Search Results
Found 2 results
510(k) Data Aggregation
(189 days)
The ADVIA Centaur® Cortisol (COR) Assay is for in vitro diagnostic use in the quantitative determination of cortisol in serum or urine using the ADVIA Centaur XP system. Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland.
The ADVIA Centaur Cortisol assay is a competitive immunoassay using direct chemiluminescent technology. Results are determined using a calibration curve that is generated specifically on each instrument by a 2-point calibration and a master curve with the reagent bar code. The ADVIA Centaur Cortisol Assay is intended for use on the ADVIA Centaur Family of analyzers. The ADVIA Centaur Calibrator E is a set of 2 level calibrators for the assay. Siemens recommends the use of commercially available quality control materials with at least two levels (low and high).
Here's a breakdown of the acceptance criteria and the study details for the ADVIA Centaur® Cortisol (COR) Assay, based on the provided FDA 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Criteria/Test | Acceptance Criteria (Implicit from satisfactory results) | Reported Device Performance (Summary) | Comment |
---|---|---|---|---|
Precision | CLSI EP05-A2 protocol (20-day study, 2 reps/run, 2 runs/day) | Within-Lab %CV should be acceptable | Serum: 4.4% - 6.0% (Controls), 4.9% - 5.3% (Samples) | |
Direct Urine: 6.8% - 9.1% | ||||
Extracted Urine: 6.8% - 9.2% | Performance appears to be within generally accepted clinical chemistry precision standards, demonstrating good reproducibility. | |||
Linearity/Assay Range | Linearity across the measuring range (EP06-A) | % Recovery should be acceptable, high r² (correlation) | Serum: Y=1.057x - 0.051, r²=0.9991, % Recovery 96.0-109.3% | |
Direct Urine: Y=1.011x + 0.090, r²=0.9975, % Recovery 94.7-119.6% | ||||
Extracted Urine: Y=0.914x + 0.017, r²=0.9997, % Recovery 82.7-100.9% | Strong linearity demonstrated with high r² values close to 1, indicating a good proportional relationship between expected and observed values over the claimed analytical range. The urine recoveries showed a slightly wider range but were still deemed acceptable. | |||
Analytical Detection Limits | LoB, LoD, LoQ (EP17-A2) | Values should be below claimed measuring range | LoB: Serum 0.06 µg/dL, Direct Urine 0.19 µg/dL, Extracted Urine 0.18 µg/dL | |
LoD: Serum 0.14 µg/dL, Direct Urine 0.45 µg/dL, Extracted Urine 0.44 µg/dL | ||||
LoQ: Serum 0.31 µg/dL, Direct Urine 0.48 µg/dL, Extracted Urine 0.44 µg/dL | The determined detection limits are well below the lower end of the claimed measuring range (0.50 µg/dL), supporting the device's ability to accurately measure low concentrations. | |||
Analytical Specificity (Interference) | Endogenous substances (EP07-A2) | % Interference ≤ 10% | All tested endogenous substances (Hemoglobin, Triglycerides, Bilirubin, Protein, NaCl, Urea, Creatinine, Glucose, Boric Acid) showed % Interference within -5% to 5%. | The device demonstrates good resistance to interference from common endogenous substances in both serum and urine, ensuring reliable results in various patient samples. |
Analytical Specificity (Cross-reactivity) | Potential cross-reactant compounds | Low cross-reactivity with structurally similar compounds | Most compounds showed low cross-reactivity ( |
Ask a specific question about this device
(145 days)
Immunological in vitro assay for the quantitative determination of cortisol in human serum, plasma and urine. The determination of cortisol is used for the recognition and treatment of functional disorders of the adrenal gland.
The electrochemiluminescence immunoassay "ECLIA" is intended for use on the Roche Elecsys® 1010 / 2010 and Modular Analytics E170 (Elecsys module) Immunoassay Analyzers.
The ELECSYS® Cortisol Assay a two step sandwich immunoassay with streptavidin microparticles and electrochemiluminescence detection. Results are determined using a calibration curve that is generated specifically on each instrument by a 2-point calibration and a master curve provided with the reagent bar code.
The information provided describes the Elecsys® Cortisol Assay and its substantial equivalence to a predicate device (Bayer Diagnostics ACS:180 Cortisol Assay). It focuses on comparing the new device's performance characteristics to the predicate device to demonstrate substantial equivalence, rather than detailing a specific study proving the device meets acceptance criteria with predefined thresholds. Therefore, some of the requested information cannot be fully extracted as it pertains to a typical clinical validation study with explicitly stated acceptance criteria and a detailed study design.
Here's the breakdown of what can be extracted and what is not available in the provided text:
1. Table of acceptance criteria and the reported device performance:
The document doesn't explicitly state "acceptance criteria" through specific numerical thresholds for each performance characteristic. Instead, it demonstrates performance by comparing the new device's data directly to that of the predicate device, implying that performance comparable to the legally marketed predicate device is the de facto acceptance criteria for substantial equivalence.
Feature | Acceptance Criteria (Implied: Comparable to Predicate) | New Device (ELECSYS Cortisol) Reported Performance | Predicate Device (Bayer ACS:180 Cortisol) Reported Performance |
---|---|---|---|
Intra-assay precision (%CV) (Urine) | Comparable to Predicate's Intra-assay precision | • 2.2% at 22.3 µg/dl | |
• 2.3% at 33.2 µg/dl | |||
• 2.9% at 41.9 µg/dl | |||
• 2.3% at 61.0 µg/dl | • 5.7% at 3.04 µg/dl | ||
• 5.1% at 5.43 µg/dl | |||
• 4.5% at 14.90 µg/dl | |||
• 6.4% at 18.98 µg/dl | |||
• 7.0% at 31.79 µg/dl | |||
• 7.5% at 38.67 µg/dl | |||
Interassay precision (%CV) (Urine) | N/A (Predicate data not fully provided) | • 2.5% at 23.2 µg/dl | |
• 3.2% at 33.4 µg/dl | |||
• 2.5% at 42.1 µg/dl | |||
• 1.8% at 58.9 µg/dl | |||
Control: 4.7% at 2.82 µg/dl | N/A | ||
Total precision (%CV) | Comparable to Predicate's Total precision | N/A | • 9.1% at 3.04 µg/dl |
• 8.0% at 5.43 µg/dl | |||
• 6.4% at 14.90 µg/dl | |||
• 8.2% at 18.98 µg/dl | |||
• 9.2% at 31.79 µg/dl | |||
• 9.7% at 38.67 µg/dl | |||
Functional sensitivity | Comparable or better than Predicate's functional sensitivity |
Ask a specific question about this device
Page 1 of 1