Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K973475
    Manufacturer
    Date Cleared
    1997-11-20

    (69 days)

    Product Code
    Regulation Number
    870.4240
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K891687, K941473

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medtronic CardioTherm™ Blood Cardioplegia System, with or without Carmeda® BioActive Surface, is intended for use in procedures requiring the mixing, cooling, warming, and delivery of oxygenated blood and/or asanguineous cardioplegia solution.

    Device Description

    The Medtronic CardioTherm™ Blood Cardioplegia System with Carmeda® BioActive Surface is coated with a non-leaching bioactive heparin surface which provides thromboresistant blood contact surfaces. The Carmeda® BioActive Surface is present on blood path materials.

    Each Medtronic CardioTherm™ Blood Cardioplegia System with Carmeda® BioActive Surface is a single use, disposable device designed to mix arterial blood from an oxygenator with asanguineous cardioplegia solution in specific ratios depending on the tubing set configurations. The blood/cardioplegia solution is then cooled/warmed and delivered to the patient.

    The Medtronic CardioTherm™ Blood Cardioplegia Heat Exchanger with Carmeda BioActive Surface consists of a polycarbonate housing which incorporates a bubble chamber. A stopcock is attached to the upper luer port of the housing lid allowing for air venting through a purge line during priming and pressure monitoring through a vented or non-vented pressure monitoring line throughout the procedure. A temperature probe port is located adjacent to the blood outlet port at the bottom of the device, allowing for temperature monitoring of the blood/cardioplegia solution prior to patient delivery.

    The heat exchanger housing contains non-porous polypropylene hollow fibers, which are supported on their exterior by a polyethylene screen. These polypropylene fibers run longitudinally through the polycarbonate chamber. Blood/cardioplegia solution flows through the polypropylene hollow fibers. The walls of these fibers provide a barrier between the blood/cardioplegia solution and the cooling/warming water.

    The flow of solution through the Medtronic CardioTherm™ Blood Cardioplegia Heat Exchanger with Carmeda® BioActive Surface is as follows:

    Blood/Cardioplegia

    The blood and/or cardioplegia solution enters the bottom of the device, flows up through the polypropylene hollow fibers, flows back down a center collecting tube, and exits the from the bottom of the device.

    Water

    The cooling/warming water enters a side port, passes around the outside walls of the polypropylene fibers, then exits through a parallel side port.

    Heat exchange occurs as the blood/cardioplegia solution passes through the hollow fibers, while the temperature regulated cooling/warming water passes around the outside walls of the same fibers. The blood/cardioplegia solution is cooled/ warmed as it flows through the heat exchanger.

    By connecting tubing of specific inner diameters to the heat exchanger, a predetermined approximate ratio of blood to cardioplegia solution may be delivered to the patient. The tubing sets are equivalent to the tubing sets which are currently used for the CardioTherm™ Blood Cardioplegia System (K960755)

    Four of the various delivery ratios available are:

    • a tubing set for 1:1 ratio of blood to asanguineous cardioplegia. 1.
    • a tubing set for 2:1 ratio of blood to asanguineous cardioplegia. 2.
    • a tubing set for 4:1 ratio of blood to asanguineous cardioplegia.
    • a tubing set for 9:1 ratio of blood to asanguineous cardioplegia. 4.

    These tubing ratios are identical to the commercially available CardioTherm™ Blood Cardioplegia Systems.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Medtronic CardioTherm™ Blood Cardioplegia System with Carmeda® BioActive Surface. This submission aims to demonstrate substantial equivalence to legally marketed predicate devices, not to prove that the device meets specific acceptance criteria through a clinical study with real-world patient data, as would typically be the case for novel AI/ML medical devices.

    Instead, the submission focuses on comparing the modified device (with the Carmeda® coating) to existing predicate devices already approved for market, primarily through in-vitro bench testing and biocompatibility testing. The "acceptance criteria" here are implicitly related to demonstrating that the new technological characteristic (the Carmeda® coating) does not negatively affect safety or effectiveness and that the modified device remains substantially equivalent to its predicates.

    Therefore, the requested information elements related to clinical studies, human readers, ground truth establishment, and training/test set sample sizes (common for AI/ML device evaluations) are not directly applicable or available in the provided document, as this is a modification to an existing device rather than a de novo submission for a new AI/ML diagnostic or therapeutic device.

    Here's an interpretation based on the provided text's context:

    Acceptance Criteria and Study Details for Medtronic CardioTherm™ Blood Cardioplegia System with Carmeda® BioActive Surface

    The "acceptance criteria" in this context are not explicit numerical thresholds from a clinical trial, but rather the demonstration of substantial equivalence to predicate devices through various tests. The device is considered to meet these criteria if the performance data from in-vitro bench testing and biocompatibility testing show that the addition of the Carmeda® BioActive Surface does not significantly affect safety and effectiveness compared to the predicate devices.

    1. Table of "Acceptance Criteria" and Reported Device Performance

    Given that this is a 510(k) for a device modification, the "acceptance criteria" are implied to be comparable performance to the predicate device across key characteristics, and absence of new safety/effectiveness questions. The "reported device performance" refers to the results of the in-vitro bench testing and biocompatibility studies conducted on the CardioTherm™ Blood Cardioplegia System with Carmeda® BioActive Surface.

    Characteristic"Acceptance Criterion" (Implied)Reported Device Performance (with Carmeda® BioActive Surface)
    Carmeda® PerformanceNo significant impact on safety/effectiveness of the device due to the coating, and
    demonstration of coating functionality as per established methods for other Carmeda®-coated products.Bioactivity: Assessed (details not provided, but implicitly met as no new safety/effectiveness questions raised).
    Coating Coverage: Assessed (details not provided, implicitly met).
    Leaching: Assessed (details not provided, implicitly met to confirm non-leaching property).
    Device PerformancePerformance characteristics (pressure drop, blood trauma, heat exchange) comparable to predicate device (K960755).Pressure Drop: Assessed (details not provided, implicitly comparable).
    Blood Trauma: Assessed (details not provided, implicitly comparable).
    Heat Exchange Performance: Assessed (details not provided, implicitly comparable).
    Physical CharacteristicsPhysical integrity (blood pathway, heat exchanger water pathway) and priming volume comparable to predicate device.Physical Integrity: Assessed (details not provided, implicitly comparable).
    Priming Volume: Assessed (details not provided, implicitly comparable).
    BiocompatibilityBiocompatibility profile consistent with existing Carmeda®-coated products and no new biocompatibility concerns.Biocompatibility testing demonstrated equivalence to predicate devices and no significant effect on safety/effectiveness.

    2. Sample Size for Test Set and Data Provenance

    • Sample Size for Test Set: Not applicable in the context of clinical patient data. The "test set" here refers to the in-vitro bench tests and biocompatibility tests conducted on the modified device and compared to the predicate. The specific number of devices or test runs for these in-vitro tests is not provided in the summary.
    • Data Provenance: The data primarily comes from in-vitro bench testing and biocompatibility testing performed by Medtronic, Inc. It is not patient data from a specific country or retrospective/prospective study in the clinical sense.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not applicable. For this type of device modification submission, "ground truth" is established through engineering and scientific test methods (e.g., measuring pressure, analyzing coating, biological compatibility assessments), not through expert clinical consensus on patient cases.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. This concept applies to expert review of ambiguous clinical cases, which is not relevant for this engineering and biocompatibility assessment.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study: No. This type of study assesses the impact of an AI algorithm on human reader performance, which is not relevant for a blood cardioplegia system with a bioactive surface coating.

    6. Standalone Performance Study (Algorithm Only)

    • Standalone Study: No. This device is not an algorithm, and the concept of standalone performance does not apply to a physical medical device in this context. The closest analogy would be the in-vitro bench testing evaluating the device's physical and functional characteristics.

    7. Type of Ground Truth Used

    • Type of Ground Truth: The "ground truth" for this submission is based on established scientific and engineering measurement standards for device performance (e.g., pressure drop, heat exchange efficiency) and biocompatibility assessment protocols. It is not pathology, expert consensus on patient outcomes, or clinical outcomes data. The acceptance is based on demonstrating that the new coating does not alter these "ground truths" of established performance and safety compared to the predicate.

    8. Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. This device is not an AI/ML algorithm that undergoes a training phase.

    9. How Ground Truth for the Training Set Was Established

    • How Ground Truth for Training Set Was Established: Not applicable, as there is no training set for an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1