Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K210797
    Manufacturer
    Date Cleared
    2022-08-18

    (520 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K972278

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bio-MEM Ti Reinforced membranes are an implantable temporary non-resorbable device (membrane) for use as a spacer creation barrier in the treatment of local defects in the oral cavity in conjunction with tissue regeneration or augmentation. Bio-MEM Ti Reinforced membranes are intended to be submerged and clinically implanted more than 30 days and no longer than 6 months.

    Device Description

    Bio-MEM Ti Reinforced membranes are non-resorbable membranes available as Tireinforced configured of a highly shapeable Ti (Reinforced only) enveloped by an embossed polytetrafluoroethylene (PTFE) membrane. Extensive documentation demonstrates tissue compatibility of PTFE. Bio-MEM Ti Reinforced membranes are designed to avoid ingrowth of gingival soft tissue into bony defects, in order to facilitate the neovascularization and bone formation during the repair process of the defect. Bio-MEM Ti Reinforced membranes are provided STERILE and in different anatomical shapes and sizes. The titanium reinforcement is intended for space creating and shapemaintaining which minimize movements and subsequent exposure during the regenerative healing.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Bio-MEM Ti Reinforced Membrane.

    Note: The provided text is a 510(k) summary for a medical device (Bio-MEM Ti Reinforced Membrane). 510(k) submissions typically demonstrate substantial equivalence to a predicate device rather than presenting acceptance criteria and studies in the traditional scientific sense for a novel AI/software device. Therefore, the information below will reflect that context. The questions asked are more aligned with AI device studies, and the response will clarify what information is available given the document type.

    1. Table of Acceptance Criteria and Reported Device Performance

    Strict "acceptance criteria" and "reported device performance" as one might see for a diagnostic AI algorithm are not explicitly stated in this 510(k) summary. Instead, the document focuses on demonstrating that the Bio-MEM Ti Reinforced Membrane is substantially equivalent to its predicate devices based on design, materials, technological characteristics, and performance testing against recognized standards.

    The performance testing sections describe the studies undertaken to ensure the device's safety and functionality, which implicitly means the device met the requirements of these tests.

    Aspect TestedStudy TypeCriteria (Implicit)Performance (Reported)
    Strength & IntegrityBend and Delamination TestingDevice maintains structural integrity and functionality after bending and resists delamination.Confirmed the strength, integrity, and functionality of the device. (Specific quantitative results not provided, but implies pass)
    SterilizationSterilization Validation (EO Gas)Device is sterile and maintains sterility over shelf-life. Meets ISO 11135[2014] standards.Validation conducted to support EO sterilization. (Implies pass)
    Packaging Shelf-lifePackaging ValidationPackaging maintains sterility and integrity for the specified shelf-life. Meets ISO, ASTM standards.Validation conducted to support packaging shelf-life of 5 years. (Implies pass)
    BiocompatibilityCytotoxicity Testing (ISO 10993-5)No unacceptable cytotoxic effects.Cytotoxicity testing conducted. (Implies pass)
    BiocompatibilitySensitization Testing (ISO 10993-10)No unacceptable sensitization.Sensitization testing conducted. (Implies pass)
    BiocompatibilityIrritation Testing (ISO 10993-10)No unacceptable irritation.Irritation testing conducted. (Implies pass)
    BiocompatibilityAcute Systemic Toxicity Testing (ISO 10993-11)No unacceptable acute systemic toxicity.Acute systemic toxicity testing conducted. (Implies pass)
    BiocompatibilityMaterial-Mediated Pyrogenicity (USP 29 )No pyrogenic response.Material-mediated pyrogenicity testing conducted. (Implies pass)
    BiocompatibilityGenotoxicity Testing (AMES, Micronucleus - ISO 10993-3)Not genotoxic.Genotoxicity (AMES and micronucleus) testing conducted. (Implies pass)
    BiocompatibilityImplantation Testing (ISO 10993-6)Acceptable local effects after implantation.Implantation testing conducted. (Implies pass)
    Ethylene Oxide ResidualsTesting against ISO 10993-7Ethylene oxide residues below acceptable limits.Testing conducted against ISO 10993-7. (Implies pass)
    Material Properties (PTFE)Testing against ASTM D4895 - 10PTFE material meets standard specifications.Testing conducted against ASTM D4895 - 10. (Implies pass)

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes material and mechanical tests, and biocompatibility studies. These types of tests do not typically use "test sets" of patient data in the same way an AI algorithm would.

    • Sample Size for Test Set: Not applicable in the context of human patient data for algorithm evaluation. For the bench and in-vitro/in-vivo biocompatibility tests, specific sample sizes are not detailed in this summary document, but the fact that the tests were conducted according to ISO/ASTM/USP standards implies appropriate sample sizes were used for those specific test methods.
    • Data Provenance: Not applicable. These tests were likely performed in a laboratory setting, not with patient data.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. The "ground truth" for these types of device performance tests (e.g., sterility, material strength, biocompatibility) is established by adherence to validated international standards and the results of laboratory measurements, not by expert human interpretation of medical images or clinical data.

    4. Adjudication Method for the Test Set

    Not applicable. This concept belongs to the evaluation of algorithms where human assessment of data needs to be considered. For a physical device like this, performance is measured against established scientific and engineering standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is a clinical study design used to compare the performance of different diagnostic methods (e.g., human readers with and without AI assistance). This 510(k) summary is for a physical medical implant (a membrane) and focuses on demonstrating its safety and substantial equivalence, not its diagnostic accuracy or the improvement of human performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No. This device is not an algorithm, so "standalone performance" in that context is not applicable. Its performance is measured by its material properties, biological interactions, and mechanical integrity.

    7. The Type of Ground Truth Used

    The "ground truth" for the various performance tests is either:

    • Physical/Chemical Standards: (e.g., strength measurements in Bend and Delamination testing, density, thickness).
    • Biological Standards: (e.g., absence of cytotoxicity, irritation, sensitization, genotoxicity, pyrogenicity as defined by ISO 10993 and USP standards).
    • Microbiological Standards: (e.g., sterility per ISO 11135).

    Essentially, the ground truth is adherence to internationally recognized and validated scientific and engineering standards.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI algorithm, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no AI algorithm and no training set, this question is not relevant to the Bio-MEM Ti Reinforced Membrane.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171774
    Date Cleared
    2017-10-19

    (127 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K972278, K050466

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RPM™ Reinforced PTFE Mesh is a temporarily implantable material (non-resorbable) indication and support of bone grafts in alveolar bony defect sites.

    Device Description

    RPM™ Reinforced PTFE Mesh is placed between bone grafts and the periosteum in dental bone grafting procedures to stabilize and support the bone graft. The PTFE mesh helps create the space needed for bone-derived cells to repopulate and repair the defect. RPM™ Reinforced PTFE Mesh is composed of proprietary 100% polytetrafluoroethylene sheets reinforced with a titanium frame. The titanium frame is embedded between two layers of PTFE. PTFE is a biologically inert and tissue-compatible material. RPM™ Reinforced PTFE Mesh is manufactured with circular 0.66 mm diameter macropores to allow direct contact between the bone graft and the periosteum. Direct contact between the periosteum and bone graft allows naturally occurring revascularization and infiltration of cells. The PTFE mesh is designed to maintain space and conform to tissue contours. RPM™ Reinforced PTFE Mesh is provided pre-shaped in a variety of shapes and sizes.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the RPM™ Reinforced PTFE Mesh, a medical device for bone grafting. The document focuses on demonstrating substantial equivalence to predicate and reference devices through non-clinical testing.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance
    Tensile StrengthTensile strength ≥ reference device (Bio-Gide®)Tensile strength ≥ reference device (Bio-Gide®)
    Lamination StrengthLamination strength ≥ reference device (Cytoplast™ Regentex™ Titanium 250)Lamination strength ≥ reference device (Cytoplast™ Regentex™ Titanium 250)
    Suture Retention ForceSuture retention force ≥ reference device (Cytoplast™ Regentex™ Titanium 250)Suture retention force ≥ reference device (Cytoplast™ Regentex™ Titanium 250)
    CytotoxicityNon-cytotoxic (passed ISO 10993-5)Non-cytotoxic
    IrritationNon-irritant (passed ISO 10993-10)Non-irritant

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the numerical sample sizes used for each in vitro bench test. It mentions that "in vitro product characterization testing was performed" and "A series of bench tests were conducted."
    The data provenance is from non-clinical testing (in vitro characterization studies and biocompatibility studies). No information about country of origin for the data is provided, but it's generated as part of a US FDA submission. The studies are prospective in the sense that they were conducted specifically for this submission to evaluate the device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    Not applicable. The "ground truth" in this context is established by the design and execution of the scientific tests based on recognized standards (ASTM, ISO). There are no human "experts" establishing ground truth in the way one would for image interpretation in AI. The testing itself is the ground truth.

    4. Adjudication Method for the Test Set:

    Not applicable. This is not a study involving human interpretation or subjective assessment that would require adjudication. The test results are objective measurements against established standards and comparisons to reference devices.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This submission is for a physical medical device (reinforced PTFE mesh), not an AI/imaging diagnostic device. Therefore, no MRMC study was conducted or is relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an algorithm or AI device.

    7. The Type of Ground Truth Used:

    The ground truth used for these non-clinical tests is based on:

    • Performance against established industry standards: ASTM D638-14 for tensile strength, and ISO 10993-5 and ISO 10993-10 for biocompatibility.
    • Comparative performance against legally marketed predicate/reference devices: The specified predicate device (Osteo-Mesh™ TM-300) and reference devices (Cytoplast™ Regentex™ Titanium 250 and Bio-Gide® Resorbable Bilayer Membrane) serve as benchmarks for tensile strength, lamination strength, and suture retention force.

    8. The Sample Size for the Training Set:

    Not applicable. This is not an AI device or a study involving a "training set" in that sense.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable. Refer to point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091120
    Date Cleared
    2010-03-19

    (336 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K922489, K051267, K984230, K972278, K970841

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MEDPOR CONTAIN Implant is intended to stabilize, support and provide space maintenance for bone graft materials in the maxilla, mandible and zygoma.

    Device Description

    MEDPOR CONTAIN Implant is a thin, highly porous, biocompatible semi-rigid sheet, made from the same biocompatible pure porous high-density polyethylene (pHDPE) material which allows for host tissue integration as other MEDPOR Surgical Implants. The MEDPOR CONTAIN Implant is designed to stabilize and support bone graft materials and provide space maintenance necessary for regenerative healing.

    AI/ML Overview

    The provided document is a 510(k) summary for the MEDPOR® CONTAIN™ Implant. It describes the device, its intended use, technological characteristics, and compares it to predicate devices to establish substantial equivalence.

    Crucially, this document is a premarket notification for a medical device and does not contain a study or data proving the device meets acceptance criteria. Instead, it is a submission to the FDA demonstrating that the new device is substantially equivalent to legally marketed predicate devices, meaning it has the same intended use and similar technological characteristics, and raises no new questions of safety or effectiveness.

    Therefore, many of the requested sections, such as acceptance criteria, reported device performance, sample sizes, ground truth establishment, and details of comparative effectiveness studies (MRMC or standalone), cannot be extracted from this document, as they are not present.

    However, I can provide information based on what is available regarding the comparison to predicate devices, which is the basis for its clearance.


    1. A table of acceptance criteria and the reported device performance

    This document does not specify "acceptance criteria" in the typical sense of a performance study with numerical targets and results. Instead, it relies on demonstrating substantial equivalence to predicate devices based on intended use, material, design, sterilization, and biocompatibility.

    The "reported device performance" in this context is implicitly the assertion that the MEDPOR CONTAIN Implant performs similarly to its predicates due to shared characteristics.

    CategoryAcceptance Criteria (Implied by Substantial Equivalence to Predicates)Reported Device Performance (as described in the document)
    Intended UseStabilize, support, and provide space maintenance for bone graft materials in maxilla, mandible, and zygoma.The MEDPOR CONTAIN Implant is intended to stabilize, support and provide space maintenance for bone graft materials in the maxilla, mandible, and zygoma. (Matches predicate intended uses or sufficiently similar)
    MaterialBiocompatible, linear, high-density polyethylene.A linear, high-density polyethylene biomaterial. (Same as Predicate #1)
    Design/FormSheets of various sizes and thicknesses; trimmable/cuttable.15mm x 30mm, 25mm x 30mm, 45mm x 30mm sheets in 0.25mm and 0.35mm thicknesses and other shapes and sizes, as well as patient specific customized shapes. Can be trimmed or cut to size and bent. (Similar to predicate forms)
    SterilitySterileSterile
    Sterilization MethodAppropriate and effective sterilization method.EtO (Same as Predicate #1)
    BiocompatibilityBiocompatibleYes

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. This document is a 510(k) premarket notification based on substantial equivalence to existing predicate devices, not a clinical study report with a test set. There are no "test sets" or "data provenance" details for a performance study in this submission.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As no performance study with a test set is reported, there is no "ground truth" established by experts within this document. The assessment is regulatory by the FDA based on the submission.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No test set or performance data is presented, so no adjudication method is relevant here.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. A MRMC comparative effectiveness study is not discussed. This device is a passive implant, not an AI-powered diagnostic tool.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a medical implant, not an algorithm, so this type of study is not relevant.


    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    Not applicable. No ground truth data from a performance study is utilized in this 510(k) summary. The "ground truth" for the FDA's decision is the established safety and effectiveness of the predicate devices to which the new device is compared.


    8. The sample size for the training set

    Not applicable. This document describes a medical device, not a machine learning model, so there is no concept of a "training set" in this context.


    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for a machine learning model, there is no associated ground truth establishment method.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1