Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K061820
    Manufacturer
    Date Cleared
    2007-02-26

    (243 days)

    Product Code
    Regulation Number
    866.3900
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K981867

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIAISON® VZV IgG Assay uses chemiluminescence immunoassay (CLIA) technology on the LIAISON® Analyzer for the qualitative detection of specific IgG antibodies to varicella-zoster virus in human serum. This assay can be used as an aid in the determination of previous infection of varicella-zoster virus.

    Device Description

    The method for the qualitative determination of specific IgG to varicella- zoster virus is an indirect chemiluminescence immunoassay (CLIA). All assay steps and incubations are performed by the LIAISON® Analyzer.

    Varicella-zoster virus antigen is used for coating magnetic particles (solid phase) and a mouse monoclonal antibody to human IgG is linked to an isoluminol derivative (isoluminol-antibody conjugate). During the first incubation, anti-VZV IgG antibodies, present in calibrators, samples or controls, bind to the solid phase. After each incubation, the unbound material is removed with a wash cycle. During the second incubation, the antibody conjugate reacts with anti-VZV IgG already bound to the solid phase. Subsequently, the starter reagents are added and a flash chemiluminescence reaction is induced. The light signal, directly related to the amount of isoluminol-antibody conjugate, is measured by a photomultiplier as relative light units (RLU) and is indicative of the presence of anti-VZV IgG in calibrators, samples or controls.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the DiaSorin LIAISON® VZV IgG device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the clinical trial results. The device aims to demonstrate substantial equivalence to the predicate device.

    MetricAcceptance Criteria (Implied)Reported Device Performance (General Samples)Reported Device Performance (Pregnant Women)
    Positive AgreementHigh agreement with predicate device for positive samples98.8% (97.7 - 99.5% 95% CI)99.2% (98.2 - 99.7% 95% CI)
    Negative AgreementHigh agreement with predicate device for negative samples84.4% (74.4 - 91.7% 95% CI)64.1% (47.6 - 78.8% 95% CI)
    Overall ReproducibilityConsistent results across runs, sites, and operators (low %CV)See detailed tables belowSee detailed tables below
    Assay PrecisionConsistent results over time on a single instrument (low %CV)See detailed tables belowSee detailed tables below

    Reproducibility (3-site, 5-day study with 3 kit lots):

    ID#NMean Index (Overall %CV)
    DiaSorin Neg Ctl6030.3 (18.6)
    DiaSorin Pos Ctl60434 (14.4)
    011006 (Cutoff Ctl)60246 (13.8)
    BR Neg Ctl (100% serum)60<10 (8.24)
    BR Pos Ctl (100% serum)60863 (12.6)
    33146060.2 (9.54)
    336060265 (12.9)
    338560242 (13.8)
    340360164 (9.14)
    349260276 (12.3)
    351560252 (14.5)
    355460291 (14.3)
    Pos 5601530 (19.0)
    Pos 960679 (12.2)

    Assay Precision (single-site, 20-day study with 1 kit lot):

    ID#NMean Index (Overall %CV)
    DiaSorin Neg Ctl4036.3 (20.05)
    DiaSorin Pos Ctl40446 (20.07)
    011006 (Cutoff Ctl)40271 (17.01)
    BR Neg Ctl (100% serum)40<10 (12.62)
    BR Pos Ctl (100% serum)40952 (19.25)
    33144060.8 (13.00)
    336040318 (16.85)
    338540292 (23.02)
    340340164 (14.80)
    349240330 (16.62)
    351540280 (14.79)
    355440286 (18.00)
    Pos 5401761 (19.79)
    Pos 940769 (16.50)

    2. Sample Size Used for the Test Set and Data Provenance

    • General Samples: 745 total samples.

      • 33 equivocal results were excluded from agreement calculations (implied by the statement: "Specimens that were equivocal by both assays were not included in the percent agreement calculation").
      • Resulting in 667 for positive agreement calculation and 77 for negative agreement calculation (after adjusting for equivocal results in the reference assay matched with positive/negative device results).
    • Pregnant Women Samples: 689 total samples.

      • 11 equivocal results in the predicate assay and 3 equivocal results in the device were handled in agreement calculation.
      • Resulting in 650 for positive agreement calculation and 39 for negative agreement calculation (after adjusting for equivocal results in the reference assay matched with positive/negative device results).
    • Reproducibility/Precision Samples:

      • Reproducibility (3 sites): 9 coded frozen repository samples, plus 4 controls (DiaSorin Neg Ctl, DiaSorin Pos Ctl, Cutoff Ctl, BR Neg Ctl, BR Pos Ctl). Each sample/control tested 60 times (implied 3 sites * X runs * Y replicates per run, totaling 60).
      • Precision (1 site): The same serum samples and controls as the reproducibility study. Each sample/control tested 40 times (quadruplicate over 20 working days).
    • Data Provenance: Prospectively-collected U.S. samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The ground truth was established by a "comparison assay" (Diamedix Is-VZV IgG Test System, K981867), which is another immunoassay. No human experts are mentioned as establishing the ground truth for the clinical agreement studies. For the reproducibility and precision studies, a coded panel of samples was used, but the initial "truth" of these samples (negative, low positive, moderate to high positive) is not attributed to experts.

    4. Adjudication Method for the Test Set

    No explicit adjudication method (like 2+1 or 3+1) is mentioned. The comparison was directly between the LIAISON® VZV IgG assay and the predicate VZV IgG ELISA. Equivocal results were handled by excluding them from percent agreement calculations or by considering them as non-agreements when the reference assay was equivocal and the LIAISON® result was positive/negative.

    5. Multi-reader Multi-case (MRMC) Comparative Effectiveness Study

    No multi-reader, multi-case (MRMC) comparative effectiveness study was done. This study focuses on the performance of an automated immunoassay device, not on human reader performance with or without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, the studies presented are all standalone. The DiaSorin LIAISON® VZV IgG is an automated chemiluminescence immunoassay (CLIA) system. Its performance is measured directly against a predicate immunoassay (clinical agreement) and its own internal consistency (reproducibility and precision). There is no human-in-the-loop component described for these performance evaluations.

    7. Type of Ground Truth Used

    The "ground truth" for the clinical agreement studies was established by a predicate immunoassay (VZV IgG ELISA). For reproducibility and precision, it was based on known control materials and coded repository samples with established characteristics (negative, low-positive, high-positive).

    8. Sample Size for the Training Set

    The document does not provide information on the sample size used for the training set of the DiaSorin LIAISON® VZV IgG assay. The provided data is for the validation/test phase of the device.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for the training set (if any was used for algorithm development) was established. The information focuses on the validation of the device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1