Search Results
Found 2 results
510(k) Data Aggregation
(193 days)
Quantum SuperPAC Tubing set is a set of tubing intended for use during cardiopulmonary bypass for a duration up to 6 hours.
Quantum SuperPAC Cardioplegia Set is a tubing set used for the infusion of cardioplegia solutions and physiological fluids during cardiac surgery procedures on the heart and great vessels for up to six hours.
Quantum SuperPAC Tubing set devices are designed to connect different devices that are not provided in the Quantum SuperPAC tubing set such as oxygenators, pumps, reservoirs, filters, and other cardiopulmonary bypass components into circuits used in surgical procedures requiring extracorporeal support, for a maximum duration of 6 hours.
Quantum SuperPAC Cardioplegia set devices are designed to connect different devices that are not provided in the Quantum SuperPAC tubing set for the infusion of cardioplegia solutions and physiological fluids during cardiac surgery procedures on the heart and great vessels for up to six hours.
Quantum SuperPAC devices (all variants) are non-toxic, non-pyrogenic, and sterilized by ethylene oxide. Devices are intended for single use only and are not to be resterilized by the user.
All the device surfaces in contact with blood are treated with a phosphorylcholine-based coating.
Quantum SuperPAC devices (all variants) are mainly constituted of polyvinyl chloride (PVC) DOP free tubing and additional components composing the set; different variants are available, varying for tubing dimension and set configuration in order to address customer and surgical procedure specifications.
This document is a 510(k) summary for the Quantum SuperPAC Tubing Set and Quantum SuperPAC Cardioplegia Set. It details the device's intended use, comparison to predicate devices, and performance testing. However, it does not contain information regarding studies that establish acceptance criteria for device performance in the manner of an AI/ML algorithm's effectiveness in diagnostic tasks, nor does it provide details on sample sizes, ground truth establishment, or expert adjudication typical for such studies.
Instead, this document focuses on demonstrating substantial equivalence to existing medical devices through non-clinical performance and biocompatibility testing for a medical device (tubing sets) rather than a diagnostic AI system.
Therefore, many of the requested categories are not applicable to the information provided in this 510(k) summary, as it pertains to a physical medical device.
Here's the breakdown based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of acceptance criteria alongside reported device performance in the context of an AI/ML diagnostic system. Instead, it states that "All testing passed by meeting the established requirements set for the devices." The performance tests conducted are listed as:
- Operating Parameters
- Mechanical Integrity
- Device pressure drop
- Spallation and Tubing Life
- Connection strength
These tests were mainly performed according to ISO 15676. Additional tests included:
- Evaluation of product shelf life (according to EP/UPS requirements)
- Validation of EtO Sterilization process (according to ISO 11135:2014)
- Packaging Validation tests (according to ISO 11607-1:2019, ASTM F1886/F1886M-16, EN 868-5 and ASTM F1929-15)
- Biocompatibility (according to ISO 10993-1:2018 and FDA Guidance)
The document asserts that the devices met these established requirements, implying that the acceptance criteria were defined by these standards and the device's performance aligned with them.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not applicable as the document describes non-clinical performance testing for a physical medical device, not a diagnostic AI/ML system requiring a test set of data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable as the document describes non-clinical performance testing for a physical medical device, not a diagnostic AI/ML system requiring expert-established ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable as the document describes non-clinical performance testing for a physical medical device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable as the document describes non-clinical performance testing for a physical medical device, not a diagnostic AI/ML system.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable as the document describes a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
This information is not applicable for the type of device described. The "ground truth" for this device would be its adherence to performance specifications and safety standards as determined by the listed non-clinical tests (e.g., mechanical integrity tests, biocompatibility tests).
8. The sample size for the training set
This information is not applicable as the document describes a physical medical device, not a machine learning model.
9. How the ground truth for the training set was established
This information is not applicable as the document describes a physical medical device, not a machine learning model.
Ask a specific question about this device
(28 days)
The SMAR, T® BCD Vanguard™ Blood Cardioplegia System is intended to mix, cool, warm and deliver oxygenated blood and cardioplegic solution for periods up to six hours. The device also allows monitoring of temperature and pressure, traps bubbles, and removes air. Blood and cardioplegic solution are delivered to the patient by a single 100 % occlusive roller pump, through the extension line and appropriate cannula.
The SMAR, T BCD Vanguard System is a cardioplegia heat exchanger with integral bubble trap, available with various tubing ratio configurations connected to the heat exchanger assembly. Surface-modifying materials have been added to the primary blood contact surfaces of the heat exchanger, tubing and connectors to improve the blood compatibility of the system. The product is sterilized by ethylene oxide, is for single use only, and has nonpyrogenic fluid pathways.
The heat exchanger assembly (heat exchanger with bubble trap) contains a hydrophobic membrane, a one-way valve, a pressure relief valve, a filter screen, a temperature monitoring port and a pressure monitoring port. The housing is designed such that air entering the device rises to the membrane and is vented to the atmosphere. The one-way valve keeps air from entering the device. The pressure relief valve is designed to relieve excess pressure in the event of outlet line clamping, and has tubing attached to it to vent the solution to the cardiotomy reservoir. The temperature monitoring port is used to measure the temperature of the blood cardionlegia mixture, and the pressure monitoring line is connected to either a pressure manometer or a transducer to measure pressure.
The provided text describes a Special 510(k) Pre-Market Notification for the SMAR,T® BCD Vanguard™ Blood Cardioplegia System, which is a modified version of an existing device. The acceptance criteria and the study that proves the device meets these criteria are not explicitly detailed in the provided document.
However, based on the information given, we can infer some aspects of the "study" conducted and the implicit acceptance criteria:
Implicit Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Inferred) | Reported Device Performance (Inferred) |
---|---|
Functional Equivalence: The modified device (SMAR,T® BCD Vanguard™) performs the same functions as the predicate device (BCD Vanguard™ Blood Cardioplegia System). | The SMAR,T® BCD Vanguard™ Blood Cardioplegia System has the "same intended use" as the predicate device. It is intended to mix, cool, warm and deliver oxygenated blood and cardioplegic solution for periods up to six hours, and also allows monitoring of temperature and pressure, traps bubbles, and removes air. This suggests its functional performance is deemed equivalent to the predicate. |
Material Biocompatibility/Safety Equivalence: The new surface-modifying materials and packaging materials do not introduce new safety concerns or adverse effects compared to the predicate. | "In-Vitro tests were performed to demonstrate that the SMAR-T BCD Vanguard Blood Cardioplegia System described in this submission is substantially equivalent to the BCD Vanguard Blood Cardioplegia System (K934847)." This implies the in-vitro tests covered safety and biocompatibility aspects relevant to the material changes. The FDA's substantial equivalence determination confirms this was met. |
Mechanical/Physical Integrity: The changes in materials and surface modification do not compromise the device's structural integrity or performance within its intended use. | "In-Vitro tests were performed to demonstrate that the SMAR-T BCD Vanguard Blood Cardioplegia System described in this submission is substantially equivalent to the BCD Vanguard Blood Cardioplegia System (K934847)." This would typically include tests to ensure the device's physical properties and mechanical performance remained acceptable after the modifications. |
"Study" Information (Inferred/Not Provided)
The document primarily focuses on demonstrating "substantial equivalence" to a predicate device (Sorin Biomedica BCD Vanguard™ Blood Cardioplegia System, K934847) rather than reporting a comprehensive clinical study. The only "study" mentioned is:
- Sample Size used for the test set and the data provenance: Not specified. The document only mentions "In-Vitro tests." These tests were likely laboratory-based, comparing the new device's performance characteristics (e.g., blood compatibility, heat exchange efficiency with new surfaces, pressure relief, bubble trapping) against the predicate device's established performance or accepted standards. The data provenance would be from laboratory testing.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable/Not specified. For in-vitro tests, the "ground truth" is typically defined by engineering specifications, regulatory standards, or predetermined performance benchmarks for the predicate device, rather than expert consensus on patient data.
- Adjudication method: Not applicable/Not specified. This is more relevant for studies involving human interpretation or clinical endpoints.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, not mentioned. This type of study is more common for diagnostic imaging AI systems or scenarios where human reader variability is a significant factor.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a medical device, not an algorithm.
- The type of ground truth used: For the "In-Vitro tests," the ground truth would be based on objective measurements against established engineering specifications, performance standards, and comparison with the predicate device's known performance characteristics. The goal was to prove the modified device was "substantially equivalent."
- The sample size for the training set: Not applicable. This is not an AI/machine learning algorithm, so there is no training set mentioned or implied.
- How the ground truth for the training set was established: Not applicable.
Summary of Device Changes and Justification for Substantial Equivalence:
The core of the submission (K021830) is a modification to an existing device (BCD Vanguard™ Blood Cardioplegia System, K934847). The changes are:
- Addition of surface-modifying material (Mimesys) to the heat exchanger.
- Substitution of SMAR,T Surface Modified Tubing and Connectors for various tubing and connectors.
- Substitution of High Impact Polystyrene (HIPS) for PolyEthylene Terephthalate Glycol (PETG) as the packaging material.
The document states that "In-Vitro tests were performed to demonstrate that the SMAR-T BCD Vanguard Blood Cardioplegia System described in this submission is substantially equivalent to the BCD Vanguard Blood Cardioplegia System (K934847)." This means the "study" was focused on proving that these modifications did not alter the fundamental safety or effectiveness of the device as compared to the previously cleared predicate device.
The FDA's letter states: "We have reviewed your Section 510(k) premarket notification... and have determined the device is substantially equivalent... You may, therefore, market the device..." This indicates that the "In-Vitro tests" successfully demonstrated that the new device met the implicit acceptance criteria for substantial equivalence to the predicate.
Ask a specific question about this device
Page 1 of 1