The Whole Body (including head, abdomen, pelvis, limbs and extremities, spine, neck, TMJ, heart, blood vessels). [Application terms include MRCP (MR Cholangiopancreatography), MR Urography, MR Myelography, MR Fluoroscopy, SAS (Surface Anatomy Scan), Dynamic Scan and Cine Imaging.] -Fluid Visualization -2D/3D Imaging Additional indications for v3.1, v3.2, & v3.3 (only) - MR Angiography/MR Vascular Imaging - Additional indication for v3.2 & v3.3 (only) - Water/Fat Imaging - Additional indication for v3.3 {only} Perfusion/Diffusion Imaging -

    Device Description

    Versions v3.0/v3.1/v3.2/v3.3 software are a combination of modifications and the addition of new sequences to the existing software, which facilitate the acquisition and reconstruction of MR images. The four versions have the same base software features with certain additional features available in each subsequent version (see Comparison Table, Appendix B, for detailed description). A brief description follows: - v3.0: Based on v2.5 (K990260) with MR Angio and FASE sequences removed - v3.1: Based on v2.5 (K990260) - v3.2: Based on v2.6 (K990260) - v3.3: Based on v2.6 (K990260) with addition of Perfusion/Diffusion imaging

    AI/ML Overview

    This 510(k) premarket notification describes an upgrade to an existing Magnetic Resonance Diagnostic Device, the OPART™ (Model MRT-600), to software versions v3.0, v3.1, v3.2, and v3.3, along with optional hardware items. The core of this submission is to demonstrate substantial equivalence to previously cleared versions and to justify new functionalities and increased safety parameters.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission focuses on safety parameters and imaging performance. The "acceptance criteria" appear to be specified maximums for safety and a general "specification volume" for imaging. The "reported device performance" is essentially that the device operates within these stated limits and produces sample images.

    ParameterAcceptance CriteriaReported Device Performance
    Maximum static magnetic field strengthSpecified (0.35 Tesla)0.35 Tesla (inherent to the device model)
    Rate of change of magnetic fieldSpecified (19 T/second)19 T/second (inherent to the device model)
    Maximum radio frequency power deposition (SAR)
    Ask a Question

    Ask a specific question about this device

    K Number
    K990620
    Device Name
    EXCELART
    Date Cleared
    1999-05-14

    (78 days)

    Product Code
    Regulation Number
    892.1000
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K983110

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Imaging of:

    • The Whole Body (including head, abdomen, breast, heart, pelvis, joints, neck, TMJ, . spine, blood vessels , limbs and extremities). [Application terms include MRCP (MR Cholangiopancreatography), MR Urography, MR Myelography, MR Fluoroscopy, SAS (Surface Anatomy Scan), Dynamic Scan, Cine Imaging, and Cardiac tagging.]
    • . Fluid Visualization
    • 2D/3D Imaging
    • MR Angiography/MR Vascular Imaging
    • Blood Oxygenation Level Dependent (BOLD) Imaging .
    Device Description

    This submission consists of a new Magnetic Resonance Imaging system which is based primarily on the cleared VISART™/Hyper system (K965068) with modifications including a short bore magnet and gantry design. The software used with the EXCELART MRI system is V4 (K983110). All other technology used for the EXCELART system is the same as that of the cleared VISART™/Hyper system.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for the EXCELART Magnetic Resonance Diagnostic Device Accessory (MRI System). This document focuses on demonstrating substantial equivalence to a predicate device, rather than providing independent clinical study results with detailed acceptance criteria and performance metrics for a new, non-inferiority claim.

    Therefore, many of the requested details about acceptance criteria, specific study designs, sample sizes, and expert adjudications are not explicitly stated or applicable in the context of this 510(k) submission.

    Here's an analysis of the provided information relative to your request:

    1. A table of acceptance criteria and the reported device performance

    The submission does not outline specific, quantitative acceptance criteria for image quality or diagnostic accuracy in the way a clinical study for a new AI device would. Instead, it demonstrates that the EXCELART system's safety and imaging performance parameters are comparable to its predicate device, the VISART™/Hyper system.

    ParameterAcceptance Criteria (Predicate)Reported Device Performance (EXCELART)
    Safety Parameters
    Maximum static field strength1.5T (VISART™/Hyper)1.5T
    Rate of change of magnetic field19.5 T/sec. (VISART™/Hyper)19.35 T/sec.
    Maximum radio frequency power
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1