Search Results
Found 1 results
510(k) Data Aggregation
(31 days)
The iScience Surgical Ophthalmic Viscoinjector is a manually operated device for precision delivery of small amounts of viscous fluid such as a sterile viscoelastic solution pre-packaged in a cartridge, for example Healon, HealonGV and Healon5 manufactured by Advanced Medical Optics Inc., and Ocucoat manufactured by Bausch and Lomb.
The iScience Surgical Ophthalmic Viscolnjector consists of a holder for a viscoelastic The loolorios bargreates a manually operated screw driven plunger. For use, a cantridge of viscoelastic solution is placed into the body of the Ophthalmic Viscolnjector. A needle inside the Ophthalmic Viscolnjector pierces the septum of the viscoelastic cartridge as it is seated within the body cavity. The cap from the Ophthalmic Cannage as it is connected to the body to secure the cartridge of viscoelastic solution inside the Viscolnjector. The thumbwheel is rotated clockwise thereby advancing the screw driven plunger to deliver viscoelastic solution. The end of the screw has a bushing that contacts the stopper in the viscoelastic cartridge. As the screw is advanced at a controlled rate the stopper displaces fluid out through the needle that has pierced the septum. The opposite, blunt end of the needle is embedded in a male Luer lock connector. This connector may be mated to a female Luer connector and the viscoelastic solution delivered. Markings on the thumbwheel, cap, and body of the Viscolnjector are provided to assist in determining the number of complete or partial thumbwheel rotations and thus relate to the amount of fluid delivered.
The provided text describes the iScience Surgical Ophthalmic Viscoinjector, a manually operated device for precision delivery of small amounts of viscous fluid. The company submitted a 510(k) premarket notification to the FDA, and the device was determined to be substantially equivalent to previously marketed devices.
Here's an analysis of the acceptance criteria and study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The text does not explicitly state numerical acceptance criteria with pass/fail values. Instead, it refers to general guidance and the intrinsic strength of materials. The performance section indicates that the device has been shown to conform to certain standards and practices.
Acceptance Criteria Category | Specific Standard/Guidance/Basis | Reported Device Performance |
---|---|---|
Sterilization | ANSI/AAMI/ISO 11137-1994 (Radiation Sterilization) | Conforms to standard. |
Shelf-Life & Packaging Integrity | ANSI/AAMI/ISO 11607-1997 & AAMI TIR 1997 | Conforms to standards. |
Performance | ISO 7886-1 "Sterile hypodermic syringes for single use- Part 1: Syringes for manual use" (for general guidance); General experience with syringe-type devices; Intrinsic strength of Viscoinjector materials; Expected load during intended use. | Evaluated for: |
- Strength of device joints
- Static pressure failure
- Delivery of fluid (e.g., viscoelastic solution) |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample size used for performance testing (e.g., number of devices tested for joint strength, static pressure, or fluid delivery).
The data provenance is not explicitly mentioned as retrospective or prospective, nor is the country of origin of the data provided.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable and not provided. The study focuses on the mechanical and functional performance of a medical device, not on diagnostic accuracy requiring expert interpretation of results. The "ground truth" in this context would be objective measurements of physical properties and fluid delivery.
4. Adjudication Method for the Test Set
This is not applicable and not provided. Performance testing involved objective measurements against established engineering principles and standards, not subjective assessments requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
This is not applicable. The device is a manually operated ophthalmic viscoinjector, a physical instrument for fluid delivery, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study comparing human reader performance with and without AI assistance was not conducted.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This is not applicable. The device is a mechanical instrument, not an algorithm, and its operation inherently involves human interaction (manual operation).
7. The Type of Ground Truth Used
The "ground truth" for the performance testing appears to be based on:
- Objective Measurements: Such as measurements of joint strength, static pressure at failure, and the volume or precision of fluid delivery.
- Established Industry Standards and Guidance: ISO 7886-1-1994, ANSI/AAMI/ISO 11137-1994, ANSI/AAMI/ISO 11607-1997, and AAMI TIR.
- Engineering Principles: Related to the intrinsic strength of materials and the expected loads during use.
8. The Sample Size for the Training Set
This is not applicable. The device is a mechanical medical instrument, not a machine learning model, so there is no concept of a "training set" in this context.
9. How the Ground Truth for the Training Set Was Established
This is not applicable for the reasons stated in point 8.
Ask a specific question about this device
Page 1 of 1