Search Results
Found 1 results
510(k) Data Aggregation
(150 days)
The SelectFlex 072 Neurovascular Access System is indicated for the introduction of interventional devices into the peripheral and neurovasculature.
The SelectFlex 072 Neurovascular Access System is a sterile, single-use intravascular catheter used to facilitate access target vasculature during interventional procedures. The system is composed of a the SelectFlex 072 Catheter, a 3cc Inflation Syringe, and a 7Fr Peel Away Introducer. The 072 SelectFlex Catheter has a usable length of 105cm. The SelectFlex 072 Catheter has variable stiffness along its length and has a dual mode stiffness mechanism on the distal portion of the catheter that is activated by the user allowing the device to transition between tracking and support modes. The distal end of the SelectFlex 072 Catheter has a hydrophilic coating.
This document, a 510(k) summary for the SelectFlex 072 Neurovascular Access System, describes various performance tests conducted to demonstrate the device's suitability for its intended use, rather than a clinical study evaluating an AI algorithm's performance.
Therefore, many of the requested fields regarding AI algorithm evaluation are not applicable (N/A) in this context. The document focuses on demonstrating substantial equivalence to a predicate device through physical and biological performance testing.
Here's the breakdown based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state numerical acceptance criteria for each test in a table, but rather indicates that all samples "met the pre-determined acceptance criteria" or "passed" the tests. The acceptance criteria are implicitly defined by the referenced ISO/ASTM standards and internal product specifications.
Test Description | Test Method | Reported Device Performance |
---|---|---|
Packaging Integrity (sterile barrier) | Tested per ISO 11607-1 and -2 | PASS All samples met the pre-determined acceptance criteria |
Visual Surface Requirements | Visual inspection of catheter surfaces | PASS All samples met the pre-determined acceptance criteria |
Dimensional Verification | Device dimensions were measured to confirm conformance to the product specification | PASS All samples met the pre-determined acceptance criteria |
Liquid Leakage Under Pressure | Tested per ISO 10555-1:2013 Annex C | PASS All samples met the pre-determined acceptance criteria |
Hub Aspiration Air Leakage | Tested per ISO 10555-1 2013 for Hub Aspiration Air Leakage | PASS All samples met the pre-determined acceptance criteria |
Simulated Use/Usability | Evaluated in a challenging neurovascular model, including usability assessment | PASS All samples met the pre-determined acceptance criteria |
Flex Fatigue | Tested per ISO 10555-1: 2013 for Flexural Fatigue | PASS All samples met the pre-determined acceptance criteria |
Tip Deflection | Evaluated and compared to the predicate device | PASS All samples met the pre-determined acceptance criteria |
Inflation Fatigue | Tested per ISO 10555-1: 2013 for Inflation Fatigue - 20 inflation cycles | PASS All samples met the pre-determined acceptance criteria |
Burst Volume | Tested per ISO 10555-1: 2013 for Inflation Fatigue - tested to 2x inflation volume | PASS All samples met the pre-determined acceptance criteria |
Torque Test | Tested per ISO 10555-1: 2013 for Torque Testing | PASS All samples met the pre-determined acceptance criteria |
Flow Rate | Tested per ISO 10555-1: 2013 for Flow rate compared to the predicate device | PASS All samples met the pre-determined acceptance criteria |
ISO 594-2: Conical Luer Fittings | Tested per ISO 594-2 for Conical Fittings | PASS All samples met the pre-determined acceptance criteria |
Corrosion Resistance | Tested per ISO 10555-1 Annex A for corrosion resistance | PASS All samples met the pre-determined acceptance criteria |
Particulate Count | Effluent tested per AAMI TIR42, USP 788 using multiple insertion and withdrawal cycles | PASS All samples met the pre-determined acceptance criteria |
Coating Integrity, Lubricity, Durability | Tested in consideration of FDA CTQ for Hydrophilic Coated vascular catheters | PASS All samples met the pre-determined acceptance criteria |
Peak Tensile Testing | Tested per ISO 10555-1 for tensile strength including all bonds/joints | PASS All samples met the pre-determined acceptance criteria |
Device Removal in Support and Tracking Modes | Removal force in both modes compared to a reference device | PASS All samples met the pre-determined acceptance criteria |
Radiopacity | Evaluated during simulated use testing confirming visualization under fluoroscopy | PASS All samples met the pre-determined acceptance criteria |
Bacterial Endotoxin | Tested in accordance with USP 40, NF 35, 2017. Bacterial Endotoxins Test | PASS All samples met the pre-determined acceptance criteria |
Shelf Life | Accelerated aging studies | PASS All samples met the pre-determined acceptance criteria |
Biocompatibility Testing | ||
Cytotoxicity | Tested in accordance with ISO 10993-5 (Neutral Red Uptake Method) | Pass Noncytotoxic according to predetermined acceptance criteria |
Intracutaneous Irritation | Tested in accordance with ISO 10993-10 | Pass Test requirements for intracutaneous reactivity were met |
Sensitization | Tested in accordance with ISO 10993-10 (Kligman Maximization Test) | Pass Did not elicit a sensitization response |
Systemic Toxicity | Tested in accordance with ISO 10993-11 | Pass Test requirements for systemic toxicity were met |
Material Mediated Pyrogenicity | Tested in accordance with ISO 10993-11 and USP 40 Pyrogen Test | Pass Nonpyrogenic, met the predetermined acceptance criteria |
Hemolysis | Tested in accordance with ASTM F756-17 and ISO 10993-4 | Pass Non-hemolytic, met the predetermined acceptance criteria |
In Vitro Hemocompatibility | Tested in accordance with ISO 10993-4 (Direct Contact Method) | Pass Not expected to result in adverse effects in vivo |
Complement Activation | Tested in accordance with ISO 10993-4 (SC5b-9 Complement Activation) | Pass Does not activate the complement system |
Un-activated Partial Thromboplastin Time | Tested in accordance with ISO 10994-4 and ASTM F2382-04 | Pass Does not affect coagulation of human plasma |
Thrombogenicity | Tested in accordance with ISO 10994-4 | Pass Demonstrates similar thromboresistance characteristics as the control device |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify the exact sample sizes used for each individual performance test (e.g., how many catheters were tested for flex fatigue). It generally states "All samples met the pre-determined acceptance criteria," implying that a sufficient number of samples were tested per the relevant standards.
- Sample Size for Test Set: Not explicitly stated for each test, but implied to be adequate for standard compliance.
- Data Provenance: Not specified. This typically refers to data like patient images or clinical records, which are not relevant to these engineering and biocompatibility tests. The tests themselves are laboratory-based.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Number of Experts: For the "Simulated Use/Usability" test, it states "a usability assessment with multiple physicians." The exact number or qualifications are not provided beyond "physicians."
- Qualifications of Experts: "Multiple physicians" for usability assessment. For other tests, "ground truth" is established by adherence to recognized international standards (e.g., ISO, ASTM, USP) and internal product specifications, carried out by qualified testing personnel, but not "experts" in the sense of clinical specialists interpreting results.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is not a clinical study involving interpretation of data by multiple readers. The assessment of performance tests is based on objective measurements against pre-defined criteria in laboratory settings.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document pertains to the performance and safety testing of a neurovascular access system, not an AI or imaging diagnostic device. No AI component is mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI algorithm.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
For the performance and biocompatibility tests, the "ground truth" is established by:
- Compliance with international standards: ISO, ASTM, AAMI, USP.
- Adherence to internal product specifications: Dimensional verification.
- Comparison to predicate/reference devices: For tip deflection, flow rate, and device removal forces.
- Usability assessment: By "multiple physicians" in a simulated environment.
No pathology or outcomes data are referenced as this is a pre-market notification for a device and not a clinical trial.
8. The sample size for the training set
Not applicable. This is not an AI algorithm, so there is no training set.
9. How the ground truth for the training set was established
Not applicable. There is no AI training set.
Ask a specific question about this device
Page 1 of 1